Transtibial prosthetic suspension: less pistoning versus easy donning and doffing.

INTRODUCTION

The main roles of the suspension systems incorporated into lower-limb prostheses are to hold the prosthesis on the residual limb and to decrease the motion that takes place at the bone-skin-liner-socket interface during ambulation (pistoning, vertical movements within the socket) [1]. Effective suspension systems and prosthetic components can improve a person with amputation's gait and decrease his or her energy expenditure [2-3]. Prosthetic limbs should have an intimate fit with the residual limb in order to replace the lost body part with a device that offers high levels of comfort and satisfaction [3-6].

Individuals with amputation believe that both the suspension method and the fitting of a prosthetic device have significant effects on their overall satisfaction with the prosthesis [6-8]. Several questionnaires have been developed and a number of prosthetics surveys have been conducted to analyze patient satisfaction with prosthetic devices. The majority of researchers prefer the Prosthesis Evaluation Questionnaire (PEQ) as a means of evaluating differences in function, performance, and satisfaction between the different components or techniques of prosthetics fabrication and adjustment (Appendix, available online only) Good reliability and validity have been reported for the PEQ [9-11].

Evidence shows that silicone liners are preferred by many people with lower-limb amputation because they offer enhanced suspension and fit within the socket as well as improved function [3,7-8,12]. Previous research on the silicone liners has found that patient comfort and satisfaction are particularly higher in contrast with other suspension systems, such as the belt for patellar tendon bearing socket [3,8,12]. Silicone liners are believed to be more effective in controlling the pistoning within the prosthetic socket than polyethylene foam (pelite) liners. Pistoning at the socket-liner interface is said to be lower with silicone liners (1-5 mm) than with pelite liners (6.0-41.7 mm) [13-21].

Based on the literature, the pistoning is correlated with the prosthetic suspension system and fit [15]. Thus, both clinicians and researchers should be able to determine the quality of suspension and prevent the negative effects of pistoning (such as gait deviation, skin breakdown, and discomfort) by pistoning measurement [13-22].

A number of methods exist to measure the pistoning of various interfaces within the socket (liner-socket) or the residual limb (bone-soft tissue). These include X-ray [12,20,23-25], spiral computerized tomography [26], and photoelectric sensors [22]. These measurement methods are mostly useful for measuring the bone movement inside the socket. Recently, two new methods were introduced for the liner-socket interface in transtibial prostheses: a photographic method and a motion analysis system [16-19]. The literature review revealed that the majority of researchers measured the pistoning during quiet standing (static) and only a few had evaluated the pistoning that occurred inside the socket during gait (dynamic) [15].

A previous study by Gholizadeh et al. revealed low levels of pistoning for the seal-in suspension (Seal-In X5 liner, Ossur; Reykjavik, Iceland) than the locking system (Dermo liner, Ossur) during standing [16]. The findings of that study motivated this current research and prompted investigation on the effects of these suspension systems during gait along with patient satisfaction. To our knowledge, no study has previously compared the quality of suspension systems during gait and the associated levels of patient satisfaction.

METHODS

Subjects

Ten subjects with unilateral transtibial amputation participated in this study. We determined the participants' mobility grade based on the guidelines of the American Academy of Orthotists & Prosthetists [27]. Table 1 lists subject characteristics.

In order to be eligible for the study, subjects with transtibial amputation were required to be unilateral, without pain or ulcer on the residual limb, and with a residual-limb length not less than 13 cm. Furthermore, they could not have volume fluctuation in the residual limb, could not depend on assistive devices such as a cane or crutches for ambulation, and had to have good upper-limb strength.

Procedures

Two transtibial prostheses (Figure 1) were manufactured for each subject. Two different suspension systems were used: Seal-In X5 liner with valve (Icelock Expulsion Valve 551, Ossur) and Dermo liner with shuttle lock (Icelock Clutch 4H 214, Ossur). All prosthetic feet were Flex-Foot Talux (Ossur) [16,18].

One of the researchers (registered prosthetist) designed, fit, and aligned all the prosthetic limbs. Two separate total surface bearing sockets were fabricated individually for each of the two liners that were used in the study. Transparent thermoplastic material (NorthPlex 12 mm, North Sea Plastics Ltd; Glasgow, United Kingdom) enabled us to check the socket fit. The subjects attended a gait training session in the Brace and Limb Laboratory (Department of Biomedical Engineering, University of Malaya, Malaysia).

The prosthetist ensured that there was no gait abnormality and that the fit of the prosthetic sockets was satisfactory. We determined prosthetic alignment through bench, static (standing in an upright position), and dynamic (during walking) alignment. All subjects had an acclimation period of 4 weeks for each prosthetic device. To ensure subject safety, one definite socket was also made for each liner type for the 4-week acclimation period. Check sockets were used only during the kinematic experiments.

[FIGURE 1 OMITTED]

Following the trial period, we performed pistoning evaluation in the motion analysis laboratory with the Vicon 612 system using seven MXF20 motion capture cameras (Vicon; Los Angeles, California), which is believed to have an accuracy level of less than [+ or -] 0.1 mm [28]. We adopted a sampling rate of 200 Hz for the data collection. The signals from the motion analysis system were filtered by a Butter-worth filter (cutoff frequency of 10 Hz).

We fixed 16 reflective markers to the subjects' lower limbs in accordance with the Helen Hayes marker set. The knee and tibia markers for the prosthetic leg were located on the lateral proximal socket wall and the lateral distal end of the socket, respectively (Figure 2). We placed two additional markers on the liner under the knee joint level (LLin1) and 5 cm below that (LLin2) [16].

Because knee joint movement could affect the actual pistoning values, we positioned the additional markers (LLin1 and LLin2), aligned by laser liner, on the liner below the knee joint. With the transparent socket, the markers were visible through the hard socket and detectable by the cameras [16]. By fixing the markers to one segment (the shank), we avoided knee movements leading to unreal displacement.

The transparent socket could create some reflections that could be mistakenly considered as markers, therefore we used paper tape (except for the areas where additional markers were located) to mask the socket wall [16]. Prior to the test, we asked subjects to walk in the motion analysis laboratory in order to accustom themselves to the environment. Afterward, the subjects walked at a self-selected speed on an 8 m walkway. We recorded five successful trials per subject with each type of liner. We considered a trial to be successful if the cameras could capture all the markers. We could measure the pistoning by analyzing the markers' positions; however, in order to detect one gait cycle in each trial, we also used two Kistler force plates. There was a 1 min rest interval between the trials. We used the distance between the markers on the liner and on the socket to identify the piston motion.

The reproducibility of measurements was evaluated by intraobserver intrasession, intraobserver intersession, and interobserver intersession variabilities. Two observers performed the experiments over two sessions with a 1 week interval.

[FIGURE 2 OMITTED]

Prosthesis Evaluation Questionnaire

Following the experiments, we asked the subjects to complete one PEQ for each studied liner. We used some parts of the PEQ to quantitatively assess patient satisfaction [10]. The PEQ consisted of the following three sections:

* Demographic data (sex, age, weight, height, time since amputation, and cause of amputation).

* Satisfaction (fit, donning and doffing, sitting, walking on level surface, walking on unlevel ground, ascending and descending stairs, cosmesis, and overall satisfaction).

* Problems (sweat, wound, skin irritation, pistoning, pain, swelling [edema], smell, and unwanted sounds).

We rated the responses on a scale from 0 to 100, where 0 indicated "dissatisfaction or extreme problems" with the system and 100 indicated "complete satisfaction or no problems."

We used SPSS 18.0 (IBM Corporation; Armonk, New York) for the data analyses, with p-values set at 0.05. A paired-samples t-test compared the effects of the two different liners on pistoning during each gait cycle. We divided the gait cycle (stance and swing) into eight phases. We divided the stance phase by initial contact, loading response, midstance, terminal stance, and pre-swing. Initial swing, midswing, and terminal swing formed the swing phase of gait. In order to analyze the data, we first calculated the peak pistoning that occurred during each phase of one gait cycle for one gait trial of each subject. Following that, we computed the average peak pistoning that occurred across five successful gait trials. Finally, we found the overall average of peak pistoning across the different phases of gait for all 10 subjects for the comparison between the liners.

RESULTS

Pistoning Evaluation

The mean time since amputation was 7 years and all subjects had undergone amputation at least 3 years prior to study participation. The reproducibility of the measurements across the different trials of one session and between two sessions by two observers was shown to be high. The intraclass correlation coefficients of intraobserver intrasession, intraobserver intersession, and interobserver intersession were 0.92, 0.87, and 0.79, respectively.

The results of the motion analysis revealed that the amount of pistoning that occurred when the Seal-In X5 liner was used was significantly less than the pistoning with the Dermo liner throughout the gait cycle (p < 0.05), with the exception of loading response (0.5 mm), midstance (0.0 mm) and terminal stance (0.0 mm). Both liners exhibited no pistoning during preswing (Table 2, Figures 3-4).

During initial contact, the Dermo liner was displaced 5.1 [+ or -] 0.7 mm (mean [+ or -] standard deviation) within the socket. However, this value decreased rapidly to 0.0 mm at the end of loading response and remained the same until the initial swing. Only 1.9 [+ or -] 0.4 mm of pistoning was found with the Seal-In X5 liner during initial contact. Maximum displacements in 10 subjects were 5.4 [+ or -] 0.6 mm for the Dermo and 2.5 [+ or -] 0.4 mm for the Seal-In X5 liners during the initial swing.

Satisfaction

The PEQ revealed that the subjects were overall more satisfied (p < 0.05) with the Dermo liner than the Seal-In X5 liner. Nevertheless, many of them mentioned increased levels of pain and pistoning when using the Dermo liner. Donning and doffing the Seal-In X5 liner was more difficult, but the satisfaction with the socket fit was higher (Table 3). The participants also stated that the prosthesis with the Seal-In X5 liner acted like a natural part of their body and that they did not experience any traction at the end of the liner.

DISCUSSION

Selecting a suitable suspension system for individuals who have undergone transtibial amputation is a critical issue in rehabilitation [7,16,18]. In this study, we evaluated two different prosthetic suspension systems in 10 subjects with transtibial amputation to compare pistoning movement and patient satisfaction with the device during ambulation. The Vicon motion system was introduced for the purpose of evaluating pistoning during gait.

The literature review revealed that the majority of existing research was based on pistoning measurement in the static position of quiet standing as opposed to walking [15]. The complications of taking such measurements during gait and concerns over subject safety by exposure to X-ray hampered such practice [22]. A few studies attempted to use videofluoroscopy [29], photoelectric sensors [22], or axial movement detectors [30] to measure the pistoning that occurred during ambulation. However, they were only able to measure vertical movement between the pelite liner and socket [22,30]. Among them, only Sanders et al. provided the value of the pistoning that occurred across different phases of gait [22].

In this study, the Vicon motion system was shown to be an efficient method of measuring the pistoning at the liner-socket interface during the gait. It also offered a harmless method of pistoning measurement [22]. However, it is unable to detect bone displacement within the soft tissue.

Pistoning

This study showed that the Seal-In X5 liner helps to decrease pistoning by developing suction against the socket wall. The resultant suction ensures firm attachment between the liner and the socket wall. The purpose of silicone liners is to provide enhanced suspension by causing less pistoning within the prosthetic socket [3,8,12,15]. The findings of this study support this statement because the pistoning values with both Seal-In X5 and Dermo liners were lower than those found with the polyethylene foam liners [22,29-30].

With the exception of the preswing phase, we found significant differences between the two liners during the gait cycle (p < 0.05) (Figures 3-4, Table 2). These significant differences can be attributed to the different elongation properties of the liners used [16,18]. The pistoning that occurred during the initial swing might have been high as a result of peak flexion in the knee joint.

[FIGURE 3 OMITTED]

[FIGURE 4 OMITTED]

Finally, as a result of centrifugal forces, the pistoning increased between the liners and socket during the terminal swing. We noted significant difference in pistoning between the studied liners during this phase of the gait (p < 0.05), which can be associated with the firm attachment between the Seal-In X5 liner and the socket.

Satisfaction

Prosthetic satisfaction is an issue influenced by several factors. Prosthetic users require more time and energy to don and doff the Seal-In X5 liner [16,18]. They also need lubricant sprays to facilitate donning. Moreover, hand dexterity is more critical for donning and doffing a Seal-In X5 liner than for the Dermo liner. All locking liners usually have an umbrella-shaped feature at the distal part that is connected distally to a pin. Weight bearing during ambulation over this rigid and small pin may result in pain at the distal end of the residual limb [31].

The Seal-In X5 liner seems to resolve the so-called problem of "milking" (distal tissue stretch caused by the pin and lock) [32]. This milking phenomenon can also result in pain, particularly at the end of the tibia and along the tibial crest. The subjects in the current study had more pain with the pin and lock suspension (Dermo liner) than the Seal-In X5 liner.

Little is known about the effects of different prosthetic components and systems on patient satisfaction with prostheses. Effortless donning and doffing does appear to have a positive effect on satisfaction with a prosthesis [6]. The participants of this study were mainly dissatisfied with the Seal-In X5 in terms of donning and doffing and many of them specified that donning and doffing was significantly easier with the Dermo liner than with the Seal-In X5 liner. As such, the subjects stated a preference for this suspension system over the Seal-In X5 liner for long-term use.

One limitation of this study was the small sample size, particularly for the satisfaction survey. In addition to this, further research is needed to compare more suspension alternatives in order to provide a better guideline for suspension system selection. Future research should also investigate and compare the effects of these suspension systems on proprioception.

CONCLUSIONS

In conclusion, amputation rehabilitation is influenced by appropriate choice of prosthetic components in accordance with the real needs of the individual. We can infer from the results of this study that the Seal-In X5 liner decreased the pistoning within the prosthetic socket significantly, possibly as a result of the strong suction seal between the liner and the socket. Nevertheless, the subjects had difficulty with donning and doffing. We can therefore conclude that pistoning may not be the main factor that determines subjects' overall satisfaction with the prosthesis.

The study introduced a new method for evaluating the pistoning at the liner-socket interface in transtibial prostheses during gait. The Vicon system has the potential to detect the pistoning during gait while also offering a safer alternative to X-ray. Further studies are needed to come to a "gold standard" for pistoning.

At a Glance

A prosthetic leg should be securely attached to the limb so that the user feels it as part of his or her own body. Different methods have been used to evaluate the pistoning movement of the residual limb inside the socket. The majority of existing research is based on pistoning measurement in quiet standing rather than walking. We introduce a new technique for evaluating prosthetic suspension during walking using a motion analysis system. This study revealed that other factors such as donning and doffing might be more clinically significant in a person with amputation's satisfaction than pistoning once a reasonable level of pistoning is reached.

Abbreviations: PEQ = Prosthesis Evaluation Questionnaire.

ACKNOWLEDGMENTS

Author Contributions:

Study concept and design: H. Gholizadeh, N. A. Abu Osman, A. Eshraghi, S. Ali, S. K. Savarsson, W. A. B. Wan Abas.

Acquisition of data: H. Gholizadeh, A. Eshraghi.

Analysis and interpretation of data: H. Gholizadeh, N. A. Abu Osman, A. Eshraghi, S. Ali.

Drafting of manuscript: H. Gholizadeh, A. Eshraghi.

Critical revision of manuscript for important intellectual content: H. Gholizadeh, N. A. Abu Osman, A. Eshraghi, S. Ali, S. K. Savarsson, W. A. B. Wan Abas.

Obtained funding: H. Gholizadeh, N. A. Abu Osman, S. K. Savarsson. Administrative, technical, or material support: N. A. Abu Osman, S. K. Savarsson, G. H. Pirouzi.

Study supervision: N. A. Abu Osman, W. A. B. Wan Abas.

Financial Disclosures: The authors have declared that no competing interests exist.

Funding/Support: This material was based on work supported by the Malaysia UM/MOHE/HIR (grant D000014-16001) and the prosthetic components were donated by Ossur.

Additional Contributions: The authors would like to thank Mrs. Elham Sadat Yahyavi, Ms. Asa Guolaug Luoviksdottir, Dr. Nader Ale Ebrahim, and Mr. Scott Elliott for providing technical advice.

Institutional Review: The ethical approval was granted from the University of Malaya Medical Centre Ethics Committee. All subjects were asked to provide written informed consent.

Participant Follow-Up: The authors do not plan to inform participants of the publication of this study due to a lack of contact information.

REFERENCES

[1.] Michael JW. Prosthetic suspensions and components. In: Smith DG, Michael JW, Bowker JH, editors. Atlas of amputations and limb deficiencies: Surgical, prosthetic, and rehabilitation principles. 3rd ed. Rosemont (IL): American Academy of Orthopaedic Surgeons; 2004. p. 409-25.

[2.] Schmalz T, Blumentritt S, Jarasch R. Energy expenditure and biomechanical characteristics of lower limb amputee gait: the influence of prosthetic alignment and different prosthetic components. Gait Posture. 2002;16(3):255-63. [PMID:12443950] http://dx.doi.org/10.1016/S0966-6362(02)00008-5

[3.] Baars EC, Geertzen JH. Literature review of the possible advantages of silicon liner socket use in trans-tibial pros theses. Prosthet Orthot Int. 2005;29(1):27-37. [PMID:16180375] http://dx.doi.org/10.1080/17461550500069612

[4.] Czerniecki JM, Gitter AJ. Gait analysis in the amputee: Has it helped the amputee or contributed to the development of improved prosthetic components? Gait Posture. 1996;4(3):258-68. http://dx.doi.org/10.1016/0966-6362(96)01073-9

[5.] Goh JC, Lee PV, Chong SY. Stump/socket pressure profiles of the pressure cast prosthetic socket. Clin Biomech (Bristol, Avon). 2003;18(3):237-43. [PMID:12620787] http://dx.doi.org/10.1016/S0268-0033(02)00206-1

[6.] Legro MW, Reiber G, del Aguila M, Ajax MJ, Boone DA, Larsen JA, Smith DG, Sangeorzan B. Issues of importance reported by persons with lower limb amputations and pros theses. J Rehabil Res Dev. 1999;36(3):155-63. [PMID:10659798]

[7.] Baars EC, Dijkstra PU, Geertzen JH. Skin problems of the stump and hand function in lower limb amputees: A historic cohort study. Prosthet Orthot Int. 2008;32(2):179-85. [PMID:18569886] http://dx.doi.org/10.1080/03093640802016456

[8.] Kristinsson O. The ICEROSS concept: a discussion of a philosophy. Prosthet Orthot Int. 1993;17(1):49-55. [PMID:8337100] http://dx.doi.org/10.3109/03093649309164354

[9.] Berke GM, Fergason J, Milani JR, Hattingh J, McDowell M, Nguyen V, Reiber GE. Comparison of satisfaction with current prosthetic care in veterans and servicemembers from Vietnam and OIF/OEF conflicts with major traumatic limb loss. J Rehabil Res Dev. 2010;47(4):361-71. [PMID:20803404] http://dx.doi.org/10.1682/JRRD.2009.12.0193

[10.] Legro MW, Reiber GD, Smith DG, del Aguila M, Larsen J, Boone D. Prosthesis evaluation questionnaire for persons with lower limb amputations: assessing prosthesis-related quality of life. Arch Phys Med Rehabil. 1998;79(8):931-38. [PMID:9710165] http://dx.doi.org/10.1016/S0003-9993(98)90090-9

[11.] Van de Weg FB, Van der Windt DA. A questionnaire survey of the effect of different interface types on patient satisfaction and perceived problems among trans-tibial amputees. Prosthet Orthot Int. 2005;29(3):231-39. [PMID:16466153] http://dx.doi.org/10.1080/03093640500199679

[12.] Narita H, Yokogushi K, Shii S, Kakizawa M, Nosaka T. Suspension effect and dynamic evaluation of the total surface bearing (TSB) trans-tibial prosthesis: a comparison with the patellar tendon bearing (PTB) trans-tibial prosthesis. Prosthet Orthot Int. 1997;21(3):175-78. [PMID:9453088]

[13.] Yigiter K, Sener G, Bayar K. Comparison of the effects of patellar tendon bearing and total surface bearing sockets on prosthetic fitting and rehabilitation. Prosthet Orthot Int. 2002;26(3):206-12. [PMID:12562067] http://dx.doi.org/10.1080/03093640208726649

[14.] Commean PK, Smith KE, Vannier MW. Lower extremity residual limb slippage within the prosthesis. Arch Phys Med Rehabil. 1997;78(5):476-85. [PMID:9161365] http://dx.doi.org/10.1016/S0003-9993(97)90160-X

[15.] Eshraghi A, Osman NA, Gholizadeh H, Karimi MT, Ali S. Pistoning assessment in lower limb prosthetic sockets. Prosthet Orthot Int. 2012;36(1):15-24. [PMID:22269941] http://dx.doi.org/10.1177/0309364611431625

[16.] Gholizadeh H, Osman NA, Kamyab M, Eshraghi A, Abas WA, Azam MN. Transtibial prosthetic socket pistoning: static evaluation of Seal-In([R]) X5 and Dermo([R]) Liner using motion analysis system. Clin Biomech (Bristol, Avon). 2012;27(1):34-39. [PMID:21794965] http://dx.doi.org/10.1016/j.clinbiomech.2011.07.004

[17.] Gholizadeh H, Abu Osman NA, Luviksdottir A, Eshraghi A, Kamyab M, Wan Abas WA. A new approach for the pistoning measurement in transtibial prosthesis. Prosthet Orthot Int. 2011;35(4):360-64. [PMID:21975850] http://dx.doi.org/10.1177/0309364611423130

[18.] Gholizadeh H, Abu Osman NA, Kamyab M, Eshraghi A, Luviksdottir AG, Wan Abas WA. Clinical evaluation of two prosthetic suspension systems in a bilateral transtibial amputee. Am J Phys Med Rehabil. 2012;91(10):894-98. [PMID:22173083] http://dx.doi.org/10.1097/PHM.0b013e31823c74d7

[19.] Gholizadeh H, Abu Osman NA, Luoviksdottir AG, Kamyab M, Eshraghi A, Ali S, Wan Abas WA. A new method for measuring pistoning in lower limb prosthetic. Proceedings of 5th Kuala Lumpur International Conference on Biomedical Engineering; 2011 Jun 20-23; Kuala Lumpur, Malaysia. Berlin Heidelberg (Germany): Springer; 2011. p. 728-31.

[20.] Lilja M, Johansson T, Oberg T. Movement of the tibial end in a PTB prosthesis socket: a sagittal X-ray study of the PTB prosthesis. Prosthet Orthot Int. 1993;17(1):21-26. [PMID:8337097] http://dx.doi.org/10.3109/03093649309164351

[21.] Newton RL, Morgan D, Schreiber MH. Radiological evaluation of prosthetic fit in below-the-knee amputees. Skeletal Radiol. 1988;17(4):276-80. [PMID:3212490] http://dx.doi.org/10.1007/BF00401811

[22.] Sanders JE, Karchin A, Fergason JR, Sorenson EA. A non-contact sensor for measurement of distal residual-limb position during walking. J Rehabil Res Dev. 2006;43(4): 509-16. [PMID:17123190] http://dx.doi.org/10.1682/JRRD.2004.11.0143

[23.] Convery P, Murray KD. Ultrasound study of the motion of the residual femur within a trans-femoral socket during gait. Prosthet Orthot Int. 2000;24(3):226-32. [PMID:11195358] http://dx.doi.org/10.1080/03093640008726552

[24.] Grevsten S, Erikson U. A roentgenological study of the stump-socket contact and skeletal displacement in the PTB-Suction Prosthesis. Ups J Med Sci. 1975;80(1):49-57. [PMID:1145905] http://dx.doi.org/10.3109/03009737509178991

[25.] Papaioannou G, Mitrogiannis C, Nianios G, Fiedler G. Assessment of amputee socket-stump-residual bone kinematics during strenuous activities using Dynamic Roentgen Stereogrammetric Analysis. J Biomech. 2010;43(5):871-78. [PMID:20047746] http://dx.doi.org/10.1016/j.jbiomech.2009.11.013

[26.] Madsen MT, Haller J, Commean PK, Vannier MW. A device for applying static loads to prosthetic limbs of transtibial amputees during spiral CT examination. J Rehabil Res Dev. 2000;37(4):383-87. [PMID:11028693]

[27.] American Academy of Orthotists & Prosthetists. Medicare guideline forms: K-level determination (PSC044). Washington (DC): American Academy of Orthotists & Prosthetists; 2010.

[28.] Jenkins SP. Sports science handbook/2, I-Z. Brentwood (UK): Multi-Science; 2005.

[29.] Bocobo CR, Castellote JM, MacKinnon D, Gabrielle-Bergman A. Videofluoroscopic evaluation of prosthetic fit and residual limbs following transtibial amputation. J Rehabil Res Dev. 1998;35(1):6-13. [PMID:9505248]

[30.] Wirta RW, Golbranson FL, Mason R, Calvo K. Analysis of below-knee suspension systems: effect on gait. J Rehabil Res Dev. 1990;27(4):385-96. [PMID:2089149] http://dx.doi.org/10.1682/JRRD.1990.10.0385

[31.] Street G. Vacuum suspension and its effects on the limb. Orthopadie Technik. 2006;IV:1-6.

[32.] Beil TL, Street GM. Comparison of interface pressures with pin and suction suspension systems. J Rehabil Res Dev. 2004;41(6A):821-28. [PMID:15685470] http://dx.doi.org/10.1682/JRRD.2003.09.0146

Submitted for publication November 21, 2011. Accepted in revised form March 23, 2012.

This article and any supplementary material should be cited as follows:

Gholizadeh H, Abu Osman NA, Eshraghi A, Ali S, Saevarsson SK, Wan Abas WA, Pirouzi GH. Transtibial prosthetic suspension: Less pistoning versus easy donning and doffing. J Rehabil Res Dev. 2012;49(9):1321-30.

http://dx.doi.org/10.1682/JRRD.2011.11.0221

ResearcherID: Hossein Gholizadeh, MEngSc: G-48382012; Noor A. Abu Osman, PhD: B-9265-2010; Arezoo Eshraghi, PhD: A-4405-2011

Hossein Gholizadeh, MEngSc; (1) * Noor Azuan Abu Osman, PhD; (1) Arezoo Eshraghi, MSc; (1) Sadeeq Ali, BSc; (1) Stefan Karl Saevarsson, MSc; (2) Wan Abu Bakar Wan Abas, PhD; (1) Gholam Hossein Pirouzi, BSc (1)

(1) Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, Kuala Lumpur, Malaysia;

(2) Department of Biomedical Engineering, University of Calgary, Calgary, Canada

* Address all correspondence to Hossein Gholizadeh, MEngSc; Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, 50603, Kuala Lumpur, Malaysia; 603-7967-4581; fax: 603-7967-4579.

Email: gholizadeh@um.edu.my

http://dx.doi.org/10.1682/JRRD.2011.11.0221
Table 1.

Subject characteristics.

Subject   Age    Height   Mass
          (yr)    (cm)    (kg)

1          45     168      75
2          35     173      90
3          22     168      60
4          71     181      75
5          49     167      64
6          37     177      99
7          51     160      57
8          52     165      60
9          62     169      72
10         34     172      86

Subject    Cause of    Amputated
          Amputation     Side

1          Diabetic      Left
2           Trauma       Left
3           Trauma       Left
4          Diabetic      Left
5           Trauma       Right
6          Diabetic      Right
7          Diabetic      Right
8          Diabetic      Left
9           Trauma       Right
10          Trauma       Left

Subject    Residual-     Mobility
          Limb Length     Grade
            (cm) *      ([dagger])

1             14            K2
2             15            K3
3             14            K3
4            13.5           K2
5             13            K3
6             17            K2
7             14            K3
8             15            K3
9             13            K2
10            16            K3

* Inferior edge of patella to distal
end of residual limb.

([dagger]) Based on American Academy
of Orthotists & Prosthetists scale.
K2 = Patient has ability or potential
for ambulation with ability to traverse
low-level environmental barriers such
as curbs, stairs, or uneven surfaces--a
typical community ambulator. K3 =
Patient has ability or potential for
ambulation with variable cadence--a
typical community ambulator with
ability to traverse most environmental
barriers and may have vacation,
therapeutic, or exercise activity that
demands prosthetic use beyond simple
locomotion.

Table 2.

Average of displacement in different phases of gait cycle (n = 10).

Phase        Suspension      Displacement,         p-Value
                            Mean [+ or -] SD
                                  (mm)

Initial      Dermo *        5.1 [+ or -] 0.7   <0.001 ([dagger])
  Contact    Seal-In X5 *   1.9 [+ or -] 0.4   <0.001 ([dagger])
Loading      Dermo          0.5 [+ or -] 0.1
  Response   Seal-In X5     1.6 [+ or -] 0.4   <0.001 ([dagger])
Midstance    Dermo                 0
             Seal-In X5     0.8 [+ or -] 0.2    0.02 ([dagger])
Terminal     Dermo                 0
  Stance     Seal-In X5     0.3 [+ or -] 0.1
Preswing     Dermo                 0                   -
             Seal-In X5            0           <0.001 ([dagger])
Initial      Dermo          5.4 [+ or -] 0.6
  Swing      Seal-InX5      2.5 [+ or -] 0.4   <0.001 ([dagger])
Midswing     Dermo          4.2 [+ or -] 1.1
             Seal-In X5     1.7 [+ or -] 0.5   <0.001 ([dagger])
Terminal     Dermo          5.1 [+ or -] 0.7
  Swing      Seal-In X5     1.9 [+ or -] 0.4

* Ossur; Reykjavik, Iceland.

([dagger]) Statistically significant.

SD = standard deviation.

Table 3.

Comparison of satisfaction and perceived problems
with Dermo (Ossur; Reykjavik, Iceland) and Seal-In
X5 (Ossur) liners.

Satisfaction   Suspension   Mean *    p-Value
               Type

Fit            Dermo        75.59      0.003
               Seal-In X5   87.09    ([double
                                     dagger])
Donning and    Dermo        87.50     <0.001
  Doffing      Seal-In X5   35.44    ([double
                                     dagger])
Sitting        Dermo        76.30        --
               Seal-In X5   77.80
Walking        Dermo        78.25      0.001
               Seal-In X5   85.80    ([double
                                     dagger])
Uneven         Dermo        75.20      0.04
  Walking      Seal-In X5   80.30    ([double
                                     dagger])
Stair          Dermo        76.50      0.09
               Seal-In X5   78.75
Overall        Dermo        85.77      0.004
               Seal-In X5   75.20    ([double
                                     dagger])
Cosmetic       Dermo        82.50      0.46
               Seal-In X5   80.75

Problem      Suspension      Mean       p-Value
             Type         ([dagger])

Sweat        Dermo           70.53       0.082
             Seal-In X5      72.50

Wound        Dermo          100.00         --
             Seal-In X5     100.00

Irritation   Dermo          100.00         --
             Seal-In X5     100.00
Pistoning    Dermo           72.50      <0.001
Within       Seal-In X5      95.75     ([double
  Socket                               dagger])
Pain         Dermo                      <0.001
             Seal-In X5                ([double
                                       dagger])
Swelling     Dermo          100.00
(edema)      Seal-In X5     100.00
Smell        Dermo           95.80       0.15
             Seal-In X5      94.00

Sound        Dermo           74.85      <0.001t
             Seal-In X5      95.50     ([double
                                       dagger])

* Greater mean indicates higher satisfaction.

([dagger]) Greater mean indicates less complaints/problems.

([double dagger]) Statistically significant.
COPYRIGHT 2012 Department of Veterans Affairs
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2012 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Author:Gholizadeh, Hossein; Osman, Noor Azuan Abu; Eshraghi, Arezoo; Ali, Sadeeq; Saevarsson, Stefan Karl;
Publication:Journal of Rehabilitation Research & Development
Article Type:Clinical report