Noninvasive, automatic optimization strategy in cardiac resynchronization therapy.

ABSTRACT

Objective: Optimization of cardiac resynchronization therapy (CRT) is still unsolved. It has been shown that optimal electrode position, atrioventricular (AV) and interventricular (VV) delays improve the success of CRT and reduce the number of non-responders. However, no automatic, noninvasive optimization strategy exists to date.

Methods: Cardiac resynchronization therapy was simulated on the Visible Man and a patient data-set including fiber orientation and ventricular heterogeneity. A cellular automaton was used for fast computation of ventricular excitation. An AV block and a left bundle branch block were simulated with 100%, 80% and 60% interventricular conduction velocity. A right apical and 12 left ventricular lead positions were set. Sequential optimization and optimization with the downhill simplex algorithm (DSA) were carried out. The minimal error between isochrones of the physiologic excitation and the therapy was computed automatically and leads to an optimal lead position and timing.

Results: Up to 1512 simulations were carried out per pathology per patient. One simulation took 4 minutes on an Apple Macintosh 2 GHz PowerPC G5. For each electrode pair an optimal pacemaker delay was found. The DSA reduced the number of simulations by an order of magnitude and the AV-delay and VV-delay were determined with a much higher resolution. The findings are well comparable with clinical studies.

Conclusion: The presented computer model of CRT automatically evaluates an optimal lead position and AV-delay and VV-delay, which can be used to noninvasively plan an optimal therapy for an individual patient. The application of the DSA reduces the simulation time so that the strategy is suitable for pre-operative planning in clinical routine. Future work will focus on clinical evaluation of the computer models and integration of patient data for individualized therapy planning and optimization.

Key words: cardiac resynchronization therapy, automatic optimization, computer models

Introduction

An individualized optimization strategy for biventricular pacing (BiVP) as cardiac resynchronization therapy is not yet feasible in daily clinical routine (1, 2). This work presents an optimization strategy with which optimal electrode position, atrio-ventricular (AV) and interventricular (VV) delays can be computed fast and automatically to make the application suitable for clinical practice.

Methods

This work builds on optimizing BiVP by using an adaptive cellular automation to compute the activation times of a cardiac cell in a computer model of the heart based on a detailed anatomical model including fiber orientation and heterogeneity. The anatomical models were given by a heart failure patient and the Visible Man data-set (National Library of Medicine, Bethesda, USA). The pathologies simulated were an AV block and a left bundle branch block (LBBB). For each, the left ventricular conduction velocity was reduced by 0%, 20% and 40%. The root mean square error [E.sub.RMS] of the activation times during sinus rhythm versus pathology/therapy was used as optimization parameter (3). Overall 12 different left ventricular electrode positions (Fig. 2) were investigated, which cover more electrode positions tested in clinical trials yet (4). Four were placed in the anterior coronary sinus branches, four were put in the posterior coronary sinus branches and four were placed on the left ventricular free wall to evaluate the electrode position independently of the coronary sinus branches. The right ventricular lead was placed in the apex. The AV and VV delays were determined by the Downhill Simplex Algorithm (DSA) (5). The DSA is moving the triangle to the minimum of the parameter space (Fig. 1). A termination criterion c-termination is defined to stop the iteration:

[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII]

with c-termination=0.0005, min(ERMS) being the lowest value of the three simplex points and max(ERMS) being the highest value after an iteration step.

[FIGURES 1-2 OMITTED]

The results were compared with sequentially searching the 2 dimensional parameter space given by AV and VV delays. The timing was set according to literature (6) by sampling the AV delay between 60--220 ms/260 ms (Visible Man/patient) with increments of 20 ms. For each AV delay, the VV delay was set to -30 -50 ms/70 ms (Visible Man/patient) with 10 ms increments. A negative VV delay in this work means that the right ventricular electrode is pacing before the left ventricular electrode.

Results

Up to 1512 simulations were carried out per pathology per patient. One simulation took 4 minutes on an Apple Macintosh 2 GHz PowerPC G5. Tables 1 and 2 summarize the results and Figure 2 gives an example of the optimal electrode positions in the patient data-set. The optimal electrode positions are in accordance with clinical practice. The optimal timing delays are within the range described in clinical studies. The DSA reduced the number of simulations to 15-25% of the simulations needed for the sequential search.

The DSA achieves a lower ERMS value for all pathologies and both anatomical data-sets except the AV block pathology in the Visible Man data-set where it performs slightly less good. But it achieves a better temporal resolution with respect to AV and VV delay setting.

Discussion

A minimal error was found for each electrode set-up. This means that the optimization could be carried out preoperatively to determine optimal pacing lead position with the respective AV and VV delays--and it can also be run postoperatively to find the optimal timing delays for a given electrode position. The method of optimization is independent on anatomical shape as well as pathophysiology so far. However, general trends could be observed: the lower the interventricular conduction delay, the lower the optimal AV delay. With respect to VV delay setting, a general rule cannot be determined. It has to be set independently with respect to electrode position and pathology.

Comparing with clinical studies the advantage of the presented strategy is that a multitude of electrode positions as well as AV/VV delay combinations can be investigated automatically and noninvasively, which has not been done in clinical studies yet.

Conclusion

The presented method reduces the number of simulations required drastically. While previous optimization simulations took around 5 days for one pathology and 12 pacing lead set-ups, the DSA reduces the simulation time to even less than 15 hours. Given a patient is admitted to the clinic the night before pacemaker implantation, the optimal AV and VV delays as well as electrode position could be computed automatically with a noninvasive strategy given the presented model is clinically validated. A current project of the Institute of Biomedical Engineering, Karlsruhe University (TH) and the Cardiology, Faculty Mannheim, and the Cardiac Surgery of the University of Heidelberg targets this issue based on this work. In future, a contraction and deformation model (7-9) will be included to compute the cardiac output.

References

(1.) Reumann M. Computer assisted optimisation of non-pharmacological treatment of congestive heart failure and supraventricular arrhythmia. Karlsruhe; Karlsruhe Transactions on Biomedical Engineering: 2007.

(2.) Kerckhoffs RCP, Neal ML, Gu Q, Bassingthwaighte JB, Omens JH, McCulloch AD. Coupling of a 3D finite element model of cardiac ventricular mechanics to lumped systems models of the systemic and pulmonic circulation. Ann Biomed Eng 2007; 35: 1-18.

(3.) Albrecht K, Reumann M, Seemann G,Reinerth G, Vahl CF, Dossel O. Computer-aided evaluation and optimisation of biventricular pacing for patients with congestive heart failure. Biomedizinische Technik 2005; 50; 701-2.

(4.) Van Campen CMC, Visser FC, de Cock CC, Vos HS, Kamp O, Visser CA. Comparison of the hemodynamics of different pacing sites in patients undergoing resynchronization therapy: need for individualization an optimal lead localization. Heart 2006; 92: 1795-800.

(5.) Nelder JA, Mead R. A simplex method for function minimization. The Computer Journal 1965; 7: 308-13.

(6.) Whinnett ZI, Davies JE, Willson K, Manisty CH, Chow AW, Foale RA, et al. Haemodynamic effects of changes in atrioventricular and interventricular delay in cardiac resynchronisation therapy show a consistent pattern: analysis of shape, magnitude and relative importance of atrioventricular and interventricular delay. Heart 2006; 92: 1628-34.

(7.) Kerckhoffs RC, Neal ML, Gu Q, Bassingthwaighte JB, Omens JH, McCulloch AD. Coupling of a 3D finite element model of cardiac ventricular mechanics to lumped systems models of the systemic and pulmonic circulation. Ann Biomed Eng 2007; 35: 1-18.

(8.) Sermesant M, Delingette H, Ayache N. An electromechanical model of the heart for image analysis and simulation. IEEE Trans Med Imag 2006; 25: 612-25.

(9.) Dossel O, Farina D, Mohr M, Reumann M, Seemann G, Weiss DL. Computer-assisted planning of cardiac interventions and heart surgery. In: Informatik 2006--Informatik fur Menschen. Gesellschaft fur Informatik e.V. (GI). Bonn: Kollen Druck-Verlag GmbH; 2006. p. 499-506.

Matthias Reumann, Brigitte Osswald *, Olaf Doessel **

Computational Biology Center, IBM TJ Watson Research Center, New York, USA

* Department of Cardiac Surgery, University of Heidelberg, Heidelberg, Germany

** Institute of Biomedical Engineering, Karlsruhe University (TH), Karlsruhe, Germany

Address for Correspondence: Matthias Reumann, PhD, Computational Biology Center, IBM TJ Watson Research Center, 1101 Kitchawan Road, Route 136 Yorktown Heights, NY 10598 USA Phone: +49-177-6727731 Fax: +1-914 945-4217 E-mail: mreumann@ieee.org
Table 1. Optimal electrode position, AV and VV delays for each
pathology using both sequential search and Downhill Simplex
Algorithm for the Visible Man model

Pathology Pacing AV delay, VV delay, ERMS,
 lead ms ms ms

AV block -0% RA 100 20 3.99
AV block -20% RL 100 10 4.59
AV block -40% RA 60 40 5.91
LBBB -0% RB 120 -10 5.28
LBBB -20% RB 100 0 6.84
LBBB -40% RK 60 30 9.50
AV block -0% RI 234 117 3.18
AV block -20% RI 217 117 4.23
AV block -40% RG 343 -134 5.19
LBBB -0% RH 220 138 5.23
LBBB -20% RE 206 13 6.77
LBBB -40% RK 188 34 9.49

Pathology Number of Optimization
 simulations algorithm

AV block -0% 972 Sequential
AV block -20% 972 Sequential
AV block -40% 972 Sequential
LBBB -0% 972 Sequential
LBBB -20% 972 Sequential
LBBB -40% 972 Sequential
AV block -0% 301 DSA
AV block -20% 280 DSA
AV block -40% 256 DSA
LBBB -0% 226 DSA
LBBB -20% 225 DSA
LBBB -40% 219 DSA

AV block-atrioventricular block, DSA-Downhill Simplex Algorithm,
ERMS-root mean square error of the activation times, LBBB-left
bundle branch block, VV-interventricular. The percentage indicates
the reduction of interventricular conduction velocity

Table 2. Optimal electrode position, AV and VV delays for each
pathology using both sequential search and Downhill Simplex
Algorithm for the patient model

Pathology Pacing AV delay, VV delay, ERMS,
 lead ms ms ms

AV block -0% RJ 260 0 3.21
AV block -20% RL 260 -20 3.28
AV block -40% RL 240 -30 4.63
LBBB -0% RK 160 70 10.12
LBBB -20% RK 160 40 13.50
LBBB -40% RK 120 20 19.66
AV block -0% RL 249 111 3.90
AV block -20% RL 234 127 4.27
AV block -40% RL 211 109 5.38
LBBB -0% RK 228 11 9.89
LBBB -20% RK 188 34 12.59
LBBB -40% RK 109 84 18.92

Pathology Number of Optimization
 simulations algorithm

AV block -0% 1452 Sequential
AV block -20% 1452 Sequential
AV block -40% 1452 Sequential
LBBB -0% 1452 Sequential
LBBB -20% 1452 Sequential
LBBB -40% 1452 Sequential
AV block -0% 282 DSA
AV block -20% 330 DSA
AV block -40% 387 DSA
LBBB -0% 265 DSA
LBBB -20% 284 DSA
LBBB -40% 212 DSA

AV block-atrioventricular block, DSA-Downhill Simplex Algorithm,
ERMS--root mean square error of the activation times, LBBB-left
bundle branch block, VV-interventricular. The percentage indicates
the reduction of interventricular conduction velocity
COPYRIGHT 2007 Galenos Yayincilik
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2007 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Original Investigation
Author:Reumann, Matthias; Osswald, Brigitte; Doessel, Olaf
Publication:The Anatolian Journal of Cardiology (Anadolu Kardiyoloji Dergisi)
Article Type:Clinical report
Geographic Code:4EUGE
Date:Jul 1, 2007
Words:1893
Previous Article:Vectorcardiographic assessment of acute hypoxia effects in pulmonary hypertension due to chronic bronchitis.
Next Article:Twenty-millisecond interventricular difference as assessed by body surface potential mapping identifies patients with clinical improvement after...
Topics:


Related Articles
MEDTRONIC SUPPORTS NEW EUROPEAN HF TREATMENT GUIDELINES.
Cardiac Resynchronization Therapy.
Twenty-millisecond interventricular difference as assessed by body surface potential mapping identifies patients with clinical improvement after...
Optimal programming in cardiac resynchronization therapy.
Cardiac resynchronization therapy: implantation tips and tricks.
Survival in cardiac resynchronization therapy. What do we know?
Repercussions of cardiac resynchronization therapy on the ventricular repolarization of heart failure patients as assessed by body surface potential...
The nuts and bolts of cardiac resynchronization therapy.
Cardiac pacing and defibrillation; principle and practice.

Terms of use | Privacy policy | Copyright © 2021 Farlex, Inc. | Feedback | For webmasters