Printer Friendly

Clinical vitamin [B.sub.6] analysis: an interlaboratory comparison of pyridoxal 5'-phosphate measurements in serum.

The subject of vitamin [B.sub.6] intake and its relationship to indicators of adequate vitamin [B.sub.6] status is a contentious issue (1,2), particularly in the absence of a true reference method and certified reference materials for vitamin [B.sub.6] measurements. Clinical vitamin [B.sub.6] analysis is accomplished at present by a variety of enzymatic assays (3-6) and HPLC methods (7-14). Although some direct comparisons of these methods can be found in the literature (usually in the form of comparing results from a new method with an established one as a part of method validation), objective comparative studies of vitamin [B.sub.6] measurements among several methods or laboratories are rare. The only previously published account of a multi-laboratory comparative study of vitamin [B.sub.6] measurements in human biological specimens was conducted by Reynolds in 1983 (15), in which a single pooled plasma sample was sent to 14 US laboratories routinely performing [B.sub.6] analyses. Results from this study showed a wide range of values for the [B.sub.6] vitamer and metabolite content of the specimen. For pyridoxal 5'-phosphate (PLP), the primary circulatory form of vitamin [B.sub.6], 6 laboratories using enzymatic methods reported concentrations of 53-65 nmol/L for the specimen; the remaining laboratories, using HPLC, found PLP concentrations of 48-1700 nmol / L.

Because much interest currently surrounds the role vitamin [B.sub.6] plays in reducing the risk of coronary heart disease (16-18) and stroke (19-21), an objective comparison of the various methods used for clinical vitamin [B.sub.6] measurements would be prudent. In response, the US Centers for Disease Control and Prevention invited domestic and international clinicians and laboratorians who routinely measure vitamin [B.sub.6] in plasma or serum, as well as manufacturers of commercially available assays, to participate in a study designed to assess the performance and comparability of methods used for measuring vitamin [B.sub.6] in serum.

Materials and Methods

PARTICIPATING LABORATORIES

Laboratories were invited to participate in an interlaboratory comparison study that would entail the analysis of ~70 serum specimens for vitamin [B.sub.6] over 3 days. The minimum requirement for participation was the ability to measure the PLP concentrations of these specimens, although measurement of other [B.sub.6] vitamers and metabolites was encouraged. This invitation was extended to a total of 25 laboratories.

SERUM SAMPLES

Serum samples distributed to participants in this study were prepared from serum pools obtained from Solomon Park Research Institute and from pools generated in house. All in-house serum collection was conducted in accordance with an internal review-board-approved human subjects protocol. In-house serum pools were created from sera collected from on-site volunteers by venipuncture into evacuated collection tubes (Vacutainer SST Tubes; Becton Dickinson). Approximately 30 mL of serum was collected from each volunteer. Sera were processed according to the instructions provided by the collection tube manufacturer with all sample processing conducted under yellow lighting at room temperature. Once processed, a 350-[micro]L aliquot of each serum sample was reserved for screening purposes, and the bulk of each sample was stored immediately at -80[degrees]C. Serum samples were screened for their PLP concentrations in house by a validated method (7), and pools were proposed based on these preliminary values. Serum samples were then thawed, combined in various proportions, and filtered through sterile gauze to remove fibrin and other particulates. Pools were then dispensed while being stirred and immediately stored at -80[degrees]C. Serum pools from Solomon Park Research Institute were received as blinded samples in 1-mL aliquots. Samples were drawn at random from these pools and screened in house by a validated method (7) for their PLP concentrations and homogeneity among aliquots (2 specimens/day x 6 days).

Participating laboratories received a total of 69 serum samples divided into 3 groups (days) according to the schedule appearing in Table 1. Each sample consisted of a 1-mL aliquot in a 2-mL cryogenic polypropylene vial (Sarstedt) labeled with an encrypted 7-digit identifier. Each day also included an aqueous sample consisting of 40.4 nmol/L PLP (Sigma-Aldrich) dissolved in distilled deionized water (AquaSolutions) filtered through a 0.45 m filter. The aqueous sample was prepared by serial dilution of a 100 mg/L aqueous stock solution prepared from dry PLP, and its concentration was verified spectro-photometrically by absorbance measurements at 388 nm ([epsilon] = 5070 L x [cm.sup.-1] x [mol.sup.-1]) (22) made on intermediate dilutions (10, 20, and 30 mg/L) of the stock solution in 25 mmol/L phosphate buffer (pH 7.0). Samples were frozen and stored at -70[degrees]C before distribution and were shipped within 48 h over dry ice to participants. Participants were instructed to treat all serum and aqueous samples as unknown specimens. Laboratories were also instructed to analyze the samples in groups (days) as indicated; however, samples within a specific group could be analyzed in any order. Beyond these instructions and some general specimen handling and storage recommendations, participants were free to analyze the samples as determined by their methods. Laboratories were given ~3 weeks from the date that samples were received to report their results.

STATISTICAL ANALYSIS

All statistical analyses were performed with SAS 9.0 (SAS Institute Inc.). Except in the calculation of consensus means for each of the serum pools (for which observations had to meet certain statistical criteria to be included), all calculations described here were performed using all reported observations.

Within-day imprecision was evaluated by determining the CVs for each of pools 21, 22, and 23. Mean within-day imprecision was calculated as the mean CV (SD) with 95% confidence limits (CLs) for these pools. Mean between-day imprecision was calculated in an analogous manner, using the CVs calculated for each of pools 1-20.

Analytical recovery was evaluated both by the addition of exogenous PLP at 2 concentrations and by mixing serum pools. Recoveries of exogenous PLP were calculated on each day as follows:

Recovery (%) = [sample + exogenous PLP] - [sample]/[added PLP]

= [pool 14] - [pool 8]/32.1 nmol/L (for concentration 1)

= [pool 16] - [pool 8]/64.2 nmol/L (for concentration 2) (1)

Serum pool mixing recovery was calculated on each day as follows:

Recovery (%) = [1:1 mixture of samples A & B]/([sample A] + [sample B])/2

= [pool 15]/([pool 4] + [pool 18])/2 (2)

Mean recoveries were expressed as the mean (SD) with 95% CLs of the individual recoveries calculated for each day.

In the absence of PLP reference values for the serum pools used in this study, consensus means were calculated for each pool and used as surrogates. Consensus means were determined by a recursive testing procedure for outlying observations. The means and SDs for serum pools 1-20 were first calculated using all reported observations. Using a 99% probability (mean [+ or -] 2.576 SD), outlying observations were rejected and the consensus means and SDs were recalculated. This process was repeated until all observations fell within the 99% probability limits of their respective consensus means, or until 3 iterations of this process had been completed, whichever occurred first.

Absolute laboratory measurement bias was calculated by determining the differences between the consensus means and the results from each laboratory for serum pools 1-20 over all 3 days. Relative bias was assessed in an identical manner by use of relative ratios. Overall absolute and relative biases were reported as the means (SDs) with 95% CLs.

PERFORMANCE CRITERIA BASED ON BIOLOGICAL VARIATION

Between-day imprecision and relative bias were evaluated against empirically derived performance criteria based on biological variation. Analytical variability ([CV.sub.A]) was assessed using the following equation (23), with the mean between-day CV taken to be equivalent to [CV.sub.A]:

[CV.sub.A] < (n)[CV.sub.within-individual] (3)

where n = 0.25 (optimum), 0.50 (desirable), or 0.75 (minimum). On the basis of a within-individual CV of 45% observed for PLP in human plasma (24), between-day CVs <11.3%, <22.5%, and <33.8% were needed to satisfy the performance criteria of optimum, desirable, and minimum, respectively.

Proportional bias ([B.sub.a]) was assessed using the following equation (23):

[B.sub.a] < (n) [square root of [CV.sup.2.sub.within-individual] + [CV.sup.2.sub.between-individual]] (4)

where n = 0.125 (optimum), 0.250 (desirable), or 0.375 (minimum). To determine the performance criteria for relative bias, a between-individual CV for PLP in serum was calculated using in-house data from a nonrepresentative population subset of 301 individuals on an uncontrolled diet (7), yielding a value of 79%. From Eq. 4, relative bias was required to be <11.4%, <22.7%, and 34.1% (absolute) to satisfy optimum, desirable, and minimum performance criteria, respectively.

Results

PARTICIPATING LABORATORIES AND DATA REPORTING

Of the 25 laboratories invited to participate in the present study, 12 agreed to participate and 10 remained in the study through completion. Participants consisted of 3 academic research institutions, 3 clinical research institutions, 3 manufacturers, and 1 government laboratory. Six of the participating laboratories were located in North America, 3 were located in Europe, and 1 was located in Asia. Eight of the laboratories used HPLC-based assays, with the remaining 2 using enzymatic methods. Participating laboratories were assigned an alphanumeric code at random, with the prefix letter indicating whether an HPLC (H) or enzymatic (E) method was used. Laboratory identifiers and respective method details are given in Table 2.

The PLP concentrations reported by each participating laboratory for the serum pools are presented in the Data Supplement that accompanies the online version of this article at http://www.clinchem.org/content/vol51/ issue7/(Table 51 for between-day serum pools, Table 52 for within-day serum pools, Table 53 for aqueous reference pool). The results reported for laboratories H2, H6, H7, E1, and E2 represent the mean of 2 replicates, whereas all other results are from single measurements. In addition to PLP, results for 4-pyridoxic acid were reported by laboratories H4, H6, and H7. Laboratory H4 also reported values for pyridoxal, pyridoxamine, and pyridoxine. Meaningful interlaboratory statistical analysis of these results was not possible because of the small number of laboratories reporting values for these analytes; consequently, these data were not included in this report.

IMPRECISION

Laboratory within- and between-day imprecision is summarized in Table 3. Mean within-day CVs of 0.6%-37% and between-day CVs of 1.4%-26% were observed among participants. The performance criteria met by the mean between-day CVs for each laboratory appear in Table 3 (if different from its corresponding mean, the criterion met by the upper 95% CL of the between-day CV appears in parentheses). All 10 laboratories met at least minimum performance requirements, with 4 HPLC laboratories (H1, H2, H6, and H8) achieving optimum status.

RECOVERY

Standard addition recoveries for PLP ranged from 56% to 128% at an added concentration of 32.1 nmol/L, and from 53% to 144% at an added concentration 64.2 nmol/L (Table 4). Recoveries observed within each laboratory were generally better at the higher of the 2 added concentrations. Of a1110 participants, only 2 laboratories (H3 and H8) achieved a mean standard addition recovery within 10% of quantitative recovery (100%) at the lower added concentration, but 5 laboratories (H2, H3, and H6-H8) were within 10% of quantitative recovery at the second concentration. Mean mixing recoveries ranged from 75% to 119% (Table 4). Six laboratories (H1-H3 and H6-H8) had mean mixing recoveries within 10% of quantitative recovery.

CONSENSUS MEANS

Because the serum pools used in this study lacked PLP reference values and a true reference method for determining PLP does not exist, consensus means were calculated from the results submitted by participating laboratories for pools 1-20 and used as surrogates. Preliminary inspection of these results indicated the possible existence of extreme outlying observations (Table S1 of the online Data Supplement). To identify outlying observations and thus eliminate the bias they may introduce into the consensus mean calculations, an iterative procedure of recursive testing of outlying observations was used. To illustrate the process of identifying outlying observations and calculating consensus means, the normalized difference of the observations from their respective consensus means are plotted in Fig. 1. Of the 20 pools for which consensus means were calculated, 8 pools had all observations fall within the 99% probability limits with the initial mean calculation, and 7 of the remaining 12 pools required only 1 iteration of the consensus means calculation procedure. Of the 10 participating laboratories, 5 using HPLC methods (H1-H3, H6, and H8) and 1 using an enzymatic assay (E1) reported results that satisfied the 99% probability inclusion criteria for the consensus mean calculations of all serum pools (Table 51 of the online Data Supplement). A total of 25 observations from the remaining 4 laboratories were rejected over the course of the consensus mean calculations; 16 observations were rejected in the first iteration, 6 in the second, and 3 in the third (Fig. 1). After 3 iterations of the consensus mean calculation procedure, only pool 2 had 1 remaining outlying observation based on 99% probability limits. PLP consensus means for the 20 serum pools ranged from 15.8 to 319 nmol/L (mean, 84.2 nmol/L; median, 55.2 nmol/L; Table S1 of the online Data Supplement). Between-laboratory variability (CV) within each consensus mean ranged from 12% to 37% (mean, 21%; median, 21%) after exclusion of outlying values.

[FIGURE 1 OMITTED]

Bias

Mean absolute and relative measurement biases among participants ranged from -7.7 nmol/L to 18.6 nmol/L and from -10.0% to 30.6%,respectively (Table 5). Relative bias was also evaluated against performance criteria related to biological variation (23). When applied to the mean relative bias, all 10 laboratories satisfied the requirements of minimum performance, with 8 of the 10 also meeting the requirements of desirable performance and 6 of the 10 achieving optimum performance (Table 5). Use of the mean relative bias in this case can be somewhat misleading, however, because Eq. 4 does not take into consideration the error associated with the mean relative bias. As a consequence, the mean relative bias of a method may satisfy a certain performance criterion although many of the individual observations made with the method may not. A more meaningful way to assess performance bias is to look at the relative bias of each individual observation and to determine the proportion of results from a laboratory that satisfy the performance criteria. These results appear in Table 5 as cumulative percentages of the observations from each laboratory that were deemed to meet the performance criteria stated above. Laboratories H2, H6, and H8 had 90% or more of their results satisfy optimum performance criteria, and 100% of the results from these 3 laboratories satisfied desirable performance criteria. All observations from H1 and H3, as well as the above-mentioned laboratories, met minimum performance criteria for bias. For laboratory E1, all but 2 observations also satisfied the minimum performance criteria (98%). The 4 remaining laboratories (H4, H5, H7, and E2) had results that failed to meet the minimum bias requirement, with the proportion of results failing to achieve the minimum varying from 25% to 53%.

Discussion

This study has afforded the opportunity to compare PLP measurements in human serum among 10 domestic (United States) and international laboratories using a variety of analytical methods. Participants represented a cross-section of academic, clinical, governmental, and commercial entities using a diverse selection of enzymatic and HPLC-based assays. The study design, which entailed analyzing 72 blinded specimens over 3 days, permitted the objective comparison of measurement imprecision, analyte recovery, and measurement bias among laboratories. Individual laboratory performance was also evaluated against empirically derived performance criteria based on biological variation. Because of the relatively small number of participants and the unique nature of each of the analysis methods used, however, laboratory-and method-specific statements and conclusions contained herein are to a certain extent presumptive.

Imprecision varied greatly among laboratories. Within-day CVs of 0.4%-37% and between-day CVs of 1.4%-26% were observed among participants. Cases were also observed in which the mean within-day CV for a laboratory apparently exceeded its mean between-day CV; these likely were artifacts arising from the relatively small number of pools used to assess within-day imprecision (n = 3) compared with between-day imprecision (n = 20). In comparison, data from several published methods for measuring PLP in human plasma or serum suggest that within- and between-day CVs are typically 0.5%-5.9% and 3.6%-11.8%, respectively, for HPLC methods (2, 7-9,13,14) and 3.0%-11% and 7.9%-13%, respectively, for enzymatic methods (3-5). It should be noted, however, that the number of within- or between-day measurements used for determining CVs in these studies was considerably greater (typically 10-20 per specimen in each case) than the number of measurements used in the present study. The 4 HPLC laboratories that achieved optimum imprecision (H1, H2, H6, and H8) had mean within-day CVs (0.6%-3.1%) and between-day CVs (1.4%-5.6%)that were consistent with the lower ends of the HPLC ranges found in the literature. Laboratory H3, which demonstrated desirable imprecision, had a within-day CV that was also consistent with the aforementioned laboratories (2.0%), but its between-day CV was toward the high end of the HPLC range (12%). Although the remaining laboratories using HPLC-based and enzymatic assays achieved either desirable (H7, E1, and E2) or minimum (H4 and H5) imprecision, their mean within- and between-day CVs exceeded the expected imprecision based on methods in the literature, suggesting that quality control issues may exist in these laboratories.

Serum pool mixing recoveries were generally quantitative and consistent among laboratories. The mean between-laboratory serum pool mixing recovery was 99.1%, and 5 HPLC laboratories (H1-H3, H6, and H8) had serum pool mixing recoveries within 3% of being quantitative (100%). Laboratories whose mixing recoveries appeared to deviate significantly from 100% (H4, E1, and E2) suggest that these assays may be subject to concomitant interferentes, particularly if the confidence interval associated with the recovery is relatively small.

Recoveries in standard addition experiments were more variable than in mixing recoveries and tended to be <100%. The mean between-laboratory recovery was 91.2% when an effective PLP concentration of 32.1 nmol/L was added to a serum pool with an endogenous PLP concentration of ~49 nmol/L. The mean between-laboratory recovery improved to 97.6% when the amount of PLP added was increased to 62.4 nmol/L. These recoveries seem to be consistent with the literature; a selection of published methods for PLP suggest that analyte addition recoveries of 88%-103% (between-method mean of 95%) are typical for HPLC methods (2, 7, 9-13, 25) and recoveries of 83%-125% (between-method mean of 99%) can be expected from enzymatic methods (3-5). Because the recovery calculation was based on a gravimetrically determined amount of PLP and the actual amount of the PLP addition was not measured in each laboratory, actual recoveries in each laboratory may vary slightly depending on calibration accuracy. Results from the analysis of the aqueous specimen seem to support this (Table S3 of the online Data Supplement).

The 4 laboratories that met optimum performance criteria for measurement imprecision (H1, H2, H6, and H8) also met desirable bias performance criteria for a high proportion (100% for H2, H6, and H8; 97% for H1) of their PLP measurements. Three of these 4 laboratories also met optimum performance criteria for a high proportion of measurements (90% H2 and H6; 95% for H8). Information on method bias in the literature is sparse, however, making comparisons with other studies difficult. Bias in PLP measurements is most frequently evaluated through the direct comparison of 2 methods, typically as part of a method validation. Consequentially, the bias of a method is usually expressed relative to measurements made with another method, as opposed to more objective benchmarks such as reference values or consensus means. Although no PLP reference value is given for the plasma specimen used in the interlaboratory comparison study conducted by Reynolds (15), the extended high range observed for the HPLC values (48-1700 nmol/L) compared with those obtained with enzymatic assays (53-65 nmol/L) implies that PLP measurements made by HPLC in the study may be inherently biased high. This appears not to be the case, however, for the HPLC laboratories in the present study that satisfy optimum performance criteria for bias. Slightly negative mean absolute (-6.1 nmol/L to -0.8 nmol/L; between-method mean of -3.4 nmol/L) and relative (-6.1% to -0.9%; between-method mean of -4.8%) biases were observed for these 3 laboratories.

Four HPLC laboratories (H1, H2, H6, and H8) were found to have very similar characteristics in terms of measurement imprecision and bias. This is interesting from the standpoint that, although these 4 methods are based on common instrumental principles, they are markedly different in terms of sample preparation, chromatographic separation schemes, and in particular, PLP derivatization approaches. This seems to suggest that the derivatization approaches used by these laboratories [bisulfite (11), chlorite (7), and semicarbazide (13,14)] to increase the fluorescence sensitivity of PLP were all equally effective for quantifying PLP in this study and that quality control within the laboratory is the single most important issue. Although a cursory review of the mean bias and imprecision statistics for laboratory H3 suggests that its overall performance was inferior to the performance of the above-mentioned laboratories, closer inspection of these data suggests otherwise. The results for days 2 and 3 from laboratory H3 correlated extremely well within each specimen pool, whereas the results from day 1 were consistently biased low relative to days 2 and 3, suggesting calibration inconsistencies between days. Interestingly, the results from day 1 for laboratory H3 appear to be consistent with those reported for laboratories H1, H2, H6, and H8 [mean (SD) absolute bias of -2.7 (3.8) nmol/L, mean relative bias of -4.6 (4.7)%], whereas the results from days 2 and 3 appear to be biased high. The HPLC laboratories that demonstrated relatively high degrees of imprecision and bias in their assays (H4, H5, and H~ may have issues in terms of quality control and, possibly, nonspecific analyte interferentes.

Imprecision in the laboratories using enzymatic methods was generally higher than that observed in the HPLC laboratories, a difference that is most likely a result of fundamental differences in analytical methodologies. All results from laboratory E1 met the inclusion criteria for the consensus mean calculations and the minimum performance standard for relative bias. This was not the case for laboratory E2, although their mean between-day imprecision was quite similar to that observed for laboratory E1. The presence of extreme outlying observations with good reproducibility (Table 51 of the online Data Supplement) suggests that concomitant analyte interferentes may exist in the assay used by laboratory E2.

The overall results of this study suggest that quality control and calibration accuracy are the 2 largest issues that laboratories performing vitamin [B.sub.6] assays face. Regular studies such as this, or the establishment of an external quality assurance program with a wide participation base, would greatly assist laboratories in identifying and resolving issues with their methods. The development of standard reference materials for vitamin [B.sub.6] would also be a more objective means of assessing method bias. The development and validation of true reference methods for vitamin [B.sub.6] that are capable of confirming the identities of analytes, such as isotope-dilution liquid chromatography-tandem mass spectrometry, should also be pursued.

We would like to sincerely thank the following individuals for their participation and support in this study: Robert M. Hoffman, Li Tang, and Quinghong Han (A/C Diagnostics, San Diego, CA); Christian Niederberger and Roger Sokoll (Buhlmann Laboratories AG, Basel, Switzerland); Donna LaVoie (US Centers for Disease Control and Prevention, Atlanta, GA); Martin Seiler and H. Fischer (Chromsystems GmbH, Munich, Germany); Yi-Chia Huang and Ping-Ting Lin (School of Nutrition, Chung Shan Medical University, Taichung, Taiwan); Christine M. Hansen and M. Beanie (Department of Food Science and Human Nutrition, Iowa State University, Ames, IA); Cynthia L. Deitrick and Anna Moore (Lipid Analysis Inc., Springfield, IL); Joseph P. McConnell, Mary Hooper, and Betty Rommell (Biochemical Genetics Laboratory, Mayo Clinic, Rochester, MN); W.H.J. Vaes and Barbara Veldhuis (Nutrition and Food Research, TNO, Zeist, The Netherlands); and Jesse F. Gregory, Steven R. Davis, and Jennifer B. Scheer (Department of Food Science and Human Nutrition, University of Florida, Gainesville, FL).

References

(1.) Leklem JE. Vitamin B-6: a status report. J Nutr 1990;120: S1503-7.

(2.) Bailey AL, Wright AJA, Southon S. High performance liquid chromatography method for the determination of pyridoxal-5-phosphate in human plasma: how appropriate are cut-off values for vitamin B-6 deficiency. Eur J Clin Nutr 1999;53: 448-55.

(3.) Han Q, Xu M, Tang L, Tan X, Tan X, Tan Y, et al. Homogeneous, nonradioactive, enzymatic assay for plasma pyridoxal 5-phosphate. Clin Chem 2002;48: 1560-4.

(4.) Shin YS, Rasshofer R, Friedrich B, Endres W. Pyridoxal-5'-phosphate determination by a sensitive micromethod in human blood, urine and tissues; its relation to cystathioninuria in neuro-blastoma and biliary atresia. Clin Chim Acta 1983;127: 77-85.

(5.) Buehring YSS, Rasshofer R, Endres W. A new enzymatic method for pyridoxal-5-phosphate determination. J Inherit Metab Dis 1981;4: 123-4.

(6.) Phan APH, Ngo TT, Lenhoff HM. Tyrosine decarboxylase spectro-photometric assay and application in determining pyridoxal-5'phosphate. Appl Biochem Biotechnol 1983;8: 127-33.

(7.) Rybak ME, Pfeiffer CM. Clinical analysis of vitamin [B.sub.6]: determination of pyridoxal 5'-phosphate and 4-pyridoxic acid in human serum by reversed-phase high-performance liquid chromatography with chlorite postcolumn derivatization. Anal Biochem 2004;333: 336-44.

(8.) Bisp MR, Bor MV, Heinsvig E-M, Kall MA, Nexo E. Determination of vitamin [B.sub.6] vitamers and pyridoxic acid in plasma: development and evaluation of a high-performance liquid chromatographic assay. Anal Biochem 2002;305: 82-9.

(9.) Deitrick CL, Katholi RE, Huddleston DJ, Hardiek K, Burrus L. Clinical adaptation of a high-performance liquid chromatographic method for the assay of pyridoxal 5'-phosphate in human plasma. J Chromatogr B 2001;751: 383-7.

(10.) Bates CJ, Pentieva KD, Matthews N, Macdonald A. A simple, sensitive and reproducible assay for pyridoxal 5'-phosphate and 4-pyridoxic acid in human plasma. Clin Chim Acta 1999;280: 101-11.

(11.) Kimura M, Kanehira K, Yokoi K. Highly sensitive and simple liquid chromatographic determination in plasma of [B.sub.6] vitamers, especially pyridoxal 5'-phosphate. J Chromatogr A 1996;722: 295-301.

(12.) Sharma SK, Dakshinamurti K. Determination of vitamin [B.sub.6] vitamers and pyridoxic acid in biological samples. J Chromatogr 1992; 578: 45-51.

(13.) Ubbink JB, Serfontein WJ, de Villers LS. Stability of pyridoxal-5'phosphate semicarbazone: applications in plasma vitamin [B.sub.6] analysis and population surveys of vitamin [B.sub.6] nutritional status. J Chromatogr 1985;342: 277-84.

(14.) Schrijver J, Speek AJ, Schreurs WHP. Semi-automated fluorometric determination of pyridoxal-5'-phosphate (vitamin [B.sub.6]) in whole blood by high-performance liquid chromatography (HPLC). Int J Vitam Nutr Res 1981;51: 216-22.

(15.) Reynolds RD. Nationwide assay of vitamin B-6 in human plasma by different methods [Abstract]. Fed Proc 1983; 42:665.

(16.) Bleie O, Refsum H, Ueland PM, Vollset SE, Guttormsen AB, Nexo E, et al. Changes in basal and postmethionine load concentrations of total homocysteine and cystathionine after B vitamin intervention. Am J Clin Nutr 2004;80: 641-8.

(17.) Friso S, Girelli D, Martinelli N, Olivieri 0, Lotto V, Bozzini C, et al. Low plasma vitamin B-6 concentrations and modulation of coronary artery disease risk. Am J Clin Nutr 2004;79: 992-8.

(18.) Stanger 0, Herrmann W, Pietrzik K, Fowler B, Geisel J, Dierkes J, et al. Clinical use and rational management of homocysteine, folic acid, and B vitamins in cardiovascular and thrombotic diseases. Z Kardiol 2004;93: 439-53.

(19.) Kelly PJ, Shih VE, Kistler JP, Barron M, Lee H, Mandell R, et al. Low vitamin [B.sub.6] but not homocyst(e)ine is associated with increased risk of stroke and transient ischemic attack in the era of folic acid grain fortification. Stroke 2003;34: E51-4.

(20.) Kelly PJ, Kistler JP, Shih VE, Mandell R, Atassi N, Barron M, et al. Inflammation, homocysteine, and vitamin [B.sub.6] status after ischemic stroke. Stroke 2004;35: 12-5.

(21.) Schwammenthal Y, Tanne D. Homocysteine, B-vitamin supplementation, and stroke prevention: from observational to interventional trials. Lancet Neurol 2004;3: 493-5.

(22.) Sigma-Aldrich. Certificate of analysis, pyridoxal 5'-phosphate. Product no. P9255; lot no. 041K1088. St. Louis, M0: Sigma-Aldrich.

(23.) Fraser CG, Petersen PH, Libeer J-C, Ricos C. Proposals for setting generally applicable quality goals solely based on biology. Ann Clin Biochem 1997;34: 8-12.

(24.) Bor MV, Refsum H, Bisp MR, Bleie O, Schneede J, Nordrehaug JE, et al. Plasma vitamin [B.sub.6] vitamers before and after oral vitamin [B.sub.6] treatment: a randomized placebo-controlled study. Clin Chem 2003;49: 155-61.

(25.) Millart H, Lamiable D. Determination of pyridoxal 5'-phosphate in human serum by reversed-phase high-performance liquid chromatography combined with spectrofluorimetric detection of 4-pyri doxic acid 5'-phosphate as a derivative. Analyst 1989;114: 1225-8.

Michael E. Rybak, Ram B. Jain, and Christine M. Pfeiffer *

Inorganic Toxicology and Nutrition Branch, Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA.

* Address correspondence to this author at: Centers for Disease Control and Prevention, 4770 Buford Hwy, Atlanta, GA 30341. Fax 770-488-4139; e-mail CPfeiffer@cdc.gov.

Received February 28, 2005; accepted May 2, 2005.

Previously published online at DOI: 10.1373/clinchem.2005.050278
Table 1. Study design: Sample pools.

 Day

Pool Purpose (a) 1 2 3

 0 R [check] [check] [check]

 1 B, D [check] [check] [check]
 2 B, D [check] [check] [check]
 3 B, D [check] [check] [check]
 4 B, D, M [check] [check] [check]
 5 B, D [check] [check] [check]
 6 B, D [check] [check] [check]
 7 B, D [check] [check] [check]
 8 B, D, S [check] [check] [check]
 9 B, D [check] [check] [check]
 10 B, D [check] [check] [check]
 11 B, D [check] [check] [check]
 12 B, D [check] [check] [check]
 13 B, D [check] [check] [check]
 14 B, D, S [check] [check] [check]

 15 B, D, M [check] [check] [check]

 16 B, D, S [check] [check] [check]

 17 B, D [check] [check] [check]
 18 B, D, M [check] [check] [check]
 19 B, D [check] [check] [check]
 20 B, D [check] [check] [check]
 21 W [check] (x3)
 22 W [check] (x3)
 23 W [check] (x3)

Pool Purpose (a) Notes

 0 R Pool 0 = 40.4 nmol/L PLP
 aqueous solution
 1 B, D
 2 B, D
 3 B, D
 4 B, D, M
 5 B, D
 6 B, D
 7 B, D
 8 B, D, S
 9 B, D
 10 B, D
 11 B, D
 12 B, D
 13 B, D
 14 B, D, S Pool 14 = pool 8 = 32.1
 nmol/L PLP
 15 B, D, M Pool 15 = 1:1 mixture of
 pools 4 and 18
 16 B, D, S Pool 16 = pool 8 = 64.2
 nmol/L PLP
 17 B, D
 18 B, D, M
 19 B, D
 20 B, D
 21 W
 22 W
 23 W

(a) B, bias assessment; D, between-day imprecision;
M, mixing recovery; R, aqueous reference;
S, addition recovery; W, within-day imprecision.

Table 2. Method characteristics.

 Derivatization
Laboratory Method type (HPLC methods) (a) Analyte(s)

 H1 HPLC-fluorescence Bisulfite, on-line PLP
 H2 HPLC-fluorescence Semicarbazide, sample PLP, PL (b)
 H3 HPLC-fluorescence Bisulfite, on-line PLP, PM
 H4 HPLC-fluorescence Bisulfite, post-column PLP, PL, PM,
 PN, 4-PA
 H5 HPLC-fluorescence PLP
 H6 HPLC-fluorescence Chlorite, post-column PLP, 4-PA
 H7 HPLC-fluorescence Cyanide, sample PLP, 4-PA
 H8 HPLC-fluorescence Semicarbazide, sample PLP
 E1 Enzymatic, PLP
 radioactive
 E2 Enzymatic, PLP
 nonradioactive

 Specimen
 volume No. of Calibration
 requirement measurements range,
Laboratory per assay per sample nmol/L PLP

 H1 500 1 0-200
 H2 400 2 10-160
 H3 500 2 5-200
 H4 300 2 25-200
 H5 1000 2 6-70
 H6 150 2 2-200
 H7 160 2 10-150
 H8 1000 1 5-75
 E1 200 2 13.5-216
 E2 5 2 10-200

 No. of
 tests
 performed
Laboratory per month Reference(s)

 H1 400 (11)
 H2 100 (13) (modified)
 H3 40 (9)
 H4 (12)
 H5 50 Proprietary
 H6 300 (7)
 H7 150 (10)
 H8 50 (14)
 E1 10-15 (4, 5) (modified, proprietary)
 E2 500 (3)

(a) Refers to derivatization technique used (if any) to enhance
the fluorescence sensitivity of PLP.

(b) PL, pyridoxal; PM, pyridoxamine; PN, pyridoxine; 4-PA,
4-pyridoxic acid.

Table 3. Laboratory imprecision.

 CV, %

 Within-day Between-day
 Performance
 Mean Mean criteria
Laboratory (SD) 95% CLs (SD) 95% CLs met (a)

 H1 3.1 (1.7) 1.1; 5.1 5.6 (3.2) 4.2; 7.1 Optimum
 H2 2.7 (0.2) 2.6; 2.9 3.1 (2.3) 2.0; 4.1 Optimum
 H3 2.0 (0.2) 1.7; 2.3 12 (0.8) 11; 12 Desirable
 H4 37 (20) -14; 60 26 (15) 19; 33 Minimum
 H5 33 (41) -15; 81 25 (15) 18; 32 Minimum
 H6 0.6 (0.4) 0.1; 1.1 1.4 (0.7) 1.1; 1.7 Optimum
 H7 9.7 (5.8) 3.0; 16 16 (8.9) 12; 20 Desirable
 H8 0.7 (0.8) -0.2; 1.6 3.9 (1.8) 3.0; 4.7 Optimum
 E1 24 (29) -9.8; 58 15 (8.3) 11; 19 Desirable
 E2 9.1 (7.0) 1.0; 17 19 (10) 14; 24 Desirable
 (minimum)

(a) Based on mean between-day imprecision. If different from mean,
performance criterion met by upper 95% CL appears in parentheses.
See Results and Discussion sections for details.

Table 4. Analytical recoveries.

 Standard addition recoveries, %

 Concentration 1 (32.1 nmol/L)

Laboratory Mean (SD) 95% CLs

 H1 75.2 (10.7) 70.5; 79.9
 H2 88.9 (4.8) 86.8; 91.0
 H3 109.3 (18.1) 101.4; 117.3
 H4 75.0 (51.5) 52.5; 97.4
 H5 128.4 (99.4) 85.1; 171.8
 H6 88.5 (5.3) 86.2; 90.9
 H7 72.8 (82.2) 36.9; 108.6
 H8 95.5 (7.2) 92.4; 98.7
 E1 122.4 (38.5) 105.6; 139.2
 E2 56.2 (12.1) 50.9; 61.5

 Standard addition recoveries, %

 Concentration 2 (64.2 nmol/L)

Laboratory Mean (SD) 95% CLs

 H1 84.8 (5.4) 82.5; 87.2
 H2 97.1 (1.8) 96.3; 97.9
 H3 106.7 (13.1) 101.0; 112.4
 H4 52.7 (28.1) 37.6; 67.7
 H5 144.2 (54.4) 120.5; 167.9
 H6 93.8 (3.0) 92.5; 95.1
 H7 94.4 (23.6) 84.1; 104.7
 H8 94.6 (6.3) 91.9; 97.4
 E1 133.8 (12.0) 128.5; 139.0
 E2 74.2 (14.6) 67.8; 80.6

 Serum pool mixing recovery, %

Laboratory Mean (SD) 95% CLs

 H1 102.8 (5.5) 100.4; 105.2
 H2 102.6 (2.8) 101.4; 103.8
 H3 100.8 (1.3) 100.2; 101.4
 H4 119.1 (13.2) 112.0; 126.1
 H5 110.9 (27.2) 99.0; 122.7
 H6 99.9 (0.7) 99.6; 100.2
 H7 92.9 (16.4) 85.7; 100.0
 H8 100.2 (2.1) 99.3; 101.1
 E1 86.9 (2.2) 85.9; 87.8
 E2 75.1 (19.6) 66.6; 83.6

Table 5. Laboratory bias.

 Absolute (SD) bias, nmol/L

Laboratory Day 1 Day 2 Day 3 Mean

 H1 6.6 (5.7) 9.0 (5.3) 6.0 (5.4) 7.7 (6.8)
 H2 3.1 (3.5) 4.4 (2.3) 3.3 (3.5) 3.4 (3.4)
 H3 2.7 (3.8) 13.5 (10.8) 16.2 (17.9) 10.1 (17.3)
 H4 9.6 (15.4) 21.6 (28.7) 19.7 (25.9) 18.6 (25.6)
 H5 10.6 (29.1) 1.6 (9.4) 29.2 (38.0) 14.1 (29.9)
 H6 6.0 (4.9) 6.3 (5.5) 5.5 (4.8) 6.1 (5.1)
 H7 16.6 (36.4) 8.3 (15.1) 27.0 (37.6) 15.6 (34.8)
 H8 0.4 (2.5) 0.4 (2.5) 3.6 (4.9) 0.8 (4.3)
 E1 5.1 (9.7) 4.5 (17.8) 1.8 (8.8) 0.4 (13.2)
 E2 4.1 (64.8) 2.2 (68.6) 10.7 (65.7) 4.9 (66.6)

 Relative (SD) bias, %

Laboratory Day 1 Day 2 Day 3 Mean

 H1 7.6 (6.2) 13.7 (5.5) 8.9 (6.3) 10.0 (6.4)
 H2 3.8 (3.2) 7.9 (5.7) 6.3 (4.8) 5.9 (4.9)
 H3 4.6 (4.7) 17.5 (6.2) 17.2 (6.4) 10.1 (12.0)
 H4 6.2 (41.5) 32.6 (48.3) 24.8 (26.5) 24.3 (36.6)
 H5 1.6 (23.7) 0.3 (14.3) 38.5 (44.9) 13.4 (34.9)
 H6 7.5 (3.6) 8.5 (3.6) 6.8 (4.1) 7.5 (3.8)
 H7 28.8 (49.8) 19.3 (33.3) 46.1 (43.3) 30.6 (43.9)
 H8 0.4 (4.1) 0.8 (4.4) 4.0 (6.0) 0.9 (5.3)
 E1 12.5 (14.7) 0.6 (22.7) 0.5 (9.9) 3.8 (17.5)
 E2 35.2 (126) 24.1 (126) 10.4 (117) 22.2 (119)

 Results meeting performance criteria, (b) %
Performance
 criteria Below
 met (a) Optimum Desirable Minimum minimum (c)

 Optimum 62 97 100 0
 Optimum 90 100 100 0
 Optimum 43 88 100 0
 Minimum 25 50 61 38 (11)
 Desirable 33 60 75 25 (5)
 Optimum 90 100 100 0
 Minimum 32 45 65 35 (12)
 Optimum 95 100 100 0
 Optimum 48 78 96 4
 Desirable 10 27 47 53 (15)

(a) Based on mean relative bias.

(b) See Results and Discussion sections for details.

(c) Percentage of laboratory observations excluded from consensus
mean calculations during recursive outlier testing appear in
parentheses.
COPYRIGHT 2005 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2005 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Automation and Analytical Techniques
Author:Rybak, Michael E.; Jain, Ram B.; Pfeiffer, Christine M.
Publication:Clinical Chemistry
Date:Jul 1, 2005
Words:6353
Previous Article:High-throughput purification of viral RNA based on novel aqueous chemistry for nucleic acid isolation.
Next Article:The REALAB project: a new method for the formulation of reference intervals based on current data.
Topics:


Related Articles
High-dose vitamin B6 decreases homocysteine serum levels in patients with schizophrenia and schizoaffective disorders: a preliminary study.
Pyridoxal phosphate decreases in plasma but not erythrocytes during systemic inflammatory response.
Routine isotope-dilution liquid chromatography--tandem mass spectrometry assay for simultaneous measurement of the 25-hydroxy metabolites of vitamins...
Homogeneous enzymatic colorimetric assay for total cysteine.
Homogeneous, nonradioactive, enzymatic assay for plasma pyridoxal 5-phosphate.
Vitamin [B.sub.12] status in the elderly as judged by available biochemical markers.
Correlation between plasma total homocysteine and copper in patients with peripheral vascular disease.
Plasma vitamin [B.sub.6] vitamers before and after oral vitamin [B.sub.6] treatment: a randomized placebo-controlled study.
Multianalyte quantification of vitamin [B.sub.6] and [B.sub.2] species in the nanomolar range in human plasma by liquid chromatography--tandem mass...
Quantification of holo-transcobalamin, a marker of vitamin [B.sub.12] deficiency.

Terms of use | Privacy policy | Copyright © 2018 Farlex, Inc. | Feedback | For webmasters