Microwave characterization of low-loss solid dielectric materials using rectangular waveguide.

1 INTRODUCTION

Application of materials in the microwave, aerospace, and communications demands an accurate knowledge of material properties such as complex permittivity and complex permeability. The active research fields like materials science and electrical and electronic engineering [1] need necessarily information about permittivity and permeability for developing novel microwave materials. In the radar absorbing material design, the complex permittivity and/or permeability are the fundamental elements that have to be taken primarily into count. In addition, the material characterization is an important issue in many industrial areas, and numerous microwave methods are proposed in literature. Each technique has its own benefits and limitations. Several factors influence the choice of technique to obtain the desired information on material under test MUT. For example, the frequency band, the sample size, contacting or non-contacting, material state (solid, liquid and powder), etc. are the major factors to take into consideration when selecting the characterization technique. In general, microwave methods can be categorized into three groups: 1) free-space techniques; 2) resonant or cavity perturbation techniques; 3) non resonant techniques based on coaxial or rectangular waveguide. The free-space methods are employed when material is available in a large sheet and when we want to characterize composite materials that their intrinsic heterogeneity does not permit to have small representative samples of the entire material. The free-space techniques are nondestructive and contactless. They are ideally suited at variable incidence of the electromagnetic wave and for high temperature measurements. But unwanted reflections surrounding objects and diffractions from the edges of the sample make the free-space measurements less accurate. The resonant methods [2] are more accurate, but they require elaborate sample preparation, and can be applied to medium- and low-loss materials at narrow frequency band. Microwave non resonant methods [3-6] are widely used over a wide range of frequencies, even though these techniques are less accurate than the resonant methods. In one- and/or two-port measurement techniques, the reflection and/or transmission coefficients of a sample material filled in the coaxial line or rectangular waveguide are measured using the Vector Network Analyzer VNA. Using the inverse procedure, the constituent parameters of the MUT are extracted. In these measurements, the cross section of the MUT is the same of the transmission line and only the dominant mode is assumed.

The work presented in this paper is a continuation of that published in [7], emphasizing the validity of the Transmission/Reflection T/R method in [7] for a variety of solid low-loss dielectric materials. Based on our knowledge, the permittivity information of these materials (Delrin, Peek, Spanish Peek, Nylatron, Vulkollan, Arnite, and Celotex) is unknown at X-band. This T/R technique is applied to uncalibrated scattering parameters (S-parameters). However, the standard calibration manipulations are not needed; these manipulation procedures produce inevitable errors affecting the final results. The calibrated and uncalibrated measurement results are compared between them and the relative errors are calculated over the X-band frequencies.

2 METHOD

We consider the published T/R method in [7] for permittivity determination. This method is more flexible; and it can be applied to the calibrated or uncalibrated S-parameter measurements. This technique eliminates mathematically the systematic errors of the experimental setup like source and load match errors, tracking (frequency) errors, effects of wires carrying interconnections, hardware imperfections of VNA, etc. [3][7][8]. Two measurements in T/R are sufficient to evaluate the complex permittivity of the dielectric sample; the first is that the sample holder filled with the reference dielectric PTFE (Polytetrafluoroethylene), and the second with the MUT. A precise location of the sample in the waveguide is not needed. The rigorous mathematical approach of this method is based on the wave cascading matrix [7, 8] conducting to a nonlinear equation which the relative complex permittivity variable of the MUT is the single unknown. The nonlinear function is resolved using any two-dimensional root finding algorithms [9]. The reader is invited to consult [7] for more details.

3 EXPERIMENTAL SETUP

We consider the measurement setup shown in figure 1. The MUT with thickness d=1 cm is imprecisely located in the WR90 rectangular waveguide of sections (22.86x10.16) [mm.sup.2]. The E8634A VNA is connected to two coaxial-to-waveguide adapters. We suppose that only the dominant mode [TE.sub.10] propagates in the structure. The dielectric samples are machined to the same waveguide sections. The uncalibrated S-parameters of PTFE and MUT are measured. Then, the computer program can determine the complex permittivity of the MUT. In the following section, the uncalibrated and calibrated results were compared for validation of the employed method [7]. The Thru-Reflect-Line TRL calibration technique [10] is utilized for calibrating the experimental setup. All measurements are performed at [8.2-12.4] GHz band with 201 frequency points.

[FIGURE 1 OMITTED]

4 RESULTS AND DISCUSSION

The method of the paper [7] was used to determine the relative complex permittivity of seven low-loss dielectric materials cited above. According to our knowledge, these materials are not cited in literature and especially their permittivity information over X-band frequencies. Given its stability, the relative complex permittivity of PTFE [epsilon]*=2.04-/0.01 is taken as a reference dielectric and the Sparameters of PTFE sample are measured only once. A single specimen of each material is necessary to extract its dielectric constant. The table 1 presents the average values of the relative complex permittivity of the MUT <[[epsilon].sup.*.sub.calib]> and <[[epsilon].sup.*.sub.uncalib]>, and the average relative error percentage on the real and imaginary parts defined as:

[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII]

[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII] are the complex relative permittivity of the MUT determined from calibrated and uncalibrated S-parameter measurements respectively. The average values are evaluated using the standard average such as the number of measurements points is equal to 201.

[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII]

Calculations based on measured information have indicated that the error in the real part is very small (close to zero), but that in the imaginary part can be large (within 7.5%) for the used low-loss materials. Like any other S-parameter based methods, this is a common phenomenon. We note according to the error values that the relative error in the imaginary part decreases considerably toward 1% for Vulkollan and Celotex samples. The latter two materials have significantly high dielectric losses compared to the other studied materials; it appears that the method is well adapted for medium-loss materials.

[FIGURE 2 OMITTED]

[FIGURE 3 OMITTED]

The figures 2 and 3 represent the variation of the real and imaginary parts of the complex permittivity over the X-band. It is seen that the permittivity values are very stable over the frequency range and near to the average values cited in the table 1; except the Celotex sample, that the real part decreases from 4.1 to 3.95. An overall view in both figures, the imaginary parts of all samples are not stable like the real parts, which explain the high average error values in the imaginary parts. Assuming that the experimental conditions are assumed the same for all samples, the errors produced between calibrated and uncalibrated measurements can be explained to the uncertainty in sample lengths.

5 CONCLUSION

We proposed a dielectric characterization for a set of dielectric materials not cited in the literature. We utilized the Transmission/Reflection method previously published. This method needs two scattering parameter measurements; the first is performed with a filled rectangular waveguide by a standard dielectric that its complex permittivity is stable and well known over the X-band, and the second is done, in the same experimental conditions, with the material under test. This technique is well suited for the calibrated or uncalibrated measurements. The relative errors between these two types of measures are calculated for each sample. Errors in the real parts are close to zero, but in the imaginary parts they are within 7% for very low-loss materials and within 1% for medium-loss.

REFERENCES

[1.] Chen. LF, Ong. CK, Neo. CP, Varadan. VV, and Varadan. VK, "Microwave electronics: measurement and materials characterization". John Wiley & Sons, West Sussex, England, 2004.

[2.] Li. E, Nie. Z.-P, Guo. G, Zhang. Q, Li. Z, and He. F, "Broadband measurements of dielectric properties of low-loss materials at high temperatures using circular cavity method", Progress In Electromagnetics Research PIER 2009; 92: 103-120.

[3.] Baker-Jarvis. J, Vanzura. EJ, and Kissick. WA, "Improved technique for determining complex permittivity with the transmission/reflection method", IEEE Trans. Microwave Theory Tech. 1990; 38(8): 1096-1103.

[4.] Boughrit. A, Legrand. C, Chapoton. A, "Noniterative stable transmission/reflection method for low-loss material complex permittivity determination", IEEE Trans. Microwave Theory Tech. 1997; 45(1): 52-57.

[5.] Belrhiti. MD, Bri. S, et al., "Complex permittivity measurement for dielectric materials at microwave frequencies using rectangular waveguide", European Journal of Scientific Research. 2011; 49(2): 234-248.

[6.] Hasar. UC, Yurtcan MT, "A microwave method based on amplitude-only reflection measurements for permittivity determination of low-loss materials", Measurement. 2010; 43: 1255-1265.

[7.] Jebbor. N, Bri. S et al., "Complex permittivity determination with the transmission / reflection method", International Journal of Emerging Sciences 2011; 1(4): 682-695.

[8.] Hasar. UC, "A new calibration-independent method for complex permittivity extraction of solid dielectric materials", IEEE Microwave and Wireless Components Letters 2008; 18(12): 788-790.

[9.] Press. WH, Teukolsky. SA, Vtterling WT, and Flannery BP, "Numerical Recipes in C: The art of scientific computing", Cambridge Univ. Press, New York, 1992.

[10.] Engen. GF, and Hoer. CA, "Thru-Reflect-Line: an improved technique for calibrating the dual six port automatic network analyzer", IEEE Microw. Theory Tech. 1979; 27(2): 987-993.

Nawfal Jebbor (1), Seddik Bri (2), Angel Mediavilla Sanchez (3), Mohamed Chaibi (4)

(1) Spectrometry Laboratory of Materials and Archeomaterials (LASMAR) Faculty of Sciences, Moulay Ismail University, Meknes-Morocco, (2) Electrical Engineering Department, Technology High School (ESTM), Moulay Ismail University, Meknes-Morocco, (3) Department of Communication Engineering, University of Cantabria Avd. Los Castros, Plaza de la Ciencia, s/n 39005 Santander, Spain, (4) Telecommunications and Information Technologies (TTI) PCTCAN, Calle Albert Einstein 14, 39011 Santander, Spain

jebbor.nawfal@gmail.com, briseddik@gmail.com, angel.mediavilla@unican.es, mchaibi@ttinorte.es
Table 1. Average relative complex permittivity <[[epsilon]*
.sub.calib]> and
<[[epsilon] *.sub.uncalib]> determined from the
calibrated and uncalibrated scattering parameter measurements,
respectively. <%Error [epsilon]'> and
<%Error [epsilon]"> are,
respectively, the average relative error percentage on the real and
imaginary parts of the complex permittivity at X-band frequencies.

               <[[epsilon] *    <[[epsilon] *
Materials      .sub.uncalib>     .sub.calib>

Delrin         2.9372-/0.0825   2.9380-/0.0805
Peek           3.2033-/0.0138   3.2032-/0.0130
Spanish Peek   3.0660-/0.0448   3.0662-/0.0431
Nylatron       3.2290-/0.0845   3.2284-/0.0804
Vulkollan      2.9892-/0.2130   2.9894-/0.2105
Arnite         3.0955-/0.0249   3.0949-/0.0232
Celotex        4.0230-/0.3526   4.0222-/0.3488

                 <%Error       <%Error
Materials      [epsilon]'>   [epsilon]">

Delrin            0.027         2.484
Peek              0.00          6.154
Spanish Peek      0.00          3.944
Nylatron          0.018         5.099
Vulkollan         0.006         1.187
Arnite            0.019         7.327
Celotex           0.019         1.089
COPYRIGHT 2012 Springfield Publishing Corporation
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2012 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Author:Jebbor, Nawfal; Bri, Seddik; Sanchez, Angel Mediavilla; Chaibi, Mohamed
Publication:International Journal of Emerging Sciences
Article Type:Report
Geographic Code:6MORO
Date:Dec 1, 2012
Words:1836