Diabetes Technol Ther. Validation of the Continuous Glucose Monitoring System (CGMS) by the use of two CGMS simultaneously in pregnant women with type 1 diabetes mellitus.

In pregnant women with type 1 diabetes tight glycemic control reduces perinatal complications. Intensive observation of glucose profiles is essential in the achievement of tight glycemic control. The recent availability of the Continuous Glucose Monitoring System (CGMS, Minimed, Sylmar, CA) creates the opportunity to obtain more complete glucose profiles. This study was aimed at evaluating the accuracy of the CGMS in pregnant women with type 1 diabetes.

Five pregnant women with type 1 diabetes were asked to use two CGMS devices simultaneously. The simultaneously measured glucose levels were analyzed using the Pearson correlation, the mean absolute difference, and Bland-Altman analysis. Second, the percentage of concordance of paired data in the hypoglycemic, normoglycemic, or hyperglycemic range was calculated. RESULTS: The correlation coefficient between simultaneously measured data was 0.94 (P < 0.001). The mean absolute difference was 1.1 +/- 0.8 mmol/L. Bland-Altman analysis shows that 95% of the data pairs have a difference < or = 1.74 mmol/L. Almost 80% of the data pairs could be classified in the same glucose range. In 81% of the non-concordant pairs, one glucose value was classified in the hypoglycemic range and one in the normoglycemic range.

This study shows that the reproducibility of the CGMS in pregnant women with type 1 diabetes is adequate. This indicates that the CGMS is a useful tool in the management of type 1 diabetes in pregnant women. However, the CGMS should only be used as a supplementary method of daily glucose level measurement as a small degree of error, mainly in the hypoglycemic range, is present.

2005 Oct;7(5):699-706

COPYRIGHT 2007 American Medical Technologists
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2007 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:ABSTRACTS FROM THE CURRENT LITERATURE
Author:Kerssen, A.; De Valk, H.W.; Visser, G.H.
Publication:Journal of Continuing Education Topics & Issues
Article Type:Brief article
Geographic Code:1USA
Date:Jan 1, 2007
Words:265
Previous Article:Diabetes Care. Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial.
Next Article:Breast Cancer Res Treat. Cigarette smoking and breast cancer risk: update of a prospective cohort study.
Topics:

Terms of use | Privacy policy | Copyright © 2021 Farlex, Inc. | Feedback | For webmasters