Total protein determination in urine: elimination of a differential response between the Coomassie Blue and Pyrogallol Red protein dye-binding assays.

The total protein concentration in urine can be determined by biuret assay (1-3), precipitation with trichloroacetic acid-Ponceau S (TCA-PS) [1] (4), turbidimetry [using TCA (5), sulfosalicylic acid (6), or benzethonium chloride (BEC) (7)], or protein dye-binding assays utilizing Coomassie Brilliant Blue (CBB) (8-19) or Pyrogallol Redmolybdate (PRM) (20-22). Urinary protein determinations are notoriously unreliable (23-27), and comparative evaluation of a range of assay methods (28-31) has failed to establish a convenient method of choice. However, protein dye-binding assays are simple, rapid, and readily automated (10, 12, 13, 15, 16, 18-20, 22) and are becoming increasingly popular for the clinical determination of total protein in urine. The present study indicates a lack of agreement in the urine protein concentrations measured by the Sigma CBB and PRM test kits as recommended by the manufacturer. The use of an alternative protein calibrator prepared from Sigma urinary protein lyophilizate (UPL) overcomes this problem and simultaneously improves the comparability of the protein concentration values obtained with the dye-binding assays relative to the TCA and BEC methods.

Materials and Methods

MATERIALS

CBB protein dye reagent (cat. no. 610-A), Microprotein-PR[TM] dye reagent (cat. no. 611-A), human albumin (HA) protein calibrators (cat. nos. 610-30 and 610-50), the albumin/ globulin (A/G) protein calibrator (cat. no. 540-10), urine control (UC) Levels 1 and 2 (cat. nos. U9506 and U9631), and UPL (cat. no. U8126) were purchased from Sigma Diagnostics. Bovine serum albumin (BSA) calibrator was purchased from Pierce & Warriner.

SAMPLES

Urine. Specimens (n = 60), covering a wide range of protein concentrations, were collected without preservatives, centrifuged (2500g for 10 min), and stored at 270[degrees]C. Immediately before protein assay, the samples were thawed at room temperature.

Controls. Sigma urine controls Levels 1 and 2 (UC 1 and UC 2) were reconstituted in 10 mL of ultrapure water (MilliQ Plus Water Purification System; Millipore) and stored at 4[degrees]C (for up to 5 days) as recommended by the manufacturer.

UPL. Sigma UPL (5 mg protein) was reconstituted in 2.5 mL of 0.15 mol/L sodium chloride containing 1 g/L sodium azide, and the mixture was stirred gently for 2 h at room temperature. After centrifugation (10 000g for 5 min), the protein content of the supernate (UPL stock solution, 1.9 g/L) was determined by Lowry assay (32), and the UPL was diluted (in saline/sodium azide) to give 300 mg/L and 500 mg/L UPL protein calibrators, which were stored at -20[degrees]C in 100-[micro]L aliquots. Immediately before protein assay, the UPL protein calibrator was thawed at room temperature and gently vortex-mixed for 1 min.

PROTEIN ASSAYS

CBB protein dye-binding assay. The assay procedure was performed manually as described by the manufacturer (33), except that the volume of the assay mixture was scaled down from 2.55 mL to 1.02 mL. Briefly, 20 [micro]L (rather than 50 [micro]L) of sample (urine, diluted urine, or UC) or protein calibrator (HA, BSA, A/G, or UPL at 300 mg/L) was gently mixed with 1.0 mL (rather than 2.5 mL) of Protein Assay Solution (Protein Dye Reagent diluted with four volumes of ultrapure water in a plastic container), and after 2-30 min, the absorbance of the assay mixture was measured ([A.sub.595 nm]) in a plastic cuvette by a Jenway 6100 spectrophotometer (Dunmow) zeroed with a reagent blank. Protein concentrations were calculated as recommended by the manufacturer from the ratio of the absorbance value of the sample and 300 mg/L calibrator. In subsequent studies, UC 1 was assayed at 20, 50, 100, 150, and 200 [micro]L, and UC 2 was assayed at 20, 40, 50, 60, 80, and 100 [micro]L; protein concentrations were derived by extrapolation of the absorbance values from a calibration curve.

PRM protein dye-binding assay. The assay procedure was performed manually as described by the manufacturer (34). Briefly, 20 [micro]L of sample (urine, diluted urine, or UC) or protein calibrator (HA, BSA, A/G, or UPL at 500 mg/L) was gently mixed with 1 mL of Microprotein-PR Reagent, and after 3-15 min, the absorbance of the assay mixture was measured ([A.sub.600 nm]) against water, followed by subtraction of the absorbance of a reagent blank. Protein concentrations were calculated as recommended by the manufacturer from the ratio of the absorbance values of the sample and the 500 mg/L calibrator. In subsequent studies, UC 1 was assayed at 20, 50, 100, 150, and 200 [micro]L, and UC 2 was assayed at 20, 40, 50, 60, 80, and 100 [micor]L; protein concentrations were derived by extrapolation of the absorbance values from a calibration curve.

Lowry protein assay (32). Reagent C (1 mL) was mixed with 0.1 mL of HA calibrator or 10 [micro]L of UPL stock solution (adjusted to 100 [micro]L with 0.15 mol/L sodium chloride containing 1 g/L sodium azide) followed, after 15 min, by 0.1 mL of reagent E. The absorbance of the assay mixture was measured ([A.sub.540 nm]) against a reagent blank after 30 min, and protein concentrations were derived by extrapolation of the absorbance values from a calibration curve.

[FIGURE 1 OMITTED]

TCA protein assay (35). TCA (0.25 mL of a 125 g/L solution) was mixed with 1 mL of sample (UC) or protein calibrator (A/G or UPL at 300 mg/L), and after 10 min, the absorbance of the assay mixture was measured ([A.sub.620 nm]) against a sample blank (1 mL of sample plus 0.25 mL of 0.15 mol/L sodium chloride containing 1 g/L sodium azide). Protein concentrations were calculated from the ratio of the absorbance values of the sample and the 300 mg/L calibrator. In subsequent studies, UCs 1 and 2 were each assayed at 0.2, 0.4, 0.6, 0.8, and 1 mL, and protein concentrations were derived by extrapolation of the absorbance values from a calibration curve.

[FIGURE 2 OMITTED]

BEC protein assay (7). Eight hundred microliters of 0.5 mol/L sodium hydroxide containing 33 mmol/L EDTA was mixed with 20 [micro]L of sample (UC) or protein calibrator (HA or UPL), followed immediately by the addition of 0.2 mL of 2 g/L BEC. The absorbance of the assay mixture was measured ([A.sub.660 nm]) against a reagent blank after 50 min. Protein concentrations were derived by extrapolation of the absorbance values from a calibration curve. In subsequent studies, UC 1 was assayed at 25, 50, 75, and 100 [micro]L, and UC 2 was assayed at 10, 20, 30, 40, and 50 [micro]L.

[FIGURE 3 OMITTED]

TCA-PS protein assay (4). TCA-PS concentrated reagent (50 [micro]L) was mixed with 0.5 mL of sample (UC 1 or UC 2 diluted 1:5 with saline/azide) or protein calibrator (A/G or UPL at 100 mg/L), and the mixture was centrifuged at 2500g for 5 min before removal of the supernate and resolubilization of the pellet in 1 mL of 0.8 g/L sodium hydroxide. The absorbance of the assay mixture was measured ([A.sub.560 nm]) against a reagent blank. Protein concentrations were derived by extrapolation of the absorbance values from a calibration curve. In subsequent studies, UC 1 was assayed at 100, 200, 300, 400, and 500 [micro]L, and UC 2 was assayed at 50, 100, 150, and 200 [micro]L.

Biuret protein assay (1). Ice-cold ethanolic phosphotungstic acid (5 mL) was added to duplicate 5-mL aliquots of ice-cold sample (0.8 mL of UPL calibrator diluted with 4.2 mL of saline/azide) or calibrator (A/G) and allowed to stand for 15 min in ice before centrifugation at 1000g for 10 min. The supernates were decanted, and the protein pellets were washed with 2.5 mL of ice-cold ethanol. After centrifugation (1000g for 10 min), the supernates were discarded, and the pellets of duplicate aliquots were dissolved in either 1 mL of biuret reagent or reagent without copper sulfate. After 15 min, the absorbances ([A.sub.540 nm]) of the assay tubes containing biuret reagent were measured against the duplicate assay tubes containing reagent without copper sulfate. Protein concentrations were derived by extrapolation of the absorbance values of the samples from a calibration curve.

[FIGURE 4 OMITTED]

Results

The Sigma CBB Microprotein assay gave consistently lower protein concentration values than the corresponding PRM assay when applied to urine and Sigma urine controls. The values for the urine were 110-13 500 mg/L (mean, 2390 mg/L) with the CBB assay compared with 160-18 300 mg/L (mean, 3470 mg/L) with the PRM assay (CBB:PRM protein concentration ratio, 0.46-0.88; mean, 0.69 [+ or -] 0.10). The effect was highly reproducible as demonstrated using Sigma urine controls UC 1 and UC 2 (Table 1). Thus, UC 1 gave within-day (between-day) protein values of 86 [+ or -] 2 mg/L (88 [+ or -] 3 mg/L) by CBB assay and 127 [+ or -] 4 mg/L (131 [+ or -] 8 mg/L) by PRM assay. The corresponding values for UC 2 were 262 [+ or -] 5 mg/L (259 [+ or -] 5 mg/L) by CBB assay and 438 [+ or -] 13 mg/L (440 [+ or -] 15 mg/L) by PRM assay. The respective within-day (between-day) CBB:PRM protein concentration ratios were 0.68 [+ or -] 0.02 (0.67 [+ or -] 0.04) for UC 1 and 0.60 [+ or -] 0.01 (0.59 [+ or -] 0.02) for UC 2. These ratios were unaffected by the use of BSA (rather than HA) as a protein calibrator, and A/G only slightly improved the comparability of the assays (Table 1). The use of UPL as a protein calibrator increased the CBB protein concentration values of the controls and decreased the corresponding PRM values to greatly improve comparability between the assays, i.e., within-day CBB:PRM ratio, 1.0 [+ or -] 0.03 for UC 1 and 0.85 [+ or -] 0.02 for UC 2 (Table 1). However, the latter was further improved when protein concentrations were derived using a calibration curve, i.e., the calculation of protein concentration from an absorbance ratio relative to a single protein calibrator (see Table 1) fails to take into account the nonlinear response of UPL to the assays (Fig. 1). Thus, when UC 1 and UC 2 were each assayed over a range of sample volumes (to give varying color yields within the calibration curves) and the protein concentration values were derived by extrapolation of absorbance from a UPL calibration curve, then agreement between the assays was obtained with UC 2 as well as UC 1 (Fig. 2). Repeat assays of the urine using a UPL calibration curve gave protein concentration values of 120-15 200 mg/L (mean, 2700 mg/L) with the CBB assay and 140-15 500 mg/L (mean, 2800 mg/L) with the PRM assay. Thus, the CBB:PRM protein concentration ratio was greatly improved (0.69-1.20; mean, 0.96 [+ or -] 0.11). The correlation plots for the CBB and PRM assay of the urine are shown in Fig. 3, and the respective statistical data are tabulated in Table 2. The agreement between the assays was dramatically improved when UPL (slope, 0.972; intercept, -16 mg/L) was used as protein calibrator rather than HA (slope, 0.685; intercept, 17 mg/L; Fig. 3).

[FIGURE 5 OMITTED]

We investigated the comparability between the CBB, PRM, TCA, BEC, and TCA-PS protein assays by determining the protein concentration of the urine controls over a range of sample volumes, using calibration curves (Fig. 4). When the recommended calibrator (HA or A/G) was used, the protein concentration values obtained with the CBB, PRM, TCA, BEC, and TCA-PS assays for UC 1 were 81 [+ or -] 7, 122 [+ or -] 5, 89 [+ or -] 4, 115 [+ or -] 8, and 79 [+ or -] 8 mg/L, respectively; and the UC 2 values were 264 [+ or -] 17, 455 [+ or -] 15, 369 [+ or -] 23, 350 [+ or -] 12, and 305 [+ or -] 13 mg/L, respectively. With UPL as a calibrator, the values obtained with the CBB, PRM, TCA, BEC and TCA-PS assays for UC 1 were 89 [+ or -] 8, 88 [+ or -] 4, 84 [+ or -] 4, 93 [+ or -] 9, and 61 [+ or -] 6 mg/L, respectively; and the UC 2 values were 295 [+ or -] 16, 303 [+ or -] 12, 322 [+ or -] 6, 289 [+ or -] 12, and 243 [+ or -] 11 mg/L, respectively. Thus, UPL improved the CBB:PRM ratio for UC 1 from 0.66 to 1.01, the TCA:PRM ratio from 0.73 to 0.95, and the BEC:PRM ratio from 0.94 to 1.06; UPL improved the CBB:PRM ratio for UC 2 from 0.58 to 0.97, the TCA:PRM ratio from 0.81 to 1.06, and the BEC:PRM ratio from 0.77 to 0.95.

The protein composition of the UPL calibrator was investigated by electrophoresis and densitometry. The UPL consisted predominantly of albumin (57.5%), [gamma]-globulin (6.5%), and transferrin (6.9%), plus a diverse range of minor proteins (29.1%), the latter including protein bands corresponding to [[micro].sub.1]-microglobulin, immunoglobulin light chain, and retinol-binding protein (36) (Fig. 5)

Discussion

Methods for the determination of total urinary protein that use precipitation of protein and indirect biuret are accurate, but they also are time-consuming and, therefore, not widely used. Protein dye-binding assays are rapid, simple, and readily automated (10, 12, 13, 15, 16, 18-20, 22), but they characteristically lack a uniform response to different proteins (9-20, 22, 28). This problem has been addressed by modification of the purity/concentration of the dye and the addition of sodium dodecyl sulfate to the reagent (14, 15, 18, 19, 22). However, the preparation of modified reagents (and the inherent problems of reproducibility and stability) is troublesome, and it is advantageous to use optimized test kits that are commercially available worldwide in a standardized form.

The results of the present study indicate that the Sigma Microprotein CBB assay and the corresponding PRM assay give different protein concentration values when applied to urine and Sigma urine controls. This effect is unrelated to freezing/thawing of the samples and is a potential source of confusion because it is desirable to obtain the same protein concentration values with both tests. Consequently, we have aimed to equalize the response of the assays (without modifying the dye reagents) by use of a urinary protein calibrator and commercial urine controls to minimize the compounding effects of pathological variation. Our results indicate that the disagreement between the protein concentration values obtained with the different assays reflects a difference in the response of the CBB and PRM assays to urinary protein (compared with HA) and that the problem is overcome by the use of UPL as a protein calibrator. Previous studies have indicated similar differences in the urine protein values obtained with other protein assay methods (28-31). Thus, it is interesting to note that in the present study, the use of a urinary protein calibrator (rather than HA or A/G) greatly improved the comparability of the results obtained with the CBB, PRM, TCA, and BEC protein assays when applied to the urine controls (and slightly improved their comparability with the TCA-PS method). This is an important observation because the urine controls are likewise recommended for assessing the reliability of other methods for total protein determination of urine. However, the potential value of UPL as a protein calibrator for other protein assays can only be fully evaluated with actual urine specimens, and this is beyond the scope of the present study.

In conclusion, our results indicate that the use of a urinary protein calibrator will improve the comparability of the Sigma CBB and PRM protein assay methods when used for the determination of total protein in urine. Ideally, such a calibrator should be commercially available in a stabilized form and precalibrated against a reference method such as the biuret. We have attempted biuret assays of submilligram quantities of UPL calibrator, and this suggests a protein concentration of 545 mg/L (rather than 500 mg/L as determined by Lowry assay). However, the biuret assay is insensitive (requiring milligram quantities of protein for accurate determination), and the response of the assay to UPL is nonlinear. Consequently, the accuracy of the UPL biuret value is questionable. This complicates the use of the biuret assay as a reference method for comparison of the CBB and PRM assays. In addition, this is impractical given the expense of UPL and its availability in only milligram amounts, i.e., each UPL calibration curve would require more than the maximum batch amount (10 mg) of UPL as currently supplied. The use of CBB:PRM protein concentration ratios and correlation plots of the CBB and PRM values eliminates any uncertainty relating to the absolute protein concentration of the UPL calibrator and is adequate, for the purpose of the present study, to demonstrate improved agreement between the assays.

Received August 10, 1999; accepted November 30, 1999.

References

(1.) Savory J, Pu PH, Sunderman FW. A biuret method for determination of protein in normal urine. Clin Chem 1968;14:1160-71.

(2.) Doetsch K, Gadsden RH. Determination of total urinary protein, combining Lowry sensitivity and biuret specificity. Clin Chem 1973;19:1170-8.

(3.) Rice EW. Improved biuret procedure for routine determination of urinary total proteins in clinical proteinuria. Clin Chem 1975;21: 398-401.

(4.) Pesce MA, Strande CS. A new micromethod for determination of protein in cerebrospinal fluid and urine. Clin Chem 1973;19: 1265-7.

(5.) Shahanigan S, Brown PI, Ash KO. Turbidimetric measurement of total urinary proteins: a revised method. Am J Clin Pathol 1984; 81:651-4.

(6.) Henry RJ, Sobel C, Seglove M. Turbidimetric determination of proteins with sulphosalicylic and trichloroacetic acids. Proc Soc Exp Biol Med 1956;92:748-51.

(7.) Iwata J, Nishikaze O. New micro-turbidimetric method for determination of protein in cerebrospinal fluid and urine. Clin Chem 1979;25:1317-9.

(8.) McIntosh JC. Application of dye-binding method to the determination of protein in urine and cerebrospinal fluid [Letter]. Clin Chem 1977;23:1939-40.

(9.) Schleicher E, Wieland OH. Evaluation of the Bradford method for protein determination in body fluids. J Clin Chem Clin Biochem 1978;16:533-4.

(10.) Heick HMC, Begin-Heick N, Acharya C, Mohammed A. Automated determination of urine and cerebrospinal fluid proteins with Coomassie Brilliant Blue and the Abbott ABA-100. Clin Biochem 1980;13:81-3.

(11.) Thomas L, Winckelmann M, Michaelis HC, Walb D. Quantitative determination of total urinary protein utilizing the principle of Coomassie Brilliant Blue G250 binding to protein. J Clin Chem Clin Biochem 1981;19:203-8.

(12.) Sano K, Kanamori K, Shiba A, Nakao M. Automated assay of urinary protein using Coomassie Brilliant Blue G-250. Anal Biochem 1981;113:197-210.

(13.) Lott JA, Stephan VA, Pritchard KA. Evaluation of the Coomassie Brilliant Blue G250 method for urinary protein. Clin Chem 1983; 29:1946-50.

(14.) Macart M, Gerbaut L. Evaluation of an improved Coomassie dye binding method for urinary protein assay. Clin Chim Acta 1984; 141:77-84.

(15.) Perini JM, Mizon C, Dehon B, Fialdes P, Lefebvre I, Mizon J, Roussel P. Urinary protein determination using Coomassie Brilliant Blue in the presence of sodium dodecyl sulphate. Clin Chim Acta 1984;143:321-8.

(16.) Goren MP, Li JT. The Coomassie Brilliant Blue method underestimates drug-induced tubular proteinuria. Clin Chem 1986;32: 386-8.

(17.) Marshall T, Williams KM. Comparison of Coomassie Brilliant Blue protein dye-binding assays for determination of urinary protein concentration. Clin Chem 1987;33:577-8.

(18.) Wimsatt DK, Lott JA. Improved measurement of urinary total protein (including light-chain proteins) with a Coomassie Brilliant Blue G-250-sodium dodecyl sulfate reagent. Clin Chem 1987;33: 2100-6.

(19.) Lim CW, Chisnall WN, Stokes YM, Pratt R, Crooke MJ. Effects of sodium dodecyl sulphate, dye concentration and paraprotein on Coomassie Blue dye-binding assays for protein in urine. Clin Biochem 1988;21:277-81.

(20.) Watanabe N, Kamel S, Ohkubo A, Yamanaka M, Ohsawa S, Makino K, Tokuda K. Urinary protein as measured with a Pyrogallol-Red-molybdate complex manually and in a Hitachi 726 automated analyzer. Clin Chem 1986;32:1551-4.

(21.) Phillipou G, James SK, Seaborn CJ, Phillips PJ. Screening for microalbuminuria by use of a rapid low cost colorimetric assay. Clin Chem 1989;35:456-8.

(22.) Orsonneau JL, Douet P, Massoubre C, Lustenberger P, Bernard S. An improved Pyrogallol Red-molybdate method for determining total urinary protein. Clin Chem 1989;35:2233-6.

(23.) Glen GC. The results of analyte enhancement and use of supplied urine protein standard on the CAP urine chemistry program. Am J Clin Pathol 1980;74:531-4.

(24.) Legg EF, Hurrell AE. External quality assessment of quantitative urinary analysis. Ann Clin Biochem 1984;21:491-3.

(25.) Shepherd MDS, Penberthy LA. Performance of quantitative urine analysis in Australasia critically assessed. Clin Chem 1987;33: 792-5.

(26.) Chambers RE, Bullock DG, Whicher JT. Urinary total protein estimation--fact or fiction? Nephron 1989;53:33-6.

(27.) Chambers RE, Bullock DG, Whicher JT. External quality assessment of total urinary protein estimation in the United Kingdom. Ann Clin Biochem 1991;28:467-73.

(28.) McElderry LA, Tarbit IF, Cassells-Smith AJ. Six methods for urinary protein compared. Clin Chem 1982;28:356-60.

(29.) Dilena BA, Penberthy LA, Fraser CG. Six methods for determining urinary protein compared. Clin Chem 1983;29:553-7.

(30.) Lim CW, Chisnall, WN, Stokes YM, Debnam PM, Crooke MJ. Effects of low and high relative molecular protein mass on four methods for total protein determination in urine. Pathology 1990; 22:89-92.

(31.) Macart M, Forzy G, Gerbaut L, Vekich AJ, Guilbaud JC. Measuring urinary protein with the new BioRad reagent kit: evaluation and comparison with five other methods. Ann Biol Clin 1994;51:355-60.

(32.) Lowry OH, Rosebrough NJ, Farr AL, Randall RJ. Protein measurement with the Folin phenol reagent. J Biol Chem 1951;193:265-75.

(33.) Sigma Diagnostics. Micro protein determination procedure 610. St Louis, MO: Sigma Diagnostics, 1997.

(34.) Sigma Diagnostics. Microprotein-PRTM procedure 611. St Louis, MO: Sigma Diagnostics, 1995.

(35.) Silverman LM, Christenson RH. Proteins in urine. In: Burtis CA, Ashwood ER, eds. Tietz textbook of clinical chemistry, 2nd ed. Philadelphia: WB Saunders, 1994:717-23.

(36.) Marshall T, Williams KM. Clinical analysis of human urinary proteins using high resolution electrophoretic methods. Electrophoresis 1998;19:1752-70.

[1] Nonstandard abbreviations: TCA, trichloroacetic acid; PS, Ponceau S; BEC, benzethonium chloride; CBB, Coomassie Brilliant Blue; PRM, Pyrogallol Red-molybdate; HA, human albumin; A/G, albumin/globulin; UC, urine control; UPL, urinary protein lyophilizate; and BSA, bovine serum albumin.

Thomas Marshall * and Katherine M. Williams

Analytical Biochemistry Group, School of Sciences, The University of Sunderland, Sunderland SR1 3 RG, UK.

* Author for correspondence. Fax 44-191-515-3747; e-mail tom.marshall@ sunderland.ac.uk.
Table 1. Protein concentrations for the urine controls (UC 1
and UC 2) as determined by the CBB and PRM protein dye-binding
assays calibrated against HA, BSA, A/G, or UPL.

