Development and validation of a virtual reality transrectal ultrasound guided prostatic biopsy simulator.

Author(s): Venu Chalasani, MD [1], Derek W. Cool, MD [2], Shi Sherebrin, MD [2], Aaron Fenster, MD [2], Joseph Chin, MD, FRCSC [1], Jonathan I Izawa, MD, FRCSC [1]

Introduction

The current gold standard and the most widely used method to diagnose prostate cancer is transrectal ultrasound (TRUS)- guided prostatic biopsy, which is commonly performed in Westernized high income countries. For instance, in Australia in 2008 a total of 26 146 TRUS guided biopsies of the prostate were performed.[sup.1]

Transrectal ultrasound-guided prostatic biopsy does have well-documented complications. A recently published Japanese nationwide survey of more than 200 000 prostatic biopsies revealed that complications occurred frequently, with hematuria in 12%, hematochezia in 5.9%, hematospermia in 1.2%, fever in 1.1%, sepsis in 0.07%, voiding symptoms in 1.9%, urinary retention in 1.1% and prostatitis in 0.9%. Hospitalization to treat complications occurred in 0.69% of the entire cohort.[sup.2] Mortality rates at 120 days of 1.3% have been also been reported from a cohort of 22 175 patients by Gallina and colleagues.[sup.3] Unfortunately, there are currently no good alternative diagnostic techniques to histologically diagnose and confirm the presence of prostate cancer, and so TRUS-guided prostatic biopsy will remain in widespread usage for the foreseeable future.

Numerous prostate biopsy schemes exist, with no universally agreed standard,[sup.4] although most authors suggest an extended biopsy scheme.[sup.5] Variations in core biopsy schemes can arise from the TRUS view chosen (axial vs. sagittal), the number of cores taken (commonly ranging from 10 to 20), and the anatomical location of the biopsies themselves (e.g., sextant vs. sextant plus alternate sites[sup.6,7]). Given the small but definite biopsy risks, sufficient training of urology and radiology residents is difficult given the great variation in procedural schemes. Current training methods for TRUS-guided prostatic biopsy are based on an apprenticeship approach, with trainees watching some procedures, and then performing the procedure under the guidance of a mentor; eventually the trainee will perform the procedure without supervision. The problem with this traditional approach to medical skills acquisition is that the patient is being exposed to the complete learning curve of the trainee. Some have argued that simulation-based training is imperative to further protect patients' interests.[sup.8]-[sup.10] Other reasons for using simulators include decreasing access to patients for training purposes, time constraints for teaching during procedures due to demands for greater efficiency and shorter working hours for residents; simulators can also be used as a tool to assess initial competence and subsequent re-certification.[sup.11] Other industries, most notably the aviation industry, have long used simulators for this purpose, where both initial training and subsequent re-certification can be performed safely without putting lives in danger.

To improve the training of TRUS-guided prostatic biopsy, we have developed and performed initial validation of a simulator for this procedure. The aim of this study was to perform intrinsic reliability testing (defined as the reproducibility or consistency of a test [sup.12,13]), to demonstrate face validity (defined as the opinion of novices with regards to realism[sup.13]), content validity (defined as the opinion of experts with regards to whether the simulator teaches what it is designed to teach[sup.13]), and finally to demonstrate construct validity (defined as the ability of the simulator to discriminate between experts and novices[sup.13]).

Methods

Patient prostate image database

3D TRUS prostate images of 50 patients were collected using a side-firing, biplane 2D TRUS probe attached to a rotational mover. [sup.14,15] All patients were scheduled for prostate biopsy for suspected prostate cancer. All 3D TRUS images were collected with approval from the local institutional ethics board.

Prostate biopsy simulator design

The TRUS-guided prostatic biopsy simulator was designed to improve the accuracy of prostatic ultrasound, specifically through the development of novel 3D ultrasound techniques.[sup.14]-[sup.16] The development of the simulator has been a collaborative effort between imaging physicists, engineers, academic urologists and academic radiologists carried out at a dedicated imaging research facility (Robarts Research, London, Ontario, Canada) in conjunction with clinical departments at the University of Western Ontario. The design of the biopsy simulator incorporates a patient's 3D TRUS prostate image into a mock pelvis allowing for multiple simulated biopsies to be performed using a standard end-fire TRUS probe. The database of patient TRUS images provides prostates of varied shape, size and pathology for biopsy training.

The hardware of the simulator is composed of a TRUS biopsy probe, a 3D magnetic tracking system, a mock pelvis and a biopsy activator foot pedal (Fig. 1). A standard clinical 2D TRUS probe is held and manipulated by the operator to perform a simulated biopsy within the mock pelvis (Fig. 2). The end-fire TRUS probes were used for our biopsy simulator; however, the simulator design allows for the use of a biplane side-fire TRUS probe, if preferred. A contoured polyoxymethylene plastic housing was manufactured to attach to an end-fire TRUS probe and contain a micro-magnetic tracking sensor (Fig. 3). The embedded magnetic sensor is 1.8 mm in diameter and allows for real-time monitoring of the TRUS probe position using a second-generation, Aurora magnetic tracking system (NDI, Waterloo, ON). The magnetic sensor can monitor the probe movement with 6 degrees of freedom (position: x, y, z and rotation: pitch, roll, yaw). The position of the magnetic sensor on the TRUS probe was calibrated with the TRUS image location, so that the position and orientation of the 2D TRUS image plane are known and can be monitored throughout the biopsy procedure.

The mock pelvis is a 6-sided (height x width x depth = 9 .5 x 14.0 x 11.0 cm[sup.3]) rectangular box, manufactured using polyoxymethylene plastic, with a single 3.1 cm diameter porthole. Large rubber bands were crossed over the hole to create a tight, elastic entry port, representative of an anus (Fig. 4). A dense elastic foam was embedded within the pelvis to create a pliable surface representative of a rectal wall. The elastic foam location within the mock pelvis is adjustable to allow for the simulated biopsy to be performed in left lateral decubitus or lithotomy positions. The patient's 3D TRUS prostate image is virtually placed behind the foam rectal wall within the mock pelvis, such that simulated transrectal biopsies are performed by inserting the tracked TRUS probe through the anus to "image" the prostate across the mock rectal wall. It is important to note that no real TRUS imaging is performed during the simulated biopsy procedure. Instead, the tracked orientation and position of the TRUS probe within the mock pelvis is used by the simulator software to dynamically slice the 3D TRUS prostate image to provide cross-sectional TRUS images that replicate those seen in a biopsy procedure (Fig. 4).

The hardware inputs are received by the biopsy simulator software and were integrated on a standard PC (Pentium 4, 3.2 GHz, 1 GB RAM). The simulator user interface provides multiple biopsy training modes including evaluation tools. The 2D TRUS mode displays standard 2D TRUS images of a patient's prostate as the operator manipulates the probe within the mock pelvis and is representative of the actual biopsy procedure. Examples of anatomy and pathology from the prostate simulator are shown in Fig. 4. The 3D TRUS mode extends 2D TRUS by providing an anatomical reference of the TRUS image location relative to 3D models of the patient's pelvic anatomy (Fig. 5). Patient-specific models of the prostate, bladder and urethra are reconstructed from the original 3D TRUS image, and the relative anatomical position of the TRUS image plane is displayed in real-time. The 3D TRUS mode user interface is an enhancement of a clinical 3D TRUS biopsy interface,[sup.16] and it allows the trainee to visualize the 3D TRUS probe location within the pelvis to aid with their orientation.

Multiple biopsy schemes can be designed and incorporated into both the 2D and 3D TRUS simulator modes (Fig. 5); however, the location of each biopsy target is not visible to the user under 2D TRUS. This allows trainees to learn biopsy schemes within the 3D TRUS mode, and then practice performing simulated schematic biopsy procedures under the 2D TRUS guidance. Activation of a foot pedal connected to the simulator software signals initiation of biopsy tissue collection and the location of the TRUS image along with the biopsy needle path are recorded simultaneously. Within the simulator, the accuracy of each biopsy is automatically calculated, based on the 3D distance of the schematic biopsy target from the biopsy needle line on the TRUS image, and the total time required to perform the biopsy is recorded. This information can be displayed to the trainee after each biopsy or summarized after completion of a full procedure.

Simulator evaluation

Intrinsic reliability of the simulator was assessed using the technique of split-halves,[sup.13] comparing the left side of the prostate with the right side of the prostate for 14 physicians with regards to time taken to complete a 12-core biopsy. The split-halves technique correlates the performance on one side of the prostate with the performance on the other side of the prostate.[sup.13]

To test, validate and gain further insight into possible improvements with the TRUS guided prostatic biopsy simulator, we developed a content and face validity questionnaire, which uses a Likert scale for evaluators of this new device (Appendix 1). This questionnaire sought to gain insights on the evaluators impressions of the simulator, in particular on the realism, the ease of use of the interface, the realism of the force feedback (haptics), the similarity of the images to a regular ultrasound machine, the range of motion of the ultrasound probe and whether the simulator would be useful for training urology and radiology residents on TRUS-guided biopsy of the prostate. The simulator was trialled and the questionnaire administered to 26 physicians (14 experts and 12 novices).

Construct validity was assessed by comparing experts and novices with regards to accuracy and time taken to perform a TRUS guided prostatic biopsy on the simulator. Simulated biopsies were performed on 3 separate prostate cases by 5 biopsy experts and a mixture of 9 radiology and urology resident trainees to evaluate the intrinsic reliability and construct validity of the biopsy simulator. The biopsies were performed in a targeted manner on 12 virtual biopsy targets that were arranged in a standard 12-core sextant scheme with medial and lateral targets defined for each region. This experiment used and expanded upon previously published work from our group, which compared simulated prostate biopsy accuracies using 2D ultrasound to 3D ultrasound systems, and involved only experts and radiology residents.[sup.17] Prior to each simulated biopsy, a single, static 2D TRUS image of the virtual target was provided to the operator for review. It is important to note that the length of time for target review was not limited and was at the discretion of the biopsy operator for each virtual target. Following sufficient review, the operator performed a simulated biopsy under 2D TRUS guidance and attempted to sample the location of the virtual target, which was no longer displayed. When the operator was confident in the position of TRUS probe, then the biopsy needle was fired to signify tissue collection. For each simulated biopsy, the simulator automatically recorded the total time for review and biopsy guidance along with the accuracy of the biopsy. Biopsy accuracy was defined and calculated within the simulator as the 3D Euclidean distance between the virtual target and the expected path of the biopsy needle on the TRUS image. Large values in biopsy accuracy suggest poor alignment or targeting of the biopsy needle and a decreased likelihood of sampling tissue surrounding the virtual target.

Statistical analysis was performed using SPSS (SPSS, Chicago, IL) and Stata (StataCorp LP, College Station, TX), utilizing parametric methods; statistical significance was defined as <0.05.

Results

All components of the software were developed and successfully integrated with the 3D magnetic tracking hardware and the biopsy activation foot pedal. Patient-specific 3D pelvic anatomy models were created for each 3D TRUS prostate image within the patient database, and each prostate was successfully incorporated with the simulator for training purposes.

Intrinsic reliability was assessed by comparing the left side with the right side of the prostate with regards to the time taken for each biopsy for the 14 physicians involved in the construct validity experiments. A total of 252 cores from the 14 physicians were used for this analysis. Mean time for the left side and right side was 62.4 seconds (95% CI 57.2-67.7 seconds) and 63.0 seconds (95% CI 58.4-67.6 seconds) respectively, with no statistically significant difference. Strong correlation was noted (r = 0.83) between the time taken for the two sides of the prostate, confirming the intrinsic reliability of the simulator.

The device was successfully assessed by both novice and expert physicians in the area of TRUS-guided biopsy of the prostate (Table 1). Physicians were equally divided between radiologists and urologists. The simulator was tested by academic and community urologists at a regional urological conference. Among these expert assessors, 93% found the simulation accurate, 100% found the interface easy to use, 100% agreed that the images on the simulator moved in a similar manner to a regular ultrasound machine, 64% regarded the haptics as lifelike, 93% rated the range of motion as realistic, and all felt that the simulator would be useful for training residents. For novice trainees, the pool of radiology and urology residents was used at the University of Western Ontario (London, Ontario, Canada) training program. Among these novice assessors, 67% found the simulation realistic, 100% found the interface easy to use, 83% rated the range of motion as realistic, 83% felt that the images on the simulator moved in a similar manner to a regular ultrasound machine, and 67% regarded the haptics as lifelike. All novices (100%) felt the simulator would be a useful tool in their training.

Construct validity was assessed by comparing 5 experts with 9 novices with regards to the time taken and accuracy for each biopsy (Table 2). This experiment used previously published work from our group, which assessed the accuracy of 5 experts using a 3-dimensional ultrasound system.[sup.17] We found differences between experts and novices using the simulator with regards to the time taken (34.7 seconds vs. 78.3 seconds, respectively); however the differences in accuracy of each biopsy between experts and residents (3.76 mm vs. 4.16 mm, respectively) did not reach statistical significance. The differences in accuracy were trending towards reaching statistical significance (p = 0.16), and we suspect that the difference could be confirmed with larger sample sizes. During construct validity testing, we also noted significant improvements amongst residents in the time taken between the first prostate biopsied and the third prostate biopsied. As expected, such an improvement was not seen among experts (Fig. 6).

Discussion

We have successfully developed a TRUS-guided prostatic biopsy simulator, and initial content and face validity testings have shown encouraging results, with all testers finding the simulator to be realistic. Of note, force feedback (or haptics), which is often very difficult to replicate in the simulation environment, was realistic in this simulator. We have also demonstrated intrinsic reliability and construct validity for this simulator. Other future potential applications of this technology lie with the training and accreditation of newer minimally invasive procedures which use prostatic ultrasound to guide therapy, as these minimally invasive options become more widely accepted with sound guidelines underscoring their usage (e.g., prostate cryosurgery[sup.18]).

There are no comparable simulators for a TRUS-guided prostatic biopsy reported in the literature.[sup.12] There are, however, numerous other simulator devices which are being developed or are in usage, particularly in the fields of laparoscopy [sup.19,20] and endoscopy.[sup.21]-[sup.25] Eventually, these simulators will hopefully be incorporated into both medical education and subsequently medical re-certification, with governments and universities around the world investing in the required infrastructure for simulation centres to improve patient safety and outcomes. All of these simulators require multiple forms of validation (e.g., predictive, content, face, concurrent) prior to being used,[sup.13] with early consensus panels noting the lack of rigorous validation prior to usage.[sup.26] Although we have now performed face, content and construct validity testing, our simulator warrants larger scale studies given the promising initial findings. These larger scale studies will need to demonstrate predictive validity (defined as the extent to which the simulator predicts future performance in the operating room), using an objective structured assessment of technical skills (OSATS) as the main outcome measure.[sup.13]

Potential limitations, in particular, the lack of actually seeing the biopsy gun firing in real-time, were identified by the testers of the device. However, this is a minor limitation given the fact that we can accurately track the exact location of where the biopsy was taken using our software, and then show the trainee where the biopsy was taken from. Other limitations noted include the fact that we have only tested the simulator with a small sample size of 26 physicians. We do plan to do further research with the simulator to demonstrate predictive/transfer validity. Also, our simulator demonstrated similar biopsy accuracy between experts and residents; this is likely to be due to the small difference in accuracy between experts and residents, which will necessitate a much larger sample size than is available in any one training program. Conversely, our simulator found significant differences in the time taken for each biopsy, suggesting that residents are able to achieve similar levels of accuracy as experts, but take longer to achieve those results. Validation of surgical simulators prior to widespread usage is required, especially if they are to be used for re-certification procedures. Similar to the aviation industry, it is not inconceivable that if the standard biopsy template or the technique change (e.g., usage of local anaesthesia), practicing physicians will have to demonstrate competence on a TRUS-guided prostatic biopsy simulator to continue practicing. Such procedures are commonplace in the aviation industry, where technology on planes and standard procedures are constantly being updated.

Conclusion

In the ideal world, all physicians would perform procedures on a simulator before performing the procedure under supervision on a patient. We believe that our TRUS-guided prostatic biopsy simulator could potentially be incorporated into training programs for radiology and urology residents. Furthermore, the simulator could also be used on a continuing basis for re-certification of practicing physicians (i.e., every 5 to 10 years). We believe that better training, along with continual improvement in the basic technology of ultrasound (e.g., 3D sonography [sup.27,28]) may lead to improved outcomes and safety for patients.

A virtual reality TRUS simulator has successfully been created; it has promising face, content and construct validity results. More predictive validity studies are needed to further develop this technology.

See related article on page 27.

Competing interests: None declared.

This paper has been peer-reviewed.

References

1. Medicare Australia. https://www.medicareaustralia.gov.au/statistics/mbs_item.shtmlAccessed January 20, 2011.

2. Kakehi Y, Naito S. Japanese Urological Association. Complication rates of ultrasound-guided prostate biopsy: a nation-wide survey in Japan. Int J Urol 2008;15:319-21.

3. Gallina A, Suardi N, Montorsi F, et al. Mortality at 120 days after prostatic biopsy: a population-based study of 22,175 men. Int J Cancer 2008;123:647-52.

4. Eichler K, Hempel S, Wilby J, et al. Diagnostic value of systematic biopsy methods in the investigation of prostate cancer: a systematic review. J Urol 2006;175:1605-12.

5. Karakiewicz PI, Perrotte P, McCormack M, et al. Early detection of prostate cancer with ultrasound-guided systematic needle biopsy. Can J Urol 2005;12Suppl 2:5-8.

6. Ochiai A, Troncoso P, Babaian RJ. Comparison of cancers detected at only a sextant or alternative location. BJU Int 2008;101:837-40.

7. Naya Y, Ochiai A, Troncoso P, et al. A comparison of extended biopsy and sextant biopsy schemes for predicting the pathological stage of prostate cancer. J Urol 2004;171:2203-8.

8. Ziv A, Wolpe PR, Small SD, et al. Simulation-based medical education: an ethical imperative. Simul Healthc 2006;1:252-6.

9. Chin MW, Forbes GM. Should simulator use become mandatory in endoscopy training. J Gastroenterol Hepatol 2008;23:996-7.

10. Nedas T, Challacombe B, Dasgupta P. Virtual reality in urology. BJU Int 2004;94:255-7.

11. Shah J, Mackay S, Vale J, et al. Simulation in urology--a role for virtual reality? BJU Int 2001;88:661-5.

12. Wignall GR, Denstedt JD, Preminger GM, et al. Surgical simulation: a urological perspective. J Urol 2008;179:1690-9.

13. McDougall EM. Validation of surgical simulators. J Endourol 2007;21:244-7.

14. Tong S, Downey DB, Cardinal HN, et al. A three-dimensional ultrasound prostate imaging system. Ultrasound Med Biol 1996;22:735-46.

15. Chin JL, Downey DB, Elliot TL, et al. Three-dimensional transrectal ultrasound imaging of the prostate: initial experience with an emerging technology. Can J Urol 1999;6:720-6.

16. Cool D, Sherebrin S, Izawa J, et al. Design and evaluation of a 3D transrectal ultrasound prostate biopsy system. Med Phys 2008;35:4695-707.

17. Cool DW, Connolly MJ, Sherebrin S, et al. Repeat prostate biopsy accuracy: simulator-based comparison of two- and three-dimensional transrectal US modalities. Radiology 2010;254:587-94.

18. Babaian RJ, Donnelly B, Bahn D, et al. Best practice statement on cryosurgery for the treatment of localized prostate cancer. J Urol 2008;180:1993-2004.

19. Hruby GW, Sprenkle PC, Abdelshehid C, et al. The EZ Trainer: validation of a portable and inexpensive simulator for training basic laparoscopic skills. J Urol 2008;179:662-6.

20. McDougall EM, Kolla SB, Santos RT, et al. Preliminary study of virtual reality and model simulation for learning laparoscopic suturing skills. J Urol 2009;182:1018-25.

21. Knudsen BE, Matsumoto ED, Chew BH, et al. A randomized, controlled, prospective study validating the acquisition of percutaneous renal collecting system access skills using a computer based hybrid virtual reality surgical simulator: phase 1. J Urol 2006;176:2173-8.

22. Schout BM, Muijtjens AM, Hendrikx AJ, et al. Acquisition of flexible cystoscopy skills on a virtual reality simulator by experts and novices. BJU Int 2009;105:234-9.

23. Michel MS, Knoll T, Kohrmann KU, et al. The URO Mentor: development and evaluation of a new computer-based interactive training system for virtual life-like simulation of diagnostic and therapeutic endourological procedures. BJU Int 2002;89:174-7.

24. Shah J, Darzi A. Virtual reality flexible cystoscopy: a validation study. BJU Int 2002;90:828-32.

25. Knoll T, Trojan L, Haecker A, et al. Validation of computer-based training in ureterorenoscopy. BJU Int 2005;95:1276-9.

26. Carter FJ, Schijven MP, Aggarwal R, et al. Consensus guidelines for validation of virtual reality surgical simulators. Simul Healthc 2006;1:171-9.

27. Shen F, Shinohara K, Kumar D, et al. Three-dimensional sonography with needle tracking: role in diagnosis and treatment of prostate cancer. J Ultrasound Med 2008;27:895-905.

28. Narayanan R, Werahera PN, Barqawi A, et al. Adaptation of a 3D prostate cancer atlas for transrectal ultrasound guided target-specific biopsy. Phys Med Biol 2008;53:N397-406.

Figures and Tables

Fig. 1: Frontal view showing simulator and magnetic tracking transmitter. [Figure omitted]

Fig. 2: Ultrasound probe in use inside the simulator. [Figure omitted]

Fig. 3: Ultrasound probe with attached micro-magnetic tracking sensor. [Figure omitted]

Fig. 4: 2D TRUS views from simulator. (a) Mid-gland image displaying verumontanum. (b) prostatic calcification in apex. (c) Right side peripheral zone hypoechoic lesion suspicious for prostate cancer. [Figure omitted]

Fig. 5: 3D view showing biopsy control screen (axial view in upper panel; 3D sagittal view bottom left panel; 3D axial view bottom right panel). Urethra is colored blue; prostate red; bladder green. Potential areas to biopsy have been marked with green spheres in lower panels. [Figure omitted]

Fig. 6: Improvement in biopsy time between first case and third case. [Figure omitted]

Table 1: Survey results [Table omitted]

Table 2: Construct validity: comparison of experts and novices [Table omitted]

Appendix 1: Content validity questionnaire: prostate biopsy simulator [Table omitted]

Author Affiliation(s):

[1] Departments of Surgery & Oncology, Divisions of Urology & Surgical Oncology, Schulich School of Medicine & Dentistry, University of Western Ontario, London, ON

[2] Robarts Research Institute, University of Western Ontario, London, ON

Correspondence: Dr. Jonathan I. Izawa, Associate Professor, Departments of Surgery & Oncology, Divisions of Urology & Surgical Oncology, London Health Sciences Centre-Victoria Hospital, 800 Commissioners Road East, Suite C3-120, London, ON N6A 4G5; fax: 519 685 8450; jonathan. izawa@lhsc.on.ca
COPYRIGHT 2011 Canadian Urological Association
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2011 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Original Research
Author:Chalasani, Venu; Cool, Derek W.; Sherebrin, Shi; Fenster, Aaron; Chin, Joseph; Izawa, Jonathan I
Publication:Canadian Urological Association Journal (CUAJ)
Article Type:Report
Geographic Code:1CANA
Date:Feb 1, 2011
Words:4408
Previous Article:High resolution analysis of wait times and factors affecting surgical expediency.
Next Article:Development and validation of a virtual reality transrectal ultrasound guided prostatic biopsy simulator.
Topics:

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters