Performance evaluation and subsequent clinical experience with the Abbott automated architect STAT troponin-I assay.

Cardiac troponins are specific biochemical markers of myocardial injury with clinical applications in the diagnosis of acute myocardial disease and cardiac risk stratification. In their consensus redefinition of myocardial infarction (1), the European Society of Cardiology and the American College of Cardiology defined an increased troponin value as one greater than the 99th percentile of a healthy reference population with the added recommendation that assay imprecision (CV) at this concentration be [less than or equal to]10%. This emphasizes the importance of performance at the low end of the measuring range if one is to avoid misclassification of patients. Currently available troponin assays fail to achieve such precision at their 99th percentile reference values (2), and laboratories using these assays quote higher decision limits. Accuracy is another issue, with past concerns regarding antibody specificity and standardization now partially addressed (3). However, reports of interference from rheumatoid factor, heterophile antibodies, autoantibodies, and hemolysis (4-11) emphasize the potential for both false-positive and false-negative results to influence therapeutic decisions and the importance of the serial rise and fall of troponin concentrations in the circulation after myocardial infarction.

The Architect STAT Troponin-I assay (Abbott Diagnostics) is a 2-step chemiluminescent microparticle immunoassay designed to detect cardiac troponin I (cTnI) in serum and plasma. We assessed the assay on the ci8200, particularly evaluating the performance at the low end of measurement and in response to interfering substances. We also report on the reference limit for a healthy population, the correlation with the Access second-generation AccuTnI (Beckman Coulter) and the AxSYM 1st-generation cTnI (Abbott Diagnostics) assay, and the detection of interference in the assay. The Architect STAT Troponin-I assay uses 3 murine monoclonal antibodies to the cTnI molecule. Two act as capture antibodies coating the paramagnetic microparticles, and the third is acridinium-labeled. The assay is traceable to the AACC candidate reference material for cTnI (human ICT tertiary complex; NIST SRM 2921) with a calibrator range of 0-50.00 [micro]g/L (as reported by the manufacturer).

Analytical evaluation was performed to Clinical and Laboratory Standards Institute (CLSI; formerly NCCLS) protocols. The analytical detection limit was defined as the mean value 2 SD above the mean of 10 replicates of calibrator A (0 [micro]g/L). Functional sensitivity was determined by measuring 10 samples between 0.01 and 0.10 [micro]g/L, assayed in duplicate daily for 10 days. For total imprecision, 3 concentrations of the manufacturer's quality-control material were measured in duplicate in 2 runs each day for 5 days. Reference intervals were determined on freshly collected serum and plasma from 480 consenting donors from the Australian Red Cross Blood Bank. The 99th percentile reference limits were calculated by a simple nonparametric approach.

Correlation experiments with the AxSYM cTnI assay were performed on 296 consecutive hospital plasma samples with cTnI requests, of which the Emergency and Cardiology Departments accounted for ~70%. When testing was not possible within 4 h, plasma was stored (-30 [degrees]C) for later analysis. For correlation with the Access AccuTnI assay, we compared plasma samples from 43 patients with cTnI concentrations [less than or equal to]0.10 [micro]g/L. In addition, 25 stored samples from 7 patients with false-positive AXSYM cTnI results (negative clinical history and undetectable cTnI in 2 other assays) were analyzed with the Architect assay to study immunoassay interference. During the study, lithium heparin was the anticoagulant used for all plasma samples, and heterophile blocking tubes (Scantibodies Clinical Laboratories) were used to investigate the presence of heterophile antibody interference. This study was approved by the Alfred Hospital ethics committee.

The Architect STAT Troponin-I assay had a limit of detection of 0.004 [micro]g/L and a CV of 10% at concentrations approaching 0.03 [micro]g/L (Fig. 1A). The total CV did not exceed 6% at concentrations of 0.12, 0.45, and 12.04 [micro]g/L. The healthy reference population had a 2:1 male preponderance (age range, 16-82 years; mean, 47 years for males and 41 years for females). Distribution of cTnI results was slightly lower in serum than for plasma (detection limit, 90th percentile for serum, 75th percentile for plasma; Wilcoxon rank test, P <0.01) with the 99th percentiles in serum (n = 475) and in plasma (n = 480) both at 0.03 [micro]g/L. Matched patient serum and plasma samples across a wide range of cTnI results (0.1-85 [micro]g/L) were in excellent agreement [Deming regression: serum cTnI = 1.03(plasma cTnI) - 0.065 [micro]g/L; correlation coefficient (r) = 0.999, n = 32].

[FIGURE 1 OMITTED]

The correlation of the Architect TnI assay with the AxSYM cTnI assay was good across the clinical measurement range [Deming regression: Architect = 0.11(AxSYM) + 0.28 [micro]g/L; correlation coefficient (r) = 0.98] but weaker at concentrations below the AxSYM cutoff (2 [micro]g/L) for acute myocardial infarction [Architect = 0.18(AxSYM) + 0.02 [micro]g/L; r = 0.65]. Of 296 consecutive patient samples, 245 (82.8%) were either positive or negative on both assays. In 51 discrepant samples (17.2%), the AxSYM assay measured increased cTnI in 3 samples ([greater than or equal to]0.6 [micro]g/L), whereas the Architect assay was below the 99th percentile value of the reference population. Conversely, the Architect assay results were increased (i.e., [greater than or equal to]0.03 [micro]g/L) for 48 samples that were negative by the AxSYM method. At the time of evaluation, clinical decisions were based on the results of the AxSYM assays. Two of the 3 increased AxSYM samples were taken 7-10 days after myocardial infarction. Time course data obtained with serial samples from patients with myocardial infarctions did not show any time difference between AxSYM and Architect results (data not shown). The other positive AxSYM result was likely a false positive (no evidence of cardiac ischemia on nuclear medicine imaging). Of the 48 samples with increased cTnI by the Architect assay only, clinical files were available for 28 samples (25 patients) and were reviewed by 2 physicians. The Architect assay identified 6 patients with a new acute coronary syndrome/acute myocardial infarction, and an additional 7 samples were obtained from patients with acute pulmonary edema, arrhythmia, or chest trauma or after cardiac procedures. In 3 samples, other AxSYM results from the same clinical episode showed increased troponin. One sample was taken from a patient with severe illness who subsequently died, and 2 samples were collected from patients with renal impairment. Six of the 48 samples (in 4 patients) were likely to be true AxSYM negatives because there were no detected cardiac events during follow-up. The Architect values in these 6 samples ranged from 0.04 to 0.09 [micro]g/L. These findings highlight the potential of the Architect assay to reclassify patients previously labeled as "normal". Of the 25 specimens with known AxSYM cTnI assay interference, all had Architect STAT cTnI concentrations below the reference population.

Deming regression for sample comparison with the AccuTnI assay at plasma concentrations <0.10 [micro]g/L showed good agreement [Architect = 1.06(Beckman) -0.01 [micro]g/L; r = 0.83; Fig. 1B]. The published limit of detection for the AccuTnI assay is <0.01 [micro]g/L, with the concentration giving a CV of 10% reported as 0.06 [micro]g/L (12). No clinical information was obtained regarding the outcome of these patients.

The Architect STAT Troponin-I assay has now been in routine clinical use in our hospital for almost 12 months with wide acceptance of the new assay and the change to a reference interval in line with the European Society of Cardiology and American College of Cardiology guidelines. Using a commercial control serum (Abacus MAS) at a concentration of 0.05 [micro]g/L, we obtained long-term CVs between 12% and 20% over 12 months. As expected, long-term CVs have been higher than that obtained in our initial evaluation. During this time, we have identified 6 patients with suspected false-positive results (range, 0.1-0.6 [micro]g/L) based on discrepancies between results and clinical information and persistently increased TnI concentrations. Normal TnI concentrations were obtained in these samples after preincubation with heterophile blocking tubes and in other troponin assays. The theoretical incidence of these heterophile antibodies in our population is 1 in 5000.

The Architect STAT Troponin-I assay provides measurement of cTnI in plasma with a CV of 10% near the 99th percentile of a reference group of healthy blood bank donors (0.03 [micro]g/L). Our 99th percentile value is higher than that reported by the manufacturer (0.012 [micro]g/L) and may be attributable to our use of freshly collected blood bank specimens.

Comparison with the clinically evaluated AxSYM cTnI assay (13-15) showed that the Architect STAT Troponin-I assay identified additional patients who had clinical evidence of cardiac damage. This is in keeping with studies showing that the AxSYM assay may miss patients who later developed poor cardiac outcomes (16). Furthermore, the Architect TnI showed good agreement with the Access AccuTnI at the lower end of the measurement range. Similar data comparing the Architect assay with the Advia Centaur assay were reported recently (17). Identification of such patients is important given the likely benefit from early intervention.

Although the Architect STAT Troponin-I assay was not affected by any of the substances that cause interference in the AxSYM cTnI assay, our 6 cases of heterophile antibody interference are reminders that interference from these and like substances remains an issue. Consistently increased troponin concentrations without a pattern of increase and decrease should be questioned by laboratory staff. Laboratories should have documented procedures to investigate suspicious results such as access to different troponin methodologies in real time.

We thank Abbott Diagnostics for supplying some of the reagents, Drs. Z. Lu and J. Newman for performing the Access AccuTnI measurements, and Dr. Al-Kindi for performing the serum troponin measurements on the Architect.

References

(1.) Alpert JS, Thygesen K, Antman E, Bassand JP. Myocardial infarction redefined: a consensus document of the Joint European Society of Cardiology/ American College of Cardiology Committee for the Redefinition of Myocardial Infarction. J Am Coll Cardiol 2000;36:959-69.

(2.) Panteghini M, Pagani F, Yeo KT, Apple FS, Christenson RH, Dati F, et al. Committee on Standardization of Markers of Cardiac Damage of the IFCC. Evaluation of imprecision for cardiac troponin assays at low-range concentrations. Clin Chem 2004;50:327-32.

(3.) Panteghini M. Current concepts in standardization of cardiac marker immunoassays. Clin Chem Lab Med 2004;42:3-8.

(4.) Eriksson S, Junikka M, Laitinen P, Majamaa-Voltti K, Alfthan H, Pettersson K. Negative interference in cardiac troponin I immunoassays from a frequently occurring serum and plasma component. Clin Chem 2003;49:1095-104.

(5.) Knoblock RJ, Lehman CM, Smith RA, Apple FS, Roberts WL. False-positive AxSYM cardiac troponin I results in a 53-year-old woman. Arch Pathol Lab Med 2002;126:606-9.

(6.) Katwa G, Komatireddy G, Walker SE. False-positive elevation of cardiac troponin I in seropositive rheumatoid arthritis. J Rheumatol 2001;28: 2750-1.

(7.) Fitzmaurice TF, Brown C, Rifai N, Wu AH, Yeo KT. False increase of cardiac troponin I with heterophilic antibodies. Clin Chem 1998;44:2212-4.

(8.) Eriksson S, Hellman J, Pettersson K. Autoantibodies against cardiac troponins. N Engl J Med 2005;352:98-100.

(9.) Eriksson S, Halenius H, Pulkki K, Hellman J, Pettersson K. Negative interference in cardiac troponin I immunoassays by circulating troponin autoantibodies. Clin Chem 2005;51:839-47.

(10.) Eriksson S, Ilva T, Becker C, Lund J, Porela P, Pulkki K, et al. Comparison of cardiac troponin I immunoassays variably affected by circulating autoantibodies. Clin Chem 2005;51:848-55.

(11.) Hawkins RC. Hemolysis interference in the Ortho-Clinical Diagnostics Vitros ECi cTnI assay [Letter]. Clin Chem 2003;49:1226.

(12.) Uettwiller-Geiger D, Wu AH, Apple FS, Jevans AW, Venge P, Olson MD, et al. Multicenter evaluation of an automated assay for troponin I. Clin Chem 2002;48:869-76.

(13.) Heeschen C, Hamm CW, Goldmann B, Deu A, Langenbrink L, White HD. Troponin concentrations for stratification of patients with acute coronary syndromes in relation to therapeutic efficacy of tirofiban. PRISM Study Investigators: Platelet Receptor Inhibition in Ischemic Syndrome Management. Lancet 1999;354:1757-62.

(14.) Apple FS, Maturen AJ, Mullins RE, Painter PC, Pessin-Minsley MS, Webster RA, et al. Multicenter clinical and analytical evaluation of the AxSYM troponin-I immunoassay to assist in the diagnosis of myocardial infarction. Clin Chem 1999;45:206-12.

(15.) Wilcox G, Archer PD, Bailey M, Dziukas L, Lim CF, Schneider HG. Measurement of cardiac troponin I concentrations in the emergency department: predictive value for cardiac and all-cause mortality. Med J Aust 2001;174: 170-3.

(16.) Venge P, Lagerqvist B, Diderholm E, Lindahl B, Wallentin L. Clinical performance of three cardiac troponin assays in patients with unstable coronary artery disease (a FRISC II substudy). Am J Cardiol 2002;89:1035-41.

(17.) Boyd JC, Henrich M, Schubert MT, Shumacher NR. Evaluation of a sensitive rapid assay for cardiac troponin I (cTnI) as implemented on the ARCHITECT ci8200 analyzer [Abstract]. Clin Chem 2005;51(Suppl 6):A19.

DOI: 10.1373/clinchem.2005.057216

Que Lam, [1] Marion Black, [1] Odette Youdell, [1] Heather Spilsbury, [1] and Hans-Gerhard Schneider [1,2] * ([1] Clinical Biochemistry Unit, Alfred Pathology Service, The Alfred Hospital, and [2] Department of Medicine, Monash University, Melbourne, Australia; * address correspondence to this author at: Clinical Biochemistry Unit, Alfred Pathology Service, Commercial Road, Melbourne, Victoria 3004, Australia; fax 61-3-92763424, e-mail schneiderh@alfred.org.au)
COPYRIGHT 2006 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2006 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Technical Briefs
Author:Lam, Que; Black, Marion; Youdell, Odette; Spilsbury, Heather; Schneider, Hans-Gerhard
Publication:Clinical Chemistry
Date:Feb 1, 2006
Words:2244
Previous Article:Feasibility study of new calibrators for thyroid stimulating hormone (TSH) immunoprocedures based on remodeling of recombinant TSH to mimic...
Next Article:Atorvastatin reduces the expression of COX-2 mRNA in peripheral blood monocytes from patients with acute myocardial infarction and modulates the...
Topics:


Related Articles
Use of the Centaur TnI-Ultra assay for detection of myocardial infarction and adverse events in patients presenting with symptoms suggestive of acute...
Point-of-care i-STAT cardiac troponin I for assessment of patients with symptoms suggestive of acute coronary syndrome.
Selection of antibodies and epitopes for cardiac troponin immunoassays: should we revise our evidence-based beliefs?
New generation cardiac troponin I assay for the access immunoassay system.
Near-bedside whole-blood cardiac troponin I assay for risk assessment of patients with acute coronary syndromes.
Clinical efficacy of three assays for cardiac troponin I for risk stratification in acute coronary syndromes: a thrombolysis in myocardial infarction...
False increase of cardiac troponin I with heterophilic antibodies.
Characteristics of an Albumin Cobalt Binding test for assessment of acute coronary syndrome patients: a multicenter study.
Multicenter evaluation of an automated assay for troponin I.

Terms of use | Privacy policy | Copyright © 2021 Farlex, Inc. | Feedback | For webmasters