X-ray microanalysis in the variable pressure (environmental) scanning electron microscope.Electron-excited x-ray microanalysis microanalysis /mi·cro·anal·y·sis/ (-ah-nal´i-sis) the chemical analysis of minute quantities of material. microanalysis the chemical analysis of minute quantities of material. performed in the variable pressure and environmental scanning Environmental scanning is a concept from business management by which businesses gather information from the environment, to better achieve a sustainable competitive advantage. electron microscopes electron microscope: see microscope. is subject to additional artifacts artifacts see specimen artifacts. beyond those encountered in the conventional scanning electron microscope scan·ning electron microscope n. Abbr. SEM An electron microscope that forms a three-dimensional image on a cathode-ray tube by moving a beam of focused electrons across an object and reading both the electrons scattered by the object and . Gas scattering scattering In physics, the change in direction of motion of a particle because of a collision with another particle. The collision can occur between two charged particles; it need not involve direct physical contact. leads to direct contributions to the spectrum from the environmental gas, as well as remote generation of x rays by electrons scattered Scattered Used for listed equity securities. Unconcentrated buy or sell interest. out of the focussed beam. The analyst can exert some degree of control over these artifacts, but depending on the exact situation, spurious spu·ri·ous adj. Similar in appearance or symptoms but unrelated in morphology or pathology; false. spurious simulated; not genuine; false. elements can appear at the trace (<0.01 mass fraction), minor (0.01 mass fraction to 0.1 mass fraction), or even major (>0.1 mass fraction) levels. Dispersed dis·perse v. dis·persed, dis·pers·ing, dis·pers·es v.tr. 1. a. To drive off or scatter in different directions: The police dispersed the crowd. b. particle samples give the least compromised results, while fine scale microstructures are the most severely compromised. Procedures to optimize the situation based upon specimen preparation as well as spectral spectral /spec·tral/ (spek´tral) pertaining to a spectrum; performed by means of a spectrum. spec·tral adj. Of, relating to, or produced by a spectrum. processing are described. Key words: energy dispersive dispersive /dis·per·sive/ (-per´siv) 1. tending to become dispersed. 2. promoting dispersion. x-ray spectrometry x-ray spectrometry n. The use of an x-ray spectrometer, especially for chemical analysis of a substance. ; environmental scanning electron microscopy electron microscopy Technique that allows examination of samples too small to be seen with a light microscope. Electron beams have much smaller wavelengths than visible light and hence higher resolving power. (ESEM ESEM Environmental Scanning Electron Microscope ESEM International Symposium on Empirical Software Engineering and Measurement ESEM Experiment of Space Environment with Materials ESEM Ethernet Service Expansion Module ); variable pressure scanning electron microscopy (VP-SEM); x-ray mapping; x-ray microanalysis; x-ray spectrometry. 1. Introduction Characterization of chemical microstructure mi·cro·struc·ture n. The structure of an organism or object as revealed through microscopic examination. microstructure Noun a structure on a microscopic scale, such as that of a metal or a cell is one of the most important applications of the conventional scanning electron microscope (SEM) equipped with an energy dispersive x-ray spectrometer x-ray spectrometer n. A spectrometer using x-rays to separate the chemical constituents of a substance into their characteristic spectral lines for identification and determination of their concentration. (EDS (Electronic Data Systems, Plano, TX, www.eds.com) Founded in 1962 by H. Ross Perot (independent candidate for the President of the U.S. in 1992), EDS is the largest outsourcing and data processing services organization in the country. ) (1). Interest in electron-excited x-ray microanalysis is potentially even greater in the variable pressure (VPSEM) and environmental scanning electron microscopes (ESEM) where dynamic chemical experiments can be conducted. The distinction between the VP-SEM and ESEM is made on the basis of achieving gas pressures that can maintain an equilibrium between water vapor and liquid water. With sample cooling to 2 [degrees]C, this equilibrium can be established at approximately 660 Pa. An arbitrary division between ESEM and VPSEM can be made at 100 Pa. X-ray microanalysis performed in the VPSEM-ESEM is subject to additional conditions and constraints that arise from the presence of the environmental gas and its influence on the primary electron beam A stream of electrons, or electricity, that is directed towards a receiving object. See electron beam imaging and electron beam lithography. . Figure 1 shows schematically sche·mat·ic adj. Of, relating to, or in the form of a scheme or diagram. n. A structural or procedural diagram, especially of an electrical or mechanical system. the effects of elastic and inelastic inelastic Of or relating to the demand for a good or service when quantity purchased varies little in response to price changes in the good or service. sc attering of the beam electrons by the gas atoms. The major consequence of inelastic scattering inelastic scattering n. The scattering of particles resulting from inelastic collision. is the generation of characteristic and continuum (bremsstrahlung bremsstrahlung (brĕm`shträ'ləng): see X ray. bremsstrahlung (German; “braking radiation”) ) x rays from the gas atoms that contribute to the measured Si-EDS spectrum. X-ray production is a relatively rare event, suffered by one in [10.sup.6] electrons or fewer. Most electrons do not suffer significant energy loss from inelastic interactions. The consequences of elastic scattering In scattering theory and in particular in particle physics, elastic scattering is one of the specific forms of scattering. In this process, the energy of the incident photon or particle (electron, positron, or neutron) is conserved and its propagating direction is changed by the are the reduction of beam current within the focused probe and redistribution of this current to form a wide "skirt" around the beam, significantly degrading TO DEGRADE, DEGRADING. To, sink or lower a person in the estimation of the public. 2. As a man's character is of great importance to him, and it is his interest to retain the good opinion of all mankind, when he is a witness, he cannot be compelled to disclose the spatial resolution (Data West Research Agency definition: see GIS glossary.) A measure of the accuracy or detail of a graphic display, expressed as dots per inch, pixels per line, lines per millimeter, etc. It is a measure of how fine an image is, usually expressed in dots per inch (dpi). of x-ray microanalysis. These gas-scattering effects can greatly alter the results achieved with x-ray microanalysis in the VPSEM-ESEM compared to performing a similar x-ray measurement in a conventional SEM under high vacuum conditions, given that the specimen would be compatible with high vacuum. This paper will consider the special aspects of x-ray spectrometry and microanalysis performed in the VPSEM-ESEM, especially the impact of gas scattering on spectrum quality, methods of specimen preparation to minimize the effects of gas scattering, practical aspects of qualitative and quantitative x-ray microanalysis, and prospects for future improvements in this area. 2. X-Ray Spectrometry in the VPSEM-ESEM Electron-excited x-ray spectrometry performed with wavelength dispersive spectrometry spectrometry /spec·trom·e·try/ (spek-trom´e-tre) determination of the wavelengths or frequencies of the lines in a spectrum. spec·trom·e·try n. (WDS Wds Words WDS Wireless Distribution System (Joint Common Database) WDS Wide-area Data Services WDS Wireless Domain Services (Cisco Systems technology) WDS Wavelength Dispersive Spectroscopy ) and/or semiconductor energy dispersive spectrometry (Si-EDS) in the SEM is a mature technique that is widely employed across many of the sciences (1). Specimen excitation excitation Addition of a discrete amount of energy to a system that changes it usually from a state of lowest energy (ground state) to one of higher energy (excited state). For example, in a hydrogen atom, an excitation energy of 10. with a focussed electron beam at a fixed position can achieve lateral spatial resolution down to approximately 1 [mu]m or less, depending on the beam energy and the exact composition of the specimen at the beam location. WDS and Si-EDS have critical strengths and weaknesses (e.g., resolution, spectral coverage, limits of detection, speed of photon processing, etc.) that are mutually supportive, so that combined Si-EDS-WDS instruments represent the most sophisticated level of this instrumentation in conventional SEM applications (1). Because of the constraints imposed by the more aggressive environment of the VPSEM-ESEM, virtually all x-ray spectrometry in these instruments has been performed with Si-EDS, usually equipped with a vacuum isolation windo w that is resistant to water vapor. WDS could, in principle, be incorporated, but the special optical focusing properties of WDS demand precise positioning Precise Positioning is a term used to describe techniques to obtain the location of an object to better than a few centimeters of accuracy. Historically precise positioning was associated with surveying and geodesy. of the electron-excited x-ray source, and the diffractors of the WDS would require special protection to avoid degradation from exposure to the environmental gas. In this paper, we will consider only Si-EDS for performing x-ray spectrometry in the VPSEM-ESEM. After recording an x-ray spectrum at a fixed beam location, the x-ray microanalysis procedure consists of two distinct stages: (1) Qualitative analysis Qualitative Analysis Securities analysis that uses subjective judgment based on nonquantifiable information, such as management expertise, industry cycles, strength of research and development, and labor relations. : The x-ray peaks are assigned to specific elemental elemental emanating from or pertaining to elements. elemental diet see elemental diet. constituents, and the broad categorization of major, minor, and trace is applied to each constituent so identified. These terms are defined (arbitrarily) as: Major: C > 0.1 mass fraction (> 10 weight percent) Minor: 0.01 [less than or equal to] C [less than or equal to] 0.1 mass fraction (1 to 10 weight percent) Trace: C < 0.01 mass fraction (< 1 weight percent) (2) Quantitative analysis Quantitative Analysis A security analysis that uses financial information derived from company annual reports and income statements to evaluate an investment decision. Notes: : A numerical value is assigned to the concentration, along with a statistical measure of the precision as a measure of repeatability and of expected accuracy. A separate procedure, x-ray mapping, involves measuring x-ray intensities while the beam is scanned in a regular array of locations to form an image that depicts the spatial distribution of elemental constituents. In the following discussion, we must also be aware that Si-EDS conducted under conventional high vacuum conditions is itself subject to artifacts (e.g., escape peaks, coincidence or sum peaks, and remote excitation due to backscattered electrons and rescattering of BSEs in the specimen chamber) that must be understood and corrected to achieve optimum results. In the following discussion, an understanding of EDS artifacts in conventional high vacuum operation will be assumed. A large literature on SEM/EDS exists that describes all aspects of the measurement science of the technique, including spectral artifacts, peak identification, various mathematical peak modeling procedures for separating peak and background, accuracy of quantitative analysis, limits of detection, etc. (for comprehensive treatments, see Refs. [1,2,3]). This literature forms the basis for proceeding with Si-EDS in the VPSEM-ESEM. Gas scattering of the primary beam is the single most important difference between performing x-ray spectrometry with the conventional low pressure (i.e., high vacuum) SEM and with the elevated pressure (low vacuum) VPSEM-ESEM instruments. X-ray spectrometry performed in the VPSEM-ESEM must inevitably be compromised because of gas scattering compared to the "ideal" situation in the conventional high vacuum SEM. The key problem to consider for practical microanalysis in the VPSEM-ESEM is determining the concentration level of the analyte in the specimen (major, minor, or trace) for which the results can be trusted. 2.1 Extraneous ex·tra·ne·ous adj. 1. Not constituting a vital element or part. 2. Inessential or unrelated to the topic or matter at hand; irrelevant. See Synonyms at irrelevant. 3. X-Ray Peak(s) Due to the Environmental Gas X-ray spectrometry in the VPSEM-ESEM is subject to additional artifacts beyond those familiar in conventional SEM/EDS. These artifacts are directly related to the presence of the environmental gas. The inevitable gas scattering, both elastic and inelastic, of a fraction of the primary beam electrons has a significant and frequently severe impact on both qualitative and quantitative Si-EDS x-ray microanalysis in the VPSEM-ESEM. Considering first the case of inelastic scattering, both characteristic and continuum (bremsstrahlung) x rays are produced by the incident beam electrons during interactions with the environmental gas atoms. Moreover, the beam electrons that backscatter backscatter in radiology, radiation deflected by scattering processes at angles greater than 90 degrees to the original direction of the beam of radiation. Important in radiotherapy when estimating surface exposure dose. from the specimen can also undergo inelastic scattering events with the environmental gas atoms, further contributing to the measured x-ray spectrum. Although the density of atoms in the gas is very low compared to the atom density in the solid specimen, the volume of the gas which lies within the solid angle of collection of the Si-EDS , even when properly collimated In a straight line. Collimated light beams are parallel rays of light. , is quite large. The EDS accepts x rays from most of the gas path length of the beam from the final pressure limiting aperture An orifice. It often refers to an opening in which light is allowed to pass in optical systems such as cameras and lasers. See f-stop and numerical aperture. to the specimen, a distance of several millimeters. The volume above the specimen into which the backscattered electrons are emitted (following a cosine cosine: see trigonometry. See sine. COSINE - Cooperation for Open Systems Interconnection Networking in Europe. A EUREKA project. distribution for a specimen surface placed normal to the beam) is also within the acceptance of the Si-EDS for the majority of BSEs, with only those lost which are emitted as a result of beam electrons scattered so far out into the skirt that they re-emerge as BSE See Bombay Stock Exchange. BSE See Boston Stock Exchange (BSE). outside the collimated acceptance area of the Si-EDS. Figure 2 (a) shows Si-EDS spectra obtained as a function of water vapor pressure vapor pressure, pressure exerted by a vapor that is in equilibrium with its liquid. A liquid standing in a sealed beaker is actually a dynamic system: some molecules of the liquid are evaporating to form vapor and some molecules of vapor are condensing to form liquid. from a pure carbon disk (2.5 cm in diameter) bombarded with 20 keV electrons with a beam gas path length of 6 mm. The artifact A distortion in an image or sound caused by a limitation or malfunction in the hardware or software. Artifacts may or may not be easily detectable. Under intense inspection, one might find artifacts all the time, but a few pixels out of balance or a few milliseconds of abnormal sound oxygen contribution is barely detectable at the base pressure ([approximately equal to] 50 Pa), but develops into an easily detectable peak at 133 Pa (1 torr torr (tōr), n a unit of pressure equivalent to 0.001316 atmosphere; named after the physicist Torricelli. Also called mm Hg. ) and above. Figure 2 (b) shows a plot of the O/C peak intensity ratio as a function of the pressure. Depending on the pressure, the environmental gas can be detected in the x-ray spectrum as an apparent major, minor, or trace constituent relative to the legitimate peak from the target. At the highest pressure used (2800 Pa = 21 torr), the oxygen peak intensity reached more than 70 % of the C K peak intensity from the carbon target. Close examination of Figure 2 (a) reveals that initially the oxygen intensity increases with increasing water vapor pressure with a gradual lowering of the carbon peak relative to the measurement at base pressure. At the highest pressure, the carbon peak is substantially reduced in intensity compared to the base level. This reduction in carbon x-ray intensity occurs because of elastic scattering into the skirt at distances beyond the acceptance of the EDS collimator collimator (kol´imātur), n a diaphragm or system of diaphragms made of an absorbent material and designed to define the dimensions and direction of a beam of radiation. (see next section) rather than from energy loss. Below approximately 10 Pa (0.1 torr), the contribution of the environmental gas to the x-ray spectrum becomes negligible. When He-H2 gas mixtures are used instead of [H.sub.2]O or air, extraneous x rays from the gas can be eliminated because of the lack of measurable x-ray emission from these atoms. There is still a contribution to the composite spectrum from continuum x rays produced from this gas mixture. For equivalent gas densities, the intensity of this extraneous continuum radiation contribution will be lower for He[H.sub.2] because of the proportional dependence of the continuum on atomic number atomic number, often represented by the symbol Z, the number of protons in the nucleus of an atom, as well as the number of electrons in the neutral atom. Atoms with the same atomic number make up a chemical element. . Does the environmental gas act to significantly absorb the x rays emitted from the specimen? Table 1 gives the results of calculations of absorption for various photon energies as a function of pressure (2500 Pa, 100 Pa, and 10 Pa) for oxygen as the environmental gas with a specimen-to-EDS path length of 4 cm. For the VPSEM pressure range (10 Pa and 100 Pa) x-ray absorption by the gas phase is not a significant effect, being only about 6 % for F K x rays, which are strongly absorbed by oxygen. For the upper end of the ESEM pressure range (2500 Pa), absorption is a significant effect, attenuating FK by 8l%, NaK by 43%, A1K by 20%, and S K by 6 % for a 4 cm specimen to EDS pathlength through the environmental gas. 2.2 Primary Beam Gas Scattering: Remote Excitation of X Rays The electrons scattered out of the beam by elastic interactions with the atoms of the gas form a broad, non-focused "skirt" around the unscattered, focused portion of the primary beam. Danilatos (1988) has described the development of the skirt with the following equation [4]: [r.sub.s] = (364/E) [(p/T).sup.1/2] [L.sup.3/2] (1) where [r.sub.s] = skirt radius, m Z = atomic number of the gas E = beam energy, eV P = pressure, Pa T = temperature, K L = beam path length in gas, m Figure 3 shows examples of the skirt radius calculated with Eq. (1) for various gases (hydrogen, water vapor, and argon argon (är`gŏn) [Gr.,=inert], gaseous chemical element; symbol Ar; at. no. 18; at. wt. 39.948; m.p. −189.2°C;; b.p. −185.7°C;; density 1.784 grams per liter at STP; valence 0. ) for a beam energy of 20 keV and gas path lengths of 5 mm [Fig. 3 (a)] and 15 mm [Fig. 3 (b)]. The scale (in linear dimensions) of the skirt relative to the beam can cover many orders of magnitude. Depending upon the beam energy, environmental gas species and pressure, and the gas path length, the skirt can have a diameter of millimeter or more [e.g., in Fig. 3 (b), Ar above a pressure of 800 Pa gives a skirt radius above 1 mm], while the focused beam diameter The beam diameter of an electromagnetic beam is the diameter along any specified line that is perpendicular to the beam axis and intersects it. For this purpose, the diameter is often defined as the distance between the two diametrically opposite points at which the irradiance is a may be 10 nm or less, giving a skirt/beam ratio of [10.sup.5]. What is the influence of the skirt on electron imaging and x-ray spectrometry performed in the VPSEM and ESEM? While a large fraction, 50 % or more, of the beam current can be transferred from the focused beam to the skirt due to gas scattering, the electron current density (A/[cm.sup.2]) at any point in the skirt is much lower than that in the focussed beam. When a conventional electron image is formed by scanning such a beam/skirt combination over an array of pixels, it is the high current density of the focused beam that produces sharp responses across fine details of the specimen topography topography (təpŏg`rəfē), description or representation of the features and configuration of land surfaces. Topographic maps use symbols and coloring, with particular attention given to the shape and elevations of terrain. and creates a high resolution image. The skirt, which can even contain the majority of the beam electrons, is so spread out and locally diffuse diffuse /dif·fuse/ 1. (di-fus´) not definitely limited or localized. 2. (di-fuz´) to pass through or to spread widely through a tissue or substance. dif·fuse adj. that during a scan, particularly at high magnification Magnification A measure of the effectiveness of an optical system in enlarging or reducing an image. For an optical system that forms a real image, such a measure is the lateral magnification m , the skirt barely moves relative to the fine features that form the interesting image details. The gas-scattered skirt component is so widely spread that the electron imaging signals it produces are completely decoupled from the local imaging environment experienced by the focused beam. Thus, the skirt merely adds a non-specific, s teady state (DC) level to the measured signal which does not vary significantly with the scan position of the primary focused beam. Local topography on the scale of the beam influences the beam related signal only. The principal effect of the skirt signal on image quality is to increase the statistical noise upon which the useful scan varying signal information rides, lowering the signal-to-noise (S/N (1) (Serial/Number) Common shorthand for serial number. (2) (Signal/Noise) As in "s/n ratio." See signal-to-noise ratio. ) ratio. A deterioration de·te·ri·o·ra·tion n. The process or condition of becoming worse. in the S/N ratio S/N ratio - signal-to-noise ratio raises the minimum level of feature contrast that can be rendered visible in the SEM image. This negative impact on signal quality can usually be overcome by increasing the pixel dwell time The time cargo remains in a terminal's in-transit storage area while awaiting shipment by clearance transportation. See also storage. so that adequate images can be obtained even from low contrast features. Useful electron imaging with gas scattering losses as high as 90 % of the total current has been reported, provided enough accumulation time is used (4). Thus, although gas scattering to form the beam skirt is deleterious deleterious adj. harmful. to electron imaging, the interaction of the skirt electrons with the sample can be largely ig nored and high quality, high resolution electron images can be regularly obtained with the VPSEM-ESEM. When we consider the impact of the beam skirt on x-ray spectrometry in the VPSEM-ESEM, we encounter a much different circumstance. The remotely scattered skirt electrons interact with the specimen atoms that they encounter, producing the characteristic and continuum x rays appropriate to each location. If these x rays produced by the skirt electrons are within the solid angle of acceptance of the x-ray spectrometer, they are indistinguishable from the characteristic and continuum x rays produced by the focussed probe within its interaction volume. The x-ray spectrum thus measured is actually a composite spectrum with contributions from the beam and the skirt, but the analyst has no immediate way to distinguish which x rays are from the focussed beam and which are from the skirt. This effect is illustrated in Fig. 4, which shows Si-EDS spectra recorded on a 500 [micro]m diameter wire of 40Cu-60Au (nominal mass The nominal mass is the integer mass of the most abundant naturally occurring stable isotope of an element. The nominal mass of a molecule is the sum of the nominal masses of the elements in its empirical formula. fraction, selected from NIST (National Institute of Standards & Technology, Washington, DC, www.nist.gov) The standards-defining agency of the U.S. government, formerly the National Bureau of Standards. It is one of three agencies that fall under the Technology Administration (www.technology. Standard Reference Material (SRM (1) (Storage Resource Management) The management of the storage resources in an organization in order to avoid duplication of files and to determine space utilization across all servers. ) 482 Copper-Gold Alloys) embedded Inserted into. See embedded system. in a large (2.5 cm diameter) aluminum disk (5). For a fixed beam gas path length of 2 mm, the skirt contribution on the aluminum holder increases as the pressure of the water vapor increases, as shown in the sequence of spectra superimposed su·per·im·pose tr.v. su·per·im·posed, su·per·im·pos·ing, su·per·im·pos·es 1. To lay or place (something) on or over something else. 2. Fig. 4 (a): 53 Pa (0.4 torr); 4 (b) 200 Pa (1.5 torr); and 4 (c) 1600 Pa (12 torr). Plotting the contribution of the A1K peak to the spectrum as the ratio AlK/CuK produces the plot as a function of pressure shown in Fig. 4 (b). A nearly linear response is observed for AlK/CuK vs pressure. This near-linear behavior is a consequence of the circularly symmetric No difference in opposing modes. It typically refers to speed. For example, in symmetric operations, it takes the same time to compress and encrypt data as it does to decompress and decrypt it. Contrast with asymmetric. (mathematics) symmetric - 1. geometry of the wire-disc composite target. From Eq. (1), the skirt radius r [approximately equal to] [p.sup.1/2]. The area of the skirt, which is proportional to [r.sup.2], is therefore proportional to p. Linear behavior vs pressure in the skirt component will therefore be observed providing the gas scattered electrons strike common material, which is the situation for the circular symmetry Circular symmetry in mathematical physics applies to a 2-dimensional field which can be expressed as a function of distance from a central point only. This means that all points on each circle take the same value. target. The linear behavior vs pressure is eventually lost at high pressures because gas scattering takes so much current out of the beam striking the alloy wire that the peak intensities of the alloy components begin to decrease significantly. The sequence of spectra shown in Fig. 4 reveals that, depending on the degree of gas scattering, the pathological 1. pathological - [scientific computation] Used of a data set that is grossly atypical of normal expected input, especially one that exposes a weakness or bug in whatever algorithm one is using. peak due to the contribution to the composite spectrum of the remote scattering onto the surrounding matrix can have the appearance of a trace, minor, or major constituent, Fig. 4 (a). Gas scattering can profoundly influence the interpretation of a spectrum and thus affect both stages of x-ray microanalysis: (1) qualitative analysis wherein where·in adv. In what way; how: Wherein have we sinned? conj. 1. In which location; where: the country wherein those people live. 2. the peaks from gas scattering are assigned to elemental constituents not actually present in the specimen sampled by the unscattered beam, and (2) quantitative analysis, in which the elements contributed by gas scattering will alter the matrix correction calculation by introducing unnecessary corrections for absorption, etc. Observing the composite spectrum obtained from specimens with complex microstructures, such as a fine scale discontinuous discontinuous /dis·con·tin·u·ous/ (dis?kon-tin´u-us) 1. interrupted; intermittent; marked by breaks. 2. discrete; separate. 3. lacking logical order or coherence. phase in a matrix phase, some early researchers quickly became dismayed with the prospects for useful x-ray microanal ysis in VPSEM-ESEM and dismissed the technique prematurely [6; E. Lifshin, State University of New York (body) State University of New York - (SUNY) The public university system of New York State, USA, with campuses throughout the state. at Albany; private communication]. Restricted as they were to using long gas paths in this early instrumentation, it is not surprising that these early workers reached this conclusion. Both elastic and inelastic gas scattering effects can be recognized in the same spectrum. Figure 5 shows a spectrum of an NIST glass (K-230) measured as a small fragment (approximately 50 [mu]m in dimensions) placed on a large (2.5 cm diameter) carbon planchet planch·et n. 1. A flat disk of metal ready for stamping as a coin; a coin blank. 2. A small shallow metal container in which a radioactive substance is deposited for measurement of its activity. at two different pressures, 266 Pa and 1330 Pa [5]. Inelastic scattering can be recognized as the increase in the relative intensity of the oxygen K-peak due to the water vapor used as the environmental gas in spectrum measured at the higher pressure. Elastic scattering manifests itself in the relative increase in the carbon K-peak in the spectrum measured at the higher pressure. A second, less obvious effect of elastic scattering is the increase in the continuum background and the lower peak-to-background ratio of the various spectral peaks (e.g., ZnL, AlK, SiK, PbM, etc.) caused when gas scattering removes electrons that interacted with the glass fragment at the lower pressure but which strike the carbon planchet at higher pressure, generating carbon K-peak x rays and continuum at all other photon energies. 2.3 Charging Effects One of the principal strengths of the VPSEM-ESEM is the possibility of examining of insulating materials without the necessity of modifying the surface with a conductive conductive having the quality of readily conducting electric current. conductive flooring flooring or floor covering made specially conductive to electrical current, usually by the inclusion of copper wiring that is earthed coating, as must be done for the conventional high vacuum SEM. In the VPSEM, the inelastic scattering of beam and backscattered electrons with gas atoms creates free electrons Noun 1. free electron - electron that is not attached to an atom or ion or molecule but is free to move under the influence of an electric field electron, negatron - an elementary particle with negative charge and positive ions. These charged species are automatically attracted to charged areas on the specimen surface, thus acting to discharge them. In the ESEM, the gas ionization ionization: see ion. ionization Process by which electrically neutral atoms or molecules are converted to electrically charged atoms or molecules (ions) by the removal or addition of negatively charged electrons. process is further augmented by secondary electrons Secondary electrons are electrons generated as ionization products. They are called 'secondary' because they are generated by other radiation (the primary radiation). This radiation can be in the form of ions, electrons, or photons with sufficiently high energy, i.e. , emitted from the specimen surface by inelastic scattering of beam and backscattered electrons, that are subsequently accelerated by the applied potential of the gaseous gas·e·ous adj. 1. Of, relating to, or existing as a gas. 2. Full of or containing gas; gassy. secondary electron secondary electron n. An electron produced in secondary emission. secondary electron An electron produced by secondary emission. detector. A cascade of ionization and secondary electron multiplication multiplication, fundamental operation in arithmetic and algebra. Multiplication by a whole number can be interpreted as successive addition. For example, a number N multiplied by 3 is N + N + N. creates a much higher density of charge carriers in the ESEM. Stable images of bare insulators, even with deep holes, can be observed with this form of charge compensation. A full understanding of charge control in the VPSEM-ESEM imaging process continues to evolve. However, even if a stable image of an insulator insulator Substance that blocks or retards the flow of electric current or heat. An insulator is a poor conductor because it has a high resistance to such flow. Electrical insulators are commonly used to hold conductors in place, separating them from one another and from can be obtained, the analyst must still be wary that charging effects can influence the x-ray spectrum. The Duane-Hunt limit, the continuum energy that corresponds to the energy of the incident beam as it reaches the surface of the specimen, is a useful diagnostic to detect charging conditions (1). Figure 6 shows a sequence of EDS spectra measured on an NIST glass, K 1070 (Mg = 0.0750 mass fraction; Si = 0.0 187 mass fraction; Ca 0.0893 mass fraction; Zn = 0.0100 mass fraction; Ba = 0.0112 mass fraction; Pb = 0.0928 mass fraction; O = 0.0343 mass fraction) as a function of pressure from 266 Pa (2 torr) down to 53 Pa (0.4 ton) (6). The spectra are displayed with a logarithmic logarithmic pertaining to logarithm. logarithmic relationship when the logs of two variables plotted against each other create a straight line. intensity axis which makes determination of the Duane-Hunt limit straightforward by extrapolating the high energy continuum to its intersection with the energy axis at 0 intensity. (Note that there will always be some counts above the true Duane Hunt limit due to pulse pile-up pile·up or pile-up n. 1. Informal A serious collision usually involving several motor vehicles. 2. An accumulation: "the pile-up of unsold autos" of lower energy photons.) At 266 pA (2 ton), the spectrum in Fig. 6 (a) shows a Duane Hunt limit of 15 keV, which is equivalent to the beam energy selected, indicating that the gas/ion/electron environment of the VPSEM-ESEM is successfully preventing the accumulation of charge on the specimen surface. When charge builds up on the surface, it acts to accelerate the electron beam. If this charge is negative, as is usually the case for an electron beam incident on an insulating surface, then the negative charge will act to slow the incoming electron velocity Electron velocity is a very important value in computing. Electron is the subatomic particle responsible for eletromagnetic field, that's the way to transmit informations in electronic hardware. so that the landing energy is less than the electron gun A device that creates a fine beam of electrons that is focused on a phosphor screen in a CRT. energy. This effect is seen in Fig. 6 (b), which shows the spectrum obtained at 67 Pa (0.5 torr) where the Duane-Hunt limit has dropped to approximately 13 keV, indicating surface charging to approximately 2000 V. Lowering the pressure to 53 Pa (0.4 torr) in Fig. 6 (c) leads to a further decrease in the Duane Hunt limit to approximately 12 keV. Close examination of Figs. 6 (b) and 6 (c) shows photons recorded between the "real" beam energy and the charge-limited Duane Hunt condition; these photons are too abundant to be solely due to pulse coincidence. These photons are indicative of the dynamic nature of charging, where there may be a momentary mo·men·tar·y adj. 1. Lasting for only a moment. 2. Occurring or present at every moment: in momentary fear of being exposed. 3. Short-lived or ephemeral, as a life. discharge of the accumulated electrons, followed by a rapid build-up build·up also build-up n. 1. The act or process of amassing or increasing: a military buildup; a buildup of tension during the strike. 2. . During the time of the discharge when the beam electrons are not decelerated, a few photons may be created near the true Duane-Hunt limit, but most of the spectrum is dominated by charging. For the spectrum of K1070, the effect of this charging is to lose the high energy characteristic x-ray peaks for Zn K and Pb L, as shown in Fig. 6 (d) with a linear intensity axis for spectra recorded at pressures of 266 Pa (2 ton) and 53 Pa (0.4 ton) with a linear intensity axis. The PbL[alpha] peak is lost entirely due to charging, and ZnK[alpha] and ZnK[beta] are also reduced in intensity. Charging effects such as those illustrated in Fig. 6 can have a catastrophic impact on both qualitative and quantitative x-ray microanalysis by severely modifying the relative peak heights and in some cases completely suppressing high energy peaks. The analyst performing x-ray microanalysis in the VPSEM-ESEM needs to be extremely careful when measuring x-ray spectra from uncoated, insulating specimens. Time dependent, dynamic charging situations are likely to occur. To detect dynamic charging during spectrum accumulation, it is useful to have within the EDS control software the capability to define a series of "ratemeters" covering specified energy ranges that are placed throughout the x-ray spectrum (covering peaks and/or continuum windows), especially in the high photon energy range. If there is no charging, the counting rates should be constant, except for normal statistical fluctuations, in all windows throughout the spectral accumulation. When charging occurs, it should affect the higher photon energy wi ndows near the Duane-Hunt limit first. 3. Strategies for Dealing With Gas Scattering Several strategies have been developed for dealing with gas scattering and the inevitable excitation of portions of the specimen remote from that being interrogated by the direct beam [7]. These strategies depend on the type of specimen to be examined and the level of compositional information sought. However, it must be recognized from the outset that x-ray spectrometry performed in the VPSEM-ESEM can never be as good as that performed in the high vacuum SEM. Nevertheless, the value of x-ray microanalysis in VPSEM-ESEM studies can be so great that it is worth the analyst's efforts to implement additional specimen preparation protocols and analysis procedures to achieve the best possible results within the limitations imposed by gas scattering. 3.1 Selection of Instrumental Parameters to Minimize Gas Scattering Equation (1) describes the impact of various parameters on elastic scattering by the gas atoms. Thus, the size of the skirt and the extent of remote scattering can be reduced by: (1) increasing the beam energy, which decreases the elastic scattering probability; (2) reducing the average atomic number of the environmental gas, e.g., using oxygen instead of argon, helium-hydrogen instead of oxygen); (3) reducing the pressure of the environmental gas to the minimum consistent with stable operation or with the requirements of the experiment; (4) increasing the temperature; (5) decreasing the gas path length. From the exponents on the various terms in equation (1), we can see that modifying the gas path length will have one of the strongest influences on the skirt and remote scattering. Of course, the experimentalist may not have the latitude latitude, angular distance of any point on the surface of the earth north or south of the equator. The equator is latitude 0°, and the North Pole and South Pole are latitudes 90°N and 90°S, respectively. to change specific parameters over the full possible range. For example, there may only be a narrow range of chamber pressure over which stable charge-free imaging can be achieved, or the particular chemical reaction under study may require a specific gas species, pressure and temperature to obtain the desired kinetics kinetics: see dynamics. Kinetics (classical mechanics) That part of classical mechanics which deals with the relation between the motions of material bodies and the forces acting upon them. . Thus, to conduct a particular experiment in the VPSEM-ESEM, the gas path length and the beam energy may be the only variables with which to work. Recognition of the importance of the gas path length in controlling the skirt diameter has led to important advances in the design of the VPSEM-ESEM, especially for operation in the highest pressure regime [4,7]. For the ESEM-class of instruments, which are capable of operating in the pressure range of 100 Pa to 2500 Pa (~ 1 ton to 20 ton) or higher, gas scattering is especially significant, and reducing the gas path length to the practical minimum is necessary to preserve as much spatial resolution as possible. To compensate for the longer working distance needed to accommodate a sidemounted EDS detector while minimizing the gas path length, the low pressure beam path has been extended into the specimen chamber through the use of a conical conical /con·i·cal/ (kon´i-k'l) cone-shaped. con·i·cal or con·ic adj. Of, relating to, or shaped like a cone. pressure limiting aperture with a large altitude-to-diameter ratio. In addition, by bending the side-loading EDS snout snout the upper lip and the apex of the nose, especially of the pig. Called also rostrum. Has a specialized skin to survive the rigors of rooting, is supported by a separate bone (the os rostri), and also has a few sensory hairs. through a large angle from the horizontal, typically 40[degrees] to 50[degrees] then a large positive x-ray take-off angle, 25[degrees] to 50[degrees] can be achieved even when the specimen is oriented o·ri·ent n. 1. Orient The countries of Asia, especially of eastern Asia. 2. a. The luster characteristic of a pearl of high quality. b. A pearl having exceptional luster. 3. so that the beam is placed normal to the surface. For x-ray microanalysis, the choice of beam energy E is frequently constrained con·strain tr.v. con·strained, con·strain·ing, con·strains 1. To compel by physical, moral, or circumstantial force; oblige: felt constrained to object. See Synonyms at force. 2. by the elemental species to be measured. The generation of characteristic x rays depends strongly on the overvoltage, which is the ratio of the incident beam energy ([E.sub.0]) to the critical excitation or edge energy ([E.sub.c]), U = [E.sub.0]/[E.sub.c]. The characteristic intensity I [approximately equal to] [(U - 1).sup.n], where 1.3 [less than or equal to] n [less than or equal to] 1.6 (1). Generally good analysis practice requires an overvoltage of at least U = 2 to produce adequate characteristic peak intensity above the continuum (bremsstrahlung) background. The photon energy range from 0.1 keV to 12 keV provides a K, L, and/or M shell x-ray for all elements in the Periodic Table with Z [less than or equal to] 4 (Be). For conventional microanalysis performed in high vacuum systems vacuum system Urology A mechanical system used to facilitate and maintain an erection; an erection erector. Cf Penile implant. , a general rule is to select an incident beam energy that is a factor of 2 greater than the most energetic elemental excitation edge to be measu red. The choice of a high primary beam energy, 20 keV or more, provides efficient excitation of the upper part of the photon energy range, and is a good choice for the VPSEM-ESEM since a high beam high beam n. The beam of a vehicle's headlight that provides long-range illumination. Noun 1. high beam - the beam of a car's headlights that provides distant illumination energy also serves to minimize gas scattering. However, a negative consequence of a high incident beam energy is the relative reduction in intensity of low photon energy x rays ([less than or equal to] 3 keV) due to increased absorption within the specimen. Absorption occurs because the x rays are generated deeper into the specimen as a result of the greater range of the beam electrons, which increases approximately as [E.sub.0.sup.1.66]. The strategy of lowering the beam energy to reduce the electron range and to lower the absorption losses for low energy photons that is available in conventional high vacuum operation may be severely restricted in VPSEM-ESEM because of the rapid increase in gas scattering as the beam energy decreases. 3.2 Pressure Variation Method Doehne and Bilde-Sorenson and Appel have suggested a method of estimating the skirt contribution to the spectrum through measurements at different pressures to predict the spectrum that would be obtained with no gas scattering (8,9,10). As described in detail by Doehne, this method requires recording two spectra with all other conditions identical except for the pressure (8). If "A" is the spectrum recorded at higher pressure and "B" the lower, then the "zero scattering" spectrum "C" is estimated by: C = B - [(A - B) * d] (2) where "d" is an empirical scaling factor. Implicit in Adj. 1. implicit in - in the nature of something though not readily apparent; "shortcomings inherent in our approach"; "an underlying meaning" underlying, inherent this method is the assumption that changes to the skirt contribution with pressure are primarily due to the skirt intensity rather than the extent of the skirt, that is, the compositional environment excited by the skirt electrons does not change significantly as the extent of the beam skirt increases with increasing gas scattering due to increased pressure (11). If a peak is entirely contributed by the skirt (i.e., the "true" spectrum only arises from the focussed beam), then there will be a distinct change is intensity of the skirt peak(s) as the pressure is changed. Doehne (1997) has demonstrated the capability of eliminating minor intensity skirt peaks by this method (8). Note that because difference methods are used, it is very important to obtain high count spectra, so that the variance in the spectra does not dominate low intensity peaks of interest in the processed spectrum. This method is demonstrated in Fig. 7 using the spectra from the Cu-Au wire in Al block from Fig. 4. Spectra at 200 Pa (1.5 torr), Fig. 7 (a), and 400 Pa (3 torr), Fig. 7 (b), show the increase in AlK with pressure. These spectra are used to form the difference spectrum, Fig. 7 (c). Using Eq. (2) and a scaling factor derived from the ratio of the pressures results in the "corrected to no scattering" spectrum shown in Fig. 7 (d). Examining this spectrum, we note the complete elimination of the spurious Al-K peak, while the fine scale structure of the nearby AuM[zeta] peak of the Au-M family is accurately retained, suggesting that, for at least this type of specimen which satisfies the Doehne-Bower criterion of no spectral change with skirt radius, major and minor peaks are well preserved (11). Figure 8 (a) shows a much more challenging analysis situation, a Raney nickel Raney nickel is a solid catalyst composed of fine grains of a nickel-aluminium alloy, used in many industrial processes. alloy microstructure with fine scale features having dimensions of approximately 10 [micro]m to 20 [micro]m, including three distinct phases with differing Al-Ni compositions, labeled "D" (dark, high aluminum), "I" (intermediate) and "B" (bright, high nickel). These three phases can be readily distinguished in the Si-EDS spectra taken at base pressure, Fig. 8 (b). When the pressure is increased to 665 Pa, Fig. 8 (c), the I and B phases can no longer be readily distinguished with the NiK x-ray peak, while the situation for the AlK peak is nearly as difficult. Table 2 shows the results of the pressure-variation correction procedure for a series of spectra measured on the highest average Z phase in Raney nickel (bright phase). The raw intensity data show the deviations observed from the "no scattering" situation as a function of pressure. As the pressure increases, the apparent phase composition deviates significantly from the "ideal" value found at low pressure. The results of corrections calculated with various choices of the pressure are also presented in Table 2. A "zero scattering" spectrum calculated with the spectra measured at 200 Pa and 400 Pa has a small error for AIK AIK As I Know AIK Assistance in Kind (host nation support) AIK Allmäna Idrottsklubben (Swedish sports club) AIK American Institute of Kenpo (Tucson, AZ marital arts) , but a larger error for NiK than either of the raw spectra. When the 400 Pa and 800 Pa spectra are used to calculate the zero scattering" spectrum, the errors are similar in magnitude but the larger error is now found for the A1K peak. Finally, when the pressure range used to calculate the "zero scattering" spectrum is expanded to 200 Pa to 800 Pa, very large errors in both the A1K and NiK are found. This complex behavior most likely arises because of the complex arrangement and relative sizes of the phases of other composition that surround the particular location used for this measurement series. The gas scattering skirt situation for this alloy differs sharply from the circular symmetric geometry of the wire/disc experiment shown in Fig. 2. The magnitude of the errors suggests that relatively small pressure changes should be used to make corrections to determine the "zero scattering" spectrum. 3.3 Intercepting the Unscattered Beam Bilde-Sorenson and Appel have described a method to estimate the contribution of the skirt based upon comparing two measurements, the first with the primary beam and the skirt striking the specimen and the second after intercepting the primary beam with a beam stop (9,10). In one version, the beam stop consisted of a fine wire composed of an element that was not present in the specimen on interest. The second spectrum with the fine wire in place consists of contributions from the wire material and the skirt contribution. The peak(s) of the wire element are stripped from the beam stop spectrum, and what remains is assumed to be the skirt contribution. This skirt spectrum is then subtracted from the original spectrum to yield the spectrum due to the unscattered beam alone. In another variation of this technique, a foil of a unique element not contained in the specimen is used to cover the half of the specimen. The first spectrum is taken with the beam placed on the specimen near the foil, giving a composite sp ectrum consisting of the location on the specimen plus the skirt. The second spectrum is taken with the beam is placed on the foil close to the bare specimen, which yields the composite spectrum of the foil element plus the skirt contribution on the specimen. Again, the foil peak(s) is stripped off, and the remaining spectrum is considered to be the skirt contribution, which can then be subtracted from the first spectrum. This second procedure is especially suited to line scans made parallel to the edge of the foil. In practical applications, the foil is easier to accommodate than the wire, since this foil can be placed on the specimen externally relative to features of interest and no further manipulation is needed within the VPSEM-ESEM. The beam stop method involves the insertion of the fine wire above the specimen during operation, which requires a micromanipulator micromanipulator /mi·cro·ma·nip·u·la·tor/ (-mah-nip´u-la-ter) an instrument for the moving, dissecting, etc., of minute specimens under the microscope. micromanipulator an instrument for the moving, dissecting, etc. , and if a gaseous secondary electron detector (GSED) is in use, the conducting wire Noun 1. conducting wire - a metal conductor that carries electricity over a distance wire conductor - a device designed to transmit electricity, heat, etc. can interfere with the collection field of the GSED. 3.4 Bremsstrahlung Normalization In relational database management, a process that breaks down data into record groups for efficient processing. There are six stages. By the third stage (third normal form), data are identified only by the key field in their record. Method Griffin and Nockolds have described a method for quantitative x-ray microanalysis in the VPSEM-ESEM that makes use of an internal normalization based upon the use of a window of x-ray continuum (12). Strictly, this method does not attempt to improve the resolution of the analysis by separating the skirt component from that of the direct beam, but rather seeks to re-establish the quantitative relationship between the intensity measured on a large area standard and on an unknown when gas scattering renders invalid the usual measures of beam current necessary for accurate dose normalization. A linear relation was observed between the mean atomic number of a target and the background counts in a specified high photon energy window. Monitoring the continuum intensity in this window provided an accurate correction for beam intensity variations of up to 50 % from a calibrated cal·i·brate tr.v. cal·i·brat·ed, cal·i·brat·ing, cal·i·brates 1. To check, adjust, or determine by comparison with a standard (the graduations of a quantitative measuring instrument): reference value in a conventional SEM. In the VPSEM-ESEM measurements, loss of intensity due to gas scattering outside of the standard would be manifested as a decrease in the intensity measured in the reference continuum window compared to the value expected in a conventional vacuum measurement Vacuum measurement The determination of a gas pressure that is less in magnitude than the pressure of the atmosphere. This low pressure can be expressed in terms of the height in millimeters of a column of mercury which the given pressure (vacuum) will without gas scattering. The authors noted that this correction resulted in VPSEM-ESEM quantitative measurements of similar accuracy as that achieved in conventional practice for silicate mineral silicate mineral Any of a large group of silicon-oxygen compounds that are widely distributed throughout much of the solar system. The silicates make up about 95% of the Earth's crust and upper mantle, occurring as the major constituents of most igneous rocks and in samples. The principal constraint on this method is the requirement that as gas scattering occurs, it must simply reduce the x-ray intensity measured on the unknown and not contribute new characteristic x rays from material of different composition where the skirt impinges. As such, it is best suited to a specimen consisting of well isolated particles on a simple carbon background, as described below. 3.5 X-Ray Focusing Optics X rays can be efficiently focused by means of tapered ta·per n. 1. A small or very slender candle. 2. A long wax-coated wick used to light candles or gas lamps. 3. A source of feeble light. 4. a. polycapillary optics (13). X rays approaching a surface below a certain critical angle are reflected with high efficiency. The critical angle is dependent on photon energy and material and is generally less than 0.01 rad. A capillary capillary (kăp`əlĕr'ē), microscopic blood vessel, smallest unit of the circulatory system. Capillaries form a network of tiny tubes throughout the body, connecting arterioles (smallest arteries) and venules (smallest veins). provides a rotationally symmetric reflector reflector: see telescope. that can propagate prop·a·gate v. 1. To cause an organism to multiply or breed. 2. To breed offspring. 3. To transmit characteristics from one generation to another. 4. x rays along its length through multiple low angle reflections. By gradually tapering Tapering Gradually reducing the amount of a drug when stopping it abruptly would cause unpleasant withdrawal symptoms. Mentioned in: Narcotics tapering, n the capillary, x rays can be made to follow a converging or diverging di·verge v. di·verged, di·verg·ing, di·verg·es v.intr. 1. To go or extend in different directions from a common point; branch out. 2. To differ, as in opinion or manner. 3. path. To maximize efficiency, the solid angle must contain as much reflecting surface as possible with as little solid glass, which acts to absorb the x rays. By bundling thin walled capillaries Capillaries The smallest arteries which, in the lung, are located next to the alveoli so that they can pick up oxygen from inhaled air. Mentioned in: Adult Respiratory Distress Syndrome, Birthmarks, Platelet Count into polycapillaries, the amount of reflecting surface is maximized, and efficient optical components can be made. Figure 9 shows results of Wollman et al. on the focussing properties of one side of a double tapered polycapillary in which the electron excited source of x rays from titanium titanium (tītā`nēəm, tĭ–) [from Titan], metallic chemical element; symbol Ti; at. no. 22; at. wt. 47.88; m.p. 1,675°C;; b.p. 3,260°C;; sp. gr. 4.54 at 20°C;; valence +2, +3, or +4. metal is moved in the plane pe rpendicular to the optic axis optic axis n. The line connecting the anterior and the posterior poles of the eye. (14). A sharp focus function is observed, such that the transmission is reduced by 60 % with approximately a 60 [micro]m movement of the source off the maximum transmission point. Moreover, these measurements show similar behavior for photons spanning a broad energy range from Ti-L at 450 eV to Ti-K at 4500 eV. The absolute efficiency is energy dependent, with efficiency increasing for lower photon energies due to an increasing critical angle. The polycapillary optic can thus act as a spatial filter A spatial filter is an optical device which uses the principles of Fourier optics to alter the structure of a beam of coherent light or other electromagnetic radiation. Spatial filtering is commonly used to "clean up" the output of lasers, removing aberrations in the beam due to on a distributed x-ray source, such as that which exists in the VPSEM-ESEM due to the gas scattering skirt so as to exclude the remotely produced x rays while efficiently collecting x rays produced by the focused probe. In practice, x-ray mapping could be used to establish the position of the maximum transmission spot relative to the specimen position to maximize the efficiency. Such an optic would have the additional benefit of increasing the solid angle of collection of the EDS relative to an ordinary detector of the same size placed at the same distance. The spectrum measured with the optic would suffer the effects of the energy dependent transmission of the optic. Measurements of the x-ray continuum with and without the optic could be used to accurately establish the transmission function. 4. X-Ray Microanalysis in VPSEM-ESEM: Influence of the Specimen Configuration 4.1 The Analytical Blank For certain specimen types, such as particles, the degree of success that can be achieved with x-ray microanalysis in the VPSEM-ESEM depends upon the configuration of the specimen. The central question is the validity of the measured EDS x-ray spectrum. When the focused beam is placed on the particle of interest, is the measured x-ray spectrum representative of the particle constituents? At what equivalent concentration level is the particle spectrum compromised by the remote scattering of beam electrons into the skirt? Can major (C > 0.1), minor (0.01 [less than or equal to] C [less than or equal to] 0.1), or trace constituents (C < 0.01) be trusted? Can the constituents, especially at minor and trace levels, be confidently assigned to the beam analysis position or must the analyst inevitably accept the much broader sampling of the beam/skirt combination? In the following discussion, progressively more complicated samples will be considered. Underlying all measurements is the critical concept of the "analytical blank," which is the irreducible irreducible /ir·re·duc·i·ble/ (ir?i-doo´si-b'l) not susceptible to reduction, as a fracture, hernia, or chemical substance. ir·re·duc·i·ble adj. 1. background level of each constituent of interest contributed by all materials present except for the specimen itself. Determining and minimizing the analytical blank is especially important to achieve robust x-ray microanalysis in the VPSEM-ESEM. 4.1.1 The Analytical Blank for Particle Analysis: Conventional SEM The "analytical blank" for particle analysis by conventional SEM microanalysis is the spectrum of the specimen support substrate exposed to all stages of the specimen preparation procedure, including transfers in the laboratory environment, but without the specimen itself applied to the substrate. The spectrum obtained from the blank thus contains x-ray peaks arising from the substrate and adhesion adhesion /ad·he·sion/ (ad-he´zhun) 1. the property of remaining in close proximity. 2. the stable joining of parts to one another, which may occur abnormally. 3. materials as well as anything added as a result of specimen preparation and inadvertent contamination. 4.1.2 The Analytical Blank for Particle Analysis: VPSEM-ESEM For VPSEM-ESEM analysis, the analytical blank is more complicated and consists of two levels, the preparation blank and the operational blank. The preparation analytical blank is identical to the conventional blank, except that the substrate must be measured in the VPSEM-ESEM under identical conditions of gas species, pressure, and gas path length as will be used for the unknowns. Thus, gas scattering effects, including excitation of x rays from the environmental gas as well as the skirt electrons striking various materials, are included with the contributions from the direct beam striking the substrate and interacting with the environmental gas. Figure 10 (a) shows the Si-EDS spectrum of a preparation blank of a carbon planchet, in which a major carbon peak is observed with a small oxygen peak arising from the water vapor used as the environmental gas (333 Pa with a beam gas path length of 3 mm and a specimen to Si-EDS distance of 15 mm). Figure 10 (b) shows the spectrum of a preparation blank of carbon tap e (with adhesive on both surfaces) mounted on a similar carbon planchet measured under the same conditions. A major carbon peak is again observed but with a substantially higher oxygen peak compared to Fig. 10 (a), with the additional oxygen arising from the adhesive and polymer tape substrate. Thus, this particular preparation is compromised with regard to the measurement of carbon at major constituent levels and oxygen at least to the minor constituent level by the constituents of the mounting materials. The operational blank for VPSEM-ESEM considers the additional contributions that arise from the actual environment of the prepared sample, for example, from the particles surrounding the particle of interest in a dispersion dispersion, in chemistry dispersion, in chemistry, mixture in which fine particles of one substance are scattered throughout another substance. A dispersion is classed as a suspension, colloid, or solution. deposited on a substrate. Figure 11 (a) shows a dispersion of particles of NIST SRM 1633 (Trace metals in fly ash fly ash n. Fine particulate ash sent up by the combustion of a solid fuel, such as coal, and discharged as an airborne emission or recovered as a byproduct for various commercial uses. Noun 1. ) deposited on double-sided adhesive carbon tape, the preparation blank for which is shown in Fig. 10 (b) (5). Figure 11 (b) shows an example of the operational analytical blank for this array of particles, as measured at the location marked in Fig. 11 (a). The consistency of the operational blank can be estimated by moving the beam off the particle of interest to a series of nearby locations, e.g., "BL1", "BL2", etc. marked in Fig. 11 (a). When this is done, the measured peak intensities represent the contributions from the skirt striking the substrate and other nearby particles. (Note that even with this careful protocol, the determination of the operational blank is necessar ily imperfect imperfect: see tense. because it does not consider the contribution due to electrons from the focussed beam that scatter scat·ter v. 1. To cause to separate and go in different directions. 2. To separate and go in different directions; disperse. 3. To deflect radiation or particles. n. off the particle of interest onto surrounding particles.) The operational blank can be expected to vary with beam position and with the exact arrangement of particles, and thus it should be measured at several locations to determine its variability. In this case, the operational blank spectra measured at "BL 1" [Fig. 11 (b)], "BL2" [Fig. 11 (c)], and further away at "BL3" [Fig. 11 (d)] are very similar, although this does not have to be the case. The low but distinct levels of aluminum, silicon, calcium, and iron are the result of the skirt electrons striking other particles in the neighborhood, while carbon and at least some of the oxygen arise from the substrate and tape, as determined from the preparation blank. After determining the operational blank, a representative spectrum should be compared to each measured particle spectrum. Examples are shown for several of the particles in Fig. 11 (a) i n Figs. 11 (e), (f), and (g). Taking into account the peak heights observed when the beam was placed on pure element standards, the operational blank indicated that several elements were compromised in particle measurements: carbon as a major constituent; oxygen, aluminum and silicon as minor constituents; and calcium and iron as trace constituents. The situation might change in another area on the same specimen where particles are more widely dispersed, or different species are present, so the operational blank must be continually updated for each local region. 4.2 Isolated Particles 4.2.1 Particles Dispersed on Carbon-Containing Substrates Particles are often collected on non-conducting filter media such as paper and plastics. Such filters may be impossible to examine directly in the conventional SEM, even with heavy coating, because the extreme topography of the filter leads to inadequate coating coverage and subsequent charging. In such cases, this necessitates removal of the substrate and/or transfer of the particles. However, because microscopic particles are often easily modified morphologically mor·phol·o·gy n. pl. mor·phol·o·gies 1. a. The branch of biology that deals with the form and structure of organisms without consideration of function. b. and chemically when exposed to the solvents needed to remove the filter medium, such removal and transfer may not be possible without compromising the information that is sought. Situations that require examination of particles as collected on the filter medium represent an ideal opportunity for LVSEM-ESEM microscopy microscopy /mi·cros·co·py/ (mi-kros´kah-pe) examination under or observation by means of the microscope. mi·cros·co·py n. 1. The study of microscopes. 2. and microanalysis. The optimum sample configuration to achieve nearly uncompromised x-ray microanalysis in LVSEM-ESEM is that of widely dispersed particles Noun 1. dispersed particles - (of colloids) a substance in the colloidal state dispersed phase phase, form - (physical chemistry) a distinct state of matter in a system; matter that is identical in chemical composition and physical state and separated from . Wide dispersal dis·per·sal n. The act or process of dispersing or the condition of being dispersed; distribution. Noun 1. dispersal minimizes contributions to the operational blank from nearby particles and decreases its variability. The particles are deposited on a simple substrate, most typically high purity carbon, at a loading density such that the particles are spaced by at least 10 to 100 times their largest lateral dimension. With such a wide dispersion, the spectrum of an individual particle should be uncompromised for major and minor constituents, with the exception of those elements noted in the preparation blank that arise from the substrate and the environmental gas. To determine the validity of trace element peaks, the analyst must be prepared to carefully examine the operational blank associated with each particle. Carlton tested this situation with artificial constellations Constellations Constellation English name Position R.A. (hours) DEC. (degrees) Andromeda Andromeda (Chained Lady) 1 +43 Antlia Air Pump 10 −33 Apus Bird of Paradise 16 −75 Aquarius1 of particles (dimensions from 100 [mu]m to 400 [mu]m) [15]. For example, with a beam energy of 20 ke y, a chamber pressure of 599 Pa (4.5 torr) of water vapor, and particles (approximately 10 [mu]m dimensions) separated by approximately 3.5 mm, the contribution of the remote particles [pure copper, glass (of unspecified Adj. 1. unspecified - not stated explicitly or in detail; "threatened unspecified reprisals" specified - clearly and explicitly stated; "meals are at specified times" composition), and the mineral cassitterite] to the central spectrum of a titanium particle occurred at the level of Cu-K/Ti-K = 0.0014 and Sn-L/Ti-K = 0.00029, which are just at the threshold At the Threshold, whose son Lil E. Tee won the 1992 Kentucky Derby for W. Cal Partee, died March 23 of a stroke at Purdue University School of Veterinary Medicine in West Lafayette, Ind. The 21-year-old stallion stood at Wayne Houston's Stoney Creek Horse Farm near Mooreland, Ind. of trace detection for EDS performed with practical dose conditions. 4.2.2 Reducing the Background Use of Thin Foil Substrates As the size of the particle is reduced, the relative proportions of the spectrum contributed by the direct beam striking the particle and by the skirt striking the bare substrate and surrounding particles change. While the skirt contribution remains constant, the x-ray intensity generated by the direct beam on the particle decreases with particle size Particle size, also called grain size, refers to the diameter of individual grains of sediment, or the lithified particles in clastic rocks. The term may also be applied to other granular materials. , especially for particles below approximately 1 [micro]m in thickness (i.e., the dimension along the beam). For sufficiently small sufficiently small - suitably small particles, the remote skirt spectrum from the bulk carbon substrate will eventually dominate the particle spectrum. After the operational parameters of beam energy, environmental gas species, gas pressure, and beam path length have been chosen, the remaining variable to reduce the skirt contribution to the composite spectrum is to modify the substrate itself. A low mass thickness substrate can be obtained with a free standing carbon film (20 nm nominal thickness) supported on a electron microscope grid (e.g., copper, nickel or carbon ). Such a thin carbon film is surprisingly strong, and particles can be deposited on the film by various methods, including by drying particle-loaded liquid droplets. This grid with the particle deposit is then placed over a deep (at least 5 mm) blind hole in a carbon block so that electrons passing through the film are likely to be absorbed by the substrate and not backscattered to strike the specimen or grid again. Figure 12 (a) shows such a preparation of NIST K309 glass particles on a carbon film (20 nm nominal thickness) supported on a copper grid with 80 [micro]m square openings. Figures 12 (b) and (c) show a comparison of EDS spectra from similar, micrometer-sized K309 particles on a bulk carbon tape substrate as compared to carbon thin film substrate. Relative to the aluminum and silicon peaks from the glass, the spectrum from the film on grid preparation shows carbon reduced by a factor of at least 10 compared to the spectrum measured on the bulk carbon tape substrate, where carbon is the highest pea pea, hardy, annual, climbing leguminous plant (Pisum sativum) of the family Leguminosae (pulse family), grown for food by humans at least since the early Bronze Age; no longer known in the wild form. k. Note that the spectral peak-to-background is higher for the thin film support. At Si, the P/B P/B See: Price to book ratio is approximately twice as high, and therefore the measured spectrum consists of a larger contribution from the particle relative to the substrate compared to the equivalent case with a bulk carbon substrate. For sub-micrometer particles, the improvement in P/B would be even greater. Note that the particle spectrum from the thin film also contains an additional artifact, the x rays emitted when the skirt electrons strike the support grid, which is copper in this case. To avoid this artifact, other metallic and nonmetallic non·me·tal·lic adj. 1. Not metallic. 2. Chemistry Of, relating to, or being a nonmetal. Adj. 1. support grids are available, including other metals (Ni, Al, stainless steel stainless steel: see steel. stainless steel Any of a family of alloy steels usually containing 10–30% chromium. The presence of chromium, together with low carbon content, gives remarkable resistance to corrosion and heat. ), carbon and nylon. 4.2.3 Alternatives to Carbon Substrates If carbon is of interest in the particles, other high purity elemental substrates such as aluminum foil Noun 1. aluminum foil - foil made of aluminum aluminium foil, tin foil foil - a piece of thin and flexible sheet metal; "the photographic film was wrapped in foil" or silicon wafers are readily available. Often Al and Si are of interest themselves, so these materials may not be satisfactory choices for the substrate. As an alternative, high purity beryllium beryllium (bərĭl`ēəm) [from beryl ], metallic chemical element; symbol Be; at. no. 4; at. wt. 9.01218; m.p. about 1,278°C;; b.p. 2,970°C; (estimated); sp. gr. 1.85 at 20°C;; valence +2. would be of particular interest as a substrate, since its characteristic x-ray is of such low energy (110 eV) that it is not detectable by most EDS systems. Unfortunately, beryllium in the form of beryllium oxide Beryllium oxide (BeO) is a white crystalline oxide. It is obtained from beryllium or beryllium compounds by ignition in the air. The sintered beryllium oxide (beryllia), which is very stable, has ceramic characteristics. is highly toxic highly toxic Occupational medicine adjective Referring to a chemical that 1. Has a median lethal dose–LD50 of ≤ 50 mg/kg when administered orally to 200-300 g albino rats 2. , and this fact greatly constrains its use, especially if the planchet surface must be polished produce a flat surface with the inevitable possibility of contamination of the laboratory. As an alternative, elemental boron boron (bōr`ŏn) [New Gr. from borax], chemical element; symbol B; at. no. 5; at. wt. 10.81; m.p. about 2,300°C;; sublimation point about 2,550°C;; sp. gr. 2.3 at 25°C;; valence +3. and its oxide are not significantly toxic, although its x-ray peak at 185 eV is detectable with high performance EDS systems. Boron is extremely hard, and a highly polished surface can be produced with an appropriate polishing protocol (16). 4.3 Fibers If the specimen is in the form of an individual fiber, it is again possible to improve the quality of the x-ray spectrum by reducing or even completely eliminating the skirt contribution from the substrate. By placing the fiber over a large diameter, deep, blind hole in a carbon block, the skirt electrons have nothing with which to interact within the solid angle of acceptance of the EDS and their contribution is effectively eliminated except for the environmental gas itself. An example of this approach is seen in the images in Fig. 13 (a), where a fiber of NIST glass K230 (O = 0.0245 mass fraction; Al = 0.0265 mass fraction; Si = 0.0140 mass fraction; Zn = 0.0402; Ba = 0.0896 mass fraction; Ta = 0.0409 mass fraction; Pb = 0.0418 mass fraction), is suspended over a deep, 3 mm diameter hole while attached on either end to a pad of carbon tape. The x-ray spectrum of a 16 [micro]m diameter fiber obtained at 266 Pa (2 torr, [H.sub.2]O) with the beam placed in the center of the hole is shown in Fig. 13 (b), while the spectrum obtained where the same fiber is attached to the carbon tape is shown in Fig. 13 (c), where the carbon peak from the skirt is prominent. Note that the carbon peak is almost completely eliminated in the 266 pA (2 torr) spectrum recorded in the center of the hole. Thus carbon-bearing fibers could be successfully characterized by x-ray spectrometry with this specimen mounting procedure, which could be improved even further through the use of a support that did not contain carbon, e.g., by supporting the fiber across a hole in an aluminum disk, or some other metal not of particular interest. A further variant of this fiber technique can be used to obtain high quality spectra of particles. A small diameter fiber with a thin layer of adhesive can be used to mount particles for suspension over a hole. While the material of the fiber will contribute to the spectrum, it may be possible to make this contribution insignificant to the particle spectrum with careful choice of the mounting fiber material. Lithium lithium (lĭth`ēəm) [Gr.,=stone], metallic chemical element; symbol Li; at. no. 3; at. wt. 6.941; m.p. about 180.54°C;; b.p. about 1,342°C;; sp. gr. .534 at 20°C;; valence +1. Lithium is a soft, silver-white metal. tetraborate glass fibers would be of particular use here, since the only significant spectral contribution would be from the oxygen component. Figure 14 (a) shows an example of a blank spectrum from a 20 [micro]m diameter fiber of lithium tetraborate glass suspended over a 3 mm diameter hole in a carbon block. The only significant characteristic peak is that of oxygen. The corresponding skirt spectrum obtained with the beam placed just off the fiber is shown in Fig. 14 (b); this spectrum is virtually identical but the oxygen peak intensity is a factor of three less. Figure 15 (a) shows a 5 [micro] m particle of unknown composition attached to this fiber. The EDS spectrum of this particle, Fig. 15 (b), reveals a large iron peak, and lower intensities for sulfur and chlorine. The particle may also contain oxygen, but the interference from the strong oxygen peak from the lithium tetraborate glass fiber precludes interpretation. 4.4 Bulk Specimen 4.4.1 Homogeneous, Single Phase Bulk specimens that are compositionally homogeneous over large regions and consist of a single phase can be analyzed in VPSEM-ESEM without significant difficulty. As long as the lateral specimen dimensions exceed the diameter of the skirt, as determined with Eq. (1) and/or with Monte Carlo simulation Monte Carlo Simulation A problem solving technique used to approximate the probability of certain outcomes by running multiple trial runs, called simulations, using random variables. , then the effect of the gas scattering is simply to degrade TO DEGRADE, DEGRADING. To, sink or lower a person in the estimation of the public. 2. As a man's character is of great importance to him, and it is his interest to retain the good opinion of all mankind, when he is a witness, he cannot be compelled to disclose the size of the probe, but in the absence of a microstructure, this spatial degradation is unimportant un·im·por·tant adj. Not important; petty. un im·por tance n. .
Since the energy lost during gas scattering is not significant, all of
the electrons in the broadened probe, including those in the skirt, are
equivalent in terms of x-ray excitation. X rays from the environmental
gas can contribute to the spectrum as the only significant artifact, and
if the gas pressure is at the high end of the ESEM range (> 1500 Pa),
absorption can occur, especially for low energy peaks. Operation in the
VPSEM pressure range or the low portion of the ESEM pressure range
should effectively eliminate both of these artifacts. Thus, e xcept for
the loss of spatial resolution, the analysis of large, homogeneous
targets in the VPSEM-ESEM is essentially equivalent to microanalysis
performed in the conventional high vacuum SEM but with a highly
defocused beam.Since the skirt electrons are energetically equivalent to the beam electrons, it should be possible to perform quantitative x-ray microanalysis in the VPSEM-ESEM for large, homogeneous targets. The most rigorous quantitative x-ray microanalysis procedure is based upon the measurement of standards (e.g., pure elements or stoichiometric stoi·chi·om·e·try n. 1. Calculation of the quantities of reactants and products in a chemical reaction. 2. The quantitative relationship between reactants and products in a chemical reaction. compounds) under the same conditions (beam energy, known dose, and spectrometer spectrometer Device for detecting and analyzing wavelengths of electromagnetic radiation, commonly used for molecular spectroscopy; more broadly, any of various instruments in which an emission (as of electromagnetic radiation or particles) is spread out according to some efficiency) as the unknown followed by the calculation of matrix correction factors to convert x-ray intensity ratios for each x-ray peak (sample/standard) into concentration ratios (1). In principle, standard intensities could be measured for the unscattered beam, while the gasscattered beam with the same total current could be used for the large, homogeneous unknown. Alternatively, very large standards could be used to accommodate the skirt electrons, but this may be procedurally difficulty when the need is for a large suite of standards. In actual practice, it is very difficult in the VPSEM-ESEM to establish this measurement equivalence between the unknown and the standards. The large lateral spread A technique used to place the mean point of impact of two or more units 100 meters apart on a line perpendicular to the gun-target line. of the gas-scattered beam results in the possibility that some portion of the skirt electrons may strike the unknown in areas where the efficiency of the EDS drops off. This situation is further complicated by the use of a support grid for the ultrathin ul·tra·thin adj. Very thin. window in the modern EDS spectrometer. The detector window efficiency depends strongly on the position at which an x-ray strikes the window. If the x-ray path intersects the grid, the low energy photon efficiency drops off severely. When all of the x rays are effectively produced as a point source with lateral dimensions of only a micrometer micrometer (mīkrŏm`ətər, mī`krōmē'tər). 1 Instrument used for measuring extremely small distances. , the illumination illumination, in art illumination, in art, decoration of manuscripts and books with colored, gilded pictures, often referred to as miniatures (see miniature painting); historiated and decorated initials; and ornamental border designs. of the EDS window is well defined. When the x rays are produced over a large spatial area as in the case of the VPSEM-ESEM, then the illumination of the EDS window is much more complex and difficult to bring under measurement control. Moreove r, if the region excited by the skirt extends outside the area of acceptance of the collimator of the EDS, then it is critical that the unknown and standard be located at the same position relative to the center of the beam of the VPSEM-ESEM and the central axis of the the diameter of the sphere which is perpendicular to the plane of the circle. See also: Axis EDS. With sufficient care in this positioning and using bremsstrahlung to compensate for dose variations, Griffin and Nockolds have demonstrated satisfactory results in the quantitative analysis of large mineral specimens with ESEM-EDS (12). As a result of this difficult measurement situation, "standardless" analysis, in which the necessary standard intensities are calculated theoretically or else estimated from remotely measured standards, is often the method of choice for the VPSEM-ESEM. It must be recognized, however, that "standardless" analysis procedures, even under optimal conditions in the conventional SEM, have been shown to produce much broader error distributions than those obtained with standards (17,18). Before reporting to a customer the results of any "standardless analysis procedure" applied in a VPSEM/ESEM, it is strongly recommended that the analyst assess the accuracy of the standardless procedure by testing it against known multi-component homogeneous standards measured under conventional high vacuum conditions. Figure 16 (a) shows an example of a homogeneous block of glass embedded in silver-loaded epoxy epoxy Any of a class of thermosetting polymers, polyethers built up from monomers with an ether group that takes the form of a three-membered epoxide ring. The familiar two-part epoxy adhesives consist of a resin with epoxide rings at the ends of its molecules and a curing in a hole drilled into a block of titanium. Figure 16 (b) shows the EDS spectrum obtained with a pressure below 50 Pa (0.4 torr), in which the only significant peaks are those for the known constituents of the glass. When this spectrum was processed through the standardless analysis procedure embedded in the commercial analytical software Analytical software is software that is designed specifically for and development of a particular environment or object. package that supported the particular EDS spectrometer used, the results given in Table 3 were obtained. The relative errors are typical of the distribution observed in a detailed test of standardless analysis performed previously (18). When the pressure was increased to 650 Pa (5 Torr), the spectrum included additional peaks arising from the silver-loaded epoxy and the titanium block. By instructing the quantitative analysis software to ignore the Ag-L and Ti-K peaks, the results listed in Table 3 were obtained, with similar but somewhat larger relative errors compared to the uns cattered case. The increase in the error observed at the higher pressure may be due to increased uncertainty in the peak fitting introduced by the additional x-ray peaks contributed by the skirt. Since normalization is forced upon the results in standardless analysis, this has the effect of distributing the error over all components. 4.4.2 Heterogeneous, Multiphase Mul´ti`phase a. 1. (Elec.) Having many phases; Adj. 1. multiphase - of an electrical system that uses or generates two or more alternating voltages of the same frequency but differing in phase angle Microstructures The most difficult analytical situation for the VPSEM-ESEM is the case of a heterogeneous microstructure with two or more phases, such as a discontinuous phase in a bulk matrix (e.g., an inclusion in a matrix), especially when the problem involves determining elemental constituents that are partitioned par·ti·tion n. 1. a. The act or process of dividing something into parts. b. The state of being so divided. 2. a. between the two phases. As usual, the effective beam footprint due to gas scattering acts to produce a composite spectrum from all of the phases. This effect is shown in Fig. 10. A three-phase microstructure, Fig. 10 (a), produces sharply different spectra when measured under pressure conditions (<50 Pa), Figs. 10 (b), (c), (d), while with significant gas scattering at 650 Pa of water vapor, the spectra begin to converge, Figs. 10 (e), (f), and (g). The bright and intermediate phases, Figs. 10 (e) and (f), are virtually indistinguishable, while the dark phase, which actually contains a very low level of nickel, as seen in Fig. 10 (d) for the no gas scattering condition, appears to contain significant nickel du e to gas scattering, Fig. 10 (g). The level of spectral differentiation that will be observed obviously depends very strongly on the exact compositional nature and scale of the of the microstructure relative to the gas scattering footprint. While a series of spectra measured as a function of pressure might make it possible to deduce de·duce tr.v. de·duced, de·duc·ing, de·duc·es 1. To reach (a conclusion) by reasoning. 2. To infer from a general principle; reason deductively: the no-gas-scattering condition, it is clear that these measurements would have to be made with extreme care to maintain a consistent gas scattering situation. 5. X-Ray Mapping in the VP-ESEM X-ray mapping provides powerful visual information on the spatial distribution of elemental constituents at micrometer lateral resolution and is one of the most widely applied qualitative analysis procedures. An x-ray map is created by assigning a gray level in the image storage/display according to according to prep. 1. As stated or indicated by; on the authority of: according to historians. 2. In keeping with: according to instructions. 3. the x-ray intensity measured at each pixel. Generally, the intensity is measured (and stored) on a 16 bit deep scale (maximum 65,536 counts), but for display purposes the intensity range is scaled to 8 bits (0-255), with the maximum intensity set to white (level 255). In the conventional mapping procedure that is incorporated in most commercial software systems, no background or peak overlap corrections are applied, so that the intensity maps may be valid only for major constituents (concentrations above 0.1 mass fraction). For minor and trace level constituents, the continuum background forms a progressively larger fraction of the measured intensity as the concentration decreases, so that the atomic number dependence of the x-ray continuum eventually dominates the contrast in the x-ray map. With background and overlap corrections, mapping can be useful for minor and even trace constituents. A good quality assurance procedure in x-ray mapping is to record x-ray spectra of each phase that apparently contains minor and trace constituents to determine the validity of any apparent contrast between phases observed in maps. If the x-ray c ontrast between any pair of phases is valid, it should be confirmed by a proportional change in the relative peak heights in spectra measured from those phases. Invalid contrast situations may arise due to the atomic number dependence of the x-ray continuum, but this situation will be revealed in spectra taken in the regions in question. X-ray mapping is subject to the additional artifacts discussed throughout this chapter that are peculiar to the VPSEM-ESEM. The most serious is the action of gas scattering of the primary beam to degrade the effective spatial resolution through remote excitation of x rays. A particular consequence of gas scattering for x-ray mapping is the decrease in contrast between phases as the beam gas path length increases. This artifact is especially serious for bulk specimens where the remotely scattered electrons excite characteristic x rays from the same elemental species as in the mapped area. Figures 17 (a) and (b) show the results of x-ray mapping for the major Al and Ni constituents of Raney nickel alloy over a range of pressures. The maps obtained at the lowest pressure [133 Pa (1 torr) and a 4 mm gas path] showed three distinct major phases, sharply defined by the variation in both the Al and Ni intensities. Quantitative EDS analysis of these phases (performed in a conventional SEM with pure element standards and quantitative calculations with Desktop Spectrum Analyzer A hardware device or software used to examine the frequency and power components of a signal. It provides more information than an oscilloscope, because it can display the signals over a range of frequencies. ) yielded the results shown in Table 4, which indicate the degree of compositional contrast.) As the chamber pressure was progressively increased from 133 Pa (1 torr), 665 Pa (5 torr), 1330 Pa (10 ton), and finally 2000 Pa (15 torr), the x-ray maps showed a decrease in resolution and in the contrast between phases. The visibility of the intermediate and bright phases, which yield a concentration contrast of about 25% from Table 4, decreased sharply and was only barely visible in the maps recorded at 1330 Pa (10 torr). At 2000 Pa (15 torr) this contrast was completely lost, and only two phases could be discerned. 6. Conclusions X-ray microanalysis in the VPSEM and ESEM can be a useful tool to complement SEM imaging, but the analyst must recognize the inevitable limitations that result from gas scattering compared to the level of analytical performance achieved in a conventional high vacuum SEM. The impact of gas scattering on both qualitative and quantitative analysis generally increases as the concentration of a constituent of interest is lowered from major (C > 0.1 mass fraction), to minor (0.01 [less than or equal to] C [less than or equal to] 0.1), to trace (C < 0.01 mass fraction) levels. While it is usually possible to achieve useful results for major constituents, minor and trace constituents are likely to be severely compromised. To minimize the effects of gas scattering, the beam gas path length must be made as short as possible, consistent with accommodation of the EDS x-ray spectrometer. For certain classes of specimens, such as particles and fibers, the analyst can also seek to minimize the contributions of the backgrou nd through the use of thin film supports (particles) and suspension over holes (fibers). Quantitative analysis of areas with micrometer dimensions is severely compromised, but methods have been developed to correct for the contributions of the electron scattering Electron scattering is the process whereby an electron is deflected from its original trajectory. Electrons are charged particles and are acted upon by the electromagnetic forces. They are scattered by other charged particles through the electrostatic Coulomb forces. skirt by measuring spectra over a range of pressures. X-ray mapping under VPSEM-ESEM conditions suffers in terms of the minimum compositional contrast that can be detected, as well as in terms of degraded de·grad·ed adj. 1. Reduced in rank, dignity, or esteem. 2. Having been corrupted or depraved. 3. Having been reduced in quality or value. spatial resolution. [FIGURE 2 OMITTED] [FIGURE 3 OMITTED] [FIGURE 4 OMITTED] [FIGURE 5 OMITTED] [FIGURE 6 OMITTED] [FIGURE 7 OMITTED] [FIGURE 8 OMITTED] [FIGURE 10 OMITTED] [FIGURE 11 OMITTED] [FIGURE 12 OMITTED] [FIGURE 13 OMITTED] [FIGURE 14 OMITTED] [FIGURE 15 OMITTED] [FIGURE 16 OMITTED]
Table 1
Characteristic x-ray transmission by the environmental gas (oxygen)
(specimen to EDS window: 4 cm)
Element/X-ray [I/I.sub.o] (2500 Pa) [I/I.sub.o] (100 Pa)
F K 0.194 0.940
NaK 0.572 0.979
AlK 0.805 0.992
SiK 0.868 0.995
S K 0.939 0.998
C1K 0.957 0.998
K K 0.986 0.999
CaK 0.990 0.9996
Element/X-ray [I/I.sub.o] (10 Pa)
F K 0.994
NaK 0.998
AlK 0.9992
SiK 0.9995
S K 0.9998
C1K 0.9998
K K 0.9999
CaK 0.9999
Table 2
Analysis of raney nickel high-Z phase with corrections applied for gas
scattering
Pressure (Pa) A1K % Deviation NiK % Deviation
ref. ref.
50 149826 153029
200 169089 +13 % 155919 +1.9 %
400 192770 +29 % 140844 -8 %
800 212216 +42 % 121235 -21 %
Corr. (400-200) 145576 -3 % 170915 +12 %
Corr. (800-400) 173384 +16 % 160437 +5 %
Corr. (800-200) 67505 -55 % 260043 +70 %
Table 3
VPSEM-ESEM EDS analysis of NIST glass K411 (SRM 470)
Element Certified Analysis 1 Relative Analysis 2
mass fraction (< 50 Pa) uncertainty (%) (650 Pa)
Mg 0.0885 0.0725 -18 % 0.058
Si 0.2538 0.2696 +6.2 % 0.2564
Ca 0.111 0.147 +32 % 0.158
Fe 0.112 0.148 +32 % 0.1731
[O.sup.l] 0.424 0.363 -14 % 0.354
Element Relative
uncertainty (%)
Mg -34 %
Si 1 %
Ca +42 %
Fe +55 %
[O.sup.l] -17 %
Table 4
Composition of phases in raney nickel alloy (phases designated according
to gray level in a BSE image)
A1 Contrast [DELTA]C/C Ni Contrast [DELTA]C/C
Dark 0.988 0.0122
Intermediate 0.588 40% 0.409 97%
Bright 0.434 26% 0.565 28%
Accepted: August 22, 2002 7. References (1.) J. I. Goldstein, D. E. Newbury, P. Echlin, D. C. Joy, A. D. Romig, Jr., C. E. Lyman, C. Fiori, and E. Lifshin, Scanning Electron Microscopy and X-ray Microanalysis, 2nd Ed., Plenum In a building, the space between the real ceiling and the dropped ceiling, which is often used as an air duct for heating and air conditioning. It is also filled with electrical, telephone and network wires. See plenum cable. Press, New York New York, state, United States New York, Middle Atlantic state of the United States. It is bordered by Vermont, Massachusetts, Connecticut, and the Atlantic Ocean (E), New Jersey and Pennsylvania (S), Lakes Erie and Ontario and the Canadian province of (1992). (2.) C. E. Fiori and D. E. Newbury, Artifacts in Energy Dispersive X-ray Spectrometry in the Scanning Electron Microscope, Scanning Electron Microscopy, II (1980) pp. 251-258. (3.) D. E. Newbury, Artifacts in Energy Dispersive X-ray Spectrometry in Electron Beam Instruments. Are Things Getting Any Better?, in X-ray Spectrometry in Electron Beam Instruments, D. B. Williams, J. I. Goldstein, and D. E. Newbury, eds., Plenum Press, New York (1995) pp. 167-201. (4.) G. D. Danilatos, Foundations of Environmental Scanning Electron Microscopy, Adv. Electronics Electron Phys. 71, 109-249 (1988). (5.) NIST Standard Reference Material (SRM) 482 Copper-Gold Alloys); Standard Reference Materials Program, National Institute of Standards and Technology National Institute of Standards and Technology, governmental agency within the U.S. Dept. of Commerce with the mission of "working with industry to develop and apply technology, measurements, and standards" in the national interest. , Gaithersburg, MD 20899. www.nist.gov (6.) R. B. Bolon, X-ray Microanalysis in the ESEM, Microbeam Analysis-1991, D. G. Howitt, ed., San Francisco San Francisco (săn frănsĭs`kō), city (1990 pop. 723,959), coextensive with San Francisco co., W Calif., on the tip of a peninsula between the Pacific Ocean and San Francisco Bay, which are connected by the strait known as the Golden Press (1991) pp. 199-200. (7.) B. J. Griffin, Effects of Chamber Pressure and Accelerating Voltage on X-ray Resolution in the ESEM, Proc. 50th Annual Meeting, Electron Microscopy Society of America, G. W. Bailey, J. Bentley, and J. A. Small, eds., San Francisco Press (1992) pp. 1324-1325. (8.) E. Doehne, A New Correction Method for High-Resolution Energy-Dispersive X-ray Analyses in the Environmental Scanning Electron Microscope, SCANNING 19, 75-78 (1997). (9.) J. Bilde-Sorenson and C. C. Appel, Energy-Dispersive X-ray Spectrometry in the Environmental Scanning Electron Microscope, abstracts of 48th meeting of the Scandinavian Society for Electron Microscopy, A. B. Maunsbach, ed., SVF SVF Serial Vector Format (ICT in circuit test) SVF Simple Vector Format SvF Sjónvarp Føroya (Television of the Faroes; Faeroe Islands) SVF State Variable Filter SVF Stoicorum Veterum Fragmenta , Aarhus, Denmark (1996). (10.) J. Bilde-Sorenson and C. C. Appel, X-ray Spectrometry in ESEM and LVSEM: Corrections for Beam Skirt Effects, abstracts of 49th meeting of the Scandinavian Society for Electron Microscopy, A. R. Tholen, ed., Svenski Tryck I, Goteborg, Sweden (1997). (11.) E. Doehne and N. Bower, Empirical evaluation of the electron skirt in the environmental SEM: Implications for energy dispersive x-ray microanalysis, Microbeam Analysis 2, S35-36 (1993). (12.) B. J. Griffin and C. E. Nockolds, Quantitative EDS Analysis in the Environmental Scanning Electron Microscope (ESEM) Using a Bremsstrahlung Intensity-Based Correction for Primary Beam Variation and Scatter, Microsc. Microanal. 842-843 (1996). (13.) M. A. Kumakov and F. F. Komarov, Multiple reflection from surface x-ray optics X-ray optics By analogy with the science of optics, those aspects of x-ray physics in which x-rays exhibit properties similar to those of light waves. , Phys. Rep. 191, 289-350 (1990). (14.) D. A. Wollman, C. Jezewski, G. C. Hilton, Q.-E. Xiao, K. D. Irwin, L. L. Dulcie, and J. M. Martinis, Use of polycapillary optics to increase the effective area of microcalorimeter spectrometers, Microsc. Microanal. Suppl. 2 3, 1075-1076 (1997). (15.) R. Carlton, The Effect of Some Instrumental operating Conditions on the X-ray Microanalysis of Particles in the Environmental Scanning Electron Microscope, SCANNING 19, 85-91 (1997). (16.) E. S. Windsor, D. E. Newbury, J. D. Kessler, and P. H. Chi, Boron Substrates for Parliculate X-ray Microanalysis, Microsc. Microanal. Suppl. 2 5, 920-921 (1999). (17.) D. E. Newbury, C. R. Swyt, and R. L. Myklebust, Standardless' Quantitative Electron Probe Microanalysis with Energy-Dispersive X-ray Spectrometry: Is It Worth the Risk?, Anal anal (a´n'l) relating to the anus. a·nal adj. 1. Of, relating to, or near the anus. 2. . Chem. 67, 1866-1871 (1995). (18.) Dale E. Newbury, Standardless Quantitative Electron-Excited X-ray Microanalysis by Energy-Dispersive Spectrometry: What Is Its Proper Role? Microsc. Microanal. 4, 585-597 (1999). About the author: Dale E. Newbury is an NIST Fellow in the Surface and Microanalysis Science Division of the Chemical Science and Technology Laboratory of NIST in Gaithersburg, MD. He was educated at Lehigh University Lehigh University, at Bethlehem, Pa.; coeducational; chartered and opened 1866 by Asa Packer. It has undergraduate colleges of arts and science, business and economics, and engineering and applied science, as well as several graduate programs. and the University of Oxford. His research interests include x-ray spectrometry, scanning electron microscopy, and Monte Carlo Monte Carlo (môNtā` kärlō`), town (1982 pop. 13,150), principality of Monaco, on the Mediterranean Sea and the French Riviera. modeling of electron beam-specimen interactions. The National Institute of Standards and Technology is an agency of the Technology Administration, U.S. Department of Commerce. |
|
||||||||||||||||||

im·por
tance n.
Printer friendly
Cite/link
Email
Feedback
Reader Opinion