Multiple Gamma-Ray Detection Capability of a Ce[Br.sub.3] Detector for Gamma Spectroscopy.

1. Introduction

Gamma-ray spectroscopy techniques are continuously progressing due to rapid advances in the development and fabrication of fast and higher light output scintillation materials [1-3]. The development of the cerium tribromide (Ce[Br.sub.3]) gamma-ray detector among the lanthanum halide detector series (La[Br.sub.3]:Ce and La[Cl.sub.3]:Ce detectors) is one such example [4-11]. The Ce[Br.sub.3] detector, which has an energy resolution comparable with that of a La[Cl.sub.3]:Ce detector, has an order of magnitude reduced intrinsic activity as compared to the La[Br.sub.3]:Ce and La[Cl.sub.3]:Ce detector [9]. Furthermore, the reduced intrinsic activity of the Ce[Br.sub.3] detector has a narrower energy range of 1400-2200keV [9, 12], as compared to 780-2200 keV energy range reported for the intrinsic activity of the La[Br.sub.3]:Ce and La[Cl.sub.3]:Ce ([1, 3, 7]). These features enable the Ce[Br.sub.3] detector to provide a larger interference-free energy range for the gamma-ray detection in prompt gamma-ray analysis of bulk samples as compared to the La[Br.sub.3]:Ce and La[Cl.sub.3]:Ce detectors.

In the present study, the multiple gamma-ray detection capabilities of a cylindrical 76 mm x 76 mm (diameter x height) Ce[Br.sub.3] detector were studied through chlorine concentration measurements in saline water samples utilizing eight chlorine prompt gamma rays [13] in conjunction with a portable neutron generator-based PGNAA setup [8, 12]. The saline water samples were chosen due to their significance as secondary calibration standards. The prompt gammas emitted from chlorine are considered as secondary standards for detector calibrations because of the high-thermal neutron capture cross section of chlorine and also because of the wide range of high-intensity gamma emissions up to 8.5 MeV [14].

Furthermore, the chlorine analysis in saline water samples is also very important for the assessment of groundwater and surface water pollution program of the Environmental Studies [15,16]. Chloride in water has adverse effects on the environment and industry. The U.S. Environmental Protection Agency (USEPA) has imposed an upper limit of 250 mg/L for maximum contaminant level (SMCL) of chloride in drinking water [16].

In the present study, the multiple gamma-ray detection capabilities of a cylindrical 76 mm x 76 mm (height x diameter) Ce:[Br.sub.3] detector has been tested through analysis of saline water samples containing 0.0-8.0wt% chlorine concentrations. The experimental data was acquired for eight chlorine prompt gamma rays and minimum detectable concentration (MDC) for each gamma-ray line was estimated. For comparison, the intensities of chlorine prompt gamma rays from the saline water samples were also calculated using the MCNP4B2 code [17].

For comparison with the Ce[Br.sub.3] detector, the prompt gamma-ray spectrum from a cylindrical 76 mm x 76 mm (diameter x height) La[Cl.sub.3]:Ce detector was acquired with the saline water samples. Details of this study are reported in the following sections of the paper.

2. Prompt Gamma Spectrum of La[Cl.sub.3]:Ce Detector with Water Sample

The prompt gamma-ray spectrum from a La[Cl.sub.3]:Ce detector was measured using the portable neutron generator-based PGNAA set up described earlier [8] and shown in Figure 1. It will be described briefly. The setup contains a high-density polyethylene cylindrical moderator with a concentric central cylindrical cavity for a cylindrical sample container with 106 mm x 1.5 mm (diameter x length). Moreover, a cylindrical 76 mm x 76 mm (diameter x length) La[Cl.sub.3]:Ce or Ce[Br.sub.3] gamma-ray detector views the sample through the cavity hole at a right angle to the neutron generator axis. The empty polyethylene plastic sample container has a mass of 96 g and a density of 0.92 g/[cm.sup.3]. The paraffin and lithium carbonate mixed neutron shield is used to protect the detector against neutrons while lead shielding provides detector protection against gamma rays. For the tests of the La[Cl.sub.3]:Ce detector, the plastic sample bottle was filled with pure water.

MeV neutrons were produced using the D(d,n) reaction using a 70 [micro]A pulsed deuteron beam of 70 keV energy, a width of 800 microseconds and a frequency of 250 Hz. The prompt gamma-ray spectrum of the La[Cl.sub.3]:Ce detector from the pure water samples was counted for 25 min. The neutron flux was monitored using a beam current monitor from the neutron generator. Any fluctuation in the beam current was taken into account while normalizing the data.

Figures 2 and 3 show the gamma-ray spectra of the La[Cl.sub.3]:Ce detector for the water sample. Figure 2 shows the gamma-ray background spectrum over 0.69-3.51 MeV energy range. The spectrum shows background gamma-ray peaks from detector material as well as shielding material used in the PGNAA setup. Noticeable capture peaks from detector material are La(920), La(926), and La(2521) peaks along with La(1468) intrinsic peak. Also shown are capture peaks Cl(1165), Cl(1602), andCl(1951,1959) due to chlorine in the detector, along with Pb(810), Pb(1060) and Pb(2620) peaks from shielding material. Also, Si(1274) and Al(1766) prompt gamma rays are shown along with hydrogen H(2223) capture peak.

Figure 3 shows the background gamma-ray spectrum from the water sample over 3.53-9.23 MeV energy range taken with the La[Cl.sub.3]:Ce detector. Noticeable capture peaks from chlorine in the detector material are Cl(5718), Cl(6111), Cl(6620, 6628), Cl(6978), Cl(7414), Cl(7790), and C(8579).

This study has shown that multiple energy chlorine background gamma-ray peaks from La[Cl.sub.3]:Ce makes its use in the detection of chlorine quite difficult and inaccurate. A Ce[Br.sub.3] detector can be a good substitute for the La[Cl.sub.3]:Ce detector in the chlorine measurements and this study was undertaken to verify this fact.

3. Prompt Gamma-Ray Intensity Calculations from Saline Water Samples

Prior to the gamma-ray measurements with a Ce[Br.sub.3] detector, the intensities of chlorine prompt gamma rays from water samples were calculated using the general purpose MCNP4B2 code [17]. The intensities of eight chlorine gamma rays, namely, 517, 788, 1165, 1951, 2863, 6110, 6619, and 8578 keV were calculated for saline water samples containing 0.0-8.0 wt% chlorine. The calculations were carried out for the 2.5 MeV neutron-based PGNAA setup using the procedure described earlier [8]. For calculation, the sample was divided into subcells of 1 cm thickness, which allowed the transport of the neutrons and gamma rays of appropriate statistical weight to the next adjacent cell, without any loss. The F4 tally was used to calculate gamma-ray intensity in the detector volume. Saline water samples were prepared by thoroughly mixing water with 0.0-8.0 wt% chlorine. Gamma-ray yields were calculated for 517, 788, 1165, 1951, 2863, 6110, 6619, and 8578 keV chlorine prompt gamma rays produced due to capture of thermal neutrons in chlorine. Thermal neutrons were produced through moderation of 2.5 MeV neutrons in the moderator as well as in the water samples.

The calculated intensities of 517, 788, 1165, 1951, 2863, 6110, 6619, 8578 keV gamma rays increase with chlorine concentration but each line has a different slope. Among the low energy 517, 788,1165, and 1951 keV chlorine gamma rays, the 788 keV line has maximum slope while for the high energy 2863, 6110, 6619, and 8578 keV chlorine gamma rays, the maximum slope has been observed for the 2863 keV gamma rays. The calculated yield curves of 517, 788,1165,1951,2863, 6110, 6619, and 8578 keV prompt gamma rays are plotted along with the experimental data in Figures 4 and 5.

4. Prompt Gamma Measurements Using the [Cebr.sub.3] Detector

Prompt gamma-ray measurements from saline water samples were carried out using the Ce[Br.sub.3] detector in conjunction with the portable neutron generator and following the procedure given in Section 2.

Prompt gamma rays from chlorine have several intense lines with energies varying from 0.5 MeV to 8.6 MeV. In order to obtain optimum resolution of these gamma rays, the measurements were carried out with two different pulse height amplifier gain settings. One amplifier setting, with larger gain, was used to acquire gamma-ray pulse height spectrum over 0.00-2.22 MeV while the other amplifier setting, with smaller gain, was used to acquire gamma-ray pulse height spectrum over 1.45-9.98 MeV.

The larger gain setting of the amplifier was used to record low energy gamma rays with energies below 2.22 MeV by adjusting the hydrogen peak location near the end channel of the spectrum, as shown in Figure 6 which shows the low energy activation spectrum of the Ce[Br.sub.3] detector due to the capture of thermal neutrons in the Ce[Br.sub.3] detector material. The prompt gamma peaks due to the capture of thermal neutrons in Ce, Br, and H (from moderator material) are quite prominent. Figure 7 shows the low energy prompt gamma ray spectrum from saline water samples containing 4.0 wt% chlorine superimposed upon the background spectrum over 0.41-2.29 MeV energy range. Chlorine peaks at 517, 788, 1165, and 1951-1959 keV along with the 2223 keV hydrogen peak from the moderator are quite prominent.

The smaller gain setting of the amplifier was used to record high energy gamma rays with 1.45-9.98 MeV energies by adjusting the location of the hydrogen peak near the beginning channels of the spectrum, as shown in Figure 8 which shows the high energy prompt gamma-ray spectrum from saline water samples containing 4 wt% chlorine along with the background spectrum over 1.75-9.98 MeV energy range. Chlorine peaks at 6111 and 8578 keV along with the 2223 keV hydrogen peak are quite prominent. Figure 9 shows the enlarged part of Figure 8 over 2.57-8.75 MeV range showing 2864, 3062, 4980, 5715, 6111, 6620, and 8578 keV chlorine peaks.

In order to calibrate the detector for the higher energy gamma rays, the detector spectrum was acquired from a NiN[O.sub.3] sample and the spectrum was calibrated using high energy Ni (8533) and Ni(8998) keV gamma rays along with the Br(7577) gamma-ray peak from the detector material.

For each amplifier setting, the detector background spectrum was recorded with an empty plastic container inserted into the moderator cavity. The background spectrum of the Ce[Br.sub.3] detector contains capture gamma-ray peaks of Br and Ce elements present in the Ce[Br.sub.3] detector material. The energies and intensities of prominent capture prompt gamma ray in cerium and bromine are listed in Table 1 [13]. The background spectrum is subtracted from the sample spectrum to obtain the difference gamma-ray spectrum of the sample. The prompt gamma-ray data from the chloride-contaminated water samples were acquired for a preset time of 20-30 min. Low and high energy gamma-ray spectra from saline water samples containing 2, 4, 6, and 8 wt% chlorine were analyzed using pulse height windows corresponding to 517, 788, 1165, 1951, 2863, 6110, 6619, and 8578 keV chlorine prompt gamma rays.

5. Results and Discussion

For each gamma ray, the energy difference spectrum was generated for 2%, 4%, 6%, and 8% chlorine by subtracting the background spectrum from the sample spectrum and normalizing to the same neutron flux. Finally, the area under each difference peak was integrated to generate the gamma yield as a function of chlorine concentration. Then, the gamma-ray yield versus concentration curve was plotted. This process was repeated for each of the eight chlorine gamma rays with 517, 786-788, 1165, 1951-1959, 2863, 6110, 6620, and 8578 keV and eight difference spectra were analyzed. In the following, only three representative gamma-ray difference spectra are shown for which we obtained the best value of MDC.

Figures 10, 11, and 12 show gamma-ray difference spectra of 517, 786-788 keV, and 1951-1959 keV gamma rays. For these three gamma rays, we have found the best values of MDC for chlorine detection in water samples. Figure 10 shows the 517 keV peak over 0.42-0.66 MeV energy range while Figure 11 shows 786-788 keV chlorine peak over 0.66-0.91 MeV energy range. Figure 12 shows the 1951-1959 keV chlorine peak over 1.78-2.13 MeV energy range.

The integrated gamma ray yields of chlorine for the eight gamma rays as a function of chlorine concentration in the saline water samples are shown in Figures 4 and 5 where the experimental data points for each gamma ray are superimposed on the Monte Carlo calculated yield of the respective gamma ray. The experimental results are in agreement with the calculated values.

Figure 4 shows the integrated yield of 517, 788, 1165, and 1951 keV gamma rays as a function of chlorine concentration. Among this group of chlorine gamma-ray lines, the maximum slope has been observed for 788 keV gamma rays while the minimum slope was for 1951 keV gamma rays. Figure 5 shows the integrated yield of 2863, 6110, 6619, and 8578 keV chlorine gamma for various chlorine concentrations. Among this group of chlorine gamma-ray lines, the maximum slope has been observed for 2863 keV gamma rays while the minimum slope has been observed for 6619 keV gamma rays.

5.1. Minimum Detection Limits for 0.5-8.58MeV Chlorine Gamma Rays. The minimum detectable concentrations (MDCs) of the Ce[Br.sub.3] detector-based PGNAA setup were determined for 517, 786-788, 1165, 1951-1959, 2863, 6111, 6619, and 8578 keV Cl gamma rays using the method published earlier [8]. Due to the limited neutron flux from the 2.5 MeV portable neutron generator, all MDCs are in the wt% range. For our cylindrical water samples, the MDCs for the eight chlorine prompt gamma rays vary from 0.07 wt% to 0.80 wt%. The values (MDCs) for each gamma ray along with its standard deviation [[sigma].sub.MDC] are listed in Table 2.

Most of the MDC values vary over 0.07 [+ or -] 0.02 wt% to 0.18 [+ or -] 0.05 wt% with an exceptionally large value of MDC of 0.80 [+ or -] 0.24 for 8578 keV gamma. The larger MDCs have been obtained for the less intense peaks and/or peaks having a higher background underneath due to gamma rays produced from the Ce[Br.sub.3] detector material. The smaller MDCs have been obtained for those gamma rays with higher intensities and relatively smaller background under the gamma-ray peak.

The optimum value of MDCs [+ or -] [[sigma].sub.MDC] (wt%) was measured to be 0.07 [+ or -] 0.02 wt% and 0.08 [+ or -] 0.02 wt% for 517 keV and 786-788 keV chlorine gamma rays, respectively.

Although the 1165 keV gamma rays have a large production cross section, yet a larger value of 0.18 [+ or -] 0.05 wt% of MDC has been measured for this gamma ray. This is due to the 1165 keV chlorine gamma-ray interference with of Ce(1107) gamma rays from the Ce[Br.sub.3] detector material.

The 1951-1959 keV and 6111 keV gamma rays have large production cross sections but, within statistical uncertainty, they have equal MDCs, that is, 0.11 [+ or -] 0.03 wt% and 0.13 [+ or -] 0.04 wt%, respectively. For the remaining gamma rays, the measured MDCs are larger because they have smaller production cross sections. This study has shown the excellent multiple gamma-ray detection capabilities of the Ce[Br.sub.3] detector for gamma-ray spectroscopy.

6. Conclusion

In this study, suitability of La[Cl.sub.3]:Ce and Ce[Br.sub.3] detectors for chlorine measurements was studied by comparing neutron beam-associated gamma-ray background of La[Cl.sub.3]:Ce and Ce[Br.sub.3] detectors. The comparison revealed very low gamma-ray background for the Ce[Br.sub.3] detector while for the La[Cl.sub.3]:Ce detector background showed additional chlorine thermal neutron for Cl(1165), Cl(1602), Cl(1951, 1959), Cl(5718), Cl(6111), Cl(6620, 6628), Cl(6978), Cl(7414), Cl(7790), and C(8579) gamma rays. This made the La[Cl.sub.3]:Ce detector not suitable for chlorine measurements while Ce[Br.sub.3] could be a good substitute of La[Cl.sub.3]:Ce detector for chlorine measurements in bulk sample.

Then multigamma-ray detection capability of a cylindrical 75 mm x 75 mm (diameter x height) Ce[Br.sub.3] detector has been tested by analyzing the chlorine concentration in water samples using eight chlorine prompt gamma rays 517, 786-788, 1165, 1951-1959, 2863, 6111, 6619, and 8578 keV chlorine over 517 to 8578 keV energies utilizing a D-D portable neutron generator-based PGNAA setup and measuring the corresponding minimum detection limit (MDC) of chlorine.

The MDC values for these gamma rays have been determined to be 0.08 [+ or -] 0.02, 0.07 [+ or -] 0.02, 0.18 [+ or -] 0.05, 0.11 [+ or -] 0.03, 0.17 [+ or -] 0.05, 0.13 [+ or -] 0.04, 0.18 [+ or -] 0.05, 0.80 [+ or -] 0.24 wt%, respectively. The experimental results agree with Monte Carlo results. The best MDC values of 0.07 [+ or -] 0.02 and 0.08 [+ or -] 0.02 wt% have obtained for 786-788 and 517keV gamma rays, respectively. The Ce[Br.sub.3] detector has proven to be an excellent multiple gamma-ray detectors.

https://doi.org/10.1155/2017/5634384

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Acknowledgments

This study was carried out under NSTIP Project no. ENV2367-04 funded by KACST, Saudi Arabia. The authors gratefully acknowledge the support provided by the Physics Department, King Fahd University of Petroleum and Minerals, Dhahran, Saudi Arabia.

References

[1] D. Alexiev, L. Mo, D. A. Prokopovich, M. L. Smith, and M. Matuchova, "Comparison of La[Br.sub.3]: Ce and La[Cl.sub.3]: Ce with NaI(Tl) and cadmium zinc telluride (CZT) detectors," IEEE Transactions on Nuclear Science, vol. 55, 2008, no. 3, pp. 1174-1177, 2008.

[2] N. J. Cherepy, S. A. Payne, S. J. Asztalos et al., "Scintillators with potential to supersede lanthanum bromide," IEEE Transactions on Nuclear Science, vol. 56, no. 3, pp. 873-880, 2010.

[3] P. R. Menge, G. Gautier, A. Iltis, C. Rozsa, and V. Solovyev, "Performance of large lanthanum bromide scintillators," Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 579, no. 1, pp. 6-10, 2007.

[4] P. Dorenbos, "Light output and energy resolution of [Ce.sup.3+]-doped scintillators," Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 486, no. 1-2, pp. 208-213, 2002.

[5] W. Drozdowski, P. Dorenbos, A. J. Bos, G. Bizarri, A. Owens, and F. G. Quarati, "Ce[Br.sub.3] scintillator development for possible use in space missions," IEEE Transactions on Nuclear Science, vol. 55, no. 3, pp. 1391-1396, 2008.

[6] P. Guss, M. Reed, D. Yuan et al., "Size effect on nuclear gamma-ray energy spectra acquired by different-sized Ce[Br.sub.3], La[Br.sub.3]:Ce, and NaI:Tl gamma-ray detectors," Nuclear Technology, vol. 185, no. 3, pp. 309-321, 2014.

[7] P. Guss, M. Reed, D. Yuan, M. Cutler, C. Contreras, and D. Beller, "Comparison of Ce[Br.sub.3] with La[Br.sub.3]:Ce, La[Cl.sub.3]:Ce, and NaI:Tl detectors," in Proceedings Volume 7805, Hard XRay, Gamma-Ray, and Neutron Detector Physics XII, 78050L, 2010.

[8] A. A. Naqvi, F. Z. Khiari, F. A. Liadi, U.-R. Khateeb-, and A. A. Isab, "Performance tests of a large volume cerium tribromide (Ce[Br.sub.3]) scintillation detector," Applied Radiation and Isotopes, vol. 114, pp. 50-56, 2016.

[9] F. G. Quarati, P. Dorenbos, J. Van der Biezen et al., "Scintillation and detection characteristics of high-sensitivity Ce[Br.sub.3] gamma-ray spectrometers," Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 729, pp. 596-604, 2013.

[10] F. G. Quarati, M. S. Alekhin, K. W. Kramer, and P. Dorenbos, "Co-doping of Ce[Br.sub.3] scintillator detectors for energy resolution enhancement," Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, vol. 735, pp. 655-658, 2014.

[11] D. N. Ter Weele, D. R. Schaart, and P. Dorenbos, "The effect of self-absorption on the scintillation properties of [Ce.sup.3+] activated La[Br.sub.3] and Ce[Br.sub.3]," IEEE Transactions on Nuclear Science, vol. 61, no. 1, pp. 683-688, 2014.

[12] A. A. Naqvi, F. Z. Khiari, F. A. Liadi, U.-R. Khateeb-, and A. A. Isab, "Energy resolution measurements of Ce[Br.sub.3] and La[Cl.sub.3]:Ce detectors," Journal of Radioanalytical and Nuclear Chemistry, vol. 310, no. 2, pp. 849-855, 2016.

[13] H. D. Choi, R. B. Firestone, R. M. Lindstrom et al., Database of Prompt Gamma-Rays from Slow Neutron Capture for Elemental Analysis, INTERNATIONAL ATOMIC ENERGY AGENCY, Vienna, Austria, 2006.

[14] G. L. Molnar, Z. Revay, and T. Belgya, "Accurate absolute intensities for the 35Cl (n,[gamma]) reaction gamma-ray standard," Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, vol. 213, pp. 32-35, 2004.

[15] J. R. Mullaney, D. L. Lorenz, and A. D. Arntson, "Chloride in groundwater and surface water in areas underlain by the glacial aquifer system, Northern United States," National Water-Quality Assessment Program Scientific Investigations Report # 2009-5086, U.S. Department of the Interior and U.S. Geological Survey, 2009, http://www.usgs.gov.

[16] U.S. Environmental Protection Agency, Secondary Drinking Water Regulations--Guidance for Nuisance Chemicals: EPA 810/K-92-001, 1992, http://www.epa.gov/safewater/consu mer/2ndstandards.html.

[17] J. F. Briesmeister, Ed., "MCNP4B2--a General Monte Carlo N-particles transport code," Los Alamos National Laboratory Report, LA-12625. Version 4C, Los Alamos National Laboratory Report, LA-12625-M, 1997.

A. A. Naqvi, (1) F. Z. Khiari, (1) F. A. Liadi, (1) Khateeb ur-Rehman, (1) M. Raashid, (1) and A. A. Isab (2)

(1) Department of Physics, King Fahd University of Petroleum and Minerals, Dhahran, Saudi Arabia

(2) Department of Chemistry, King Fahd University of Petroleum and Minerals, Dhahran, Saudi Arabia

Correspondence should be addressed to A. A. Naqvi; aanaqvi@kfupm.edu.sa

Received 8 June 2017; Accepted 20 September 2017; Published 21 December 2017

Academic Editor: Javier Garcia-Guinea

Caption: Figure 1: Schematic representation of the MP320 portable neutron generator-based PGNAA setup used to measure the prompt gamma-ray yields.

Caption: Figure 2: Prompt gamma-ray spectrum due to activation of the LaCl3:Ce detector caused by the capture of thermal neutrons in La, Ce and Cl in LaCl3:Ce detector material over 0.69-3.51 MeV range.

Caption: Figure 3: Prompt gamma-ray spectrum due to activation of the LaCl3:Ce detector caused by the capture of thermal neutrons in La, Ce, and Cl in LaCl3:Ce detector material over 3.53-9.23 MeV range.

Caption: Figure 4: Integrated yield of 517, 788, 1165, and 1951 keV prompt gamma-ray spectra of the Ce[Br.sub.3] detector from chlorine contaminated water samples plotted as a function of chlorine concentration. The lines show normalized-calculated yields of the gamma rays obtained through Monte Carlo calculations.

Caption: Figure 5: Integrated yield of 2863, 6110, 6619, and 8578 keV prompt gamma-ray spectra of the Ce[Br.sub.3] detector from chlorine-contaminated water samples plotted as a function of chlorine concentration. The lines show normalized-calculated yields of the gamma rays obtained through Monte Carlo calculations.

Caption: Figure 6: Prompt gamma-ray spectrum due to activation of the Ce[Br.sub.3] detector caused by the capture of thermal neutrons in Br and Ce elements present in the Ce[Br.sub.3] detector along with hydrogen capture peak from the moderator.

Caption: Figure 7: Low-energy pulse height spectrum of the Ce[Br.sub.3] detector for a saline water sample over 0.45-2.29 MeV range from a 4% chlorine-contaminated water sample superimposed upon background spectrum.

Caption: Figure 8: High-energy pulse height spectrum of the Ce[Br.sub.3] detector for a saline water sample over 1.75-9.98 MeV range from a 4% chlorine-contaminated water sample superimposed upon background spectrum.

Caption: Figure 9: Enlarged high-energy pulse height spectrum of the Ce[Br.sub.3] detector for a saline water sample over 2.57-8.75 MeV range from a 4% chlorine-contaminated water sample superimposed upon background spectrum.

Caption: Figure 10: Pulse height difference spectra of the Ce[Br.sub.3] detector for 517 keV chlorine gamma rays from 2%, 4%, 6%, and 8% chlorine-contaminated water samples over 0.42-0.66 MeV range superimposed upon each other.

Caption: Figure 11: Pulse height difference spectra of the Ce[Br.sub.3] detector for 776-778 keV chlorine gamma rays from 2%, 4%, 6%, and 8% chlorine-contaminated water samples over 0.66-0.91 MeV range superimposed upon each other.

Caption: Figure 12: Pulse height difference spectra of the Ce[Br.sub.3] detector for 1951-1959 keV chlorine gamma rays from 2%, 4%, 6%, and 8% chlorine-contaminated water samples over 1.78-2.13 MeV range superimposed upon each other.
Table 1: Energies and partial elemental cross section
[[sigma].sub.[gamma].sup.z]([E.sub.[gamma]])
barns of prominent capture gamma rays of bromine,
cerium, and chlorine [13].

Element   Gamma-ray energy (keV)     [[sigma].sub.
                                    [gamma].sup.z]
                                   ([E.sub.[gamma]])
                                         barns

                   196                   0.43
                   271                   0.46
                   275                   0.16
                   315                   0.46
                   367                   0.23
Br                 513                   0.21
                   661                   0.08
                   828                   0.28
                   1248                  0.053
                   7577                  0.10
                   475                   0.082
Ce                 662                   0.241
                   1107                  0.040
Ni                 8533                  0.721
                   8998                  1.49
                   517                   7.58
                   786                   3.42
                   788                   5.42
                   1165                  8.91
                   1601                  1.21
                   1951                  6.33
                   1959                  4.10
Cl                 2864                  1.82
                   3062                  1.13
                   4980                  1.23
                   5715                  1.82
                   6111                  6.59
                   6619                  2.53
                   7414                  3.29
                   7790                  2.66
                   8578                  0.88

Table 2: Minimum detectable concentration (MDC) of chlorine in
water samples for various gamma rays using Ce[Br.sub.3]
detector along with portable neutron generator-based
PGNAA setup.

Cl-gamma ray      Ce[Br.sub.3] detector
energy (keV)
               MDC (wt%)   [[sigma].sub.MDC]
                                 (wt%)

517              0.08            0.02
786-788          0.07            0.02
1165             0.18            0.05
1951/1959        0.11            0.03
2863             0.17            0.05
6111             0.13            0.04
6619             0.18            0.05
8578             0.80            0.24
COPYRIGHT 2018 Hindawi Limited
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2018 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Research Article
Author:Naqvi, A.A.; Khiari, F.Z.; Liadi, F.A.; ur-Rehman, Khateeb; Raashid, M.; Isab, A.A.
Publication:Journal of Spectroscopy
Date:Jan 1, 2018
Words:4420
Previous Article:Different Spectrophotometric Methods Manipulating Ratio Spectra Applied for the Analysis of Aclidinium in Duaklir[R] Genuair[R] Inhalation Powder.
Next Article:Influence of the Uniaxial Hot-Pressing Sintering Condition on [Bi.sub.4][Ge.sub.3][O.sub.12] Ceramic Scintillators.
Topics:

Terms of use | Privacy policy | Copyright © 2021 Farlex, Inc. | Feedback | For webmasters |