Continuous blood glucose monitoring with a thin-film optical sensor.

Conventionally, glucose is measured electrochemically via enzymes on an intermittent basis determined by patient compliance. Consequently, excursions into hyper- or hypoglycemia may be missed, with potential long-term health implications (1). Newer devices in which the electrode is inserted subdermally over a 3- to 5-day period allow continuous monitoring but are susceptible to drift and may require recalibration. These devices are not approved for therapeutic intervention, and a fingerstick measurement must be taken before administration of insulin (2). Some of the drift seen with these devices may be attributed to their reliance on enzymes (e.g., glucose oxidase) and the well-known tendency of proteins to denature or aggregate over time, which may affect their catalytic performance (3-5).

Alternative approaches being actively developed, with potential to provide more reliable long-term continuous monitoring for people with diabetes, use the group of synthetic receptors called phenylboronic acids (6-9). These Lewis acids can bind the cis-1,2- or -1,3diols of glucose covalently to form 5- or 6-part rings. Unusually, the binding of saccharides is reversible, allowing a direct measurement of glucose, in contrast to enzyme-based methods.

The present study is the 1st time that glucose has been measured in a real physiologic liquid (human blood plasma) using phenylboronic acids. Other phenylboronic acid-based sensor studies have resorted to using simple buffer solutions with selected additives such as sodium chloride or proteins (10-12). This study is also the 1st to test a phenylboronic acid-based sensor for suitable temporal response characteristics that would enable following changes in blood plasma glucose in real time. We used a novel holographic princi ple, first developed by Lowe and colleagues (13, 14), wherein glucose is monitored using phenylboronic acids covalently immobilized within a thin-film biocompatible hydrogel, into which a holographic grating has been recorded. In response to analyte-receptor binding and fluctuating glucose concentrations, hydrogel swelling and contraction increase or decrease the spacing between the holographic fringes of the diffraction grating, thereby modulating the color of the light that is diffracted in a manner determined by Bragg's law. The advantage of this novel detection method over optical techniques such as fluorescence is the long-term stability of the sensor and the ease with which the wavelength may be tuned to suit the application (see the Data Supplement that accompanies the online version of this at article http: / /www. clinchem.org/content/vol53/issuelO). Fluorophores, in contrast, often have physical characteristics (e.g., excitation and emission maxima) inappropriate for real applications for which light attenuation, background fluorescence, and photobleaching are possible (15,16).

The physical and chemical robustness of these holographic sensors allows them to be used continuously for extended periods without the need for recalibration. In addition, these sensors are format-flexible and potentially may be incorporated into strips, catheters, contact lenses, and subdermal implants, thereby providing an attractive alternative to conventional electrochemical enzyme-based methods of glucose monitoring for people afflicted with diabetes.

Materials and Methods

The materials and equipment used to make the holograms and the materials for the buffer solutions are listed in Supplemental Data. Human blood plasma (single donor and pooled) was purchased from Scipac, Patricell, Cambridge Biosciences, and Valley Biomedical. Samples were supplied with either Na EDTA or [K.sub.3] EDTA as the anticoagulant, divided into aliquots, and stored at -20[degrees]C on receipt (17,18). We used fresh plasma samples for each experiment. In accordance with guidelines, the samples were thawed from frozen through incubation at 37 [degrees]C in a water bath to prevent cryoprecipitation (19). We measured sample pH by use of a Hanna Instruments pH 213 Microprocessor pH meter or an ELIT 8 channel pH analyzer (NICO2000 Ltd.) equipped with a Hamilton Biotrode or a Jencons P 16 microelectrode designed for use in small-volume proteinaceous samples. The pH microelectrodes were calibrated with a 2-point calibration using fresh 20-mL sachets of ready-mixed pH buffer (Hanna Instruments, NIST traceable) at pH 7.01 and either pH 4.01 or 10.01 as appropriate before measurement. We used peristaltic pumps (Ismatec Reglo digital ISM 834) fitted with Tygon tubing (Ismatec S00061 and S00068 set for flow rates of 78.4/[micro]L/min and 175 /[micro]L/min, respectively) for glucose-tracking experiments.

SENSOR HOLOGRAM PREPARATION

We prepared 3 mol% NN'-methylene bisacrylamide (MBA), [1] 12 mol% 3-acrylamidophenylboronic acid (3-APB), 12 mol% (3-acrylamidopropyl)trimethylammonium chloride (ATMA) holograms on glass microscope slides as described (20) with a few modifications (see Supplemental Data). The formulation of the polymer hydrogel is expressed in terms of mol%, which reflects the molar amounts of each monomer in the dry thin-film hydrogel. To obtain the hydrogels the appropriate quantities of acrylamide, MBA, 3-APB, and ATMA to give the desired molar ratios are dissolved in a solution of 2% (wt/vol) 2,2-dimethoxy-2-phenylacetophenone in DMSO at a ratio of 1:2.21 (wt/vol) of monomers to solvent. After polymerization the DMSO is removed and the hydrogel is dried down. We assume that there is 100% conversion of the monomers to polymer.

GLUCOSE MONITORING IN BLOOD PLASMA SAMPLES

Plasma sample preparation and pH control. Ex vivo plasma becomes more alkaline when exposed to the atmosphere owing to the continual loss of carbon dioxide and its role in the carbonic acid-bicarbonate buffering mechanism (21). To limit the effects of sample pH fluctuations during use and mimic conditions in vivo, we used 2 methods for pH control. The method used to create samples for glucose-tracking experiments required buffering the sample within the physiologic pH range with sodium phosphate, producing a final phosphate concentration of 0.1 mol/L. The buffered solution was then exposed to the atmosphere for 16 h to equilibrate P[C0.sub.2], before heating to 37 [degrees]C and measuring the final pH. The sample was subsequently sterilized by passing through a 0.45-[micro]m Minisart High Flow syringe filter (Sartorius). In the 2nd method, used for static cell experiments, [CO.sub.2] was infused directly into the sample using a syringe to lower the sample pH to 7.2, and the diffraction wavelength of the hologram was read when the sample pH attained pH 7.4. Commercially available human plasma is typically supplied within the normal range of glucose concentrations for adults, 4.1 to 5.9 mmol/L (19). To create a panel of samples covering the pathologic range of glucose concentrations, we enriched samples with additional glucose. Samples were allowed to equilibrate for a minimum of 45 min before use, in accordance with ISO guidelines to

allow time for mutarotation of glucose (22).

Hologram glucose interrogation and system-specific sensor calibration. Sensor holograms were interrogated from the glass substrate side as described (20) with a few modifications (see Supplemental Data). We calibrated the sensors by comparing sensor diffraction wavelength with sample glucose concentration measured using a YSI 2300 STATplus glucose analyzer. The holographic sensor directly detects analyte molality (moles of analyte per volume of water) of the undiluted sample, and the YSI 2300 measures analyte molarity (moles of analyte per volume of solution) in a diluted sample. For calibration, we used a plasma standard to decrease the systematic error produced through the difference in water content on the YSI 2300-measured glucose concentration, which would occur if an aqueous buffer were used (18,23).

Static calibration. Each hologram slice was equilibrated in 1.5 mL PBS (10 mmol/L phosphate, 2.7 mmol/L potassium chloride, 137 mmol/L sodium chloride solution, pH 7.40 at 25 [degrees]C) to establish a 0 glucose baseline before removal of the PBS and addition of 1.5 mL plasma. Sample pH was manipulated as described above; after the sample attained pH 7.4, the sensor diffraction wavelength was measured and a 60-[micro]L aliquot of plasma was removed from the cuvette for YSI 2300 analysis to establish sample glucose concentration. We added a 60-[micro]L aliquot of 0.1 mol/L glucose-containing plasma. The sample pH was then lowered, and this process was repeated until a glucose concentration >30 mmol/L was recorded on the YSI 2300. The nonlinear calibration data were fitted using the modified single 3-parameter exponential decay function in SigmaPlot 9.

Flow calibration. The sensor cuvette was filled with 1.5 mL buffered plasma, and a peristaltic pump was primed with 1.7 mL buffered plasma. During the assay, plasma was circulated to and from the cuvette at a rate of 175 [micro]L/min, mimicking the system under operation in a glucose-tracking experiment. At 45-min intervals, a 60-[micro]L sample of plasma was removed from the cuvette for analysis on YSI 2300, and a 60-[micro]L aliquot of 0.1 mol/L glucose in plasma was added to the cuvette. Between each addition and removal, the sensor was allowed to equilibrate to a stable diffraction wavelength. This procedure was repeated until a 17 mmol/L glucose concentration was recorded, producing a calibration plot.

Ex vivo glucose sensing. Hologram slices were equilibrated in 1.5 mL PBS before complete solution exchange into sample plasma. We performed sample pH manipulation as stated above and recorded the sensor wavelength when pH 7.4 was attained. Simultaneously, we measured the sample glucose concentration on the YSI 2300. We used the plot equation function in SigmaPlot 9 to interpolate sample concentration from sensor response.

Holographic-based continuous glucose sensing in plasma. Glucose tracking was achieved through a fluidic system using 2 peristaltic pumps (see Supplemental Data). Both peristaltic pumps were primed and initialized 10 min before interrogation of the sample, enabling the system to equilibrate. The plasma glucose concentration in the cuvette was increased through the infusion of high-glucose (44 mmol/L) stock plasma at a rate of 78.4 [micro]L/min into a stirred plasma reservoir containing 15 mL plasma. The reservoir had an initial theoretical rate of change of 0.23 [mmol.sup.-1] * [L.sup.-1] * [min.sup.-1]. The reservoir plasma was fed at a rate of 175 [micro]L/min into the sensor cuvette, which was emptied to waste at a rate of 175 [micro]L/min, retaining a constant cuvette volume of 1.5 mL. When the appropriate plasma glucose concentration was attained in the reservoir, the pump infusing high-glucose stock plasma was stopped. The plasma glucose concentration within the cuvette was decreased through replacement of the reservoir every 8.5 min with a plasma sample containing a 1.39 mmol/L lower glucose concentration, aiming for a rate of change of 0.17 [mmol.sup.-1] * [L.sup.-1] * [min.sup.-1]. Throughout the procedure, 40-[micro]L plasma samples were collected every 2 to 5 min from the cuvette for glucose measurement by YSI 2300, and the sensor diffraction wavelength was recorded. After a complete cycle of glucose concentrations, the sensor was washed into PBS for overnight storage. We collected daily glucose excursions in plasma over a 3-day period without further calibration. Glucose rate error grids were constructed following methods used in the original report and a subsequent appraisal (24, 25). We collected and paired sensor readings for glucose concentration and their reference blood glucose counterparts every 2 to 5 min over the course of the experiment. Glucose rate of change was calculated using 3- and 15-min time intervals interpolated from the fitted data. The glucose concentration vs time curves for both the sensor and the reference were fitted using the 5-parameter Weibull regression analysis in SigmaPlot 9.0. The sensor and reference plasma glucose rate of change was calculated as the change in glucose concentration of the sample divided by the elapsed time ([mmol.sup.-1] * [L.sup.-1] * [min.sup.-1]).

Results

In Fig. 1, we show the color of a 3 mol% MBA, 12 mol% 3-APB, 12 mol% ATMA hologram in human blood plasma at pH 7.4 and 37 [degrees]C as a function of the glucose concentration. The results of using holographic glucose sensors with this hydrogel formulation to measure 46 samples with different glucose concentrations at pH 7.4 and 37 [degrees]C are plotted in Fig. 2 against the true reference values obtained using a YSI 2300 (26). The data pairs are plotted as a Clarke error grid, which separates the results into zones reflecting their clinical significance (27). Of 46 measurements, 42 (91.3%) are located in zone A (i.e., accurate), whereas 4 of 46 measurements (8.7%) are located in zone B (i.e., benign outcome). Table 1 shows the results of an experiment performed after collection of the Clarke error grid data to assess measurement imprecision.

The measurements performed under flow at pH 7.4 and 37[degrees]C while varying the blood plasma glucose concentration at approximately 0.17 to 0.28 [mmol.sup.-1] * [L.sup.-1] * [min.sup.-1] (i.e., at approximately the maximum rate of change of blood glucose in vivo) are represented in the form of rate error grids in Fig. 3 (24, 25). The number of observations made at each rate is listed in Table 2. Using data interpolation and a 15-min time interval, 34 of 34 readings (100%) are located in zone [A.sub.R] (i.e., accurate) with 0 of 34 measurements located in the [uB.sub.R] (i.e., upper benign) or 1BR (i.e., lower benign) zones.

[FIGURE 1 OMITTED]

Fig. 4 shows how the peak diffraction wavelength of the hologram changes with glucose concentration for different pH values of human blood plasma.

Discussion

Based on the results of a previous study looking at sugar selectivity, we chose to prepare thin-film holographic sensors composed of 73 mol% acrylamide, 3 mol% MBA, 12 mol% 3-APB, and 12 mol% ATMA (20) (see Supplemental Data). The 3-APB Lewis acid receptor endows the hydrogel with saccharide responsivity, and ATMA is included at the specified concentration because of its ability to modify the sugar response of the sensor and increase its selectivity for glucose over other saccharides possessing cis-1,2- or -1,3-diols, including the main potential interferent for phenylboronic acids, fructose.

[FIGURE 2 OMITTED]

When illuminated with ordinary white light, the holographic fringes within the gel collectively diffract a narrow band of wavelengths in a manner governed by the Bragg equation ([[lambda].sub.max] = 2nd sin [theta]). The end result is a characteristic spectral reflection peak, the wavelength of which can be correlated with glucose concentration. Using the recording conditions, we obtained holograms that replayed in the red to yellow region of the visible spectrum (Fig. 1). With this sensor, the presence of glucose specifically causes the diffraction peak to blue-shift, indicative of hydrogel contraction of the film perpendicular to the fringe planes. Previously, we proposed that the contraction is caused by cross-linking of 2 boronic acid receptors with favorable stereochemistry by glucose to give a bis-boronate-glucose complex (20).

We generated calibration plots by incrementally increasing the blood plasma glucose concentration and recording the color of the test holograms spectrophoto-metrically. The glucose concentration of 46 unknowns was then determined from their respective hologram colors using the fit parameters from the calibration curves. Fig. 2 displays the data in the form of a Clarke error grid. The data pairs are scattered about the 45degree line, demonstrating a good correlation between the test and reference measurement methods. Because 91.3% of the points fall in zone A, the test values deviate by <20% from the true glucose value or both the test value and the true glucose value are <3.9 mmol/L. Because 8.7% of the data points fall in zone B, the deviation from the true glucose value is >20%, but clinically no treatment is needed or, if treated, the result is benign. No points fall in zones C through E, which correspond to errors that are potentially dangerous.

[FIGURE 3 OMITTED]

The %CV for the test sensor compares favorably with the YSI reference technique (Table 1). It is anticipated that improvements to the experimental protocols, for instance used to account for the significant pH variation of the samples ex vivo, will lead to further measurement accuracy and precision (see Supplemental Data). One benefit of the holograms is that they behave extremely similarly (see Supplemental Data). In the future, this finding raises the possibility of being able to mass-manufacture materials that do not require calibration before use.

[FIGURE 4 OMITTED]

To see if the sensors would be suitable for real-time continuous glucose monitoring, particularly useful for patients with type 1 (insulin-dependent) diabetes, we conducted flow experiments to see if the sensor could track changes in glucose occurring up to the maximum rate of change in vivo. The rate of change of blood glucose and its distribution for the diabetic population has been characterized by Dunn et al. (28). Rates in excess of 0.17 [mmol.sup.-1] * [L.sup.-1] * [min.sup.-1] (upward and downward) are observed in <1% of observations, whereas the majority of the time (approximately 37% of observations), blood glucose changes in the body at <0.06[mmol.sup.-1] * [L.sup.-1] * [min.sup.-1].

The number of observations at each different rate is listed in Table 2. Fig. 3 shows the rate error grids obtained from the data for 3 consecutive up and down concentration-time curves (see Supplemental Data) assuming fixed time intervals of 3 and 15 min for analysis. The plot reflects the difference in rate between the test sensor and corresponding reference measurements and indicates how well the holographic sensor is able to follow changes in the direction and tempo of plasma glucose. Using a 15-min interval, good accuracy is obtained with 34 readings (100%) located in zone AR (i.e., accurate readings) and 0 of 34 measurements in the upper and lower benign zones (uBR and 1BR). Better rate accuracy is obtained if longer time intervals between consecutive measurements are used (see Supplemental Data). Part of this relates to the larger measurement error associated with smaller glucose changes obtained over shorter time intervals.

A common concern regarding hydrogels based on phenylboronic acids is their relatively slow swelling times. Often, rather than quoting equilibrium values, response times are given in terms of the time taken for the signal to reach 90% of its maximum (29). There has been much discussion about the physical form of the glucose binding to the phenylboronic acid (30). Glucose undergoes mutarotation between its different anomers. It is postulated that the glucose a-anomers bind faster than the [beta\-anomers, but the overall response is poor because the relative abundance of [alpha]-anomers is low and interconversion between the different diastereoisomers is slow. In this study, we show for the 1st time that the response characteristics of phenylboronic acid holographic glucose sensors is perfectly adequate for real applications for which continuous monitoring is a benefit, for instance in type 1 diabetes or to facilitate intensive insulin therapy in critical care settings (31).

The ability of these holographic sensors to monitor in real time over long periods without hysteresis makes them ideally suited to continuous glucose monitoring. A benefit of these hydrogels for such applications is their inherent simplicity and biocompatibility (32-34), which is expected to lead to more reliable monitoring. Temperature effects on the signal will be minimized if used in vivo, because the temperature remains essentially constant between 36.1 and 37.8 [degrees]C (35). Our hydrogels showed limited temperature sensitivity. Tests measurements made at 36 and 38 [degrees]C of human blood plasma containing glucose at 30, 8, and 4 mmol/L all fell within zone A of the Clarke error grid. The plasmas used in this work were at normal physiologic ionic strength (275-300 mOsmol/kg) and are expected to contain normal levels of potentially interfering substances. However, zone A results were also obtained with samples whose ionic strength was adjusted to critical or panic alert levels reflecting a life-threatening situation (250-325 mOsmol/kg) (18).

Another possible cause of signal drift and measurement error is pH. Again the pH of whole blood/plasma is usually tightly controlled between 7.35 and 7.45. However, conditions such as diabetic ketoacidosis and ketoalkalosis may cause pH to fall outside the normal range (36). To investigate how changes in blood/plasma pH affect the glucose measurement, samples of human blood plasma at 6 different glucose concentrations were incubated at 37 [degrees]C, and the pH was allowed to drift from 7.2 to 7.7 (Fig. 4). From this data, it is very simple to build a calibration plot for every glucose concentration at every pH value and use this result to obtain an accurate glucose measurement. Many boronic acids have [pK.sub.a] values above physiologic pH (37). Therefore, as the pH of the surrounding medium increases towards the [pK.sub.a] of the phenylboronic acid in the hydrogel (3-APB, approximately 8.2) (38), the phenylboronic acids switch to their tetrahedral boronate form. This configuration is thought to be better able to bind the vicinal diols of glucose on steric grounds. Thus, the degree of bis-boronate-glucose cross-linking at a given glucose concentration increases with pH, leading to increased hydrogel contraction. Strategies to minimize the pH dependence of phenylboronic acids include modification of the aromatic ring with electron-withdrawing groups so that the tetrahedral form is predominant at physiologic pH and the incorporation of electron donors adjacent to the boron to force the boron into its tetrahedral boronate form (39). The latter method is not as effective, because the electron donors can become protonated or deprotonated if the pH of the surrounding medium changes greatly. Future work will examine the behavior of sensors incorporating 2-acrylamidophenylboronic acid derivatives. These phenylboronic acids have an oxygen electron donor intramolecular to the boron and have been shown to exist in a zwitterionic tetrahedral form that is insensitive to changes in pH (40). It is anticipated that the use of these receptors will remove the need outside the normal physiologic pH range for a separate pH measurement as a control.

In conclusion, phenylboronic acid-based sensors can accurately measure glucose in real biological fluids and track changes in concentration occurring at rates mimicking those in vivo. Future work will examine the behavior of these sensors in the presence of different therapeutic substances under physiologic conditions corresponding to different disease states using phenylboronic acid receptors with reduced pH sensitivity.

Grant/funding support: We thank the U.K. Department of Trade and Industry for funding via the Knowledge Transfer Partnership scheme for Dr. Michael Thatcher, a University of Bath KTP Associate with Smart Holograms Ltd.

Financial disclosures: None declared.

Acknowledgments: We are grateful to Prof. Chris Lowe (Institute of Biotechnology, University of Cambridge) and Drs. Tony James and Steve Bull (Department of Chemistry, University of Bath) for useful discussions.

Received May 7, 2007; accepted August 2, 2007. Previously published online at DOI: 10.1373/clinchem.2007.091629

References

(1.) The Diabetes Control and Complications Trial (DCCT) Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993;329:977-86.

(2.) DEXCOM[TM] STS[TM] Continuous glucose monitoring system packaging insert. DexCom 2006;LBL 9551, Rev. 04.

(3.) Rauf S, Ihsan A, Akhtar K, Ghauri MA, Rahman M, Anwar MA, et al. Glucose oxidase immobilization on a novel cellulose acetate-polymethylmeth acrylate membrane. J Biotechnol 2006;121:351-60.

(4.) Babu VRS, Kumar MA, Karanth NG, Thakur MS. Stabilization of immobilized glucose oxidase against thermal inactivation by silanization for biosensor applications. Biosens Bioelectron 2004; 19:1337-41.

(5.) Akhtar MS, Ahmad A, Bhakuni V. Divalent cation induced changes in structural properties of the dimeric enzyme glucose oxidase: dual effect of dimer stabilization and dissociation with loss of cooperative interactions in enzyme monomer. Biochemistry 2002; 41:7142-9.

(6.) Takahashi S, Anzai J. Phenylboronic acid monolayer-modified electrodes sensitive to sugars. Langmuir 2005;21:5102-7.

(7.) Brewer SH, Allen AM, Lappi SE, Chasse TL, Briggman KA, Gorman CB, et al. Infrared detection of a phenylboronic acid terminated alkane thiol monolayer on gold surfaces. Langmuir 2004;20: 5512-20.

(8.) Lee YJ, Pruzinsky SA, Braun PV. Glucose-sensitive inverse opal hydrogels: analysis of optical diffraction response. Langmuir 2004;20:3096-106.

(9.) Karnati W, Gao XM, Gao SH, Yang WQ, Ni WJ, Sankar S, et al. A glucose-selective fluorescence sensor based on boronic acid-diol recognition. Bioorg Med Chem Lett 2002;12:3373-7.

(10.) Badugu R, Lakowicz JR, Geddes CD. Fluorescence sensors for monosaccharides based on the 6-methylquinolinium nucleus and boronic acid moiety: potential application to ophthalmic diagnostics. Talanta 2005;65:762-8.

(11.) Alexeev VL, Das S, Finegold DN, Asher SA. Photonic crystal glucose-sensing material for noninvasive monitoring of glucose in tear fluid. Clin Chem 2004;50:2353-60.

(12.) Suri JT, Cordes DB, Cappuccio FE, Wessling RA, Singaram B. Continuous glucose sensing with a fluorescent thin-film hydrogel. Angew Chem Int Ed Engl 2003;42:5857-9.

(13.) Millington RB, Mayes AG, Blyth J, Lowe CR. A holographic sensor for proteases. Anal Chem 1995;67:4229-33.

(14.) Mayes AG, Blyth J, Millington RB, Lowe CR. A holographic sensor based on a rationally designed synthetic polymer. J Mol Recognit 1998;11:168-74.

(15.) Ziegler KJ. Developing implantable optical biosensors. Trends Biotechnol 2005;23:440-4.

(16.) Sapsford KE, Pons T, Medintz IL, Mattoussi H. Biosensing with luminescent semiconductor quantum dots. Sensors 2006;6: 925-53.

(17.) Art to Science. Freezing and thawing serum and other biological materials: optimal procedures minimize damage and maximize shelf-life. HyClone Laboratories, Inc. 1992;11:1-8.

(18.) Burtis CA, Ashwood ER, Bruns DE, eds. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics, 4th ed. Philadelphia: Elsevier Saunders, 2003:56pp.

(19.) Duguid J, O'Shaughnessy DF, Atterbury C, Maggs PB, Murphy M, Thomas D, et al. Guidelines for the use of fresh-frozen plasma, cryoprecipitate and cryosupernatant. British Committee for Stan dards in Haematology, Blood Transfusion Task Force. Br J Haematol 2004;126:11-28.

(20.) Horgan AM, Marshall AJ, Kew SJ, Dean KES, Creasey CD, Kabilan S. Crosslinking of phenylboronic acid receptors as a means of glucose selective holographic detection. Biosens Bioelectron 2006;21:1838-45.

(21.) Fura A, Harper TW, Zhang HJ, Fung L, Shyu WC. Shift in pH of biological fluids during storage and processing: effect on bioanalysis. J Pharm Biomed Anal 2003;32:513-22.

(22.) ISO 15197:2003(E). In vitro diagnostic test systems: requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. Switzerland: ISO, 2003:15pp.

(23.) Fogh-Andersen N, D'Orazio P. Proposal for standardizing direct-reading biosensors for blood glucose. Clin Chem 1998;44:655-9.

(24.) Wentholt IM, Hoekstra JB, DeVries JH. A critical appraisal of the continuous glucose-error grid analysis. Diabetes Care 2006;29: 1805-11.

(25.) Kovatchev BP, Gonder-Frederick LA, Cox DJ, Clarke WL. Evaluating the accuracy of continuous glucose-monitoring sensors. Diabetes Care 2004;27:1922-8.

(26.) Chua KS, Tan IK. Plasma glucose measurement with Yellow Springs glucose analyzer. Clin Chem 1978;24:150-2.

(27.) Clarke WL, Cox D, Gonderfrederick LA, Carter W, Pohl SL. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care 1987;10:622-8.

(28.) Dunn TC, Eastman RC, Tamada JA. Rates of glucose change measured by blood glucose meter and the Glucowatch Biographer during day, night, and around mealtimes. Diabetes Care 2004; 27:2161-5.

(29.) Gamsey S, Suri JT, Wessling RA, Singaram B. Continuous glucose detection using boronic acid-substituted viologens in fluorescent hydrogels: linker effects and extension to fiber optics. Langmuir 2006;22:9067-74.

(30.) Ben-Moshe M, Alexeev VL, Asher SA. Fast responsive crystalline colloidal array photonic crystal glucose sensors. Anal Chem 2006;78:5149-57.

(31.) Garber AJ. American College of Endocrinology and American Diabetes Association consensus statement on inpatient diabetes and glycemic control: a call to action. Diabetes Care 2006;29: 1955-62.

(32.) Peppas NA, Bures P, Leobandung W, Ichikawa H. Hydrogels in pharmaceutical formulations. Eur J Pharmaceut Biopharmaceut 2000;50:27-46.

(33.) Final report of the safety assessment of polyacrylamide. J Am Coll Toxicol 1991;10:193-203.

(34.) Hall DG. Structure, properties, and preparation of boronic acid derivatives: overview of their reactions and applications. In: Hall DG, ed. Boronic Acids, Chapter 1. Weinheim: Wiley-VCH, 2005:199.

(35.) Simmers, L. Diversified Health Occupations, 2nd ed. Canada: Delmar, 1988:150-1.

(36.) Trachtenbarg DE. Diabetic ketoacidosis. Am Fam Phys 2005;71: 1705-14.

(37.) Yan J, Springsteen G, Deeter S, Wang BH. The relationship among pK(a), pH, and binding constants in the interactions between boronic acids and diols: it is not as simple as it appears. Tetrahedron 2004;60:11205-9.

(38.) Matsumoto A, Yoshida R, Kataoka K. Glucose-responsive polymer gel bearing phenylborate derivative as a glucose-sensing moiety operating at the physiological pH. Biomacromolecules 2004;5: 1038-45.

(39.) Wiskur SL, Lavigne JJ, Ait-Haddou H, Lynch V, Chiu YH, Canary JW, et al. pK(a) values and geometries of secondary and tertiary amines complexed to boronic acids: implications for sensor design. Org Lett 2001;3:1311-4.

(40.) Yang XP, Lee MC, Sartain F, Pan XH, Lowe CR. Designed boronate ligands for glucose-selective holographic sensors. Chem Eur J 2006;12:8491-7.

[1] Nonstandard abbreviations: MBA, NN'-methylene bisacrylamide MBA; 3-APB, 3-acrylamidophenylboronic acid; ATMA, (3-acrylamidopropyl)trimethyl-ammonium chloride.

GRAHAM J. WORSLEY, GUILHEM A. TOURNIAIRE, KATHRYN E. S. MEDLOCK, FELICITY K. SARTAIN, HAZEL E. HARMER, MICHAEL THATCHER, ADRIAN M. HORGAN, AND JOHN PRITCHARD *

SmartHolograms Ltd., Cambridge, United Kingdom.

* Address correspondence to this author at: Smart Holograms Ltd., 291 Cambridge Science Park, Milton Rd., Cambridge CB4 OWF, United Kingdom. Fax 44 (0) 1223 393401; e-mail john.pritchard@smartholograms.com.
Table 1. Results of an experiment to assess the
measurement precision (reproducibility and repeatability)
associated with the experimental test system. (a)

 Reproducibility

Glucose, %CV, %CV,
mmol/L hologram YSI

4 5.4 0.7
8 4.4 2.0
30 1.9 1.7

 Repeatability

Glucose, %CV, %CV,
mmol/L hologram YSI

4 4.8 1.7
8 1.1 2.1
30 1.5 0.7

(a) Experiments were performed using human blood plasma at
pH 7.4 and 37 [degrees] C and a 3 mol% MBA, 12 mol% 3-APB,
12 mol% ATMA hologram.

Table 2. The number of observations made at each
different rate for the fitted data at 2 time intervals.

Rate, mmol 15 min 3 min

0-0.06 7 35
0.06-0.11 8 47
0.11-0.17 12 54
0.17-0.22 6 34
COPYRIGHT 2007 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2007 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Automation and Analytical Techniques
Author:Worsley, Graham J.; Tourniaire, Guilhem A.; Medlock, Kathryn E.S.; Sartain, Felicity K.; Harmer, Haz
Publication:Clinical Chemistry
Date:Oct 1, 2007
Words:5009
Previous Article:Serum proprotein convertase subtilisin kexin type 9 is correlated directly with serum LDL cholesterol.
Next Article:Methanol-associated matrix effects in electrospray ionization tandem mass spectrometry.
Topics:


Related Articles
Abbott Receives European CE Mark Approval for FreeStyle Navigator(R) Continuous Glucose Monitoring System.
Medtronic Mediterranean Launches World's First Arabic-Language Integrated Insulin Therapy for Improved Diabetes Management.
Should continuous glucose monitors be used routinely in adolescent patients?
Smiths Medical to Promote and Distribute Abbott's FreeStyle Navigator Continuous Glucose Monitoring System.
Solianis' Non-Invasive Glucose Monitor Tracks Blood Glucose in Outpatient Study.
Non-Invasive Glucose Monitoring - Solianis Monitoring AG.
Advances in photoacoustic noninvasive glucose testing.
Materials for fabricating biosensors for transdermal glucose monitoring.

Terms of use | Copyright © 2018 Farlex, Inc. | Feedback | For webmasters