Comparative study of glucometer and laboratory glucose oxidase method for the estimation of blood glucose levels in neonates.

INTRODUCTION: Hypoglycemia is one of the most common metabolic problems encountered in the newborns. The term "hypoglycemia" refers to a reduction in the glucose concentration of the circulating blood. It is almost 100 years since hypoglycemia was first described in children and over 50 years since it was recognized in newborns and infants. (1)

Variable incidence has been reported by various authors in different weight and gestational age groups. (2) The overall incidence of hypoglycemia in neonates varies from 0.2 to 11.4%. (3,4) However in the presence of certain risk factors i.e. small for date, large for date, infants of diabetic mothers, prematurity etc., the probability of hypoglycemia increases many folds. (3)

Hypoglycemia in neonates can be symptomatic and asymptomatic. (5,6,7) The most common symptoms such as jitteriness, convulsions, apathy, hypotonia, coma, refusal to feed, cyanosis, high pitched cry, hypothermia are very nonspecific and especially in small sick infants, these symptoms may be easily missed. Therefore hypoglycemia must always be confirmed biochemically and by response to treatment. Hypoglycemia is known to be associated with brain dysfunction and neuromotor developmental retardation in both symptomatic and asymptomatic cases. (1,3,8)

The definition of clinically significant hypoglycemia remains one of the most confused and contentious issues in contemporary neonatology. (9) It is not possible to define a blood glucose level that requires intervention in every newborn infant because there is uncertainty over the level and duration of hypoglycemia that cause brain damage and little is known of the vulnerability of the brain of infants at different gestational ages for such damage. In symptomatic infant, plasma glucose concentration should be measured and if the value is <45mg% [2.5m mol/l], clinical interventions aimed at increasing blood glucose concentrations are indicated. (9)

The methods for determining blood/plasma glucose concentration include reductiometric method, glucose oxidase method and hexokinase method. (1,10) The glucose oxidase method used in the laboratory for determining the blood glucose concentration is precise and specific for glucose. (1) As it is usually performed in the main laboratory, the results are not available quickly enough for timely appropriate management. The development of reagent strip blood glucose tests in the 1970s facilitated the practice of screening for hypoglycemia in newborns. (1) Currently used glucometers were initially developed for glucose monitoring in adult diabetics. These glucometers are often used for blood glucose estimation in NICU. Many studies have shown that their results co-relate well with the laboratory measured glucose levels in the normoglycemic and hyperglycemic range but are not satisfactory in the lower range. (11-18)

OBJECTIVES: 1] To determine the efficacy of glucometer in the estimation of blood glucose levels in newborns, in comparison with the glucose oxidase method used in the laboratory.

REVIEW OF LITERATURE: HISTORICAL BACKGROUND: It is almost a century sine hypoglycemia was first described in children and over 50 years since it was first described in newborns. Given the numerous advances which have since occurred in the care of the newborns it is surprising that so much controversy still surrounds the definition, significance, detection and management of neonatal hypoglycemia. Paradoxically, technological developments in the form of bedside glucose monitoring have exacerbated rather than eased the problem by facilitating screening for an ill-characterized clinical entity.

METHODS FOR MEASURING BLOOD/PLASMA GLUCOSE CONCENTRATION:

1. REDUCTIOMETRIC METHODS: Traditional methods for measurement of blood glucose depend on the reducing property of glucose. An example is the ferricyanide method. These methods measure total reducing sugar concentrations. (1) Enzymatic methods have largely superseded reductiometric methods in clinical practice.

