1. Introduction

Radioactive-contaminated sites can be released from regulatory control only if the remaining radioactivity is less than the regulatory limit [1]. Prior to its decontamination, the radiation from the site needs be identified, which provides accurate information on the radionuclides to be decontaminated. The radionuclides frequently found at the contaminated sites of foreign nuclear facilities include Am-241, Cs-137, Co-60, H3, I-129, I-131, Pu, Ra, Rn, Sr-90, Tc-99, Th, and U [2]. The radionuclides found at the decommissioned site of an old Korean research reactor, TRIGA MARK III, were Co-60, Cs-134, Cs-137, Sr-90, Cr-51, and Fe-59 [3].

Most radionuclides found at the contaminated site are gamma-emitters, but some are beta-emitters that need to be considered in the subsequent decontamination stage. Sr-90, which emits a 0.546 MeV beta particle, has chemical similarity to calcium. The radionuclide accumulates in bones and irradiates the bone marrow causing its high radiotoxicity [4]. Accordingly, this study selected Sr-90 as the target radionuclide to be detected by a fiber-optic radiation sensor (FORS).

Several fiber-optic radiation sensors have been developed for the remote and real-time measurements. Park and Joo developed an underwater beta radiation monitoring sensor using a CaF2 : Eu, YAG:Ce, YAP: Ce scintillator [5]. Seo et al. developed a thin film type of plastic scintillation detector [6]. Jang et al. fabricated a fiber-optic radiation sensor for the detection of tritium [7]. Kim et al. developed a ZnS(Ag)/BC-408 phoswich detector for monitoring radioactive contamination inside pipes [8]. Hong et al. developed a fiber-optic dual detector to measure beta- and gamma-rays simultaneously [9].

In this study, a FORS was developed for the remote and real-time measurement of beta radiation from radioactive-contaminated soil. The sensing probe consisted of a bundle of organic scintillators and a mixture of epoxy resin to improve its detection efficiency of lights. Several sensing probes, whose dead-end diameters were different, were prepared and characterized to select the best sensing probe in terms of its radiation detection efficiency.

2. Experimental Setup

The organic scintillator used was a plastic scintillating fiber (BCF12, Saint-Gobain). Despite being more expensive than a liquid scintillator, the plastic scintillating fiber is chemically stable as a solid material and is easy to fabricate [10]. Table 1 lists its physical properties [11].

The light generated from scintillation was transmitted to the light guide, which was fabricated with transparent epoxy resin that can transmit the light efficiently to the optical fiber owing to its small optical loss resulting from the absorption and leakage of the light. Its characteristics depend on the epoxy resin to hardener mixing ratio. To meet the characteristics required for the light guide, the epoxy resin (YD-128, Kukdo., Korea) and hardener (D-230, Kukdo., Korea) were mixed at a ratio of 5: 1 [12]. The mixture of epoxy resin was prepared in a truncated conical shape.

Table 2 shows cross-sectional views of two parts of the FORS sensing probes. The overall diameters of the sensing probes were 26 mm (Bundle type I), 36 mm (Bundle type II), and 46 mm (Bundle type III). The bundles of the scintillating fibers were attached to the broader end of the epoxy resin mixture. The sides of the sensing probes were wrapped with aluminium foil (3 [micro]m, Samwha) as a reflector to reduce light leakage.

A plastic optical fiber (#53-833, Edmund optics) was used as a transmitting optical fiber. The plastic optical fiber has small transmission loss with a peak value of 435 nm. Table 3 lists the properties of the plastic optical fiber.

Figure 1 shows the experimental setup for measuring beta radiation, which consisted of a photomultiplier (PMT, R1924, Hamamatsu), preamplifier (Canberra amplifier model 2006), multichannel analyzer (MCA, Canberra Multiport II), and a laptop computer that was powered by a high-voltage power supplier (Canberra model 3102D). The light transmitted through the transmitting optical fiber to the PMT tube was converted to an electrical signal. This electrical signal was amplified by the preamplifier. The signals were acquired by MCA and analyzed using the GENIE 2000 program installed at the laptop computer.

