Diagnosing diabetes mellitus: performance of hemoglobin [A.sub.1c] point-of-care instruments in general practice offices.

An expert committee officially recommended the use of hemoglobin Hb [A.sub.1c] (Hb [A.sub.1c]) (4) for the diagnosis of diabetes in 2009 (1). There are several advantages to using Hb [A.sub.1c] rather than glucose measurements, including the preanalytical stability of Hb [A.sub.1c] in the sample and low within-subject biological variation of Hb [A.sub.1c] (1). The stability of Hb [A.sub.1c] throughout the day also obviates the fasting and dietary restrictions associated with glucose measurements. Thus, the use of Hb [A.sub.1c] as a diagnostic test likely will lead to an easier way to identify people with diabetes compared to the use of glucose (2). The WHO recommends an Hb [A.sub.1c] level of 6.5% (48 mmol/mol) as the cutoff point for diabetes (3). The assays must be "standardized to criteria aligned to the international reference values" (3). In addition, the American Diabetes Association (ADA) and the National Academy of Clinical Biochemistry (NACB) state that assays used for diagnosis should be National Glycohemoglobin Standardization Program (NGSP) certified (4, 5). The College of American Pathologists (CAP) recommends that the external quality assurance (EQA) acceptable limits for the accuracy compared to a target value should be 7% in 2012 and 6% in 2013 (6). Furthermore, the NACB recommends a withinlaboratory CV of <2% and abetween-laboratory CV of <3% for a single method (5). All CVs are based on the Diabetes Control and Complications Trial (DCCT)/ NGSP units (5). The ADA and WHO still recognize diagnostic criteria based on fasting plasma glucose measurements or the oral glucose tolerance test (3, 4), and each country should decide what diagnostic criteria they will use to diagnose diabetes (3). In Norway, the Norwegian Directorate of Health recommends that Hb [A.sub.1c] should be the primary diagnostic criterion (7, 8). NACB recommends using hospital laboratory Hb [A.sub.1c] instruments for the diagnosis of diabetes because point-of-care (POC) Hb A1C assays are "currently not sufficiently accurate for this purpose" (5). The quality of evidence for this recommendation is rated moderate, meaning that "further research is likely to have an important impact on the recommendation." The WHO allows the use of POC Hb [A.sub.1c] assays when this is the "only option available or when a stringent quality assurance program is in place" (3). It would be a great advantage at general practice (GP) offices to use POC instruments for this purpose and have the Hb [A.sub.1c] results during the consultation. All Hb [A.sub.1c] instruments used in Norway are traceable to the IFCC reference method (9). Furthermore, 99% of the Norwegian primary care offices and all hospital laboratories voluntarily participate in the same EQA programs through the Norwegian Quality Improvement of Primary Care Laboratories (Noklus) (www.Noklus.no). To our knowledge, no studies have investigated the performance of Hb [A.sub.1c] POC instruments based on results from EQA surveys (10).

In this study we used data from 13 EQA Hb [A.sub.1c] surveys from 2006 to 2012 and compared the results of POC instruments at GP offices with the analytical performance criteria for the diagnosis of diabetes, which are based on CAP (6), NACB (5) recommendations, and with the performance of hospital laboratory instruments.

Material and Methods

PARTICIPANTS

From 2006 to 2012, Noklus distributed 13 Hb[A.sub.1c]EQA surveys to all GP offices and hospital laboratories in Norway that analyze Hb [A.sub.1c]. In this longitudinal multicenter study, data from these EQA surveys were used. The POC instruments used at the GP offices were the Afinion (Axis-Shield PoC), DCA 2000 Analyzer, DCA 2000+ Analyzer, DCA VantageTM Analyzer (Siemens Healthcare Diagnostics), and Nycocard Hb [A.sub.1c] Reader (Axis-Shield PoC). All participants with the different DCA models were merged to one instrument group. The hospital laboratories participated with the Architect c8200, Architect ci8200 (Abbott), Advia 1200, Advia 1650 (Siemens), Bio-Rad D10 (Bio-Rad), Cobas 6000, Cobas Integra, Cobas Integra 400, Cobas Integra 400 Plus, Cobas Integra 800 (Roche Diagnostics), Dimensions RXL, Dimensions Vista 1500 (Siemens), Hitachi 902, Hitachi 912 (Roche Diagnostics), Modular P (Roche Diagnostics), Tosoh HPLC G7, Tosoh HPLC G8 (Tosoh Bioscience), Variant Classic HPLC, and Variant II HPLC, (Bio-Rad Laboratories). The hospital laboratory instruments were merged into a hospital laboratory group because there were only 1-9 participants with each instrument in each survey. The number of participants in the 13 Hb [A.sub.1c] EQA surveys is shown in Fig. 1. In this report we use the term "laboratory" for both GP offices and hospital laboratories.

EQA MATERIAL

Venous blood was collected into tubes containing EDTA. Samples were obtained from healthy volunteers (with Hb [A.sub.1c] levels within reference intervals) and from patients with diabetes (Hb [A.sub.1c] values above the upper limit of the reference interval) and mixed gently. Blood from a minimum of 2 donors with Hb [A.sub.1c]at each level were distributed fresh in cryovials (Sarstedt) within 12 h after sampling. The stability of the EQA material was tested at the Laboratory of Clinical Biochemistry (Haukeland University Hospital, Bergen, Norway) based on ISO 13528 (11). Two cryovials at each Hb [A.sub.1c] level were kept first at room temperature, and then 1 vial was frozen 1 day after the collection of the blood. The remaining vials were kept at 4 [degrees]C from day 2 until they were frozen after 4 days. The EQA materials were thawed and the Hb [A.sub.1c] analyzed at the same time on the Variant II HPLC (Bio-Rad Laboratories) in 6 replicates. The Hb [A.sub.1c] concentration was stable in this period. Because the stability of the EQA material can be method dependent, the results from the different POC instruments analyzed the first, second, third, or fourth day after the participants received the EQA material were examined. There were no differences in the median values, indicating that the EQA material was stable for this time period (data not shown). Furthermore, it has been found that Hb [A.sub.1c] values in EDTA whole blood samples were stable at both room temperature and 4[degrees]C for more than 4 days for measurement using the Siemens DCA2000 +, Tosoh G7 and G8, Bio-Rad Variant IINU, and Trinity Biotech ultra2 (12).

TARGET VALUE

The target values for the EQA materials were established at the European Reference Laboratory for Glycohemoglobin (the Netherlands). Values were assigned in triplicate with 2 IFCC secondary reference measurement procedures that were IFCC calibrated (13). Method 1 is HA 8160, HPLC, ionic exchange (Menarini) and method 2 is PDQ, HPLC, boronat affinity Principal (Primus/Trinity Biotech). The results were given in both % Hb [A.sub.1c] (DCCT/NGSP unit) and mmol/mol Hb [A.sub.1c] (IFCC unit). In Norway the DCCT/ NGSP unit is used. The target value is the mean from triplicate measurements from the 2 methods (see Table 1 in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/ content/vol59/issue12). The master equation from Hoelzel et al. (14) was used for calculations from millimoles per mole Hb [A.sub.1c] (IFCC unit) to % Hb [A.sub.1c] (DCCT/NGSP unit).

EQA SURVEY

The GP offices and hospital laboratories received 500 IjL of the fresh EQA material in each level, with an Hb [A.sub.1c] level from 5.0% to 5.8% Hb [A.sub.1c] (30.9-39.6 mmol/ mol Hb [A.sub.1c]) in the level within the reference interval and from 6.8% to 7.9% Hb [A.sub.1c] (51.3-62.6 mmol/mol Hb [A.sub.1c]) for the level above the upper limit of the reference interval in the 13 surveys (see online Supplemental Table 1). The samples were shipped without temperature control, and the participants kept the EQA materials at 4[degrees]C until analysis. Participants analyzed each EQA material on 2 different days (1 x 2) within a 4-day period. In the response forms, the participants were asked to return the analytical results with information about the instrument used and the lot number of the reagent as well as the date of analysis. The scheme was performed twice a year.

STATISTICS

All calculations were performed using the % Hb [A.sub.1c] (DCCT/NGSP) unit. The participant's performance was evaluated with respect to trueness based on CAP limits for accuracy (6). An estimate of the trueness was defined here as the degree of closeness of the mean of the participant duplicate measurements and the target value. To obtain a measure of the trueness of a method, the mean Hb [A.sub.1c] was calculated from the mean of duplicate measurements from each participant for each instrument group at the 2 Hb [A.sub.1c] levels, and the percentage deviation from the target value was calculated. The deviation of results from the target value was evaluated against the criteria of 5.0%, 6.0%, and 7.0% (6). Whereas the CAP deals with quality specifications for 1 result from a target value, we used the deviation of the mean of duplicate results for 2 different samples. This makes our quality specifications a little bit wider than the CAP specifications, but it is then an estimate of the trueness.

Because it was not possible to obtain precision measures in the form of CVs from internal QC for all the participants, we used the difference between duplicate measurements as estimates of the precision. To be able to differentiate between the upper limit of the reference interval (6.0% Hb [A.sub.1c]) and the cutoff 6.5% Hb [A.sub.1c] for diagnosis of diabetes, the difference between duplicates must be between 0.2% and 0.3% Hb [A.sub.1c] (15). This corresponds to the NACB limit for imprecision of <2% (5). These differences between duplicates were therefore evaluated against the quality specifications of 0.2%, 0.3%, and 0.4% Hb [A.sub.1c] (5, 15). In addition, CVs were calculated using duplicate measurements from the last 6 surveys that each GP office and hospital laboratory participated in, using the equation for [CV.sub.within] given below We excluded outliers before calculating a pooled within-laboratory analytical imprecision and between-laboratory imprecision. Values more than 3 SDs from the mean difference between duplicates and values more than 3 SDs from the mean of means were excluded. Percentages of participants that were excluded in each survey were [mean (95% CI)] 1.5% (0.9%-2.2%), 1.3% (1.0%-1.6%), 2.6% (1.5%-3.6%), and 2.6% (1.6%-3.5%) for the Afinion, DCA, Nycocard, and hospital laboratory instruments, respectively.

The median of the differences was calculated, and the pooled within-laboratory analytical imprecision ([CV.sub.within]) was calculated for each Hb [A.sub.1c] EQA survey and each POC instrument group and the hospital laboratory group in the 2 Hb [A.sub.1c] levels, using the equation:

[CV.sub.within] (%) = [square root of ([summation][diff.sup.2]/2n)]/[bar.x] x 100% (1)

where "diff" is the difference between duplicates from each participant, "n" is the number of differences, and x is the median of the means.

The variation between laboratories based on duplicate measurements (CV inter) was calculated for each Hb [A.sub.1c] EQA survey and each POC instrument group in the 2 Hb [A.sub.1c] levels after exclusion of outliers, using the equation:

[CV.sub.inter] (%) = [summation ove ([summation][([[bar.x].sub.i] - [[bar.x].sub.mean dup]).sup.2])/n - 1]/[bar.x] x 100 (2)

where [[bar.x].sub.i] is the mean of duplicates from each participant, [[bar.x].sub.mean dup] is the mean of duplicates for all participants, "n" is the number of participants, and X is the median of the means.

Because the participants measured Hb [A.sub.1c] in duplicate, the between-laboratory CV was calculated using the following equation (corresponding to 1 measurement):

[CV.sub.between laboratory] (%) = [square root of ([([CV.sub.inter]).sup.2] + [([CV.sub.within]).sup.2]/2)] (3)

Computations were performed using SPSS version 20.0 and Excel version 14.2.4.

Results

The total number of GP participants in the EQA surveys for Hb [A.sub.1c] increased from 1034 (59% of the total of 1758 GP offices in Norway) in the 2006-2 survey to 1288 (75% of the total of 1727 GP offices in Norway) in the 2012-2 survey (Fig. 1). All GPs performing Hb [A.sub.1c] analyses in Norway participated in the EQA survey. The number of participants using the Afinion instrument increased the most, whereas the Nycocard seemed to disappear from the market. In this period Noklus advised GPs to use either the DCA or the Afinion. The number of hospital laboratory participants using laboratory instruments increased from 23 in the 2006-2 survey to 52 in the 2012-2 survey (Fig. 1). In each survey, between 75% and 96% of the participants returned duplicate results for Hb [A.sub.1c] values both within the reference interval and above the upper limit of the reference interval in each survey and are included in the calculations.

SYSTEMATIC DEVIATION FROM TARGET VALUE

Between 72% and 96% of the GPs using the Afinion had a systematic deviation of the results from the target value of [less than or equal to] 6.0% for Hb [A.sub.1c] values both within the reference interval and above the upper limit of the reference interval (Table 1 and Fig. 2). The corresponding values for DCA users were between 77% and 96%, for Nycocard users between 27% and 59%, and for hospital laboratory participants between 54% and 84% (Table 1 and Fig. 2). The mean deviation from target value varied between -0.29% and 0.23% Hb [A.sub.1c], -0.30% and 0.28% Hb [A.sub.1c], -0.20% and 0.67% Hb [A.sub.1c], and 0.03% and 0.63% Hb [A.sub.1c] for the Afinion, DCA, Nycocard, and hospital laboratory instruments, respectively (Fig. 2). The systematic deviation did not change significantly during the observed period (Fig. 2), and the laboratories did not have a systematic high or low bias (data not shown). The percentage of participants that were within <6.0% in deviation from target value for Hb [A.sub.1c] values both within the reference interval and above the upper limit of the reference interval in 1-6 of the last 6 surveys they participated in is given in Fig. 3A. For Afinion, DCA, and Nycocard users and hospital laboratories, 37%, 58%, 1%, and 27%, respectively, were within [less than or equal to] 6.0% in all 6 surveys (Fig. 3A).

WITHIN- AND BETWEEN-LABORATORY VARIATION

The percentage of participants for which the absolute difference between duplicate Hb [A.sub.1c] measurements was [less than or equal to] 0.3% Hb [A.sub.1c] is given in Table 1. The absolute difference between the duplicate results did not change significantly during the observed period (see online Supplemental Fig. 1). Of the participants with Afinion, DCA, Nycocard, and hospital laboratory instruments, 27%, 36%, 10%, and 78%, respectively, obtained a deviation between duplicate Hb [A.sub.1c] measurement [less than or equal to] 0.3% Hb [A.sub.1c] for Hb [A.sub.1c] values both within the reference interval and above the upper limit of the reference interval in the last 6 Hb [A.sub.1c] surveys they participated in (Fig. 3B).

Twenty-four percent, 37%, 9%, and 82% of the Afinion, DCA, and Nycocard users and hospital laboratories, respectively, had a CV [less than or equal to] 2.0% for Hb [A.sub.1c] values both within the reference interval and above the upper limit of the reference interval calculated from duplicate measurements from the last 6 surveys they participated in (see online Supplemental Table 2). If the requirement is set to [less than or equal to] 2%, as given in Sacks et al. (5) (meaning that participants with CV <2.5% fulfill the recommendation), the corresponding percentages is increased to 51%, 63%, 11%, and 93%, respectively (see online Supplemental Table 2).

The pooled within-laboratory variations of the Afinion, DCA, and hospital laboratory instruments were below the recommended limit of 2% (5) in the majority of the surveys (Fig. 4). Nycocard did not meet the requirement for within-laboratory variation in any of the EQA surveys (Fig. 4).

The Afinion and DCA instruments were also below the recommended limit of 3% for within-method between-laboratory variations (5) in the majority of the surveys (see online Supplemental Fig. 2), whereas the Nycocard instrument did not meet the 3% recommendation in any of the surveys (see online Supplemental Fig. 2).

COMBINED TRUENESS AND IMPRECISION

Thirty-eight percent, 60%, 2%, and 45% of the Afinion, DCA, and Nycocard users and hospital laboratories, respectively, were within both the 6.0% limit for trueness and the 0.3% Hb [A.sub.1c] limit for imprecision in the last 5 and 6 surveys they participated in (Fig. 3C). Corresponding numbers for other quality specifications are also given in Table 2. Use of the 7.0% limit for trueness from 2012 (6) led to a slight increase in the percentage of participants passing the limits. However, if the limits for trueness and imprecision were reduced to 5.0% and 0.2% Hb [A.sub.1c], respectively, a very small portion of the participants were able to fulfill the requirements over time (Table 2).

Discussion

The NACB recommends using only hospital laboratory Hb [A.sub.1c] instruments for the diagnosis of diabetes, mainly because there is little information about the analytical quality of POC Hb [A.sub.1c] assays in the clinical setting (5). The main finding in the present study is that about 60%-90% of GPs using Afinion or DCA instruments performed measurements that met the quality specifications for both trueness ([less than or equal to] 6.0%) and imprecision ([less than or equal to] 0.3% Hb [A.sub.1c]) for diagnosing diabetes mellitus in each EQA survey (Table 1). This performance was similar to that of hospital laboratories (Table 1). Adding the 2 last columns in Fig. 3C shows that 60% of GPs using DCA instruments were within the quality specifications in 5 or 6 of the 6 last surveys they participated in, whereas the corresponding numbers for the Afinion users and hospital laboratories were 38% and 45%, respectively (Fig. 3C). These numbers are close to what can be expected from the numbers given in Table 1, which highlights the finding that the group of GP offices using one type of instrument is rather homogenous and that, for example, DCA users have about 90% probability of passing the quality specifications in each survey. The percentage of laboratories that perform within acceptable quality specifications will of course vary depending on the criteria given (Table 2). For instance, the percentage of laboratories passing the quality specifications for trueness in the last 6 years of participation increased by 5%-10% when the limit for trueness was expanded to 7.0%, which was the CAP limit in 2011 and 2012 (6). Furthermore, DCA and also to some degree Afinion users more often fulfilled the quality specification for trueness than the specification for precision, whereas the opposite was true for hospital laboratories (Fig. 3, A and B, and Tables 1 and 2). It is possible, however, that the overall quality of the trueness of the hospital laboratory methods is not as good as that for most of those laboratories included in the CAP surveys, of which more than 90% had an absolute deviation from the target value of [less than or equal to] 6.0% (16). The within-method between-participant CVs, including those for between-lot variations, were less than the 3% limit recommended by the NACB (5) for Afinion in the majority of the surveys and in all 13 of the surveys for DCA (see online Supplemental Fig. 2).

The NACB states that the quality specifications for within-lot variation should be <2% (5). Our results show that this quality specification was fulfilled for most of the users of Afinion, DCA, and hospital instruments on a group basis in each survey (Fig. 4) and for about 50%-90% of the participants when the precision was calculated from the 6 last surveys of participation for each laboratory (see online Supplemental Table 2). In addition, we used the difference between the 2 duplicates ([less than or equal to] 0.3% Hb [A.sub.1c] units) as a measure of the precision on a group basis in each survey (Table 1) and for each individual laboratory (Fig. 3B), and these results are in agreement with the results using pooled (Fig. 4) and individual CVs (see online Supplemental Table 2), respectively. Thus, different measures for precision show that at least 70% of Afinion or DCA users or hospital laboratories fulfilled the precision criteria for being able to diagnose diabetes in each survey (Table 1). Our findings are similar to results with Afinion and DCA instruments in the hands of trained personnel (17, 18). These studies, however, were laboratory experiments following the CLSI EP-5 guidelines (19) and not longitudinal results from clinical practice like the present study.

The strength of the present study is that it includes a large number of GP offices and that it follows their performance in numerous EQA surveys over many years. In addition, in the present study the participants have been evaluated regarding both trueness and imprecision simultaneously, and their performance in clinical practice has been evaluated against criteria based on the current CAP and NACB quality specifications for Hb [A.sub.1c]. The EQA material used is fresh native blood and is similar for all types of instruments. The target values are set by 2 reference methods.

A limitation of the present study is that the estimates of trueness and imprecision might be overly optimistic because they were calculated from duplicates over a short period of time and there is a possibility that the participants might have reanalyzed a "strange" result and removed the results that did not fit. However, we think that the results in general are representative for the practices and that it is not worthwhile to increase the number of replicates in the EQA scheme. Furthermore, the number of hospital laboratories is small and does not permit comparisons of performances between methods. The performance of the overall group of hospital laboratories is therefore dependent on the composition of methods used. However, the aim with the present study was not to compare specifically the performances of different hospital laboratory methods with POC methods in GP offices but to see if the performance of methods used in general practices was as good as the "average" hospital method in Norway. But our findings cannot be generalized to the quality of hospital laboratory instruments in other countries. Nevertheless, the pooled [CV.sub.within] results for hospital laboratories (Fig. 4) are similar to those of other studies performed in one laboratory with experienced users (20-24).

The performance of the participants in this study has been evaluated using the quality specifications from CAP (6) and NACB (5), which are given in DCCT units. It is important to underline that the use of IFCC units will give different quality specifications when calculated in percentages (25).

A presupposition for high-quality performance of Hb [A.sub.1c] measurements in GP offices is the existence of a stringent quality assurance program (QAP). The WHO has not defined what elements should be included in a stringent QAP. In our opinion, EQAs are an important part of this, and in the Noklus EQA schemes for Hb [A.sub.1c] participants receive EQA materials on a regular basis, the EQA material are commutable (fresh native human blood) at 2 different levels, and a secondary reference laboratory has set the target value. Equally important is that all GP offices should receive regular guidance in the form of visits and courses from laboratory consultants (e.g., biomedical laboratory scientists). When the participants obtain an evaluation of their EQA result outside the "acceptability" limits, the consultants must provide them with advice and instruction regarding their measurement procedures, and they can also recommend what instruments to use. Such recommendations should be based on reports from organizations that perform supplier-independent evaluations of the analytical quality and user friendliness of POC equipment, for example, SKUP (the Scandinavian evaluation of laboratory equipment for primary healthcare) (26). Noklus runs the EQA surveys for both hospital laboratories and GP offices in Norway, and from the autumn of 2012, the within-laboratory day-to-day variation has been recorded for all participants, and the number of EQA surveys will increase to 4 per year. In the feedback reports it will be stated if the results of the laboratories satisfy the quality specifications to diagnose diabetes mellitus, and help will be offered to GP offices if requested. Participation in the Noklus EQA programs is voluntary, and because 99% of Norwegian GP offices participate, the participants are representative for POC users. We think that our results are generalizable to other countries or areas where GPs analyzing Hb [A.sub.1c] participate in a stringent QAP.

In conclusion, a large proportion of GP offices in Norway using Afinion and DCA for measurements of Hb [A.sub.1c] fulfill the quality specifications based on the CAP and NACB requirements for trueness and precision. Furthermore, their analytical quality is comparable to that of hospital laboratory instruments.

About 75% of GP offices in Norway have Hb [A.sub.1c] POC instruments, thus the instruments are easily available. Considering the large number of persons with undiagnosed type 2 diabetes (5), it is a great advantage to have the possibility of using POC instruments at GP offices for this purpose and to have the results during the consultation. Since a national stringent QAP is established that closely monitors the laboratory performance in GP offices in Norway (3), it is therefore recommended that GP offices that participate in this QAP and fulfill the quality specifications can use their POC instruments to diagnose diabetes mellitus.

Author Contributions: All authors confirmed they have contributed to the intellectual content of this paper and have met the following 3 requirements: (a) significant contributions to the conception and design, acquisition of data, oranalysis and interpretation of data; (b) drafting or revising the article for intellectual content; and (c) final approval of the published article.

Authors' Disclosures or Potential Conflicts of Interest: No authors declared any potential conflicts of interest.

Role of Sponsor: No sponsor was declared.

Acknowledgments: The authors thank Per Hyltoft Petersen and Gunn Kristensen for valuable comments on the manuscript and the biomedical laboratory scientists at Noklus conducting the external quality assessment surveys.

References

(1.) The International Expert Committee. International expert committee report on the role of the A1C assay in the diagnosis of diabetes. Diabetes Care 2009; 32:1327-34.

(2.) Saudek CD, Herman WH, Sacks DB, Bergenstal RM, Edelman D, Davidson MB. A new look at screening and diagnosing diabetes mellitus. J Clin Endocrinol Metab 2008; 93:2447-53.

(3.) WHO. Use of glycated haemoglobin (Hb[A.sub.1c]) in the diagnosis of diabetes mellitus. Abbreviated report of a WHO consultation. WHO/NMH/CHP/ CPM/11.1 ed. Geneva: WHO; 2011.

(4.) American Diabetes Association. Diagnosis and classification of diabetes mellitus. Diabetes Care 2010; 33(Suppl 1):S62-9.

(5.) Sacks DB, Arnold M, BakrisGL, Bruns DE, Horvath AR, Kirkman MS, et al. Guidelines and recommendations for laboratory analysis in the diagnosis and management of diabetes mellitus. Clin Chem 2011; 57:e1-47.

(6.) NGSP. NGSP news. http://www.ngsp.org/news. asp (Accessed June 2013).

(7.) Berg JP, Julsrud Berg T, Bjerve KS, Claudi T, Dahl-Jorgensen K, Fougner KJ, et al. Rapport fra arbeidsgruppe for vurdering av diagnostisk bruk av Hb[A.sub.1c] ved diabetes. http://legeforeningen.no/ PageFiles/104318/Hb[A.sub.1c]%20til%20diagnostikk% 20av%20diabetes.pdf (Accessed June 2013).

(8.) Helsedirektoratet. Hb[A.sub.1c] som diagnostikum for diabetes. http://helsedirektoratet.no/Om/nyheter/ Sider/helsedirektoratet-anbefaler-blodprove-for-a finne-pasienter-med-diabetes-.aspx (Accessed June 2013).

(9.) Jeppsson JO, Kobold U, Barr J, Finke A, Hoelzel W, Hoshino T, et al. Approved IFCC reference method for the measurement of Hb[A.sub.1c] in human blood. Clin Chem Lab Med 2002; 40:78-89.

(10.) Little RR. Analysis of the accuracy and precision of the Axis-Shield Afinion hemoglobin [A.sub.1c] measurement device. J Diabetes Sci Technol 2012; 6: 387-8.

(11.) International Organization for Standardization. Statistical methods for use in proficiency testing by interlaboratory comparisons. ISO 13528. Geneva: International ISO; 2005.

(12.) Rohlfing CL, Hanson S, Tennill AL, Little RR. Effects of whole blood storage on hemoglobin a1c measurements with five current assay methods. Diabetes Technol Ther 2012; 14:271-5.

(13.) IFCC. Approved laboratories of the IFCC Network Laboratories for Hb[A.sub.1c]. http://www.ifcchba1c.net/ network/approved?page=1&order=field_country_ app&sort=asc (Accessed July 2013).

(14.) Hoelzel W, Weykamp C, Jeppsson JO, Miedema K, Barr JR, Goodall I, et al. IFCC reference system for measurement of hemoglobin [A.sub.1c] in human blood and the National standardization schemes in the United States, Japan, and Sweden: a methodcomparison study. Clin Chem 2004; 50:166-74.

(15.) Goodall I, Colman PG, Schneider HG, McLean M, Barker G. Desirable performance standards for HbA(1c) analysis--precision, accuracy and standardisation: consensus statement of the Australasian Association of Clinical Biochemists (AACB), the Australian Diabetes Society (ADS), the Royal College of Pathologists of Australasia (RCPA), Endocrine Society of Australia (ESA), and the Australian Diabetes Educators Association (ADEA). Clin Chem Lab Med 2007; 45:1083-97.

(16.) NGSP. College of American pathologists (CAP) survey data http://www.ngsp.org/CAPdata.asp (Accessed June 2013).

(17.) Lenters-Westra E, Slingerland RJ. Six of eight hemoglobin [A.sub.1c] point-of-care instruments do not meet the general accepted analytical performance criteria. Clin Chem 2010; 56:44-52.

(18.) Little RR, Lenters-Westra E, Rohlfing CL, Slingerland R. Point-of-care assays for hemoglobin A(1c): is performance adequate? Clin Chem 2011; 57:1333-4.

(19.) CLSI. Evaluation of precision performance of clinical chemistry devices: approved guideline. NCCLS/CLSI document EP5-A. Wayne (PA): CLSI; 1992.

(20.) John WG, Braconnier F, Miedema K, Aulesa C, Piras G. Evaluation of the Menarini-Arkray HA 8140 hemoglobin Au analyzer. Clin Chem 1997; 43:968-75.

(21.) Higgins TN, Blakney GB, Dayton J. Analytical evaluation of the Bio-Rad variant II automated HbA(1C) analyzer. Clin Biochem 2001; 34:361-5.

(22.) Moridani MY, Verjee Z, Allen LC. Analytical evaluation of hemoglobin A(1c) dual kit assay on Bio-Rad Variant II: an automated HPLC hemoglo bin analyzer for the management of diabetic patients. Clin Biochem 2003; 36:317-20.

(23.) Chapelle JP, Teixeira J, Maisin D, Assink H, Barla G, Stroobants AK, et al. Multicentre evaluation of the Tosoh Hb[A.sub.1c] G8 analyser. Clin Chem Lab Med 2010; 48:365-71.

(24.) Meurice J, Guillard E, Jaisson S, Leroy N, Gillery P. [Evaluation of the new kit HbA(1c) Analyzer 2.0 Variant II Turbo (Bio-Rad).] Ann Biol Clin 2011; 69:561-9. [French]

(25.) Weykamp CW, Mosca A, Gillery P, Panteghini M. The analytical goals for hemoglobin A(1c) measurement in IFCC units and National Glycohemoglobin Standardization Program Units are different. Clin Chem 2011; 57:1204-6.

26. SKUP. Scandinavian evaluation of laboratory equipment for primary health care. http:// www.skup.nu (Accessed June 2013).

Una Orvim Solvik, [1] * Thomas Roraas, [2] Nina Gade Christensen, [2] and Sverre Sandberg [1,2,3]

[1] Department of Global Health and Primary Care, Faculty of Medicine and Dentistry, University of Bergen, Norway; [2] The Norwegian Quality Improvement of Primary Care Laboratories (Noklus), Haraldsplass Deaconess Hospital, Bergen, Norway; [3] Laboratory of Clinical Biochemistry, Haukeland University Hospital, Bergen, Norway.

[4] Nonstandard abbreviations: Hb [A.sub.1c], hemoglobin [A.sub.1c]; ADA, American Diabetes Association; NACB, National Academy of Clinical Biochemistry; NGSP, National Glycohemoglobin Standardization Program; CAP, College of American Pathologists; EQA, External Quality Assurance; DCCT, Diabetes Control and Complications Trial; POC, point of care; GP, general practice; Noklus, Norwegian Quality Improvement of Primary Care Laboratories; [CV.sub.within], pooled within-laboratory analytical imprecision; QAP, quality assurance program.

* Address correspondence to this author at: Noklus, P.O. Box 6165, N-5892 Bergen, Norway. Fax +47-55-97-95-10; e-mail una.solvik@isf.uib.no.

Received June 4, 2013; accepted July 30, 2013.

Previously published online at DOI: 10.1373/clinchem.2013.210781

Table 1. Percentage of participants within the quality specifications
for trueness and imprecision for Hb [A.sub.1c] levels within and
above reference intervals in the 13 Hb [A.sub.1c] EQA surveys from
2006 to 2012.

                                          Survey

Quality specification   2006-2   2007-1   2007-2   2008-1   2008-2
and instrument

Trueness (a)
  Afinion                 80       84       73       73       77
  DCA                     77       91       89       90       84
  Nycocard                42       37       58       27       48
  Hospital laboratory     84       77       67       56       64
    instruments
Imprecision (b)
  Afinion                 85       94       88       94       95
  DCA                     89       91       91       87       84
  Nycocard                48       49       55       56       56
  Hospital laboratory     95       86       94       88       95
    instruments
Trueness and
imprecision (c)
  Afinion                 70       79       71       72       73
  DCA                     68       83       82       81       72
  Nycocard                21       19       34       16       30
  Hospital laboratory     84       64       61       56       62
    instruments

                                          Survey

Quality specification   2009-1   2009-2   2010-1   2010-2   2011-1
and instrument

Trueness (a)
  Afinion                 89       77       96       72       85
  DCA                     84       93       94       79       91
  Nycocard                41       53       49       46       32
  Hospital laboratory     54       82       81       62       71
    instruments
Imprecision (b)
  Afinion                 96       92       88       90       85
  DCA                     93       90       80       87       90
  Nycocard                54       61       53       63       44
  Hospital laboratory    100       98       95       98       96
    instruments
Trueness and
imprecision (c)
  Afinion                 87       72       84       65       72
  DCA                     79       84       77       69       83
  Nycocard                22       34       30       34       12
  Hospital laboratory     54       80       81       62       71
    instruments

                                 Survey

Quality specification   2011-2   2012-1   2012-2
and instrument

Trueness (a)
  Afinion                 73       82       91
  DCA                     95       95       96
  Nycocard                59       54       57
  Hospital laboratory     69       81       79
    instruments
Imprecision (b)
  Afinion                 80       74       82
  DCA                     86       86       90
  Nycocard                55       77       57
  Hospital laboratory     98      100       88
    instruments
Trueness and
imprecision (c)
  Afinion                 59       61       77
  DCA                     83       83       88
  Nycocard                27       38       43
  Hospital laboratory     67       81       67
    instruments

(a) The absolute deviation from target value [less than or equal to]
6.0%.

(b) The absolute difference between duplicate Hb Au measurements
[less than or equal to] 0.3% Hb [A.sub.1c].

(c) Within the limits for both trueness ([less than or equal to]
6.0%) and imprecision ([less than or equal to] 0.3% Hb [A.sub.1c]).

Table 2. Percentage of participants that were within the limits for
both trueness ([less than or equal to] 5.0%, [less than or equal to]
6.0%, and [less than or equal to] 7.0%) and imprecision ([less than
or equal to] 0.2%, [less than or equal to] 0.3%, and [less than or
equal to] 0.4% Hb [A.sub.1c]) for Hb [A.sub.1c] levels within and
above reference intervals in the last 4 and 6 surveys they
participated in.

                                               Trueness

                                             Last 4 surveys

Instrument and imprecision        [less than or    [less than or
                                  equal to] 5.0%   equal to] 6.0%
Afinion
  [less than or equal to] 0.2%    1.1              1.8
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    11.0             18.6
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    17.3             28.7
    Hb [A.sub.1c]
DCA
  [less than or equal to] 0.2%    3.5              4.1
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    31.8             40.7
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    39.8             51.2
    Hb [A.sub.1c]
Nycocard
  [less than or equal to] 0.2%    0.0              1.1
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    0.6              1.1
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    0.6              1.1
    Hb [A.sub.1c]
Hospital laboratory instruments
  [less than or equal to] 0.2%    8.2              18.4
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    14.3             36.7
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    14.3             36.7
    Hb [A.sub.1c]

                                            Trueness

                                         Last 4 surveys

Instrument and imprecision        [less than or    n (a)
                                  equal to] 7.0%
Afinion                                            571
  [less than or equal to] 0.2%    2.1
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    25.0
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    37.7
    Hb [A.sub.1c]
DCA                                                691
  [less than or equal to] 0.2%    4.9
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    47.9
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    60.8
    Hb [A.sub.1c]
Nycocard                                           180
  [less than or equal to] 0.2%    1.1
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    1.1
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    2.2
    Hb [A.sub.1c]
Hospital laboratory instruments                    49
  [less than or equal to] 0.2%    22.4
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    46.9
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    49.0
    Hb [A.sub.1c]

                                               Trueness

                                           Last 6 surveys

Instrument and imprecision        [less than or    [less than or
                                  equal to] 5.0%   equal to] 6.0%
Afinion
  [less than or equal to] 0.2%    0.7              1.1
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    4.5              10.0
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    8.2              17.3
    Hb [A.sub.1c]
DCA
  [less than or equal to] 0.2%    1.2              1.5
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    16.0             22.7
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    24.2             34.0
    Hb [A.sub.1c]
Nycocard
  [less than or equal to] 0.2%    0.0              0.0
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    1.1              1.1
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    1.1              1.1
    Hb [A.sub.1c]
Hospital laboratory instruments
  [less than or equal to] 0.2%    0.0              13.3
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    2.2              26.7
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    2.2              26.7
    Hb [A.sub.1c]

                                            Trueness

                                        Last 6 surveys

Instrument and imprecision        [less than or    n (a)
                                  equal to] 7.0%
Afinion                                            440
  [less than or equal to] 0.2%    1.4
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    16.4
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    26.6
    Hb [A.sub.1c]
DCA                                                582
  [less than or equal to] 0.2%    1.9
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    30.9
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    45.0
    Hb [A.sub.1c]
Nycocard                                           87
  [less than or equal to] 0.2%    0.0
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    1.1
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    1.1
    Hb [A.sub.1c]
Hospital laboratory instruments                    45
  [less than or equal to] 0.2%    15.6
    Hb [A.sub.1c]
  [less than or equal to] 0.3%    33.3
    Hb [A.sub.1c]
  [less than or equal to] 0.4%    35.6
    Hb [A.sub.1c]

(a) A laboratory may have participated with 2 different instruments
during the time period and thus participated 4 and 6 times in 2
different time periods. The total numbers may therefore be higher
than those given in Fig. 1.
COPYRIGHT 2013 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2013 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Point-of-Care Testing
Author:Solvik, Una Orvim; Roraas, Thomas; Christensen, Nina Gade; Sandberg, Sverre
Publication:Clinical Chemistry
Article Type:Report
Geographic Code:4EXNO
Date:Dec 1, 2013
Words:6644