Printer Friendly

FDA Warns Magellan Diagnostics of Possible Federal Law Violations.

The U.S. Food and Drug Administration (FDA) issued a warning letter to Magellan Diagnostics Inc. for several violations of federal law, including marketing significantly modified versions of two of its blood lead testing systems without the FDA's required clearance or approval and failing to submit medical device reports to the agency after becoming aware of customer complaints involving discrepancies in blood lead test results.

In May 2017, the FDA advised Americans that Magellan's LeadCare test systems performed on blood drawn from the vein (venous) may provide falsely low results. At that time, the agency launched an investigation into the issue to determine the cause of the inaccurate results, which included an inspection (May 10-June 29) of Magellan's facility.

"The FDA has serious concerns about Magellan Diagnostics' actions after learning about inaccuracies in results using its blood lead testing devices--tests that the American people depend on," said Donald St. Pierre, acting director of the Office of In Vitro Diagnostics and Radiological Health and deputy director of new product evaluation in the FDA's Center for Devices and Radiological Health. "The evidence uncovered during the inspection shows the company put patients at risk after it recognized that its tests could provide inaccurate results and failed to take appropriate steps to report this issue and work through a strategy to effectively mitigate problems with the FDA. We take the violations described in the warning letter very seriously and are continuing to work with the company to investigate the root cause of the falsely low test results."

During the inspection, FDA investigators observed numerous violations, including failing to submit medical device reports to the agency regarding discrepancies in test results. In addition, the inspection showed that Magellan altered two of its blood lead testing systems after they were already FDA cleared by changing the amount of time the mixture of blood samples with the reaction substance (reagent) must sit before analysis in an effort to reduce the risk of falsely low results; however, Magellan did not report these significant changes to the FDA for review and evaluation as required.

Magellan Diagnostics was issued a deadline of Nov. 10 to notify the FDA about specific steps the company has taken to address these violations and to prevent them from recurring. If the company fails to promptly correct these violations, the FDA may take additional action, such as seizure, injunction and civil money penalties.

The FDA continues to encourage people to follow the Centers for Disease Control and Prevention's re-testing recommendations and the FDA's recommendations for healthcare professionals. In addition, the FDA is continuing its investigation into the root cause of the inaccurate lead test results and is working with the CDC on an independent analysis of Magellan Diagnostics' LeadCare System tests.

COPYRIGHT 2017 Rodman Publishing
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2017 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:TOP OF THE NEWS
Publication:Medical Product Outsourcing
Date:Nov 1, 2017
Words:456
Previous Article:Abbott Launches Smartphone Compatible Insertable Cardiac Monitor in U.S.
Next Article:Boston Scientific Buys Apama Medical for $300M.

Terms of use | Privacy policy | Copyright © 2020 Farlex, Inc. | Feedback | For webmasters