The FORSs were applied to measure beta radiation from a Sr-90 disk source (0.1 [micro]Ci, 2014.09, Spectrum Technique). The FORSs were installed in a dark room to reduce the influences of external light. The sensing probes of the FORS were located 30 mm above the beta disk source to reduce the effects of backscattered radiation [13]. For comparison, beta radiation from a Sr-90 disk source was also detected using a commercial GM counter (GM214007, NEO).

Because it cannot help being affected by the background radiation, the radiation measurement system should distinguish between background radiation and the radiation to be measured. The critical level ([L.sub.C]) is defined as the net count rate which must be exceeded before the sample can be said, with a certain level of confidence, to contain measurable radioactivity above background. For the critical level, this paper used (1) to check if the FORS can measure the radiation from the beta source distinguishable from the background radiation, as suggested by [14].

[L.sub.C] [congruent to] 2.33 [absolute value of ([n.sub.b])], (1)

where [n.sub.b] is the total background counts.

3. Results and Discussion

Figure 2 shows the measurements by the FORSs with the three different diameters. Each measurement by each FORS was made for over 600 seconds and repeatedly 20 times. The total counts in Table 4 are the average of the 20 measurements obtained by using the FORSs. The total counts by the FORSs with 26 mm, 36 mm, and 46 mm in diameter were 15341, 16543, and 17004, respectively, which are summarized in Table 4, along with other measurement values. The FORS detected more radiation from the source as the diameter of its sensing probe was increased. For the bundle type III of the sensing probe, the peak values were higher than the other two sensing probes. Compared to that by using a commercial GM counter for over 600 seconds, the relative total counts obtained by using the FORSs with a diameter of 26 mm, 36 mm, and 46 mm were found to be 26.9%, 29.0%, and 29.8%, respectively.

The reproducibility of the FORS was tested by applying a [chi square]-test to the measurements as a minimum requirement for a radiation sensor. Because radioactive decay is a random process, any measurement of radiation is subject to a degree of the statistical fluctuation. A [chi square]-test is constructed from a sum of squared errors or through the sample variance. If the probability for the estimated [chi square] is in the range of 0.1 < p < 0.9, it can be determined that the radiation detection system produces reproducible measurement reflecting the statistical fluctuations of radioactive decay appropriately [15-18].

Table 5 lists the 20 measurements by each FORS and the relevant statistics for the [chi square]-test. The [chi square]s for the measurements by the FORSs were 19.72, 18.55, and 14.70 for a diameter of 26 mm, 36 mm, and 46 mm, respectively. The probabilities of [chi square]'s with 19 degrees of freedom were 0.54, 0.51, and 0.76, respectively. Figure 3 shows the distributions of a statistic ([N.sub.i] - [N.sub.av])/[sigma], where [N.sub.i] is ith measurement by the FORSs and GM counter, Nav is the average, and a is the standard deviation of the 20 measurements. Figure 3 helps to check intuitively if the FORSs generate the statistically significant measurements, showing the difference between the distributions of the statistic and the normal distribution.

4. Conclusion

This study developed and characterized a FORS for the remote and real-time measurements of beta radiation. The FORS consisted of the sensing probe with an aluminium foil reflector, transmitting plastic optical fiber, and light-measuring system that comprised a photomultiplier tube, preamplifier, multichannel analyzer, and laptop computer. The measurements showed that the FORS could be used as a radiation sensor and that the FORS sensing probe with a diameter 46 mm has the best detection performance.

https://doi.org/10.1155/2017/1452765

Conflicts of Interest

The authors declare that they have no conflicts of interest.

Acknowledgments

This study was supported by a National Research Foundation of Korea (NRF) grant funded by the Korea government, Ministry of Science, ICT and Future Planning (MSIP, Research Project no. 2012M2B2B1055499).

References

[1] "U. S. Nuclear Regulatory Commission," 10 CFR part 20-standards for protection against radiation.

[2] L. E. Boing, Introduction to decommissioning, Argone National Laboratory, 2013.

[3] G.N. Kim et al., "Methods of recycling soil washing wastewater for volume reduction," Journal of KoSSGE, vol. 8, pp. 17-26, 2003.

[4] S. K. Sahoo, N. Kavasi, A. Sorimachi et al., "Strontium-90 activity concentration in soil samples from the exclusion zone of the Fukushima daiichi nuclear power plant," Scientific Reports, vol. 6, Article ID 23925, pp. 1-10, 2016.

[5] H. M. Park and K. S. Joo, "Feasibility Study on Development of an Underwater Beta-ray Monitoring Sensor," Journal of Sensor Science and Technology, vol. 25, no. 5, pp. 333-336, 2016.

[6] B. K. Seo, Z. H. Woo, W. Z. On, K. W. Lee, and M. J. Han, "Preparation of a thin film type of plastic scintillation detector for beta-ray detection," Analytical science and Technology, vol. 18, pp. 495-499, 2005.

[7] K. Jang, D. Cho, W. Yoo et al., "Fabrication and Characterization of a Fiber-optic Radiation Sensor for Detection of Tritium," Korean Journal of Optics and Photonics, vol. 20, no. 4, pp. 201-206, 2009.

[8] G. H. Kim, C. H. Park, C. H. Jung, K. W. Lee, and B. K. Seo, "Development of the ZnS(Ag)/BC-408 phoswich detector for monitoring radioactive contamination inside pipes," Journal of the Korean association for radiation protection, vol. 31, pp. 123-128, 2006.

[9] S. Hong, S. H. Shin, H. I. Sim et al., "Feasibility study on development of a fiber-optic dual detectorto measure beta- And gamma-rays simultaneously," The Transactions of Korean Institute of Electrical Engineers, vol. 63, no. 2, pp. 284-290, 2014.

[10] K. Han, W. Yoo, S. Shin et al., "Development of Fiber-optic Radiation Sensor Using LYSO Scintillator for Gamma-ray Spectroscopy," Journal of Sensor Science and Technology, vol. 21, no. 4, pp. 287-292, 2012.

[11] D. L. Chichester, S. M. Watson, and J. T. Johnson, "Comparison of bcf-10, bcf-12, and bcf-20 scintillating fibers for use in a 1-dimensional linear sensor," IEEE Transactions on Nuclear Science, vol. 60, no. 5, pp. 4015-4021, 2013.

[12] C. H. Park, J. H. Moon, and B. K. Seo, "Development and characterization of the integrated fiber-optic sensor for remote detection of alpha radiation," Journal of the Korean Physical Society, vol. 63, no. 9, pp. 1720-1723, 2013.

[13] H. Joo, R. Kim, and J. H. Moon, "Radioactivity measurement of radioactive contaminated soil by using a fiber-optic radiation sensor," Journal of the Korean Physical Society, vol. 68, no. 11, pp. 1287-1290, 2016.

[14] J. E. Martin, Hysics for Radiation Protection: A Handbook, Wiley-VCH, Berlin, Germany, 2 edition, 2006.

[15] G. F. Knoll, Radiation Detection and Measurement, John Wiley and Sons, New York, NY, USA, 3 edition.

[16] N. Tsoulfanidis, Measurement and Detection of Radiation, Hemisphere, Washington DC, USA, 1983.

[17] P. R. Bevington, Data reduction and error analysis for the physical sciences, McGraw-Hill, New York, NY, USA, 1969.

[18] D. M. Namuth-covert, "Chi-square test for goodness of fit in a plant breeding example," Plant & Soil Sciences [eLibrary.sup.PRO].

Rinah Kim, Sang Bin Lee, Jae Wook Kim, and Joo Hyun Moon

Dongguk University, Gyeongju Campus, 123 Dongdae-ro, Gyeongbuk 38066, Republic of Korea

Correspondence should be addressed to Joo Hyun Moon; jhmoon86@dongguk.ac.kr

Received 20 April 2017; Revised 2 September 2017; Accepted 14 September 2017; Published 18 October 2017

Academic Editor: Andrea Cusano

Caption: FIGURE 1: Experimental setup for measuring beta radiation.

Caption: FIGURE 2: Comparison of (a) pulse height spectra and (b) measurements (cps) by the FORSs with different diameters and a commercial GM counter.

Caption: FIGURE 3: Distributions of a statistic ([N.sub.i] - [N.sub.av])/[sigma] for 20 measurements by the FORSs and a commercial GM counter.
TABLE 1: Properties of the plastic scintillating fiber (BCF12, Saint-
Gobain).

Emission    Decay time   1/e length      Number of
peak (nm)      (ns)         (m)       photons per MeV

435            3.2          2.7            ~8000

TABLE 2: Cross-sectional views of two parts of the FORS sensing probes
(unit: mm).

                 (a) bundle   (b) bundle   (c) bundle
                   type I      type II      type III

Epoxy resin       [phi]26      [phi]36     [phi]46
mixture

Bundle of         [phi]26      [phi]36     [phi]46
scintillating
fibers

TABLE 3: Properties of the plastic optical fiber
(#53-833, Edmund optics).

Outer diameter   Core diameter      Attenuation
(mm)              ([micro]m)          (dB/km)

3                    2,950       150~300 (@ 650 nm)

Outer diameter     Numerical     Operating temperature
(mm)             aperture (NA)       ([degrees]C)

3                    0.50               -55~70

TABLE 4: Comparison of the measurements by the commercial
detector and FORSs.

Detector              Background   Critical level
                        counts      ([L.sub.c])

Commercial detector      273           38.49
  (GM counter)
FORS with 26 mm         12367          259.11
FORS with 36 mm         12915          264.79
FORS with 46 mm         13274          268.45

Detector              Total counts    Counts per second

Commercial detector      56967       94.49 [+ or -] 0.45
  (GM counter)
FORS with 26 mm          15341       25.57 [+ or -] 0.40
FORS with 36 mm          16543       27.57 [+ or -] 0.32
FORS with 46 mm          17004       28.34 [+ or -] 0.20

TABLE 5: The 20 measurements by the FORSs with different
diameters and the relevant statistics.

Number of      FORS with   FORS with   FORS with
measurements     26 mm       36 mm       46 mm

(1)              15244       16521       16781
(2)              15369       16406       17039
(3)              15540       16553       17120
(4)              15188       16744       16919
(5)              15406       16556       17012
(6)              15256       16491       16859
(7)              15340       16781       17011
(8)              15233       16512       17142
(9)              15499       16352       17096
(10)             15533       16652       17102
(11)             15171       16581       17017
(12)             15396       16566       16988
(13)             15128       16279       16965
(14)             15273       16522       17108
(15)             15351       16588       17165
(16)             15406       16535       16948
(17)             15359       16571       16759
(18)             15552       16751       17004
(19)             15351       16409       16926
(20)             15226       16498       17114

Average          15341       16543       17004
[chi square]     19.72       18.55       14.70
P                0.54        0.51        0.76
COPYRIGHT 2017 Hindawi Limited
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2017 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Research Article
Author:Kim, Rinah; Lee, Sang Bin; Kim, Jae Wook; Moon, Joo Hyun
Publication:Journal of Sensors
Date:Jan 1, 2017
Words:2351
Previous Article:Sequential Monte Carlo Localization Methods in Mobile Wireless Sensor Networks: A Review.
Next Article:Automatic Supervision of Temperature, Humidity, and Luminance with an Assistant Personal Robot.
Topics:

Terms of use | Privacy policy | Copyright © 2020 Farlex, Inc. | Feedback | For webmasters