Simultaneous Testing for 6 Lysosomal Storage Disorders and X-Adrenoleukodystrophy in Dried Blood Spots by Tandem Mass Spectrometry.

In March 2015, Pompe disease [acid [alpha]-glucosidase deficiency; Online Mendelian Inheritance in Man (OMIM) [2] 232300), and in February 2016, mucopolysaccharidosis type I (MPS I) ([alpha]-L-iduronidase deficiency; OMIM 607014, 607015, and 607016), and X-adrenoleukodystrophy (ALD) (OMIM 300100) were included in the US Recommended Uniform Screening Panel of the US Secretary of Health and Human Services (1). Other lysosomal storage disorders (LSD) have also been suggested for inclusion in the Recommended Uniform Screening Panel (2). These requests are driven by the increased availability of treatment and screening test options, and by the emerging evidence of the benefit of early intervention preceding the onset of irreversible symptoms. Successful state-based advocacy efforts have led to several newborn screening programs already screening for some of these conditions (3). In 2006, the state of New York started screening for Krabbe disease (galactocerebrosidase deficiency; OMIM 245200), and in 2014 added ALD and Pompe disease to its newborn screening panel (4, 5, 6). Two other US states, Missouri and Illinois, have been screening for 5 LSDs since 2012 and 2015 respectively; Kentucky began screening for Pompe disease, MPS I, and Krabbe disease in 2016. A few other states already have passed legislation to include LSDs in their newborn screening panels. In addition to MPS I, Pompe, and Krabbe diseases, screening may include Fabry disease ([alpha]-galactosidase A deficiency; OMIM 301500), NiemannPick type A and B disease (NPA/B) (acid sphingomyelinase deficiency; OMIM 257200 and 607616), mucopolysaccharidosis type II (iduronate-2-sulfatase deficiency; OMIM 309900), and Gaucher disease ([beta]-glucosidase deficiency; OMIM 230800). Outside the US, screening programs in Taiwan have implemented newborn screening for Pompe and Fabry diseases (2). A major finding from past and ongoing newborn screening projects is the much higher than expected prevalence of infants with low lysosomal enzyme deficiencies and genotypes suggestive of later onset disease variants (2). Notably, these LSDs are likely underdiagnosed in adult patients because they present with symptoms different from the classic, earlier onset phenotypes.

Given this context, our aim was to develop an assay that would allow for the simultaneous high-throughput testing of dried blood spots (DBS) for 6 of the abovementioned LSDs, ALD, and peroxisomal biogenesis disorders (PBD) by combining previously described methods for the measurement of lysosomal enzyme activities and lysophosphatidylcholines (7-12). Following validation of this multiplex assay, we have applied it for screening of at-risk patients of any age.

Materials and Methods

MATERIALS

All lysophosphatidylcholine (LPC) species were purchased from Avanti Polar Lipids. Chloroform (ACS reagent grade) and methanol ([greater than or equal to] 99.9% purity) were purchased from EMD chemicals. Ammonium formate, silica, sodium phosphate monobasic, sodium citrate tribasic dehydrate, sodium acetate, taurocholic acid sodium salt hydrate, N-acetylgalactosamine and D-saccharic acid 1,4 lactone monohydrate were purchased from SigmaAldrich. Glacial acetic acid and zinc chloride were purchased from Mallinckrodt. Acarbose was purchased from Toronto Research Chemicals. Substrates (S) and internal standards (IS) for the 6 lysosomal enzymes were manufactured by Genzyme Pharmaceutical and provided by the Newborn Screening Branch of the CDC (8, 13).

Preparation of calibrators and controls. LSD controls at 4 levels of enzyme activity were prepared and supplied by the CDC (10, 13). These controls include a base pool control (CDC BP) deficient in enzyme activities, a low control (CDC low) with reduced enzyme activities, a medium control (CDC med) with moderate enzyme activities and a high control (CDC high) with normal enzyme activities. QC materials with LPCs at 2 levels were prepared in-house. For level 1 (ALD QC 1), EDTA blood was spotted on Whatman 903 filter paper and allowed to dry for at least 3 h at ambient conditions. For level 2 (ALD QC 2), EDTA blood was enriched with [C.sub.20]-, [C.sub.22]-, [C.sub.24]-, and [C.sub.26]-LPC, spotted on Whatman 903 filter paper, and allowed to dry for at least 3 h at ambient conditions. The spotted cards were then transferred to a zip-lock bag with desiccator and stored at -20 [degrees]C. Liquid calibrators with product (P) to IS ratios (P/IS) of 0.00, 0.05, 0.5, 1.0, 2.0 and 5.0 for each enzyme supplied by the CDC, as well as liquid calibrators corresponding to 0.00, 0.58, 0.97, 1.16, 1.74 and 2.32 mg/L of [C.sub.20]-, [C.sub.22]-, [C.sub.24]-, and [C.sub.26]-LPC in a 3-mm DBS were used for linearity studies.

Samples. Reference intervals were established from DBS of 550 controls. In addition, DBS samples from confirmed cases of MPS I (n = 5), Gaucher disease (n = 5), NPA/B disease (n = 2), Pompe disease (n = 5), Krabbe disease (n = 5), Fabry disease (n = 11), Fabry carriers (n = 17), ALD (n = 9), ALD carriers (n = 5), and PBD (n = 5) were analyzed. These studies were approved by the Mayo Clinic Institutional Review Board (10000292 and 10-006866).

METHODS

A 5-plex cocktail to assay [beta]-glucosidase (ABG), acid [alpha]-glucosidase (GAA), galactocerebrosidase (GALC), [alpha]-galactosidase A (GLA), and [alpha]-L-iduronidase (IDUA) activity was based on the method described by Orsini et al. (9). The vials containing S and IS for ABG and GAA were reconstituted with methanol and transferred to the GALC S + IS vial followed by evaporation. The GLA S + IS vial was reconstituted with 1.8 mL of 96 g/L sodium taurocholate in water and then transferred to the dry GALC S+ IS vial, which was then heated to 60 [degrees]C to dissolve all solids. To the GALC S + IS vial the following reagents were added: 0.3 mL of 0.8 mmol/L acarbose in water (GAA inhibitor) and 2.88 mL of 1 mol/L ^-acetylgalactosamine (GLA inhibitor) in buffer (0.2 mol/L sodium phosphate + 0.1 mol/L sodium citrate, pH = 4.4). The IDUA S + IS vial was then reconstituted with 12.52 mL of buffer (0.2 mol/L sodium phosphate + 0.1 mol/L sodium citrate, pH = 4.4) and 0.5 mL of 3 mmol/L D-saccharic acid 1,4-lactone in water. The reconstituted IDUA S + IS vial was transferred to the GALC S + IS vial. An acid sphingomyelinase (ASM) cocktail was prepared separately by reconstituting the ASM S + IS vial with 0.15 mL of 96 g/L sodium taurocholate in water and then adding 17.85 mL of buffer (0.85 mol/L sodium acetate + 0.604 mmol/L zinc chloride, pH = 5.7). After mixing, 1-mL aliquots of these reagents were stored at -20 [degrees]C for up to 2 months.

Three 3-mm discs were excised from a DBS and placed into individual microtiter plates. The first disc was treated with 30 [micro]L of ASM cocktail, the second one with 30 [micro]L 5-plex cocktail containing S and IS for ABG, GAA, GLA, GALC, and IDUA. The 2 enzyme plates were sealed, centrifuged for 2 min at 493g, and incubated for 19 h at 37[degrees]C. The third disc was extracted with 300 [micro]L of methanol containing 12 ng/mL [d.sub.4]-[C.sub.26] LPC as previously described (11). The extract was then evaporated under heated nitrogen, reconstituted in 130 [micro]L of mobile phase (800 mL methanol/200 mL water with 5 mmol/L ammonium formate) and stored refrigerated. Following the 19-h incubation, the 5-plex and ASM reactions were stopped by adding 200 [micro]L of 1:1 ethyl acetate: methanol to each plate. The 5-plex and ASM plates were combined into a single deep-well plate and liquid-liquid extraction was performed by adding 400 [micro]L of ethyl acetate and then 400 [micro]L of water to each well. The plate was sealed and centrifuged at 493g for 2 min, and 150 [micro]L of the organic layer was transferred to a new plate, evaporated under nitrogen, and reconstituted in 150 [micro]L of 19:1 ethyl acetate:methanol. The reconstituted extract was added to a silica-containing filter plate that had been washed previously with 200 [micro]L of 19:1 ethyl acetate. The sample moved through the filter plate via positive pressure and the plate was eluted with an additional 200 [micro]L of 19:1 ethyl acetate. The 350 [micro]L of eluent collected was evaporated under a stream of nitrogen and then reconstituted with the stored ALD plate. The samples were subjected to flow injection tandem mass spectrometry (FIA-MS/MS) analysis. The total time to process a plate of 96 samples was 60 and 70 min, respectively, before and after the 19 h incubation.

MS/MS procedure. A triple-quadrupole MS/MS system (Applied Biosystems/MDS Sciex API3200) operated in positive ion mode (source voltage, 5500 V) was used. Mass calibration and resolution of both resolving quadrupoles were optimized with a poly(propylene)glycol solution introduced by an infusion pump. Method optimization was performed by infusing a solution containing the measured enzyme products, LPC species and IS at 0.6 mL/h. The instrument was optimized to monitor the transitions (see Table 1 in the Data Supplement that accompanies the online version of this article at http:// www.clinchem.org/content/vol62/issue9). The single reaction monitoring experiments (100 ms dwell, each experiment) were added to the MS/MS method. Sample introduction into the Turbo V[TM] atmospheric pressure ionization source was achieved by a CTC Leap autosampler and Shimadzu HPLC system. Autosampler injections of 20 [micro]L per sample were made into the liquid chromatography (LC) mobile phase (800 mL methanol/ 200 mL water with 5 mmol/L ammonium formate) flow of 0.250 mL/min without the use of a chromatographic column. Between injections, analysis time was 1 min per sample. Because an unidentified substance that was inconsistently present in DBS interfered with accurate quantification of [C.sub.26]-LPC by the flow injection tandem mass spectrometry (FIA-MS/MS) method. Each pre pared sample with an apparently increased concentration of [C.sub.26]-LPC was verified by a reflex analysis by use of LC-MS/MS as described previously (11, 12) to minimize false-positive results.

Results

LINEARITY AND IMPRECISION

Liquid calibrators for ABG, ASM, GAA, GALC, GLA, and IDUA at 7 different P/IS ratios (0, 0.05, 0.10, 0.50, 1.0, 2.0, and 5.0 P/IS) showed reproducible signals with a linear response (mean n = 3 each analyte; ABG, m = 1.021,b = 0.047, [R.sup.2] = 0.999; ASM, m = 1.338,b = 0.028, [R.sup.2] = 0.999; GA,A m = 0.965, b = 0.036, [R.sup.2] = 0.999; GALC, m = 0.788, b = 0.033, [R.sup.2] = 0.998; GLA, m = 0.900, b = 0.045, [R.sup.2] = 0.999; IDUA, m = 0.990, b = 0.039, [R.sup.2] = 0.999). Liquid calibrators for [C.sub.20]-, [C.sub.22]-, [C.sub.24]-, and [C.sub.26]-LPC at 6 different concentrations (0.000, 0.581, 0.968, 1.161, 1.742, and 2.322 [micro]g/mL) also showed reproducible signals with a linear response (mean n = 3 each analyte, [C.sub.20]-LPC, m = 0.707, b = 0.029, [R.sup.2] = 0.994; [C.sub.22]-LPC, m = 0.740, b = 0.018, [R.sup.2] = 0.995; [C.sub.24]-LPC, m = 0.700, m = 0.022, [R.sup.2] = 0.994; [C.sub.26]-LPC, m = 0.785, b = 0.020, [R.sup.2] = 0.993). Intra- and interassay imprecision data for each enzyme activity and LPC concentration are provided in online Supplemental Tables 2 and 3.

SAMPLE STABILITY

The stability of prepared samples was assessed by analysis of 6 QCs and 10 randomly selected DBS samples, 1, 2 and 3 days after preparation. Prepared samples returned the same clinical result after 2 days of storage at 10[degrees]C or after 3 days of storage at -20[degrees]C.

METHOD COMPARISON

A method comparison was performed by use of DBS from previously diagnosed individuals with MPS I (n = 5), Gaucher disease (n = 5), NPA/B disease (n = 2), Pompe disease (n = 5), Krabbe disease (n = 5), Fabry disease (n = 11), ALD (n = 9), PBD (n = 5), and female carriers of GLA (n = 17) and ABCD1 (n = 5) mutations, as well as controls (n = 20). Alternative assays replicated in our laboratory were another MS/MS multiplex method for ABG, ASM, GAA, GALC, and GLA analysis (10), a fluorometric method for IDUA analysis (14), and an LC-MS/MS method for [C.sub.20]-, [C.sub.22]- [C.sub.24]-, and [C.sub.26]-LPC analysis (11, 12). Analytical agreement between methods varied by analyte (see online Supplemental Table 4); however, all methods were clinically concordant with the exception of Fabry carriers, for which the selected cutoff value resulted in lower clinical sensitivity (35% vs 82%) when compared to the reference method. False-negative results for Fabry carriers were not surprising because Fabry carriers may have considerable residual activity that overlaps with unaffected controls (15).

INTERFERENCE

Potential interference due to the presence of anticoagulants (EDTA, acid citrate dextrose and sodium heparin) was evaluated. Blood was collected from 2 healthy donors via a finger stick and from vein into tubes containing acid citrate dextrose, EDTA and sodium heparin. The blood was spotted on filter paper, allowed to dry and then analyzed. Results were less than [+ or -] 20% different between these collection methods.

FILTER PAPER STUDY

Whatman 903, Alhstrom 226, and Munktell filter papers were compared by spotting blood collected from 10 healthy donors on each type of paper. The mean difference from the mean value in each donor was measured 0.8%, 5.0%, -6.5% (donor mean range = 2.8-7.3 nmol * m[L.sup.-1] * [h.sup.-1]) for ABG; 2.2%, 8.2%, -10.4% (4.9 -15.2; range and unit as shown above) for ASM; -2.6%, 3.7%, -1.1% (4.2-13.7) for GAA; -1.3%, 0.5%, 0.8% (0.4-1.4) for GALC; -0.5%, 1.6%, -1.1% (4.4-11.0) for GLA; -4.9%, 3.6%, 1.3% (2.66.8) for IDUA; -1.6%, 11.7%, -10.1% (0.07-0.18 mg/L) for [C.sub.20]-LPC; 3.0%, 12.1%, -15.0% (0.04-0.08) for [C.sub.22]-LPC; 1.0%, 10.6%, -11.5% (0.05-0.09) for [C.sub.24]-LPC; and 5.3%, 6.1%, -11.4% (0.09-0.22 mg/L) for [C.sub.26]-LPC in Whatman 903, Alhstrom 226 and Munktell filter papers, respectively. These findings were consistent with other enzyme assays measured in DBS on filter paper from different manufacturers (16).

REFERENCE RANGES

Reference intervals were determined according to Morkrid et al. (17) following the analysis of 550 DBS samples collected from healthy adult donors (n = 280 male; n = 270 female; age range: 1 day to 86 years, median age 23 years). Cutoff values for ABG (<3.0 nmol * m[L.sup.-1] * [h.sup.-1]), ASM (<5.5 nmol * m[L.sup.-1] * [h.sup.-1]), GAA (2<4.0 nmol * m[L.sup.-1] * [h.sup.-1]), GALC (2<0.4 nmol * m[L.sup.-1] * [h.sup.-1]), GLA (2<4.0 nmol * m[L.sup.-1] * [h.sup.-1]), IDUA (2<2.0 nmol * m[L.sup.-1] * [h.sup.-1]), [C.sub.20]-LPC (2>0.26 mg/L), [C.sub.22]-LPC (2>0.11 mg/L), [C.sub.24]-LPC (2>0.20 mg/L) and [C.sub.26]-LPC (20.30 mg/L) were determined by comparing relevant markers measured in patient samples to those observed in controls (Fig. 1).

PROSPECTIVE CLINICAL TESTING

Following the analytical and clinical validation of the assay using samples from known patients, we began testing clinical samples prospectively with the goal of providing an efficient and cost-effective diagnostic approach to patients who present with signs and/or symptoms that could be caused by deficiencies of one of the LSD enzymes analyzed or by a peroxisomal disorder. Table 1 summarizes the assay's performance after testing 578 samples between September 2011 and December 2015.

Discussion

To facilitate the rapid diagnosis of patients of all ages with treatable LSDs and ALD, we developed and implemented a multiplex MS/MS assay to screen for 6 LSDs, ALD, and PBD in DBS. Late-onset variants of these conditions often exhibit a nonspecific phenotype, making it particularly difficult to consider or justify a request for several single enzyme activity tests or single gene analyses. Compared to single-enzyme assays or molecular genetic analysis of relevant genes or gene panels, the DBS-based, multiplex-test approach is less expensive and allows for fast turnaround time and laboratory efficiency (approximately 3 h of technologist time for up to 96 samples) because patients with various differential diagnoses can be tested simultaneously and a sufficient number of samples can be batched more quickly for analysis. Furthermore, the ability to compare enzyme and LPC results to each other has advantages because disease and control profiles can be determined and compared, which has been shown to markedly improve clinical specificity in newborn screening (18, 19). The finding of multiple low enzyme activities in the same specimen can also help avoid false-positive results because such findings suggest an artifact likely caused by improper handling or shipment of the specimen (i.e., exposure of the sample to heat during transit to the laboratory).

Between September 2011 and December 2015 we analyzed a total of 578 samples from patients of all ages (median 2.8 years; range birth to 86 years; female: n = 256; male: n = 322). A majority of samples were submitted because these patients displayed symptoms suggestive of one of the lysosomal or peroxisomal conditions detected by this assay. Abnormal results were reported for all disorders tested, and the number of male and female patients was similar for the autosomal recessive conditions. Except for 18 cases that were lost to follow up, the tentative diagnoses were confirmed or excluded by additional biochemical and/or molecular genetic testing. Excluding the 18 cases lost to follow up, clinical sensitivity, clinical specificity, and the positive predictive value were very high for a screening test, but the false-positive rate was high at 4% (Table 1). The latter is not entirely surprising because unaffected carriers and patients deemed pseudo-deficient are considered false positive. The identification of carriers and pseudo-deficient patients is expected because it reflects both the known limitation of enzyme assays and the experience from new born screening for these conditions (2). Abnormal results must be followed up with additional confirmatory testing, such as molecular genetic analysis of relevant genes or testing of other disease-specific biomarkers. The same DBS specimen can be used for such analyses. Except for Pompe disease, biomarkers for the other LSDs included in this assay can be measured in DBS, specifically glycosaminoglycans for MPS I, psychosine for Krabbe disease, glucosylsphingosine for Gaucher disease, globotriaosylsphingosine for Fabry disease, and sphingomyelin for NPA/B disease (20-23). Because [C.sub.26]-LPC can be falsely increased in DBS due to an unknown compound in some specimens, we apply reflex testing of samples with increased [C.sub.26]-LPC by LC-MS/MS as previously reported (11), which explains the absence of false-positive results for ALD and other peroxisomal conditions (Table 1).

This DBS-based assay can also be applied to newborn screening. To this end, our laboratory is currently completing a study of 100000 de-identified newborn screening samples to evaluate this assay's appropriateness for population wide screening. However, going forward, the test method requires modifications because the LSD substrates and IS currently provided by the CDC will be discontinued by the end of 2016 (C. Cuthbert, personal communication, May 2016). New reagents based on the work by Gelb and colleagues are now available from PerkinElmer, which may allow additional efficiencies because only two 3-mm DBS may be required for the determination of the lysosomal enzyme activities and LPC concentrations, and the solid phase extraction step could be omitted (24).

In conclusion, the assay described here allows for a more efficient and cost-effective evaluation of patients of all ages with symptoms suggestive of the tested conditions.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article.

Authors' Disclosures or Potential Conflicts of Interest: Upon manuscript submission, all authors completed the author disclosure form. Disclosures and/or potential conflicts of interest:

Employment or Leadership: All authors are employed by Mayo Clinic.

Consultant or Advisory Role: None declared.

Stock Ownership: None declared.

Honoraria: None declared.

Research Funding: Eunice Kennedy Shriver National Institute of Child Health and Human Development, National Institutes of Health (HHSN275201000017C); P. Rinaldo, the T. Denny Sanford Professorship fund; D. Matern, Newborn Screening Translational Research Network (HHSN275200800001C 01), Mayo Clinic, The Legacy of Angels Foundation, Glaxo Smith Kline, and Hunter's Hope Foundation.

Expert Testimony: None declared.

Patents: None declared.

Role of Sponsor: The funding organizations played no role in the design of study, choice of enrolled patients, review and interpretation of data, and final approval of manuscript.

Acknowledgments: We are grateful to Dr. Joseph J. Orsini for sharing his multiplexed MS/MS assay for 5 LSDs (9), to Drs. Robert Vogt and Hui Zhou of CDC's Newborn Screening Branch for providing relevant reagents, as well as to Dr. Gerald Raymond, Ann Moser, Dr. Amber Salzman, and Dr. Fred Lorey for sharing DBS of known patients with ALD and other peroxisomal disorders.

References

(1.) U.S. Department of Health & Human Services, Advisory Committee on Heritable Disorders in Newborns and Children. http://www.hrsa.gov/advisorycommittees/ mchbadvisory/heritabledisorders/index.html (Accessed May 2016).

(2.) Matern D, Gavrilov D, Oglesbee D, Raymond K, Rinaldo P, Tortorelli S. Newbornscreening for lysosomal storage disorders. Sem Perinatol 2015; 39:206-16.

(3.) The Association of Public Health Laboratories, Newborn Screening Technical assistance and Evaluation Program (NewSTEPs). https://data.newsteps.org/newsteps-web/ reports/screenedConditions/list.action (Accessed May 2016).

(4.) New York State Department of Health, Wadsworth Center. http://www.wadsworth.org/programs/newborn/ screening/screened-disorders (Accessed May 2016).

(5.) Orsini JJ, Morrissey MA, Slavin LN, Wojcik M, Biski C, Martin M, et al. Implementation of newborn screening for Krabbe disease: population study and cutoff determination. Clin Biochem 2009; 42:877-84.

(6.) Vogel BH, Bradley SE, Adams DJ, D'Aco K, Erbe RW, Fong C, et al. Newborn screening for X-linked adrenoleukodystrophy in New York State: diagnostic protocol, surveillance protocol and treatment guidelines. Mol Genet Metab 2015; 114:599-603.

(7.) Li Y, Scott CR, Chamoles NA, Ghavami A, Pinto BM, Turecek F, Gelb MH. Direct multiplex assay of lysosomal enzymes in dried blood spots for newborn screening. Clin Chem 2004; 50:1785-96.

(8.) Zhang XK, Elbin CS, Chuang WL, Cooper SK, Marashio CA, Beauregard C, Keutzer JM. Multiplex enzyme assay screening of dried blood spots for lysosomal storage disorders by using tandem mass spectrometry. Clin Chem 2008; 54:1725-8.

(9.) Orsini JJ, Martin MM, Showers AL, Bodamer OA, Zhang XK, Gelb MH, Caggana M. Lysosomal storage disorder 4+1 multiplex assay for newborn screening using tandem mass spectrometry: application to a small-scale population study for five lysosomal storage disorders. Clin Chim Acta 2012; 413:1270-3.

(10.) DeJesusVR, Zhang XK, Keutzer J, Bodamer OA, Muhl A, Orsini JJ, et al. Development and evaluation of quality control dried blood spot materials in newborn screening for lysosomal storage disorders. Clin Chem 2009; 55:158-64.

(11.) Turgeon CT, Moser AB, Morkrid L, Magera MJ, Gavrilov DK, Oglesbee D, et al. Streamlined determination of lysophosphatidylcholines in dried blood spots for newborn screening of X-linked adrenoleukodystrophy. Mol Genet Metab 2015; 114:46-50.

(12.) Hubbard WC, Moser AB, Liu AC, Jones RO, Steinberg SJ, Lorey F, et al. Newborn screening for X-linked adrenoleukodystrophy (X-ALD): validation of a combined liquid chromatography-tandem mass spectrometric (LC-MS/MS) method. Mol Genet Metab 2009; 97: 212-20.

(13.) De Jesus VR, Zhou H, Vogt RF. Dried blood spot quality control materials for newborn screening to detect lysosomal storage disorders. Clin Chem 2013; 59:1275-6.

(14.) Chamoles NA, Blanco M, Gaggioli D. Diagnosis of alpha-L-iduronidase deficiency in dried blood spots on filter paper: the possibility of newborn diagnosis. Clin Chem 2001; 47:780-1.

(15.) Weidemann F, Niemann M, Sommer C, Beer M, Breunig F, Wanner C. Interdisciplinary approach towards female patients with Fabry disease. Eur J Clin Invest2012; 42:455-62.

(16.) Stevens S. Filter paper comparison study. Newborn Screening Quality Assurance Program. Atlanta: CDC; 2009. p 1-15.

(17.) Morkrid L, Rowe AD, Elgstoen KB, Olesen JH, Ruijter G, Hall PL, et al. Continuous age- and sex-adjusted reference intervals of urinary markers for cerebral creatine deficiency syndromes: a novel approach to the definition of reference intervals. Clin Chem 2015; 61:760-8.

(18.) Hall PL, MarquardtG, McHugh DM, Currier RJ, Tang H, Stoway SD, Rinaldo P. Postanalytical tools improve performance of newborn screening by tandem mass spectrometry. Genet Med 2014; 16:889-95.

(19.) Marquardt G, Currier R, McHugh DM, Gavrilov D, Magera MJ, Matern D, et al. Enhanced interpretation of newborn screening results without analyte cutoff values. Genet Med 2012; 14:648-55.

(20.) Turgeon CT, Orsini JJ, Sanders KA, Magera MJ, Langan TJ, Escolar ML, et al. Measurement of psychosine in dried blood spots-a possible improvement to newborn screening programs for Krabbe disease. J Inherit Metab Dis 2015; 38:923-9.

(21.) Johnson B, Mascher H, Mascher D, Legnini E, Hung CY, Dajnoki A, et al. Analysis of lyso-globotriaosylsphingosine in dried blood spots. Ann Lab Med 2013; 33:274-8.

(22.) Raymond KM, Turgeon C, Ory D, Lourenco C, Giugliani R, Rinaldo P, et al. Combined analysis of plasma oxysterol and lysosphingomyelin for Niemann-Pick types A, Band C diagnosis. J Inherit Metab Dis 2015; 38:S36.

(23.) de Ruijter J, de Ru MH, WagemansT, Ijlst L, Lund AM, Orchard PJ, et al. Heparan sulfate and dermatan sulfate derived disaccharides are sensitive markers for newborn screening for mucopolysaccharidoses types I, II and III. Mol Genet Metab 2012; 107:705-10.

(24.) Gelb MH, Scott CR, Turecek F. Newborn screening for lysosomal storage diseases. Clin Chem 2015; 61: 335-46.

Silvia Tortorelli, [1] Coleman T. Turgeon, [1] Dimitar K. Gavrilov, [1] Devin Oglesbee, [1] Kimiyo M. Raymond, [1] Piero Rinaldo, [1] and Dietrich Matern [1] *

[1] Biochemical Genetics Laboratory, Department of Laboratory Medicine and Pathology, Mayo Clinic, Rochester, MN.

* Address correspondence to thisauthorat: Biochemical Genetics Laboratory, Department of Laboratory Medicine and Pathology, Mayo Clinic, 200 First St. SW, Rochester, MN 55905. Fax 507-266-2888; e-mail matern@mayo.edu.

Received February 17, 2016; accepted May 26, 2016.

Previously published online at DOI: 10.1373/clinchem.2016.256255

[2] Nonstandard abbreviations: OMIM, Online Mendelian Inheritance in Man; MPS I, mucopolysaccharidosis type I; ALD, X-adrenoleukodystrophy; LSD, lysosomal storage disorders; NPA/B, Niemann-Pick type A and B disease; DBS, dried blood spot; PBD, peroxisomal biogenesis disorder; LPC, lysophosphatidylcholines; IS, internal standard; S, substrate; BP, base pool; P, product; ABG, [beta]-glucosidase; GAA, acid [alpha]-glucosidase; GALC, galactocerebrosidase; ASM, acid sphingomyelinase; GLA, [alpha]-galactosidase A; IDUA, [alpha]-L-iduronidase; FIA-MS/MS, flow injection tandem mass spectrometry; LC, liquid chromatography.
Table 1 Assay performance based on prospective clinical testing of 578
patients. (a)

                                             TP
Condition          Count   LTFU (b)   (age range, years)   FP   FN

All                 578    18         24 (0.02-62.1)       23   0
Normal              430
Artifact (c)         63
POX                   3    0          3(0.6-15.7)          0    0
Fabry (m)             8    1          1 (17.8)             2    0
GLA gene carrier      5    0          1 (30.6)             4    0
Pompe                11    2          4(0.4-62.1)          2    0
NPA/B                 8    1          3 (0.4-2.2)          4    0
Krabbe               16    2          3(0.5-1.9)           4    0
Gaucher              30    12         8(0.02-10.1)         7    0
MPS I                 1    0          1 (6.4)              0    0

Condition          TN    TOT--LTFU   PPV, %   NPV, %   FPR, %

All                513   560          51.1     100      4.1
Normal
Artifact (c)
POX                --    3           100        --       0
Fabry (m)          --    7            33.3      --      28.5
GLA gene carrier   --    5            20.0      --      80.0
Pompe              --    9            66.7      --      22.2
NPA/B              --    7            42.9      --      57.1
Krabbe             --    14           42.9      --      28.6
Gaucher            --    18           53.3      --      38.9
MPS I              --    1           100        --       0

(a) Excluding the 18 cases lost to follow-up overall clinical
sensitivity was 100% and clinical specificity 95.7%.

(b) LTFU, lost to follow up; TP, true positive; FP, false positive;
FN, false negative; TN, true negative; TOT, total cases; PPV, positive
predictive value; NPV, negative predictive value; FPR, false-positive
rate [FP/(TOT--LTFU)]; POX, X-linked ALD or peroxisomal biogenesis
disorder; m, male.

(c) Artifacts were those samples that yielded multiple low enzyme
activities most likely due to heat exposure of the samples during
transport to the laboratory.
COPYRIGHT 2016 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2016 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Endocrinology and Metabolism
Author:Tortorelli, Silvia; Turgeon, Coleman T.; Gavrilov, Dimitar K.; Oglesbee, Devin; Raymond, Kimiyo M.;
Publication:Clinical Chemistry
Date:Sep 1, 2016
Words:4787
Previous Article:Multiplex Droplet Digital PCR Quantification of Recurrent Somatic Mutations in Diffuse Large B-Cell and Follicular Lymphoma.
Next Article:Analytical Bias Exceeding Desirable Quality Goal in 4 out of 5 Common Immunoassays: Results of a Native Single Serum Sample External Quality...
Topics:

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters