Systematic evaluation of sanger validation of next-generation sequencing variants.

BACKGROUND: Next-generation sequencing (NGS) data are used for both clinical care and clinical research. DNA sequence variants identified using NGS are often returned to patients/participants as part of clinical or research protocols. The current standard of care is to validate NGS variants using Sanger sequencing, which is costly and time-consuming.

METHODS: We performed a large-scale, systematic evaluation of Sanger-based validation of NGS variants using data from the ClinSeq[R] project. We first used NGS data from 19 genes in 5 participants, comparing them to high-throughput Sanger sequencing results on the same samples, and found no discrepancies among 234 NGS variants. We then compared NGS variants in 5 genes from 684 participants against data from Sanger sequencing.

RESULTS: Of over 5800 NGS-derived variants, 19 were not validated by Sanger data. Using newly designed sequencing primers, Sanger sequencing confirmed 17 of the NGS variants, and the remaining 2 variants had low quality scores from exome sequencing. Overall, we measured a validation rate of 99.965% for NGS variants using Sanger sequencing, which was higher than many existing medical tests that do not necessitate orthogonal validation.

CONCLUSIONS: A single round of Sanger sequencing is more likely to incorrectly refute a true-positive variant from NGS than to correctly identify a false-positive variant from NGS. Validation of NGS-derived variants using Sanger sequencing has limited utility, and best practice standards should not include routine orthogonal Sanger validation of NGS variants.

Massively parallel sequencing technologies have revolutionized medical genetics. More than 200000 genomes and even more exomes have been sequenced to date (1). However, it is still widely accepted that variants found using next-generation sequencing (NGS)3 should be validated with the current gold standard for DNA sequencing, Sanger dideoxy terminator sequencing (2), before returning or publishing results. There have been several reports suggesting that NGS data used in clinical and research settings are at least as accurate--or in some cases more accurate--than Sanger sequencing (3-6). However, several of these studies used small sample sets (168, 37, and 110 variants sequenced, respectively) of secondary data from applied clinical research. Another recent, larger-scale study included 2 separate comparisons of NGS variants with Sanger sequencing results (443 variants and 762 variants, respectively), but this study was performed using target-capture gene panels, which are not representative of the overall genomic landscape, and the authors did not specify if any variants were included in both comparisons (6).

In the current study, we set out to determine the utility Of Sanger validation using a subset of data from 684 exomes and 2793 321 Sanger sequencing reads from the ClinSeq[R] cohort (7). The ClinSeq project was initiated in 2006, well before NGS was widely available, and began with semi-automated high-throughput Sanger sequencing. By the time that millions of Sanger reads had been generated for the ClinSeq project, NGS displaced Sanger sequencing as a less expensive, higher-throughput tool, which was then applied to the same samples that had already been Sanger sequenced, thus making this data set ideal for evaluating the utility of orthogonal Sanger validation of NGS variants.

Materials and Methods

DNA ISOLATION

DNA was isolated from whole blood using the salting-out method (Qiagen), followed by phenol--chloroform extraction using a Manual Phase Lock Gel extraction kit (5Prime) and rehydration with DNA hydration solution (Qiagen).

NEXT-GENERATION SEQUENCING

Solution-hybridization exome capture was performed with the SureSelect All Exon System, the SureSelect ICGC System (Agilent Technologies), or theTruSeq system, V1 or V2 (Ilumina). Flow-cell preparation and paired-end read sequencing were performed with either the GAIIx or HiSeq 2000 sequencer (Illumina) as previously described (8). Image analyses and base calling were performed as described (9). Reads were aligned to hg19 (NCBI build 37) using NovoAlign (Novocraft Technologies). For exome sequencing, samples were sequenced to sufficient coverage such that 85% of all targeted bases were called with a minimum Most Probable Genotype (MPG) score of at least 10 (10). The MPG genotype caller uses a Bayesian model that calculates the posterior probability of all possible genotypes at a position and reports the most likely genotype with a corresponding score calculated as the natural log of the quotient of the relative probability of the most likely genotype to the next most likely genotype (10). This means that an MPG score of 10 estimates the probability that the next most likely genotype is correct at [e.sup.-10] or 4.54 X [10.sup.-5]. The MPG score is dependent upon both the high-quality sequencing read depth and the zygosity at that base and is correlated linearly to the overall sequencing read depth (see Fig. 1 in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/ content/vol62/issue4). When a base was covered by more than 200 reads, MPG was applied to a random subset of 200 reads. We evaluated a given variant position in 684 exomes if there was at least 1 variant "sentinel" call at that position with an MPG score of [greater than or equal to] 10. If at least 1 sample met that threshold, then all calls at that position were considered, irrespective of their MPG score, so long as there were at minimum 10 reads covering that position. Structural variants and indels were excluded from analysis.

SANGER SEQUENCING

Candidate genes for Sanger sequencing were selected on the basis of evidence for association with development of coronary artery calcification and/or atherosclerosis (7).

That list was expanded to include genes associated with heart disease identified through the use of mouse models, gene family analyses, and pathway analyses, among others. This resulted in a list of 308 genes sequenced using 16371 pairs of sequencing primers (see online Supplemental Table 1). PCR and sequencing primers were generated using PrimerTile, an automated primer design program which used the most recent version of the dbSNP database (version 130) to omit common variants from designed primers (11). Amplicons were sequenced as described (7) with a mean amplicon length of 648.8 bp. A subset of 5 genes found in the Genetic Testing Registry (http://www.ncbi.nlm.nih.gov/gtr/) were chosen for analysis in 684 ClinSeq samples that had been exome sequenced. These genes were chosen to be representative of the genome on the basis of criteria that included coding DNA sequence (CDS) length, number of exons (minimum 4, maximum 20), GC content, and the presence or absence of pseudogenes in the genome (Table 1, Fig. 1). A larger subset of 19 genes, chosen similarly, were interrogated in 5 ClinSeq samples that had undergone exome, genome, and Sanger sequencing. Validation of variants identified by NGS was simulated using large-scale Sanger sequencing data, which were generated as part of the ClinSeq project (7). All bases with a Phred quality score of Q20 or greater within covered regions were aligned and interrogated using the Consed graphical sequence editor, and genotypes were verified by manual observation of fluorescence peaks (12). Only variants with Sanger data for both forward and reverse read alignments were used in the analysis.

Variants from the exome data that were not validated by the Sanger data were resequenced using the original primers and manually optimized primers designed using Primer3 software (13, 14). Resequencing was performed on a 3130x sequencer using the BigDye 3.1 sequencing kit (Applied Biosystems). Reads generated by resequencing were aligned to genome build hg19 (NCBI build 37) using Sequencher (Gene Codes Corporation). Primer sequences are provided in online Supplemental Table 2.

STATISTICAL ANALYSIS

The replication failure rates in the 19-gene and 5-gene sets from NGS and Sanger sequencing data were compared using the Fisher exact test (15). The extracted data were then subjected to the Jaccard sameness test (16), calculated using all data points with a given minimal MPG score threshold or higher, and that threshold was iterated across the entire range of MPG scores (Fig. 2). Box-and-whisker plots of GC content, CDS length, and exon count for all genes used in this study were plotted using R (Fig. 1). The 95% CI of the accuracy estimate was calculated using the Jeffreys interval calculation (17).

DATA ACCESS

Exome sequencing data from ClinSeq participants are available from NCBI's dbGAP database, accession num ber phs000971.v1.p1. Sanger traces are available from the NCBI SRA database (18) (see online Supplemental Table 1 for associated accession numbers and query strings).

Results

We simulated Sanger validation of NGS data by selecting 2 representative subsets of genes from our data set. Our objective was to include a range of gene attributes, using GC content, CDS length, and exon count as selection criteria (Table 1, Fig. 1). We also included several genes with known pseudogenes to address the challenges of sequence alignment (Table 1).

The mean number of variants with exome coverage of [greater than or equal to] 10 reads per kilobase of interrogated DNA in the exome data was 0.8041 (14258/17732 kb). Of these, 5660 nonreference variants were covered bidirectionally by the Sanger sequencing data. For all interrogated variants, the Sanger reads were evaluated manually to emulate techniques typically used in a clinical setting. Among these 5660 variants, 19 were identified by NGS but not by Sanger sequencing, representing 13 unique single-nucleotide variants.

We next set out to address the possibility that this set of 5 genes was in some way not representative of the wider universe of gene attributes by evaluating a larger set of genes, which necessarily had to be performed on a smaller set of samples. We examined 19 genes (including the 5 used in our initial analysis) from 5 samples using both exome and genome sequencing (Table 1). Within these 19 genes, we identified 714 nonreference variants with coverage of at least 10 reads, with a mean variant/kb rate of 0.1256 (714/5686 kb). There was a strong linear correlation of MPG score and read depth coverage of these variant positions (r2 = 0.8978; see online Supplemental Fig. 1). Of these variants, 234 were covered bidi-rectionally by Sanger sequencing data, and all of those variants were present in exome, genome, and Sanger sequencing data. The replication failure rate in the 19-gene data set (0/234) is not significantly different from the 5-gene dataset (19/5660; P = 1.000, [chi square]).

We further evaluated the 19 discrepant results by performing another round of Sanger sequencing using both the original primers designed through automated primer design and new, optimized sets of sequencing primers designed using primer3 software (13). In 4 cases, sequencing with the original primers yielded a reference (nonvariant) genotype, and the newly designed primers validated the variant found via NGS. The original discrepant Sanger results for these cases could be due to polymorphisms within the sample DNA sequence complementary to the sequence of one of the original primers. In 2 other cases, resequencing with the original primers confirmed the variant, but the newly designed primers yielded no usable sequence. Finally, in 11 cases resequencing with both sets of primers confirmed the original NGS variant call.

Resequencing reconciled all but 2 differences between the NGS and Sanger sequencing data. The remaining 2 discrepancies were found in noncoding regions of the genes apolipoprotein A-V (APOA5) [4] and platelet-derived growth factor receptor beta (PDGFRB), and had MPG quality scores of 4 and 10, respectively (10).

That only 2 of 5660 variants were truly discrepant represents an agreement rate of 0.99965 (95% CI, 0.99887--0.99993). Jaccard sameness scores were plotted against each possible minimum MPG score threshold from the NGS data and the resulting index ranged from 0.99965 to 1.00000, corresponding to a minimum of 99.965% accuracy of NGS compared to Sanger sequencing (Fig. 2).

Discussion

The power and utility of NGS is based on its massively parallel interrogation of nucleic acids. The ability to simultaneously evaluate millions of base pairs allows clinicians and researchers to ask and answer novel and important questions. However, requiring relatively low-throughput dideoxy sequencing as a validation of high-throughput NGS interrogation severely limits the utility of NGS. With the consistently decreasing costs of NGS, the expense and time required to validate variants found in NGS data using Sanger sequencing can quickly outpace the cost of generating the initial NGS data.

Previous studies have provided preliminary evidence that Sanger sequencing validation may not represent the best practice for clinical NGS validation; however, these studies were relatively small in scale and used secondary data from clinical diagnostic laboratories (3-5).

In 2013, Sikkema-Raddatz and colleagues (3) evaluated NGS variants in 84 individuals using a targeted panel including 48 genes, validating 168 novel variants using Sanger sequencing, including 7 indels. They reported nearly 100% Sanger validation of variants identified through their NGS panel. Notably, the single variant that was not initially validated using Sanger sequencing was validated by a subsequent Sanger sequencing run. They concluded that targeted NGS could be reliably implemented as a stand-alone test, with no orthogonal validation required.

McCourt and colleagues (4) then used a combination of NGS technologies to interrogate variants in a host of cancer-related genes. Of the identified NGS variants, 37 were confirmed by Sanger sequencing validation, leading the authors to conclude that existing NGS technologies perform well in detecting known clinically relevant mutations.

In 2014, Strom and colleagues (5) addressed the question of Sanger validation using data from 144 clinical exomes, from which they attempted to Sangervalidate 110 total single nucleotide variants. Of these 110 variants, 109 were validated by Sanger sequencing, and the 1 variant which was not validated had an exome quality score below their quality threshold.

More recently, Baudhuin and colleagues (6) performed a larger-scale study in which data from targeted NGS panels were compared to either Sanger sequence data or data from the 1000 Genomes Project. Sanger sequencing verified 100% of 919 variants identified from the targeted panels.

Combining the data from these 4 studies yields a total of 1234 variants, only 1 of which was not validated by Sanger sequencing. These data, although compelling, are not sufficient to conclude that routine Sanger validation is unnecessary, partially because the largest study included only data from targeted panels with at least 100X coverage in >99.7% of captured bases (6), which is markedly higher coverage than can be expected from current exome sequencing technologies.

To address the need for systematic and large-scale evaluation of orthogonal Sanger validation of NGS, we used a data set of 684 exomes comprising approximately 21 terabases of sequence and matching Sanger data comprising 2.9 million reads from the same samples. We began with the detection of variants from NGS data generated with well-known exome capture kits (Agilent and TruSeq) and Illumina sequencing, coupled with our well-established variant calling process described in a number of prior successful genetic analysis efforts (19-26). We endeavored to select a range of genes that had attributes that were similar to, or were in some aspects more challenging sequencing targets than, a typical gene. This gene set provided us with 5660 variants that we could validate with our Sanger data, a sample set much larger than prior analyses (3-6). Although our Sanger data set included millions of reads that could potentially be compared with the NGS variants, we limited our analysis to model a clinical orthogonal testing scenario as closely as possible. To that end, all variants that met our criteria for interrogation were manually evaluated from the Sanger traces. We also limited our analysis to germline variants from leukocyte DNA, because NGS-based discovery of somatic variants or from formalin-fixed, paraffin-embedded tissue would likely require a separate validation process.

Using this approach, we found only 2 variants among 5660 that were not validated by Sanger sequencing. Those 2 variants had relatively low quality scores for their NGS calls, with one of the variant quality scores being exactly at our standard NGS base-calling quality threshold (MPG = 10), and the other being well below that threshold (MPG = 4). This suggested that, even without setting a minimum quality threshold for accepting NGS variant calls, 99.965% of those calls would be true positives, based on the lower limit of a 95% CI for this large sample. In addition, our data suggested that, through application of a conservative score threshold, a single high-quality (in our case, MPG >10) "sentinel" call in any sample leads to the same variant being more likely correctly called in other samples, irrespective of quality score for the variant in that sample. Our sentinel call approach resulted in 583 variant calls with an MPG score of <10 being validated by Sanger sequencing. If a flat quality score threshold was applied to all of the data, these variants would have been missed through NGS screening, which could lead to variants that might impact a patient's health being undetected. Furthermore, if a minimum quality threshold of MPG score 7 was applied to these data [which represents a very conservative threshold approximating to a GATK (Genome Analysis Toolkit) Q30 score] there would have been only a single nonconfirmation, leading to a confirmation rate of 99.9823%. Though MPG is not the most widely used quality metric for NGS data, our results can easily be extended to NGS data using any quality metric by using this sentinel call approach with equivalent score thresholds.

The more striking conclusion came from resequencing the 19 originally discrepant variants, in that the majority of these orthogonal Sanger validations were themselves incorrect. Seventeen of the NGS variants would have been considered false positives if a single round of Sanger sequencing were used as a validation criteria. Our results suggest that if such practice were used in a clinical setting, more positive NGS variants would be discarded as (incorrectly designated) false positives, as compared to using the NGS data directly. Jaccard index analysis supported this assertion, showing no appreciable difference between NGS and Sanger sequencing with respect to variants within our data set, and complete agreement of the 2 sequencing methods at an MPG score threshold higher than 10.

Our measured validation rate of at least 99.965% for NGS data across a large data set with no established minimum quality threshold represents higher accuracy than many medical tests currently used by clinicians. Results of such tests are routinely used to determine the course of treatment for a patient without any expectation of orthogonal validation. Given these data, we conclude that Sanger validation of NGS variants that are associated with robust quality scores should not be performed routinely. At the same time, we recognize that some variants detected using NGS technology can have serious medical implications for the tested probands and their family members. In such cases, performing a second orthogonal validation may be appropriate. One can envision a future in which such determinations are made by the ordering clinician on the basis of the presenting findings and the intended or anticipated clinical use of the genetic testing result. Although this assertion will be controversial, we have been unable to find reference to a clinical laboratory test that boasts a 99.965% or higher analytical confirmation rate for which orthogonal confirmatory testing is routinely mandated, and we suggest that leaving the question of confirmatory testing in the hands of the ordering physician is most appropriate. This would align clinical genomics with the practices across many fields of medicine, reduce overall costs of genomic testing, and potentially reduce the error rate of inappropriately labeling a NGS variant as a false positive due to failure of the orthogonal assay. The fact that some CLIA-approved laboratories are already returning NGS variant results for clinically relevant variants in certain subsets of genes, such as the ACMG (American College of Medical Genetics and Genomics)-established list of genes for return of secondary findings (27), suggests that the field is already moving toward this practice. We therefore recommend that NGS testing results should be treated as many other clinical tests are treated: imperfect, but highly reliable.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contribution to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article.

Authors' Disclosures or Potential Conflicts of Interest: Upon manuscript submission, all authors completed the author disclosure form. Disclosures and/or potential conflicts of interest:

Employment or Leadership: None declared.

Consultant or Advisory Role: L.G. Biesecker, Illumina Corp.

Stock Ownership: None declared.

Honoraria: L.G. Biesecker, Wiley-Blackwell, Inc.

Research Funding: The Intramural Research Program of the National Human Genome Research Institute.

Expert Testimony: None declared.

Patents: None declared.

Other Remuneration: L.G. Biesecker, Genentech Corp.

Role of Sponsor: The funding organizations played a direct role in the design of study, choice of enrolled patients, review and interpretation of data, and final approval of manuscript.

Acknowledgments: The authors acknowledge Dr. Niraj Trivedi of NHGRI for statistical analyses, and NancyHansen of the Comparative Genomic Analysis Unit, NHGRI, for lending her bioinformatics expertise. The authors thank Dr. Steve Lincoln of Invitae for helpful discussions.

References

(1.) de Souza F. EmTech: Illumina says 228,000 human genomes will be sequenced this year. MIT Technology Review. http://www.technologyreview.com/news/ 531091/emtech-illumina-says-228000-humangenomes-will-be-sequenced-this-year/ (Accessed July 2015).

(2.) Sanger F, Nicklen S, Coulson AR. DNA sequencing with chain-terminating inhibitors. Proc Natl Acad Sci U S A 1977;74:5463-7.

(3.) Sikkema-Raddatz B, Johansson LF, de Boer EN, Almomani R, Boven LG, van den Berg MP, et al. Targeted next-generation sequencing can replace Sanger sequencing in clinical diagnostics. Hum Mutat 2013;34: 1035-42.

(4.) McCourt CM, McArt DG, Mills K, Catherwood MA, Maxwell P, Waugh DJ, et al. Validation of next generation sequencing technologies in comparison to current diagnostic gold standards for BRAF, EGFR and KRAS mutational analysis. PLoS ONE 2013;8:e69604.

(5.) Strom SP, Lee H, Das K, Vilain E, Nelson SF, Grody WW, Deignan JL. Assessing the necessity of confirmatory testing for exome-sequencing results in a clinical molecular diagnostic laboratory. Genet Med 2014;16: 510-5.

(6.) Baudhuin LM, Lagerstedt SA, Klee EW, Fadra N, Oglesbee D, Ferber MJ. Confirming variants in next-generation sequencing panel testing by Sanger sequencing. J Mol Diagn 2015;17:456-61.

(7.) Biesecker LG, Mullikin JC, Facio FM, Turner C, Cherukuri PF, Blakesley RW, et al. The ClinSeq Project: piloting large-scale genome sequencing for research in genomic medicine. Genome Res 2009;19:1665-74.

(8.) Johnston JJ, Teer JK, Cherukuri PF, Hansen NF, Loftus SK; NIH Intramural Sequencing Center (NISC), et al. Massively parallel sequencing of exons on the X chromosome identifies RBM10 as the gene that causes a syndromic form of cleft palate. Am J Hum Genet 2010; 86:743-8.

(9.) Johnston JJ, Rubinstein WS, Facio FM, Ng D, Singh LN, Teer JK, et al. Secondary variants in individuals undergoing exome sequencing: screening of 572 individuals identifies high-penetrance mutations in cancer susceptibility genes. Am J Hum Genet 2012;91:97108.

(10.) Teer JK, Bonnycastle LL, Chines PS, Hansen NF, Aoyama N, Swift AJ, et al. Systematic comparison of three genomic enrichment methods for massively parallel DNA sequencing. Genome Res 2010;20:1420-31.

(11.) Chines PS, Swift AJ, Bonnycastle LL, Erdos M, Mullikin JC; NIH Intramural Sequencing Center, et al. PrimerTile: designing overlapping PCR primers for resequencing. Am J Hum Genet 2005;77:A1257.

(12.) Gordon D, Green P. Consed: a graphical editor for next-generation sequencing. Bioinformatics 2013;29: 2936-7.

(13.) Untergasser A, Cutcutache I, Koressaar T, Ye J, Faircloth BC, Remm M, et al. Primer3-new capabilities and interfaces. Nucleic Acids Res 2012;40:e115.

(14.) Koressaar T, Remm M. Enhancements and modifications of primer design program Primer3. Bioinforma Oxf Engl 2007;23:1289-91.

(15.) Fisher RA. On the interpretation of [chi square] 2 3 4 5 6 7 from contingency tables, and the calculation of P. J R Stat Soc 1922:85:8794. Available from: https://digital.library.adelaide. edu.au/dspace/handle/2440/15173 (Accessed December 2015).

(16.) Jaccard P. The distribution of the Flora in the Alpine Zone. 1. New Phytol 1912;11:37-50.

(17.) Brown LD, Cai TT, DasGupta A. Interval estimation for a binomial proportion. Stat Sci 2001;16:101-33.

(18.) Leinonen R, Sugawara H, Shumway M. The sequence read archive. Nucleic Acids Res 2011;39:D19-21.

(19.) Johnston JJ, Lewis KL, Ng D, Singh LN, Wynter J, Brewer C, et al. Individualized iterative phenotyping for genome-wide analysis of loss-of-function mutations. AmJ Hum Genet 2015;96:913-25.

(20.) Sen SK, Boelte KC, Barb JJ, Joehanes R, Zhao X, Cheng Q, et al. Integrative DNA, RNA, and protein evidence connects TREML4 to coronary artery calcification. Am J Hum Genet 2014;95:66-76.

(21.) Pierson TM, Yuan H, Marsh ED, Fuentes-Fajardo K, Adams DR, Markello T, et al. GRIN2A mutation and early-on-set epileptic encephalopathy: personalized therapy with memantine. Ann Clin Transl Neurol 2014;1: 190-8.

(22.) Yuan H, Hansen KB, Zhang J, Pierson TM, Markello TC, Fajardo KVF, et al. Functional analysis of a de novo GRIN2A missense mutation associated with early-onset epileptic encephalopathy. Nat Commun 2014;5:3251.

(23.) Gonsalves SG, Ng D, Johnston JJ, Teer JK, Stenson PD, Cooper DN, et al. Using exome data to identify malignant Hyperthermia susceptibility mutations. Anesthesiology 2013;119:1043-53.

(24.) Gartner JJ, Parker SCJ, Prickett TD, Dutton-Regester K, Stitzel ML, Lin JC, et al. Whole-genome sequencing identifies a recurrent functional synonymous mutation in melanoma. Proc Natl Acad Sci U S A 2013;110: 13481-6.

(25.) Ng D, Johnston JJ, Teer JK, Singh LN, Peller LC, Wynter JS, et al. Interpreting secondary cardiac disease variants in an exome cohort. Circ Cardiovasc Genet 2013; 6:337-46.

(26.) Le Gallo M, O'Hara AJ, Rudd ML, Urick ME, Hansen NF, O'Neil NJ, et al. Exome sequencing of serous endometrial tumors identifies recurrent somatic mutations in chromatin-remodeling and ubiquitin ligase complex genes. Nat Genet 2012;44: 1310-5.

(27.) Green RC, Berg JS, Grody WW, Kalia SS, KorfBR, Martin CL, et al. ACMG recommendations for reporting of incidental findings in clinical exome and genome sequencing. Genet Med 2013;15:565-74.

(28.) Kent WJ, Sugnet CW, Furey TS, Roskin KM, Pringle TH, Zahler AM, et al. The human genome browser at UCSC. Genome Res 2002;12:996-1006.

(29.) Pruitt KD, Tatusova T, Maglott DR. NCBI reference sequences (RefSeq): a curated non-redundant sequence database of genomes, transcripts and proteins. Nucleic Acids Res 2007;35:D61-5.

Tyler F. Beck,* [1] James C. Mullikin on behalf of the NISC Comparative Sequencing Program, [1,2] and Leslie G. Biesecker [1] *

[1] National Human Genome Research Institute, NIH, Bethesda, MD; [2] NIH Intramural Sequencing Center, Rockville, MD.

* Address correspondence to this author at: National Human Genome Research Institute, NIH, 49 Convent Dr., 4A56, Bethesda, MD 20830. Fax 301-402-2170; e-mail lesb@mail.nih.gov.

Received September 16, 2015; accepted January 7, 2016.

Previously published online at DOI: 10.1373/clinchem.2015.249623

[3] Nonstandard abbreviations: NGS, next-generation sequencing; MPG, MostProbable Genotype; CDS, coding DNA sequence.

[4] Human genes: APOA5, apolipoprotein A-V; PDGFRB, platelet-derived growth factor receptor beta; LDLRAP1, low density lipoprotein receptor adaptor protein 1; MMP9, matrix metallopeptidase 9; VEGFA, vascular endothelial growth factor A; ACTA2, actin, alpha 2, smooth muscle, aorta; APOC3, apolipoprotein C-III; CAV1, caveolin 1; CD40, CD40 molecule; CETP, cholesterol ester transfer protein, plasma; CIITA, class II, major histocompatibility complex, transactivator; FGG, fibrinogen gamma chain; GPX1, glutathione peroxidase 1; LPL, lipoprotein lipase; MBL2, mannose-binding lectin (protein C)2,soluble; MVK, mevalonate kinase; PITX2, paired like homeodomain 2; TNFRSF1A, tumor necrosis factor receptor superfamily, member 1A; UCP2, uncoupling protein 2 (mitochondrial, protein carrier).

Caption: Fig. 1. Box-and-whisker plots of gene statistics. Box-and-whisker plots show the distribution of genes in each candidate set used in this analysis across GC content (A), CDS length (B), and exon count(C). Data on these genes were collected using the UCSC Genome Browser [Kent et al. (28)] or NCBI's Entrez [Pruitt et al. (29)]. In the case of multiple transcripts, the transcript encoding the longest protein isoform was used.

Caption: Fig. 2. Jaccard Index with Increasing MPG Cutoff. The Jaccard sameness index was used to evaluate the agreement of variants discovered using NGS vs Sanger sequencing, correlated with increasing MPG thresholds, then plotted using R. Data from both sequencing methods were incomplete agreement at an MPG higher than 10, resulting in a score of 1.000 at all points thereafter.
Table 1. Genes used in these analyses. (a)

Gene name    Chromosome    Exon count    CDS length    Pseudogene?

APOA5 (b)       Chr11           4           1101          False
LDLRAP1         Chr1            9            927          False
MMP9            Chr20          13           2124          False
PDGFRB          Chr5           23           3321          False
VEGFA           Chr6            7           1239          False
ACTA2           Chr10           9           1134           True
APOC3           Chr11           4            300          False
CAV1            Chr7            3            537          False
CD40            Chr20           9            834          False
CETP            Chr16          16           1482          False
CIITA           Chr16          20           3396          False
FGG             Chr4           10           1362           True
GPX1            Chr3            1            612           True
LPL             Chr8           10           1428          False
MBL2            Chr10           4            747           True
MVK             Chr12          11           1191          False
PITX2           Chr4            6            954          False
TNFRSF1A        Chr12          10           1368          False
UCP2            Chr11           8            930          False

Gene name    GC content

APOA5 (b)       0.64
LDLRAP1         0.60
MMP9            0.64
PDGFRB          0.58
VEGFA           0.62
ACTA2           0.53
APOC3           0.59
CAV1            0.47
CD40            0.55
CETP            0.54
CIITA           0.62
FGG             0.41
GPX1            0.72
LPL             0.50
MBL2            0.53
MVK             0.60
PITX2           0.59
TNFRSF1A        0.57
UCP2            0.58

(a) The top 5 genes listed in this table (above the dark line) were
interrogated via exomeand Sanger sequencing in 684 samples, whereas
all genes in the table were interrogated in 5 samples via exome,
genome, and Sanger sequencing.

(b) APOA5, apolipoprotein A-V; LDLRAP1, low density lipoprotein
receptor adaptor protein 1; MMP9, matrix metallopeptidase 9;
PDGFRB, platelet-derived growth factor receptor beta; VEGFA,
vascular endothelial growth factor A; ACTA2, actin, alpha 2, smooth
muscle, aorta; APOC3, apolipoprotein C-III; CAV7, caveolin 1; CD40,
CD40 molecule; CETP, cholesteryl ester transfer protein, plasma;
CIITA, class II, major histocompatibility complex, transactivator;
FGG, fibrinogen gamma chain; GPX1, glutathione peroxidase 1; LPL,
lipoprotein lipase; MBL2, mannose-binding lectin (protein C) 2,
soluble; MVK, mevalonate kinase; PITX2, paired like homeodomain 2;
TNFRSF1A, tumor necrosis factor receptor superfamily, member 1A;
UCP2, uncoupling protein 2 (mitochondrial, protein carrier).
COPYRIGHT 2016 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2016 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Molecular Diagnostics and Genetics
Author:Beck, Tyler F.; Mullikin, James C.; Biesecker, Leslie G.
Publication:Clinical Chemistry
Article Type:Report
Date:Apr 1, 2016
Words:5010
Previous Article:Minimizing matrix effects for the accurate quantification of 25-hydroxyvitamin D metabolites in dried blood spots by LC-MS/MS.
Next Article:Downregulation of miR-15a and miR-16-1 at 13q14 in chronic lymphocytic leukemia.
Topics:

Terms of use | Privacy policy | Copyright © 2018 Farlex, Inc. | Feedback | For webmasters