Role of the internal superior laryngeal nerve in the motor responses of vocal cords and the related voice acoustic changes.

Intoduction

The scientific application of surface electrical stimulation (SES) to treat dysphagia and especially voice disorders is new. (1) LaGorio (2) used VitalStim to treat swallowing disorder in a single individual and reported that, in conjunction with improvement in dysphagia symptoms, the voice functions improved. Voice is one of the most important instruments for communication. (3) Voice has some acoustic parameters that play an important role in its characterization as normal or abnormal, including fundamental frequency (F0), intensity, shimmer, jitter, and harmonic-to-noise ratio (HNR). (4-7) Accordingly, it is essential that new methods to treat voice impairments be devised.

SES studies were commenced on the subject of dysphagia and its resultant voice changes. (2) Nonetheless, the unknown details about the procedure and its effectiveness prompted researchers to investigate SES effects on normal subjects' voice. The existing literature, however, suffers from a dearth of published data on this topic. In addition to SES studies, Kempster (8) applied implanted electrical stimulation in normal individuals' internal laryngeal muscles, which resulted in increased F0. Although implanted electrical stimulation evokes laryngeal muscles more specifically, it is invasive and not widely applicable. Some researchers were, therefore, prompted to draw upon SES. (9-13)

Previous studies have demonstrated that SES is ineffective in causing changes in acoustic voice parameters and also in closing the vocal folds. (9,11,13-15) Flower (10) found that SES resulted in statistically significant unfavorable effects on voice parameters in experimental participants.

According to the published studies on SES in normal individuals, generally SES is not effective in inducing changes in voice parameters. This finding may be attributed to the application of SES protocols, which are designed for dysphagia therapy and are not useful for voice and larynx specifically. Consequently, this method cannot induce a remarkable effect on voice in normal subjects, although we should not discount SES as an important agent.

A possible method to impose changes on the internal laryngeal muscles and create subsequent acoustic changes may be the use of the potential capability of the internal superior laryngeal nerve (ISLN) to move the internal laryngeal muscles. Contrary to earlier studies that knew only a sensory role for the ISLN, some neuroanatomical studies in the last 3 decades have found that the ISLN in addition to its sensory role also contributes to the movements of the internal laryngeal muscles by its distributed motor branches to those muscles. (16-19) However, in awake humans, the literature contains only 2 studies that stimulated the ISLN with needle electrodes inserted around the ISLN and induced motor reactions, which were monitored via flexible laryngeal endoscopy and also electromyographic recordings. The results indicated that ISLN stimulation caused movement and electromyographic responses in the internal laryngeal muscles. (20,21) Although the findings asserted the motor role of the ISLN, (20, 21) the stimulus instrument used (i.e., needle electrodes) are invasive and not applicable at clinics for repeated efforts.

Accordingly, in this pre-experimental study, we employed SES as a noninvasive, cost-effective, and repeatable stimulus modality, to evoke the ISLN in awake subjects and then recorded the reaction of their internal laryngeal muscles in terms of the resultant acoustic changes. To our knowledge, the present study is the first research to apply SES in order to evoke vocal cords through the ISLN in awake normal humans. The aim of this study, as part of a larger research project, was to investigate the probability of utilizing the motor potentials of the ISLN on the vocal cords via SES, which can be used in future studies as an electrotherapy modality to treat patients with recurrent laryngeal damage.

Materials and Methods

The current study was performed on the students of School of Rehabilitation Sciences, Iran University of Medical Sciences in 2014. Twenty-two female and 10 male individuals (n=32) between 20 and 33 years old (mean age=23.87) participated in this study. All the subjects were selected according to the accessible sampling method, although participation was on a completely voluntary basis. Because of the lack of a previous similar study, the sample size was defined according to the data gathered from our study's 10 pilot subjects. According to the applied formula to calculate the sample size presented below, the sufficient number was 18 persons:

n = [[([Z.sub.1-[alpha]/2 + [Z.sub.1-[beta]]).sup.2] [[sigma].sup.2.sub.d]]/[d.sup.2] + (Z.sub.1-[alpha]/2])/2]

First, all the subjects completed a consent form approved by the Ethics Committee of Iran University of Medical Sciences. All the study participants were nonsmokers and reported a negative history of neck and head surgery and neurological, respiratory, serious cardiac, psychiatric, swallowing, hearing loss, voice, and speech disorders. There were also no systemic diseases such as diabetes. According to the subjects' negative self-reported voice complaints and the perceptual judgment of a speech-language pathology expert, who was experienced in voice disorder evaluation and treatment, all of the participants' voice was diagnosed normal at the time of test. In addition, during the study, any subject who could not tolerate a gradual rise in SES intensity was excluded from the study. Likewise, inability to prolong the/a/vowel at least for 10 seconds without any changes in voice acoustic parameters led to the exclusion of the subject from the study.

Procedure

SES was presented to the subjects via 2 self-adhesive circular active monopolar silver electrodes, 0.7 cm in diameter (F-55). They were placed on the thyrohyoid membrane between the hyoid bone and the thyroid cartilage at the entrance location of the ISLN into the larynx on both sides. Additionally, one 6*5 cm rectangular farther dispersive electrode was used on the back of the neck. The electrical current was generated by a commercially available electrotherapy system (ELPHA II 3000, Danmeter A/S, Denmark). The system was set to produce a 3-second current with 5-second rests between each 2 stimuli. The current parameters were programmed on a 200-microsecond duration of pulses and a frequency of 100 Hz. Amplitude was started from 0 mA and increased gradually according to the participant's tolerance. Before electrode placement, the skin of all the subjects was cleaned with alcohol. Additionally, in some males, the extra hair on the site of electrode placement was shaved.

Controlling volume conduction (diffusing the current to the contiguous tissues, especially the vocal cords and the laryngeal recurrent nerve) was a challenge. Thus, after the termination of ISLN stimulation via electrodes, the 2 active electrodes were once again placed on the thyroid lamina (the nearest place to the vocal cords and the recurrent nerve) and the previous current was repeated precisely. The objective was to define the amount of the direct conduction of the current to the vocal cords and the recurrent nerve and to exclude evoked ISLN effects.

Voice Recordings

As the intensity of the stimulant reached up to 10% of the subjects' ultimate tolerance, they were asked to prolong the/a/vowel for 13 seconds: 5 seconds without stimulation (rest period), 3 seconds with stimulation (SES period), and finally another 5 seconds' rest period. The second rest period was meant to coordinate the time of phonation with the rest periods of the electrical stimulation. Hence, only 2 seconds of phonation from the mid-point of the first rest period and the SES period were compared with each other. The subjects were free to practice during stimulation and were instructed to prolong a comfortable and consistent natural vowel. Voice recordings were obtained using a condenser microphone (Rode NT6, Australia), connected to a preamplifier (M-Audio, Avid technology, Inc.). The distance of mouth to microphone was 20 centimeters. The collected voice samples were analyzed using Praat (version 5.3.13; http:// www.fon.hum.uva.nl/praat/download_win.html). The acoustic parameters digitized comprised F0, standard deviation of F0 (SD F0), shimmer, jitter, voice break, HNR, and intensity of voice.

Statistical Analysis

The digitized acoustic parameters of rest and SES periods were analyzed using SPSS (version 17) and were compared in terms of the induced changes across the 2 conditions of phonation. The paired t-test was utilized to Wcompare the means oWf the acoustic parameters, namely mean F0, minimum F0, maximum F0, SD F0, jitter, shimmer, HNR, mean intensity, maximum intensity, and minimum intensity at rest and SES periods.

Results

Figure 1 demonstrates mean F0, maximum (max) F0, and minimum (min) F0 for the rest and SES periods related to ISLN electrode placement. The paired sample t-test showed no statistically significant effects of SES on mean (m) F0. As the figure indicates, mean F0 for rest was 198.35 (SD=49.93) while it was 193.97 (SD=50.15) for the SES period, which did not constitute significant statistical difference (t=1.03; P=0.31). Also, the means of max F0 during rest (m=202.28, SD=50.78) and SES (m=203.93, SD=52.64) were not significantly different (t=0.35; P=0.72). However, the mean of min F0 for the rest time was 194.54 (SD=49.19) versus 182.62 (SD=53.39) for SES, duration which the difference between the 2 means was significant (t=2.61; P=0.01).

The SD F0 means of the rest and SES durations related to electrode placement on the ISLN are shown in figure 2. The rest SD F0 mean was 1.54 (SD=0.55) versus 4.15 (SD=3.00) for SES. The paired t-test indicated significant differences between the SD F0 means (t=-4.95; P<0.001).

Table 1 demonstrates jitter, shimmer, HNR, mean intensity, maximum intensity, and minimum intensity results related to electrode placement on the ISLN. As the table presents, the means of rest compared to those of the SES period for both jitter and shimmer increased: The changes in the means were 0.09 and 0.9, respectively, which were statistically significant (jitter: P<0.001 and shimmer: P<0.001). HNR mean during rest versus SES decreased by 2.21 dB, with the change being significant according to the paired f-test (P<0.001). However, there were only 2 participants who demonstrated phonation break during the SES phase.

The means of the 3 token values of intensity decreased from rest to the SES phase. Mean intensity changed 1.48, minimum intensity 3.99, and maximum intensity 0.38. The paired f-test indicated that the changes in both mean intensity and minimum intensity were statistically significant (P<0.001), whereas the changes in maximum intensity were not statistically significant (f=0.92; P=0.36).

Table 2 demonstrates the acoustic changes created by SES when active electrodes were placed on both sides of the thyroid lamina. As the table shows, the means of all the investigated acoustic parameters did not differ significantly between the rest and SES periods.

Discussion

The present study examined the possibility of imposing changes on the vocal cords via SES on ISLN topical location on skin. The results showed that an evoked ISLN did not give rise to a significant change in mean F0. Given that the contraction of the cricothyroid muscle heightens mean F0, it can be concluded that there were no descending branches or anastomoses from the ISLN to the cricothyroid muscle, which is in agreement with the results of previous anatomical studies. (16-17) Also, as the stimulation did not evoke the external SLN, which innervates the cricothyroid muscle, the stimulated location on skin was correct. It is not possible to exclude the external SLN from stimulation as Ludlow et al. (20) asserted in their study even after using needle electrodes around the ISLN. Nevertheless, in our study, as SES did not lead to an increase in mean F0, SES effects on the external SLN were too little to be taken into account. The same result was observed for maximum F0, which again confirms the above conclusions. This result is in accordance with the previous studies that indicated insignificant effects of SES on mean F0. (9,13) However, those studies reported increased F0 after SES, which may be due to their larger electrodes and consequent decreased current density (22) distributed to the external laryngeal muscles. SES resulted in decreased minimum F0 compared to the rest state. This may have been due to the effects of SES on the thyroarytenoid muscle via the stimulation of the ISLN, lowering the frequency of voice. (23) SD F0 was increased by SES. By taking into consideration that SD F0 reflects frequency variability for a large time segment, (24) it can be concluded that SES plays an important role in imposing variability on muscles and thus contributes to muscular tension stability. Our findings do not chime in with the results of the previous studies performed on the effects of SES on voice parameters inasmuch as in none of those investigations SD F0 and minimum F0 changed significantly after SES. (9-11, 13)

Perturbation, including jitter and shimmer, is another acoustic factor which increased during the SES phase. The irregularity of the vibratory behavior of the vocal cords is referred to as perturbation. (24-25) As a result, SES led to higher irregularity during the vibration of the vocal cords. This finding is not in accordance with the previous studies insofar as they did not show significant differences in their subjects' perturbation of voice. (13) Therefore, it is impossible to create perturbation changes via SES directly throughout the thyroid cartilage and into the vocal cords. Presumably, it is because of the lack of stimulus penetration due to incorrect electrode placement and electrical stimulation parameters. Nonetheless, our findings demonstrated that it was possible through the stimulation of the vocal cords directly via the ISLN.

Greater aperiodicity of vibration may show up as greater noise in the spectrum and lesser HNR. (26, 27) In the present study, HNR decreased during the SES period. As a result, heightening of aperiodicity can be concluded consequent to electrical stimulation on the ISLN. Gorham-Rowan et al. (13) reported similar results regarding the change direction of jitter, shimmer, and HNR after SES, but the changes were not statistically significant. It may be attributed to the differences in methodology between the study by Gorham-Rowan et al. and the present study.

In spite of these reported aperiodicity and instability in the participants' voice during the SES period, SES was able to cause only voice break in 2 subjects. Ludlow CL et al. (20) stimulated the ISLN by inserting needle electrodes around the ISLN for normal awake participants and reported voice break during phonation in 1 of their 3 subjects. The authors attributed it to increased vocal fold adduction, which disrupted vibration. Similarly in our research, voice breaks occurred in 2 subjects but the other 30 participants did not show any voice break during phonation at SES time. This suggests that SES, unlike needle electrode stimulation, is not strong enough to interrupt the phonation process completely.

In addition to frequency and perturbation, the results showed that while mean and minimum intensity decreased with SES, maximum intensity decreased but it was not statistically significant. As the average intensity is a criterion to present the strength of the vibratory behavior of the vocal cords, (24) the effects of SES in terms of reducing vocal intensity may be due to glottal closure interruption consequent to SES presentation. This result was in contrast to the results of a study by Flower and et al., (9) who did not report significant changes in voice intensity after SES in healthy individuals.

There are some reasons for reporting nonsignificant changes in acoustic parameters after using SES in the previous studies. (9-11,13,14) The most important reason relates to the electrical stimulation parameters applied in those studies compared to our own. The previous investigations applied an electrical current with a low frequency (80 Hz) and long duration pulses (700 microseconds) compared to our study with a high pulse frequency (100 Hz) and short duration pulses (200 microseconds). It is possible to increase the intensity of pulses with short durations and high frequencies well above the motor threshold without stimulating pain endings. (22) Stimulation of the deeply placed nerves (such as the ISLN) will, therefore, be feasible. (22) Accordingly, the current method applied in the present research had more power to penetrate deeply and reach the ISLN.

The other reasons include different methods in electrode placement, different stimulated organs (internal laryngeal muscles versus ISLN in our study), different sizes of electrodes, and dissimilar time of recording voice (after or during SES presentation, as was the case in the present study).

As the results of electrode placement on the thyroid lamina indicated, SES did not bring about significant changes in any of the acoustic parameters. This finding confirms the notion that by dispossessing and excluding the ISLN from the SES-affected area and also by delivering the current at the nearest place to the vocal cords and the recurrent nerve, SES cannot stimulate the vocal cords or the recurrent nerve. It also shows that all the findings on delivering SES with electrodes placed on the thyrohyoid hint at the stimulation of the ISLN and not the direct conduction of the current to the vocal cords or the recurrent nerve. The results of electrode placement on the thyroid lamina are in agreement with those reported by the previous studies whose method of electrode placement was aimed at stimulating the vocal cords directly. (9-11,13) This confirms the importance of correct electrode placement when seeking to impact voice through SES.

Although the findings on perturbation and intensity suggested exacerbation of voice during SES presentation, they indicated the effectiveness of SES in impacting the internal laryngeal muscles from the ISLN as a gate for motor responses in the vocal cords. The results of electrode placement on the thyroid lamina indicated the impossibility of stimulating the recurrent laryngeal nerve by SES. Therefore, the clinical application of this finding would see SES utilized to evoke the motor potential of the ISLN in order to treat patients suffering from recurrent laryngeal nerve damage.

The current study has some limitations, first and foremost among which were problems in fixing electrodes on a defined location, not least when the subjects sweated and the test had to be repeated after cleaning the skin. In addition, some subjects feared the first sensation of tingling caused by SES and were, thus, excluded from the study.

Conclusion

The results of the present study demonstrated the significant impact of SES on voice changes in healthy participants. To the best of our knowledge, ours is the first study of its kind to report significant voice changes in consequence of SES application. The reason for the dissimilarities between our results and those of the previous studies may be attributed to the methodology of stimulation and the time of recording voice. Our results indicated that the application of this method via the ISLN at least in healthy people could induce motor pulses in the vocal cords and cause voice changes. The importance of this finding is that it hints at the possibility of SES application in patients who need to have the motor neurons of their vocal folds stimulated when there is no accessibility owing to damage to the recurrent nerve (e.g., patients with vocal paralysis). Confirmation of the applicability of this method, however, requires further research.

What's Known

* Neuroanatomic studies have indicated axons descending from the internal superior laryngeal nerve to the internal laryngeal muscles.

What's New

* Previous attempts to impose significant changes on acoustic parameters through surface electrical stimulation (SES) on the larynx have all failed.

* We are the first to significantly change acoustic parameters (e.g., shimmer, jitter, harmonic-to-noise ratio, and intensity of voice) by applying SES on the vocal folds.

Acknowledgement

We appreciate all the participants who assisted us in performing this study.

Conflict of Interest: None declared.

References

(1.) Freed ML, Freed L, Chatburn RL, Christian M. Electrical stimulation for swallowing disorders caused by stroke. Respir Care. 2001;46:466-74. PubMed PMID: 11309186.

(2.) Lagorio LA, Carnaby-Mann GD, Crary MA. Cross-system effects of dysphagia treatment on dysphonia: a case report. Cases J. 2008;1:67. doi: 10.1186/1757-1626-1-67. PubMed PMID: 18667069; PubMed Central PMCID: PMC2515091.

(3.) Seifpanahi S, Jalaie S, Nikoo MR, Sobhani-Rad D. Translated Versions of Voice Handicap Index (VHI)-30 across Languages: A Systematic Review. Iran J Public Health. 2015;44:458-69. PubMed PMID: 26056664; PubMed Central PMCID: PMC4441958.

(4.) Naufel de Felippe AC, Grillo MH, Grechi TH. Standardization of acoustic measures for normal voice patterns. Braz J Otorhinolaryngol. 2006;72:659-64. PubMed PMID: 17221059.

(5.) Seifpanahi S, Bakhtiar M, Salmalian T. Objective vocal parameters in Farsispeaking adults with Down syndrome. Folia Phoniatr Logop. 2011;63:72-6. doi: 10.1159/000316326. PubMed PMID: 20926889.

(6.) Eadie TL, Stepp CE. Acoustic correlate of vocal effort in spasmodic dysphonia. Ann Otol Rhinol Laryngol. 2013;122:169-76. doi: 10.1177/000348941312200305. PubMed PMID: 23577569.

(7.) Rothman HB, Brown WS, Jr., Sapienza CM, Morris RJ. Acoustic analyses of trained singers perceptually identified from speaking samples. J Voice. 2001;15:25-35. doi: 10.1016/S0892-1997(01)00004-2. PubMed PMID: 12269632.

(8.) Kempster GB, Larson CR, Kistler MK. Effects of electrical stimulation of cricothyroid and thyroarytenoid muscles on voice fundamental frequency. J Voice. 1988;2:221-9.

(9.) Fowler LP, Gorham-Rowan M, Hapner ER. An exploratory study of voice change associated with healthy speakers after transcutaneous electrical stimulation to laryngeal muscles. J Voice. 2011;25:54-61. doi: 10.1016/j.jvoice.2009.07.006. PubMed PMID: 20080024.

(10.) Fowler LP, Awan SN, Gorham-Rowan M, Morris R. Investigation of fatigue, delayedonset muscle soreness, and spectral-based cepstral measurements in healthy speakers after neuromuscular electrical stimulation. Ann Otol Rhinol Laryngol. 2011;120:641-50. PubMed PMID: 22097149.

(11.) Yan JCW [Internet]. Effects of surface electrical stimulation on voice production in non-dysphonic adults. The University of Hong Kong. 2009. Available from: http://hdl. handle.net/10722/173661.

(12.) Kagaya H, Baba M, Saitoh E, Okada S, Yokoyama M, Muraoka Y Hyoid bone and larynx movements during electrical stimulation of motor points in laryngeal elevation muscles: a preliminary study. Neuromodulation. 2011;14:278-83. doi: 10.1111/j.1525-1403.2011.00331.x. PubMed PMID: 21992253.

(13.) Gorham-Rowan M, Fowler L, Hapner E. Acoustic analysis of voice change in normal speakers following transcutaneous electrical stimulation to the laryngeal area. Open Rehabil J. 2010;3:67-74.

(14.) Hapner ER, Gorham-Rowan M, Morris RJ, Fowler LP. S190-Voice Spectral Analysis After Surface Electrical Stimulation. Otolaryngology--Head and Neck Surgery. 2008;139:P139.

(15.) Humbert IA, Poletto CJ, Saxon KG, Kearney PR, Ludlow CL. The effect of surface electrical stimulation on vocal fold position. Laryngoscope. 2008;118:14-9. doi: 10.1097/MLG.0b013e318155a47d.

(16.) Sanders I, Mu L. Anatomy of the human internal superior laryngeal nerve. Anat Rec. 1998;252:646-56. doi: 10.1002/(SICI)10970185(199812)252:4%3C646:AID AR15%3E3.0.CO;2-E. PubMed PMID: 9845215.

(17.) Maranillo E, Leon X, Orus C, Quer M, Sanudo JR. Variability in nerve patterns of the adductor muscle group supplied by the recurrent laryngeal nerve. Laryngoscope. 2005;115:358-62. doi: 10.1097/01. mlg.0000154745.78808.02. PubMed PMID: 15689766.

(18.) Krammer EB, Streinzer W, Millesi W, Ellbock E, Zrunek M. Central localization of motor components in the internal branch of the superior laryngeal nerve: a horseradish peroxidase study in the rat. Laryngol Rhinol Otol (Stuttg). 1986;65:617-20. PubMed PMID: 3807597.

(19.) Wustrow TP, Martin F, Hollander H. Auxiliary motor innervation of the laryngeal muscles via the internal branch of the superior laryngeal nerve. Laryngol Rhinol Otol (Stuttg). 1988;67:275-81. PubMed PMID: 3043125.

(20.) Ludlow CL, Van Pelt F, Koda J. Characteristics of late responses to superior laryngeal nerve stimulation in humans. Ann Otol Rhinol Laryngol. 1992;101:127-34. doi: 10.1177/000348949210100204. PubMed PMID: 1739256.

(21.) Barkmeier JM, Bielamowicz S, Takeda N, Ludlow CL. Modulation of laryngeal responses to superior laryngeal nerve stimulation by volitional swallowing in awake humans. J Neurophysiol. 2000;83:1264-72. PubMed PMID: 10712454.

(22.) Robertson V, Ward A, Low J, Reed A. Electrotherapy explained: principles and practice. In: Electrical stimulation - currents and parameters. 4th ed. Butterworth Heinemann: Elsevier; 2006. p. 45-89.

(23.) Sataloff RT. Clinical assessment of voice. Oxford: Plural publishing; 2005.

(24.) Colton RH, Casper JK. Understanding voice problems: a physiological perspective for diagnosis and treatment. In: Differential diagnosis of voice problems. Philadelphia: Lippincott Williams & Wilkins; 2006. p: 12-63.

(25.) Shao J, MacCallum JK, Zhang Y, Sprecher A, Jiang JJ. Acoustic analysis of the tremulous voice: assessing the utility of the correlation dimension and perturbation parameters. J Commun Disord. 2010;43:35-44. doi: 10.1016/j.jcomdis.2009.09.001. PubMed PMID: 19909966; PubMed Central PMCID: PMC2813418.

(26.) Ferrand CT. Harmonics-to-noise ratio: an index of vocal aging. J Voice. 2002;16:480-7. doi: 10.1016/S0892-1997(02)00123-6. PubMed PMID: 12512635.

(27.) Olszewski AE, Shen L, Jiang JJ. Objective methods of sample selection in acoustic analysis of voice. Ann Otol Rhinol Laryngol. 2011;120:155-61. doi: 10.1177/000348941112000303. PubMed PMID: 21510140; PubMed Central PMCID: PMC3212407.

Sadegh Seifpanahi [1], PhD; Farzad Izadi [2], MD; Ali-Ashraf Jamshidi [3], PhD; Farhad Torabinezhad [1], PhD; Javad Sarrafzadeh [3], PhD; Siavash Mohammadi [1], MS

[1] Department of Speech and Language Pathology, School of Rehabilitation Sciences, Iran University of Medical Sciences, Tehran, Iran;

[2] Division of Laryngology, Rasoul Akram Medical Complex, Iran University of Medical Sciences, Tehran, Iran;

[3] Department of Physical Therapy, School of Rehabilitation Sciences, Iran University of Medical Sciences, Tehran, Iran

Correspondence:

Ali-Ashraf Jamshidi, PhD; Department of Physical Therapy, School of Rehabilitation Sciences, Shahnazari St., Mirdamad Blvd., P.O. Box: 15875-4391, Tehran, Iran

Tel/Fax: +98 21 22228051-2 (345)

Email: alijamshidipt@gmail.com

Received: 28 April 2015

Revised: 01 July 2015

Accepted: 02 August 2015
Table 1: Means, t-statistics (t), and P values of jitter, shimmer,
harmonic-to-noise ratio (HNR), mean intensity, maximum intensity, and
minimum intensity at rest and surface electrical stimulation (SES)
periods related to electrode placement on the internal superior
laryngeal nerve

                          Rest               SES                 t

Jitter (%)                0.31 (SD: 0.11)    0.4 (SD: 0.13)      3.3
Shimmer (%)               2.91 (SD: 1.03)    3.81 (SD: 1.41)     3.49
HNR (dB)                  21.68 (SD: 3.26)   19.46 (SD: 2.8)     5.82
Mean intensity (dB)       62.02 (SD: 7.67)   60.53 (SD: 7.75)    3.99
Minimum intensity (dB)    59.81 (SD: 8.35)   55.81 (SD: 10.73)   3.42
Maximum intensity (dB)    64.03 (SD: 7.24)   63.65 (SD: 7.47)    0.92

                          P value

Jitter (%)                <0.001
Shimmer (%)               <0.001
HNR (dB)                  <0.001
Mean intensity (dB)       <0.001
Minimum intensity (dB)    <0.001
Maximum intensity (dB)    0.36

Table 2: Means, t-statistics (t), and P values of mean fundamental
frequency (F0), minimum F0, maximum F0, F0 standard deviation (SD F0),
jitter, shimmer, harmonic-to-noise ratio (HNR), mean intensity,
maximum intensity, and minimum intensity at rest and during the
surface electrical stimulation (SES) period related to electrode
placement on the thyroid cartilage

                         Rest                 SES

Mean F0 (Hz)             197.52 (SD: 50.02)   197.46 (SD: 50.69)
Minimum F0 (Hz)          194.02 (SD: 48.93)   193.79 (SD: 49.13)
Maximum F0 (Hz)          201.47 (SD: 50.71)   201.28 (SD: 50.75)
SD F0 (Hz)               1.57 (SD: 0.55)      1.6 (SD: 0.57)
Jitter (%)               0.3 (SD: 0.11)       0.3 (SD: 0.11)
Shimmer (%)              2.93 (SD: 1.01)      2.94 (SD: 1.1)
HNR (dB)                 21.47 (SD: 3.15)     21.43 (SD: 3.13)
Mean intensity (dB)      61.75 (SD: 7.4)      61.74 (SD: 7.47)
Minimum intensity (dB)   58.71 (SD: 8.54)     58.77 (SD: 8.51)
Maximum intensity (dB)   63.41 (SD: 7.26)     63.41 (SD: 7.16)

                         t       P value

Mean F0 (Hz)             0.21    0.83
Minimum F0 (Hz)          0.67    0.5
Maximum F0 (Hz)          0.47    0.64
SD F0 (Hz)               0.62    0.53
Jitter (%)               0.3     0.76
Shimmer (%)              0.21    0.82
HNR (dB)                 0.46    0.64
Mean intensity (dB)      0.05    0.95
Minimum intensity (dB)   0.74    0.46
Maximum intensity (dB)   0.02    0.98

Figure 1: The figure presents mean fundamental frequency (F0),
maximum F0, and minimum F0 for the rest and surface
electrical stimulation (SES) periods related to electrode
placement on the internal superior laryngeal nerve.

                  Rest         SES

Mean FO          198.35      193.97
Minimum FO       194.54      182.62
Maximum FO       202.28      203.93

Figure 2: The figure presents the means of the standard
deviation of fundamental frequency (SD F0 in Hz) at
rest and surface electrical stimulation (SES) durations
related to electrode placement on the internal superior
laryngeal nerve.

Rest    4.15 (HZ)
SES     1.54 (HZ)

Note: Table made from bar graph.

Figure 2: The figure presents the means of the standard deviation
of fundamental frequency (SD F0 in Hz) at rest and surface
electrical stimulation (SES) durations related to electrode
placement on the internal superior laryngeal nerve.

SES    4.15 (HZ)
REST   1.54 (HZ)

Note: Table made from pie chart.
COPYRIGHT 2016 Shiraz University of Medical Sciences
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2016 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Original Article
Author:Seifpanahi, Sadegh; Izadi, Farzad; Jamshidi, Ali-Ashraf; Torabinezhad, Farhad; Sarrafzadeh, Javad; M
Publication:Iranian Journal of Medical Sciences
Article Type:Report
Geographic Code:7IRAN
Date:Sep 1, 2016
Words:4821
Previous Article:Prevalence of the most common virulence-associated genes among Brucella Melitensis isolates from human blood cultures in Hamadan Province, West of...
Next Article:SDS-PAGE analysis of the outer membrane proteins of uropathogenic Escherichia coli isolated from patients in different wards of Nemazee Hospital,...
Topics:

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters