Analytical quality of near-patient blood cholesterol and glucose determinations.

Healthcare worldwide is moving in a direction of prevention through screening and educational programs as part of primary care, rather than treatment of an established condition. Although prevention is not always possible in all cases, complications and the cost of healthcare can nevertheless be reduced and quality of life can be improved by early diagnosis and treatment.

An awareness of the importance and possible advantages of population screening for diabetes and hypercholesterolemia has increased in recent years. According to the Expert Committee on the Diagnosis and Classification of Diabetes Mellitus, all individuals 45 years and above should be screened for diabetes mellitus, and testing should be repeated every 3 years if initially normal. In certain high-risk patients, testing should be carried out at a younger age and with shorter intervals (1). Early detection, and consequently early treatment, might well reduce the burden of especially type 2 diabetes and its complications because epidemiological evidence suggests that retinopathy begins to develop at least 7 years before the clinical diagnosis of diabetes (2). Although extra-laboratory blood glucose measurement with a test strip system is not recommended for the primary diagnosis of diabetes mellitus, it may be used for screening purposes, followed by laboratory confirmation (3).

In the second report of the National Cholesterol Education Program (Adult Treatment Pane12), it is stated that serum total cholesterol should be measured in all adults 20 years of age and older at least once every 5 years. Both blood cholesterol and other coronary heart disease risk factors are considered in classification of individuals and in deciding on follow-up intervals and appropriate intervention (4).

Recent advances, in particular the development of the so-called "reagent test strip", have produced a proliferation of inexpensive and convenient systems for the measurement of blood glucose and cholesterol outside the laboratory (5). The analytical performance of the reagent strips and meters can be evaluated easily in the laboratory, but this gives no information about the performance of the individual user who lacks professional training. The establishment of a quality-control program adapted to this new approach to biochemical testing remains an important challenge because conventional quality-control procedures with lyophilized materials or liquid ethylene glycol-based materials often are inappropriate (6).

In cost-effective population screening, extra-laboratory testing by, for example, primary care clinics, pharmacies, or general practitioners, could play a major part, provided that their results are reliable, accurate, and reproducible. Although the accuracy and precision of high-quality near-patient testing devices in the hands of trained laboratory technicians are excellent, little information exists on the accuracy and reproducibility of results obtained under day-to-day conditions at various point-of-care centers, e.g., pharmacies.

The objective of this study was to assess the analytical quality (accuracy and precision) and interchangeability of capillary blood total cholesterol and glucose assays as performed in 12 different pharmacies in Pretoria, South Africa.

Materials and Methods

STUDY DESIGN

Accuracy and within-subject total variance for capillary whole blood and venous serum/plasma total cholesterol and glucose measurements were determined under the following circumstances: (a) in our laboratory using a single point-of-care analyzer; (b) using 12 randomly selected pharmacies that offer near-patient testing facilities; (c) in our laboratory using a single clinical chemistry analyzer; and (d) using all 8 diagnostic clinical chemistry laboratories in the Pretoria area.

Precision was determined for a single point-of-care instrument and single laboratory analyzer.

POINT-OF-CARE CENTER EVALUATION

During the first part of the study, six apparently healthy volunteers visited independently and unannounced each of 12 randomly selected pharmacies in Pretoria to have their fasting capillary blood glucose and cholesterol concentrations determined. Visits took place randomly during a 3-week period according to a predetermined schedule.

After each pharmacy visit, the participants had the same determinations done by one trained person at the Institute of Pathology. Either a right- or left-handed finger-prick blood sample was obtained according to appropriate guidelines for reliable finger-stick collection (7); the blood sample was analyzed immediately for glucose and cholesterol, using one near-patient device (Accutrend GCT; Roche Diagnostics, Mannheim, Germany). To minimize and standardize biological variation, all blood samplings were performed between 0900 and 1100 after an overnight fast of at least 10 h.

Two serum pools with concentrations at important decision cutoff values for glucose (2.5 and 7.0 mmol/L) and total cholesterol (5.2 and 6.2 mmol/L) were prepared from fresh human serum received at the Institute of Pathology for routine analyses. After being appropriately aliquoted, serum pools were frozen at -70 [degrees]C until being assayed. The respective in-house serum pools were assayed daily for 12 consecutive days on the single point-of-care instrument to assess between-day analytical variation.

After this 3-week period, each pharmacy was informed by telephone about the objective of the project. Permission was requested to assess the accuracy of each pharmacy's testing device. This was performed by using three human-serum-based Standard Reference Materials (SRMs) [1] with certified concentrations for glucose (SRM 965) and cholesterol (SRM 1952a) obtained from NIST (Gaithersburg, MD). The certified concentrations of glucose were determined using the NIST definitive method for glucose (isotope dilution/gas chromatography/mass spectrometry), whereas total cholesterol was determined by a modification of the isotope dilution/mass spectrometric definitive method.

The various point-of-care instruments usually measure glucose and cholesterol concentrations in capillary whole blood, whereas the reference material utilized in this study was human serum. For instruments using test strip systems incorporating a porous outer layer that allows only plasma to penetrate into the reagent layer, the same reference values are valid for whole blood, serum, and plasma. However, in instruments without this plasma separation procedure and which were calibrated to yield plasma results, when whole blood was used as the specimen for analysis, the value obtained from the instrument for the serum-based reference material was divided by 1.12 (8) before calculation of the percentage of bias.

The following point-of-care instruments were used by participating pharmacies. For cholesterol assays, nine pharmacies used an Accutrend analyzer (Roche Diagnostics), two pharmacies used the Lipotrend C (Roche Diagnostics), and one pharmacy used the Reflotron (Roche Diagnostics). For glucose assays, eight pharmacies used an Accutrend analyzer (Roche Diagnostics). The remaining pharmacies used either the Precision (Novo Nordisk), Glucometer Elite (Bayer), One Touch basic (Lifescan), or Reflotron analyzer (Roche Diagnostics).

LABORATORY EVALUATION

Evaluation of the eight clinical chemistry laboratories took place 2 months after the evaluation of the point-of-care centers. The six volunteers visited each of the eight laboratories independently and unannounced over a 2-week period for venous blood sampling to determine their fasting serum total cholesterol and plasma glucose concentrations. After each visit, the participant went to the Institute of Pathology to have the same analyses done on a single laboratory instrument (Synchron CX7; Beckman Instruments). The above-mentioned in-house serum controls were also assayed for 12 consecutive days on the single laboratory analyzer to assess between-day analytical variance.

After completion of the laboratory visits, the accuracy of each laboratory's clinical chemistry analyzer was assessed as described previously, using the NIST SRM. The following instruments were used by the participating laboratories: six used Beckman analyzers (Synchron CX7 or LX20) for both glucose and total cholesterol analyses, whereas the other two laboratories used the Hitachi 917 (Roche Diagnostics) and the Olympus AU 600 (Olympus Diagnostica), respectively.

ETHICS

All participants gave written consent for participation in the study. Ethical approval was obtained from the Ethics Committee of the Faculty of Medicine, University of Pretoria.

CALCULATIONS AND STATISTICS

Results from repeat analyses of the serum pools on the in-house instruments were used to calculate the between-day analytical variance ([[sigma].sub.A.sup.2]) for both the point-of-care ([[sigma].sub.AC.sup.2]) and laboratory analyzer ([[sigma].sub.AV.sup.2]).

The mean within-subject total variance ([[sigma].sub.S.sup.2]) was determined using data obtained from the in-house instruments for capillary blood ([[sigma].sub.SC.sup.2]) and venous serum/plasma ([[sigma].sub.SV.sup.2]) cholesterol and glucose assays, respectively, as described by Fraser and Harris (9).

The average within-subject biological variance ([[sigma.sub.I.sup.2]) was then calculated by adjusting within-subject total variance for analytical variance. Because the determinations for this study were carried out in singletons, the general formula suggested by Holzel (10) was used for calculation of within-subject biological variance.

The within-subject biological variance for capillary blood ([[sigma].sub.IC.sup.2]) was determined as follows:

[[sigma].sub.IC.sup.2] = [[sigma].sub.SC.sup.2] - [[sigma].sub.AC.sup.2]

The within-subject biological variance for venous serum or plasma ([[sigma].sub.IV.sup.2]) was calculated similarly:

[[sigma].sub.IV.sup.2] = [[sigma].sub.SV.sup.2] - [[sigma].sub.AV.sup.2]

Using the data from all of the participating pharmacies for each individual, we again calculated the within-subject total variance. This variable was designated [[sigma].sub.SP.sup.2]. By correcting [[sigma].sub.SP.sup.2] for the capillary blood within-subject biological variance ([[sigma].sub.IC.sup.2]), we could calculate the "between-pharmacy analytical variance" ([[sigma].sub.AP.sup.2]) according to the following formula:

[[sigma].sub.AP.sup.2] = [[sigma].sub.SP.sup.2] - [[sigma].sub.IC.sup.2]

The between-pharmacy analytical variance included true within-instrument analytical variance, between-instrument variance, sampling variance, and operator-specific variance (i.e., variance introduced by having multiple operators of a given analyzer, or variance attributable to different analyzers being used by different single operators).

The appropriate CVs were subsequently calculated for each of the above variables.

For a capillary blood glucose or cholesterol determination done at any of the participating pharmacies, the percentage of closeness to the homeostatic set point (95% dispersion of the mean value) was calculated as follows (9,11):

[D.sup.2] = [Z.sup.2] (C[V.sub.AP.sup.2] + C[V.sub.IC.sup.2])/n

Where D is the percentage of closeness to the homeostatic set point; Z is the number of standard deviations required for a stated probability (for p = 0.05, Z = 1.96); n is the number of specimens (in this case, n = 1); C[V.sub.AP] is the between-pharmacy analytical CV; and C[V.sub.IC] is the within-subject biological CV, determined on capillary whole blood.

The magnitude of a significant (P = 0.05) difference between two subsequent cholesterol determinations (critical difference) was calculated (12 ) for the following scenarios:

(a) using the same point-of-care center:

CD = K x [(C[V.sub.AC.sup.2] + C[V.sub.IC.sup.2]).sup.1/2]

Where CD is the critical difference; K = 1.96 x [(2).sup.1/2] = 2.77 for P = 0.05; C[V.sub.AC] is analytical CV, as determined on a single instrument; and C[V.sub.IC] is within-subject biological CV for capillary whole blood.

(b) using different point-of-care centers:

CD = K x [(C[V.sub.AP.sup.2] + C[V.sub.IC.sup.2]).sup.1/2]

Where C[V.sub.AP] is between-pharmacy analytical CV.

Similar calculations were performed using data from the participating laboratories for each individual. In this case, [[sigma].sub.SL.sup.2] designates within-subject total variance calculated from the data of all participating laboratories for each individual, and [[sigma].sub.AL.sup.2] indicates between-laboratory analytical variance, i.e., a combination of true within-instrument analytical variance, between-instrument variance, sampling variance, and operator-specific variance.

Accuracy was determined by means of duplicate analyses of the SRMs for glucose and cholesterol, respectively. The results were used to calculate the percentage of bias according to the following formula:

Bias (%) = Value obtained - Certified value/Certified value x 100

Results

Using the CVs as summarized in Table 1, we calculated that the percentage of closeness to the homeostatic set point was 24.6% for a single glucose determination done on capillary whole blood in any one of the participating pharmacies, whereas it was 16.9% if performed on plasma in any one of the participating laboratories. For a total cholesterol assay, the percentage of closeness to the homeostatic set point was 23.6% when performed on capillary whole blood in any one of the participating pharmacies and 15.6% when performed on serum in any one of the participating laboratories.

The critical differences for cholesterol determinations were 33.3% and 22.1% when two subsequent determinations were done in any of the pharmacies or any of the laboratories, respectively. For two subsequent determinations performed at the same pharmacy or the same laboratory, the critical differences were 16.2% and 16.9%, respectively. The latter two values are based on the in-house evaluation of a point-of-care instrument and a clinical chemistry analyzer and will vary from pharmacy to pharmacy or laboratory to laboratory, depending on the analytical CV for the instruments in use.

Figs. 1 and 2 summarize the observed bias for glucose measurements in pharmacy point-of-care centers and in laboratories. Bias in pharmacies was -48.1% to 16.2%, whereas bias for laboratory measurements was -1.0% to 7.4%.

Because of the limited analytical range of the Accutrend analyzer (3.9-7.8 mmol/L), only the medium concentration of the SRM for cholesterol (certified value, 6.04 [+ or -] 0.04 mmol/L) was used to assess bias. Bias was -5.6% to 16.6% in pharmacies and -10.6% to 3.7% in laboratories for the cholesterol assay (Figs. 1 and 2).

[FIGURE 1 OMITTED]

Discussion

This study is an attempt to quantify the variation between point-of-care centers for two important blood components, total cholesterol and glucose. The data on variation allowed us to assess the utility of a screening test at a point-of-care center (pharmacy) in terms of the expected closeness of the result to the individual's homeostatic set point. Similar data were also obtained for all eight of the clinical chemistry laboratories in Pretoria. Data obtained from the laboratories served as a reference with which the pharmacies could be compared.

For cholesterol analyses, the instrument most popular in pharmacies (Accutrend) yielded a C[V.sub.A] of 2.5%, whereas a C[V.sub.A] of 3.3% was achieved with the in-house clinical chemistry analyzer (Synchron CX7).

The within-subject biological CVs for capillary and venous specimens for total cholesterol (5.3% and 5.1%, respectively) confirmed the previous observations by Kafonek et al. (13 ) that there is no significant difference in the coefficient of biological variation between capillary and venous specimens with respect to total cholesterol concentration. The results also compare well with the average value from previous studies by Fraser (C[V.sub.I] = 5.8%) (14).

[FIGURE 2 OMITTED]

The C[V.sub.AP] for total cholesterol was 11% compared with a C[V.sub.AL] of 6.1%, indicating a higher between-pharmacy variation than experienced between laboratories. This implies an increased uncertainty of the significance of a pharmacy cholesterol measurement, as reflected by a percentage of closeness to the homeostatic set point of 23.6% if a single capillary blood specimen was analyzed for cholesterol. The corresponding value obtained in laboratories was 15.6% (relative reduction of 50% from pharmacy value).

Given that the within-instrument analytical performance for point-of-care and laboratory analyzers is similar, it seems that the higher between-pharmacy variation may be attributed to differences in sampling (sampling variance) or nonuniformly calibrated instruments.

Further evidence supporting that calibration bias may have contributed to a higher between-pharmacy variation compared with between-laboratory variation is derived from analyses of certified serum material obtained from the NIST. For the cholesterol assay, the bias range spanned 22.2% for pharmacy analyzers compared with 14.3% for laboratory analyzers. Between-analyzer matrix effects cannot be excluded for the pharmacy analyses (designed to assay whole blood samples), and these effects could have contributed to the wider bias range observed for the point-of-care instruments.

Although finger-stick sampling is considered easier to perform, requires less training, and is less invasive than conventional venous collection, there is nevertheless considerable potential for technique-related variability, which often is responsible for the discrepant results between finger-stick and venous results. Finger-stick specimens are subject to dilution by interstitial fluid, giving a lower cholesterol concentration. Hemolysis is more likely to occur with finger-stick collection. Sensitivity to hemolysis varies by method but is usually the cause of falsely increased values (7). In recent years, improved lancets and collection devices have become available, and with appropriate techniques, finger-stick results can be equivalent to venous results (7).

The National Cholesterol Education Program recommends a performance of 3% CV with a 3% bias for determination of total cholesterol concentration (15,16). This becomes a total allowable error of 8.9% [total error = absolute bias + (1.96 X C[V.sub.A])] vs the true value in situations where cholesterol is determined as a single measurement affected by both systematic and random error (17). This implies that only two pharmacies did not meet the 8.9% total error goal for determination of total cholesterol.

Despite the larger variation in pharmacy determinations, they should nevertheless be adequate to screen for hypercholesterolemia, provided that an abnormal value is confirmed by repeat testing. As stated previously by Fraser (14), our results also stress that a patient cannot be classified into a certain risk category based on a single cholesterol determination, not even when determined in a clinical chemistry laboratory (15.6% closeness to the homeostatic set point).

Changes in sequential results from a patient may be caused by analytical imprecision, within-subject biological variation, or by a change in the patient's clinical condition. The "critical difference" between results gives an indication of the magnitude of the change that should occur before it is considered (with 95% confidence) to be attributable to an altered clinical condition or to treatment of the patient.

Our results indicate that cholesterol testing at different pharmacies currently may have a limited utility in monitoring the efficacy of treatment for hyperlipidemia. The difference between two subsequent cholesterol measurements at two different pharmacies should be at least 33.3% before it may be considered (with 95% confidence) to be the result of intervention. For two different laboratories, the critical difference is substantially less at 22.1%.

When the same pharmacy or laboratory is used to follow the result of intervention, the critical difference declines to ~17% in both cases. This again serves to illustrate how between-pharmacy variation contributes to the uncertainty of results.

To assess pharmacy glucose screening, we compared our results to the American Diabetes Association (ADA) guidelines applicable to self-monitoring of blood glucose (SMBG). According to the ADA guidelines of 1986, the goal of all future SMBG systems should be to achieve a variability (system + user) of <10% at glucose concentrations of 1.7-22.2 mmol/L. In 1986, the ADA stated that with current systems, SMBG measurements should be within 15% of the results of the reference measurement (18). Later, at a second conference in 1993, they recommended tightening the limits on "analytic error" to within 5% (19). Applying the 1986 ADA guidelines, only three pharmacies had results for all three SRMs within 10% of the certified concentrations. In contrast, all of the laboratories reported values for the SRM materials within 10% of the certified values.

For glucose determinations, the larger biological variance (C[V.sub.IC]) as well as analytical imprecision for capillary specimens (C[V.sub.AC]) contributed to the larger total variance. However, a significant portion of the larger between-pharmacy variance could probably also be attributed to differences in sampling (sampling variance) and to the variety of instruments from different manufacturers being used.

The previously published within-subject biological CV (C[V.sub.I]) for fasting plasma glucose of 4.75% (14, 20) was confirmed by our value of 4.0% for plasma glucose on venous specimens. However, the value of 6.9% for fasting capillary whole-blood glucose was considerably larger. No published information could be found for this specific type of specimen. The reason for the larger variance observed for capillary whole-blood glucose vs venous plasma glucose is unknown but could in part be attributable to shifts over the red cell membrane and varying degrees of lymphatic dilution.

It has repeatedly been shown that a scheme involving training, internal quality control, and external quality assessment substantially improves the quality of near-patient testing (21-23 ). If this is combined with a universal system for calibration of near-patient devices and uniformity in reporting of results, pharmacy determinations should yield results comparable to the clinical chemistry laboratory. Unfortunately, in contrast to diagnostic clinical chemistry laboratories, near-patient testing in South Africa (as in many other countries) is not subject to compulsory external quality assessment programs or accreditation with a relevant authority. Appropriate quality-control schemes and accreditation requirements for point-of-care laboratory services will enhance quality and accuracy of blood test results.

We thank the University of Pretoria for financial support and Roche Diagnostics for the provision of an Accutrend GCT analyzer and reagent test strips used for the in-house evaluation of near-patient testing on a single instrument.

References

(1.) Expert Committee on the Diagnosis and Classification of Diabetes Mellitus. Report of the Expert Committee on the Diagnosis and Classification of Diabetes Mellitus. Diabetes Care 1997; 20:1183-201.

(2.) Harris MI. Undiagnosed NIDDM: clinical and public health issues. Diabetes Care 1993;16:642-52.

(3.) Price CP, Burrin JM, Nattrass M. Extra-laboratory blood glucose measurement: a policy statement. Diabet Med 1998;5:705-9.

(4.) Summary of the second report of the National Cholesterol Education Program (NCEP) Expert Panel on detection, evaluation, and treatment of high blood cholesterol in adults (Adult Treatment Panel II). JAMA 1993;269:3015-23.

(5.) Steinhausen RL, Price CP. Principles and practice of dry chemistry systems. In: Price CP, Alberti KGMM, eds. Recent advances in clinical biochemistry, Vol. 3. Edinburgh: Churchill Livingstone, 1985:273-96.

(6.) Forest J, Rousseau F, Carrier R, Hivon P, Gosselin M. Quality-control scheme for blood glucose measured outside the laboratory. Clin Chem 1987;33:1233-5.

(7.) Warnick GR, Leary ET, Ammirati EB, Allen MP. Cholesterol in fingerstick capillary specimens can be equivalent to conventional venous measurements. Arch Pathol Lab Med 1994;118:1110-4.

(8.) Sacks DB. Carbohydrates. In: Burtis CA, Ashwood ER, eds. Tietz textbook of clinical chemistry, 3rd ed. Philadelphia: WB Saunders, 1999:777.

(9.) Fraser CG, Harris EK. Generation and application of data on biological variation in clinical chemistry. Crit Rev Clin Lab Sci 1989; 27:409-37.

(10.) Holzel WGE. Intra-individual variation of some analytes in serum of patients with insulin-dependent diabetes mellitus. Clin Chem 1987;33:57-61.

(11.) Fraser CG, Petersen PH. The importance of imprecision. Ann Clin Biochem 1991;28:207-11.

(12.) Fraser CG. Data on biological variation: essential prerequisites for introducing new procedures? [Editorial]. Clin Chem 1994;40: 1671-3.

(13.) Kafonek SD, Donovan L, Lovejoy KL, Bachorik PS. Biological variation of lipids and lipoproteins in fingerstick blood. Clin Chem 1996;42:2002-7.

(14.) Fraser CG. Biological variation in clinical chemistry. An update: collated data, 1988-1991. Arch Pathol Lab Med 1992;116:916-23.

(15.) Laboratory Standardization Panel. Recommendations for improving cholesterol measurement. A report from the Laboratory Standardization Panel of the National Cholesterol Education Program. NIH Publication No. 90-2964. Bethesda, MD: NIH, 1990.

(16.) National Cholesterol Education Program. Recommendations on lipoprotein measurement from the Working Group on Lipoprotein Measurement. NIH Publication No. 95-3044. Bethesda, MD: National Heart, Lung and Blood Institute, 1995.

(17.) Rogers EJ, Misner L, Ockene IS, Nicolosi RJ. Evaluation of seven Cholesterh L.D.X analyzers for total cholesterol determinations. Clin Chem 1993;39:860-4.

(18.) American Diabetes Association. Consensus statement on self-monitoring of blood glucose. Diabetes Care 1987;10:95-9.

(19.) American Diabetes Association. Consensus statement on self-monitoring of blood glucose. Diabetes Care 1994;17:81-6.

(20.) Cummings ST, Fraser CG. Variability of capillary plasma glucose in healthy individuals in repeated 75 g oral glucose tolerance tests. Ann Clin Biochem 1988;25:634-7.

(21.) RumIeyAG.Improving the quality of near-patient glucose measurement. Ann Clin Biochem 1997;34:281-6.

(22.) Hamilton AD, Allen KR, Waters AK. The introduction of a quality assessment scheme for extra-laboratory blood glucose analysis. Diabet Med 1989;6:444-7.

(23.) Stahl M, Brandslund I, Iversen S, Filtenborg JA. Quality assessment of blood glucose testing in general practitioners' offices improves quality. Clin Chem 1997;43:1926-31.

Marita du Plessis, * Johan B. Ubbink, and W.J. Hayward Vermaak

Department of Chemical Pathology, University of Pretoria, Institute of Pathology, PO Box 2034, Pretoria 0001, South Alrica.

* Author for correspondence. Fax 27-12-328-3600; e-mail mdupless@ medic.up.ac.za.

[1] Nonstandard abbreviations: SRM, Standard Reference Material; ADA, American Diabetes Association; and SMBG, self-monitoring of blood glucose.

Received February 29, 2000; accepted April 27, 2000.
Table 1. Components of variation for point-of-care and
laboratory glucose and cholesterol analyses.

 Capillary blood

 Component Glucose Cholesterol

[CV.sub.AC], % 5.7 2.5
[CV.sub.SC], (a) % 8.6 6.0
[CV.sub.IC], % 6.9 5.3
[CV.sub.SP], % 12 12
[CV.sub.AP], % 10 11

 Venous plasma/serum

 Component Glucose Cholesterol

[CV.sub.AV], % 3.0 3.3
[CV.sub.SV], % 5.1 6.4
[CV.sub.IV], % 4.0 5.1
[CV.sub.SL], % 8.5 7.8
[CV.sub.AL], % 7.6 6.1

(a) [CV.sub.SC], within-subject total CV for capillary blood;
[CV.sub.SP], within-subject total CV for capillary blood analyses
between pharmacies; [CV.sub.AV], analytical CV for venous
serum/plasma as obtained from a single laboratory instrument;
[CV.sub.SV], within subject total CV for venous serum/plasma;
[CV.sub.IV], within-subject biological CV for venous serum/plasma;
[CV.sub.SL], within-subject total CV for venous serum/plasma analyses
between laboratories; [CV.sub.AL], between-laboratory analytical
CV for venous serum/plasma.
COPYRIGHT 2000 American Association for Clinical Chemistry, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2000 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Evidence-based Laboratory Medicine and Test Utilization
Author:Plessis, Marita du; Ubbink, Johan B.; Vermaak, W.J. Hayward
Publication:Clinical Chemistry
Date:Aug 1, 2000
Words:4233
Previous Article:Genetic analysis of dna excreted in urine: a new approach for detecting specific genomic DNA sequences from cells dying in an organism.
Next Article:Opportunistic discovery of occult disease by use of test panels in new, symptomatic primary care outpatients: yield and cost of case finding.