PRESSURE DROP OF FILTERING FACEPIECE RESPIRATORS: HOW LOW SHOULD WE GO?

INTRODUCTION

Filtering facepiece respirators (FFRs) are the most commonly used respiratory protective devices in U.S. private industry and healthcare, and the class N95 FFR (essentially equivalent to the FF P2 class specified in European standard EN 149) is the single most used version [1]. The United States National Institute for Occupational Safety and Health (NIOSH) N95 FFR certification testing (utilizing a filter tester at 85 l/min of constant airflow) specifies peak average inhalation and exhalation resistance to airflow (i.e., filter airflow resistance--[R.sub.filter]) limits of 35 mm (343.2 Pa) and 25 mm (245.1 Pa) [H.sub.2]O pressure, respectively, for this class of respirators [2]. Despite these relatively low [R.sub.filter] (defined as the pressure drop/flow rate) limits, complaints of difficulty breathing when wearing N95 FFRs have been reported by a sizeable number of healthcare workers [3-5].

Unfortunately, there are limited human data available with regard to actual [R.sub.filter] encountered across N95 FFR filters. Jones [6] reported that the peak average inhalation and exhalation [DELTA]P of 1 early model FFR was related to work load and ranged 0-20 mm (0-196.1 Pa) [H.sub.2]O pressure in the subjects tested at rest and during mild, moderate and heavy workloads, but the route of breathing (i.e., nasal, oral, oronasal) was not identified.

This investigation, part of a larger study and some of the results which have previously been reported [7], was undertaken by the National Personal Protective Technology Laboratory (NPPTL) of NIOSH and a respirator manufacturer (3M Company, St. Paul, Minnesota, U.S.) to determine mean peak inhalation and exhalation [R.sub.filter] during oral and nasal breathing by subjects wearing a prototype FFR configured in 3 low filter resistances. These data could be of interest to researchers, manufacturers, standards development organizations, respiratory protection program managers, and end users.

MATERIAL AND METHODS

Ten healthy, non-smoking subjects (7 men, 3 women) participated in this study, the majority of whom (8/10) were experienced N95 FFR users. Demographic mean values ([+ or -] standard deviation) were: age--24.5 years ([+ or -] 3.8), height--179 cm ([+ or -] 11), weight--75.3 kg ([+ or -]12.4), and body mass index--23.4 ([+ or -] 2.9). On the day of testing, the subjects underwent a screening history and physical examination by a licensed physician. The subjects were dressed in athletic shorts or pants, tee shirts and athletic shoes during testing. The study procedures were in accordance with the ethical standards of the Helsinki Declaration of 1975, as revised in 1983, and were approved by the NIOSH Human Subjects Review Board, with all the subjects providing oral and written informed consent.

A cup-shaped, prototype FFR (PR) from 3M Company (St. Paul, MN, U.S.) was supplied to NPPTL in 3 different nominal [R.sub.filter] of 29.4 Pa, 58.8 Pa and 88.2 Pa at 85 l/min constant airflow (hereafter referred to as PR3, PR6, and PR9, respectively), achieved through modifications of the filter material. None of the PRs was equipped with an exhalation valve. Prior to the study trials, [R.sub.filter] were verified at NPPTL by testing with a TSI 8130 automated filter tester (TSI, Shoreview, MN, U.S.), the same equipment utilized by 3M Company for its determination of the PR resistances. Results indicated [R.sub.filter] of 35.3 Pa, 63.7 Pa and 91.2 Pa for PR3, PR6 and PR9 at 85 l/min constant airflow, respectively. The subjects underwent standard U.S. Occupational Safety and Health Administration (OSHA) respirator quantitative fit testing exercises [8] for each of the PRs with the TSI Portacount[R]Pro+ Respirator Fit Tester Model 8038 (TSI, Shoreview, MN, U.S.), in the "plus" mode, that counts individual airborne particles to determine a quantitative estimate of respirator fit. The ratio of measured ambient particles to within-respirator particles is termed the "fit factor" and is, in part, a measure of the adequacy of the seal of the respirator to the face. In order to pass fit testing, the subjects had to achieve a score of [greater than or equal to] 100, the passing score for an OSHA respirator quantitative fit test that is indicative of [less than or equal to] 1% entry of particles into the respirator wearer's breathing zone [8]. All the subjects passed quantitative fit testing on each of the PRs. A Validyne (Northridge, CA, U.S.) Model DP45 low pressure transducer, calibrated to a water column, was connected to a centrally-placed metal grommet in the PRs via 1/8 inch internal diameter, flexible, plastic tubing (150 cm in length and supported by a stand midway between the transducer and the subject to prevent kinking) and collected [R.sub.filter] measurements at 10 samples-per-second.

The transducer measured sample [R.sub.filter] relative to ambient atmospheric pressure and the transducer's output was converted to an electrical signal by a Validyne CD23 digital transducer connected to a computer with a dedicated LabView[R] data collection program (National Instruments Corporation, Austin, TX, U.S.). The [R.sub.filter] was recorded on LabView[R] as negative numbers for inhalation and positive numbers for exhalation and the reported [R.sub.filter] represents the average of the pressures during a breath. The subjects donned the individual PRs as per the manufacturer's instructions, adjusted the pliable nose bar, performed negative and positive user seal checks to evaluate the seal of the respirator to the face [8], and underwent a 2 min acclimatization period to reach steady state [9].

The [R.sub.filter] measurements were then taken with the sedentary subjects standing and instructed to breathe only nasally for 30 s followed immediately by only oral breathing for 30 s. The subjects were then seated on a Kettler RX7 reclining bicycle ergometer (Ense-Parsit, Germany) and pedaled at a low-moderate work rate (50 watt resistance, 60 revolutions-per-minute) for 2 min to achieve stabilization of the respiratory rate and, with continued pedaling, were then instructed to breathe only nasally over a 30 s period followed immediately by only oral breathing for 30 s while [R.sub.filter] measurements were taken. This scenario was repeated for each of the 3 PRs, with a minimum 5 min respite between each of the trials.

Statistical analysis

Mean peak inhalation and exhalation [R.sub.filter] values, measured for 30 s, were separated and first averaged for statistical analysis, followed by the calculation of the group mean, standard deviation and 95% confidence interval (CI). Two-way ANOVA (3 resistance levels x 2 exercise states) was carried out for oral and nasal breathing to determine the main effect of the 3 filter resistances on the subjects' [R.sub.filter] during sedentary activity and exercise. For a significant F value, post-hoc multiple comparisons for the observed means were subsequently performed with Sidak corrections for 95% confidence intervals. Mean peak inhalation and exhalation mean [R.sub.filter] comparisons between nasal and oral breathing were carried out by paired samples t-test. All statistical analyses were performed using SPSS v.19 (IBM, Somers, New York, U.S.).

RESULTS

Nasal breathing

The mean peak [R.sub.filter] of the PRs had a significant impact on the subjects' [R.sub.filter] during nasal inhalation (F = 29.66, p < 0.001) and nasal exhalation (F = 11.01, p < 0.001). The mean peak [R.sub.filter] was greater during exercise trials than during sedentary activity for nasal inhalation (F = 25.77, p < 0.001) and nasal exhalation (F = 16.62, p < 0.001). Pairwise comparisons (mean difference, CIs) of [R.sub.filter] during nasal inhalation were 29.4 Pa/lx[s.sup.-1] < 58.8 Pa/lx[s.sup.-1] (24.1, 15.2-32.8) and 29.4 Pa/lx[s.sup.-1] < 88.2 Pa/lx[s.sup.-1] (23.5, 14.7-32.2), and during nasal exhalation they were 29.4 Pa/lx[s.sup.-1] < 58.8 Pa/lx[s.sup.-1] (16.6, 7.0-26.3) and 29.4 Pa/lx[s.sup.-1] < 88.2 Pa/lx[s.sup.-1] (15.2, 5.58-24.9). Pair-wise comparisons of mean peak [R.sub.filter] during activity are Sedentary < Exercise (14.8, 8.9-20.5) for nasal inhalation and Sedentary < Exercise (13.0, 6.6-19.5) for nasal exhalation (Table 1).

Oral breathing

The mean peak [R.sub.filter] of the PRs had a significant impact on subjects' [R.sub.filter] during oral inhalation (F = 31.28, p < 0.001) and oral exhalation (F = 23.23, p < 0.001). The mean peak [R.sub.filter] was greater during exercise trials than during sedentary activity for oral inhalation (F = 40.18, p < 0.001) and oral exhalation (F = 33.42, p < 0.001).

Pairwise comparisons (mean difference, CIs) of mean peak [R.sub.filter] during oral inhalation were 29.4 Pa/lx[s.sup.-1] < 58.8 Pa/lx[s.sup.-1] (24.8, 15.3-34.2) and 29.4 Pa/lx[s.sup.-1] < 88.2 Pa (27.3, 17.9-36.7), and during oral exhalation they were 29.4 Pa/lx[s.sup.-1] < 58.8 Pa/lx[s.sup.-1] (19.5, 10.4-28.5) and 29.4 Pa/lx[s.sup.-1] < 88.2 Pa/lx[s.sup.-1] (23.3, 14.3-32.3).

Pairwise comparisons of mean peak [R.sub.filter] during activity are Sedentary < Exercise (19.8, 13.5-25.9) for oral inhalation and Sedentary < Exercise (17.3, 11.3-23.3) for oral exhalation (Table 1).

DISCUSSION

The mean peak [R.sub.filter] measurements in the current study represent the actual resistance to airflow through the filter material of the PRs during subject wear. The study data indicate that there were no significant differences in nasal and oral mean peak [R.sub.filter] when comparing PR6 and PR9 (Figure 1), whereas PR3 demonstrated significantly lower mean peak [R.sub.filter] compared with PR6 and PR9 (p < 0.000 for both comparisons) (Table 1; Figures 1 and 2). Some inter-subject variability was noted and is represented in the width of the confidence intervals (Table 1) and visually in Figure 1. Nonetheless, our previously reported data from this study [7] showed that there were no significant differences in physiological (i.e., respiratory rate, heart rate, oxygen saturation, transcutaneous carbon dioxide) and subjective responses (i.e., exertion, thermal comfort, inspiratory effort, expiratory effort, overall breathing discomfort) among the 3 PRs during 1 h of treadmill exercise at a low-moderate work rate (5.6 km/h).

The reason for this lack of difference in these responses may be related to the fact that the 3 PRs had mean peak [R.sub.filter] (Table 1) that was at, or below, the 58.8-74.5 Pa/lx[s.sup.-1] threshold level of detection for inspiratory resistance [10-15]. This suggests that the relatively common complaint by healthcare workers of difficulty breathing when wearing FFRs [3-5] with [R.sub.filter] similar to PR9 [16,17] are related to issues other than FFR-associated [R.sub.filter]. This further indicates that lowering [R.sub.filter] on FFRs below 88.2 Pa/lxs-1 may not be advantageous in terms of user physiological responses or subjective measures of comfort [7], parameters that are intimately connected to respirator compliance issues. The significantly higher mean peak [R.sub.filter] of low-moderate exercise compared with the sedentary state (Figure 2) when wearing a respirator, as reported in the current study, has been previously noted [18] and is a reflection of the increased airflow requirements of exercise.

The mean peak [R.sub.filter] during oral breathing was universally higher than during nasal breathing in the current study (Figures 1 and 2). Although the oral airway is generally considered a lower resistance breathing route than the nasal airway, this is not necessarily always the case [19]. In research studies, airway pressures are frequently measured using either a mouthpiece or a mask (as in the current study), with the former being associated with lower pressures due to its creating a fairly wide opening of the mouth [19,20]. Also, the function of the bottom strap to ensure that the FFR is drawn down and over the jawbone [21] may result in restriction of jaw articulation leading to narrowing of the oral aperture that significantly affects the resistance of the oral route [18]. The same mechanism of respirator-related restriction in jaw articulation has also been postulated as the cause of speech impairment when wearing various respirator facemasks [22,23]. Additionally, during exercise, nasal resistance can fall due to sympathetic vasoconstriction of the nasal mucosa [24]. Lastly, the trajectory of the forces generated by the upper strap of FFRs is directed upward and outward at the malar (cheekbone) regions of the face [25], placing variable tension upon the paranasal facial skin that can increase the opening of the nasal valve and decrease nasal resistance (lateral traction on the paranasal facial skin is the basis of the Cottle test for evaluation of the nasal valve patency) [26].

The similarity in the mean peak [R.sub.filter] of PR6 and PR9 was an unexpected finding. The [R.sub.filter] across a filter fabric, such as N95 FFR filters, is airflow velocity dependent, as expressed in the following formula: [DELTA][P.sub.f] = [K.sub.1]V (where K1 = fabric resistance and V = airflow velocity) [27,28]. When humans experience mild-to-moderate inspiratory resistive loads (as with wearing an FFR), the main immediate neural mechanism for stabilizing their tidal volume is prolongation of the inspiratory portion of the breathing cycle at the expense of expiration, termed an increase in the duty cycle (ratio of total time of inspiration to total respiratory time) [29-32]. This increase in the duty cycle may have a role in minimizing the tension of respiratory muscles [9] and results in almost no change in minute ventilation [30].

We have previously reported that the respiratory rate, tidal volume and minute ventilation of the current study's subjects at a low-moderate work rate (5.6 km/h) did not differ significantly while wearing the same 3 PRs [7]. Given that the [R.sub.filter] of filter fabrics is related to the air flow velocity [27], the similarity in inspiratory mean peak [R.sub.filter] of PR9 and PR6 indicates that the subjects wearing PR9 are sacrificing peak inspiratory airflow by prolonging the duty cycle to compensate for the greater mean peak [R.sub.filter] [33], a feature that may be imperceptible to the user. The similarity in exhalation mean peak [R.sub.filter] between the PR6 and PR9 may be related to the higher positive pressure generated during exhalation to overcome PR9 resistance that results in relatively greater disruption of the faceseal [34] and allows more exhaled air to follow the path of least resistance [35], thereby lowering mean peak [R.sub.filter] to the levels seen with PR6.

Prior research, using thermal imaging, showed that leakage at a single or multiple sites occurred during exhalation in majority of the study subjects wearing N95 FFRs during fit testing [36]. Thus, further study of human subjects to determine the impact of FFR [R.sub.filter] on faceseal leakage during exhalation may be warranted. Exhalation valveequipped FFRs limit the buildup of positive pressure during exhalation [34], such that comparison of FFR with and without an exhalation valve might elucidate the contribution of leaks to exhalation [DELTA]P.

Interestingly, several of the study subjects conveyed that their molding of the deformable nose bar to the nasal contour was perceived as creating some nasal blockage and thereby promoting oronasal or oral breathing when the PRs were 1st donned. Previous research has indicated that respirators can distort the nasal alae [18], an area that accounts for the major contribution of nasal airway resistance [37], increasing the work of breathing and leading in a switch to oral or oronasal breathing [38]. Thus, if pliable nose bars cause nasal obstruction of any significant degree, it is possible that the nasal airway inhalation and exhalation mean peak [R.sub.filter] values we reported could have been artificially elevated. A human subject study comparing FFRs with and without nasal bars would be required to definitively answer this question. If pliable nasal bars on FFRs are shown to affect nasal respiration, alternative design features to allow for conformity to the face without impingement (e.g., face seal adhesives, pre-molded nasal contours, etc.) might be an alternative, but this assumption requires further investigation. Additionally, given the current study findings of lower [DELTA]P for nasal breathing and the additional physiological benefits ascribed to nasal breathing (e.g., decrease in water loss, lower FFR deadspace humidity, air filtration, transport of nitric oxide to the lungs, etc.) [39], worker education in the proper use of FFRs might include information that the preferential route of breathing at low and moderate work rates is nasal, if tolerable.

Limitations of the current study include the relatively small number of participants (N = 10). However, the majority of the subjects (8/10) were experienced FFR users, thus offering some measure of confidence in the reliability of findings. We did not examine the impact of high workloads on the oral and nasal mean peak [R.sub.filter]; however, most current workers experience low and moderate work rates [40,41]. Also, none of the PRs was equipped with an exhalation valve that would likely have impacted exhalation measurements [34]. We tested only cupshaped respirators and cannot comment on the impact of different respirator designs (e.g., duckbill, flat fold, etc.) on [R.sub.filter]. We tested only N95 FFR class respirators and cannot comment on [R.sub.filter] of classes with higher filter resistances. Lastly, we did not measure the rate or depth of breathing to assist in determining differences between nasal and oral breathing so that the reported mean peak [R.sub.filter] values could have been influenced either by functional filter resistance (pressure drop) or differences in airflow velocity.

CONCLUSIONS

Mean peak [R.sub.filter] during oral and nasal inhalation and exhalation was significantly lower for PR3 compared with PR6 and PR9 at sedentary and low-moderate work rates. However, mean peak [R.sub.filter] for all 3 PRs was at, or below, the threshold limit for detection of inspiratory resistance. The nasal route of breathing was associated with lower [R.sub.filter] at both work rates for the 3 PRs. The route of respiration (oral versus nasal) when wearing an FFR impacts [R.sub.filter], and efforts to promote nasal breathing with the use of FFRs should be considered. Decreasing [R.sub.filter] of N95 FFRs below 88.2 Pa (measured at a constant airflow of 85 l/min) may not be of additional physiological or subjective benefit to the user [7].

http://dx.doi.org/10.13075/ijomeh.1896.00153

ACKNOWLEDGMENTS

The authors thank Richard Vojtko for his assistance with the pressure transducer setup and measurements and also Andrew Palmiero, MS and Drs. William King, Aitor Coca, and W. Jon Williams of NPPTL and Andrew Viner, MS, of 3M Company, for their manuscript reviews and suggestions. Jung-Hyun Kim, Raymond J. Roberge, Jeffrey B. Powell, and Ronald Shaffer of the National Personal Protective Technology Laboratory and Caroline M. Ylitao and John M. Sebastian of the 3M Company conducted the study, analyzed the results and prepared the manuscript.

REFERENCES

[1.] National Institute for Occupational Safety and Health. Respirator usage in private sector firms, 2001 [cited 2013 Feb 8]. Available from: http://www.cdc.gov/niosh/docs/respsurv.

[2.] Code of Federal Regulations (42 CFR 84.180). Airflow resistance tests [cited 2013 May 16]. Available from: http://www. gpo.gov/fdsys/pkg/CFR-2007-title42-vol1/pdf/CFR-2007-title42-vol1-sec84-180.pdf.

[3.] Bryce E, Forrester L, Scharf S, Eshqhpour M. What do healthcare workers think? A survey of facial protection equipment user preferences. J Hosp Infect. 2007;68(3):241-7, http:// dx.doi.org/10.1016/j.jhin.2007.12.007.

[4.] Baig A, Knapp C, Eagan AE, Radonovich LJ Jr. Health care workers' views about respirator use and features that should be included in the next generation of respirators. Am J Infect Control. 2010;38(1):18-25, http://dx.doi.org/10.1016/ j.ajic.2009.09.005.

[5.] Mitchell R, Ogunremi T, Astrakianakis G, Bryce E, Gervais R, Gravel D, et al. Impact of the 2009 influenza A (H1N1) pandemic on Canadian health care workers: A survey on vaccination, illness, absenteeism, and person protective equipment. Am J Infect Control. 2012;40(7):611-6, http://dx.doi. org/10.1016/j.ajic.2012.01.011.

[6.] Jones JG. The physiological cost of wearing a disposable respirator. Am Ind Hyg Assoc J. 1991;52(6):219-25, http:// dx.doi.org/10.1080/15298669191364631.

[7.] Roberge RJ, Kim J-H, Powell JB, Shaffer RE, Ylitalo CM, Sebastian JM. Impact of different breathing resistances on subjective and physiological responses to filtering facepiece respirators. PLoS One. 2013;8(12):e84901, http://dx.doi. org/10.1371/journal.pone.0084901.

[8.] Occupational Safety and Health Administration. Respiratory Protection Standard 1910.134 (1998) [cited 2013 Feb 11]. Available from: http://www.osha.gov/pls/oshaweb/owadisp. show_document?p_id=12716&p_table=standards.

[9.] Nishino T, Kochi T. Breathing route and ventilator responses to inspiratory resistive loading in humans. Am J Resp Crit Care Med. 1994;150(3):742-6, http://dx.doi.org/10. 1164/ajrccm.150.3.8087346.

[10.] Silverman L, Lee G, Plotkin T, Sawyers LA, Yancey AR. Air flow measurement on human subjects with and without respiratory resistance at several work rates. Arch Ind Hyg Occup Med. 1951;3(5):61-478.

[11.] Wiley RL, Zechman FW Jr. Perception of added airflow resistance in humans. Resp Physiol. 1966;2(1):73-87, http:// dx.doi.org/10.1016/0034-5687(66)90039-9.

[12.] Gottfried SB, Altose MD, Kelsen SG, Fogarty CM, Cherniak NS. The perception of changes in airflow resistance in normal subjects and patients with chronic airways obstruction. Chest. 1978;73(2 Suppl):286-8.

[13.] Campbell EJM, Gandevia SC, Killian KJ, Mahutte CK, Rigg JRA. Changes in the perception of inspiratory restive loads during partial curarization. J Physiol. 1980;309:93-100.

[14.] Killian KJ, Mahutte CK, Howell JBL, Campbell EJM. Effect of timing, flow, lung volume, and threshold pressures on restive load detection. J Appl Physiol. 1980;49(6):958-63.

[15.] Burki NK. Effects of added inspiratory loads on load detection thresholds. J Appl Physiol. 1981;50(1):162-4.

[16.] Vojtko MR, Roberge MR, Vojtko RJ, Roberge RJ, Landsittel DP. Effect on breathing resistance of a surgical mask worn over a N95 filtering facepiece respirator. J Intern Soc Resp Protect. 2008;25(1):1-8.

[17.] Sinkule EJ, Powell JB, Goss FL. Evaluation of N95 respirator use with a surgical mask cover: Effects on breathing resistance and inhaled carbon dioxide. Ann Occup Hyg. 2013;57:384-98, http://dx.doi.org/10.1093/annhyg/mes068.

[18.] Harber PJ, Beck J, Luo J. Study of respirator effect on nasal-oral flow partition. Am J Ind Med. 1997;32(4):408-12, http://dx.doi.org/10.1002/(SICI)1097-0274(199710)32:4% 3C408::AID-AJIM12%3E3.0.CO;2-7.

[19.] Amis TC, O'Neill N, Wheatley JR. Oral airway flow dynamics in healthy humans. J Physiol. 1999;515(Pt 1):293-8, http://dx.doi.org/10.1111/j.1469-7793.1999.293ad.x.

[20.] Wong LS, Johnson AT. Decrease of resistance to air flow with nasal strips as measured with the airflow perturbation device. Biomed Eng Online. 2004;3(1):38, http://dx.doi. org/10.1186/1475-925X-3-38.

[21.] Yassi A, Bryce E, Moore D. Protecting the faces of health care workers: Knowledge gaps and research priorities for effective protection against occupationally-acquired respiratory infectious diseases. Occupational Health and Safety Agency for Healthcare in British Columbia (2004) [cited 2013 May 2]. Available from: https://circle.ubc.ca/ bitstream/handle/2429/849/Protecting_Faces_Final_Report. pdf?sequence=1.

[22.] Lubker JF, Moll KL. Simultaneous oral-nasal air flow measurements and cinefluorographic observations during speech production. Proceedings of the American Speech and Hearing Association Convention; 1963 Nov 3-6; Chicago (IL). Iowa: University of Iowa; 1964. Available from: http://digital.library.pitt.edu/c/cleftpalate/pdf/ e20986v02n3.06.pdf.

[23.] Wetherell A. The UK general service respirator. Porton Down, Salisbury: Defence Scientific and Technical Laboratory; 2003 [cited 2013 May 2]. Available from: http://www. dtic.mil/cgi-bin/GetTRDoc?AD=ADA452235.

[24.] Gehring JM, Garlick SR, Wheateley JR, Amis TC. Nasal resistance and flow resistive work of nasal breathing during exercise: Effects of a nasal dilator strip. J Appl Physiol. 2000;89(3):1114-22.

[25.] Niezgoda G, Benson SM, Eimer BC, Roberge RJ. Forces generated by N95 filtering facepiece respirator straps. J Intern Soc Resp Protect. 2013;30(2):31-40.

[26.] Tikanto J, Pirila T. Effects of the Cottle's maneuver on the nasal valve as assessed by acoustic rhinometry. Am J Rhinol. 2007;21(4):456-9, http://dx.doi.org/10.2500/ ajr.2007.21.3040.

[27.] Koehler JL, Leith D. Model calibration for pressure drop in a pulse-jet cleaned fabric filter. Atmos Environ. 1983;17(10):1909-13, http://dx.doi.org/10.1016/00046981(83)90348-7.

[28.] Newnum JD. The effects of relative humidity on respirator performance. University of Iowa, Iowa Research Online, Theses and Dissertations; 2010 [cited 2014 Mar 12]. Available: http://ir.uiowa.edu/cgi/viewcontent. cgi?article=2046&context=etd.

[29.] Arad M, Heruti R, Shaham E, Atsmon J, Epstein Y. The effects of powered air supply to the respiratory protective device on respiration parameters during rest and exercise. Chest. 1992;102(6):1800-4, http://dx.doi.org/10.1378/ chest.102.6.1800.

[30.] Gozal D, Omidvar O, Kirlew KAT, Hathout GM, Hamilton R, Lufkin RB, et al. Identification of human brain regions underlying responses to resistive inspiratory loading with functional magnetic resonance imaging. Proc Natl Acad Sci. 1995;92(14):6607-11, http://dx.doi.org/10.1073/ pnas.92.14.6607.

[31.] Bansal S, Harber P, Yun D, Liu D, Liu Y, Wu S, et al. J Occup Environ Hyg. 2009;6:221-7, http://dx.doi.org/ 10.1080/15459620902729218.

[32.] Harber P, Shimozaki S, Barrett T, Fine G. Effect of exercise level on ventilator adaptation to respirator use. J Occup Med. 1990;32:1042-6.

[33.] Hlavac MC, Catcheside PG, Adams A, Eckert DJ, McEvoy RD. The effects of hypoxia on load compensation during sustained incremental resistive loading in patients with obstructive sleep apnea. J Appl Physiol. 2007;103(1):234-9, http://dx.doi.org/10.1152/japplphysiol. 01618.2005.

[34.] Roberge RJ. Are exhalation valves on N95 filtering face-piece respirators beneficial at low-moderate work rates: An overview. J Occup Environ Hyg. 2012;9(11):617-23, http://dx.doi.org/10.1080/15459624.2012.715066.

[35.] Chen CC, Willeke K. Characteristics of face seal leakage in filtering facepieces. Am Ind Hyg Assoc. 1992;53(9):533-9, http://dx.doi.org/10.1080/15298669291360120.

[36.] Roberge RJ, Monaghan WD, Palmiero AJ, Shaffer R, Bergman MS. Infrared imaging for leak detection of N95 filtering facepiece respirators: A pilot study. Am J Ind Med. 2011;54(8):628-36, http://dx.doi.org/10.1002/ ajim.20970.

[37.] McNichols WT. The nose and OSA: Variable nasal obstruction may be more important in pathophysiology than fixed obstruction. Eur Resp J. 2008;32(1):3-8, http://dx. doi.org/10.1183/09031936.00050208.

[38.] Barry PW, Mason NP, Richalet J-P. Nasal peak inspiratory flow at altitude. Eur Resp J. 2002;19(1):16-9, http://dx.doi. org/10.1183/09031936.02.00096902.

[39.] Roberge RJ, Kim J-H, Coca A. Protective facemask impact on human thermoregulation: An overview. Ann Occup Hyg. 2010;56(1):102-12, http://dx.doi.org/10.1093/annhyg/ mer069.

[40.] Harber P, Bansal S, Santiago S, Liu D, Yun D, Ng D, et al. Multidomain subjective response to respirator use during simulated work. J Occup Environ Med. 2009;51(1):38-45, http://dx.doi.org/10.1097/JOM.0b013e31817f458b.

[41.] Meyer JP, Hery M, Herrault J, Hubert G, Francois D, Hecht G, et al. Field study of subjective assessment of negative pressure half-masks. Influence of the work conditions on comfort and efficiency. Appl Ergon. 1997;28(5-6):331-8, http://dx.doi.org/10.1016/S0003-6870(97)00007-0.

JUNG-HYUN KIM (1), RAYMOND J. ROBERGE (1), JEFFREY B. POWELL (1), RONALD E. SHAFFER (1), CAROLINE M. YLITALO (2), and JOHN M. SEBASTIAN (2)

(1) Centers for Disease Control and Prevention, Pittsburgh, Pennsylvania, United States of America National Institute for Occupational Safety and Health, National Personal Protective Technology Laboratory, Technology Research Branch

(2) 3M Company, St. Paul, Minnesota, United States of America 3M Personal Safety Division

This work was supported by internal operating funds of the National Institute for Occupational Safety and Health (NIOSH), part of the U.S. Centers for Disease Control and Prevention. The findings and conclusions in this report are those of the author(s) and do not necessarily represent the views of the National Institute for Occupational Safety and Health. Mention of any product name does not imply endorsement.

Received: December 4, 2013. Accepted: July 1, 2014.

Corresponding author: R.J. Roberge, National Personal Protective Technology Laboratory / NIOSH / CDC, Technology Research Branch, 626 Cochrans Mill Road, Pittsburgh, PA, U.S. 15236 (e-mail: dtn0@cdc.gov).

Caption: Fig. 1. Mean peak oral and nasal inhalation and exhalation airflow filter resistance of prototype respirators (PR) at sedentary and low-moderate work rates: a) PR3, b) PR6, c) PR9

Caption: Fig. 2. Differences in mean peak oral and nasal respirator inhalation and exhalation filter airflow resistance at sedentary and low-moderate work rates for prototype respirators (PR): a) PR3, b) PR6, c) PR9
Table 1. Mean peak nasal and oral inhalation and exhalation airflow
resistance (Rfilter) of prototype respirators (PR) at sedentary and
low-moderate work rates

                               Airflow resistance
Trial                          (Pa/lx[s.sup.-1])

                             sedentary standing

                        M [+ or -] SD        95% CI

PR3 [R.sub.filter]
  nasal inhalation   -11.6 [+ or -] 5.5    -7.6-15.5
  nasal exhalation    +4.9 [+ or -] 6.4     +2.7-9.5
  oral inhalation    -16.8 [+ or -] 6.5    -12.1-21.4
  oral exhalation     +9.2 [+ or -] 6.3    +4.7-13.8

PR6 [R.sub.filter]
  nasal inhalation   -33.4 [+ or -] 11.1   -25.3-41.3
  nasal exhalation   + 19.9 [+ or -] 7.5   +14.6-25.3
  oral inhalation    -36.0 [+ or -] 11.6   -27.6-44.4
  oral exhalation    +24.3 [+ or -] 9.3    +17.5-30.9

PR9 [R.sub.filter]
  nasal inhalation   -31.8 [+ or -] 11.8   -23.3-40.3
  nasal exhalation   + 16.9 [+ or -] 7.4   +11.5-22.3
  oral inhalation    -38.0 [+ or -] 15.7   -26.8-49.3
  oral exhalation    +26.5 [+ or -] 11.3   +18.4-34.7

                               Airflow resistance
Trial                           (Pa/lx[s.sup.-1])

                      reclining bicycle     exercise

                        M [+ or -] SD        95% CI

PR3 [R.sub.filter]
  nasal inhalation   -22.6 [+ or -] 8.1    -16.8-28.5
  nasal exhalation   +14.8 [+ or -] 19.8   +6.1-28.9
  oral inhalation    -28.8 [+ or -] 9.7    -21.8-35.7
  oral exhalation    +19.6 [+ or -] 10.2   +12.1-26.9

PR6 [R.sub.filter]
  nasal inhalation   -49.1 [+ or -] 13.4   -10.0-58.7
  nasal exhalation   +33.0 [+ or -] 14.3   +22.8-43.3
  oral inhalation    -59.2 [+ or -] 8.9    -52.9-65.6
  oral exhalation    +43.5 [+ or -] 11.4   +35.4-51.7

PR9 [R.sub.filter]
  nasal inhalation   -49.4 [+ or -] 14.7   -38.9-60.0
  nasal exhalation   +33.1 [+ or -] 13.2   +23.6-42.6
  oral inhalation    -62.1 [+ or -] 16.4   -50.4-74.0
  oral exhalation    +48.9 [+ or -] 17.7   +36.1-61.5

M--mean; SD--standard deviation; CI--confidence interval.
COPYRIGHT 2015 Nofer Institute of Occupational Medicine
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2015 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:ORIGINAL PAPER
Author:Kim, Jung-hyun; Roberge, Raymond J.; Powell, Jeffrey B.; Shaffer, Ronald E.; Ylitalo, Caroline M.; S
Publication:International Journal of Occupational Medicine and Environmental Health
Geographic Code:1U2PA
Date:Jan 1, 2015
Words:5233
Previous Article:IMPAIRMENT OF PULMONARY FUNCTION AND CHANGES IN THE RIGHT CARDIAC STRUCTURE OF PNEUMOCONIOTIC COAL WORKERS IN CHINA.
Next Article:EXHALED BREATH MALONDIALDEHYDE, SPIROMETRIC RESULTS AND DUST EXPOSURE ASSESSMENT IN CERAMICS PRODUCTION WORKERS.
Topics:

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters