Timing and operating mode design for time-gated fluorescence lifetime imaging microscopy.

1. Introduction

Fluorescence imaging techniques are widely used in biomedical applications. Fluorescence can be classified in autofluorescence and extrinsic fluorescence. Autofluorescent substances in biological tissues like several amino acids, NADPH, and flavin can be utilized for fluorescence imaging [1]. External fluorescence is generated by the combination of fluorophore and relative target molecule. Common fluorophores include organic dyes, fluorescent protein, quantum dots, and long lifetime metal-ligand complexes (MLCs). Organic dyes such as Cy5, rhodamine series dyes have been fully used for fluorescence imaging experiments. Their lifetimes range from ns level to hundreds of ns. Fluorescent proteins are a powerful tool for biomedical imaging. One of the most famous delegates is GFP. Quantum dots can be excited by laser from 300 nm to 1000 nm according to different types and they express lifetimes between ns to hundred ns [2, 3]. MLCs possess long stokes shift and long lifetime from to ms level. One important application is intracellular oxygen molecule imaging [4, 5].

Fluorescence will decay after being excited according to a specific pattern which is often fitted by negative exponent functions [6]. Fluorescence decay process can be assorted to steady-state and time-resolved (transient) categories [7]. Target molecule barely changes during steady fluorescence decays. Differentiation of Cancer lesion area from normal tissues is one representative application of steady fluorescence measurement. Time-resolved measurement shows the dynamic process of the target molecule during imaging. Important biochemical processes like protein folding, carbon fixation in photosynthesis, and some FRET processes all take place in ms to several hundred ms [8].

Fluorescence intensity [9], polarization [10], and lifetime which carry the microenvironment changing information of the target molecule can all be utilized for imaging. Fluorescence lifetime [tau] is definite as the time when fluorescence intensity attenuates to the 1/e of the initial intensity. Fluorescence lifetime is an absolute value which is independent of excitation laser intensity, fluorophore concentration, and photobleaching. Furthermore, fluorescence lifetime can be applied to distinguish the molecules which are indiscernible using fluorescence intensity imaging methods. Flavin adenine dinucleotide (FAD) and flavin mononucleotide (FMN) share the same optimum excitation wavelength and emission wavelength; however, the different lifetimes of the two provide one solution for this dilemma [11].

Compared with fluorescence intensity microscopy (FIM), fluorescence lifetime imaging microscopy (FLIM) has the advantage that it is less influenced by dye or target molecule concentration, photo bleaching, and excitation laser intensity. Typical methods to realize FLIM include time-gated FLIM, time correlated single photon counting (TCSPC), and Frequency Modulation.

As the needs for high frame frequency and highresolution microscopy increase, the low-cost advantage of frequency modulation FLIM becomes less dominant compared with that of temporal FLIM including time-gated FLIM and TCSPC. Table 1 compares the time resolution and imaging frame of the two mainstream temporal domain methods. Mature TCSPC products can be provided by many institutes and companies. TCPSC-FLIM can achieve very high temporal resolution, as in Table 1. However, due to its scanning processing pattern, TCPSC-FLIM cannot provide as high frame frequency as the one that time-gated FLIM can realize. TCPSC-FLIM is suitable for steady fluorescence imaging which requires high time resolution such as cancer detection. As for important intraocular transient process imaging such as [Ca.sup.2+] movement and protein fusion, high frame frequency time-gated FLIM can supply a potential method for these applications. Imperial College London, UC Davis [12], University Michigan [13], and Utrecht University [14] have applied time-gated FLIM to endoscope FLIM clinical surgery, in vitro or in vivo imaging, and so forth. Important areas of time-gated FLIM focus on improvement of temporal and spatial resolution and improvement of image speed to satisfy the need for transient fluorescence imaging. Time-gated FLIM can be a complimentary method for FIM to provide a clearer and faster way of tracking the microcosms.

In all, time-gated FLIM can be performed for steady and transient fluorescence imaging applications.

2. Methods

Time-gated FLIM acquires different images at different delays according to laser excitation (Figure 1(a)), and thus lifetime images are computed through specific inversion algorithm [17]. Due to different types of time-gated FLIM systems, a simple abstract model was given to explain how the system works. Sample is excited by short pulse laser. Fluorescence images are coupled into ICCD by light path design. Control devices realize synchronization of laser and ICCD, and thus gates open at different delay according to laser. Lifetime image will be computed using processed intensity images under certain algorithms.

Lifetime algorithm was designed on Matlab R2007b, Math Works. Timing was realized using a homemade Time Control Unit (TCU) based on ARM (STM32F103VE, ST), FPGA (XC3S400, Spartan-3, Xilinx), and ISE platform [12]. TCU generates two timing sequences to be utilized to trigger pulse laser and ICCD separately. The main technical parameter of the TCU can be seen in Table 2.

3. Results and Discussion

3.1. System Control and Temporal Parameters. To acquire different intensity images at different delay in order to compute lifetime image, it is crucial to control the ICCD and laser synchronously. FLIM system can be built up by inner trigger control [18], external trigger control [19], or optical delay control [20] according to different types of ICCD and laser utilized. Figure 1(b) shows an abstract system model of different control methods and important temporal parameters involved. [t.sub.p] shows laser pulse, [tau] stands for sample lifetime, and [t.sub.g] means gate width of gated optical intensifier (GOI). Moreover, N, [d.sub.t], [DELTA]t, M, and [delta]t which represent number of gates opened in one excitation, gate interval of N gates, delaystepofdifferentexcitations, numberofexcitations and the interval between laser excitation and opening of the first gate, respectively determine different operating modes. St should be short enough to collect enough fluorescence to increase the first delay signal which is crucial for lifetime computing but not too short to saturate ICCD because of the strong laser [21]. In addition, [delta]t, dt, and [DELTA]t are defined according to the rising edge of located gate. Gated function can be summarized as a universal equation after the definition of these important temporal parameters. Timing design was based on this equation for steady and transient operating-mode of time-gated FLIM. N, dt, [DELTA]t, M, and [delta]t can either be fixed or variable according to different operating modes:

[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII]. (1)

3.2. Lifetime Computing Algorithm Design. Lifetime computing algorithm development can be seen in our former proceeding papers [6]. Parts of gated intensity images of a nematode containing GFP provided by Becker & Hickl GmbH using TCSPC method [15] can be seen in Figure 2. Original images contain a series of 8 images recorded at 1 ns to 8 ns. To realize high-speed computing under transient fluorescence circumstances, lifetime will be computed by two intensity images before next excitation under rapid lifetime determination (RLD) method. Lifetime is computed based on (2), where the interval between two images equals gate interval dt, I1 and I2 represent image intensity:

[tau] = dt/ln ([I.sub.2]/[I.sub.1]) (2)

For steady fluorescence lifetime computing, a least square algorithm is sufficient for a precise inversion. Fluorescence decay pattern is inverted per pixel through least square fit, and thus lifetime is computed by its definition in Section 1. Both RLD method and LS method results based on former GFP images can be seen in Table 3 and Figure 3.

3.3. System Operating Mode Design. Timing design was based on (1). In this paper, two timings are designed for steady-state and transient fluorescence recording.

For steady-state fluorescence, an operating mode and timing are designed in Figure 4(a). For each excitation, only one intensity image will be recorded(IV = 1) to accommodate with low frame frequency ICCD, namely, one-excitation onesampling mode. M and At are determined by laser RF. At is fixed while St is variable. This timing can be explained by (3) which is degenerated from (1):

g[(t).sub.s] = [(M -1)summation over (j = 0)] w (t + [DELTA]t + [[delta]t.sub.(j)]). (3)

Delay among each excitation and opening of its gate [delta]t during one sampling determines delays of each intensity image. Imaging inversion speed is not as important as transient fluorescence lifetime in steady case; therefore, algorithms with high accuracy will be more suitable for steady fluorescence lifetime applications. In which case, LS method can be utilized to process different delay images after the former procedure.

However, for transient fluorescence process such as [Ca.sup.2+] movement which usually occurs in a time order of hundreds or even tens of milliseconds, high-frequency imaging frame and high-speed lifetime computing algorithm are highly emphasized. In which case, Figure 4(b) provides a feasible mode and timing to realize transient fluorescence recording. Several gates (N > 1) will be opened in one excitation at the same time, and thus N intensity images are at different delays recorded simultaneously, namely, one-excitation multiple-sampling mode:

g[(t).sub.t] = [(M - 1)summation over (j = 0)] [(N - 1)summation over (i = 0)] w (t + idt + j[DELTA]t + [delta]t). (4)

Main parameters in (4) are all fixed. Here, an embodiment is shown in Figure 4(b) where N = 2 and Gate A and Gate B are opened in one excitation with a specific delay (dt). RLD algorithm utilizing two intensity images was developed to realize high-speed lifetime image inversion. Lifetime image can be inverted before next excitation. Two images acquired simultaneously based on fibre bundles gave a method to realize this timing [20]. Therefore, lifetime can be computed in one single excitation since two synchronous images are acquired and processed in real time. Every CCD frame will be fully exploited for lifetime image acquired with the coordination of laser RF and CCD frame by this timing and operating mode.

4. Conclusion

To sum up, aiming at steady and transient fluorescence lifetime imaging, we present feasible operating modes and lifetime computing algorithms (Table 4). High resolution and high frame can be achieved by one-excitation one-sampling mode and least square algorithm for steady imaging applications. Correspondingly, one-excitation two-sampling mode and rapid lifetime determination algorithm contribute to transient fluorescence situations. The universal timing equation (1) can be used for another application after certain temporal parameters changed. The abstract model (Figure 1(b)) can provide a basis for time-gated FLIM system buildup.

http://dx.doi.org/ 10.1155/2013/801901

Received 9 March 2013; Accepted 2 June 2013

References

[1] M. Y. Berezin and S. Achilefu, "Fluorescence lifetime measurements and biological imaging," Chemical Reviews, vol. 110, no. 5, pp. 2641-2684, 2010.

[2] F. Pinaud, X. Michalet, L. A. Bentolila et al., "Advances in fluorescence imaging with quantum dot bio-probes," Biomaterials, vol. 27, no. 9, pp. 1679-1687, 2006.

[3] J. M. Klostranec and W. C. W. Chan, "Quantum dots in biological and biomedical research: recent progress and present challenges," Advanced Materials, vol. 18, no. 15, pp. 1953-1964, 2006.

[4] H. C. Gerritsen, R. Sanders, A. Draaijer, C. Ince, and Y. K. Levine, "Fluorescence lifetime imaging of oxygen in living cells," Journal of Fluorescence, vol. 7, no. 1, pp. 11-15,1997

[5] D. Sud and M. Mycek, "Calibration and validation of an optical sensor for intracellular oxygen measurements," Journal of Biomedical Optics, vol. 14, no. 2, Article ID 020506, 2009.

[6] C. Liu, Y. Zhou, X. Wang, and Y. Liu, "Lifetime computing algorithms based on exponential pattern retrieve and polynomial fitting in fluorescence lifetime imaging microscopy," in Proceedings of the SPIE International Conference on Optical Instruments and Technology: Optoelectronic Imaging and Processing Technology, vol. 8200, November 2011.

[7] J. Lakowicz, Principles of Fluorescence Spectroscopy, Springer, New York, NY, USA, 3rd edition.

[8] J. Widengren, "Fluorescence-based transient state monitoring for biomolecular spectroscopy and imaging," Journal of the Royal Society Interface, vol. 7, no. 49, pp. 1135-1144, 2010.

[9] K. Konig, "Multiphoton microscopy in life sciences," Journal of Microscopy, vol. 200, no. 2, pp. 83-104, 2000.

[10] J. A. Levitt, D. R. Matthews, S. M. Ameer-Beg, and K. Suhling, "Fluorescence lifetime and polarization-resolved imaging in cell biology," Current Opinion in Biotechnology, vol. 20, no. 1, pp. 28-36, 2009.

[11] C. W. Chang, D. Sud, and M. A. Mycek, "Fluorescence lifetime imaging microscopy," Methods in Cell Biology, vol. 81, pp. 495-524, 2007.

[12] J.-B. Yang, Y.-N. Cao, S.-T. Fan et al., "Research of hundred ps level timing control system in 3D range-gated imaging," Infrared and Laser Engineering, no. 7, 2012.

[13] P. Urayama, W. Zhong, J. A. Beamish et al., "A UV-visible-NIR fluorescence lifetime imaging microscope for laser-based biological sensing with picosecond resolution," Applied Physics B, vol. 76, no. 5, pp. 483-496, 2003.

[14] A. V. Agronskaia, L. Tertoolen, and H. C. Gerritsen, "Fast fluorescence lifetime imaging of calcium in living cells," Journal of Biomedical Optics, vol. 9, no. 6, pp. 1230-1237, 2004.

[15] "Becker & Hickl GmbH, Leica Microsystems Heidelberg GmbH, Picosecond Laser Scanning Microscopy Upgrade Your Leica TCS SP-2 for Lifetime Imaging," http://www.beckerhickl.de/.

[16] O. Holub, M. J. Seufferheld, C. Gohlke, G. Govindjee, and R. M. Clegg, "Fluorescence lifetime imaging (FLI) in real-time-a new technique in photosynthesis research," Photosynthetica, vol. 38, no. 4, pp. 581-599, 2000.

[17] C. Liu, Y. Zhou, X. W. Wang et al., "Fluorescence lifetime imaging microscopy and its research progress," Laser & Optoelectronics Progress, vol. 48, 2011.

[18] Y. Sun, N. Hatami, M. Yee et al., "Fluorescence lifetime imaging microscopy for brain tumor image-guided surgery," Journal of Biomedical Optics, vol. 15, no. 5, Article ID 056022, 2010.

[19] X. F. Wang, T. Uchida, D. Coleman et al., "A two-dimensional fluorescence lifetime imaging system using a gated image intensifier," Applied Spectroscopy, vol. 45, no. 3, pp. 360-366, 1991.

[20] C. Liu, Y. Zhou, X. W. Wang et al., "Fluorescence lifetime imaging device and method based on optical delay," CN201210184350.9.

[21] R. Cubeddu, D. Comelli, C. D'Andrea, P. Taroni, and G. Valentini, "Time-resolved fluorescence imaging in biology and medicine," Journal of Physics D, vol. 35, no. 9, pp. R61-R76,2002.

Chao Liu, Xinwei Wang, Yan Zhou, and Yuliang Liu

Optoelectronic System Laboratory, Institute of Semiconductors, CAS, Beijing 100083, China

Correspondence should be addressed to Chao Liu; lcphs820@gmail.com

Academic Editors: K. Wang and J. Zhao

Table 1: Parameters comparison of different FLIM methods.

                  Time-gated FLIM   TCSPC [15]   Frequency
                                                 modulation

Time resolution   50 ps [13]        1ps          --
Frame frequency   100 Hz [14]       10 Hz        1 Hz [16]

Table 2: Technical parameters of homemade TCU.

Repetitive            Minimum delay   Minimum pulse   Delay revise
frequency             step            width           precision

1 Hz-100 kHz          500ps           2 ns            150ps

Table 3: Comparison of exponential pattern model and polynomial model.

Average         Standard    Computing
lifetime (ns)   deviation   time (s) *

RLD 5.27        0.79        0.04
LS 4.30         1.32        36.8

* Intel Pentium T3200 @ 2 GHz, RAM 2 GB, Matlab R2007b, Math Works.

Table 4: Operating mode and lifetime computing algorithm for
different applications.

                  Operating             Lifetime
                  mode                  computing
                                        algorithm

Steady            One-excitation        LS
  fluorescence      one-sampling mode
Transient         One-excitation        RLD
  fluorescence      two-sampling mode
COPYRIGHT 2013 Hindawi Limited
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2013 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Research Article
Author:Liu, Chao; Wang, Xinwei; Zhou, Yan; Liu, Yuliang
Publication:The Scientific World Journal
Article Type:Report
Date:Jan 1, 2013
Words:2478