A Retrospective Study of Cochlear Re-Implantations-Experience from a Large Centre in India.

Introduction

Cochlear implantation (CI) in contemporary world is accepted as the standard of care for patients with severe cochlear hearing loss. With increasing number of CI being carried out at numerous centres across the world, more and more of these implantees (more so children) will require cochlear re-implantation (CRI) in their lifetime. Device failures (hard failure or soft failure), medical conditions (infection, acute or chronic otitis media, and implantation cholesteatoma) and electrode array extrusion or misplacement are frequent causes of CRI (1). Studies have shown that CRI is a safe surgery with maintenance of audiological performance (2, 3). However, studies regarding CRI are lacking from developing countries such as India.

Maurer et al. (4) in their study concluded that cochlear implant reliability data should be considered during the choice of implant for each individual. Since such data are lacking in literature across the world, including India, it becomes imperative that centres with a high number of cochlear implant surgery need to report the cumulative survival rate (CSR) of cochlear implants at their respective centres.

The aim of the study was to present the data of CRI and the reliability data of cochlear implants based on our experience. The objectives were to study the incidence and indications of cochlear re-implantation and determine CSR of cochlear implantation.

Methods

Our study was a retrospective one, conducted at a large cochlear implant centre in India. A total of 1500 consecutive cochlear implanted ears (bilateral implants were counted as two implanted ears) from 1997 to 2016 were studied. All patients who underwent CRI in this period were included in the study. All cases of revision cochlear implantation where cochlear explantation was not carried out were excluded from the study. The relevant data was retrieved from our medical records department. Informed consent was taken. Institutional research ethics board approval was obtained. Appropriate statistical analysis was performed, using Statistical Package for the Social Sciences (SPSS) version 20.0 (IBM Corp.; Armonk, NY, USA).

Results

In total there were 53 ears (3.53% of the 1500 consecutive implanted ears) who had undergone CRI at our centre. Out of the 53 ears, 31 CRI were male patients and remaining 22 were female patients. The mean age of primary implantation, failure and CRI was 4.7 years, 6.2 years and 6.3 years, respectively. The mean duration of implant usage before undergoing re-implantation was 1.6 years. All patients requiring CRI were less than 8 years of age, except one who was 28 years of age requiring CRI after 10 years because of hard failure of implanted device.

The most common indication of CRI was device failure in 39 ears (34 ears with hard failure and 5 ears with soft failure) contributing to 73.6% of the total re-implantation. The next common indication of CRI was medical causes (16.98%), which included surgical site infection (4 ears), chronic otitis media-active squamous disease (3 ears) as shown in Figure 1 and middle ear infection (2 ears). The other causes of CRI in our study included electrode array extrusion (3 ears) and electrode array malposition (3 ears, 2 in hypotympanum and 1 in Eustachian tube) as shown in Figure 2. One of the patients had both electrode array extrusion and chronic otitis media-active squamous disease as cause of re-implantation at the time of presentation. It was also noted that of the two patients with bilateral simultaneous cochlear implantation had undergone re-implantation because of hard failure (one after 10 years and the other after three years) while the contralateral implant was functional. The different etiology necessitating CRI are tabulated in Table 1. It would be interesting to note that there were total of 15 patients with surgical site infection (SSI), all of whom were hospitalized and initially managed with wound debridement, local flap rotation and injectable antibiotics. Only four patients (26.67%) eventually required cochlear explantation and re-implantation because of SSI.

We further divided the cochlear implantation failure cases into two groups. Group A comprised of implantees in whom CRI was required within one year of primary implantation. While Group B comprised of implantees in whom CRI was required after one year of primary implantation.

There were total of 34 ears (64.15% of total CRI) in Group A as detailed in Table 1. The mean age in this group of patients was 4.2 years for primary implantation and 5 years for CRI. The mean duration of implant usage before undergoing re-implantation was 0.8 year (ranging from one month to 12 months). The most common indication for CRI in this group was device failure in 26 ears (22 ears with hard failure and four ears with soft failure) contributing to 76.5% of the total CRI. The other causes of CRI included SSI (four ears), electrode array extrusion (two ears) and electrode array malposition (two ears). None of the patients had chronic otitis media (active squamous disease) or middle ear infection as an etiology for CRI in this group.

The remaining 19 implanted ears (35.84% of total CRI) requiring CRI belonged to Group B as detailed in Table 1. The mean age of primary implantation and CRI in this group was 5.5 years and 8.5 years, respectively. The mean duration of implant usage before undergoing re-implantation was three years (ranging from 1.25 years to 10 years). The most common indication for CRI in this group was also device failure in 13 ears (12 ears with hard failure and one ear with soft failure) contributing to 68.4% of the total CRI. The other causes of CRI in this group was chronic otitis media--active squamous disease (three ears), middle ear infection (two ears), electrode array extrusion (one ear which coexisted with chronic otitis media--active squamous disease) and electrode array malposition (one ear).

Cumulative survival rate is the cumulative percentage of functioning implants over time and can be used to predict the reliability of the device within a given time frame. The overall CSR of cochlear implants in pediatric population was 96.5% over a period of 20 years at our centre.

Out of the total 1500 consecutive cochlear implantation included in the study, 419 implants belonged to Cochlear Limited (Australia) performed over 20 years (1997-2016). 10 of these implantees underwent CRI. This gives a CSR of 97.61% over 20 years. 56 implants were from Advanced Bionics (United States of America) performed between 2008 and 2016 (9 years period). Only one of these implants required CRI. This gives a CSR of 98.21% over 9 years. The remaining 1025 implants were from MED EL (Austria) performed over 10 years (2007-2016). Forty two of these patients underwent CRI. This gives a CSR of 95.9% over 10 years. The details of number of implants used over the years with the failures and CSR for each of the three United States Food and Drug Administration (USFDA) approved companies is given in Table 2.

Discussion

With increasing number of CI being performed throughout the world including developing countries like ours, most of these implantees, especially the pediatric population, will require CRI. Although several articles have been published on CRI (2, 3), similar studies regarding incidence and etiology are lacking from developing countries like India.

The reported incidence of revision CI surgeries from various studies around the world range from 4.1% to 18.5% with a higher incidence in children compared to adult implantees (1, 5-9). The incidence of CRI in the present study was 3.53% over a period of 20 years. The marginally lower incidence in our study could be justified by the fact that only cases of CRI were included in the present study, and patients who underwent revision surgery without cochlear explantation (e.g. surgical site infection managed without explantation) were excluded. Secondly, the lower incidence can be explained by the fact that most of the implants done at our centre belonged to the newer version of the implants. This is supported by studies having observed that cochlear implant failure rates have been reduced with newer implants (1, 6). In the present study the mean duration of device usage before explantation was 1.6 years (ranging from one month to 10 years), similar to studies available in the literature (1, 5, 8). Thirdly, the lower rate could be because at our centre we follow a standard technique of creating a receiver stimulator bed with tie-down holes and use of sutures to fix the implant. Receiver stimulator recess bed creation with sutures to secure the implant has been shown to be associated with a lower incidence of revision cochlear implantation compared to subperiosteal tight pocket technique in literature (10).

Studies have shown that the most common indication for CRI across the world is device failure (58-78%) followed by medical causes (3-37%) and electrode displacement (6-16%) (1-3, 5-7, 11). The most common indication of CRI in the present study was device failure (73.6%) followed by medical causes (16.98%), similar to world literature. It was noted that device failure was the most common cause for CRI both as an early presentation (64.1%) and as a delayed presentation (76.5%). A recent study by Gardner et al. (9) reported device failure as the most common cause for long term complication of pediatric cochlear implantation. However, none of the cases of early CRI were due to medical causes, probably because medical causes (like chronic otitis media squamous and middle ear infection) have a silent period before they become symptomatic and progress to a stage requiring CRI.

Electrode array malpositioning, though a rare complication of CI, is important because of its serious consequences. It can lead to both injury to important adjacent neurovascular structures that are within millimetres from the cochlea (vestibular system, neural structures within the internal auditory canal, facial nerve, and major vessels) and poor audiological outcomes (12-14). Hence, confirmation of the position of the electrodes intraoperatively by electrophysiologic measures (electrical impedance and neural response telemetry) and/or imaging becomes essential. In contrast, the absence of a detectable intraoperative neural response telemetry threshold has been observed in some patients even with a functional device in the correct location (15). We, at our centre, routinely use intraoperative electrophysiological measures to confirm functional status of electrodes. Imaging is not done intraoperatively at our centre. Post-operative imaging (X-ray) is resorted to in patients with abnormal cochlear anatomy or in doubtful cases where we do not get satisfactory intraoperative neural response telemetry.

The two most common anatomic sites for malposition of electrode arrays are the superior semi-circular canal, followed by the vestibule; while, the most frequent error is inadvertent implantation of a hypotympanic air cell, which is more likely to occur if the round window niche is not clearly identified (10, 14). In our experience, we had three cases of electrode malposition, two in the hypotympanum and one in the Eustachian tube. This may occur even in experienced hands if there is fibrous or bony obliteration of the niche, and thereby reliance on other landmarks (i.e., oval window position and stapedial tendon) becomes essential and important (16).

In the present study we also calculated the cumulative survival rate (CSR) of the cochlear implants used at our centre. The CSR of the three USFDA approved cochlear implants was 97.61% for Cochlear Nucleus over 20 years, 98.21% for Advanced Bionics over 9 years and 95.9% for MED EL implants over 10 years. The overall CSR of cochlear implants was 96.5% over 20 years at our centre, similar to the available literature (4). In a study by Maurer et al. (4) on the reliability of cochlear implants in both adults and children, the CSR was 91.7% over a period of 11 years.

Conclusion

This study from a cochlear implant centre in India provides objective evidence of incidence and various etiology for CRI. The CRI incidence and etiology at our centre appears to reflect the findings of the broader literature. With increasing number of cochlear implantations worldwide, there is going to be an increase in the requirement of CRI. Knowledge about the various causes of CRI and the temporal relation of causes of CRI will help in early identification of the etiology and its management. It will also help in adequate counselling of the patients both before implantation and in the follow-up of the implanted patients. The study also provides the CSR of the three USFDA approved implants performed at our centre. The reliability data can be utilized as an important factor in choice of implant by both doctors and in turn patients.

Ethics Committee Approval: Ethics committee approval was received for this study from the Ethics Committee of Madras ENT Research Foundation (MERF/EC-AUG16/15).

Informed Consent: Written informed consent was obtained from the patients who participated in this study.

Peer-review: Externally peer-reviewed.

Author Contributions: Concept - M.K., S.G.; Design - S.G., S.V.A., N.C., V.V.; Supervision - S.G., A.K.; Resource - M.K.; Materials - S.V.A., V.V.; Data Collection and/or Processing - A.K., N.C., V.V.; Analysis and/or Interpretation - S.G., N.C., V.V.; Literature Search - S.G., S.V.A.; Writing - S.G., N.C., A.K., V.V.; Critical Reviews - M.K.

Conflict of Interest: The authors have no conflicts of interest to declare.

Financial Disclosure: The authors declared that this study has received no financial support.

References

(1.) Masterson L, Kumar S, Kong JH, Briggs J, Donnelly N, Axon PR, et al. Cochlear implant failures: lessons learned from a UK centre. J Laryngol Otol 2012; 126: 15-21. [CrossRef]

(2.) Cullen RD, Fayad JN, Luxford WM, Buchman CA. Revision cochlear implant surgery in children. Otol Neurotol 2008; 29: 214-20. [CrossRef]

(3.) Marlowe AL, Chinnici JE, Rivas A, Niparko JK, Francis HW. Revision cochlear implant surgery in children: the Johns Hopkins experience. Otol Neurotol 2010; 31: 74-82. [CrossRef]

(4.) Maurer J, Marangos N, Ziegler E. Reliability of cochlear implants. Otolaryngol Head Neck Surg 2005; 132: 746-50. [CrossRef]

(5.) Brown KD, Connell SS, Balkany TJ, Eshraghi AE, Telischi FF, Angeli SA. Incidence and indications for revision cochlear implant surgery in adults and children. Laryngoscope 2009; 119: 152-7. [CrossRef]

(6.) Cote M, Ferron P, Bergeron F, Bussieres R. Cochlear reimplantation: causes of failure, outcomes, and audiologic performance. Laryngoscope 2007; 117: 1225-35. [CrossRef]

(7.) Lassig A-AD, Zwolan TA, Telian SA. Cochlear implant failures and revision. Otol Neurotol 2005; 26: 624-34. [CrossRef]

(8.) Kim C-S, Kim D-K, Suh M-W, Oh SH, Chang SO. Clinical outcomes of cochlear reimplantation due to device failure. Clin Exp Otorhinolaryngol 2008; 1: 10-4. [CrossRef]

(9.) Gardner PA, Shanley R, Perry BP. Failure rate in pediatric cochlear implantation and hearing results following revision surgery. Int J Pediatr Otorhinolaryngol 2018; 111: 13-5. [CrossRef]

(10.) Pamuk AE, Pamuk G, Jafarov S, Bajin MD, Sarac S, Sennaroglu L. The effect of cochlear implant bed preparation and fixation technique on the revision cochlear implantation rate. J Laryngol Otol 2018; 132: 534-9. [CrossRef]

(11.) Gosepath J, Lippert K, Keilmann A, Mann WJ. Analysis of fifty-six cochlear implant device failures. ORL J Otorhinolaryngol Relat Spec 2009; 71: 142-7. [CrossRef]

(12.) Ying YL, Lin JW, Oghalai JS, Williamson RA. Cochlear implant electrode misplacement: incidence, evaluation, and management. Laryngoscope 2013; 123: 757-66. [CrossRef]

(13.) Jain R, Mukherji SK. Cochlear implant failure: imaging evaluation of the electrode course. Clin Radiol 2003; 58: 288-93. [CrossRef]

(14.) Cosetti MK, Troob SH, Latzman JM, Shapiro WH, Roland JT, Waltzman SB. An evidence-based algorithm for intraoperative monitoring during cochlear implantation. Otol Neurotol 2012; 33: 169-76. [CrossRef]

(15.) Cosetti MK, Shapiro WH, Green JE, Roman BR, Lalwani AK, Gunn SH, et al. Intraoperative neural response telemetry as a predictor of performance. Otol Neurotol 2010; 31: 1095-9. [CrossRef]

(16.) Woolford TJ, Saeed SR, Boyd P, Hartley C, Ramsden RT. Cochlear reimplantation. Ann Otol Rhinol Laryngol Suppl 1995; 166: 449-53.

Abha Kumari (1) [iD], Sunil Goyal (2) [iD], Senthil Vadivu Arumugam (3) [iD], Viswanathan Vasumathi (3), Neha Chauhan (3) [iD], Mohan Kameswaran (3) [iD]

(1) Department of Otorhinolaryngology and Head & Neck Surgery, Command Hospital, Kolkatta, India

(2) Department of Otorhinolaryngology and Head & Neck Surgery, Army Hospital, Delhi, India

(3) Department of Otorhinolaryngology and Head & Neck Surgery, Madras ENT Research Foundation, Chennai, India

Corresponding Author:

Sunil Goyal; drsunilgoyal@yahoo.co.in

Received Date: 29.1.2018

Accepted Date: 20.02.2019

ORCID IDs of the authors:

A.K. 0000-0001-9589-3187; S.G. 0000-0002-0929-8443; S.V.A. 0000-0001-5403-7667; N.C. 0000-0001-7151-3311; M.K. 0000-0003-4429-4146.

Cite this article as: Kumari A, Goyal S, Arumugam SV, Vasumathi V, Chauhan N, Kameswaran M. A Retrospective Study of Cochlear Re-Implantations--Experience from a Large Centre in India. Turk Arch Otorhinolaryngol 2019; 57(1): 14-8.

DOI: 10.5152/tao.2019.3976
Table 1. Distribution of number of cochlear implantation (CI) failed
based on time frame and etiological factor

Etiology                      Number of implants failed based on time
                              frame after cochlear implantation (CI)
                                 requiring re-implantation (CRI)
                              Group A
                              Less than         Between
                              1 year after CI   1-2 years after CI

Hard failure                     22                 9
Soft failure                      4                 1
Cholesteatoma                     0                 1
Infection of middle ear           0                 0
Cholesteatoma + Electrode         0                 1
array extrusion
Electrode array extrusion         2                 0
Surgical site infection           4                 0
Electrode array malposition       2                 1
Total                            34                13
Percentage of total              64.15             24.53
Cumulative percentage            64.15             88.68

Etiology                         Number of implants failed based on time
                                 frame after cochlear implantation (CI)
                                    requiring re-implantation (CRI)
                             Group B
                             Between             More than         Total
                             2-5 years after CI  5 years after CI

Hard failure                       2                   1            34
Soft failure                       0                   0             5
Cholesteatoma                      1                   0             2
Infection of middle ear            1                   1             2
Cholesteatoma + Electrode          0                   0             1
array extrusion
Electrode array extrusion          0                   0             2
Surgical site infection            0                   0             4
Electrode array malposition        0                   0             3
Total                              4                   2            53
Percentage of total                7.55                3.77
Cumulative percentage             96.23              100

Table 2. Failure rates and cumulative survival rate (CSR) from
1997-2016 of the three USFDA approved cochlear implants from a large
cochlear implantation centre in south India

                      Cochlear Ltd   Advance Bionic  Med El

Period                1997-2016        2008-2016       2007-2016
Duration                20 years          9 years        10 years
Cumulative Total       419               56            1025
no of implants
Number of revisions     10                1              42
Failure rate             2.39%            1.79%           4.1%
Cumulative survival     97.6%            98.2%           95.9%
rate (CSR)

                      Grand total

Period                1997-2016
Duration                20 years
Cumulative Total      1500
no of implants
Number of revisions     53
Failure rate             3.53%
Cumulative survival     96.5%
rate (CSR)
COPYRIGHT 2019 AVES
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2019 Gale, Cengage Learning. All rights reserved.

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Original Investigation
Author:Kumari, Abha; Goyal, Sunil; Arumugam, Senthil Vadivu; Vasumathi, Viswanathan; Chauhan, Neha; Kameswa
Publication:Turkish Archives of Otorhinolaryngology
Article Type:Report