Reactivity of urinary albumin (microalbumin) assays with fragmented or modified albumin.

Increased urinary excretion of albumin is recognized as an early indicator of glomerular injury in diabetes, before injury is indicated by assays for total urinary protein (1). Recently, albumin excretion, even within the normal range of_30 mg/day, has been identified as an indicator for risk of mortality and cardiovascular disease (2-4). These findings heighten interest in the accurate measurement of urinary albumin within the normal range as well as in the microalbuminuric range (30-300 mg/day or 20-200 mg/g creatinine). At the same time, however, there is increased interest in accurately measuring low concentrations of urinary albumin.

Two issues have recently generated questions and controversy about the accuracy of urinary albumin measurements. First, there has been growing recognition of the occurrence of modified or fragmented forms of albumin in urine. Proportions of modified albumins such as glycated albumin and a C-terminal truncation product are higher in urine than serum, and isoelectric focusing shows that urinary albumin from people with diabetes appears to be more heterogeneous and acidic (5-7). Analysis of urinary proteins by Western blotting of 1-dimensional gels or by 2-dimensional gel electrophoresis shows the presence of large albumin fragments (8-11). Also, multiple small fragments of albumin have been detected in urine (12-14). Although urine may contain modified albumin or a variety of albumin fragments, assays for urine albumin are calibrated using serum albumin, and data regarding the influence of albumin fragments on assays for urine albumin are lacking.

A second source of uncertainty in urine albumin measurement is the lack of agreement of immunoassays for urine albumin with a size-exclusion chromatographic assay, which yields substantially higher values than immunoassays for many specimens from healthy individuals or diabetes patients (15-22). This bias has been hypothesized to result from modified forms of urinary albumin that are detected by the size-exclusion assay but not by immunoassays. Our laboratory, how ever, determined that size-exclusion chromatography does not resolve albumin from a number of other urinary proteins and is an assay for a mixture of albumin-sized molecules (ranging from about 40 000 to 80 000 Da) rather than an assay for albumin (21). This finding offers an alternative explanation for the difference between immunoassay and size-exclusion chromatography but does not completely rule out some contribution from modified forms of albumin with decreased immunoreactivity.

In this study, we sought to gain insights into the cross-reactivity of albumin fragments in immunoassays for urine albumin and to examine whether it is possible to modify albumin so as to markedly reduce its immunoreactivity.

Materials and Methods

Proteins, including human albumin, and reagents were obtained from Sigma-Aldrich, except as noted; cyanogen bromide (CNBr) [1] was from Acros Chemicals and guanidine hydrochloride from Calbiochem. Other chemicals were reagent grade. Gels (4% to 12% gradient BisTris gels), reagents, and SimpleBlue stain for SDS-PAGE were from Invitrogen. Western blotting on nitrocellulose used rabbit antiserum to human albumin and alkaline phosphatase-conjugated sheep antirabbit IgG. After blocking with SuperBlock T20 from Pierce Chemical, blots were visualized with 5-bromo-4-chloro-3-indolyl phosphate/nitroblue tetrazolium substrate.

Total protein (pyrogallol red) and immunoturbidimetric urine albumin assay were performed on the Beckman-Coulter LX20[R] analyzer. A competitive immunoassay for urine albumin analysis was performed on the Siemens Immulite[R] analyzer.

CNBr cleavage of albumin was conducted in 6 mol/L guanidine hydrochloride, 10% acetic acid, and 10 g/L cyanogen bromide (23) for 24 h at room temperature before dialysis vs PBS. Other incubations were conducted with 10 g/L albumin. Citraconylation used 0.1 mol/L citraconic anhydride in 1 mol/L N-methyl-morpholine, pH 9.0. Incubations at 37 [degrees]C with 1% hydrogen peroxide or 10 mmol/L sodium ascorbate were in PBS, pH 7.4. Other incubations were at pH 2 (0.1 mol/L glycine HCl buffer) or pH 8 with 0.5 g/L trypsin (100 mmol/L Tris, 10 mmol/L Ca[Cl.sub.2]). Incubations for 5 days at 37 [degrees]C were in PBS containing 0.01% sodium azide or in acetate-buffered saline, pH 5.7. Additives were 10% glucose, 10 mmol/L sodium nitrite, or 10 mmol/L sodium ascorbate. Modification reactions were diluted with PBS before analysis.

For preparative size-exclusion chromatography of CNBr-digested albumin, we used a Pharmacia fast protein liquid chromatography system with a 1.6 by 60 cm Sephacryl S-200 column eluted at room temperature with PBS. This purified fractions enriched in fragments A and C, identified as residues 1-123 (A) and 299-584 (C) by electrophoresis and mass spectrometry. We performed reversed-phase chromatography of a size-exclusion chromatography fraction enriched in fragment B (residues 124-298) on aWaters 3.9 by 150 mm Symmetry 300 C4 column eluted at 1 mL/min with a linear gradient of acetonitrile and 0.1% trifluoroacetic acid, yielding purified fragment B. MALDI-TOF mass spectrometry of albumin fragments was performed using sinapinic acid as matrix (24).

Urine from 24-h collections submitted to the clinical laboratory were used anonymously as approved by an institutional review board. Specimens for Western blots consisted of 10 freshly collected specimens with [greater than or equal to]300 mg/L protein by dipstick and 5 specimens stored at -20 [degrees]C for 3 years, with albumin concentrations from 100 to 1700 mg/L (by LX20 assay), and 9 random urine specimens with albumin concentrations from 2 to 8 g/L from patients with focal segmental glomerulosclerosis, including 2 with diabetes, obtained from Dr. Jeffrey Kopp (protocol 94KD0127 approved by an institutional review board).

We examined the degree of identity of serum albumin sequences for different species in the UniProt database with the selected ion monitoring alignment tool using sequences in the UniProt database (www.expasy.org).

Results

Albumin was subjected to a variety of treatments that potentially alter its conformation, chemically modify amino acid side chains, or cleave its polypeptide chain. Analysis by SDS-PAGE, with approximately equivalent inputs of protein or peptide in each lane, examined products of the modification reactions (Fig. 1, top panel without reduction and bottom panel with reduction).

SDS-PAGE indicated that almost no intact albumin remained after extensive digestion with trypsin (lane 2) or CNBr (lane 3), which should cleave albumin at 6 methionine residues, yielding 3 major fragments from residues 1-123 (A), 124-298 (B), and 299-585 (C) (25). A and C fragments should contain 1 and 3 internal cleavages, respectively, but are held together by disulfide bonds. The 3 CNBr fragments obtained by chromatographic purification are shown (lanes 4, 5, and 6). Based on amino acid sequence, the C fragment should yield 4 fragments following reduction, but 6 fragments were observed (bottom panel, lane 4), possibly related to incomplete cleavage at some sites or at residue 365, at an acid-labile Asp-Pro bond. Citraconylation of albumin resulted in altered electrophoretic mobility, showing that all albumin molecules were modified (lane 7). Incubation of albumin with sodium ascorbate for 1 day led to cleavage of some molecules, evident mainly on reduction (lane 8, bottom panel), indicating that most cleavages yielded nicked molecules held together by disulfides. Incubation with ascorbate for 5 days yielded extensive fragmentation of albumin, particularly at neutral pH (not shown). Cleavage of albumin by ascorbate possibly results from free radicals generated by trace amounts of metals as described (26).

[FIGURE 1 OMITTED]

We examined how various treatments of albumin affected its reactivity in 2 assays for urinary albumin (Table 1). Exposure to low pH, heat, and chemical modifications, such as citraconylation and exposure to hydrogen peroxide, had only a small effect on the reaction of albumin in the LX20 assay. The total digestion of albumin by CNBr also exhibited little change in assay response vs intact albumin. Even isolated CNBr fragments cross-reacted strongly in the LX20 assay. The only treatment that severely decreased measured albumin was extensive digestion with trypsin. The decrement in measured albumin became pronounced only with extensive digestion of albumin, with a high ratio of trypsin to albumin (1:20 by weight). Trypsin-treated albumin retained >20% cross-reactivity in both urine albumin assays, although electrophoresis indicated that virtually no intact albumin remained; thus trypsin-cleaved albumin cross-reacted at least partially in the assays. Less extensive tryptic digestion yielded larger fragments and less effect on immunoreactivity, as described in a preliminary report (27).

[FIGURE 2 OMITTED]

There is the potential for fragmented antigens to inhibit immunoturbidimetric reactions. To test this possibility, we prepared small tryptic fragments of albumin by ultrafiltration through a 5000-Da cutoff membrane. These fragments had little measurable cross-reactivity in the LX20 assay. Results are not included in Table 1, because total protein content of small peptides could not be standardized using the pyrogallol red assay. Small albumin fragments were mixed with albumin to examine whether the fragments would inhibit the LX20 assay, and no evidence of inhibition was obtained (data not shown).

The Immulite assay generally showed greater sensitivity to modification or fragmentation of albumin than the LX20 assay, except in the case of albumin digested by trypsin, for which severe reduction in cross-reactivity occurred for both assays (Table 1). Several treatments, such as digestion with CNBr, citraconylation, or incubation with ascorbate, glucose, or nitrite, affected the Immulite assay more than the LX20 assay. The greatest difference between assays was evident for albumin cleaved by CNBr and prolonged incubation with ascorbate.

It was considered that modified albumins might change response curves owing to altered affinity for antibodies or loss of reactivity of subsets of reactive epitopes. However, analysis of a series of dilutions of the CNBr digest of albumin indicated that responses were linear throughout the assay ranges (Fig. 2).

Serum albumins from different animal species were examined as another model for structural modification of albumin (Table 2). The LX20 assay, which uses polyclonal goat antiserum, had strong cross-reactivity with baboon albumin, 30% to 40% reactivity with albumins from rat, mouse, pig, and rabbit, low cross-reactivity with albumin from cow, and no measurable reactivity with sheep albumin.

The Immulite assay, which uses a mouse monoclonal antibody, strongly cross-reacted with albumin from baboon and rabbit, but did not cross-react with albumins from other species. Thus, the monoclonal antibodybased assay tended to respond to sequence changes in an all-or-none rather than a graded fashion. By serial dilution of albumins from baboon, rabbit, and pig, we found that LX20 responses were linear over a wide concentration range (data not shown).

As described in the Introduction, urine has been noted to contain albumin fragments (8-14). We analyzed 24 urine specimens by Western blotting to examine the occurrence of large albumin fragments. Comparison of analyses under reducing and nonreducing conditions examined the occurrence of internally cleaved forms of albumin held together by disulfide bonds. Results are shown for a representative sampling of 6 frozen (at -80 [degrees]C) and thawed specimens from patients with focal segmental glomerulosclerosis [Fig. 3, lanes 1-6, panels A (nonreduced) and B (reduced)]. Only trace amounts of fragments were observed, almost exclusively under reducing conditions. Also, small amounts of higher-molecular-weight forms were observed under nonreducing conditions. Disappearance of these components under reducing conditions suggested that they were disulfide-linked dimers.

Fresh urine specimens randomly selected for proteinuria also showed low amounts of albumin fragments [lanes 10-12, panels C(nonreduced) and F (reduced)]. Analyses of 3 proteinuric specimens frozen for 3 years at -20 [degrees]C showed variable degradation of albumin [lanes 7-9, panels C (nonreduced) and D (reduced)]. One specimen consisted primarily of intact albumin and the other 2 specimens showed mostly albumin fragments.

[FIGURE 3 OMITTED]

The 3 specimens in lanes 7-9, with prolonged storage at -20 [degrees]C, all yielded high measured values of albumin by the LX20 assay--1700, 1690, and 202 mg/L--and responses were approximately linear with dilution (Fig. 4). Total protein measurements by pyrogallol red assay were 2000, 1600, and 550 mg/L, respectively. The Immulite assay found a high concentration of albumin--1510 mg/L--and linear responses for 1 sample but values below the limit of detection in the other 2 specimens at all assayed dilutions. Apparently, the Immulite assay had low cross-reactivity with fragmented albumin in the latter 2 specimens, whereas the albumin fragments were detected by the LX20 assay.

Discussion

There are numerous ways in which urinary albumin could differ structurally from plasma albumin, so as to alter the protein's immunoreactivity. These include changes of conformation, chemical modifications of amino acid side chains, cleavage of the polypeptide chain, covalent homo- or heterodimer formation via an unpaired cysteine residue, noncovalent aggregation, changes in ligand concentration, or proteolytic fragmentation. In the urine environment, albumin can be exposed to a wide range of ionic strength, pH, urea concentrations approaching 1 mol/L, high concentrations of glucose and ascorbic acid, aerobic oxidizing conditions, increased amounts of nitrite in cases of urinary tract infections, and a variety of proteases along the urinary epithelium (14). We examined the effects of exposures of albumin to a variety of denaturing conditions or reactants on its cross-reactivity in assays for urinary albumin.

Albumin consists of 3 globular domains stabilized by 17 internal disulfide bonds (25, 28). Albumin resists denaturation during heating for 10 h at 60 [degrees]C as performed for commercial albumin preparations (25). Exposure to low pH or to high concentrations of chaotropes results in reversible denaturation (25), and we observed no effects of these treatments on measured albumin. Therefore, exposure of albumin to the usual range of urinary pH, urea, and salt concentrations appears unlikely to induce an irreversible conformational change. When albumin does denature at extremely high temperature or pH, it aggregates irreversibly; this leads to precipitation of albumin rather than additional forms in solution (25). Some dimerization of albumin is observed with many preparations, including purified plasma albumin (21, 25), but as in previously reported studies, we found albumin dimers to be minor components in urine. Aggregation of albumin with other components such as the Tamm-Horsfall protein would result in elution in size-exclusion chromatography as amuch larger complex, and that has not been observed (15, 21). Albumin binds many ligands, and ligand binding may induce a change in conformation of albumin (reviewed in 25). Such factors may be of interest in examining matrix effects of urine on albumin measurements, but analyses in this study used a simple saline matrix.

Urinary albumin is enriched for glycated forms (5, 6), but enrichment for other modifications of amino acid side chains has not been described. We found only small effects of exposure to high concentrations of glucose on cross-reactivity of albumin in the assays for urine albumin. Another reagent, citraconic anhydride, which also modifies lysine residues, had little effect on the LX20 assay but decreased cross-reactivity in the Immulite assay by approximately 50%.

[FIGURE 4 OMITTED]

We anticipated that albumin modification or fragmentation would diminish reactivity in immunoassays for urine albumin owing to loss of functional epitopes and decreased ability to form cross-linked complexes in immunoturbidimetric assays. Surprisingly, the immunoturbidimetric LX20 assay reacted nearly equivalently with albumin and with most preparations of fragmented or chemically modified albumin. Even isolated CNBr fragments representing 3 separate segments of albumin had nearly similar reactivity. These findings indicate a broad distribution of reactive epitopes across the albumin molecule that still allow antibody-mediated cross-linking reactions to occur for fragments representing only about one-third of the intact molecule. Proteolysis and chemical modification of albumin probably destroy some epitopes, but, apparently, polyclonal antibodies still react with enough epitopes to sustain turbidimetric reactions. A previous study of trypsin digestion of albumin (29), using a lower ratio of trypsin to albumin, concluded, "Many techniques used to assay patient urine samples are unable to detect fragmented albumin.... " In one respect, our results agree with that conclusion; low cross-reactivity was observed for small fragments isolated by ultrafiltration. However, our results differ in noting significant cross-reactivity of tryptic digests containing very little intact albumin. Therefore, we conclude that large fragments and internally nicked forms of albumin from trypsin digestion retained significant cross-reactivity in 2 assays for urine albumin. The experiments with CNBr cleavage and incubation with ascorbate provide even clearer demonstration of the reactivity of albumin fragments in an immunoturbidimetric assay and the differential reactivity of albumin assays based on different assay principles.

The mechanism that we consider most likely to influence the measurement of urinary albumin by immunoassays is cleavage of its polypeptide chain, and this was previously suggested by Osicka and Comper (15) as a mechanism for generating internally nicked forms of albumin with decreased immunoreactivity. Several studies have noted the presence of albumin fragments in urine (8-14), although there has been controversy regarding the abundance and sources of small peptides in urine (30-32). Our results suggest that most freshly collected urine specimens or specimens frozen at _80 [degrees]C do not contain substantial amounts of large albumin fragments detectable by Western blotting, but extensive cleavage of albumin can occur in some urine specimens stored frozen for an extended period at _20 [degrees]C. This result bears further investigation, as we had only a few such samples for analysis. Albumin degradation could result from either action of proteases or free radicals. Several reports offer varying views regarding the stability of urine specimens stored at _20 [degrees]C (33-37). The recent report by Brinkman et al. (37) notes a progressive decline in results by immunonephelometric assay for specimens stored at -20 [degrees]C, but an even greater decline for results by size-exclusion chromatography. Those results suggest that the size-exclusion assay is more sensitive to structural changes of albumin during storage than is the immunonephelometric assay, countering a previous hypothesis that analysis by size exclusion is less sensitive to albumin degradation (15, 16).

Our experiments demonstrate that it is extremely difficult to modify albumin so as to abrogate immunoreactivity with polyclonal antibodies. The amino acid sequence of albumin diverges substantially through phylogeny (25), although there is conservation of the cysteine residues that form disulfides that serve as stabilizing cross-links for albumin's globular structure. Human albumin has about 94% sequence identity with albumin from other primates such as the macaque. In the present study, albumin from gorilla was observed to have strong cross-reactivity in the 2 assays for urine albumin (human). Sequence identity with human albumin decreases to 72%-77% for albumins from rabbit, sheep, cow, rat, and mouse, and amino acid substitutions between species are relatively evenly distributed through the albumin molecule. Therefore, goats immunized with human albumin will be presented with an antigen that differs in about 150 of 585 amino acid residues relative to goat albumin. (The sequence of albumin from sheep, which phylogenetically is closely related to goat, was used to estimate identity between goat and human albumin, since goat albumin is not in the sequence database.) The > 100 amino acid differences should lead goats to develop a diverse repertoire of antibodies directed at multiple sites in human albumin, and it explains how a urine albumin assay using goat antibodies vs human albumin can react with a variety of modified albumins, albumins from other animal species, or albumin fragments. It is not plausible, as recently reported, that there is a modified form of albumin failing to cross-react with 20 different polyclonal antisera (16). Those findings are strong evidence that the putative "immunounreactive albumin" actually is a different protein than albumin.

Use of a monoclonal antibody in an assay for urine albumin leads to a very different sensitivity to structural modification of albumin, as it depends on reactivity with a single specific epitope expressed by albumin. Consequently, much greater sensitivity to albumin modification was observed for the Immulite assay, which uses a monoclonal antibody, than for the LX20 assay using polyclonal goat antibodies. The differential reactivity of assays to modified albumins is likely to be primarily related to the type of antibodies employed, although there may also be differences between competitive and noncompetitive assay formats. Previously, differential reactivity of monkey albumin was noted in 2 urine albumin assays. Using the particle-enhanced turbidity inhibition assay (a competitive assay) for human albumin on the DuPont aca[R], response curves for monkey albumin were shifted about 10-fold relative to human albumin, whereas the LX20 assay had nearly equivalent reactivity with monkey and human albumin (38). Preliminary results with another competitive urinary albumin assay on the Dade Dimension[R] analyzer suggest a loss of reactivity with albumin fragments similar to the competitive assay on the Immulite (27). Those reports, together with findings in this report, suggest that assay design probably influences the reactivity of urine albumin assays with modified or fragmented albumin, and this is a potential source of nonequivalence between assays in the analysis of urinary albumin, although assays may react equivalently with serum albumin added to calibrators, control, or proficiency testing materials.

There are many commercially available assays for urine albumin. The present study focused primarily on model systems to examine the sensitivity of urine albumin assays to structural modification or fragmentation of albumin. Our results suggest a need for evaluation of the reactivity of different assays with modified or fragmented albumin and for further investigation into occurrence and formation of modified or fragmented albumin in urine specimens. Further development of reference methods for the measurement of urinary albumin such as recent efforts to apply new electrophoretic (39) or mass spectrometric (40) methods might contribute to progress in this area.

Financial Disclosures: No other external support or potential conflicts of interest are noted for the authors.

Grant Support: Studies were supported by the intramural program of Clinical Center, National Institutes of Health, US Department of Health and Human Services.

Received June 2, 2007; accepted October 15, 2007.

Previously published online at DOI: 10.1373/clinchem.2007.092825

References

(1.) Viberti GC, Hill RD, Jarrett RJ, Argyropoulos A, Mahmud U, Keen H. Microalbuminuria as a predictor of clinical nephropathy in insulin-dependent diabetes mellitus. Lancet 1982;i:1430 -2.

(2.) Arnlov J, Evans JC, Meigs JB, Wang TJ, Fox CS, Levy D, et al. Low-grade albuminuria and incidence of cardiovascular disease events in nonhypertensive and nondiabetic individuals: the Framingham Heart Study. Circulation 2005;112: 969-75.

(3.) Wang TJ, Gona P, Larson MG, Tofler GH, Levy D, Newton-Cheh C, et al. Multiple biomarkers for the prediction of first major cardiovascular events and death. N Engl J Med 2006;355:2631-9.

(4.) Xu J, Knowler WC, Devereux RB, Yeh J, Umans JG, Begum M, et al. Albuminuria within the "normal" range and risk of cardiovascular disease and death in American Indians: the Strong Heart Study. Am J Kid Dis 2007;49:208 -16.

(5.) Ghiggeri GM, Candiano G, Delfino G, Queirolo C. Electrical charge of serum and urinary albumin in normal and diabetic humans. Kidney Int 1985;28: 168-77.

(6.) Cha T, Tahara Y, Yamato E, Yoneda H, Ikegami H, Noma Y, et al. Renal handling of glycated albumin in non-insulin-dependent diabetes mellitus with nephropathy. Diabetes Res Clin Pract 1991; 12:149 -56.

(7.) Bar-Or D, Rael LT, Bar-Or R, Slone DS, Craun ML. The formation and rapid clearance of a truncated albumin species in a critically ill patient. Clin Chim Acta 2006;365:346 -9.

(8.) Wiggins RC, Kshrisagar B, Kelsch RC, Wilson BS. Fragmentation and polymeric complexes of albumin in human urine. Clin Chim Acta 1985;149: 155-63.

(9.) Candano G, Musante L, Bruschi M, Petretto A, Santucci L, Del Boccio P, et al. Repetitive fragmentation products of albumin and _1-antitrypsin in glomerular diseases associated with nephritic syndrome. J Am Soc Nephrol 2006;17: 3139-48.

(10.) Lafitte D, Dussol B, Andersen S, Vazi A, Dupuy P, Jensen ON, et al. Optimized preparation of urine samples for two-dimensional electrophoresis and initial application to patient samples. Clin Biochem 2002;35:581-9.

(11.) Khan A, Packer NH. Simple urinary sample preparation for proteomic analysis. J Proteome Res 2006;5:2824 -38.

(12.) Wittke S, Mischak H, Walden M, Kolch W, Radler T, Wiedemann K. Discovery of biomarkers in human urine and cerebrospinal fluid by capillary electrophoresis coupled to mass spectrometry: towards new diagnostic and therapeutic approaches. Electrophoresis 2005;26:1476-87.

(13.) Jurgens M, Appel A, Heine G, Neitz S, Menzel C, Tammen H, et al. Towards characterization of the human urinary peptidome. Comb Chem High Throughput Screening 2005;8:757- 65.

(14.) Hortin GL, Sviridov D. Diagnostic potential for urinary proteomics. Pharmacogenomics 2007;8: 237-55.

(15.) Osicka TM, Comper WD. Characterization of immunochemically nonreactive urinary albumin. Clin Chem 2004;50:2286 -91.

(16.) Clavant SP, Sastra SA, Osicka TM, Comper WD. The analysis and characterization of immunounreactive urinary albumin in healthy volunteers. Clin Biochem 2006;39:143-51.

(17.) Brinkman JW, Bakker SJL, Gansevoort RT, Hillege HL, Kema IP, Gans ROB, et al. Which method for quantifying urinary albumin excretion gives what outcome? A comparison of immunonephelometry with HPLC. Kidney Int 2004;66(Suppl 92):S69S75.

(18.) Owen WE, Roberts WL. Performance characteristics of an HPLC assay for urinary albumin. Am J Clin Pathol 2005;124:219 -25.

(19.) Contois JH, Hartigan C, Rao LV, Snyder LM, Thompson MJ. Analytical validation of an HPLC assay for urinary albumin. Clin Chim Acta 2006; 367:150 -5.

(20.) Polkinghorne KR, Su Q, Chadban SJ, Shaw JE, Zimmet PZ, Atkins RC. Population prevalence of albuminuria in the Australian diabetes, obesity, and lifestyle (AusDiab) study: immunonephelometry compared with high-performance liquid chromatography. Am J Kidney Dis 2006;47:604 -13.

(21.) Sviridov D, Meilinger B, Drake SK, Hoehn GT, Hortin GL. Co-elution of other proteins with albumin during size-exclusion HPLC: implications for urine albumin analysis. Clin Chem 2006;52: 389-97.

(22.) Peters T. How should we measure albumin? [Editorial] Clin Chem 2006;52:555- 6.

(23.) Hortin G, Fok KF, Toren PC, Strauss AW. Sulfation of a tyrosine residue in the plasmin-binding domain of _-2-antiplasmin. J Biol Chem 1987;262: 3082-5.

(24.) Hortin GL, Meilinger B, Drake SK. Size-selective extraction of peptides from urine for mass spectrometric analysis. Clin Chem 2004;50:1092-5.

(25.) Peters T. All about albumin. Academic Press, San Diego, CA, 1996.

(26.) Lee HJ, Sok DE. Cu2_-catalyzed oxidative degradation of thyroglobulin. Free Radic Res 2000;33: 359-68.

(27.) Sviridov D, Drake SK, Hortin, GL. Variable reactivity of different microalbumin assays with fragmented or modified albumin [Abstract]. Clin Chem 2007;53(Suppl):A81.

(28.) Curry S, Mandelkow H, Brick P, Franks N. Crystal structure of human serum albumin complexed with fatty acid reveals an asymmetric distribution of binding sites. Nature Struct Biol 1998;5:82735.

(29.) Eppel GA, Nagy S, Jenkins MA, Tudball RN, Daskalakis M, Balazs NDH, et al. Variability of standard clinical protein assays in the analysis of a model urine solution of fragmented albumin. Clin Biochem 2000;33:487-94.

(30.) Strong KJ, Osicka TM, Comper WD. Urinary-peptide excretion by patients with and volunteers without diabetes. J Lab Clin Med 2005;145:23946.

(31.) Hortin GL, Meilinger B. Cross-reactivity of amino acids and other compounds in the biuret reaction: interference with urinary peptide measurements. Clin Chem 2005;51:1411-9.

(32.) Norden AGW, Sharratt P, Cutillas PR, Cramer R, Gardner SC, Unwin RJ. Quantitative amino acid and proteomic analysis: very low excretion of polypeptides _ 750 Da in normal urine. Kidney Int 2004;66:1994 -2003.

(33.) Osberg I, Chase HP, Garg SK, DeAndrea A, Harris S, Hamilton R, et al. Effects of storage time and temperature on measurement of small concentrations of albumin in urine. Clin Chem 1990;36: 1428-30.

(34.) MacNeil MLW, Mueller PW, Caudill SP, Steinberg KK. Considerations when measuring urinary albumin: precision, substances that may interfere, and conditions for sample storage. Clin Chem 1991;37:2120 -3.

(35.) Giampietro O, Penno G, Clerico A, Cruschelli L, Cecere M. How and how long to store urine samples before albumin radioimmunoassay: a practical response. Clin Chem 1993;39:533- 6.

(36.) Tencer J, Thysell H, Andersson K, Grubb A. Long-term stability of albumin, protein HC, immunoglobulin G, kappa- and lambda-chain-immunoreactivity, orosomucoid, and alpha 1-antitrypsin in urine stored at -20 degrees C. Scand J Urol Nephrol 1997;31:67-71.

(37.) Brinkman JW, de Zeeuw D, Lambers Heerspin HJ, Gansevoort RT, Kema IP, De Jong PE, et al. Apparent loss of urinary albumin during longterm frozen storage: HPLC vs immunonephelometry. Clin Chem 2007;53:1520-6.

(38.) Hortin GL, King C, Kopp J. Quantification of rhesus monkey albumin with assays for human microalbumin [Abstract]. Clin Chem 2000;46:A140-1.

(39.) Chan OTM, Herold DA. Chip electrophoresis as a method for quantifying total microalbuminuria. Clin Chem 2006;52:2141- 6.

(40.) Babic N, Larson TS, Grebe SK, Turner ST, Kumar R, Singh RJ. Application of liquid chromatography-mass spectrometry technology for early detection of microalbuminuria in patients with kidney disease. Clin Chem 2006;52:2155-7.

[1] Nonstandard abbreviations: CNBr, cyanogen bromide.

Denis Sviridov, [1] Steven K. Drake, [2] and Glen L. Hortin [1] *

Departments of [1] Laboratory Medicine and [2] Critical Care Medicine, Warren Magnuson Clinical Center, National Institutes of Health, Bethesda, MD.

* Address correspondence to this author at: Department of Laboratory Medicine, NIH, Building 10, Room 2C-407, Bethesda, MD 20892-1508. Fax 301-4021885; e-mail ghortin@mail.cc.nih.gov.
Table 1. Comparison of albumin:total protein ratios
for modified or fragmented albumins analyzed by
2 different assays for urinary albumin.

 Albumin:total protein

Albumin preparation (a) LX20 Immulite

Treatment for 24 h
Albumin (saline control) 1.03 0.98
CNBr digest, total 1.09 0.17
CNBr fragment A 0.85 0.04
CNBr fragment B 0.93 0.04
CNBr fragment C 1.58 0.16
Trypsin digest, total 0.21 0.23
Hydrogen peroxide 1.04 1.17
Citraconic anhydride 0.94 0.45
Incubated at pH 2 1.05 1.08
Incubated at 65 [degrees]C 0.95 0.94
Ascorbate 0.98 0.78
Treatment for 5 d
Albumin (control), pH 7.2 1.04 1.03
Albumin (control), pH 5.7 0.92 0.88
10% glucose, pH 7.2 1.01 0.83
10% glucose, pH 5.7 0.94 0.86
Nitrite, pH 7.2 1.03 0.91
Nitrite, pH 5.7 0.96 0.68
Ascorbate, pH 7.2 0.65 0.02
Ascorbate, pH 5.7 0.94 0.51

(a) Modified albumins were all prepared from purified
plasma albumin.

Table 2. Cross-reactivity of serum albumin from
different animal species in assays for human
urine albumin.

 Albumin:total protein
Animal
species LX20 Immulite

Baboon 0.82 1.19
Cow 0.08 <0.01
Rabbit 0.33 1.06
Sheep <0.01 <0.01
Pig 0.33 <0.01
Rat 0.38 <0.01
Mouse 0.33 <0.01
COPYRIGHT 2008 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2008 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Proteomics and Protein Markers
Author:Sviridov, Denis; Drake, Steven K.; Hortin, Glen L.
Publication:Clinical Chemistry
Date:Jan 1, 2008
Words:5034
Previous Article:SELDI-TOF MS whole serum proteomic profiling with IMAC surface does not reliably detect prostate cancer.
Next Article:Mass spectrometry: a tool for enhanced detection of hemoglobin variants.
Topics:

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters