The facemask produces higher peak minute ventilation and respiratory rate measurements compared to the mouthpiece.

Dear Editor-in-Chief,

The mouthpiece is the gold-standard for measuring oxygen uptake (V[O.sub.2]) during incremental exercise tests, however this device alters breathing patterns during exercise (Amis et al., 1999) and often causes subjects to experience dry-mouth, dysphagia and throat irritation (Baran et al., 2001; Bart and Wolfel, 1994). This is a problem in clinical populations: discomfort due to the mouthpiece often causes incremental exercise tests to be stopped prematurely, limiting the ability to measure a true peak oxygen uptake (V[O.sub.2peak]). The Hans-Rudolph facemask is a comfortable alternative that permits a more normal breathing pattern and reduces irritation so that patients may exercise longer and a true or predicted V[O.sub.2peak] can be accurately assessed. As exercise intervention studies and exercise prescriptions become more common in clinical populations such as cancer patients and heart failure patients, the need for feasible and accurate exercise testing becomes more relevant. The facemask has been validated in clinical populations and elite athletes, however these populations typically exercise at very low or very high intensities, which may not necessarily represent the average healthy population. The facemask has not been validated in a group of healthy adults, nor has it been validated across a spectrum of exercise intensities so that it can be used across a broader range of studies. This study aimed to examine the validity and reliability of the facemask compared to the mouthpiece in healthy individuals at peak power, 75% peak power, and 25% peak power.

Twenty-eight healthy adults (15 males and 13 females, age: 25 [+ or -] 11 yrs [mean [+ or -] s], BMI: 23.67 [+ or -] 2.57 kg-m-2) consented to complete 3 incremental exercise tests on a cycle ergometer (Ergometrics er800s; Ergoline, Bitz, Germany). One test was completed using a mouth piece (Hans Rudolph T-Shape Series 2700 two-way non-rebreathing valve), 2 tests were completed using a facemask (Hans Rudolph 7400 Vmask Series Oro-Nasal MASK) and all tests were conducted using the Vmax breath-by-breath system (Vmax; SensorMedics, Yorba Linda, CA). Each test consisted of a 2-minute rest period, followed by a 2-minute warm-up at 25 W. Power was increased by 25-75 W every 2 minutes thereafter, depending on the participant's rating of perceived exertion, until participants reached exhaustion (i.e. cadence fell below 50 rpm for longer than 30 seconds). We compared V[O.sub.2], minute ventilation ([V.sub.E]) and a variety of other ventilatory parameters at peak power, 75% of peak power and 25% of peak power using 2-way repeated measures ANOVA and Bland-Altman plots (Bland and Altman, 1986). We assessed validity of the facemask by comparing the mouthpiece trials to the first facemask trials; we assessed reliability by comparing the 2 facemask trials. Further, we stratified participants into high or low cardiovascular fitness categories depending on whether their V[O.sub.2peak] greater or less than the group median and reassessed validity and reliability within each group. Statistical significance was accepted at p < 0.05.

[FIGURE 1 OMITTED]

Participants' [V.sub.Epeak] with the mouthpiece was 10% lower than their [V.sub.Epeak] with the facemask (79.4 [+ or -] 19.9 L x [min.sup.-1] vs. 88.1 [+ or -] 24.7 L x [min.sup.-1]; p < 0.001), as shown in Table 1. When participants were stratified by cardiovascular fitness level, VEpeak remained lower with the mouthpiece compared to the facemask for participants in the high cardiovascular fitness group (n = 14) ([V.sub.Epeak] 90.7 [+ or -] 19.8 L x [min.sup.-1] vs. 106.0 [+ or -] 18.5 L x [min.sup.-1]; p < 0.001). Although [V.sub.Epeak] was lower for the mouthpiece compared to the facemask, Bland-Altman analysis did not reveal any significant differences between apparatuses. As seen in Figure 1, ninety-five percent of the differences (i.e. 27 participants) fell between +/- 1.96 standard deviations of the bias. There were no significant differences between the 2 facemask trials for any of the ventilatory parameters at peak power, at 75% peak power or at 25% peak power.

The difference in [V.sub.E] may be attributed to the fact that symptoms of discomfort associated with the mouthpiece become exacerbated at high power when subjects are undergoing strenuous exercise. Subjects have reported actively attempting to lower their ventilation rate to alleviate this discomfort (Askanazi et al., 1980). Furthermore, since nasal breathing is occluded when using the mouthpiece and breathing is restricted to the oral pathway, it may be more difficult for participants to reach greater VE values (O'Kroy et al., 2001). The mouthpiece also prevents pursed-lips breathing, a natural response to strenuous exercise that relieves dyspnea. A study on chronic obstructive pulmonary disease (COPD) patients revealed that participants who were permitted to engage in pursed-lips breathing were able to exercise for a longer duration compared to those instructed to use a mouthpiece (Faager et al., 2008). We observed no difference in V[O.sub.2peak] between the facemask and the mouthpiece, despite the marked difference in [V.sub.Epeak]. This may be attributed to the fact that although [V.sub.E] and V[O.sub.2] are tightly correlated at low-moderate exercise intensities, at peak power larger increases in [V.sub.E] are required to stimulate a concurrent increase in V[O.sub.2] (Gastinger et al., 2010). It is possible that in subjects who are able to obtain a higher [V.sub.E] at peak power than was observed in this study, V[O.sub.2] differences might become significant.

In the present study we observed that the facemask allowed participants to attain consistently higher peak power during the incremental exercise test compared to the mouthpiece (data not presented). The differences in peak power between facemask and mouthpiece trials in the present study were not statistically significant but may have required participants to exchange gas at different rates, contributing to significant differences in [V.sub.E]. Taking these differences into consideration, we conclude that the facemask is an appropriate alternative for exercise testing in healthy populations.

References

Amis, T.C., O'Neill, N. and Wheatley, J.R. (1999) Oral airway flow dynamics in healthy humans. Journal of Physiology 515(Pt1), 293-298.

Askanazi, J., Silverberg, P.A., Foster, R.J., Hyman, A.I., Milic-Emili, J. and Kinney, J.M. (1980) Effects of respiratory apparatus on breathing pattern. Journal of Applied Physiology 48(4), 577-580.

Baran, D.A., Rosenwinkel, E., Spierer, D.K., Lisker, J., Whelan, J., Rosa, M. and Goldsmith, R.L. (2001) Validating Facemask use for gas exchange analysis in patients with congestive heart failure. Journal of Cardiopulmonary Rehabilitation 21, 94-100.

Bart, B.A. and Wolfel, E.E. (1994) Method of expired gas collection during cardiopulmonary exercise testing does not affect respiratory gas exchange measurements in patients with heart failure. Journal of Cardiac Failure 1(1), 91-96.

Bland, J.M. and Altman., D.G. (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1(8476), 307-310.

Faager, G., Stahle, A. and Larsen, F.F. (2008) Influence of spontaneous pursed lips breathing on walking endurance and oxygen saturation in patients with moderate to severe chronic obstructive pulmonary disease. Clinical Rehabilitation 22(8), 675-683.

Gastinger, S., Sorel, A., Nicolas, G., Gratas-Delamarche, A., & Prioux, J. (2010) A comparison between ventilation and heart rate as indicator of oxygen uptake during different intensities of exercise. Journal of Sports Science and Medicine 9, 110-118.

O'Kroy, J.A., James, T., Miller, J.M., Torok, D. and Campbell, K. (2001) Effects of an External nasal dilator on the work of breathing during exercise. Medicine and Science in Sports and Exercise 33(3), 454-458.

Kirsten Bell, Megan Bedbrook, Tri-Tue Nguyen and Marina Mourtzakis ([mail])

Department of Kinesiology, University of Waterloo, Waterloo, ON, Canada

([mail]) Dr. Marina Mourtzakis

Burt Matthews Hall, Rm 1117, Department of Kinesiology, University of Waterloo, 200 University Ave. W, Waterloo, ON, Canada N2L 3G1

E-mail: mmourtza@uwaterloo.ca
Table 1. Ventilatory parameters. Data are means ([+ or -]s).

                                                  Peak power

                                               M              F1

Power (W)                                 240 (75)        243 (77)
V[O.sub.2peak] (mL x [kg.sup.-1] x       42.0 (9.9)      42.3 (9.0)
  [min.sup.-1])
[V.sub.Epeak] (L x [min.sup.-1])         79.1 (20.2) *   87.1 (24.5)
[Vt.sub.peak] (L)                       2.712 (.559)    2.687 (.571)
[RR.sub.peak] (breaths x [min.sup.-1])     38 (7) *        41 (7)
Exercise time (min)                     10.32 (2.59)    10.58 (2.35)
HR (beats x [min.sup.-1])                 172 (16)        176 (16)
RER                                      1.24 (.13)      1.29 (.25)

75% peak power

Power (W)                                 169 (49)        175 (53)
V[O.sub.2] (mL x [kg.sup.-1] x           31.4 (7.0)      32.0 (7.5)
  [min.sup.-1])
[V.sub.E] (L x [min.sup.-1])             49.2 (11.5)     51.7 (13.8)
Vt (L)                                  2.333 (.505)    2.363 (.493)
RR ([breaths x [min.sup.-1])               27 (7)          28 (7)
Exercise time (min)                      7.83 (1.92)     8.01 (1.90)
HR (beats x [min.sup.-1])                 151 (16)        156 (15)
RER                                      1.10 (.10)      1.14 (.17)

25% peak power

Power (W)                                  63 (43)         67 (45)
V[O.sub.2] (mL x [kg.sup.-1]             15.6 (5.5)      15.5 (5.2)
  [min.sup.-1])
[V.sub.E] ([L x [min.sup.-1])            22.6 (8.7)      21.9 (8.5)
Vt (L)                                  1.416 (.549)    1.470 (.553)
RR (breaths x [min.sup.-1])                21 (5)          20 (4)
Exercise time (min)                      4.00 (2.12)     3.95 (2.14)
HR (beats x [min.sup.-1])                 107 (12)        109 (14)
RER                                       .91 (.09)       .89 (.09)

                                          Peak power

                                              F2

Power (W)                                  253 (82)
V[O.sub.2peak] (mL x [kg.sup.-1] x        42.8 (9.3)
  [min.sup.-1])
[V.sub.Epeak] (L x [min.sup.-1])          87.2 (25.0)
[Vt.sub.peak] (L)                        2.656 (.570)
[RR.sub.peak] (breaths x [min.sup.-1])      41 (8)
Exercise time (min)                      10.74 (2.52)
HR (beats x [min.sup.-1])                  175 (16)
RER                                       1.26 (.12)

75% peak power

Power (W)                                  173 (53)
V[O.sub.2] (mL x [kg.sup.-1] x            31.9 (7.4)
  [min.sup.-1])
[V.sub.E] (L x [min.sup.-1])              51.5 (14.7)
Vt (L)                                   2.295 (.506)
RR ([breaths x [min.sup.-1])                29 (7)
Exercise time (min)                       8.06 (2.06)
HR (beats x [min.sup.-1])                  155 (18)
RER                                       1.11 (.09)

25% peak power

Power (W)                                   66 (44)
V[O.sub.2] (mL x [kg.sup.-1]              15.5 (5.0)
  [min.sup.-1])
[V.sub.E] ([L x [min.sup.-1])             22.3 (7.9)
Vt (L)                                   1.468 (.504)
RR (breaths x [min.sup.-1])                 20 (4)
Exercise time (min)                       4.17 (2.23)
HR (beats x [min.sup.-1])                  111 (16)
RER                                        .91 (.08)

M: mouthpiece; F1: first facemask trial; F2: second facemask trial;
Vt: tidal volume; HR: heart rate; RER: respiratory exchange ratio. *
denotes a significant difference from F1 and F2.
COPYRIGHT 2012 Journal of Sports Science and Medicine
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2012 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Author:Bell, Kirsten; Bedbrook, Megan; Nguyen, Tri-Tue; Mourtzakis, Marina
Publication:Journal of Sports Science and Medicine
Article Type:Report
Geographic Code:1CANA
Date:Sep 1, 2012
Words:1830
Previous Article:Longitudinal study of repeated sprint performance in youth soccer players of contrasting skeletal maturity status.
Next Article:Do toning pants help to give you a better workout?
Topics:

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters