Can dipstick method help in bacterial detection in platelet bags? A tertiary care hospital's blood bank review.

Byline: Hammad Tufail Chaudhary and Shahida Hasnain

Abstract

Objective: To find the bacterial and biochemical details of bags used in platelet transfusion.

Methods: This cross-sectional study was carried out at a tertiary care hospital of Saudi Arabia (King Khalid Hospital, Najran) from January to June 2012, and comprised platelet bags. Samples for bacterial detection and biochemical testing of platelet bags were taken from blood bags on Day 6 of donation. Bacterial detection was done by using aerobic culture bottle, different gram stain, cultures and analytical profile index strips. Glucose, pH and protein measurements were done by Multistix dipstick method. SPSS 16 was used for data analysis.

Results: Of the 352 platelet bags, 1(0.28%) showed bacterial growth on Day 6 of collection. That bacterium was Staphylococcus epidermidis. Glucose content and pH of that platelet bag was 144.14mg/dl and 5, respectively. The overall mean pH of platelet bags was 6.690.55 (range: 3-7). Moreover, 255(72.4%) bags showed pH of 7, 90(25.5%) of 6, 5(1.4%) of 5 and 2(0.57%) showed pH of 3 on Day 6. The overall mean protein level was 6.1620.204g/dl (range: 5.8-6.6). Pearson bivariate correlation between platelet bag's pH and glucose content was 0.707 (p=0.001).

Conclusion: Positive correlation was found between platelet bag's glucose and pH levels.

Keywords: Platelets; Glucose; Bacteria. (JPMA 66: 1258; 2016)

Introduction

Blood transfusion is one of the important aspects in patient care. Viral screening and nucleic acid testing remained the most common methods for detection of Human Immunodeficiency Virus (HIV), Hepatitis B and C.1 But still, blood-transmitted infections like viruses, protozoans, helminths, prions and bacteria are some of the biggest problems of transfusion medicine.2 This concern becomes even more serious in the case of platelets as they are the particular component of the blood which is ideal for the growth of bacteria.3 Since 1990, several methods have been used to minimise the risk of bacterial transfusion reaction. Among these measures, donor screening and skin decontamination by iodine or chlorhexidine were important steps, as majority of bacteria isolated from blood bags had been skin flora.4

That is why pathogens are mostly gram-positive like Staphylococcus aureus, coagulase negative Staphylococci, Corynebacteria, Bacillus spp. viridans group Streptococci and anaerobic diphteroid gram-positive bacilli, for example Propionibacterium acnes.3

Some platelet storage solutions are being introduced which reduce the plasma and prevent bacteria to grow.5 Chances of bacteria avoidance in platelet bag is not reduced to zero, so detection of bacteria in blood bags has great importance. Culture-based methods have gained the great significance to detect bacteria.4 Gram stain, accrine orange and Wright's stain are used to detect bacteria in blood bags, but with high false positive results. Inexpensive but insensitive and non-specific measures like pH, glucose level, oxygen level and other metabolic changes are also used as an indication of growth of bacteria.3 Fluorescent cytometric techniques and nucleic acid polymerase chain reaction (PCR) methods are future consideration to detect bacteria in blood bags.3,6

Platelet bag quality assurance is also a vital issue in the field of blood transfusion. There are several factors which affect the platelet quality in platelet concentrates, including method of forming units, temperature at which it is stored, platelet count of platelet bags and sufficient oxygen to carry on metabolism. Platelet concentrates need constant agitation to optimise the transport of oxygen and carbon dioxide through the container. Many morphological, biochemical and functional changes occur during this storage period of platelets, which are termed as platelet storage lesions. These lesions are important because they can affect post-transfusion survival of cells.7

The significance and applicability of this study can be assessed by findings of Eder et al., 2007,8 who estimated the ratio of septic reaction to transfusion at 1:25,000 and the number of platelet units found to cause bacterial contamination after applying detection techniques at 1:2,000 to 1:3,000. Also, lots of blood component recipients are neutropenic and febrile patients, who are already taking antibiotics. So, whole clinical scenario can be changed after bacterial detection techniques implication.8 Bacterial detection has not been applied as a policy in many hospitals of Saudi Arabia. The current study was planned to determine microbiological and biochemical changes in platelet bags after storage for 5 days, so that the quality of platelet bags and bacterial growth could be assessed.

Materials and Methods

This cross-sectional study was carried out at a tertiary care hospital (King Khalid Hospital, Najran) of Saudi Arabia, from January to June 2012, and comprised platelet bags. Approval was obtained from the ethics committee of the hospital. Blood was drawn from the donors as mentioned in the guidelines of the formerly American Association of Blood Banks (AABB).9 Then 450ml of whole blood was collected in a 450-ml quadruple bag containing 63ml of citrate phosphate dextrose (CPD) anticoagulant, with additive solution (SAGM) (TERUMO PENPOL Limited). The whole bloods were separated into platelets, fresh frozen plasma and packed red blood cells by hard spin and then light spin method, as described in AABB guidelines.10 Platelet bags were kept in a Helmer agitator for 5 days.

Samples were selected by simple random sampling technique. Samples for bacterial detection and biochemical testing of platelet bags were taken from expired blood bags i.e. on Day 6 of donation (as the platelet bag were considered expired on Day 5). Under aseptic measures, 10ml sample was taken from platelet bag. Four ml sample was inoculated in aerobic culture bottle. Three ml sample was taken in sterile plastic tubes for pH, glucose and protein measurement by Multistixdipstick method. Three ml sample was taken in sterile red top tube for glucose and protein measurement by Roche kit using COBAS instrument. Bacteria detection was carried out by inoculation of platelets from platelet bag in aerobic bottle on Day 6 of blood component preparation as recommended by BACTEC.11 Bacteria were then further characterised by different gram stain, cultures and analytical profile index (API) strips.

Statistical analysis was done using SPSS 16. Descriptive statistics were obtained and bivariate Pearson correlation was found for different parameters.

Results

Of the 352 platelet bags, 1(0.28%) showed bacterial growth on Day 6 of collection. That bacterium was Staphylococcus epidermidis. Glucose content and pH of that platelet bag was 144.14 mg/dl and 5, respectively. The overall mean pH of platelet bags was 6.690.55 (range: 3-7) (Table-1). Moreover, 255(72.4%) bags showed pH of 7, 90(25.5%) showed pH of 6, 5(1.4%) showed pH of 5 and 2(0.57%) showed pH of 3 on Day 6. The mean glucose content of the bags was 256.93136.21 (8.46-484.86) (Table-1)

Table-1: Means and ranges of pH, glucose and protein.

###MeanStandard Deviation###Range

pH [N=352 (100%)]###6.69710.55###3-7

Glucose determined by Roche kit (mg/dl)

[N=352 (100%)]###256.93136.21###8.46-484.86

Protein (g/dl) [N=352 (100%)]###6.16192.04###5.8-6.6

The glucose content of 211(60%) bags was between 180.18-360.36 mg/dl. Moreover, 46(13%) platelet bags showed glucose value <54.05 mg/dl. Moreover, 137(39%) and 167(47.7%) of platelet concentrates had 4+ and 3+ of glucose (Table-2)

Table-2: Glucose and protein (by dipsticks) percentages.

Scores###1+N(%)###2+ N(%)###3+ N(%)###4+ N(%)

Glucose by dipstick

[N=352 (100%)]###12 (3.2%)###36 (10.1%)###167 (47.7%)###137 (39%)

Protein by dipstick

[N=352 (100%)]###0 (0%)###2 (0.5%)###276 (78.2%)###74 (21.3%)

The overall mean protein level was 6.1620.204g/dl (range: 5.8-6.6)

Pearson bivariate correlation between platelet bag's pH and glucose content was 0.707 (Table-3) (p=0.001).

Table-3: Pearson correlation between pH and glucose content of platelet bags (determined by Roche Kit).

###Pearson###p- value

###bivariate###(significant

###Correlation###if <0.05)

pH and Glucose content of platelet bags (mg/dl)###0.707###< .001

Table-4: Dipstick glucose relationship to other parameters.

Glucose by dipstick###1+###2+###3+###4+

Mean Glucose (determined by Roche Kit) Standard Deviation(mg/dl) [N=352 (100%)]###16.9311.89###141.08116.03###268.29108.64###276.57121.98

Mean pH Standard Deviation [N=352 (100%)]###6.170.41###6.350.67###6.720.58###6.790.41

Discussion

Platelets are an excellent growth medium for bacterial growth. Bacteria can grow in platelet concentrates and can cause transfusion reactions. In some studies it is mentioned that there are chances of 1:2,000 to 1:5,000 of bacterial contamination in platelet concentrates. Among them some can cause sepsis i.e.1:20,000 to 1:50,000, out of which 10% can become fatal.12 It is reported that 1 platelet bag is usually infected among 1000-2000 platelet bags.13 In our study, only one bag showed bacterial growth among 352 bags on Day 6 of collection, which is equal to 0.28%. In a Taiwan study, Hsueh JC et al. found that the infection rate of platelet bags was 0.31% (7 in 2,256).5 Lee CK et al. established detection rate of 0.15% (3 in 10,035 platelet units)in Hong Kong.14 Another study carried out by American Red Cross showed the results of 0.019% (226 of 350,658 platelet bags).

In that study, 47% of bacteria isolates were of staphylococcal species.15 Similarly, other studies showed prevalence of bacterial contamination from 0.14%-1.41%, with the mean value being 0.43%. These studies also mentioned that most common bacteria isolated from platelet concentrates was staphylococcus, as in our study.16

Wagner SJ concluded that 56% of times bacteria isolated from transfusion-related sepsis was aerobic gram-positive.17 But gram-negative organisms are also found in blood bag contamination and they are more morbid than gram-positive. Yomtovian RA et al. showed division of the cases in sepsis as Staphylococcus spp. in 42% of cases, Escherichia coli in 9%, Bacillus spp. in 9%, Salmonella spp. in 9%, Streptococcus spp. in 12%, Serratia spp. in 8%, Enterobacter spp. in 7%, and other organisms in 4%.18 Donor bacteraemia is one of the less frequent causes of bacterial transmission through blood components. Ness P et al. concluded that 62.5% and 34.8%of cases of septic reaction by transfusion were shown to be related to skin flora bacteria and transient donor bacteraemia, respectively.19

Moreover, pH has been used as a quality control of platelet concentrates. Its value below 6.8 has been declared as a bad indicator of in-vivo platelets survival. Platelets release different chemicals including lactate which lowers the pH.20 Similarly, growth of bacteria can affect pH of platelet bags due to production of acids. It has been mentioned in study that dipstick accuracy for pH detection is good at pH around 7.21 Therefore, we used dipsticks to judge the pH of platelet bags.

In our study, the majority (72.6%) of bags showed pH of 7 on Day 6. Besides, 25.5% of bags showed pH of 6 and only 1.4 % and 0.5% showed pH of 5 and 3, respectively. In one study, pH of 7.125 and pH of 7.140 was detected on Days 5 and 7 of platelet collection, respectively.7 Doescher A et al. detected pH of 7 on Day 6.20 In another study, pH of 7.73 was the result on Day 5 of collection.22 This shows our study results were mostly in concordance with other studies.

Furthermore, 27.4% of platelet concentrates showed pH of 6 and less. This is an indicator of non-viable platelets in platelet bags. Platelets become spherical at pH of 6.8 and metabolism of platelets is captured at pH of 6 or less.22 The pH of the platelet bag which was found infected with bacterium was 5. This might be due to acid released by bacterium.

To maintain the good quality of platelets, glucose should be maintained during storage period. In our study, the mean glucose content of 352 platelet bags was 256.93mg/dl, with a range of 8.46-484.86. Besides, 60% of platelet concentrates showed glucose content between 180.18-360.36 mg/dl. Chandra T et al. found that glucose concentration decreased from 225.58mg/dl to 197.47mg/dl from the day of collection to Day 7.23 We came to know that 39% and 47.7% of platelet concentrates had 4+ and 3+ of glucose with dipstick method. A significant positive correlation was found between pH and glucose content of platelet bags. The same correlation was also found for glucose measured by dipstick method and pH of platelet bags. As the pH of platelet bags decreased, glucose content also decreased. This explains the fact that as platelet bags get older, more glucose is converted into acids, which results in reduction of pH.11,20

In our study, protein content (range: 5.8-6.6g/dl) in all of the platelet bags was near the lower normal range of plasma protein (6.4-8.2 g/dl) (Table-1). This indicates that protein content of the healthy donor was diluted in plasma and citrate of blood bags. Easy and cheap determination of platelet bag quality is an area of interest in field of transfusion medicine.20 In our study, we found a positive correlation between glucose determined by dipstick and glucose determined by Roche Kit and, pH (Figure-1 and 2, Table-4). Glucose determination with dipstick is an easy technique. It gives an idea about the quality of platelet bag contents according to our analysis in this study. It is a known fact that bacterial growth in platelet bags causes decrease in pH and glucose content.20 Therefore, we can suggest that dipstick technique can help in detection of bacterial growth.

Sample can be taken from platelet bag tubing and can be tested for glucose content by dipstick during storage period. Further studies are needed to establish the relationship between glucose content in tubing and platelet bags during the storage period of platelet bags.

Conclusion

The dipstick technique was found to be helpful in detection of pH and glucose content of platelet bags. Glucose and pH content of platelet bags were found to be good indicators of bacterial growth.

Disclaimer: None.

Conflict of Interest: None.

Source of Funding: None.

References

1. Andrade RG, Ribeiro MA, Namen-Lopes MS, Silva SM, Basques FV, Ribas JG, et al. Evaluation of the use of real-time PCR for human T cell lymphotropic virus 1 and 2 as a confirmatory test in screening for blood donors. Rev Soc Bras Med Trop. 2010; 43:111-5.

2. Acar A, Kemahli S, Altunay H, Kosan E, Oncul O, Gorenek L,et al.HBV, HCV and HIV seroprevalence among blood donors in Istanbul, Turkey: how effective are the changes in the national blood transfusion policies? Braz J Infect Dis.2010;14: 41-6.

3. Montag T. Strategies of bacteria screening in cellular blood components. Clin Chem Lab Med. 2008;46:926-32.

4. Martini R, Kempfer CB, Rodrigues MA, Kuhn FT, Rigatti F, Ratzlaff V,et al. Bacterial contamination on platelet concentrates: identification, antimicrobial susceptibility profile and transfusion-related sepsis. Rev Soc Bras Med Trop. 2010;43:682-5.

5. Hsueh JC, Ho CF, Chang SH,Pan FZ, Chen SC, Shi MD,et al. Blood surveillance and detection on platelet bacterial contamination associated with septic events. Transfus Med. 2009;19:350-6.

6. Karo O, Wahl A, Nicol SB, Brachert J, Lambrecht B, Spengler HP,et al. Bacteria detection by flow cytometry. Clin Chem Lab Med. 2008;46:947-53.

7. Sandgren P, Saeed K. Storage of buffy-coat-derived platelets in additive solution: in vitro effects on platelets of the air bubbles and foam included in the final unit. Blood Transfus. 2011;9:182-8.

8. Eder AF, Kennedy JM, Dy BA, Weiss JW, Fang CT, Wagner S,et al. Bacterial screening of apheresis platelets and the residual risk of septic transfusion reactions: the American Red Cross experience (2004-2006). Transfusion. 2007; 47:1134-42.

9. Brecher M. Technical Manual. 15th ed. Bethesda, Md: American Association of Blood Banks; 2005; p. 178.

10. Brecher M. Technical Manual. 15th ed. Bethesda, Md: American Association of Blood Banks; 2005, pp 179.

11. Riedel S, Siwek G, Beekmann SE, Richter SS, Raife T, Doern GV.Comparison of the BACTEC 9240 and BacT/Alert Blood Culture Systems for Detection of Bacterial Contamination in Platelet Concentrates. J Clin Microbiol. 2006;44:2262-4.

12. Schlenke P. Pathogen Inactivation Technologies for Cellular Blood Components: an Update. Transfus Med Hemother. 2014;41:309-25.

13. Brecher ME. Bacterial contamination of blood products, In: Simon TL, eds. Rossi's principles of transfusion medicine. 3rd ed. Baltimore: Lippincott; 2002, pp 789.

14. Lee CK, Ho PL, Lee KY, Tsui GT, Chua E, Tsoi WC,et al. Value of anaerobic culture in bacterial surveillance program for platelet concentrates. Transfusion. 2008;48:2606-11.

15. Fang CT, Chambers LA, Kennedy J,Strupp A, Fucci MC, Janas JA, et al.Detection of bacterial contamination in apheresis platelet products: American Red Cross experience, 2004. Transfusion. 2005;45:1845-52.

16. Carl PM, Blajchman MA. Bacterial contamination in blood and blood components. In: John Barbara JAJ, Regan FAM, Contreras DM, eds.Transfusion Microbiology. New York: Cambridge University Press, 2008; pp89.

17. Wagner SJ. Transfusion-transmitted bacterial infection: risks, sources and interventions. Vox Sang. 2004;86:157-63.

18. Yomtovian RA, Palavecino EL, Dysktra AH, Downes KA, Morrissey AM, Bajaksouzian S, et al. Evolution of surveillance methods for detection of bacterial contamination of platelets in a university hospital, 1991 through 2004. Transfusion. 2006; 46:719-30.

19. Ness P, Braine H, King K, Barrasso C, Kickler T, Fuller A, et al. Single-donor platelets reduce the risk of septic platelet transfusion reactions. Transfusion. 2001;41:857-61.

20. Doescher A, Muller TH. Noninvasive pH Monitoring in Platelet Concentrates. Transfus Med Hemother. 2013;40:88-92.

21. Kwong T, Robinson C, Spencer D, Wiseman OJ, KaretFrankl FE.Accuracy of urine pH testing in a regional metabolic renal clinic:is the dipstick accurate enough? Urolithiasis.2013;41:129-32.

22. Coelho MJ, Monteiro Tde C, Vasquez FG, Silva KL, Dos Santos KS, de Oliveira VM,et al. Platelet aggregation and quality control of platelet concentrates produced in the Amazon Blood Bank. Rev Bras Hematol Hemoter. 2011;33:110-4.

23. Chandra T, Gupta A, Kumar A.Extended Shelf Life of Random Donor Platelets Stored for 7 Days in Platelet Additive Solution at Different Temperatures. Biomed J. 2014;37:211-7.
COPYRIGHT 2016 Asianet-Pakistan
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2016 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Publication:Journal of Pakistan Medical Association
Geographic Code:7SAUD
Date:Oct 31, 2016
Words:3287
Previous Article:The effect of ethanol vapour exposure on atrial and ventricular walls of chick embryos.
Next Article:The impact of hypertension on lipid parameters in type 2 diabetes.
Topics:

Terms of use | Privacy policy | Copyright © 2019 Farlex, Inc. | Feedback | For webmasters