D-dimer testing for deep venous thrombosis: a metaanalysis.

Deep venous thrombosis (DVT) (4) is a common condition with significant morbidity and mortality if not diagnosed and treated in a timely manner. The clinical signs and symptoms of DVT are nonspecific, and objective testing is required for diagnosis. Diagnostic imaging with contrast venography or compression ultrasonography has important limitations. Venography is invasive and expensive, is contraindicated in some patients, and requires a radiologist to perform. Compression ultrasound also requires a radiologist's interpretation and has poor sensitivity for detection of calf vein thrombosis, and serial testing is often required when the initial scan is negative.

The measurement of D-dimer, degradation products of circulating cross-linked fibrin formed during activation of the coagulation system, has been studied extensively as an adjuvant test in the diagnosis of DVT. D-Dimer testing has become rapid, simple, and inexpensive, and it has the potential to detect thrombosis in any part of the venous system. If the sensitivity of the D-dimer test for DVT is consistently very high, its negative predictive value will also be high and reliably exclude the presence of disease. These are characteristics of a good "rule-out" test. As such, the use of D-dimer assays has been suggested as an initial test to rule out DVT to reduce the number of patients requiring diagnostic imaging. In 1996, our systematic review did not find sufficient evidence to support the use of D-dimer as a diagnostic test for DVT (1). However, since that time several new D-dimer assays have been introduced, and >50 studies and 8 reviews on the subject have been published (2-9). Because none of the reviews critically appraised the primary studies or discussed the potential for bias in their results, we undertook this metaanalysis of the D-dimer accuracy literature published since our last systematic review to clarify the role of the test in the diagnosis of lower extremity DVT.

Methods

SEARCH STRATEGY

Using the MEDLINE database we searched for articles published in English from February 1995 through October 2003 that compared D-dimer assay results with lower extremity ultrasound or venography in symptomatic patients with suspected acute DVT. We used the following search strategy: venous thrombosis (MeSH heading) AND fibrin fibrinogen degradation products (MeSH heading). We reviewed the bibliographies of articles obtained through the initial search for additional pertinent references.

INCLUSION CRITERIA AND ASSESSMENT OF STUDY QUALITY

We included for review studies that calculated sensitivities and specificities or provided paired test data for at least one D-dimer assay. We excluded abstracts, subgroup analyses of previously published studies, and studies reporting combined data of patients with suspected DVT and pulmonary embolism (PE).

Two of the authors (S.W.H. and J.T.P.) reviewed each study independently according to previously established methodologic standards for diagnostic test research (Table 1) (10, 11). A "Level 1" designation was given to those studies judged to be of higher quality, fulfilling standards 1, 3, 6, and 7. Standards 1 and 7 were designed to exclude studies with potential for work-up and incorporation bias (12). Standards 3 and 6 were designed to exclude studies with poorly described and biased methods of patient assembly (13).

Our search strategy identified 52 studies. Twenty-three studies were designated Level 1, and their results form the basis for this review (14-36). We excluded those studies that failed to adequately describe the method of patient selection (Standard 3) (37-51), studied a population that did not represent the full spectrum of disease (Standard 6) (40, 42, 46, 52, 53), or based the performance of the reference standard on the results of the D-dimer testing (Standard 7) (40, 41, 44, 46, 48, 51, 52, 54-56).

DATA ABSTRACTION

Test characteristics of D-dimer assays from the Level 1 studies were abstracted by both reviewers, and discrepancies were resolved by consensus. In addition, we recorded details of the study population, including setting, participant recruitment, and patient demographics, as well as details of the study methods, including reference standard and assay performance. All studies used as reference standard ultrasound, venography, or a combination of the two. Because the algorithm for combining/ sequencing the two modalities was often poorly described, for analysis purposes we contrasted studies in which venography was performed on all patients vs studies utilizing ultrasound or an ultrasound/venography combination. "Patient mix" was defined as outpatient if the study specified that all study participants were enrolled either from outpatient offices or through emergency departments. The patient mix was classified as "mixed" if it included both outpatients and inpatients. Because the number of purely inpatient studies was small, they were combined with the mixed category for analysis purposes. "Study setting" was defined as the method of assembly of the patient cohort and classified as inpatient, outpatient, emergency department, or referred. The "referred setting" included study cohorts assembled on presentation to radiology departments, specialized venous thromboembolism (VTE) services, and vascular medicine departments.

STATISTICAL ANALYSIS

A marginal logistic regression with exchangeable correlation matrix was used to model the diagnostic odds ratio (DOR) as a function of D-dimer assay while controlling for the reference standard, sample size, overall prevalence of disease, and patient mix of each study (57, 58). The study setting was not clearly distinct from patient mix and was not included in the model. These covariates were chosen a priori. Because each study evaluated from 1 to 13 different assays on the same sample, a repeated-measures approach was used with each study defining a cluster of nonindependent assay measures. The actual model is illustrated below with the parameter subscripts indicating the indexing of assay (a) within study (s):

logit([Diagnosis.sub.sa]) = [[beta].sub.0] + [beta] * [Disease.sub.sa] + [[beta].sub.2] * [Assay.sub.sa] + [[beta].sub.3] * [Disease.sub.sa] * [Assay.sub.sa]

In this model, "Diagnosis" refers to the D-dimer assay result and "Disease" refers to the actual disease state (i.e., DVT or no DVT as assessed by the reference standard). The data are organized in a grouped binomial structure such that assay results are presented in separate rows of data for each disease state. In this format, the "Disease * Assay" interaction represents the covariate (Assay) effect on the log DOR. We expanded the model in a similar fashion to include the additional covariates (59).

Our model assessed the independent effect of each covariate on diagnostic discrimination as assessed by the DOR, the ratio of the positive and negative likelihood ratios. The DOR is a summary measure of the ability of the diagnostic test to discriminate between diseased and nondiseased individuals and reflects the trade-off between sensitivity and specificity. It is closely related to the ROC curve and the area under the curve (AUC). A DOR of 1 is equivalent to an AUC of 50%. The bigger the DOR, the bigger the AUC, and as the AUC grows to 100%, the DOR grows to infinity. It is an accepted and commonly used single measure of diagnostic accuracy in the comparison of diagnostic tests (60, 61). As an aggregate measure, it has utility as a dependent variable in the analysis of predictors of test performance and diminishes heterogeneity attributable to test threshold effects or other study factors that might inflate sensitivity or specificity. Similar to other metaanalytic methods (62, 63), this statistical modeling approach can derive summary ROC curves, test hypotheses, and provide effect estimates. However, it requires fewer assumptions, uses all available observations, and takes into account data clustering. The criterion used for statistical significance in this metaanalysis was a two-tailed a of 0.05. SAS (Ver. 8.01) was used to fit the statistical model (PROC GENMOD).

Results

Details of the 23 Level 1 studies, including reference standard, patient mix, exclusion criteria, demographics, and the numbers of standards satisfied, are presented in Table 2. These 23 studies were performed in 11 countries. Ultrasound was used as the reference standard in eight studies, venography in seven, and a combination of ultrasound and venography in eight. Many of the studies were small; the median number of patients was 132, and the range was 30-474 patients.

Exclusion criteria varied, with 8 studies excluding patients with a prior episodes of VTE, 13 studies excluding patients on anticoagulants, and 5 studies excluding patients with suspected PE. Only five studies explicitly examined D-dimer assays in patients with no previous history of VTE and not taking anticoagulants. Only one study did not adequately describe methods for independent, blinded test performance and comparison (Standard 2) (14). Although all Level 1 studies described the method of patient selection (Standard 3), three did not report simple demographic descriptions of their patients (Standard 4) (14, 20, 24). Results of eight studies were stratified by proximal and distal DVT (Standard 5) (19, 20, 25, 2830,34, 36). All 23 studies described methods (Standard 8) and reported results (Standard 9) for at least one D-dimer assay, but only 3 reported data on the reproducibility of the test results (Standard 10) (18, 26, 27).

Fourteen of 23 studies were performed exclusively in outpatient populations. The prevalence of DVT in these 14 studies ranged from 20% to 68%, with a median of 35%. We classified the other nine studies as mixed populations, with four exclusively inpatient, three including both inpatients and outpatients, and two not reporting the patient mix. The four studies performed exclusively on inpatient populations had a prevalence range of 28-69% and a median of 60%.

The studies evaluated 21 different D-dimer assays. Fourteen of the assays were included in more than one study, with 6 assays examined by four or more studies. The data on D-dimer assay performance are listed in Table 3. Only 27% of these evaluations were performed on samples with DVT prevalence <40%, and only two assays (SimpliRED and VIDAS) were examined by more than one such lower prevalence study.

The results of the multivariate analysis are presented in Table 4. The VIDAS assay DOR was used as the reference for test performance comparison. The multivariate analysis identified three assays with DORs that were significantly (P <0.05) different (all lower) from the VIDAS assay. The DORs of the other 17 assays were not significantly different from the VIDAS assay DOR, although most trended toward lower discriminant ability. Increasing prevalence of DVT in the study population was independently associated with poorer assay performance (P = 0.01), whereas the choice of venography as the reference standard was associated with better assay performance (P <0.005). Our model found no independent effect on assay performance by sample size or patient mix (outpatient or other).

Discussion

We reviewed 23 studies on the use of D-dimer assays in the diagnosis of lower extremity DVT. All of these studies met methodologic standards designed to limit bias and were published since the date of our last review. We found wide variation in the primary test characteristics of sensitivity and specificity. Because the marked heterogeneity among studies precluded pooling the results, we used a multivariate modeling approach to control for factors related to patient population and study design as well as for assay specific factors. Our metaanalysis found large variability in the DORs of the commercially available D-dimer assays. The differences in performance among the majority of the assays were not statistically significant. However, two factors, the prevalence of DVT in the study population and the choice of reference standard, were significantly associated with D-dimer assay performance. We believe that prevalence is a reflection of the clinical heterogeneity of patient groups included in the various studies. These two factors, in addition to the basic chemical properties unique to each assay, contribute to the marked variability in D-dimer test characteristics, and they help guide our recommendations for current clinical use and future research. These issues and others are discussed in the following sections.

DIFFERENCES AMONG D-DIMER ASSAYS

To be useful for clinical decision-making, D-dimer assays must be rapid, accurate, and reliable in the result range used for DVT diagnosis. Not all assays have these characteristics because of important differences in both assay methodology and in the monoclonal antibodies used (5, 9).

Assay methodology. The 21 D-dimer assays can be divided into one of six different method categories (Table 3), each of which has advantages and disadvantages. The microplate ELISA tests, because of their accuracy, quantitative results, and ability to measure low concentrations of D-dimer, have long been considered the standard against which newer assays should be compared. They have the disadvantage of taking 2-4 h to perform, limiting their clinical usefulness in the evaluation of thromboembolic diseases. In addition, they are labor-intensive and must be run in batches rather than on single samples.

The first-generation latex agglutination assays are based on the visible agglutination of antibody-coated latex particles. Positive samples may be serially diluted to provide a semiquantitative estimate of D-dimer concentrations. Although these assays are rapid and easy to perform, the results are qualitative, observer-dependent, and limited in their ability to detect minimally increased D-dimer concentrations. The second-generation latex immunoassays use the same basic technique as the firstgeneration but use a photometric analyzer to provide a quantitative measure of D-dimer and are generally able to reproducibly measure lower concentrations of D-dimer.

The membrane ELISAs use a monoclonal antibody that is chemically tagged to produce a visible color change in the presence of increased plasma concentrations of Ddimer. Both the NycoCard and the Instant I.A. assays can be interpreted manually, but an automated reflectometer can be used with the NycoCard assay to provide a quantitative result. These assays are usually performed in a centralized laboratory and can provide results within 20 min.

The SimpliRED D-dimer assay is an erythrocyte agglutination assay performed on whole blood. Similar to the first-generation latex assays, the qualitative presence or absence of visible agglutination is determined by an observer. Because the SimpliRED assay requires no plasma separation or specialized equipment, it can be performed at the point of care with results available in less than 2 min.

The VIDAS D-dimer assay is a fully automated rapid enzyme-linked fluorescent assay (ELFA) that produces quantitative results within 35 min. This assay uses the same biochemical principles as the microplate ELISA tests, but also has a single-sample cartridge system.

Assays of all method categories are capable of correctly identifying samples with very high concentrations of D-dimer and can also correctly give "negative" results for samples with very low or "normal" concentrations of D-dimer. However, a substantial proportion of patients with DVT have D-dimer concentrations that are only slightly increased (17, 20, 31, 36, 64). Many of the assays perform less well when required to accurately and reproducibly measure these slightly increased D-dimer concentrations (65), providing lower sensitivities and poor interand intraassay reproducibility (66, 67). This is especially true for those tests that require an observer to determine the presence of agglutination (17, 66).

Monoclonal antibodies. Although several latex agglutination assays use a common monoclonal antibody, most D-dimer assays use a unique monoclonal antibody. Each monoclonal antibody has a unique affinity for D-dimer fragments and unique testing properties. The affinity of the monoclonal antibody in each assay for D-dimer fragments directly determines the assay result, and each assay is affected to various degrees by factors in individual patient samples, such as the presence of anticoagulants (51, 68). The quantitative results of one D-dimer assay cannot be directly compared with those of a different assay. Consequently, an assay with a cut point of 300 jug/L fibrinogen equivalent units (FEU) is not necessarily more sensitive than an assay with a cut point of 500 jug/L FEU. Additionally, the units of measurement are not standardized across assays, with some assay results expressed as FEU and others as D-dimer units (one FEU is approximately one-half of a D-dimer unit).

There is a lack of consensus on cut points for individual assays (Table 3). Our review found results for the quantitative assays frequently reported at different cut points. In several studies the cut point was not determined a priori but instead was chosen retrospectively to maximize the sensitivity of the assay (19,25-27).

Recommended assays. Although the second-generation latex, the membrane ELISA, and the automated rapid ELFA methods are all rapid, simple, and quantitative, no single method is clearly best. Within each method category, individual assays demonstrate wide variability in sensitivity and specificity. Our multivariate analysis identified the VIDAS assay as statistically significantly better than three of the other assays, but it was indistinguishable from assays representing all of the other five method categories. Interestingly, the three assays that the VIDAS assay "outperformed" were from a mixture of methods: a membrane ELISA, a first-generation latex agglutination assay, and a second-generation latex agglutination assay. Unfortunately, our analysis lacked power to detect small to moderate differences among assays. Our metaanalysis has not led to a definite recommendation with regard to specific assay.

HETEROGENEITY OF PATIENT GROUPS

Our review shows that there were major differences among the patient groups. This is not surprising considering the variety of countries, settings, and exclusion criteria used in the 23 studies. The variation in DVT prevalence is the most striking measure of this clinical heterogeneity. Prevalence ranged from 20% to 69%, with a median of 37%. Although this prevalence is similar to that reported in previous research on DVT diagnosis with ultrasound testing (69, 70), it is considerably higher than other studies of outpatients referred for evaluation of suspected DVT (16, 71-73). Test performance often varies with patient characteristics and severity of disease. Patients with more severe disease are more likely to have positive tests than those with mild or early disease; healthy patients are more likely to have negative tests than those with significant comorbidities, particularly those with conditions known to cause false-positive tests (12, 74). In the case of D-dimer and DVT, studies in which patients with more extensive clots are enrolled are likely to find higher D-dimer concentrations and to report higher sensitivities. This was confirmed in our review by the eight studies that stratified results by extent of disease (19, 20, 25, 28-30, 34, 36). All reported higher sensitivities for patients with clot extending to the thigh than for clot limited to the calf (75).

D-Dimer assays are also often positive in common conditions other than DVT, including cancer, recent trauma, or surgery (76). Studies in which many patients with these and other conditions associated with increased D-dimer concentrations are enrolled are likely to find higher D-dimer concentrations in patients classified as free of DVT and to report lower specificities. In our metaanalysis, prevalence of DVT, a proxy for severity of illness and comorbidity of the patient population, was an independent predictor of test performance, with the relative DOR decreasing by 22% for every 10% increase in prevalence of DVT. D-Dimer concentrations also vary with duration of symptoms (77), use of anticoagulants (78, 79), and age of the patient (80). However, we were unable to explore the effect of clinical heterogeneity on test discrimination because few studies stratified their results by patient characteristics and none provided disaggregated patient-level data. This variation in test performance across population subgroups, or "spectrum effect", makes it difficult to predict D-dimer assay performance in clinical settings with a broad disease spectrum (81).

RANDOM VARIATION

A rule-out test must have a high sensitivity. However, estimates of sensitivity and specificity can be unstable with associated wide confidence intervals. For example, assuming point estimates of D-dimer assay sensitivity of 95% and specificity of 80%, a study would need a sample size of 395 individuals to ensure that the sensitivity 95% confidence interval lower limit would be >90%. Only three of the reviewed studies enrolled this number of patients (16,20, 29).

REFERENCE STANDARD BIAS

Our metaanalysis found that the choice of reference standard biased study results. In research on lower extremity DVT, the use of ultrasound has disadvantages as a reference standard because of its lower sensitivity in detecting calf DVT (69, 70). A reference standard, such as venography, that fully images both calf and thigh DVT will enable all patients with DVT to be properly counted as diseased. To the extent that ultrasound fails to image the calf veins, some diseased patients will be counted in the nondiseased category. Because the sensitivity of Ddimer for thigh DVT is higher than for isolated calf DVT, studies using ultrasound as the reference standard tend to have higher sensitivities, lower specificities, and higher negative predictive values than studies using venography (75). Our metaanalysis confirmed this bias. Reference standard was an independent predictor of test performance, with use of venography alone associated with a 2.7-fold higher DOR than reference standards using ultrasound. In future research, investigators must choose the reference standard with care.

To minimize heterogeneity of study results from multiple reference standards, we compared D-dimer test results with a single ultrasound and/or venogram, all tests performed at the same time. We did not adjust results when serial ultrasounds were done nor for added information from clinical outcomes on follow-up. These strategies, more often used in management trials, can partly compensate for not using venography by revealing the presence of previously undiagnosed calf DVT.

Conclusion

Our review of the accuracy literature does not support general use of D-dimer assays as a stand-alone rule-out test for DVT. We found wide variation in the sensitivities of D-dimer assays for diagnosing DVT and even wider variation in specificities. Test characteristics were frequently inconsistent with those of a good rule-out test. One-third of the time, reported sensitivities were <90%. There was wide variation in the key index of negative predictive value, frequently reported as <90%. Some D-dimer assays, such as the VIDAS assay, which had uniformly high sensitivities and was evaluated by nine studies, might appear well suited for a rule-out test role because of excellent sensitivities. However, often these high sensitivities were achieved by choosing the necessary cut points after the fact, which led to both lower specificities and uncertainty in the cut point that should be used.

Almost one-half of the assays had reported specificities of <0.50; the resulting number of false-positive D-dimer tests therefore often exceeded the total number of negative tests. Although the intent of D-dimer testing is to rule out DVT, this situation can lead to increased rather than decreased diagnostic evaluation (82). The easy availability of the D-dimer test may induce clinicians to order it more frequently than they would have ordered diagnostic imaging. Increased utilization of ultrasonography and/or venography may then result from the obligation to pursue the work-up of patients with positive D-dimer tests.

More research is needed to determine the place of D-dimer testing in clinical practice. This could be undertaken in two ways. The first is additional accuracy studies that, similar to the reports included in this review, compare selected D-dimer assay results with the best possible reference standard. Although we believe that venography is the preferable reference standard, it has been replaced in clinical practice by ultrasonography. Strategies to improve the performance of ultrasound as a reference standard, such as serial ultrasounds or clinical follow-up, are better than a single ultrasound but are likely to uncover only a minority of calf DVT. Future accuracy studies should collect sufficient clinical information about enrolled patients to allow for stratification by risk of DVT as well as for conditions known to cause an increased D-dimer test result other than DVT. Adequate numbers of patients should be enrolled to achieve sufficient statistical power for the primary analysis as well as subgroup analysis. This is the approach that was successful in establishing the role of the ventilation-perfusion lung scan for the diagnosis of PE (83).

The second approach is the use of management trials, the type of research that established the safety of the serial noninvasive testing approach to patients with suspected lower extremity DVT (84-86). In management trials, patients are evaluated through diagnostic algorithms with outcome assessment to determine safety and cost-effectiveness. This approach has the advantage of avoiding the difficulty of performing venography on all patients. Accuracy studies and management trials are not mutually exclusive. The two types of research are usually done in sequence, with management trials being performed once promising results have been obtained from accuracy studies. To date, most management trials for suspected DVT have begun with a probability assessment followed by D-dimer testing, with the results determining whether ultrasonography would be performed (55, 87-91). These studies show that outpatients or emergency department patients with a low pretest clinical probability of DVT as determined by a validated clinical prediction rule (73) and a normal D-dimer test result can be safely managed without diagnostic imaging and without anticoagulation. Other management trials suggest that repeated ultra sound examinations may be safely avoided in patients with an initially negative ultrasound and a negative D-dimer test result (92, 93). Interestingly, these studies have used various D-dimer assays from several different method categories, including second-generation latex agglutination assays (87, 91), membrane ELISA assays (94), and the erythrocyte agglutination assay (55, 90, 92, 93). To firmly establish the safety of these and other approaches, large trials are needed that have adequate numbers of patients at the ends of the branching pathways (95).

Although much progress has been made in our knowledge of D-dimer testing since our last systematic review in 1996, many of the same questions that we raised then remain unanswered. Published estimates of the accuracy of D-dimer assays have limited generalizability to clinical practice, and no one D-dimer assay has emerged clearly as the best. Management studies have established a role for several D-dimer assays in an outpatient or emergency department population when coupled with a validated clinical prediction rule. The clinical utility of D-dimer testing for the diagnosis of DVT in most other clinical situations is still uncertain.

This work was supported by a grant from the US Department of Health and Human Services, Health Resources and Services Administration, Academic Administrative Units in Primary Care (Grant 1 D12 HP 001 48-01).

Received January 20, 2004; accepted April 26, 2004.

Previously published online at DOI: 10.1373/clinchem.2004.031765

References

(1.) Becker DM, Philbrick JT, Bachhuber TL, Humphries JE. D-Dimer testing and acute venous thromboembolism. A shortcut to accurate diagnosis? Arch Intern Med 1996;156:939-46.

(2.) Bounameaux H, de Moerloose P, Perrier A, Miron MJ. D-Dimer testing in suspected venous thromboembolism: an update. Q J Med 1997;90:437-42.

(3.) Lee AY, Ginsberg JS. Laboratory diagnosis of venous thromboembolism. Baillieres Clin Haematol 1998;11:587-604.

(4.) Brill-Edwards P, Lee A. D-Dimer testing in the diagnosis of acute venous thromboembolism. Thromb Haemost 1999;82:688-94.

(5.) Dempfle CE. Use of D-dimer assays in the diagnosis of venous thrombosis. Semin Thromb Hemost 2000;26:631-41.

(6.) Anderson DR, Wells PS. D-Dimer for the diagnosis of venous thromboembolism. Curr Opin Hematol 2000;7:296-301.

(7.) Sadosty AT, Goyal DG, Boie ET, Chiu CK. Emergency department D-dimer testing. J Emerg Med 2001;21:423-9.

(8.) Michiels JJ, Kasbergen H, Oudega R, Van Der Graaf F, De Maeseneer M, Van Der Planken M, et al. Exclusion and diagnosis of deep vein thrombosis in outpatients by sequential noninvasive tools. Int Angiol 2002;21:9-19.

(9.) Kelly J, Rudd A, Lewis RR, Hunt BJ. Plasma D-dimers in the diagnosis of venous thromboembolism. Arch Intern Med 2002; 162:747-56.

(10.) Irwig L, Tosteson AN, Gatsonis C, Lau J, Colditz G, Chalmers TC, et al. Guidelines for meta-analyses evaluating diagnostic tests. Ann Intern Med 1994;120:667-76.

(11.) Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig LM, et al. Towards complete and accurate reporting of studies of diagnostic accuracy: the STARD Initiative. Ann Intern Med 2003;138:40-4.

(12.) Ransohoff DF, Feinstein AR. Problems of spectrum and bias in evaluating the efficacy of diagnostic tests. N Engl J Med 1978; 299:926-30.

(13.) Lijmer JG, Mol BW, Heisterkamp S, Bonsel GJ, Prins MH, van der Meulen JH, et al. Empirical evidence of design-related bias in studies of diagnostic tests. JAMA 1999;282:1061-6.

(14.) Brenner B, Pery M, Lanir N, Jabareen A, Markel A, Kaftori JK, et al. Application of a bedside whole blood D-dimer assay in the diagnosis of deep vein thrombosis. Blood Coagul Fibrinolysis 1995;6:219-22.

(15.) Khaira HS, Mann J. Plasma D-dimer measurement in patients with suspected DVT--a means of avoiding unnecessary venography. Eur J Vase Endovasc Surg 1998;15:235-8.

(16.) Perrier A, Desmarais S, Miron MJ, de Moerloose P, Lepage R, Slosman D, et al. Non-invasive diagnosis of venous thromboembolism in outpatients. Lancet 1999;353:190-5.

(17.) Carter CJ, Serrano K, Breen DJ, Buckley AR, Devine DV. Rapid fibrin D-dimer tests for deep venous thrombosis: factors affecting diagnostic utility. J Emerg Med 1999;17:605-10.

(18.) D'Angelo A, D'Alessandro G, Thomassini L, Pittet JL, Dupuy G, Crippa L. Evaluation of a new rapid quantitative D-dimer assay in patients with clinically suspected deep vein thrombosis. Thromb Haemost 1996;75:412-6.

(19.) Elias A, Aptel I, Hue B, Chale JJ, Nguyen F, Cambus JP, et al. D-Dimer test and diagnosis of deep vein thrombosis: a comparative study of 7 assays. Thromb Haemost 1996;76:518-22.

(20.) Escoffre-Barbe M, Oger E, Leroyer C, Grimaux M, Le Moigne E, Nonent M, et al. Evaluation of a new rapid D-dimer assay for clinically suspected deep venous thrombosis (Liatest D-dimer). Am J Clin Pathol 1998;109:748-53.

(21.) Farrell S, Hayes T, Shaw M. A negative SimpliRED D-dimer assay result does not exclude the diagnosis of deep vein thrombosis or pulmonary embolus in emergency department patients. Ann Emerg Med 2000;35:121-5.

(22.) Fiessinger JN, Heron E, Jacq F, Ranee A, Emmerich J. Rapid blood test for the exclusion of venous thromboembolism in symptomatic outpatients. Thromb Haemost 1997;77:1042-3.

(23.) Funfsinn N, Caliezi C, Biasiutti FD, Korte W, Z'Brun A, Baumgartner I, et al. Rapid D-dimer testing and pre-test clinical probability in the exclusion of deep venous thrombosis in symptomatic outpatients. Blood Coagul Fibrinolysis 2001;12:165-70.

(24.) Harper P, Marson C, Grimmer A, Monahan K, Humm G, Baker B. The rapid whole blood agglutination d-dimer assay has poor sensitivity for use as an exclusion test in suspected deep vein thrombosis. N Z Med J 2001;114:61-4.

(25.) Janssen MC, Heebels AE, de Metz M, Verbruggen H, Wollersheim H, Janssen S, et al. Reliability of five rapid D-dimer assays compared to ELISA in the exclusion of deep venous thrombosis. Thromb Haemost 1997;77:262-6.

(26.) Legnani C, Pancani C, Palareti G, Guazzaloca G, Fortunato G, Grauso F, et al. Comparison of new rapid methods for D-dimer measurement to exclude deep vein thrombosis in symptomatic outpatients. Blood Coagul Fibrinolysis 1997;8:296-302.

(27.) Legnani C, Pancani C, Palareti G, Guazzaloca G, Coccheri S. Contribution of a new, rapid, quantitative and automated method for D-dimer measurement to exclude deep vein thrombosis in symptomatic outpatients. Blood Coagul Fibrinolysis 1999;10:6974.

(28.) Lennox AF, Delis KT, Serunkuma S, Zarka ZA, Daskalopoulou SE, Nicolaides AN. Combination of a clinical risk assessment score and rapid whole blood D-dimer testing in the diagnosis of deep vein thrombosis in symptomatic patients. J Vase Surg 1999;30: 794-804.

(29.) Leroyer C, Escoffre M, Le Moigne E, Grimaux M, Cagnioncle 0, Oger E, et al. Diagnostic value of a new sensitive membrane based technique for instantaneous D-dimer evaluation in patients with clinically suspected deep venous thrombosis. Thromb Haemost 1997;77:637-40.

(30.) Permpikul C, Chantavatcharakorn K, Bouranasompop C, Chaiyasoot W. Whole blood agglutination D-dimer test for the diagnosis of deep vein thrombosis. J Med Assoc Thai 2000;83:732-6.

(31.) Sadouk M, Desmarais S, Patenaude JV, Lepage R. Comparison of diagnostic performance of three new fast D-dimer assays in the exclusion of deep vein thrombosis. Clin Chem 2000;46:286-7.

(32.) Scarano L, Bernardi E, Prandoni P, Sardella C, Rossi L, Carraro P, et al. Accuracy of two newly described D-dimer tests in patients with suspected deep venous thrombosis. Thromb Res 1997;86: 93-9.

(33.) van der Graaf F, van den Borne H, van der Kolk M, de Wild PJ, Janssen GWT, van Uum SHM. Exclusion of deep venous thrombosis with D-dimer testing. Thromb Haemost 2000;83:191-8.

(34.) Wells PS, Brill-Edwards P, Stevens P, Panju A, Patel A, Douketis J, et al. A novel and rapid whole-blood assay for D-Dimer in patients with clinically suspected deep vein thrombosis. Circulation 1995; 91:2184-7.

(35.) Wells PS. Application of a diagnostic clinical model for the management of hospitalized patients with suspected deep-vein thrombosis. Thromb Haemost 1999;81:493-7.

(36.) Wijns W, Daoud N, Droeshout I, Pradier O, Wautrecht JC, Golzarian J, et al. Evaluation of two D-dimer assays in the diagnosis of venous thromboembolism. Acta Clin Belg 1998;53:270-4.

(37.) Killick SB, Pentek PG, Mercieca JE, Clarke MF, Bevan DH. Comparison of immunofiltration assay of plasma D-dimer with diagnostic imaging in deep vein thrombosis. Br J Haematol 1997;96:846-9.

(38.) Cornuz J, Ghali WA, Hayoz D, Stoianov R, Depairon M, Yersin B. Clinical prediction of deep venous thrombosis using two risk assessment methods in combination with rapid quantitative Ddimer testing. Am J Med 2002;112:198-203.

(39.) Wildberger JE, Vorwerk D, Kilbinger M, Piroth W, Hunter DW, Wienert V, et al. Bedside testing (SimpliRED) in the diagnosis of deep vein thrombosis. Evaluation of 250 patients. Invest Radiol 1998;33:232-5.

(40.) Knecht MF, Heinrich F. Clinical evaluation of an immunoturbidimetric D-dimer assay in the diagnostic procedure of deep vein thrombosis and pulmonary embolism. Thromb Res 1997;88: 413-7.

(41.) Mayer W, Hirschwehr R, Hippmann G, Odpadlik H, Bayer P, Partsch H. Whole-blood immunoassay (SimpliRED) versus plasma immunoassay (NycoCard) for the diagnosis of clinically suspected deep vein thrombosis. Vasa 1997;26:97-101.

(42.) Lindahl TL, Lundahl TH, Ranby M, Fransson SG. Clinical evaluation of a diagnostic strategy for deep venous thrombosis with exclusion by low plasma levels of fibrin degradation product D-dimer. Scand J Clin Lab Invest 1998;58:307-16.

(43.) Bradley M, Bladon J, Barker H. D-Dimer assay for deep vein thrombosis: its role with colour Doppler sonography. Clin Radiol 2000;55:525-7.

(44.) Freyburger G, Trillaud H, Labrouche S, Gauthier P, Javorschi S, Grenier N. Rapid ELISA D-dimer testing in the exclusion of venous thromboembolism in hospitalized patients. Clin Appl Thromb Hemost 2000;6:77-81.

(45.) Gosselin RC, Owings JT, Utter GH, Jacoby RC, Larkin EC. A new method for measuring D-dimer using immunoturbidometry: a study of 255 patients with suspected pulmonary embolism and deep vein thrombosis. Blood Coagul Fibrinolysis 2000;11:715-21.

(46.) Hein-Rasmussen R, Tuxen CD, Wiinberg N. Diagnostic value of the Nycocard, Nycomed D-dimer assay for the diagnosis of deep venous thrombosis and pulmonary embolism: a retrospective study. Thromb Res 2000;100:287-92.

(47.) Villa P, Ferrando F, Serra J, Faus H, Mira Y, Vaya A, et al. Quantification of D-dimer using a new fully automated assay: its application for the diagnosis of deep vein thrombosis. Haematologica 2000;85:520-4.

(48.) Bates SM, Grand'Maison A, Johnston M, Naguit I, Kovacs MJ, Ginsberg JS. A latex D-dimer reliably excludes venous thromboembolism. Arch Intern Med 2001;161:447-53.

(49.) Bucek RA, Quehenberger P, Feliks I, Handler S, Reiter M, Minar E. Results of a new rapid d-dimer assay (cardiac d-dimer) in the diagnosis of deep vein thrombosis. Thromb Res 2001;103:1723.

(50.) Shitrit D, Heyd J, Raveh D, Rudensky B. Diagnostic value of the D-dimer test in deep vein thrombosis: improved results by a new assay method and by using discriminate levels. Thromb Res 2001;102:125-31.

(51.) Schutgens RE, Esseboom EU, Haas FJ, Nieuwenhuis HK, Biesma DH. Usefulness of a semiquantitative D-dimer test for the exclusion of deep venous thrombosis in outpatients. Am J Med 2002;112:617-21.

(52.) Anderson DR, Wells PS, Stiell I, MacLeod B, Simms M, Gray L, et al. Management of patients with suspected deep vein thrombosis in the Emergency Department: combining use of a clinical diagnosis model with D-dimer testing. J Emerg Med 2000;19:225-30.

(53.) Wahlander K, Tengborn L, Hellstrom M, Olmarker AH, Peterson LE, Stigendal L, et al. Comparison of various D-dimer tests for the diagnosis of deep venous thrombosis. Blood Coagul Fibrinolysis 1999;10:121-6.

(54.) AguilarC, Martinez A, Martinez A, Del Rio C, Vazquez M, Rodriguez FJ. Diagnostic value of d-dimer in patients with a moderate pretest probability of deep venous thrombosis. Br J Haematol 2002;118: 275-7.

(55.) Kearon C, Ginsberg JS, Douketis J, Crowther M, Brill-Edwards P, Weitz JI, et al. Management of suspected deep venous thrombosis in outpatients by using clinical assessment and D-dimer testing. Ann Intern Med 2001;135:108-11.

(56.) LeBlanche AF, Siguret V, Settegrana C, Bohus S, de Chermont ELM, Andreux J-P, et al. Ruling out acute deep vein thrombosis by ELISA plasma D-dimer assay versus ultrasound in inpatients more than 70 years old. Angiology 1999;50:873-82.

(57.) Diggle P, Heagerty P, Kung-Yee L, Zeger S. Analysis of longitudinal data, 2nd ed. New York: Oxford University Press, 2002:379pp.

(58.) Hosmer DW, Lemeshow S. Applied logistic regression, 2nd ed. New York: Wiley, 2000:373pp.

(59.) Siadaty MS, Philbrick JT, Heim SW, Schectman JM. Comparing performance of competing diagnostic tests in partially overlapping papers: a new meta-analytic method. J Clin Epidemiol 2004;in press.

(60.) Irwig L, Macaskill P, Glasziou P, Fahey M. Meta-analytic methods for diagnostic test accuracy. J Clin Epidemiol 1995;48:119-30.

(61.) Lijmer JG, Bossuyt PM, Heisterkamp SH. Exploring sources of heterogeneity in systematic reviews of diagnostic tests. Stat Med 2002;21:1525-37.

(62.) Littenberg B, Moses LE. Estimating diagnostic-accuracy from multiple conflicting reports: a new meta-analytic method. Med Decis Making 1993;13:313-21.

(63.) Midgette AS, Stukel TA, Littenberg B. A meta-analytic method for summarizing diagnostic test performances: receiver-operatingcharacteristic summary point estimates. Med Decis Making 1993;13:253-7.

(64.) Freyburger G, Trillaud H, Labrouche S, Gauthier P, Javorschi S, Bernard P, et al. D-Dimer strategy in thrombosis exclusion--a gold standard study in 100 patients suspected of deep venous thrombosis or pulmonary embolism: 8 DD methods compared. Thromb Haemost 1998;79:32-7.

(65.) de Maat MP, Meijer P, Nieuwenhuizen W, Haverkate F, Kluft C. Performance of semiquantitative and quantitative D-dimer assays in the ECAT external quality assessment program. Semin Thromb Hemost 2000;26:625-30.

(66.) de Monye W, Huisman MV, Pattynama PM. Observer dependency of the SimpliRed D-dimer assay in 81 consecutive patients with suspected pulmonary embolism. Thromb Res 1999;96:293-8.

(67.) van Beek EJ, van den Ende B, Berckmans RJ, van der Heide YT, Brandjes DP, Sturk A, et al. A comparative analysis of D-dimer assays in patients with clinically suspected pulmonary embolism. Thromb Haemost 1993;70:408-13.

(68.) Couturaud F, Kearon C, Bates SM, Ginsberg JS. Decrease in sensitivity of D-dimer for acute venous thromboembolism after starting anticoagulant therapy. Blood Coagul Fibrinolysis 2002; 13:241-6.

(69.) Becker DM, Philbrick JT, Abbitt PL. Real-time ultrasonography for the diagnosis of lower extremity deep venous thrombosis. The wave of the future? Arch Intern Med 1989;149:1731-4.

(70.) White RH, McGahan JP, Daschbach MM, Hartling RP. Diagnosis of deep-vein thrombosis using duplex ultrasound. Ann Intern Med 1989;111:297-304.

(71.) Wells PS, Hirsh J, Anderson DR, Lensing AW, Foster G, Kearon C, et al. Accuracy of clinical assessment of deep-vein thrombosis. Lancet 1995;345:1326-30.

(72.) Anderson DR, Wells PS, Stiell I, MacLeod B, Simms M, Gray L, et al. Thrombosis in the emergency department: use of a clinical diagnosis model to safely avoid the need for urgent radiological investigation. Arch Intern Med 1999;159:477-82.

(73.) Wells PS, Anderson DR, Bormanis J, Guy F, Mitchell M, Gray L, et al. Value of assessment of pretest probability of deep-vein thrombosis in clinical management. Lancet 1997;350:1795-8.

(74.) Hlatky MA, Pryor DB, Harrell FE Jr, Califf RM, Mark DB, Rosati RA. Factors affecting sensitivity and specificity of exercise electrocardiography. Multivariable analysis. Am J Med 1984;77:64-71.

(75.) Philbrick JT, Heim S. The d-dimer test for deep venous thrombosis: gold standards and bias in negative predictive value. Clin Chem 2003;49:570-4.

(76.) Raimondi P, Bongard O, de Moerloose P, Reber G, Waldvogel F, Bounameaux H. D-Dimer plasma concentration in various clinical conditions: implication for the use of this test in the diagnostic approach of venous thromboembolism. Thromb Res 1993;69: 125-30.

(77.) Heit JA, Minor TA, Andrews JC, Larson DR, Li H, Nichols WL. Determinants of plasma fibrin D-dimer sensitivity for acute pulmonary embolism as defined by pulmonary angiography. Arch Pathol Lab Med 1999;123:235-40.

(78.) Markers of hemostatic system activation in acute deep venous thrombosis-evolution during the first days of heparin treatment. The DVTENOX Study Group. Thromb Haemost 1993;70:909-14.

(79.) Lip GY, Lip PL, Zarifis J, Watson RD, Bareford D, Lowe GD, et al. Fibrin D-dimer and /3-thromboglobulin as markers of thrombogenesis and platelet activation in atrial fibrillation. Effects of introducing ultra-low-dose warfarin and aspirin. Circulation 1996;94:42531.

(80.) Righini M, GoehringC, Bounameaux H, Perrier A. Effects of age on the performance of common diagnostic tests for pulmonary embolism. Am J Med 2000;109:357-61.

(81.) Mulherin SA, Miller WC. Spectrum bias or spectrum effect? Subgroup variation in diagnostic test evaluation. Ann Intern Med 2002;137:598-602.

(82.) Goldstein NM, Kollef MH, Ward S, Gage BF. The impact of the introduction of a rapid D-dimer assay on the diagnostic evaluation of suspected pulmonary embolism. Arch Intern Med 2001;161: 567-71.

(83.) Value of the ventilation/perfusion scan in acute pulmonary embolism. Results of the prospective investigation of pulmonary embolism diagnosis (PIOPED). The PIOPED Investigators. JAMA 1990; 263:2753-9.

(84.) Huisman MV, Buller HR, ten Cate JW, Vreeken J. Serial impedance plethysmography for suspected deep venous thrombosis in outpatients. The Amsterdam General Practitioner Study. N Engl J Med 1986;314:823-8.

(85.) Huisman MV, Buller HR, ten Cate JW, Heijermans HS, van der Laan J, van Maanen DJ. Management of clinically suspected acute venous thrombosis in outpatients with serial impedance plethysmography in a community hospital setting. Arch Intern Med 1989;149:511-3.

(86.) Hull RD, Hirsh J, Carter CJ, Jay RM, Ockelford PA, Buller HR, et al. Diagnostic efficacy of impedance plethysmography for clinically suspected deep-vein thrombosis. A randomized trial. Ann Intern Med 1985;102:21-8.

(87.) Schutgens RE, Ackermark P, Haas FJ, Nieuwenhuis HK, Peltenburg HG, Pijlman AH, et al. Combination of a normal D-dimer concentration and a non-high pretest clinical probability score is a safe strategy to exclude deep venous thrombosis. Circulation 2003;107:593-7.

(88.) Wells PS, Anderson DR, Rodger M, Forgie M, Kearon C, Dreyer J, et al. Evaluation of D-dimer in the diagnosis of suspected deepvein thrombosis. N Engl J Med 2003;349:1227-35.

(89.) Anderson DR, Kovacs MJ, Kovacs G, Stiell I, Mitchell M, Khoury V, et al. Combined use of clinical assessment and d-dimer to improve the management of patients presenting to the emergency

department with suspected deep vein thrombosis (the EDITED Study). J Thromb Haemost 2003;1:645-51.

(90.) Janes S, Ashford N. Use of a simplified clinical scoring system and D-dimer testing can reduce the requirement for radiology in the exclusion of deep vein thrombosis by over 20%. Br J Haematol 2001;112:1079-82.

(91.) Bates SM, Kearon C, Crowther M, Linkins L, O'Donnell M, Douketis J, et al. A diagnostic strategy involving a quantitative latex D-dimer assay reliably excludes deep venous thrombosis. Ann Intern Med 2003;138:787-94.

(92.) Tick LW, Ton E, van Voorthuizen T, Hovens MM, Leeuwenburgh I, Lobatto S, et al. Practical diagnostic management of patients with clinically suspected deep vein thrombosis by clinical probability test, compression ultrasonography, and D-dimer test. Am J Med 2002;113:630-5.

(93.) Kraaijenhagen RA, Piovella F, Bernardi E, Verlato F, Beckers EA, Koopman MM, et al. Simplification of the diagnostic management of suspected deep vein thrombosis. Arch Intern Med 2002;162: 907-11.

(94.) Bernardi E, Prandoni P, Lensing AW, Agnelli G, Guazzaloca G, Scannapieco G, et al. D-dimer testing as an adjunct to ultrasonography in patients with clinically suspected deep vein thrombosis: prospective cohort study. The Multicentre Italian D-dimer Ultrasound Study Investigators Group. BMJ 1998;317:1037-40.

(95.) Hull RD, Stein PD, Ghali WA, Cornuz J. Diagnostic algorithms for deep vein thrombosis: work in progress. Am J Med 2002;113: 687-8.

(4) Nonstandard abbreviations: DVT, deep venous thrombosis; PE, pulmonary embolism; VTE, venous thromboembolism; DOR, diagnostic odds ratio; AUC, area under curve; ELFA, enzyme-linked fluorescent assay; and FEU, fibrinogen equivalent unit(s).

Steven W. Heim, [1,3]* Joel M. Schectman, [2] Mir S. Siadaty, [3] and John T. Philbrick [2,3] Departments of [1] Family Medicine, [2] Internal Medicine, and [3] Health Evaluation Sciences, University of Virginia Health System, Charlottesville, VA.

* Address correspondence to this author at: University of Virginia Health System, Department of Family Medicine, PO Box 800729, Charlottesville, VA 22908. Fax 434-243-4800; e-mail heims@virginia.edu.
Table 1. Standards for research on D-dimer diagnostic testing. (a)

Standard

1 Reference standard D-Dimer results should be
 compared with an accepted
 reference standard, including
 venography or compression
 ultrasonography

2 Independent D-Dimer and reference standard
 interpretation results should be interpreted
 of test results independently

3 Patient selection The patient selection process
 should be described in enough
 detail to allow a similar group
 of patients to be enrolled in
 subsequent studies

4 Patient Descriptions of the study
 characteristics population should, at a
 minimum, include simple
 demographic characteristics

5 Disease severity Results of tests should be
 stratified by the extent and
 severity of DVT, including at
 least calf and thigh location

6 Patient spectrum There should not be exclusions
 that would prevent the patients
 studied from representing the
 complete spectrum of patients
 with suspected DVT, including
 those with and without disease

7 Diagnostic process The decision to perform the
 reference standard test for
 each individual should have
 been made independently of the
 D-dimer result

8 Test details Test descriptions (whether for
 D-dimer or the reference
 standard) should be
 sufficiently detailed to permit
 replication.

9 Test results Sensitivity and specificity (or
 the raw data for these
 calculations) for at least one
 cutoff point should be
 presented

10 Test reproducibility The reproducibility and
 interpretation of D-dimer
 results should be evaluated in
 a setting where this test is
 likely to be used

(a) Adapted from Becker et al. (1).

Table 2. Study characteristics.

 Reference
Study, year (Ref.) Country standard(s) Setting

Brenner et al., 1995 (14) Israel US (b) Referred

Carter et al., 1999 (17) Canada US Referred

D'Angelo et al., 1996 (18) Italy US Referred

Elias et al., 1996 (19) France US Referred

Escoffre-Barbe et al., 1998 France US, Venog Inpatient
(20)

Farrell et al., 2000 (21) United US ED
 States
Fiessinger et al., 1997 (22) France US Referred

Funfsinn et al., 2001 (23) Switzerland US, Venog Referred

Harper et al., 2001 (24) New Zealand US, Venog ED

Janssen et al., 1997 (25) The US, Venog ED;
 Netherlands outpatient

Khaira and Mann, 1998 United US, Venog Referred
(15) Kingdom

Legnani et al., 1997 (26) Italy Venog Referred

Legnani et al., 1999 (27) Italy Venog Referred

Lennox et al., 1999 (28) United US Referred
 Kingdom
Leroyer et al., 1997 (29) France US, Venog Inpatient

Permpikul et al., 2000 Thailand Venog Inpatient
(30)

Perrier et al., 1999 (16) Switzerland, US, Venog ED;
 Canada outpatient

Sadouk et al., 2000 (31) Canada US ED;
 outpatient

Scarano et al., 1997 (32) Italy Venog Referred

van der Graaf et al., 2000 The Venog ED
(33) Netherlands

Wells et al., 1995 (34) Canada Venog, Referred
 IPG

Wells, 1999 (35) Canada US, Venog Referred

Wijns et al., 1998 (36) Belgium Venog Referred

Study, year (Ref.) Exclusion criteria

Brenner et al., 1995 (14) Previous VTE; symptoms <7 days;
 postoperative <10 days;
 thrombolysis <30 days

Carter et al., 1999 (17) Previous VTE; postoperative <21 days

D'Angelo et al., 1996 (18) Symptoms <15 days

Elias et al., 1996 (19) Previous VTE

Escoffre-Barbe et al., 1998 Symptoms >7 days; suspected PE;
(20) caval filters

Farrell et al., 2000 (21) NR

Fiessinger et al., 1997 (22) Anticoagulation; pregnancy

Funfsinn et al., 2001 (23) Anticoagulation; pregnancy;
 hospitalized within 3 days
Harper et al., 2001 (24) Anticoagulation

Janssen et al., 1997 (25) NR

Khaira and Mann, 1998 NR
(15)

Legnani et al., 1997 (26) Anticoagulation; symptoms >30 days

Legnani et al., 1999 (27) Anticoagulation; previous VTE;
 symptoms >30 days;
 contraindications to Venog
Lennox et al., 1999 (28) Anticoagulation; previous VTE;
 symptoms >30 days; suspected PE
Leroyer et al., 1997 (29) Symptoms >7 days; suspected PE;
 caval filters
Permpikul et al., 2000 Symptoms >14 days; DIC; previous
(30) trauma, sepsis, surgery, or arterial
 embolism; MI <14 days
Perrier et al., 1999 (16) Anticoagulation; expected survival <3
 months; hospitalized >24 h before
 symptom onset
Sadouk et al., 2000 (31) Anticoagulation; recent surgery;
 expected survival <3 months;
 hospitalized >24 h before symptom
 onset
Scarano et al., 1997 (32) Anticoagulation; previous VTE;
 suspected PE; pregnancy;
 contraindications to Venog; cancer
van der Graaf et al., 2000 Anticoagulation; postoperative <8 days;
(33) contraindications to Venog
Wells et al., 1995 (34) Anticoagulation; previous VTE;
 contraindication to Venog
Wells, 1999 (35) Anticoagulation; previous VTE;
 suspected PE; asymptomatic;
 imminent death
Wijns et al., 1998 (36) Anticoagulation

 Patients, Prevalence, Outpatients,
Study, year (Ref.) n % %

Brenner et al., 1995 (14) 86 58 100

Carter et al., 1999 (17) 199 28 74

D'Angelo et al., 1996 (18) 103 21 NR

Elias et al., 1996 (19) 171 44 17

Escoffre-Barbe et al., 1998 464 60 0
(20)

Farrell et al., 2000 (21) 173 33 100

Fiessinger et al., 1997 (22) 30 23 100

Funfsinn et al., 2001 (23) 106 44 100

Harper et al., 2001 (24) 235 22 100

Janssen et al., 1997 (25) 132 68 100

Khaira and Mann, 1998 79 37 100
(15)

Legnani et al., 1997 (26) 81 52 100

Legnani et al., 1999 (27) 99 39 100

Lennox et al., 1999 (28) 200 23 41

Leroyer et al., 1997 (29) 448 59 0

Permpikul et al., 2000 65 69 0
(30)

Perrier et al., 1999 (16) 474 23 100

Sadouk et al., 2000 (31) 177 20 100

Scarano et al., 1997 (32) 126 24 100

van der Graaf et al., 2000 99 51 100
(33)
Wells et al., 1995 (34) 214 25 100

Wells, 1999 (35) 150 28 0

Wijns et al., 1998 (36) 74 43 NR

 Mean
 age, Women, Standards
Study, year (Ref.) years % satisfied (a)

Brenner et al., 1995 (14) NR NR 1, 3, 6-9

Carter et al., 1999 (17) 58 58 1-4, 6-9

D'Angelo et al., 1996 (18) 59 63 1-4, 6-10

Elias et al., 1996 (19) 64 54 1-9

Escoffre-Barbe et al., 1998 63 NR 1-3, 5-9
(20)

Farrell et al., 2000 (21) 55 59 1-9

Fiessinger et al., 1997 (22) 57 NR 1-4, 6-9

Funfsinn et al., 2001 (23) 56 51 1-4, 6-9

Harper et al., 2001 (24) NR NR 1-3, 6-9

Janssen et al., 1997 (25) 59 63 1-9

Khaira and Mann, 1998 NR 61 1-4, 6-9
(15)

Legnani et al., 1997 (26) NR 44 1-10

Legnani et al., 1999 (27) NR 57 1-10

Lennox et al., 1999 (28) 58 63 1-9

Leroyer et al., 1997 (29) 62 53 1-9

Permpikul et al., 2000 57 58 1-9
(30)

Perrier et al., 1999 (16) NR 62 1-4, 6-9

Sadouk et al., 2000 (31) 56 66 1-4, 6-9

Scarano et al., 1997 (32) 62 55 1-4, 6-9

van der Graaf et al., 2000 57 62 1-4, 6-9
(33)

Wells et al., 1995 (34) 56 62 1-9

Wells, 1999 (35) 64 51 1-4, 6-9

Wijns et al., 1998 (36) 57 66 1-9

(a) See Table 1 for description of standards.

(b) US, ultrasonography; NR, not reported; Venog, venography;
ED, emergency department; DIC, disseminated intravascular
coagulopathy; MI, myocardial infarction; IPG, impedence
plethysmography.

Table 3. D-Dimer assay characteristics.

Assay (manufacturer) Study, year (Ref.)

First-generation latex
 agglutination assays
 D-dimer test (Diagnostica Stago) Elias et al., 1996 (19)
 Dimertest (Agen Biomedical) Janssen et al., 1997 (25)
 Dimertest II (Agen Biomedical) Brenner et al., 1995 (14)
 Minutex (Biopool) van der Graaf et al., 2000 (33)
 Legnani et al., 1997 (26)
 Janssen et al., 1997 (25)
 Nephelotex (Biopool) Legnani et al., 1997 (26)
Second-generation latex
 agglutination assays
 Auto Dimertest (Agen Biomedical) Funfsinn et al., 2001 (23)
 BC D-Dimer (Dade Behring) Legnani et al., 1999 (27)
 van der Graaf et al., 2000 (33)
 IL Test (Instrumentation Harper et al., 2001 (24)
 Laboratory, SpA) van der Graaf et al., 2000 (33)
 Liatest (Diagnostica Stago) van der Graaf et al., 2000 (33)
 Escoffre-Barbe et al., 1998 (20)
 LPIA (Mitsubishi Kasei) Legnani et al., 1997 (26)
 Tinaquant (Roche Diagnostics) Sadouk et al., 2000 (31)
 Funfsinn et al., 2001 (23)
 van der Graaf et al., 2000 (33)
 Janssen et al., 1997 (25)
 Turbiquant (Dade Behring) Sadouk et al., 2000 (31)
 van der Graaf et al., 2000 (33)
Membrane ELISAs
 Instant I.A. (Diagnostica Stago) Scarano et al., 1997 (32)
 Wijns et al., 1998 (36)
 Elias et al., 1996 (19)
 van der Graaf et al., 2000 (33)
 Legnani et al., 1997 (26)
 Leroyer et al., 1997 (29)
 NycoCard (Nycomed Pharma AS) Scarano et al., 1997 (32)
 Khaira and Mann, 1998 (15)
 Elias et al., 1996 (19)
 van der Graaf et al., 2000 (33)
 Legnani et al., 1997 (26)
Erythrocyte agglutination assay
 SimpliRED (Agen Biomedical) Harper et al., 2001 (24)
 Fiessinger et al., 1997 (22)
 Lennox et al., 1999 (28)
 Wells et al., 1995 (34)
 Carter et al., 1999 (17)
 Wells, 1999 (35)
 Farrell et al., 2000 (21)
 van der Graaf et al., 2000 (33)
 Brenner et al., 1995 (14)
 Permpikul et al., 2000 (30)
 Janssen et al., 1997 (25)
Automated rapid ELFA
 VIDAS (bioMerieux) D'Angelo et al., 1996 (18)
 Perrier et al., 1999 (16)c
 Legnani et al., 1999 (27)
 Wijns et al., 1998 (36)
 Elias et al., 1996 (19)
 Funfsinn et al., 2001 (23)
 van der Graaf et al., 2000 (33)
 Legnani et al., 1997 (26)
 Janssen et al., 1997 (25)
Microplate ELISAs
 Asserachrom (Diagnostica Stago) Wijns et al., 1998 (36)
 Elias et al., 1996 (19)
 Funfsinn et al., 2001 (23)
 van der Graaf et al., 2000 (33)
 Leroyer et al., 1997 (29)
 Dimertest EIA (Agen Biomedical) Brenner et al., 1995 (14)
 Dimertest GOLD EIA (Agen Legnani et al., 1999 (27)
 Biomedical) Legnani et al., 1997 (26)
 Enzygnost (Dade Behring) Scarano et al., 1997 (32)
 Elias et al., 1996 (19)
 van der Graaf et al., 2000 (33)
 Fibrinostika (Organon Teknika) Elias et al., 1996 (19)
 van der Graaf et al., 2000 (33)
 Janssen et al., 1997 (25)

 Cut point,
 [micro]g/L Prevalence,
Assay (manufacturer) (units) %

First-generation latex
 agglutination assays
 D-dimer test (Diagnostica Stago) [+ or -] (b) 44
 Dimertest (Agen Biomedical) [+ or -] 68
 Dimertest II (Agen Biomedical) [+ or -] 58
 Minutex (Biopool) [+ or -] 51
 [+ or -] 52
 [+ or -] 68
 Nephelotex (Biopool) [+ or -] 52
Second-generation latex
 agglutination assays
 Auto Dimertest (Agen Biomedical) 100 (FEU) 44
 BC D-Dimer (Dade Behring) 155 (NR) 39
 300 (DD) 51
 IL Test (Instrumentation 250 (DD) 22
 Laboratory, SpA) 250 (DD) 51
 Liatest (Diagnostica Stago) 500 (FEU) 51
 400 (FEU) 60
 LPIA (Mitsubishi Kasei) 1100 (FEU) 52
 Tinaquant (Roche Diagnostics) 500 (FEU) 20
 500 (FEU) 44
 500 (FEU) 51
 500 (FEU) 68
 Turbiquant (Dade Behring) 250 (DD) 20
 300 (DD) 51
Membrane ELISAs
 Instant I.A. (Diagnostica Stago) [+ or -] 24
 [+ or -] 43
 [+ or -] 44
 [+ or -] 51
 [+ or -] 52
 [+ or -] 59
 NycoCard (Nycomed Pharma AS) 500 (DD) 24
 500 (NR) 37
 500 (NR) 44
 500 (DD) 51
 250 (DD) 52
Erythrocyte agglutination assay
 SimpliRED (Agen Biomedical) [+ or -] 22
 [+ or -] 23
 [+ or -] 23
 [+ or -] 25
 [+ or -] 28
 [+ or -] 28
 [+ or -] 33
 [+ or -] 51
 [+ or -] 58
 [+ or -] 66
 [+ or -] 68
Automated rapid ELFA
 VIDAS (bioMerieux) 500 (FEU) 21
 500 (NR) 23
 700 (FEU) 39
 400 (FEU) 43
 400 (FEU) 44
 500 (FEU) 44
 500 (FEU) 51
 500 (FEU) 52
 500 (FEU) 68
Microplate ELISAs
 Asserachrom (Diagnostica Stago) 400 (FEU) 43
 500 (FEU) 44
 500 (FEU) 44
 500 (FEU) 51
 500 (FEU) 59
 Dimertest EIA (Agen Biomedical) 500 (NR) 58
 Dimertest GOLD EIA (Agen 70 (NR) 39
 Biomedical) 50 (DD) 52
 Enzygnost (Dade Behring) 80 (NR) 24
 40 (NR) 44
 70 (DD) 51
 Fibrinostika (Organon Teknika) 350 (NR) 44
 500 (FEU) 51
 500 (NR) 68

 Sensitivity, Specificity,
Assay (manufacturer) % %

First-generation latex
 agglutination assays
 D-dimer test (Diagnostica Stago) 73 62
 Dimertest (Agen Biomedical) 51 47
 Dimertest II (Agen Biomedical) 80 67
 Minutex (Biopool) 80 90
 95 59
 77 64
 Nephelotex (Biopool) 98 65
Second-generation latex
 agglutination assays
 Auto Dimertest (Agen Biomedical) 98 44
 BC D-Dimer (Dade Behring) 97 63
 77 91
 IL Test (Instrumentation 94 52
 Laboratory, SpA) 90 78
 Liatest (Diagnostica Stago) 96 47
 95 35
 LPIA (Mitsubishi Kasei) 95 69
 Tinaquant (Roche Diagnostics) 83 58
 98 56
 100 39
 99 33
 Turbiquant (Dade Behring) 92 40
 89 60
Membrane ELISAs
 Instant I.A. (Diagnostica Stago) 93 94
 81 67
 93 19
 94 63
 95 76
 92 37
 NycoCard (Nycomed Pharma AS) 90 80
 97 40
 80 38
 98 31
 95 40
Erythrocyte agglutination assay
 SimpliRED (Agen Biomedical) 66 76
 71 87
 91 82
 89 77
 87 79
 87 47
 56 72
 80 94
 94 61
 84 45
 61 90
Automated rapid ELFA
 VIDAS (bioMerieux) 96 44
 98 34
 97 82
 88 31
 97 26
 100 53
 100 41
 95 55
 100 19
Microplate ELISAs
 Asserachrom (Diagnostica Stago) 91 31
 97 22
 96 61
 98 42
 93 35
 Dimertest EIA (Agen Biomedical) 87 48
 Dimertest GOLD EIA (Agen 97 82
 Biomedical) 95 59
 Enzygnost (Dade Behring) 100 80
 93 30
 94 62
 Fibrinostika (Organon Teknika) 93 37
 100 36
 98 36

 PPV, (a) NPV,
Assay (manufacturer) % %

First-generation latex
 agglutination assays
 D-dimer test (Diagnostica Stago) 60 75
 Dimertest (Agen Biomedical) 94 47
 Dimertest II (Agen Biomedical) 73 76
 Minutex (Biopool) 89 81
 71 92
 82 56
 Nephelotex (Biopool) 76 96
Second-generation latex
 agglutination assays
 Auto Dimertest (Agen Biomedical) 58 96
 BC D-Dimer (Dade Behring) 63 97
 90 79
 IL Test (Instrumentation 35 97
 Laboratory, SpA) 80 88
 Liatest (Diagnostica Stago) 65 92
 68 82
 LPIA (Mitsubishi Kasei) 77 93
 Tinaquant (Roche Diagnostics) 34 93
 64 97
 63 100
 76 93
 Turbiquant (Dade Behring) 28 95
 70 84
Membrane ELISAs
 Instant I.A. (Diagnostica Stago) 82 98
 65 82
 49 77
 72 91
 81 93
 68 76
 NycoCard (Nycomed Pharma AS) 59 96
 48 95
 50 71
 59 94
 64 88
Erythrocyte agglutination assay
 SimpliRED (Agen Biomedical) 43 89
 63 91
 60 97
 56 95
 62 94
 41 90
 50 77
 93 82
 77 88
 78 56
 93 52
Automated rapid ELFA
 VIDAS (bioMerieux) 32 97
 31 98
 78 98
 49 76
 51 93
 63 100
 63 100
 70 91
 72 100
Microplate ELISAs
 Asserachrom (Diagnostica Stago) 50 81
 50 91
 66 95
 64 95
 67 77
 Dimertest EIA (Agen Biomedical) 68 75
 Dimertest GOLD EIA (Agen 78 98
 Biomedical) 71 92
 Enzygnost (Dade Behring) 61 100
 51 85
 72 90
 Fibrinostika (Organon Teknika) 54 88
 62 100
 77 88

(a) PPV, positive predictive value; NPV, negative predictive value;
NR, not reported; DD, D-dimer units.

(b) [+ or -], qualitative dichotomous result.

(c) Because of partial overlap with patients from Sadouk et al. (31),
only Perrier et al. (16) data for VIDAS are presented.

Table 4. Results of multivariate analysis.

Variable Relative DOR (a) P

Assay
Asserachrom 0.63 NS (b)
Auto Dimertest 1.67 NS
BC D-dimer 0.38 NS
D-dimer test 0.37 NS
Dimertest 0.08 <0.0001
Dimertest EIA 0.44 NS
Dimertest GOLD EIA 1.31 NS
Dimertest II 0.53 NS
Enzygnost 0.76 NS
Fibrinostika 1.03 NS
IL Test 0.43 NS
Instant I.A. 0.67 NS
LPIA 1.01 NS
Liatest 0.88 NS
Minutex 0.45 NS
Nephelotex 1.68 NS
NycoCard 0.28 0.03
SimpliRED 0.46 NS
Tinaquant 1.07 NS
Turbiquant 0.20 <0.001
VIDAS 1.00 -(c)
Reference standard
US or US/Venog 0.37 <0.005
Venog 1.00 -(c)
Patient mix
Mixed 0.61 NS
Outpatient 1.00 -(c)
Prevalence 0.78 0.01

(a) Ratio of the DOR of each variable to that of the reference
category, or for prevalence, DOR of assay per 10% increase in
prevalence.

(b) NS, not significant; US, ultrasonography; Venog, venography.

(c) Reference category.
COPYRIGHT 2004 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2004 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Author:Heim, Steven W.; Schectman, Joel M.; Siadaty, Mir S.; Philbrick, John T.
Publication:Clinical Chemistry
Article Type:Clinical report
Date:Jul 1, 2004
Words:9938
Previous Article:Rapid intraoperative immunoassay of parathyroid hormone and other hormones: a new paradigm for point-of-care testing.
Next Article:Denaturing HPLC-based approach for detecting RYR2 mutations involved in malignant arrhythmias.
Topics:

Terms of use | Privacy policy | Copyright © 2021 Farlex, Inc. | Feedback | For webmasters