Terbium and rhodamine as labels in a homogeneous time-resolved fluorometric energy transfer assay of the [beta] subunit of human chorionic gonadotropin in serum.

The measurement of peptide hormones, such as the [beta] subunit of human chorionic gonadotropin ([beta]hCG), [1] in biological samples requires a highly sensitive assay technology. Analyses routinely are performed with immunometric assays in which the combination of two specific antibodies provides the specificity and a sensitive label, such as an enzyme, radionuclide, or a fluorescent compound, allows detection.

Immunometric assays usually are heterogeneous assays in which the high excess of reagents ensures rapid bindings, efficient washing decreases nonspecific signal, and a high-specific activity label provides the sensitivity and wide dynamic range (1). These performance characteristics have made immunometric assays the methods of choice in routine diagnostics of peptide hormones, cancer antigens, low concentrations of specific antibodies, and other clinically relevant indicators (2). In some cases, however, the required separation is too laborious to perform (e.g., from the point of view of automation) or for other reasons undesired (e.g., need for kinetic data or low affinity binding). The screening of thousands or millions of samples in a high-throughput format requires a simple and preferably a non-separation assay design (3).

Several homogeneous assay principles have been published. Only a few of them, however, are applicable to macromolecules (4,5). In diagnostics, homogeneous assays are used primarily for therapeutic drug monitoring and other low-molecular weight substances, using methods such as fluorescence polarization (6, 7). Similarly to most of the homogeneous methods, fluorescence polarization assay has inherent limitations, both in terms of applicability and sensitivity. Fluorescence resonance energy transfer (FRET) is a widely used technology in biomedical studies (8) and has also been applied to the field of diagnostic immunoassays (9).

Time-resolved (TR) fluorometry adds another dimension to homogeneous assays. Because all matrix-related components are present during the fluorometric measurement, it is only through temporal resolution that the background interference can be eliminated (10). In FRET, the temporal resolution is important both in avoiding background problems and in enabling more specific detection of energy transfer-excited acceptor emission (11). Because the excited state lifetime of the donor probe is longer than that of the acceptor probe or the background, the energy transfer can be distinguished from the emission emanating from direct excitation of acceptor molecules by the use of a sufficiently long delay time before gating the detection. The acceptor emission at a longer wavelength can be separated from donor emission by spectral resolution (filters).

Lanthanide chelate labels exhibiting decay times from a few microseconds to several milliseconds are perfectly suited for TR-FRET technologies (12). Recently, this technology has been applied to various binding assays in the field of diagnostics and research (13-16). In addition to the Eu-cryptate applied by Mathis (13), several fluorescent lanthanide chelates have been developed (17,18). Donor-acceptor pairs, such as [Eu.sup.3+] with CY-5 (15, 19), and [Tb.sup.3+] with rhodamine (20), have been published. Numerous chelates have been synthesized and used in different types of assays, from heterogeneous and homogeneous immunoassays to microscopic imaging (21). Highly fluorescent chelates of [Eu.sup.3+] and [Tb.sup.3+] are particularly well suited for TR-FRET assays. Optimization of label pairs for TR-FRET generally requires an optimally fluorescent donor chelate, i.e., a chelate with a high quantum yield, efficient light harvesting, suitable coupling system with retained biocompatibility, and a suitable emission distribution (22).

We studied the use of a fluorescent [Tb.sup.3+] chelate in a homogeneous sandwich-type immunometric assay of [beta]hCG. The assay is based on two specific monoclonal antibodies against different epitopes on [beta]hCG. One antibody was labeled with an energy-donating [Tb.sup.3+] chelate and the other with an acceptor, tetramethylrhodamine (Alexa [TM] 546), a sulfonated rhodamine derivative. The resulting homogeneous TR-FRET assay was validated by comparison with a heterogeneous assay.

Materials and Methods

REAGENTS AND BUFFERS

The [Tb.sup.3+] chelate, an isothiocyanate-activated highly fluorescent and stable [Tb.sup.3+] chelate (W14016) was obtained from Wallac Labeling Service (Wallac Oy, Turku, Finland). The anti-[beta]hCG antibodies, clones 2764 and 3341, DELFIA [R] enhancement solution and DELFIA enhancer were from Wallac Oy. The [beta]hCG calibrators were from the DELFIA Free hCG[beta] Immunoassay Kit (Wallac). Isothiocyanate-activated tetramethylrhodamine (TMR) and the succinimidyl ester of Alexa546, a sulfonated rhodamine derivative, were purchased from Molecular Probes Inc. The assay buffer consisted of 50 mmol/L Tris, pH 7.4, with 9 g/L NaCl and 1 g/L bovine serum albumin.

EQUIPMENT

Absorption spectra were recorded with a Pharmacia Biochrom 4060 spectrophotometer. Fluorescence excitation and emission spectra were recorded with a Perkin-Elmer LS 5 fluorescence spectrometer. TR measurements of [Tb.sup.3+] concentrations and energy transfer emissions were performed with a 1420 Victor multilabel counter (Wallac). The interference from quenching compounds in the samples, the [Tb.sup.3+] signal at the acceptor wavelength, and the background of plastics material were corrected for by a prototype quench correction algorithm (QCA) from Wallac.

ANTIBODY LABELING

Anti-[beta]hCG antibody 2764 was labeled with the [Tb.sup.3+] chelate. The antibody at a concentration of 5 g/L was incubated with a 30-fold molar excess of the chelate in 50 mmol/L carbonate buffer, pH 9.5, overnight at 4 [degrees]C. Labeled antibody was separated from unreacted chelates by gel filtration (Sepharose 6B with Sephadex G50 overlay, 0.5 X 70 cm; Pharmacia) with a buffer containing 50 mmol/L Tris-HCI, pH 8, containing 9 g/L NaCl as elution buffer. When the labeling degree ([Tb.sup.3+]/IgG) was determined by measurement of the [Tb.sup.3+] concentration of conjugated antibody against a [Tb.sup.3+] calibrator (Aldrich Chemie) with a DELFIA system (23), a conjugation degree of 6 was achieved. Anti-[beta]hCG antibody 3341 was coupled with TMR or Alexa546 using a similar technology with a 15-fold molar excess of the fluorochromes. The labeling degrees, determined photometrically according to the reagent supplier's instructions, were 2.2 (TMR/IgG) and 2.8 (Alexa/IgG). Antibodies with these labeling degrees were used in the assay if not stated otherwise. Other labeling degrees were obtained by varying the amount of [Tb.sup.3+] or rhodamine excess in the described protocols. Labeled antibodies were found to be stable for several months when stored in Tris buffer, pH 7.4, containing 1 g/L bovine serum albumin.

ASSAY PRINCIPLE

In the immunocomplex, Tb-antibody-[beta]hCG-rhodamine-antibody, the distance between [Tb.sup.3+] and rhodamine is short enough to allow excitation energy to be transferred from [Tb.sup.3+] to rhodamine through Forster-type FRET (24). The decay time of the energy transfer-excited rhodamine emission is directly related to the decay time of the donor chelate and inversely related to the energy transfer efficiency. Compared with the natural decay time of rhodamine (3.0 ns), the energy transfer signal has a greatly prolonged decay time, enabling its measurement with TR fluorometry, thus avoiding the interference derived from background and the emission of rhodamine excited by pulsed ultraviolet light. The interference from the long-lived emission of unbound [Tb.sup.3+] donor can be avoided by spectral resolution using a high-quality interference filter at a wavelength where [Tb.sup.3+] background is minimal (570 nm; Fig. 1). The measured energy transfer signal is directly proportional to the concentration of [beta]hCG in the sample.

ASSAY PROTOCOL

In the assay, 50 [micro]L of [beta]hCG calibrator or serum sample in the wells of black 96-well microtitration plates (Nunc) were incubated with a mixture of 100 ng of [Tb.sup.3+]-labeled antibody (donor) and 100 ng of rhodamine-labeled antibody (acceptor) in 200 [micro]L of assay buffer. Reagent mixtures were incubated either on a plate shaker for 60 min at room temperature or on the laboratory bench after 2 min of rapid shaking. Energy transfer and [Tb.sup.3+] emissions were measured using excitation at 340 mn, emission filters at 570 nm for rhodamine and 545 for [Tb.sup.3+], a delay time of 50 [micro]s, a window time of 100 [micro]s, and a cycling time of 1 ms. Interfering processes in the assay were corrected with QCA, using wells containing buffer only, wells containing [Tb.sup.3+]-labeled antibody only, and wells containing the highest concentration of calibrator (200 [micro]g/L) as external standards (25, 26). QCA subtracts the donor interference seen at the emission wavelength of the acceptor and the background from plastics from the acceptor signal before calculating the acceptor-to-donor ratio. The ratio, which corrects for absorption of excitation energy by the serum sample and competing energy transfer processes (quenching), is then multiplied with the donor signal measured from the highest calibrator. This blank-corrected, normalized ratio was used as response parameter in the assay.

Results

SPECTRA

The emission spectra of the donor and acceptor, and the absorbance spectrum for the acceptor are depicted in Fig. 1. For energy transfer to occur, the emission spectrum of the donor and the absorbance spectrum of the acceptor need to overlap.

ASSAY CONDITIONS

Antibodies of different conjugation degrees (label/IgG) were tested in the assay to find the degrees that produced the highest signal-to-noise ratios (Table 1). The fluorescence signal for the calibrator containing 200 [micro]g/L [beta]hCG divided by the signal for the zero calibrator was used to calculate the signal-to-noise ratios. The fluorescence of calibrators ranging from 0 to 200 [micro]g/L [beta]hCG with signals from the acceptor and uncomplexed donor as well as QCA-corrected signals are presented in Tables 2 and 3. The energy transfer efficiency in the assay remained below 15%, indicating a relatively large distance between the labels in the formed complexes and an excess of [Tb.sup.3+]-labeled antibody in the assay. The decay profiles of [Tb.sup.3+] (donor) and the acceptor after energy transfer are shown in Fig. 2. The donor signal shows a clear decay of unquenched [Tb.sup.3+] (1087 pts), which to a great extent results from uncomplexed antibodies and from the fraction of [Tb.sup.3+] chelates at a more distant location from acceptors. The decay of the energy transfer (180-390 pts), measured at 570 mn, is clearly shorter and has a multicomponent nature that can be attributed to the different donor-acceptor distances in complexes formed of randomly labeled antibodies. The decay time, however, is long enough for efficient TR detection.

[FIGURE 1 OMITTED]

PERFORMANCE CHARACTERISTICS

Assay kinetics were investigated by incubating plates for up to 180 min. The binding kinetics obtained in the assay with constant shaking or no shaking are presented in Fig. 3. The reaction had almost reached equilibrium after 60 min of incubation. The detection limit, i.e., precision of the measurement of the zero-dose calibrator, was 0.95 [micro]g/L for TMR and 0.43 [micro]g/L for Alexa546, defined as the value 2 SD above the mean of the zero-calibrator values (n = 24). According to the kit insert, the detection limit of the DELFIA Free hCG[beta] assay is better than 0.2 [micro]g/L. Five serum samples were assayed four times in three different assay runs to estimate the imprecision of the [beta]hCG assay with TMR as acceptor. The within-run, between-run, and total imprecision calculated with ANOVA (27) are presented in Table 4. Sixty-six serum samples relatively evenly distributed over the measurement range were assayed, and the concentrations were compared with those obtained with the DELFIA assay. The developed TR-FRET method correlated well with the DELFIA Free hCG[beta] assay (Fig. 4).

Discussion

Assay miniaturization and automation are major trends in diagnostics today. Homogeneous assays provide a key to adapting assays for automation because complicated coating procedures and slow surface reactions can be avoided. When developing simpler assays in a miniaturized format, however, it is important to maintain high assay sensitivity. Most homogeneous assays are compromised in terms of sensitivity and dynamic range, partly because of matrix-related problems (sample autofluorescence) during endpoint detection. In this context, time resolution is an enabling technology because it allows the measurement of a specific response without background interference (10). With fluorescent lanthanide chelates as the long-lifetime labels, sensitive homogeneous assays have also been developed and applied to serum thyroxine (28) and steroid glucuronides in urine (29). A majority of the homogeneous assay techniques are for small molecules and are less suitable for assays of macromolecules. Nonseparation and homogeneous immunoassay techniques and their applications and limitations have been discussed in detail elsewhere (4).

FRET is a homogeneous assay technology that, unlike most others, is well suited for the study of the binding of large components, such as proteins in various kinds of protein-protein interaction assays (9). It is not primarily the size of the proteins, but the distance between donor and acceptor that is critical in FRET assays. The shorter the distance, the more efficient the energy transfer. Therefore, it is beneficial to label proteins in a way that produces a short distance between the donor and acceptor molecules in the formed complexes and to choose a donor-acceptor pair with a high [R.sub.0] value (the distance at which the energy transfer efficiency is 50%). Probes with high [R.sub.0] values allow relative long between-probe distances in the assay. Addition of temporal resolution and long-lifetime donor probes can greatly improve the sensitivity and specificity of FRET (8). In TR-FRET, the specific energy transfer in the complex can be distinguished from donor emission by spectral filtration and from the direct emission of the acceptor by temporal filtration. Recently, TR-FRET has been exploited commercially both in diagnostics (13) and as a screening tool in drug discovery (14,16).

[FIGURE 2 OMITTED]

[FIGURE 3 OMITTED]

Regardless of temporal resolution, most of the TRFRET assays described above have limitations in both sensitivity and in dynamic range when compared with heterogeneous TR-fluoroimmunoassays. For assays requiring high sensitivity and an excess of donor-labeled reagents, the donor signal is a limiting factor. The working range is a direct function of the distribution of donor emission, i.e., to what degree the donor emits long-lived background at the emission wavelength of the acceptor. One way to alleviate the problem is to use low-quantum yield chelates (30), which hence also give less background. We looked for optimal donors amongst chelates having the following major properties: (a) fluorescence intensity as high as possible (the product of absorptivity X quantum yield); (b) fully coordinated nonadentate chelating structures with no solvent perturbation; (c) very long lifetimes (1-2 ms); and (d) optimal emission distribution, i.e., chelates exhibiting major emission at donating wavelength and minimal emission at the wavelength used to measure energy transfer.

[FIGURE 4 OMITTED]

The choice between [Eu.sup.3+] and [Tb.sup.3+] depends on the availability of high-quantum yield chelates and respective acceptors. Highly sensitive [Tb.sup.3+] chelates have been discovered recently, and today, stable fluorescent [Tb.sup.3+] chelates exhibiting quantum yields >40% are available (17, 18). Compared with [Eu.sup.3+], which emits at 613-620 nm, efficient acceptor probes are easier to find at the excitation range of 545 nm. Use of [Eu.sup.3+] requires probes emitting in the near-infrared range, which are less common (31). [Tb.sup.3+] and rhodamine have been used in several energy transfer experiments (20, 32) because of the relatively high [R.sub.0] values (6-7 nm). Thus, a [Tb.sup.3+] chelate exhibiting a high fluorescence intensity and the low-molecular weight acceptor rhodamine were used as labels in the presented sandwich assay for [beta]hCG.

In the model assay, two monoclonal antibodies recognizing different epitopes on [beta]hCG were labeled with [Tb.sup.3+] and rhodamine. The antibodies were not optimized to give a short distance between the donor and acceptor, but merely to give a good binding performance. Both antibodies were randomly labeled with 1-10 label molecules/IgG. For TMR labeling, the relative insolubility of TMR and its negative effect on antibody functionality and affinity restrict optimal labeling. However, according to the experiments, on the average, 4-8 [Tb.sup.3+] and ~2 T MR per IgG seemed to give the best response. More Alexa546 than TMR can be conjugated to proteins, probably because of lower self-quenching effects and better solubility. Alexa546 has also been reported to be more fluorescent than TMR and other spectrally compatible fluorochromes (33). These observations were confirmed in this study. Alexa546 was threefold more fluorescent than TMR in the TR-FRET assay. The degrees of labeling producing the highest response were 2-5 Alexa546 and 2-8 [Tb.sup.3+] per IgG. In the actual assay, the labels closest to each other in the complex probably give the largest response, and labels in distant positions only contribute to the background

The sandwich assay for [beta]hCG can be performed using either a direct or indirect labeling approach. The indirect way, i.e., using biotinylated antibodies and labeled streptavidin, is a more generic approach, but tends to give decreased sensitivity because of longer between-probe distances (data not shown). In FRET assays, the fluorescence decay time of the energy transfer-excited acceptor is always shorter than that of the donor. Therefore, a short delay time and counting window relative to that optimal for the measurement of free donor gives an improved signal-to-noise ratio. The shortest possible delay time depends on the phosphorescence of the plastics material used and its decay unless nonfluorescent (black) plates are applied.

The sensitivity of the developed [beta]hCG assay compared well with the commercially available [Sm.sup.3+]-based DELFIA [beta]hCG assay. Bearing in mind that the [Sm.sup.3+] label is 10-100 times less sensitive than [Eu.sup.3+] label in DELFIA assays, it can be concluded that the homogeneous assay format is one to two orders of magnitude less sensitive than a well-optimized, [Eu.sup.3+]-based, heterogeneous DELFIA assay (34). The lower sensitivities of homogeneous assays can be attributed to the differences in quantum yields, the differences in labeling degrees, the relatively high backgrounds in TR-FRET, and to the way homogeneous assays must be optimized. Highly sensitive heterogeneous assays generally apply a strategy of reagent excess (1) so that the high excess of labeled antibodies guarantees rapid equilibrium, and efficient washing on the other hand ensures low background. Because of the constant background produced by [Tb.sup.3+]-labeled antibodies and the diffusion background seen at high donor-to-acceptor concentrations, in homogeneous assays there is much less freedom in choosing the concentrations of the labeled components. It is possible to adjust the distance between labels, at least to some extent, with reagents that permit site-specific labeling. Labels in close proximity ensure an efficient energy transfer, and some of the background (interference) can be reduced because donors at distant positions, contributing only to the background, can be avoided. Heterogeneous assays frequently use antibody-coated surfaces such as beads and other solid phases to obtain convenient separation of bound and free fractions. Antibody-coated solid phases require efficient shaking or stirring of the assay mixture to speed up assay kinetics. In homogeneous assays, shaking or stirring of the assay mixture is not necessary because shaking has no effect on the reaction velocity in a homogeneous solution. The omission of solid phases and shaking means fewer assay steps, which is advantageous in automation.

Regardless of the compromises one must make when transferring a heterogeneous assay into a homogeneous format, TR-FRET has proven potential, even in two-site assay design. The assay of [beta]hCG has a good sensitivity, and the assay correlates perfectly with respective heterogeneous assays. Homogeneous TR-FRET assays provide a tool for assay simplification and enable low affinity binding assays. They are easy to perform, rapid, involve fewer steps, and are well suited for systems requiring easy automation, high throughput, or miniaturization.

Received December 22, 1998; accepted March 5, 1999.

References

(1.) Ekins R, Jackson T. Non-isotopic immunoassay--an overview. In: Bizallon CA, ed. Monoclonal antibodies and new trends in immunoassays. Amsterdam: Elsevier Science, 1984:149-63.

(2.) Diamandis EP. Immunoassays. London: Academic Press, 1996: 579pp.

(3.) Devlin JP. High throughput screening. New York: Marcel Dekker, 1997:673pp.

(4.) Hemmila IA. Applications of fluorescence in immunoassays. New York: Wiley Interscience, 1991:343pp.

(5.) Christopoulos TK, Diamandis EP. Fluorescence immunoassays. In: Diamandis EP, Christopoulos TK, eds. Immunoassay. London: Academic Press, 1996:309-36.

(6.) Dandliker WP, Feigen GA. Quantification of the antigen-antibody reaction by the polarization of fluorescence. Biochem Biophys Res Commun 1961;5:299-304.

(7.) Jolley MEJ. Fluorescence polarization immunoassay for the determination of therapeutic drug levels in human plasma. Anal Toxicol 1981;5:236-40.

(8.) Morrison LE. Detection of energy transfer and fluorescence quenching. In: Kricka U, ed. Nonisotopic probing blotting and sequencing. San Diego, CA: Academic Press, 1995:429-70.

(9.) Ullman EF, Schwarzberg M, Rubenstein KEJ. Fluorescent excitation transfer immunoassay, a general method for determination of antigens. J Biol Chem 1976;251:4172-8.

(10.) Soini E, Hemmila I. Fluoroimmunoassays: present status and key problems. Clin Chem 1979;25:353-61.

(11.) Morrison L. Time-resolved detection of energy transfer: theory and application to immunoassays. Anal Biochem 1988;174:101-20.

(12.) Hemmila I, Webb S. Time-resolved fluorometry: an overview of the labels and core technologies for drug screening applications. Drug Discov Today 1997;2:373-81.

(13.) Mathis G. Probing molecular interactions with homogeneous techniques based on rare earth cryptates and fluorescence energy transfer. Clin Chem 1995;41:1391-7.

(14.) Kolb JM, Yamanaka G, Manly SP. Use of a novel homogeneous fluorescent technology in high throughput screening. J Biomol Screen 1996;1:203-10.

(15.) Selvin PR, Rana TM, Hearst JE. Luminescence resonance energy transfer. J Am Chem Soc 1994;116:6029-30.

(16.) Stenroos K, Hurskainen P, Eriksson S, Hemmila I, Blomberg K, Lindqvist C. Homogeneous time-resolved IL-2-IL-2 Ra assay using fluorescence resonance energy transfer. Cytokine 1998;10:495-9.

(17.) Hemmila I, Mukkala V-M, Takalo H. Development of luminescent lanthanide chelate labels for diagnostic assays. J Alloys Compd 1997;249:158-62.

(18.) Latva M, Takalo H, Mukkala V-M, Matachescu C, Rodriguez-Ubis JC, Kankare J. Correlation between the lowest triplet state energy level of the ligand and lanthanide(III) luminescence quantum yields. J Lumin 1997;75:149-69.

(19.) Li M, Selvin PR. Luminescent polyaminocarboxylate chelates of terbium and europium: the effect of chelate structure. J Am Chem Soc 1995;117:8132-8.

(20.) Selvin PR, Hearst JE. Luminescence energy transfer using a terbium chelate: improvements on fluorescence energy transfer. Proc Natl Acad Sci U S A 1994;91:10024-8.

(21.) Hemmila I, StahlbergT, Mottram P, eds. Bioanalytical applications of labelling technologies, 2nd ed. Turku, Finland: Wallac Oy, 1995:394pp.

(22.) Hemmila I. Time-resolved fluorometry: advantages and potentials. In: Devlin JP, ed. High throughput screening. New York: Marcel Dekker, 1997:361-76.

(23.) Hemmila I, Mukkala V-M, Latva M, Kiilholma P. Di- and tetracarboxylate derivatives of pyridines, bipyridines and terpyridines as luminogenic reagents for time-resolved fluorometric determination of terbium and dysprosium. J Biochem Biophys Methods 1993; 26:283-90.

(24.) Forster T. Zwischen molekulare energiewanderung and fluoreszenz. Ann Physik 1948;2:55-75.

(25.) Schiller PW. The measurement of intramolecular distances by energy transfer. In: Chen FC, ed. Biochemical fluorescence: concepts, Vol. 1. New York: Marcel Decker, 1975:285-303.

(26.) Lim CS, Miller JN, Bridges JW. Energy-transfer immunoassay: a study of the experimental parameters in an assay for human serum albumin. Anal Biochem 1980;108:176-84.

(27.) National Committee for Clinical Laboratory Standards. User evaluation of precision performance and of clinical chemistry devices. Tentative guideline EP5-T. Villanova, PA: NCCLS, 1984.

(28.) Hemmila I, Malminen 0, Mikola H, Lovgren T. Homogeneous time-resolved fluoroimmunoassay of thyroxin in serum. Clin Chem 1988;34:2320-2.

(29.) Barnard J, Kohen F, Mikola H, Lovgren T. Measurement of estrone-3-glucuronide in urine by rapid, homogeneous time-resolved fluoroimmunoassay. Clin Chem 1989;35:555-9.

(30.) Aspe D, Dumont C, Foyentin M, Jolu EJP, Mathis G, Nuti D. Homogeneous luminescence immunoassays--based on energy transfer from donor label with low quantum yield to acceptor label. Patent PCT WO 92-01,225, 1992.

(31.) Miller JN. Long wavelength fluorescence spectroscopy. Spectroscopy 1993;5:4-8.

(32.) Thomas DD, Carlsen WF, Stryer L. Fluorescence energy transfer in the rapid diffusion limit. Proc Natl Acad Sci U S A 1978;75:5746-50.

(33.) Landon TM, ed. BioProbes 26. Eugene, OR: Molecular Probes, 1997.

(34.) Hemmila I. Photoluminescence immunoassays. In: Johnstone AP, Turner MW, eds. Immunochemistry 1: a practical approach. Oxford: IRL Press, 1997:193-214.

KAJ BLOMBERG, * PERTTI HURSKAINEN, and ILKKA HEMMILA

Wallac Oy, P.O. Box 10, FIN-20101 Turku, Finland.

* Author for correspondence. Fax 358-2-2678357; e-mail kaj.blomberg@wallac.fi.

[1] Nonstandard abbreviations: [beta]hCG, /3 subunit of human chorionic gonadotropin; FRET, fluorescence resonance energy transfer; TR, time-resolved; TMR, tetramethylrhodamine; and QCA, quench correction algorithm.
Table 1. Effect of the labeling degree (label/IgG) on the
signal-to-noise ratio.

 [Tb.sup.3+]/IgG

 2 4 8 10.8
TMR/IgG
 1.8 5.4 5.3 5.3 2.3
 2.2 5.3 5.8 5.8 2.5
 3.2 1.5 4.2 3.7 1.7
 4.5 1.0 1.6 1.4 1.0
Alexa546/IgG
 2.8 13.0 14.0 14.3 3.4
 4.8 12.3 12.8 14.0 3.4
 8.8 8.0 8.4 9.0 2.2

Table 2. Calibrators run with the TR-FRET [beta] hCG assay with
TMR as acceptor molecule.

 Fluorescence, counts

[beta] hCG, TMR (a) [Tb.sup.3+], (b) Normalized ratio (c)
[micro] g/L

 0 2917 378 223 0
 2 3099 377 293 179
 5 3479 378 068 532
 20 5087 379 573 2032
104 10 995 365 567 7976
202 16 035 356 943 13 270

(a) Measured fluorescence from energy transfer-excited TMR in
formed immunocomplexes.

(b) Fluorescence from [Tb.sup.3+] on unbound antibodies and
[Tb.sup.3+] on bound antibodies not donating energy.

(c) QCA-corrected signal of energy transfer-excited TMR.

Table 3. Calibrators run with the TR-FRET [beta] hCG assay with
Alexa546 as acceptor molecule.

 Fluorescence, counts
[beta] hCG,
mg/L Alexa546 (a) [Tb.sup.3+] (b) Normalized ratio (c)

 0 2473 347 870 0
 2 3228 347 646 665
 5 4323 347 905 1603
 20 9563 348 856 6620
104 29 480 337 405 26 305
202 45 274 328 431 42 877

(a) Measured fluorescence from energy transfer-excited TMR in formed
immunocomplexes.

(b) Fluorescence from [Tb.sup.3+] on unbound antibodies and
[Tb.sup.3+] on bound antibodies not donating energy.

(c) QCA-corrected signal of energy transfer-excited Alexa546.

Table 4. Within-run, between-run, and total assay
imprecision for [beta] hCG.

 CV, %

Sample Mean, [micro] g/L Within-run Between-run Total

1 10.7 5.0 5.6 7.5
2 25.4 4.9 3.2 5.9
3 51.9 1.9 3.2 3.8
4 103.7 2.3 2.0 3.0
5 185.4 2.4 1.4 2.8
COPYRIGHT 1999 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 1999 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Automation and Analytical Techniques
Author:Blomberg, Kaj; Hurskainen, Pertti; Hemmila, Ilkka
Publication:Clinical Chemistry
Date:Jun 1, 1999
Words:4428
Previous Article:Improved immunoradiometric assay for plasma renin.
Next Article:Quantification of riboflavin, flavin mononucleotide, and flavin adenine dinucleotide in human plasma by capillary electrophoresis and laser-induced...
Topics:

Terms of use | Privacy policy | Copyright © 2021 Farlex, Inc. | Feedback | For webmasters |