Quantification by liquid chromatography tandem mass spectrometry of mycophenolic acid and its phenol and acyl glucuronide metabolites.

Mycophenolic acid acyl glucuronide (AcMPAG), a secondary metabolite of mycophenolic acid (MPA), has pharmacologic and toxicologic properties that may contribute to the gastrointestinal side effects such as diarrhea that are associated with MPA (1-3). For >7 years we have used HPLC with an ultraviolet detector (HPLC-UV) in our laboratory to measure MPA and its metabolites (4, 5). Although we rarely observed interference with MPA and its inactive phenol glucuronide metabolite (MPAG), we recently found that, because of interference, AcMPAG could not be quantified in ~26% of samples. Therefore, we developed and validated an LC-MS/MS assay for MPA, MPAG, and AcMPAG that was based on our previously published assay (6).

We purchased HPLC-grade methanol, acetonitrile, acetic acid, ammonium acetate, sodium tungstate, and perchloric acid from Merck. MPA, MPAG, AcMPAG, and the carboxybutoxy ether of MPA (MPAC) were kind gifts from Roche Pharmaceuticals.

We prepared the following stock solutions: 3.0 g/L AcMPAG in dimethyl sulfoxide, 1.0 g/L MPA in acetonitrile, 5.0 g/L MPAG in acetonitrile and water (80 mL acetonitrile, 20 mL water), and 1.0 g/L of MPAC in acetonitrile. The stock solutions were prepared twice (for calibrators and quality controls) and stored at 4[degrees]C (AcMPAG) or -20[degrees]C (MPA, MPAG, and MPAC). We prepared 2 serum calibrators each for MPA (3.0 and 10.0 mg/L), MPAG (10.0 and 100.0 mg/L), and AcMPAG (1.0 and 3.0 mg/L). Because of the inherent instability of AcMPAG at physiologic pH, the AcMPAG calibrators were prepared by adding stock solution to EDTA-plasma from healthy blood donors (who gave informed consent), acidified with orthophosphoric acid (14.5 g/L). Serum quality control samples were prepared in drug-free serum (0.2, 1.0, and 25.0 mg/L MPA and 10.0, 50.0, and 250.0 mg/L MPAG; Bio-Rad) or acidified plasma (0.1, 0.6, and 1.5 mg/L AcMPAG).

Sample preparation was similar to our previous protocol (6), except that the volumes were decreased: 100 [micro]L of calibrator, quality control or patient sample (EDTA plasma) was treated with 10 [micro]L of perchloric acid (150 g/L),10 [micro]L sodium tungstate (250 g/L), and 50 [micro]L of internal calibrator solution (15 mg/L MPAC in acetonitrile). After vigorous mixing and centrifugation (5 min at 14 000g), the supernatants were transferred into 500-[micro]L polypropylene tubes (Eppendorf), and 5 [micro]L was injected into the chromatographic system. For quantification of MPAG, the samples were diluted 1:10 with drug-free serum before extraction.

The chromatographic system consisted of a binary pump, a quaternary pump, an autosampler with a 200-[micro]L sample loop, a decreased pressure degasser, and a Peltier column oven (a115eries 200, Perkin-Elmer). The analytical column was an Aqua Perfect 150 x 4 mm C18 reversedphase column, 5 [micro]m particle size (MZ-Analysentechnik). A Waters Oasis HLB Extraction Column (25 [micro]m, 2 X 15 mm) served as precolumn for online solid-phase extraction. The compounds were injected onto the precolumn in water (pH 3.0 with acetic acid) at 3 mL/min. After 0.4 min, the samples were eluted and separated on the analytical column, maintained at 30[degrees]C (1.3 mL/min; 520 mL methanol, 170 mL acetonitrile, 310 mL water, 0.2 mmol/L ammonium acetate, and 0.244 g/L formic acid, pH 3.0). For washing and equilibration, the injection pump was operated as follows: 0.0 min 3.0 mL/min (water, pH 3.0), 2.0 min 1.0 mL/min (methanol), 3.0 min 2.0 mL/min (methanol), 3.5 min 5.0 mL/min (water, pH 3.0), 4.0 min 3.0 mL/min (water, pH 3.0). The valve was switched at 0.0 min to B (injection), at 0.4 min to A (elution), at 2.5 min to B, at 3.0 min to A, and at 3.5 min to B.

A PE SCIEX API 2000 triple-quadrupole mass spectrometer with a turbo-ion spray interface was used for detection. Conditions were as follows: split 1:10; positive ion spray 5000 V; source 250[degrees]C; 50 psi for ion-source gas 1; 70 psi for ion source gas 2; entrance potential 8.5 V; declustering potentia120 V; and collision energies of 30 eV (MPAC), 25 eV (MPA), and 35 eV (MPAG and AcMPAG). Multiple-reaction monitoring was performed for the ammonium adducts and the product ions of MPAC (m/z 438.0/207.0) and MPA (m/z 338.1/207.1) and the metabolites AcMPAG and MPAG (m/z 514.2/207.1). Drug concentrations were quantified through the peak area ratios.

For examples of the ion chromatograms for MPA, MPAG, and AcMPAG, see Fig. 1 in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol52/issue10. Insource fragmentation occurred with ~1% of MPAG, but because of chromatographic separation, this MPAG fragmentation did not affect MPA quantification. The retention times were 2.75 min for MPA, 1.8 min for MPAG, 2.2 min for AcMPAG, and 2.7 min for MPAC.

The assay was validated in accordance with Food and Drug Administration guidelines (7) (Table 1). Within-run and between-run imprecision were acceptable (CV, <10%; n = 15) for each compound and each quality control. The lower limit of quantification (defined as the concentration at which the CV does not exceed 20%; n = 15) was 0.1 mg/L for MPA, 1.0 mg/L for MPAG, and 0.05 mg/L for AcMPAG. The mean analytical recovery (n = 15 for each quality control) was 90% to 110%. The assay was linear from 0.1 to 50 mg/L for MPA (r = 0.9987), from 1 to 500 mg/L for MPAG (r = 0.9999), and from 0.05 to 10 mg/L for AcMPAG (r = 0.9988). Assessment of carryover effects showed <0.5% for each compound (0.41% after 10 mg/L MPA, 0.23% after 100 mg/L MPAG, 0.06% after 3 mg/L AcMPAG; each n = 10). Mean extraction efficiencies were 81% for MPA and 107% for MPAG and AcMPAG (each n = 4). Lower extraction efficiency for MPA compared with its metabolites was consistent with previous results (4). Evaluation of matrix effects through a postcolumn infusion experiment (8) showed that suppression of the ion generation occurred only at 0.7-1.0 min (see Fig. 2 in the online Data Supplement). No ion suppression was present during the elution times for the compounds of interest (1.5-3.5 min).

To compare our LC-MS/MS method with a validated HPLC-UV method (4, 5), we analyzed specimens obtained from routine analysis of mainly kidney transplantation patients (n = 46 for MPA and MPAG; n = 28 for AcMPAG). Regression analysis was performed by use of the nonparametric procedure of Passing and Bablok (9) and EVAPAK 3.1 (Roche Diagnostics) software (Fig. 1). The comparison showed excellent agreement for MPA and MPAG. In addition, the use of proficiency testing scheme samples and the method mean data (mean deviation, 5.8%; n = 20) (10) showed good congruence between the MPA concentrations determined with our LC-MS/MS method. For AcMPAG measurements, we observed a slight bias at concentrations <0.3 mg/L, for which the LC-MS/MS yielded higher results, probably attributable to the relatively low signal-to-noise ratio at low AcMPAG concentrations in the HPLC-UV assay, leading to increasing impairment by frequent interferences, which are not always easy to identify.

At physiologic pHs, AcMPAG has been found to undergo degradation (4) attributed to different reagents used for sample preparation (11). To verify the degradation of AcMPAG in plasma before and after precipitation of the samples, we supplemented EDTA-plasma with AcMPAG and compared the concentrations during various incubation times with and without acidification. The storage of plasma samples at ambient temperature without prior acidification led to a degradation of >50% of AcMPAG after 44 h, whereas almost no degradation occurred in acidified samples (see Fig. 3 in the online Data Supplement). In contrast, after precipitation, the AcMPAG concentration remained stable for at least 44 h even at 4[degrees]C, regardless of any acidification before preparation. Therefore, we can confirm the results of Shipkova et al. (5) indicating that prior sample acidification has a substantial effect on the stability of AcMPAG in plasma samples and thus is essential for AcMPAG quantification.

Many assays have been described that use HPLC-UV (11,12,13) or mass spectrometry (6,14,15) for the quantification of MPA and/or MPAG. The use of HPLC-UV methods to measure AcMPAG has been reported (11,12), but the use of mass spectrometry has not. Advantages of this LC-MS/MS assay for routine use include the small sample volume requirement (100 [micro]L) and ease of sample preparation. Chromatographic separation decreases insource fragmentation of MPAG, which can lead to false-high MPA concentration results (16). The assay is not impaired by ion suppression during compound detection. Problems with AcMPAG quantification caused by interferences in the HPLC-UV assay can be overcome by specific mass detection of the compound. Furthermore, the assay time allows a maximum throughput of 240 samples per day and system. We found this speed of processing advantageous because of the increasing number of requests for the quantification of MPA and its metabolites for routine use and clinical trials.

[FIGURE 1 OMITTED]

We thank Roche Pharmaceuticals for providing MPA, MPAG, and AcMPAG. M.O. and V.W.A. have previously acted as advisors for Hoffmann-La Roche.

References

(1.) Shipkova M, Wieland E, Schutz E, Wiese C, Niedmann PD, Oellerich M, et al. The acyl glucuronide metabolite of mycophenolic acid inhibits the proliferation of human mononuclear leukocytes. Transplant Proc 2001;33:1080-1.

(2.) Wieland E, Shipkova M, Schellhaas U, Schutz E, Niedmann PD, Armstrong VW, et al. Induction of cytokine release by the acyl glucuronide of mycophenolic acid: A link to side effects? Clin Biochem 2000;33:107-13.

(3.) Shipkova M, Armstrong VW, Oellerich M, Wieland E. Acyl glucuronide drug metabolites: toxicological and analytical implications. Ther Drug Monit 2003;25:1-16.

(4.) Shipkova M, Niedmann PD, Armstrong VW, Schutz E, Wieland E, Shaw LM, et al. Simultaneous determination of mycophenolic acid and its glucuronide in human plasma using a simple high-performance liquid chromatography procedure. Clin Chem 1998;44:1481-8.

(5.) Shipkova M, Schutz E, Armstrong VW, Niedmann PD, Oellerich M, Wieland E. Determination of the acyl glucuronide metabolite of mycophenolic acid in human plasma by HPLC and Emit. Clin Chem 2000;46:365-72.

(6.) Streit F, Shipkova M, Armstrong VW, Oellerich M. Validation of a rapid and sensitive liquid chromatography-tandem mass spectrometry method for free and total mycophenolic acid. Clin Chem 2004;50:152-9.

(7.) US Food and Drug Administration. Guidance for industry bioanalytical method validation. http://www.fda.gov/cder/guidance/4252fnl.htm (accessed February 2006).

(8.) Annesley TM. Ion suppression in mass spectrometry. Clin Chem 2003;49: 1041-4.

(9.) Passing H, Bablok W. A new biometrical procedure for testing the equality of measurements from 2 different analytical methods--application of linear-regression procedures for method comparison studies in clinical-chemistry. 1. J Clin Chem Clin Biochem 1983;21:709-20.

(10.) Analytical Services International. Bioanalytics for the pharmaceutical and diagnostic industries. http://www.bioanalytics.co.uk/ (accessed February 2006).

(11.) Patel CG, Akhlaghi F. High-performance liquid chromatography method for the determination of Mycophenolic acid and its acyl and phenol glucuronide metabolites in human plasma. Ther Drug Monit 2006;28:116-22.

(12.) Westley IS, Sallustio BC, Morris RG. Validation of a high-performance liquid chromatography method for the measurement of mycophenolic acid and its glucuronide metabolites in plasma. Clin Biochem 2005;38:824-9.

(13.) Wiwattanawongsa K, Heinzen EL, Kemp DC, Dupuis RE, Smith PC. Determination of mycophenolic acid and its phenol glucuronide metabolite in human plasma and urine by high-performance liquid chromatography. J Chromatogr B Biomed Sci Appl 2001;763:35-45.

(14.) Annesley TM, Clayton LT. Quantification of mycophenolic acid and glucuronide metabolite in human serum by HPLC-tandem mass spectrometry. Clin Chem 2005;51:872-7.

(15.) Patel CG, Mendonza AE, Akhlaghi F, Majid 0, Trull AK, Lee T, et al. Determination of total mycophenolic acid and its glucuronide metabolite using liquid chromatography with ultraviolet detection and unbound mycophenolic acid using tandem mass spectrometry. J Chromatogr B Biomed Sci Appl 2004;813:287-94.

(16.) Vogeser M, Zachoval R, Spohrer U, Jacob K. Potential lack of specificity using electrospray tandem-mass spectrometry for the analysis of mycophenolic acid in serum. Ther Drug Monit 2001;23:722-4.

Previously published online at D01: 10.1373/clinchem.2006.074336

Gunnar Brandhorst, * Frank Streit, Sandra Goetze, Michael Oellerich, and Victor William Armstrong (Universitatsklinikum Gottingen, Abteilung Klinische Chemie/Zentrallabor, 37075 Gottingen, Germany; * address correspondence to this author at: Universitatsklinikum Gottingen, Abteilung Klinische Chemie/Zentrallabor, 37075 Gottingen, Germany; fax 49-551-39-12771, e-mail gunnar.brandhorst~med.unigoettingen. de)
Table 1. Assay validation: analytical recovery and imprecision.

 Within-run CV Between-run CV Recovery
mg/L (n = 15), % (n = 15), % (n = 15), %

MPA
 0.2 8.5 7.0 110
 1.0 5.5 6.8 100
 25.0 4.2 5.5 91
MPAG
 10.0 5.5 6.5 100
 50.0 4.7 6.0 100
 250.0 6.2 8.1 96
AcMPAG
 0.1 9.8 8.4 98
 0.6 6.6 7.2 96
 1.5 5.0 6.7 96
COPYRIGHT 2006 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2006 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Technical Briefs
Author:Brandhorst, Gunnar; Streit, Frank; Goetze, Sandra; Oellerich, Michael; Armstrong, Victor William
Publication:Clinical Chemistry
Date:Oct 1, 2006
Words:2231
Previous Article:In situ amplified chemiluminescent detection of DNA and immunoassay of IgG using special-shaped gold nanoparticles as label.
Next Article:Performance and clinical utility of a commercial von Willebrand factor collagen binding assay for laboratory diagnosis of Von Willebrand disease.
Topics:


Related Articles
HPLC-tandem mass spectrometric method to characterize resveratrol metabolism in humans.
Rapid 2nd-tier test for measurement of 3-OH-propionic and methylmalonic acids on dried blood spots: reducing the false-positive rate for...
Total polyphenol intake estimated by a modified folin-ciocalteu assay of urine.
Analysis of methylmalonic acid in plasma by liquid chromatography-tandem mass spectrometry.
Quantification of mycophenolic acid and glucuronide metabolite in human serum by HPLC-tandem mass spectrometry.
Comprehensive detection of disorders of purine and pyrimidine metabolism by HPLC with electrospray ionization tandem mass spectrometry.
Routine isotope-dilution liquid chromatography--tandem mass spectrometry assay for simultaneous measurement of the 25-hydroxy metabolites of vitamins...
Sensitive gas chromatography-mass spectrometry method for simultaneous measurement of MDEA, MDMA, and metabolites HMA, MDA, and HMMA in human urine.
Validation of a rapid and sensitive liquid chromatography-tandem mass spectrometry method for free and total mycophenolic acid.
Rapid diagnosis of MCAD deficiency: quantitative analysis of octanoylcarnitine and other acylcarnitines in newborn blood spots by tandem mass...

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters