Validation of commonly used rectangular nitinol archwire dimensions using stereomicroscope an invitro study.

INTRODUCTION

Orthodontic archwires are essential components of modern fixed appliances. The brackets used nowadays are of Pre-adjusted Edgewise appliance. The placing of maximum prescription archwires in a pre-adjusted bracket is designed to produce three-dimensional tooth moving forces. These forces are created as a result of intimate fit of the wire into the bracket slot and any 'play' or 'slop' between these components will result in incomplete transmission of the bracket prescription to the tooth and its supporting tissues. Use of undersized archwires may directly and adversely affect the three dimensional tooth positioning. (1)

Achieving a satisfactory inclination or torque of the incisors is important for the final esthetic result. Torque expression depends upon a number of factors including the size of the bracket slots and archwires. Full torque expression should potentially be achieved by using an archwire of the appropriate size to fill the bracket slot; however, to insert a full size rectangular archwire it is necessary that there is a certain amount of 'play'. This means that the vertical dimension or height of the bracket slot must be greater than the height of the archwire. If there is a larger discrepancy between the bracket slot and the archwire dimension, then there will be inconsistencies in the amount of torque expressed. (2)

Siatkowski illustrated that there is loss of anterior torque with use of undersize archwires. (3) Kusy and Whitley stated that variations in archwire dimensions may account for failure of protraction mechanics, as well as for the loss of incisor axial inclination control during other forms of tooth movement. (4) Hence it is of utmost importance that the wires be precisely manufactured to have correct prescription.

The purpose of the present study carried out in the Department of Orthodontics and Dentofacial Orthopaedics at Kamineni Institute of Dental Sciences, Nalgonda is to validate the dimensions of rectangular nitinol archwires and, if any, between reported and actual dimensions

Materials and Methods:

Materials:

The materials used for the study:

* 10 nitinol archwires of each 0.017[conjunction] x 0.025[conjunction], 0.019[conjunction] x 0.025[conjunction] dimensions from three different companies (Modern Orthodontics, Prime orthodontics and Captain Orthodontics) weretaken for the study.

* Stereomicroscope (Magnus MSZ-Bi) was used for measurement of archwire dimensions (Figure 1).

Methodology:

The selected wires were viewed under a stereomicroscope at 4x magnification. Each selected wire was scanned and captured individually by using stereomicroscope to produce a digital image. The images were exported and calibrated using Image Pro insight Software and height and width of wires were measured (Figure 2).

The values obtained were compared to the dimensions published by each manufacturer. Comparisons were also made between wires of different manufacturers.

RESULTS:

The values obtained were subjected to statistical analysis by using SPSS version 22.0 software. The values obtained were subjected to multiple comparisons (Post Hoc Tests) of height and width of archwires and have been tabulated (Tables 1-5).

DISCUSSION

Standardization is an essential tool requirement for technological progress. Edward H Angle introduced Edgewise bracket of 0.022 inch as the standard slot size for brackets. The 0.022 inch system offers more options in archwire size selection. With the use of undersized archwire, one can facilitate the free sliding of archwire through the bracket slot. Being able to use larger diameter archwires for treatment provides increase stiffness and allows to keep the teeth upright during space closure with different retraction mechanics. Use of undersized archwires may directly and adversely affect the three dimensional tooth positioning.

The play between the archwire and bracket slot is of fundamental importance in clinical orthodontics, as it indicates how many degrees the archwire must be rotated within the bracket before its edges come into contact with the slot walls, enabling it thus to transmit third-order information to the tooth. The degree of play depends entirely on geometric parameters, namely the real slot height, the dimensions of the archwire and the beveling of its edge. However, on the products on the market, these do not always conform to the measurements declared by the manufacturer. (2)

Siatkowski noted that maxillary and mandibular incisors may suffer unexpected loss of torque while using smaller-dimension rectangular wires or when inadvertently using undersized maximum-prescription archwires or wires with an excessive edge bevel. (3)

Helge Fischer-Brandies et al conducted a study to investigate the influence of cross section, edge geometry and structural hardness on torque transmission between square wire and bracket. They found that as a result of excessively small wire dimensions and plastic deformation of the brackets, a relatively large torque play occurs. (5)

Iosif Sifakakis et al compared the archwires inserted during the final stages of the orthodontic treatment with the generated moments at 0.018and 0.022-inch brackets. The same prescription bracket type, was evaluated in both 0.017 x 0.025 and

0.019 x 0.025 dimensions with TMA and stainless steel wires. This difference is exaggerated in steel archwires, in comparison with the TMA, because of differences in stiffness. The differences of maximum moments between the archwires of the same cross-section but different alloys were statistically significant at both slot dimensions. (6)

Luca Lombardo et al evaluated the degree to which the height, width, and cross-section of rectangular and square orthodontic archwires affect the play between the archwires and the bracket slot. The curvature (radius) of the edge bevels was also measured to calculate the play within the slot, and this measurement was compared with the ideal value. They found that the real height and width of the archwires differed from those stated by the manufacturers. (7)

The present study was done to compare the archwire dimensions between three different companies (Modern Orthodontics, Prime orthodontics and Captain Orthodontics and compare the dimensions with the standard value published by the manufacturer.

When 0.017 x 0.025 NiTi wires were compared between three companies the results showed a statistical significant difference in terms of height of the archwire (p=0.000) indicating that the height of the archwires were different among all three companies (Table 1 and 2).

In terms of width of the archwire the results were insignificant indicating that all three brands are similar among themselves in terms of width of the archwire (Table 1 and 3).

When 0.019 x 0.025 NiTi wires were compared between three companies the results showed a statistical significant difference in terms of height of the archwire (p=0.000) indicating that the height of the archwires were different among all three companies (Table 4 and 5).

In terms of width of the archwire the results were insignificant indicating that all three brands are similar among themselves in terms of width of the archwire (Table 3 and 4).

When 0.017 x 0.025 NiTi wires were compared to its standard value (0.431mm, 0.0635mm), all are undersized but Modern Orthodontics company wires showed dimensions close to that of standard value prescribed by manufacturer.

When 0.019 x 0.025 NiTi wires were compared to its standard value (0.482mm, 0.0635mm), all are undersized but Modern Orthodontics company wires showed dimensions close to that of standard value prescribed by manufacturer.

Conclusion

The results of this study pointed to the following conclusions:

1. The archwire dimensions were different than stated by the manufacturers.

2. The dimensions of Modern Orthodontics were close to the standard values.

3. The dimensions of Prime Orthodontics and Captain Orthodontics archwires were less than the standard values.

doi: 10.5866/2017.9.10077

REFERENCES

(1.) Karan Tangri, Piush kumar, Payal sharma, Kiran kumar. A Comparison of the Accuracy of 0.022 Slots at Face, Base and Mesial and Distal Surface of Brackets marketed by Different Manufacturers. J Ind Orthod Soc 2012; 46(3):132-6.

(2.) Joch A, Pichelmayer M, Weiland F. Bracket slot and archwire dimensions: manufacturing precision and third order clearance. J Orthod 2010; 37:241-9.

(3.) Raymond. E. Siatkowski. Loss of anterior torque control due to variations in bracket slot and archwire dimensions J Clin Orthod 1999; 33(9):508-10.

(4.) Kusy RP, Whitley JQ. Assessment of second-order clearances between orthodontic archwires and bracket slot via the critical contact angle for binding Angle Orthod 1999; 69(1):71-80.

(5.) Fischer-Brandies, Es-Souni M, Meyer. Torque transmission between square wire and bracket as a function of measurement, form and hardness parameters. J Orofac Orthop 2000; 61(4):258-64.

(6.) Sifakakis I, Pandis N, Makou M, Eliades T, Katsaros C, Bourauel C. Torque efficiency of different archwires in 0.018 and 0.022 inch conventional brackets. Angle Orthod 2014; 84(1):149-54.

(7.) Lombardo L, Arreghini A, Bratti E, Mollica F, Spedicato G, Merlin M et al. Comparative analysis of real and ideal wireslot play in square and rectangular archwires. Angle Orthod 2015; 85(5):848-58.

Vasu Murthy S [1], Pavan Kumar M [2], Praveen Kumar N [3], Shiva Prasad M [4], Sreekanth K [5], Samatha Reddy V ([6])

[1] Professor & HOD

[2] Professor

[3]) Reader

[4,5] Senior Lecturer

[6] Post Graduate Student Department of Orthodontics Kamineni Institute of Dental Sciences Sreepuram, Narketpally, Nalgonda (Dt)

Article Info:

Received: April 8, 2017

Review Completed: May 10, 2017

Accepted: June 11, 2017

Available Online: June, 2017 (www.nacd.in)

Email for correspondence: vasuortho@gmail.com

Caption: Figure 1: Archwire seen under Stereomicroscope and subjected for analysis.

Caption: Figure 2: Photographs of archwire height and width dimensions (A. Height and B. Width)
Table 1: Depicts mean, standard deviation and level of significance
(p-value) of 0.017 x 0.025 Nitinol wires between three different
companies.

         COMPANY                MEAN    SD       P-value

Height   Modern Orthodontics    0.386   0.006    0.000
         Prime Orthodontics     0.371   0.005
         Captain Orthodontics   0.378   0.006
Width    Modern Orthodontics    0.557   0.004    0.05
         Prime Orthodontics     0.544   0.0066
         Captain Orthodontics   0.553   0.004

Table 2: Multiple comparison (Post Hoc Tests)--Height

COMPANY                         P-Value

Modern Ortho    Prime Ortho     0.000
                Captain Ortho   0.027
Prime Ortho     Modern Ortho    0.00284
                Captain Ortho   0.0028
Captain Ortho   Modern Ortho    0.008
Prime Ortho                     0.007

Table 3: Multiple comparison (Post Hoc Tests)--Width

COMPANY                         P-Value

Modern Ortho    Prime Ortho     0.000
                Captain Ortho   0.374
Prime Ortho     Modern Ortho    0.000
                Captain Ortho   0.000
Captain Ortho   Modern Ortho    0.374
Prime Ortho                     0.000

Table 4: Depicts mean, SD and p-value of 0.019 x 0.025 Nitinol wires
between three different companies.

         COMPANY         MEAN    SD        P-value

Height   Modern Ortho    0.426   0.0078    0.001
         Prime Ortho     0.423   0.008
         Captain Ortho   0.417   0.006
Width    Modern Ortho    0.55    0.0086    0.527
         Prime Ortho     0.545   0.00666
         Captain Ortho   0.548   0.0066

Table 5: Multiple comparison (Post Hoc Tests)--Height

COMPANY                         P-Value

Modern Ortho    Prime Ortho     0.558
                Captain Ortho   0.001
Prime Ortho     Modern Ortho    0.558
                Captain Ortho   0.034
Captain Ortho   Modern Ortho    0.001
Prime Ortho                     0.034
COPYRIGHT 2017 National Academy of Dentistry
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2017 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:ORIGINAL RESEARCH
Author:S., Vasu Murthy; M., Pavan Kumar; N., Praveen Kumar; M., Shiva Prasad; K., Sreekanth; V., Samatha Re
Publication:Indian Journal of Dental Advancements
Article Type:Report