 Protein concentration, (a) mg/L

 UC 1

 Sample CBB PRM

Within-run HA 86 [+ or -] 2 127 [+ or -] 4
 variation BSA (c) 98 [+ or -] 2 149 [+ or -] 5
 (n=20) A/G (c) 111 [+ or -] 2 145 [+ or -] 5
 UPL 1 (d) 110 [+ or -] 2 110 [+ or -] 4
 UPL 2 (d) 103 [+ or -] 2 100 [+ or -] 3

Between-run HA 88 [+ or -] 3 131 [+ or -] 8
 variation BSA (c) 93 [+ or -] 6 139 [+ or -] 14
 (n = 10) A/G (c) 105 [+ or -] 7 138 [+ or -] 11
 UPL 1 (d) 104 [+ or -] 7 116 [+ or -] 11
 UPL 2 (d) 100 [+ or -] 7 108 [+ or -] 11

 Protein concentration, (a) mg/L

 UC 1 UC 2

Sample CBB:PRM (b) CBB

HA 0.68 [+ or -] 0.02 262 [+ or -] 5
BSA (c) 0.66 [+ or -] 0.02 277 [+ or -] 5
A/G (c) 0.76 [+ or -] 0.03 320 [+ or -] 6
UPL 1 (d) 1.00 [+ or -] 0.03 316 [+ or -] 11
UPL 2 (d) 1.03 [+ or -] 0.03 297 [+ or -] 6

HA 0.67 [+ or -] 0.04 259 [+ or -] 5
BSA (c) 0.67 [+ or -] 0.05 273 [+ or -] 11
A/G (c) 0.77 [+ or -] 0.04 310 [+ or -] 13
UPL 1 (d) 0.90 [+ or -] 0.07 307 [+ or -] 12
UPL 2 (d) 0.92 [+ or -] 0.07 303 [+ or -] 31

 Protein concentration, (a) mg/L

 UC2

Sample PRM CBB:PRM (b)

HA 438 [+ or -] 13 0.60 [+ or -] 0.01
BSA (c) 486 [+ or -] 15 0.57 [+ or -] 0.01
A/G (c) 493 [+ or -] 31 0.65 [+ or -] 0.04
UPL 1 (d) 379 [+ or -] 11 0.84 [+ or -] 0.02
UPL 2 (d) 345 [+ or -] 10 0.86 [+ or -] 0.02

HA 440 [+ or -] 15 0.59 [+ or -] 0.02
BSA (c) 463 [+ or -] 24 0.59 [+ or -] 0.03
A/G (c) 465 [+ or -] 29 0.67 [+ or -] 0.04
UPL 1 (d) 369 [+ or -] 16 0.83 [+ or -] 0.06
UPL 2 (d) 346 [+ or -] 14 0.88 [+ or -] 0.11

(a) Protein concentrations were determined with the Sigma CBB and PRM
test kits as recommended for urine, using HA as a protein calibrator.
Protein concentrations were calculated (as recommended by the
manufacturer) from the ratio of the absorbance values of the sample
and the 300 mg/L (CBB) or 500 mg/L (PRIM) calibrator.

(b) The CBB:PRM protein concentration ratio was used as an index of
comparability between the assays.

(c) BSA and A/G are alternative commercially available calibrators.

(d) UPL 1 and UPL 2 are protein calibrators prepared from two batches
of the UPL.

Table 2. Correlation between the CBB and PRM protein dye-binding
assays using HA or UPL as calibrator for the assay of human
urine (n = 60).

 PRM (x)

 Median, Range,
 mg/L mg/L

Calibrator
 HA 1950 160-18300
 UPL 1650 140-15500

 CBB (y)

 Median, Range,
 mg/L mg/L

Calibrator
 HA 1300 110-13500
 UPL 1550 120-15200

 Intecept,
 Slope mg/L r

Calibrator
 HA 0.685 17 0.98
 UPL 0.972 -16 0.99
COPYRIGHT 2000 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2000 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Automation and Analytical Techniques
Author:Marshall, Thomas; Williams, Katherine M.
Publication:Clinical Chemistry