2. GLUCOSE OXIDASE METHOD: Glucose oxidase catalyses the oxidation of glucose to yield glucoronic acid and hydrogen peroxide. The concentration of hydrogen peroxide liberated is measured using a peroxidase step coupled to a colored oxygen acceptor or an electrode. These reactions form the basis of both reagent strips and bench top glucose electrode methods. (1)

3. HEXOKINASE METHOD: Hexokinase catalyses the phosphorylation of glucose by ATP. Glucose-6-phosphate is then reduced by glucose dehydrogense yielding NADPH/ [H.sup.+] which can be measured using a suitable spectrophotometric indicator system. This method is precise and highly specific for glucose but the main drawback is its cost. (10)

4. REAGENT STRIPS: These were initially developed for monitoring blood glucose concentration in diabetics and not intended for detection of hypoglycemia. Care must be taken to avoid contamination by alcohol skin-cleansers, to cover the whole surface of the test pad and to time the reaction precisely before wiping the strip. Even when all these precautions are taken, they tend to under-estimate systematically the mean of a series of measurements in the range of glucose concentrations relevant to the diagnosis of neonatal hypoglycaemia. (1)

Several commercially available systems are available and have been evaluated for neonatal use. Reagent strip methods are prone to many errors when used to screen for neonatal hypoglycemia. Confirmation with laboratory measured plasma glucose is still of up most importance. (11-18)

Several commercially available systems are available and have been evaluated for neonatal use including:

Dextrostix, BM-test-glycemie, Chemstrip bG, Glucostix.

Care must be taken to avoid contamination by alcohol skin-cleansers, to cover the whole surface of the test-pad, and to time the reaction precisely before wiping the strip. Reagent strip methods are prone to many errors when used to screen for neonatal hypoglycemia.

Reagent strips are subject to false positive and false negative results.

FALSE POSITIVE:

* Low hematocrit values (<35%).

* Contamination with iso propoyl alcohol.

FALSE NEGATIVE:

* High hematocrit values (>55%).

* Glucose values > 200mg/dl.

* Hyperglycemic-hyperosmolar states with or without ketosis.

* Delay in lab analysis.

VARIATIONS AND ERRORS IN MEASUREMENTS:

Properties of the sample and sources of error: Arterial blood has a slightly higher glucose concentration than venous. The magnitude of this difference varies with tissue glucose demands and will be greatest under anaerobic conditions.

Capillary sampling is unreliable if peripheral blood flow is reduced. Samples must be always be free-flowing as squeezing the heel causes haemolysis which interferes with the assay unless deproteinisation is performed. The sample should either be analyzed immediately or deproteinised (for example using perchloric acid) and chilled. Glycolysis otherwise continues. Commercially available sodium fluoride coated tubes do not always ensure a fluoride concentration sufficient to inhibit glycolysis

Contamination by alcohol used for skin preparation leads to erroneously high values.

One of the problems with neonatal samples is that haematocrit may vary from <40 to >70%. Red cells contain less water than an equivalent volume of plasma though the glucose concentration in red cell water is the same as that in the plasma. Plasma glucose concentration is therefore higher than that of whole blood, on average by about 18%.

All methods employing paper reagent strips are subject to an intrinsic haematocrit bias; the higher the haematocrit value, the lower the result. Possible reasons include discoloration of the test-pad and resistance to wiping or washing before reading. Also the higher sample viscosity impedes diffusion of plasma into the test-pad of the strip.

Bilirubin also interferes with glucose oxidase-peroxidase based strip methods. Bilirubin inhibits both steps of the assay leading to falsely low values.

Haemolysis also produces falsely low values. This may be attributable to presence of hemoglobin or to release of reduced glutathione which competes with the chromogen for hydrogen peroxide released in the assay.

5. GLUCOSE ELECTRODE SYSTEMS: Based on the glucose oxidase method. The device measures plasma glucose concentration on a whole blood uncentrifuged 25microlitre sample. (1)

6. OTHER BEDSIDE SYSTEMS: The Hemo Cue beta-glucose photometer is an optical method measuring whole blood glucose on small [5microlitre] samples utilizing disposable cuvettes. Blood is hemolysed in the cuvette and the NADH formed by enzymatic glucose oxidation reduces methylthiazolyldiphenyl tetrazolium to produce a formazan dye, the concentration of which is determined spectrophotometrically. Only one study has evaluated its application to neonatal samples.

MATERIALS AND METHODS:

1. SOURCE OF DATA [SAMPLE]: 250 neonates getting admitted as inpatients in NICU in KIMS Hospital, Bangalore during the study period of december 2013 to november 2014 were taken as the source of data for the study.

2. METHOD OF COLLECTION OF DATA: 250 admitted neonates satisfying the inclusion criteria were included in the study. Collection of data was by relevant investigations i.e. glucose estimation by glucometer and laboratory glucose oxidase method by using the same venous sample. For 50 of these cases, glucose estimation of both venous and capillary blood was done by using the glucometer along with the routine laboratory venous blood glucose estimation. Blood glucose estimation was done at the time of admission.

3. INCLUSION CRITERIA: 250 Neonates admitted in NICU, KIMS Hospital, Bangalore for varied symptomatology during the study period.

4. EXCLUSION CRITERIA: Neonates admitted in postnatal ward who were healthy, term and breast-fed.

Preterm and SGA newborns with birth weight more than 1.8kgs, active and healthy with no risk factors were shifted to mother side and were not included in the study.

5. METHODOLOGY:

Our study had 2 parts: First, estimation of venous blood level by glucometer and by laboratory glucose oxidase method. Laboratory glucose oxidase method was taken as the gold standard for the glucose estimation. For the last 50 cases, estimation of the capillary blood glucose level by glucometer was also done.

Second part of the study was to compare the glucometer values with laboratory values. For the last 50 cases, capillary and venous blood glucometer values were compared with laboratory values as gold standard.

Statistical analysis was done by using 'Pearson Correlation'.

1. GLUCOMETER: Used is ACCU-CHEK Advantage/ Sensor glucometer, manufactured by Roche Diagnostics, 2003.

TEST PRINCIPLE: Bioamperometry-glucose dehydrogenase in the strip converts the glucose in the blood sample to gluconolactone. This reaction creates a harmless electrical current that the glucometer interprets for that blood glucose.

PROCEDURE: Touch the drop of the blood collected, to the curve at the edge of the test strip. No part of the yellow color on the strip should be visible after applying the initial drop of blood. Blood will be drawn into the strip automatically. Do not place the blood drop on the top of the strip. Test result will appear within 30 seconds.

2. LABORATORY METHOD: Enzymatic calorimetric test, GOD-POD i.e. glucose oxidase-peroxidase method.

PRINCIPLE:

* Glucose + [O.sub.2]--Gluconic acid + H2O2

* [H.sub.2][O.sub.2] + 4-Aminoantipyrine + Phenol--Chinomine + 4[H.sub.2][O.sub.2]

The first step is catalyzed by glucose oxidase and the second by peroxidase enzyme. The reagents and standard are ready to use.

PROCEDURE:

* Monoreagent--1000microlitre.

* Sample--10micolitre.

* Mix well; incubate for 5 minutes at 37[degrees]c.

* Read the absorbance AA against reagent blank.

Calculation--glucose mg dl = [DELTA]A SAMPLE/[DELTA]A Standard X 100

RESULTS:

STUDY DESIGN: A Prospective correlation study consisting of 250 cases was undertaken to study the correlation of Glucometer and laboratory glucose oxidase method for estimating blood glucose levels in newborns.

Every increase of one unit of Glucometer value there is an increase of 0.68 units in Lab value (with 62.5% accurately).

Table 2 and figure 2a & 2b, shows that there is a large correlation [0.756] between the glucometer and lab values when the blood glucose values, as determined by the gold standard is >45mg%. But when the lab values [gold standard] are <45mg% then there is just moderate correlation [0.417] between the two.

Table 3 and figure 3 shows the pick-up rate of hypoglycemia by glucometer using venous blood and capillary blood in relation to the gold standard-lab value.

Pick-up rate of hypoglycemia by glucometer using capillary blood was the least [22%] followed by glucometer estimation using venous blood [28%].

DISCUSSION: In this study we evaluated the efficacy of glucometers in estimating the blood glucose levels in newborns, in comparison with laboratory glucose oxidase method. Main stress was laid on detection of hypoglycemia by both the methods and to know if glucometer is a good screening tool to detect hypoglycemia. Laboratory glucose oxidase method of blood glucose estimation was taken as gold standard.

In this prospective correlation study, 250 neonates who were admitted in NICU with varied symptomatology were enrolled. Of these majority [87.6%] were admitted in the first 7 days of life while the rest 12.4% admitted were more than 7 days old. This is due to the fact that majority of the clinical conditions associated with hypoglycemia such as septicemia [early onset], birth asphyxia, prematurity, respiratory distress etc. manifest in the early neonatal period and seek admission.

Hypoglycemia was defined as blood glucose level less than 45mg% as defined by the study by Cornblath et al. (9) 103 [41.2%] were found to be hypoglycemic by laboratory glucose oxidase method whereas by glucometer screening 52 [20.8%] were found to be hypoglycemic. Thus the incidence of hypoglycemia as detected by the gold standard method i.e. laboratory glucose oxidase method in our study is 41.2%.

This wide variation in the incidence of hypoglycemia could be attributed to the lack of uniform definition of hypoglycemia, varied sample size and risk factors. In the PK Singhal et al study hypoglycemia was defined as blood glucose level <30 mg% (3) while in the PK Mishra and Bina Sharma study it was taken as 20mg% (2). In our study the cut-off value for hypoglycemia was taken as 45mg%. Also our study group consisted of sick/ septic/preterm/lbw newborns referred from other hospitals/ nursing homes without I.V. fluids or feeds during transportation and the duration of transport was usually >8hours.This has probably given rise to a higher incidence of hypoglycemia.

HYPOGLYCEMIA AND GLUCOMETERS: The results depict the poor pick-up rate of hypoglycemia by glucometer in comparison with the laboratory values. It shows that nearly half of the cases of proven hypoglycemia [by lab method] were missed by glucometer, indicating that glucometers are not good screening tool for detecting hypoglycemia in newborns.

STUDIES                  SENSITIVITY   SPECIFICITY

1. Dahlberg et al (14)      100%          84%
2. Mehta et al (15)          86%          89%
3. H O HT et al (16)        92.3%
4. M Ellis et al (17)        83%          62%
5. Hamid MH et al (18)       98%          93%
6. Our Study               41.75%        93.88%


Our study shows that glucometers have a specificity of 93.88% to detect hypoglycemia which is in correlation with few of the other studies mentioned. But in contrast with the other studies, our study showed that glucometers had a very poor sensitivity of 41.75% to detect hypoglycemia. This may be due to the glucometer brand used.

Our study also showed that glucometers had a positive predictive value of 82.69% and a negative predictive value of 69.7%, which in contrast to the results obtained by H O HT et al (16) and Dahlberg et al (14), is low. The overall accuracy of glucometers to detect hypoglycemia in newborns in comparison with the gold standard i.e. laboratory value is 72.40%.

CORRELATION OF GLUCOMETER AND LAB VALUE [OVER ALL BLOOD GLUCOSE
LEVELS]:

Hamid et al (18)   0.976
Our study          0.771


Like other studies our study shows a good correlation between glucometer and laboratory values when all blood glucose values were considered irrespective of hypoglycemia.

We can also infer that for every increase of one unit of glucometer value there is an increase of 0.68 units in the lab value. Thus lab value can be calculated by the formula:

Lab value = 0.68 x Glucometer + 6.64.

Example: if the glucometer value is 80 mg% then by this formula lab value will be 0.68 x 80 + 6.64 = 61mg%

Accuracy of this formula is only 62.5%.

The results show that the correlation between the lab and glucometer is good when the gold standard value is >45mg% [0.756] and when the value is <45mg%, there is just a moderate correlation [0.417]. This further stresses the inefficacy of glucometers in detecting hypoglycemia in newborns.

CAPILLARY BLOD GLUCOSE ESTIMATION: For the last 50 cases, blood glucose estimation was done by laboratory method [venous sample] and by glucometer [both venous and capillary blood]. This comparison was done because in many centers, capillary blood glucose estimation by glucometer is being done to screen neonatal hypoglycemia.

Results show that blood glucose estimation using capillary blood has a poor pick-up rate of detecting neonatal hypoglycemia. More than 50% of the proven cases of hypoglycemia [proved by lab method] were missed when capillary blood glucose estimation with glucometer was done. Though the pick-up rate of hypoglycemia by venous blood glucometer [28%] was slightly better compared to the capillary blood glucometer value [22%], the overall pick-up rate of hypoglycemia by glucometer is very low compared to the laboratory method.

CONCLUSION: There is a variable detection rate of hypoglycemia by glucometer in the studies mentioned and also in our study. These 'point of care' devices had a very low sensitivity and negative predictive value and for detecting hypoglycemia in newborn but had good specificity and positive predictive value. The correlation between the glucometer and laboratory values is good when the gold standard i.e. lab value is >45mg% and when the value is <45mg%, there is just a moderate correlation. Thus excluding hypoglycemia, these glucometers had a very large correlation with the laboratory measured values. These glucometers as a sole measuring device to screen neonatal hypoglycemia is not satisfactory and confirmation with the laboratory measurements of plasma glucose is still of up most importance.

DOI: 10.14260/jemds/2015/383

BIBLIOGRAPHY:

(1.) A. F. Williams. Hypoglycemia in newborns: a review. BULL WHO 1997; 75: 261-90.

(2.) P. K. Mishra, & Bina Sharma. Hypoglycemia in Newborns-A Prospective Study. INDIAN PEDIATRICS 1977; 14: 129-32.

(3.) P.K Singhal, M. Singh, V. K. Paul, A. K. Deodari, M.G Ghorpade, A. Malhotra. Neonatal Hypoglycemia--Clinical profile and glucose requirements. INDIAN PEDIATRICS 1992; 29: 167-71.

(4.) A.K Dutta, & Anu Aggarwal. Neonatal Hypoglycemia-Controversies. PAEDIATRICS TODAY 2000; 3: 740-42.

(5.) Charles AS & Eugina KP: Disorders of Carbohydrate Metabolism. In: Avery Disease of the Newborn. Teuscch HW, Ballard RA & Gleason CA. 8th ed. Philadelphia: SAUNDERS, 2005; 1410-1420.

(6.) Kleigman RM: Problems In Metabolic Adaptation: Glucose, Calcium and Magnesium. In: Care Of the High Risk Neonate. Klaus MH & Faranoff AA. 5th ed. Philadelphia: SAUNDERS, 2000; 304-309.

(7.) Barbara JS and Robert MK: The Endocrine System. In: Nelson Text Book of Pediatrics. Behrman RE, Kliegman RM, Jenson HB. 17th ed. Philadelphia: SAUNDERS, 2004; 614-616.

(8.) Alkalay, Arie l, et al. Brain Imaging Findings in Neonatal Hypoglycemia: Case Report and Review of 23 Cases. Clin Pediatr (Phila) 2005; 44(9): 783-90.

(9.) Marvin Cornblath, Jane M Hawdon, et al. Controversies Regarding Definition of Neonatal Hypoglycemia: Suggested Operational Thresholds. Pediatrics 2000; 105: 1141-1145.

(10.) Gregory A. Threatte, John Bernard henry: Carbohydrates. In: Clinical Diagnosis and management by laboratory methods. JB Henry. 19th ed. Noida: Harcourt Asia PTE Ltd, 1999; 194-207.

(11.) Shankar Narayan, Rajiv Agarwal, Ashok K, K.P Vinod. Hypoglycemia in the newborn. Indian J Pediatr 2001; 68(10): 963-965.

(12.) Michel A, Kuster H et al. Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycemia. Eur J Pediatr 2005; 164(11): 660-664.

(13.) Pamela Kirkham, Andrew Watkins. Comparison of two reflectance photometers in the assessment of neonatal hypoglycemia. Arch Dis Child Fetal Neonatal Ed 1995; 73: F170-73.

(14.) Dahlberg, Marianne, et al. Comparison of a new instant blood glucose method with conventional laboratory analysis in diagnosis of neonatal hypoglycemia. Pediatr Res 1996; 39(4)307.

(15.) Mehta, Nilesh M, Whitelaw, Andrew, Stone, Janet. Diagnosis of neonatal hypoglycemia at the cot side. A comparison of precision Q.I.D., Hemocue and Hexokinase Methods. Pediatr Res 1999; 4594): 211A.

(16.) H T HO, W K Y YEUNG and B W Y Young. Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice. Arch Dis Child Fetal Neonatal Ed 2004; 89: F356.

(17.) Ellis M, Manadar DS, Manadar N, JM Land, NPatel, AM de L Costello. Comparision of two cotside methods for detection of hypoglycemia among neonates in Nepal. Arch Dis Child Fetal Neonatal Ed 1996; 75: F122-5.

(18.) Hamid MH, Christi AL, Maqbool S. Clinical utility and accuracy of a blood glucose meter for the detection of neonatal hypoglycemia. J Coll Physicians Surg Pak. 2004; 14(4): 225-8.

Harish J [1], Srinivas H. A [2], Soumya A [3]

AUTHORS:

[1.] Harish J.

[2.] Srinivas H. A.

[3.] Soumya A.

PARTICULARS OF CONTRIBUTORS:

[1.] Assistant Professor, Department of Paediatrics, Kempegowda Institute of Medical Sciences & Research Institute, Bangalore.

[2.] Assistant Professor, Department of Paediatrics, Kempegowda Institute of Medical Sciences & Research Institute, Bangalore.

FINANCIAL OR OTHER COMPETING INTERESTS: None

[3.] Post Graduate, Department of Paediatrics, Kempegowda Institute of Medical Sciences & Research Institute, Bangalore.

NAME ADDRESS EMAIL ID OF THE CORRESPONDING AUTHOR:

Dr. Soumya A, # 23, Flat No. 204, Lakshmi Paradise, 11th Cross, 5th Main, L. N. Pura, Bangalore-21. E-mail: soumyaasokan@ymail.com

Date of Submission: 11/12/2014.

Date of Peer Review: 12/12/2014.

Date of Acceptance: 13/02/2015.

Date of Publishing: 20/02/2015.

Table 1: Diagnostic value of Glucometer in relation to
Lab value in detecting the hypoglycemia

Value            Percentage

True Positive        43
True Negative       138
False Positive       9
False Negative       60
Sensitivity        41.75
Specificity        93.88
PPV                82.69
NPV                69.70

Table 2: Pearson correlation between Glucometer and Lab
value with lab value >45mg% and <45mg% groups

                                Number     Pearson      P value
                                         correlation
Glucometer vs. Lab (<45 mg %)    103        0.417      P<0.001 **
Glucometer vs. (>45 mg %)        147        0.758      P<0.001 **

Table 3: Pick-up rate of venous blood [Glucometer],
capillary blood [Glucometer] and Lab value

Methods                      Number    Pick-up rate
                                      of hypoglycemia

Venous blood-glucometer        50       14 (28.0%)
Capillary blood-glucometer     50       11 (22.0%)
Lab value                      50       28 (56.0%)

Comparison of overall incidence of hypoglycemia with other studies

STUDIES                       INCIDENCE

PK SINGHAL et al (3)            4.8%
PK MISHRA & BINA SHARMA (2)     9.7%
OUR STUDY                       41.2%
COPYRIGHT 2015 Akshantala Enterprises Private Limited
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2015 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Author:Harish, J.; Srinivas, H.A.; Soumya, A.
Publication:Journal of Evolution of Medical and Dental Sciences
Article Type:Report
Date:Feb 23, 2015
Words:3630
Previous Article:Role of sonography in evaluation of cervical lymph nodes and comparison with their cytological findings.
Next Article:Complication trends of thyroid surgery in a teaching institute.
Topics:

Terms of use | Privacy policy | Copyright © 2021 Farlex, Inc. | Feedback | For webmasters |