Design and performance of laser-pumped Cs-magnetometers for the planned UCN EDM experiment at PSI.

We designed laser-pumped cesium vapor magnetometers in the [M.sub.x] configuration for the control and stabilization of magnetic field fluctuations and gradients in a new experiment searching for a permanent electric dipole moment of the neutron. The intrinsic sensitivity of the device was determined to be 30 fT in a measurement bandwidth of 1 Hz, limited by laser noise. In the shot noise limit the magnetometer can reach a sensitivity of 7 fT for 1 s integration time. Test measurements of the stability of a 2 [micro]T magnetic field in a threefold magnetic shield have revealed fluctuations on the order of 200 fT to 300 fT with integration times in the range of 2 s to 100 s. Those fluctuations were traced back to the stability of the power supply used to generate the magnetic field. The laser-pumped magnetometer fulfills the requirements set by the planned neutron electric dipole moment experiment.

Key words: high precision magnetometry; laser spectroscopy; optical detected magnetic resonance; optical pumping.

1. Introduction

The precise measurement and control of magnetic fields and magnetic field fluctuations is an important step for experiments searching for a permanent electric dipole moment (EDM) of the neutron, and is one of the main factors limiting the accuracy. In a project approved by Paul Scherrer Institute (PSI) in Switzerland, a neutron EDM spectrometer is proposed [1] in which the neutron spin flip transition frequency is measured in four UCN storage chambers exposed to a homogenous 2 [micro]T magnetic field. In addition, each neutron chamber has two compartments in which the neutrons are exposed to a static electric field of 15 kV/cm oriented parallel/ antiparallel to the magnetic field. The signature of a finite EDM is a change of the neutron Larmor frequency synchronous with the rever sal of the relative orientations of the magnetic and electric fields. This experiment imposes very stringent constraints on the homogeneity and stability of the magnetic field.

Fluctuations of the magnetic field will be monitored by a set of 16 optically-pumped alkali-vapor magnetometers (OPM), using the property that the oscillation frequency of an OPM is proportional to the modulus of the local magnetic field. The magnetometers considered for the first generation of the neutron EDM experiment are self-oscillating Cs vapor magnetometers (OPM) in the Mx configuration [2,3], optically pumped by spectral discharge lamps (LpOPM). However, it was previously shown that the replacement of the lamp by a resonant laser can lead to an appreciable gain in magnetometric sensitivity [3,4]. In that spirit we have designed and tested a laser-pumped OPM (LsOPM) with a geometry compatible with the neutron EDM experiment under construction. In case the laser version shows superior performance it might be considered as an alternative to the lamp-pumped magnetometers in the EDM experiment. Here we present the design and discuss the performance of Cs-LsOPMs operated in a phase-stabilized mode.

2. The Magnetometer Setup

The magnetometer consists of three parts (Fig. 1): a) a sensor head containing no metallic parts except the r.f. coils, b) a base station mounted in a portable 19" rack drawer, which contains the frequency stabilized laser and all optics, and c) the frequency locking electronics. The sensor head is designed to fit into a tube of 104 mm diameter, coaxial with the 2 [micro]T field, and has a total length of 242 mm. Its main component is a paraffin-coated, 7 cm diameter glass cell containing a droplet of cesium. The Cs atoms in the vapor are optically pumped by a circularly polarized light beam, oriented at 45[degrees] to the magnetic field. An extended-cavity diode laser, which is stabilized on the F = 4 [right arrow] F = 3 hyperfine component of the [D.sub.1] line at 894 nm wavelength by the dichroic atomic vapor laser lock (DAVLL) method [5], provides the resonant light which is guided from the base station to the sensor head via a 10 m long multimode fiber (800 [micro]m core diameter). Note that a single diode laser can produce sufficient light power to drive all sensors in the experiment. The light transmitted through the cell is carried back to the detection unit by a similar fiber. The pumping process produces a polarization (magnetization) in the sample which undergoes a Larmor precession with the frequency [[omega].sub.L]. The precession is resonantly driven by a weak radio-frequency (rf) field (of amplitude [B.sub.I] and frequency [[omega].sub.rf]) applied with a pair of circular coils surrounding the cell. The absorption of the pumping light depends on the direction of the magnetization with respect to the light direction. Therefore the resonant precession frequency appears as an amplitude modulation on the light power transmitted through the cesium vapor. The detection of the modulated transmission by a photodetector allows a direct and real time measurement of the magnetic field in terms of the Larmor frequency.

[FIGURE 1 OMITTED]

The system behaves like a classical oscillator: when an rf field is applied, the phase of the response, i.e., the AC component of the transmitted laser intensity depends on the detuning [delta][omega] = [[omega].sub.rf] - [[omega].sub.L]. For [delta][omega] [much less than] 0 the atoms follow the driving field adiabatically and thus the radio frequency and the light modulation oscillate in phase. For [delta][omega] [much greater than] 0 the phase shift is -180[degrees], while on resonance ([delta][omega] = 0) it becomes -90[degrees], while the modulation amplitude of the transmitted light reaches a maximum. In the phase-stabilized mode this dependency is used to lock the radio frequency to the Larmor frequency. The photodiode signal is analyzed by a dual-channel lock-in amplifier, referenced by the driving frequency of the rf coils produced by a voltage controlled oscillator (VCO). Either the phase or the dispersive in-phase output of the lock-in amplifier can be used in a feedback loop to lock the driving frequency to the Larmor frequency by stabilizing the phase shift to -90[degrees]. The temporal response of the phase-stabilized magnetometer to field changes, i.e., its bandwidth is determined by the response time of the feedback loop and was measured to be in the order of 1 kHz.

3. Performance of the Magnetometer

3.1 Intrinsic Magnetometric Sensitivity

We define the intrinsic magnetometric sensitivity of the LsOPM in terms of the noise equivalent magnetic field (NEM). It is the square root of the power spectral density of magnetic field fluctuations at the magnetometer frequency. The field fluctuations produce a photodetector signal equal to the power spectral density of the photocurrent at the same frequency, each integrated over a bandwidth [f.sub.bw]. The NEM [delta]B is given by:

[delta] B = [1/[gamma]] * [[[DELTA][v.sub.HWHM]]/[S/[N.sub.int]]], (1)

where [gamma] is a constant of approximately 3.5 Hz/nT for [.sup.133]Cs. The half width of the resonance is [DELTA][v.sub.HWHM] and S/[N.sub.int] the signal-to-noise ratio of the photocurrent modulation in the phase-stabilized mode. The signal S is the amplitude of the modulation. When measuring the intrinsic noise level, [N.sub.int], care has to be taken to eliminate external contributions, such as drifts and noise from external magnetic field sources. We therefore infer the intrinsic noise [N.sub.int] from a Fourier analysis of the photodiode signal when the magnetometer is operated under optimal conditions (Fig. 2). The central peak (the so-called carrier) is the oscillating magnetometer signal. It is superposed on a broad pedestal, which results mainly from a continuous distribution of side-bands due to imperfectly shielded low-frequency field fluctuations. The two discrete side-bands originate from interference of magnetic fields oscillating at the line frequency. The intrinsic noise level [N.sub.int] is given by the noise floor in Fig. 2, integrated over the detection bandwidth [f.sub.bw]. It lies 50% above the shot noise level [DELTA]I = [square root of (2e[I.sub.pc][f.sub.bw])] of the DC photocurrent [I.sub.pc] in a band-width [f.sub.bw] of 1 Hz.

[FIGURE 2 OMITTED]

The optimum operating point was found for a laser intensity of 9 [micro]W/[mm.sup.2] and a [B.sub.1] field amplitude of 2.7 nT, which yielded a linewidth [v.sub.HWHM] = 3.4(1) Hz and a signal-to-noise ratio S/[N.sub.int] = 97 dB = 66 000. The corresponding intrinsic sensitivity is [delta][B.sub.int] = 14.5 fT. If one assumes the noise to be white, the sensitivity scales with the square root of the bandwidth. Assuming the photodiode shot noise as ultimate sensitivity limit the LsOPM should reach a sensitivity of [delta][B.sub.SN] = 10 fT in 1 Hz bandwidth, which corresponds to [delta][B.sub.SN] = 7 fT for a 1 s integration time. The laser generated low frequency 1/f noise is mixed to the Larmor frequency and thus contributes to the noise spectrum. It was found to exceed the white noise level [N.sub.int] by a factor 2 at the carrier frequency, which yields an intrinsic sensitivity of [delta][B*.sub.int] = 29 fT. This contribution can be suppressed using an active stabilization of the laser power.

3.2 Application: Field Fluctuations in a Magnetic Shield

We used the LsOPM to measure residual field fluctuations inside a three-layer magnetic shield. The Larmor frequency was recorded as multiple time series of several hours with a sampling rate of 0.1 s. From each time series the Allan standard deviation [6] of the corresponding magnetic field inside the shield was calculated. A typical result is shown in Fig. 3. The observed fluctuations (curve a) are well above the intrinsic sensitivity level of the magnetometer (curves b). For integration times up to 1 s to 2 s the noise amplitude decreases as [[tau].sup.1/2], where [tau] is the integration time, indicating the presence of white field-amplitude noise. It can be characterized by a spectral amplitude of 413 fT/H[z.sup.1/2]. With the estimated signal-to-noise ratio S/[N.sub.ext] = 2600 (cf. Fig. 2) we calculate from Eq. (1) a field stability of 370 fT in a measurement bandwidth of 1 Hz, which is in good agreement with the value in the Allan standard deviation plot. The minimum field fluctuations were found to be slightly larger than 200 fT for an integration time of 4 s. The central region of the Allan standard deviation plot shows a plateau for integration times of 2 s to 100 s. It could be traced back to fluctuations of the 8 mA current producing the 2 [micro]T bias field and corresponds to a relative stability of [10.sup.-7] for the power supply. The Allan standard deviations for integration times exceeding 100 s are due to slow drifts of laboratory fields which are not completely suppressed by the three-layer shield, which has a measured longitudinal shielding factor of [10.sup.3]. The influence of light shift effects [7] due to light intensity fluctuations are measured in an additional experiment. Figure 3c shows that those effects are negligible at the present level of field stability.

[FIGURE 3 OMITTED]

4. Summary and Conclusion

We have described the design and performance of a phase-stabilized cesium vapor magnetometer with a measurement bandwidth of 1 Hz. The magnetometer has an intrinsic sensitivity of 29 fT, defined as the Allan standard deviation for a bandwidth of 1 Hz. This value can be reduced by a factor of 2 if the 1/f noise of the laser power can be lowered, e.g., by an active power stabilization. If the excess white noise can be reduced to the shot noise level a further increase of 1.5 can be obtained. Under these optimal conditions the LsOPM could reach a sensitivity of 7 fT for an integration time of 1 s. Field fluctuations of 200 fT to 300 fT were measured in a three-layer magnetic shield for integration times between 2 s and 100 s. Light shift fluctuations, against which no particular precautions, except for the active frequency stabilization of the laser, were taken, are one to two orders of magnitude smaller than the residual field fluctuations in the shield. The laser-pumped OPM described here will thus be a valuable tool for fundamental physics experiments and compares very favorably with state-of-the-art lamp-pumped magnetometers as it will be demonstrated elsewhere.

Acknowledgements

We are indebted to E. B. Aleksandrov, A. S. Pazgalev, and Robert Wynands for numerous fruitful discussions. We acknowledge financial support from Schweizerischer Nationalfonds, INTAS, and Paul Scherrer Institute (PSI). We thank PSI for the loan of the high-stability current source.

5. References

[1] E. B. Aleksandrov et al., A new precision measurement of the neutron electric dipole moment (EDM), PSI Proposal R-00-05.2 (2002).

[2] L. Bloom, Appl. Opt. 1, 61 (1962).

[3] E. B. Aleksandrov et al., Opt. Spectrosc. 78, 292 (1995).

[4] E. B. Aleksandrov et al., Sov. Phys. Tech. Phys. 35, 371 (1990).

[5] V. V. Yashchuk, D. Budker, and J. R. Davis, Rev. Sci. Instrum. 71, 341 (2000).

[6] J. A. Barnes et al., IEEE Trans. Instrum. Meas. 20, 105 (1971).

[7] C. Cohen-Tannoudji, Ann. Phys. 7, 423 (1962); Ann. Phys. 7, 469 (1962).

About the authors: Antoine Weis is full professor at the Physics Department of the University of Fribourg and leader of the Fribourg Atomic Physics group (FRAP) at the Physics Department of the University of Fribourg, Stephan Groeger and Georg Bison are Ph.D. students in the FRAP group.

S. Groeger

Universite de Fribourg, Chemin de Musee 3, 1700 Fribourg, Switzerland

and

Paul Scherrer Institute, 5232 Villingen PSI, Switzerland

G. Bison, and A. Weis

Universite de Fribourg, Chemin de Musee 3, 1700 Fribourg, Switzerland

Accepted: August 11, 2005

Available online: http://www.nist.gov/jres
COPYRIGHT 2005 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2005, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

Article Details
Printer friendly Cite/link Email Feedback
Author:Weis, A.
Publication:Journal of Research of the National Institute of Standards and Technology
Date:May 1, 2005
Words:2296
Previous Article:Magnetic field stabilization for magnetically shielded volumes by external field coils.
Next Article:Superconducting UCN polarizer for a new EDM spectrometer.
Topics:


Related Articles
The fundamental neutron physics beamline at the Spallation Neutron Source.
Search for a neutron electric dipole moment.
Magnetic field stabilization for magnetically shielded volumes by external field coils.
Superconducting UCN polarizer for a new EDM spectrometer.
UCN production with a single crystal of ortho-deuterium.
Measurement of the loss and depolarization probability of UCN on Beryllium and diamond like carbon films.
First tests of [.sup.6]Li doped glass scintillators for ultracold neutron detection.
A silicon UCN detector with large area and with analysis of UCN polarization.
Precision measurements with slow neutrons: April 5-7, 2004.
Neutron lifetime experiment based on an accordion-like UCN storage volume coated with "low temperature Fomblin".

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters