Evaluation of Body Composition in Hemodialysis Thai Patients: Comparison between Two Models of Bioelectrical Impedance Analyzer and Dual-Energy X-Ray Absorptiometry.

1. Introduction

The prevalence of chronic kidney disease (CKD) in the Thai population becomes much higher than previously known [1]. As a consequence, the patients undergoing renal replacement therapy of hemodialysis have been rising [2]. Regular hemodialysis causes the decrease of both FM and FFM over times which is independently associated with a higher mortality and a tendency toward a worse quality of life [3]. These also increase the prevalence of protein-energy malnutrition [4] and morbidity [5].

Early detection of malnourishment and optimized nutritional care can improve the outcomes [6]. Therefore, the body composition analysis is one of the most important strategies to assess and monitor the nutritional status. Searching for a practical and accurate tool for body composition evaluation is essential.

Dual-energy X-ray absorptiometry (DXA), based on the signals from two energy sources to provide a three compartment model of body composition, is taken place and has become a gold standard test [7]. DXA is a reproducible and reliable technique for measuring fat mass in healthy [8] as well as in hemodialysis patients [9]. Unfortunately, this costly device, which is nonportable and depended on proficiency, cannot be used as a practical or accessible bedside tool. Additionally, the body compartments, in particular total body water (TBW) in chronic hemodialysis population, are significantly altered comparing to healthy population [10]. Bioelectrical impedance analysis (BIA) with multifrequency technique has been proven to be one of the most valid methods comparing to DXA with high correlation in the healthy population [11]. However, the estimation by the BIA analyzers compared with that measured by DXA in hemodialysis patients was found slightly higher in fat mass (FM) and slightly lower in fat-free mass (FFM), but significantly [12]. The error of the BIA was found greater in patients with CKD than in healthy subjects [13]. Because of its readily accessible, low cost, and quickly assessing procedure, the BIA method has become widely used in clinical practice, in sport medicine, and also in weight reduction programs [7, 14].

To date, none of the studies have investigated the agreement between BIA and DXA in measurement body composition in hemodialysis Thai population. Due to the effect of race [15], the accuracy of BIA must be evaluated. The aims of the present study were to compare the agreement between BIA and DXA in measurement of the body composition and to evaluate their associated factors.

2. Materials and Methods

2.1. Subjects. This study was a cross-sectional study conducted in the hemodialysis unit of Ramathibodi Hospital, Mahidol University, Bangkok, Thailand. All subjects, aged more than 18 years, who had regularly been on maintenance hemodialysis for at least 3 months prior to the study were included. The study was performed during October 2013 and May 2014. Patients were treated with three sessions of dialysis per week on schedule, lasting 4 hours per period. The percentage of renal replacement therapy methods was found 53% with hemodialysis and 47% with hemodiafiltration. This study excluded all subjects who were unstable or currently on medications affecting metabolic rate or admitted in the hospital during the period of study. All of the measurements were done on the dialysis day. Baseline data including age, sex, nutritional information, education background, socioeconomic status, physical examinations, and anthropometric measurements were performed and completed by the well-trained physician.

The protocol was approved by the Institutional Review Board, Faculty of Medicine, Ramathibodi Hospital, Mahidol University (Approval number MURA2013/317 [Ns.sub.1] Feb 17). Written informed consent was obtained from each participant.

2.2. Measurement and Laboratory Determinations. Anthropometric parameters including weight, height, and waist circumference were measured twice with standard techniques by the same skillful physician. All measurements were performed in the same day after the subjects had fasted for 12 hours and within 30 minutes after termination of hemodialysis period. All participants dressed in light clothes without shoes. Body mass index (BMI) was calculated by the postdialysis body weight (kilogram (kg)) divided by the height squared (meter [(m).sup.2]). BIAs were measured by 2 models of multifrequency impedance analyzers (model InBody S10, Biospace Co., Ltd., Seoul, Korea, and model InBody 720 Biospace Co., Ltd., Seoul, Korea) which provide 6 different frequency impedance measurements (1, 5, 50, 250, 500, and 1000 kHz) and 3 different frequencies of phase angle measurement (5, 50, and 250 kHz) at each 5 segments (right arm, left arm, trunk, right leg, and left leg). Model InBody S10 was conducted while the patient was lying supine for 15 minutes on a bed with legs apart and arms not touching the torso after all metals were removed. The touch type electrodes were placed following the manuscript of the model, whereas model InBody 720 was performed while the patient was standing upright: hands hold the electrodes and feet on the electrodes, with 8-point tactile electrode method. The output values included the intracellular fluid, extracellular fluid, FM, FFM, and %FM. By extracellular fluid (ECF)/(TBW), edematous state was detected by the BIA method [16]. In addition, the FM index (FMI) was determined by the postdialysis FM (kg) divided by the height squared ([m.sup.2]), and FFM index (FFMI) was determined by the postdialysis FFM (kg) divided by the height squared ([m.sup.2]).

DXA was performed, immediately after BIA measurement in the same day, using the Hologic Discovery A instrument (Hologic Inc., Waltham, MA, USA). All scans were performed by the same trained technician and analyzed by the same radiologist. The calibration was done each day prior to start of testing by the standard technique. The assessed data were FM, FFM, %FM, and bone mass.

MIS (Malnutrition-Inflammation Score) is the scoring system more comprehensive and quantitative evaluation criteria composed of 10 components: the 7 Dialysis Malnutrition Score components and 3 new items (body mass index (BMI), serum albumin level, and total iron-binding capacity (TIBC)) have been added. In a recent prospective study, the MIS was found to be a comprehensive scoring system which had the significant associations with prospective hospitalization and mortality [17, 18].

2.3. Statistical Methods. All statistical analyses were performed using STATA 12.0 software (StataCorp. 2011: Stata Statistical Software: Release 12. College Station, TX: StataCorp LP). Mean [+ or -] standard deviation (SD) or median (interquartile range (IQR)) for continuous variable and frequency (%) for binary or categorical variable were presented. Paired t-tests were used to compare mean of FM, %FM, FFM, FMI, and FFMI measured by BIA and DXA. The correlation between FM, %FM, FFM, FMI, and FFMI predicted by BIA and those measured by DXA was determined by Pearson's correlation coefficient (r). The Bland and Altman analysis [19] was used to compare the agreement between the measurement techniques. The limits of agreement between methods were defined as the mean difference [+ or -] 1.96 SD (95% limits of agreement). The correlation between the intermethod differences and each body parameter was obtained by Pearson's correlation coefficient (r) test. A statistical significance was attained when a p value was less than 0.05.

3. Results

Sixty-nine measurements were performed. A total of 66% of patients were male, and the mean age was 59.66 [+ or -] 11.28 years, ranging from 40 to 87 years. Patient BMIs were between 17.41 and 35.76 kg/[m.sup.2]: 19% of them had a BMI of <18.5 kg/[m.sup.2], and 47% had a BMI of >23 kg/[m.sup.2]. By MIS, 64% of patients were defined as severe malnutrition. Edematous state was detected by both BIA methods (ECF/TBW). The averages of normalized protein catabolic rate (nPCR), Kt/v, and dialysis vintage were 0.91 [+ or -] 0.20 g/kg, 1.78 [+ or -] 0.39, and 5.21 (2.22-12.74) years, respectively (Table 1).

In both genders, no significant intermethod difference of FM, %FM, FFM, FMI, and FFMI measured by InBody S10 and DXA was found ; on the other hand, means of FM, % FM, and FMI measured by BIA 720 were significantly higher than when measured by BIA S10 and DXA (Table 2). The significant difference in means of FM, %FM, and FMI measured by all 3 methods was not observed between genders. Nevertheless, all methods showed significantly higher FFM and FFMI in men (p < 0.001).

The correlation coefficients (r) of body compositions between DXA and InBody S10 were high (Fm: r = 0.93; % FM: r = 0.85; FFM: r = 0.88; FMI: r = 0.95; FFMI: r = 0.78) with p < 0.001. Strong correlations were also found between DXA and InBody 720 (FM: r = 0.94; %FM: r = 0.88; FFM: r = 0.89; FMI: r = 0.96; FFMI: r = 0.81) with p < 0.001. The correlation coefficients of these measurements between DXA and both BIA methods in each gender are illustrated in Figures 1-5.

By Bland and Altman analysis [19], the differences of FM, %FM, and FMI between DXA and InBody S10 method were smaller than that between DXA and InBody 720. However, both BIA methods had wide 95% limits of agreement with DXA (Table 3). These intermethod differences did not significantly differ between male and female. The agreements of all measures in both genders illustrated by Bland and Altman plots also revealed the same direction: between BIA S10 and DXA (Figure 6) and between BIA 720 and DXA (Figure 7). With regard to the measures between BIA S10 and DXA, the differences of % FM, FFM, and FFMI were significantly correlated with age (%FM: r = 0.36, p = 0.002; FFM: r = -0.34, p = 0.004; FFMI: r = -0.35, p = 0.003), body weight (%FM: r = 0.25, p = 0.039; FFM: r = -0.26, p = 0.034; FFMI: r = -0.27, p = 0.025), and edematous stage (%FM: r = 0.39, p = 0.001; FFM: r = -0.32, p = 0.008; FFMI: r = -0.32, p = 0.008), whereas the differences of FM and FFM were associated with only age (FM: r = 0.40, p = 0.001; FMI: r = -0.39, p = 0.001) and edematous stage (FM: r = 0.39, p = 0.001; FMI: r = 0.38, p = 0.001). Nonetheless, any relationship between age and edematous stage and the difference measured by BIA 720 and DXA was not observed. Only body weight was found to have a significant association with the difference of % FM (r = 0.39, p = 0.002), FFM (r = -0.26, p = 0.038), FMI (r = 0.27, p = 0.034), and FFMI (r = -0.31, p = 0.016).

4. Discussion

The deterioration of body compositions, both FM and FFM, is strongly correlated with morbidity and mortality and represents a poor prognostic marker [20]. Hence, early nutritional care is very important to prevent this morbidity. Nevertheless, the measurement of FFM, which is predominantly composed of muscle mass, body water, and minerals, is affected by abnormal fluid and electrolyte distribution, and commonly observed in patients undergoing renal replacement therapy.

In agreement with the previous reports [7, 12], the present study observed highly significant correlations in the measurement of FM and FFM between DXA and both BIA methods. The FM measured by both BIA methods was lower than the DXA, and this could be the result of edema in hemodialysis patients in this study. The FFM of both genders was underestimated by InBody S10 but overestimated by InBody 720. As a result of gravity, the water distribution in the supine position differs from the upright position in the edematous state, [21] and the variation of body water distribution by BIA measurement of body composition is also affected by different positions [22]. All methods showed significantly higher FFM in men according to the normal physiology. The mean differences of FM, %FM, and FMI between DXA and InBody S10 method for both genders were small and not significant, whereas the mean differences between DXA and InBody 720 were higher. These results support the idea that the measurement of FM, %FM, and FFM by both BIA methods could be clinically, practically, and reasonably used in follow-up. BIA can also measure the edematous stage in this population. However, these devices cannot provide the accuracy of DXA when measuring FM and FFM.

The results between body compositions (FM and FFM) and their calculated indices (FMI and FFMI) measured by both BIA and DXA were in the same direction in this study. FMI and FFMI may be beneficial for nutritional assessment and easier recognition [23], but FM and FFM are also useful and familiar due to the direct report from the devices. The limitation of this study is the modest sample size. Nevertheless, this is the first study which compares DXA and both of 2 BIA methods in hemodialysis Thai patients.

5. Conclusions

In conclusion, the present study depicted that body composition values measured by DXA and both BIA methods had strongly significant correlations. However, significant differences between measurement by DXA and InBody 720 methods were also found but not between measurement by DXA and InBody S10. As a result, both BIA methods could be practically used as an instrument to follow FM, FFM, FMI, and FFMI in the same individuals. The body weight was an associated factor with the difference of FFM and FFMI when measured by DXA and both BIA methods. Notwithstanding, age and edematous stage were correlated with the intermethod difference when measured by DXA and InBody S10. Further study with a larger number of hemodialysis Thai patients should be warranted.
Abbreviations

BIA:       Bioelectrical impedance analysis
BMI:       Body mass index
BW:        Body weight
CKD:       Chronic kidney disease
DXA:       Dual-energy X-ray absorptiometry
ECF:       Extracellular fluid
FFM:       Fat-free mass
FFMI:      Fat-free mass index
FMI:       Fat mass index
FM:        Fat mass
ICF:       Intracellular fluid
IQR:       Interquartile range
kg:        Kilogram
m:         Meter
MIS:       Malnutrition-Inflammation Score
nPCR:      Normalized protein catabolic rate
SD:        Standard deviation
TBW:       Total body water
TIBC:      Total iron-binding capacity.


https://doi.org/10.1155/2018/4537623

Data Availability

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Ethical Approval

The protocol was approved by the Ethical Clearance Committee on Human

Rights Related to Researches Involving Human Subjects, Faculty of Medicine, Ramathibodi Hospital, Mahidol University (Approval number MURA2013/317 [Ns.sub.1] [Feb.sub.17]).

Consent

Before participation in this study, written informed consent was obtained from each participant.

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Authors' Contributions

Kulapong Jayanama conducted the research, conceived the study, designed the research, analyzed and interpreted the data, and drafted the manuscript. Supanee Putadechakun conducted the research, conceived the study, and revised the manuscript. Praopilad Srisuwarn conducted the research and conceived the study. Sakda Arj-Ong Vallibhakara analyzed and interpreted the data. Prapimporn Chattranukulchai Shantavasinkul conceived the study and revised the manuscript. Chanika Sritara conceived the study. Surasak Kantachuvesiri revised the manuscript. Surat Komindr conceived the study, designed the research, interpreted the data, and revised the manuscript. All authors read and approved the final manuscript.

Acknowledgments

This study was supported by a research grant of Faculty of Medicine Ramathibodi Hospital, Mahidol University. The authors are grateful to all patients who participated in this study, all staff members of Division of Nephrology, Department of Medicine, Faculty of Medicine, Ramathibodi Hospital, Mahidol University, and all nurses from the hemodialysis unit of Ramathibodi Hospital, Mahidol University.

References

[1] A. Ingsathit, A. Thakkinstian, A. Chaiprasert et al., "Prevalence and risk factors of chronic kidney disease in the Thai adult population: Thai SEEK study," Nephrology Dialysis Transplantation, vol. 25, no. 5, pp. 1567-1575, 2010.

[2] L. Ong-Ajyooth, K. Vareesangthip, P. Khonputsa, and W. Aekplakorn, "Prevalence of chronic kidney disease in Thai adults: a national health survey," BMC Nephrology, vol. 10, no. 1, p. 35, 2009.

[3] K. Kalantar-Zadeh, N. Kuwae, D. Y. Wu et al., "Associations of body fat and its changes over time with quality of life and prospective mortality in hemodialysis patients," American Journal of Clinical Nutrition, vol. 83, no. 2, pp. 202-210, 2006.

[4] C. M. Piratelli and R. Telarolli Junior, "Nutritional evaluation of stage 5 chronic kidney disease patients on dialysis," Sao Paulo Medical Journal, vol. 130, no. 6, pp. 392-397, 2012.

[5] W. M. McClellan, C. Anson, K. Birkeli, and E. Tuttle, "Functional status and quality of life: predictors of early mortality among patients entering treatment for end stage renal disease," Journal of Clinical Epidemiology, vol. 44, no. 1, pp. 83-89, 1991.

[6] Y. P. Jadeja and V. Kher, "Protein energy wasting in chronic kidney disease: an update with focus on nutritional interventions to improve outcomes," Indian Journal of Endocrinology and Metabolism, vol. 16, no. 2, pp. 246-251, 2012.

[7] N. Gupta, G. Balasekaran, V. Victor Govindaswamy, C. Y. Hwa, and L. M. Shun, "Comparison of body composition with bioelectric impedance (BIA) and dual energy X-ray absorptiometry (DEXA) among Singapore Chinese," Journal of Science and Medicine in Sport/Sports Medicine Australia, vol. 14, no. 1, pp. 33-35, 2011.

[8] J. Wang, S. B. Heymsfield, M. Aulet, J. C. Thornton, and R. N. Pierson Jr., "Body fat from body density: underwater weighing vs. dual-photon absorptiometry," American Journal of Physiology-Endocrinology and Metabolism, vol. 256, no. 6, pp. E829-E834, 1989.

[9] F. F. Horber, F. Thomi, J. P. Casez, J. Fonteille, and P. Jaeger, "Impact of hydration status on body composition as measured by dual energy X-ray absorptiometry in normal volunteers and patients on haemodialysis," British Journal of Radiology, vol. 65, no. 778, pp. 895-900, 1992.

[10] V. Jha, A. Jairam, M. C. Sharma, V. Sakhuja, A. Piccoli, and S. Parthasarathy, "Body composition analysis with bioelectric impedance in adult Indians with ESRD: comparison with healthy population," Kidney International, vol. 69, no. 9, pp. 1649-1653, 2006.

[11] G. Sun, C. R. French, G. R. Martin et al., "Comparison of multifrequency bioelectrical impedance analysis with dualenergy X-ray absorptiometry for assessment of percentage body fat in a large, healthy population," American Journal of Clinical Nutrition, vol. 81, no. 1, pp. 74-78, 2005.

[12] C. Donadio, A. B. Halim, F. Caprio, G. Grassi, B. Khedr, and M. Mazzantini, "Single- and multi-frequency bioelectrical impedance analyses to analyse body composition in maintenance haemodialysis patients: comparison with dual-energy X-ray absorptiometry," Physiological Measurement, vol. 29, no. 6, pp. S517-S524, 2008.

[13] G. Woodrow, B. Oldroyd, J. H. Turney, P. S. Davies, J. M. Day, and M. A. Smith, "Measurement of total body water by bioelectrical impedance in chronic renal failure," European Journal of Clinical Nutrition, vol. 50, no. 10, pp. 676-681, 1996.

[14] A. Sartorio, M. Proietti, P. G. Marinone, F. Agosti, F. Adorni, and C. L. Lafortuna, "Influence of gender, age and BMI on lower limb muscular power output in a large population of obese men and women," International Journal of Obesity, vol. 28, no. 1, pp. 91-98, 2004.

[15] B. M. Prior, K. J. Cureton, C. M. Modlesky et al., "In vivo validation of whole body composition estimates from dualenergy X-ray absorptiometry," Journal of Applied Physiology, vol. 83, no. 2, pp. 623-630, 1997.

[16] G. Bedogni, M. Malavolti, S. Severi et al., "Accuracy of an eight-point tactile-electrode impedance method in the assessment of total body water," European Journal of Clinical Nutrition, vol. 56, no. 11, pp. 1143-1148, 2002.

[17] K. Kalantar-Zadeh, J. D. Kopple, G. Block, and M. H. Humphreys, "A malnutrition-inflammation score is correlated with morbidity and mortality in maintenance hemodialysis patients," American Journal of Kidney Diseases, vol. 38, no. 6, pp. 1251-1263, 2001.

[18] L. C. Ho, H. H. Wang, Y. S. Peng et al., "Clinical utility of malnutrition-inflammation score in maintenance hemodialysis patients: focus on identifying the best cut-off point," American Journal of Nephrology, vol. 28, no. 5, pp. 840-846, 2008.

[19] J. M. Bland and D. G. Altman, "Statistical methods for assessing agreement between two methods of clinical measurement," The Lancet, vol. 327, no. 8476, pp. 307-310, 1986.

[20] A. Bonanni, I. Mannucci, D. Verzola et al., "Protein-energy wasting and mortality in chronic kidney disease," International Journal of Environmental Research and Public Health, vol. 8, no. 5, pp. 1631-1654, 2011.

[21] G. Medrano, S. Leonhardt, and P. Zhang, "Modeling the influence of body position in bioimpedance measurements," in Proceedings of the Annual International Conference of the IEEE Engineering in Medicine and Biology Society IEEE Engineering in Medicine and Biology Society Annual Conference, pp. 3934-3937, Sydney, NSW, Australia, 2007.

[22] B. R. Di Iorio, V. Terracciano, and V. Bellizzi, "Bioelectrical impedance measurement: errors and artifacts," Journal of Renal Nutrition, vol. 9, no. 4, pp. 192-197, 1999.

[23] T. B. VanItallie, M. U. Yang, S. B. Heymsfield, R. C. Funk, and R. A. Boileau, "Height-normalized indices of the body's fatfree mass and fat mass: potentially useful indicators of nutritional status," American Journal of Clinical Nutrition, vol. 52, no. 6, pp. 953-959, 1990.

Kulapong Jayanama (iD), (1) Supanee Putadechakun (iD), (2) Praopilad Srisuwarn (iD), (2) Sakda Arj-Ong Vallibhakara (iD) (3), Prapimporn Chattranukulchai Shantavasinkul, (2) Chanika Sritara (iD), (4) Surasak Kantachuvesiri, (5) and Surat Komindr (iD) (2)

(1) Chakri Naruebodindra Medical Institute, Faculty of Medicine Ramathibodi Hospital, Mahidol University, 270 Rama 6 Road Thungphayathai, Ratchathewi, Bangkok 10400, Thailand

(2) Division of Nutrition and Biochemical Medicine, Department of Medicine, Faculty of Medicine Ramathibodi Hospital, Mahidol University, 270 Rama 6 Road Thungphayathai, Ratchathewi, Bangkok 10400, Thailand

(3) Section for Clinical Epidemiology and Biostatistics, Faculty of Medicine Ramathibodi Hospital, Mahidol University, 270 Rama 6 Road Thungphayathai, Ratchathewi, Bangkok 10400, Thailand

(4) Division of Nuclear Medicine, Department of Radiology, Faculty of Medicine Ramathibodi Hospital, Mahidol University, 270 Rama 6 Road Thungphayathai, Ratchathewi, Bangkok 10400, Thailand

(5) Division of Nephrology, Department of Medicine, Faculty of Medicine Ramathibodi Hospital, Mahidol University, 270 Rama 6 Road Thungphayathai, Ratchathewi, Bangkok 10400, Thailand

Correspondence should be addressed to Kulapong Jayanama; kulapong.jay@mahidol.ac.th and Surat Komindr; surat.kom@ mahidol.ac.th

Received 30 March 2018; Revised 4 June 2018; Accepted 24 June 2018; Published 5 August 2018

Academic Editor: Michael B. Zemel

Caption: FIGURE 1: Correlation between fat mass in hemodialysis patients: (a) measured by DXA and BIA S10 in male, (b) measured by DXA and BIA S10 in female, (c) measured by DXA and BIA 720 in male, and (d) measured by DXA and BIA 720 in female.

Caption: FIGURE 2: Correlation between percentages of fat mass in hemodialysis patients: (a) measured by DXA and BIA S10 in male, (b) measured by DXA and BIA S10 in female, (c) measured by DXA and BIA 720 in male, and (d) measured by DXA and BIA 720 in female.

Caption: FIGURE 3: Correlation between fat-free mass in hemodialysis patients: (a) measured by DXA and BIA S10 in male, (b) measured by DXA and BIA S10 in female, (c) measured by DXA and BIA 720 in male, and (d) measured by DXA and BIA 720 in female.

Caption: FIGURE 4: Correlation between fat mass index in hemodialysis patients: (a) measured by DXA and BIA S10 in male, (b) measured by DXA and BIA S10 in female, (c) measured by DXA and BIA 720 in male, and (d) measured by DXA and BIA 720 in female.

Caption: FIGURE 5: Correlation between fat-free mass index in hemodialysis patients: (a) measured by DXA and BIA S10 in male, (b) measured by DXA and BIA S10 in female, (c) measured by DXA and BIA 720 in male, and (d) measured by DXA and BIA 720 in female.

Caption: FIGURE 6: The agreement between DXA and BIA S10 in (a) fat mass, (b) percentage of fat mass, (c) fat-free mass, (d) fat mass index, and (e) fat-free mass index. Black filled-in triangles represent values of male, and black circles represent values of female.

Caption: FIGURE 7: The agreement between DXA and BIA 720 in (a) fat mass, (b) percentage of fat mass, (c) fat-free mass, (d) fat mass index, and (e) fat-free mass index. Black filled-in triangles represent values of male, and black circles represent values of female.
Table 1: Demographic, anthropometric, and nutritional status
parameters of the hemodialysis patients (mean [+ or -] SD).

Characteristics                         Hemodialysis patients
                                         (mean [+ or -] SD)

Male, n (%)                                    24 (66)
Age (years)                             59.15 [+ or -] 10.67
BW (kg)                                 61.51 [+ or -] 12.40
BMI (kg/[m.sup.2]), median               22.65 (19.31-25.90)
(IQR)
  BMI > 23 kg/[m.sup.2], n (%)                 17 (47)
Waist circumference (cm)                88.90 [+ or -] 12.81
Mid-upper arm                            28.13 [+ or -] 4.10
circumference (cm)
Triceps skinfold thickness,              13.60 (11.45-19.20)
mm, median (IQR)
MIS
  A, n (%)                                      2 (5)
  B, n (%)                                     11 (31)
  C, n (%)                                     23 (64)
BIA (InBody S10)
  ECF (kg)                               12.26 [+ or -] 2.17
  ICF (kg)                               19.10 [+ or -] 4.06
  ECF/TBW                                0.39 [+ or -] 0.02
BIA (InBody 720)
  ECF (kg)                               13.67 [+ or -] 2.68
  ICF (kg)                                21.28[+ or -]3.95
  ECF/TBW                                0.39 [+ or -] 0.02
nPCR (g/kg)                              0.91 [+ or -] 0.20
Kt/V                                     1.78 [+ or -] 0.39
Dialysis vintage (years),                 5.21 (2.22-12.74)
median (IQR)

BIA = bioelectrical impedance analysis; BW = body weight; ECF =
extracel-lular fluid; ICF = intracellular fluid; IQR = interquartile
range; MIS = Malnutrition-Inflammation Score; SD = standard deviation;
TBW = total body water; nPCR = normalized protein catabolic rate.

Table 2: Body composition parameters of the hemodialysis patients
with BIA S10, BIA 720, and DXA (mean [+ or -] SD).

                                            DXA

Fat mass (kg)
  Male                           19.28 [+ or -] 7.30 (a)
  Female                         20.33 [+ or -] 8.94 (a)
  Total                          19.32 [+ or -] 7.94 (a)
Percentage of fat mass (%)
  Male                           27.24 [+ or -] 6.48 (a)
  Female                         34.11 [+ or -] 8.70 (a)
  Total                          29.63 [+ or -] 7.98 (a)
Fat-free mass (kg)
  Male                           47.56 [+ or -] 5.88
  Female                         36.66 [+ or -] 5.01
  Total                          43.77 [+ or -] 7.63 (c,d)
Fat mass index (kg/[m.sup.2])
  Male                            6.83 [+ or -] 2.80 (a)
  Female                          8.37 [+ or -] 3.91 (a)
  Total                           7.37 [+ or -] 3.29 (a)
Fat-free mass index (kg/[m.sup.2])
  Male                           17.18 [+ or -] 1.88
  Female                         15.10 [+ or -] 2.42
  Total                          16.46 [+ or -] 2.29 (e)

                                          BIA S10

Fat mass (kg)
  Male                           19.16 [+ or -] 8.29 (b)
  Female                         19.88 [+ or -] 11.32 (b)
  Total                          18.91 [+ or -] 9.36 (b)
Percentage of fat mass (%)
  Male                           27.53 [+ or -] 10.00 (b)
  Female                         33.68 [+ or -] 11.22 (b)
  Total                          29.67 [+ or -] 10.77 (b)
Fat-free mass (kg)
  Male                           46.46 [+ or -] 7.35 (a)
  Female                         35.35 [+ or -] 3.96 (a)
  Total                           42.6 [+ or -] 8.29 (a,c)
Fat mass index (kg/[m.sup.2])
  Male                            6.72 [+ or -] 3.35 (b)
  Female                          8.27 [+ or -] 4.75 (b)
  Total                           7.26 [+ or -] 3.93 (b)
Fat-free mass index (kg/[m.sup.2])
  Male                           16.73 [+ or -] 1.95 (a)
  Female                         14.50 [+ or -] 1.52 (a)
  Total                          15.95 [+ or -] 2.10 (a,e)

                                           BIA 720

Fat mass (kg)
  Male                           16.68 [+ or -] 9.07 (a,b)
  Female                         16.40 [+ or -] 11.28 (a,b)
  Total                          16.82 [+ or -] 9.46 (a,b)
Percentage of fat mass (%)
  Male                           23.68 [+ or -] 10.20 (a,b)
  Female                         29.30 [+ or -] 11.57 (a,b)
  Total                          25.49 [+ or -] 10.90 (a,b)
Fat-free mass (kg)
  Male                           48.88 [+ or -] 7.24 (a)
  Female                         37.11 [+ or -] 4.35 (a)
  Total                          45.08 [+ or -] 8.48 (a,d)
Fat mass index (kg/[m.sup.2])
  Male                            6.05 [+ or -] 3.45 (a,b)
  Female                          7.30 [+ or -] 4.77 (a,b)
  Total                           6.46 [+ or -] 3.93 (a,b)
Fat-free mass index (kg/[m.sup.2])
  Male                           17.63 [+ or -] 1.70 (a)
  Female                         15.19 [+ or -] 1.80 (a)
  Total                          16.84 [+ or -] 2.07 (a)

BIA = bioelectrical impedance analysis; DXA = dual-energy X-ray
absorp-tiometry; ((a,b)) significant difference between tools of
measurement at p < 0.001 in each line; (c) significant difference
between tools of measurement at p = 0.014 in each line; (d)
Significant difference between tools of measurement at p = 0.049 in
each line; (e) significant difference between tools of measurement at
p = 0.006 in each line.

Table 3: Mean differences and limits of agreement for fat mass,
percentage of fat mass, fat-free mass, fat mass index, and fat-free
mass index of the hemodialysis patients with BIA S10, BIA 720, and
DXA.

                                      InBody S10 versus DXA
Methods
                              Mean difference           95% limits
                                [+ or -] SD            of agreement
Fat mass (kg)
  Male                   -0.42 [+ or -] 3.56         -7.40 to 6.56
  Female                 -0.41 [+ or -] 3.67         -7.60 to 6.78
  Total                  -0.41 [+ or -] 3.57         -7.42 to 6.59
Percentage of fat mass (%)
  Male                    0.28 [+ or -] 5.24         -9.99 to 10.55
  Female                 -0.42 [+ or -] 6.93         -14.00 to 13.16
  Total                   0.04 [+ or -] 5.84         -11.41 to 11.48
Fat-free mass (kg)
  Male                   -1.10 [+ or -] 4.00         -8.94 to 6.74
  Female                 -1.32 [+ or -] 3.74         -8.65 to 6.01
  Total                  -1.17 [+ or -] 3.88         -8.79 to 6.44
Fat mass index (kg/[m.sup.2])
  Male                   -0.11 [+ or -] 1.28         -2.62 to 2.40
  Female                 -0.10 [+ or -] 1.52         -3.08 to 2.88
  Total                  -0.11 [+ or -] 1.36         -2.78 to 2.56
Fat-free mass index (kg/[m.sup.2])
  Male                   -0.46 [+ or -] 1.45         -3.30 to 2.38
  Female                 -0.60 [+ or -] 1.56         -3.66 to 2.46
  Total                  -0.51 [+ or -] 1.48         -3.41 to 2.39

                                      InBody 720 versus DXA
Methods
                              Mean difference           95% limits
                                [+ or -] SD            of agreement
Fat mass (kg)
  Male                   -2.57 [+ or -] 3.74         -9.90 to 4.76
  Female                 -2.70 [+ or -] 2.92         -8.42 to 3.02
  Total                   -2.6 [+ or -] 3.47         -9.41 to 4.19
Percentage of fat mass (%)
  Male                   -3.62 [+ or -] 5.24         -13.89 to 6.65
  Female                 -5.19 [+ or -] 5.74         -16.44 to 6.06
  Total                  -4.13 [+ or -] 5.41         -14.73 to 6.48
Fat-free mass (kg)
  Male                    0.98 [+ or -] 4.21         -7.27 to 9.23
  Female                  0.96 [+ or -] 2.93         -4.84 to 6.76
  Total                   0.98 [+ or -] 3.82         -6.51 to 8.46
Fat mass index (kg/[m.sup.2])
  Male                   -0.88 [+ or -] 1.30         -3.43 to 1.67
  Female                 -1.06 [+ or -] 1.16         -3.33 to 1.21
  Total                  -0.94 [+ or -] 1.25         -3.39 to 1.51
Fat-free mass index (kg/[m.sup.2])
  Male                    0.29 [+ or -] 1.49         -2.63 to 3.21
  Female                  0.35 [+ or -] 1.19         -1.98 to 3.52
  Total                   0.31 [+ or -] 1.39         -2.41 to 3.03

BIA = bioelectrical impedance analysis; DXA = dual-energy X-ray
absorptiometry.
COPYRIGHT 2018 Hindawi Limited
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2018 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Research Article
Author:Jayanama, Kulapong; Putadechakun, Supanee; Srisuwarn, Praopilad; Vallibhakara, Sakda Arj-Ong; Shanta
Publication:Journal of Nutrition and Metabolism
Date:Jan 1, 2018
Words:5081
Previous Article:Acute Consumption of Bordo Grape Juice and Wine Improves Serum Antioxidant Status in Healthy Individuals and Inhibits Reactive Oxygen Species...
Next Article:Efficacy of Diacetate Esters of Macular Carotenoids: Effect of Supplementation on Macular Pigment.
Topics:

Terms of use | Privacy policy | Copyright © 2020 Farlex, Inc. | Feedback | For webmasters