Plasma homocysteine determined by capillary electrophoresis with laser-induced fluorescence detection.

Homocysteine (Hcy) has emerged as another risk factor for the development of coronary heart disease (1). Genetic abnormalities of the enzymes cystathionine-[beta]-synthase and methylene tetrahydrofolate reductase (folic acid and vitamin [B.sub.6] and [B.sub.12] cofactors) can cause raised plasma Hcy concentrations. Deficiencies in folic acid and vitamins [B.sub.6] and [B.sub.12] may also contribute to this increased concentration (1-3). Consequently, patients with coronary heart disease have been treated successfully with folic acid, vitamin [B.sub.12], or vitamin [B.sub.6] (1), but the benefit of such treatment in reducing morbid cardiovascular endpoints is presently unknown (3).

A widely used technique for measuring total plasma Hcy is reversed-phase HPLC with fluorescence detection after derivatization of plasma thiols (4-6). Some studies use gas chromatography-mass spectrometry techniques (7) with various derivatization protocols (8).

The increasing demand for the determination of plasma total Hcy prompted us to develop a rapid, automated method based on capillary electrophoresis (CE) and laser-induced fluorescence detection (LIF), in which all specific thiols are detected selectively.

Blood was collected by venipuncture into evacuated tubes containing EDTA from patients with episodes of deep-vein thrombosis or vascular disease. The plasma was separated by centrifugation and was stored at -20[degrees]C until analysis.

Plasma samples (100 [micro]L) were treated for 30 min at room temperature with 10 [micro]L of a novel thiol-reducing agent [35 mmol/L tris(2-carboxyethyl)phosphine; Molecular Probes], as reported recently by Gilfix et al. (9). Compared with the widely employed tributyl phosphine, this new phosphine is nonvolatile, stable, soluble in water, and without disagreeable odor (5,6). When we used this procedure, total Hcy (mixed and symmetric disulfides, including protein-bound Hcy) was analyzed in its reduced form.

All patient samples and plasma-based calibrators were incubated with tris(2-carboxyethyl)phosphine under the same conditions. The internal standard [5 [micro]L of 10 mmol/L D-penicillamine (DP)] was mixed with plasma. The solution was deproteinized with 60 [micro]L of 0.8 mol/L 5-sulfosalicylic acid with vortex-mixing, followed by centrifugation at 10 000g for 15 min.

We mixed 100 [micro]L of the supernatant with 50 [micro]L of 0.3 mol/L carbonate buffer, pH 9.5, and 5 [micro]L of 5 mol/L sodium hydroxide. We then added 50 [micro]L of a 1 g/L 6-iodoacetamidofluorescein solution (6-IAF) in dimethyl sulfoxide (Sigma Chemical Co.), and the mixture was incubated overnight in the dark at room temperature to allow analysis after the reaction has plateaued (>2 h). IAF-labeled samples were stored at -20 [degrees]C until analysis; they could be preserved for up to 3 months. Each derivatized plasma was analyzed directly after being diluted 5000-fold before the CE injection. The analyses were performed on a Zeta CE instrument (Zeta Technology) equipped with a modular LIF detector and an argon ion laser (wavelength, 488 nm). The different thiols were separated on a 75 cm x 50 [micro]m (i.d.) fused-silica capillary (Polymicro Technology) with an effective length of 43 cm and a total length of 75 cm. The separation buffer consisted of 10 mmol/L sodium dodecyl sulfate, 50 mmol/L boric acid, and 20 mmol/L 3-(cyclohexylamino)-1-propanesulfonic acid adjusted to pH 9.5 by addition of a sodium hydroxide solution (10 mol/L). The separation voltage was +25 kV, producing an electrophoretic current of 25 [micro]A.

All samples were assayed in triplicate. Peaks were identified by use of solutions of Hcy, Cys, or glutathione, and the internal standards cysteamine or DP. Fig. 1A shows the thiol solutions prepared in carbonate 0.3 mmol/L buffer, pH 9.5, and then diluted in deionized water. Under the electrophoretic conditions used, the Hcy derivative is well separated with a migration time of 7.2 min.

Electropherograms of a human plasma sample with added DP as internal standard produced several peaks (Fig. 1, B and C). Known quantities of Hcy, Cys, or DP solutions were added to the plasma sample to identify Hcy, Cys, and DP peaks.

Hcy calibrators were prepared from a plasma pool (30, 15, 10, 5, and 2 [micro]mol/L), and the concentrations were calculated according to peak heights. The calibration curves for Hcy quantification were obtained by plotting the peak height ratio of Hcy/DP and Hcy/Cys. The equations, as a function of the Hcy concentration x (expressed in [micro]mol/L) were, respectively, as follows: y = 0.033x + 0.013 (r = 0.9998) for Hcy/DP and y = 0.011x + 0.003 (r = 0.9992) for Hcy/Cys.

The detection response was linear over the concentration range between 2 and 200 [micro]mol/L for all the plasma thiols. The detection limit for Hcy, defined as a signal-to-noise ratio >3, was -0.25 [micro]mol/L in plasma and was quite identical for the other thiols. The detection limit of the diluted sample injected (after the 5000 fold-dilution) was 50 pmol/L, corresponding to a quantity of 1.1 attomoles.

The within-day reproducibility of the assay for Hcy was determined on aliquots (n = 10) prepared independently from the same plasma sample, and the between-day reproducibility was calculated from the analysis of the same plasma sample that was derivatized each day during 10 consecutive days. The coefficients of variation (CVs) which take into account every step (i.e., reduction, precipitation, centrifugation, derivatization, and injection into the CE-LIF) were 4.9% and 7.8%, respectively, whereas the CVs for injection only were 1.8% and 2.6%.

[FIGURE 1 OMITTED]

To determine the recovery of the method, we added 10 [micro]L of different concentrations of Hcy (2, 5, 10, 15, 20, 25, 30, 35, 40, 45, and 50 [micro]mol/L) to 90 [micro]L of the same pooled plasma. The average recovery was 100.5% [+ or -] 4.87% (mean [+ or -] SD). The recovery (mean [+ or -] SD) of Hcy added to six plasma samples analyzed in triplicate (Table 1) was 101.1% [+ or -] 6.99% (n = 18).

We used this CE-LIF technique to analyze serum samples from 55 fasting subjects with episodes of deep-vein thrombosis or arterial occlusive disease, and we compared the results with those obtained from an ion-exchange amino acid analyzer (Beckman 6300 analyzer). The mean control values (n = 27) for the amino acid analyzer method (y) were 8.89 [+ or -] 2.85 [micro]mol/L vs 11.73 [+ or -] 4.45 [micro]mol/L for CE-LIF (x). The regression equation was as follows: y = 0.91x + 2.26; n = 55; [S.sub.y|x] =7.49; r = 0.87, which is comparable to previous observations (5, 8,10).

The same calibrator, i.e., homocystine (Sigma), was used in both methods: in the ion-exchange chromatography method, homocystine was diluted in lithium citrate buffer (0.2 mol/L, pH 7.0), and the calibrator curve in aqueous solution was determined using the ninhydrin reaction (specific for amine groups); in the CE-LIF method, homocystine was diluted in carbonate buffer (pH 9.5), and the calibrator curve in plasma solution was determined using IAF, which reacts specifically with thiol groups.

The major advantage of this IAF method is its ability to quantify Hcy, Cys, and glutathione simultaneously in a single analysis. Our derivatized thiols were stable for several months. The method is thiol-specific and the detection limit (0.25 [micro]mol/L Hcy in plasma) is similar to other HPLC or amino acid analyzer procedures (4,5). The IAF method saves time compared with the HPLC method, and we encountered no problems with interference from the reducing agent. When compared with the fluorescein isothiocyanate method, described previously (11), which quantifies Hcy and all amino acids through the reaction with amine groups, the IAF-based procedure allows a more selective detection of thiol-containing amino acids. In addition, in the IAF method, the migration times are shorter, the peak resolution is better, and co-injection of a calibrator is not required. The use of an internal standard [DP or homocysteic acid (11)] also improves the accuracy and the standardization (12) of the method.

The IAF method is simple, low in reagent cost (11), sensible, reproducible, and suitable for routine determination of serum/plasma Hcy concentrations in a clinical laboratory. This method will allow us to investigate and evaluate the potential atherogenic properties of Hcy.

This work was supported by grants from the Conseil Regional Midi-Pyrenees. We thank L. Casey, Washington University, St. Louis, MO for help with manuscript preparation.

References

(1.) Boushey C, Beresford S, Omenn G, Motulsky A. A quantitative assessment of plasma homocysteine as a risk factor for vascular disease. Probable benefits of increasing folic acid intakes. JAMA 1995;274:1049-57.

(2.) Robinson K, Mayer EL, Miller DP, Green R, van Lente F, Gupta A, et al. Hyperhomocysteinemia and low pyridoxal phosphate. Common and independent reversible risk factors for coronary artery disease. Circulation 1995; 92:2825-30.

(3.) Rasmussen K, Moller J, Lyngbak M, Pedersen A-M, Dybkjaer L. Age- and gender-specific reference intervals for total homocysteine and methylmalonic acid in plasma before and after vitamin supplementation. Clin Chem 1996;42:630-6.

(4.) Pastore A, Massoud R, Motti C, Lo Russo A, Fucci G, Cortese C, Federici G. Fully automated assay for total homocysteine, cysteine, cysteinylglycine, glutathione, cysteamine, and 2-mercaptopropionylglycine in plasma and urine. Clin Chem 1998;44:825-32.

(5.) te Poele-Pothoff MT, van den Berg M, Franken DG, Boers GH, Jakobs C, Kroon IF, et al. Three different methods for the determination of total homocysteine in plasma. Ann Clin Biochem 1995;32:218-20.

(6.) Feussner A, Rolinski B, Weiss N, Deufel T, Wolfram G, Roscher AA. Determination of total homocysteine in human plasma by isocratic high-performance liquid chromatography. Eur J Clin Chem Clin Biochem 1997; 35:687-91.

(7.) Pietzsch J, Julius U, Hanefeld M. Rapid determination of total homocysteine in human plasma by using N(O,S)-ethoxycarbonyl ethyl ester derivatives and gas chromatography-mass spectrometry. Clin Chem 1997;43:2001-4.

(8.) Ueland PM, Refsum H, Stabler SP, Malinow MR, Andersson A, Allen RH. Total homocysteine in plasma or serum: methods and clinical applications. Clin Chem 1993;39:1764-79.

(9.) Gilfix BM, Blank DW, Rosenblatt DS. Novel reductant for determination of total plasma homocysteine. Clin Chem 1997;43:687-8.

(10.) Cobbaert C, Arentsen J, Mulder P, Hoogerbrugge N, Lindemans J. Significance of various parameters derived from biological variability of lipoprotein(a), homocysteine, cysteine, and total antioxidant status. Clin Chem 1997;43:1958-64.

(11.) Causse E, Terrier R, Champagne S, Nertz M, Valdigui6 P, Salvayre R, Couderc F. Quantitation of homocysteine in human plasma by capillary electrophoresis and laser-induced fluorescence detection. J Chromatogr A 1998;817:181-5.

(12.) Kuo K, Still R, Cale S, McDowell I. Standardization (external, internal) of HPLC assay for plasma homocysteine. Clin Chem 1997;43:1653-5.

Elizabeth Causse, [1] * Nathalie Siri, [1] Helene Bellet, [2] Sandrine Champagne, [1] Christophe Bayle, [1] Pierre Valdiguie, [1] Robert Salvayre, [1] and Francoise Couderc [4] ([1] Laboratoire de Biochimie/Institut National de la Sante et de la Recherche Medicale, U466, Centre Hospitalier Rangueil, 1 avenue J. Poulhes, 31403 Toulouse Cedex 04, France; [2] Laboratoire de Biochemie, Centre Hospitalier St Elio, 34295 Montpellier, France; [3] ZETA Technology, Parc Technologique du Canal, 31520 Ramonville, France; [4] Universite Paul Sabatier, Institut de Pharmacologie et de Biologie Structurale, 205 Route de Narbonne, 31077 Toulouse, France; author for correspondence: fax 33 5 61 322953)
Table 1. Recovery of the assay.

 Concentration, [micro]mol/L Mean
Patient Sample Added Measured recovery, %

1 9.6 5 14.0 96.0
2 28.5 5 31.3 93.4
3 5.6 10 16.6 106.4
4 68.5 20 84.4 95.4
5 9.6 25 36.2 104.6
6 20.3 40 66.6 110.4
COPYRIGHT 1999 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 1999 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Technical Briefs
Author:Causse, Elizabeth; Siri, Nathalie; Bellet, Helene; Champagne, Sandrine; Bayle, Christophe; Valdiguie
Publication:Clinical Chemistry
Date:Mar 1, 1999
Words:1950
Previous Article:The international normalized ratio during concurrent warfarin and argatroban anticoagulation: differential contributions of each agent and effects of...
Next Article:Validation of a point-of-care assay for the urinary albumin:creatinine ratio.
Topics:


Related Articles
Urinary excretion of homocysteine-thiolactone in humans.
Stabilization of homocysteine in unseparated blood over several days: a solution for epidemiological studies.
Quantification of thiol-containing amino acids linked by disulfides to LDL.
Comparison of three different plasma homocysteine assays with gas chromatography--mass spectrometry.
Plasma homocysteine and L-DOPA metabolism in patients with Parkinson disease.
Riboflavin as a determinant of plasma total homocysteine: effect modification by the methylenetetrahydrofolate reductase C677T polymorphism.
New enzymatic colorimetric assay for total homocysteine.
Total homocysteine in pediatric patients.
Homocysteine and vitamins in cardiovascular disease.
Analysis of cyanine dye-labeled PCR product and restriction fragments by capillary electrophoresis and laser-induced fluorescence.

Terms of use | Privacy policy | Copyright © 2020 Farlex, Inc. | Feedback | For webmasters