Printer Friendly
The Free Library
14,558,467 articles and books
Member login
User name  
Password 
 
Join us Forgot password?

Understanding the IRHD and durometer methods used in rubber hardness testing.


Hardness is one of the most widely measured properties used to characterize rubber. The IRHD IRHD International Rubber Hardness Degree  (international rubber hardness degree) scale and the durometer scale are widely used. A number of instrument types exists for both - the IRHD Micro/Dead Load and type A durometer scales are most commonly used for rubber. Both methods are described in international standards (refs. 1, 2 and 22).

The two test methods use totally different indentor geometries, applied forces, test times and procedures. The IRHD test is usually non-destructive, and as such has to be the preferred method for final product inspection; the test takes 35 seconds. In contrast, the durometer method is often destructive (leaving a permanent indentation in·den·ta·tion
n.
A notch, a pit, or a depression.
), but the test only takes 1-3 seconds. This article begins with a historical look at the instruments and the degree of correlation between them.

Instruments exist for most of the IRHD and durometer scales, both as tabletop and hand held versions. The IRHD dead load has a micro counterpart counterpart n. in the law of contracts, a written paper which is one of several documents which constitute a contract, such as a written offer and a written acceptance. , which has had an established standard for over 30 years (ISO (1) See ISO speed.

(2) (International Organization for Standardization, Geneva, Switzerland, www.iso.ch) An organization that sets international standards, founded in 1946. The U.S. member body is ANSI.
 48 [ref. 1] and ASTM ASTM
abbr.
American Society for Testing and Materials
 1415 [ref. 2]). The proposed micro durometer does not yet have a released standard. (Since this article was written, the M type durometer has been included in the current durometer hardness standard [ref. 23]). The micro IRHD instrument was introduced in the 1950s as a scaled down version of the IRHD dead load -- used for testing thinner, smaller samples and products. The micro IRHD results are generally comparable to those given by the standard IRHD dead load instrument. In contrast, there are various type M durometers appearing on the market, some constructed quite differently. The type M durometer results are generally not comparable to those obtained from the type A scale.

Experience has shown that a degree of confusion exists between some users of the two scales. This article highlights the merits of each instrument and test type. In cases where measurements must be made on small, awkward or curved production samples, the test methods and sample dimensions may be highly significant and it is important to know the limitations of each test method.

In summary, this article aims to create a clear understanding of common hardness testing methods and the results they provide

Historical perspective

According to according to
prep.
1. As stated or indicated by; on the authority of: according to historians.

2. In keeping with: according to instructions.

3.
 Bassi bas·si  
n.
A plural of basso.
 et al (ref. 3) the durometer had historical priority over the IRHD instruments by more than 30 years. Gurney gurney /gur·ney/ (gur´ne) a wheeled cot used in hospitals.

gur·ney
n. pl. gur·neys
A metal stretcher with wheeled legs, used for transporting patients.
 (ref. 4) reported both instruments in use by the early 1920s, together with other spring and dead load (weight) variants. Results from the spring type varied with the user (Gurney [ref. 4], The Rubber Age [ref. 5]). This lead to the adoption of the dead load instrument where the indentation depth was largely user independent. After Scott (ref. 6) stressed the need for a standard to give results some common meaning in 1935, the first British Standard (BS) was introduced in 1940. At the same time, Scott and Newton (ref. 7) reported on a reliable pocket type hardness gauge that conformed to this new standard. After a comparison with the type A durometer, they concluded that the advantage was always with the BS hardness meter. Work was then carded out looking at different instrument types (Daynes and Scott [ref. 8] and the new standard [Scott, ref. 9]. They both agreed that there was some correlation between the type A durometer and BS hardness scales Hardness scales

Arbitrarily defined measures of the resistance of a material to indentation under static or dynamic load, to scratch, abrasion, or wear, or to cutting or drilling.
.

The accuracy of a range of hardness testers (Newton, [ref. 10]) was investigated, concluding that the main limitations were associated with the operator. Instruments with a spherical spher·i·cal
adj.
Having the shape of or approximating a sphere; globular.
 indentor and foot gave the smallest errors; the largest errors were associated with the durometer. The largest source of variation reported by Scott (ref. 11) was the lack of agreement between laboratories.

The micro hardness tester, a (1/6th) scaled down version of the IRHD dead load hardness tester was introduced in the 1950s to test thinner and small production samples. Scott and Soden (ref. 12) reported results comparable between the micro and dead load tests noted for rubbers of greater than 65 [degrees] hardness.

Several papers (refs. 3, 13, 14 and 15) published in the 1960s, 70s and 80s and two books, Rubber and Plastics Testing (ref. 16) and Physical Testing of Rubbers (ref. 17) stated that the most widely used instrument was then the type A durometer even though the IRHD method produced more repeatable results between operators with higher accuracy, reproducibility reproducibility Lab medicine  The degree of agreement among repeated measurements of a particular parameter, presented in terms of a standard deviation or coefficient of variation of the results in a set of measurements  and precision. However, type A has a less critical dependence than IRHD on sample thickness (Bassi et al, [ref. 3]). Comparative work by Brown and Soekarnein in 1991 (ref. 18) between the IRHD dead load, IRHD micro and type A durometer instruments demonstrated that inter-laboratory repeatability was likely to be best for the IRHD dead load and micro instruments. In 1993, Briscoe and Sebastian (ref. 19) analyzed an·a·lyze  
tr.v. an·a·lyzed, an·a·lyz·ing, an·a·lyz·es
1. To examine methodically by separating into parts and studying their interrelations.

2. Chemistry To make a chemical analysis of.

3.
 the durometer indentation, providing an approximate relationship between IRHD and durometer type A of (IRH IRH Institute for Reproductive Health
IRH Inverclyde Royal Hospital
IRH Institute for Research in Human Happiness
IRH Inspection Requirements Handbook
 [approximately equals] [H.sub.A] + 4), although this is very dependent on the sample compound.

Many contemporary hardness testers have improved accuracy due to the automatic nature of the test, requiting minimal operator intervention A procedure used in a lawsuit by which the court allows a third person who was not originally a party to the suit to become a party, by joining with either the plaintiff or the defendant. . Bench mounted instruments (IRHD dead load and micro and type A durometer scales) produce the most repeatable and reliable results. Pocket meters are much improved, but do rely entirely on the operator's hand pressure and reliable angular angular /an·gu·lar/ (ang´gu-lar) sharply bent; having corners or angles.  application for repeatable results (variations can be extreme).

In recent years, there has been an increased interest in the type M durometer, although there is no published standard. There are a variety of these instruments on the market from several suppliers and some are constructed quite differently. Recently, Wallace Wal·lace , Alfred Russel 1823-1913.

British naturalist who developed a concept of evolution that paralleled the work of Charles Darwin.
 has manufactured type M durometers to the best draft information available. The results are not comparable to those obtained from a type A durometer.

Differences between IRHD instruments and durometers and relationships between scales

There are four IRHD methods in use: the normal-hardness test (dead load), high-hardness test, low-hardness test and the micro test. The normal test is used for samples greater than or equal to 4 mm thick and preferably pref·er·a·ble  
adj.
More desirable or worthy than another; preferred: Coffee is preferable to tea, I think.



pref
 used for rubbers in the 35-85 IRHD range (but with reservation, may be used for the 30-95 IRHD range). The high-hardness test is used for testing samples of the same dimensions as the normal test, but in the 85-100 IRHD range. The low-hardness test is used for testing samples greater than or equal to 6 mm thick and hardnesses in the 10-35 IRHD range. The micro tests samples less than 4 mm thick and is used for rubber in the 35-85 IRHD range (but with reservation may be used for the 30-95 IRHD range). All four methods use a spherically spher·i·cal   also spher·ic
adj.
1.
a. Having the shape of a sphere; globular.

b. Having a shape approximating that of a sphere.

2. Of or relating to a sphere.

3.
 tipped indentor. The diameters of the ball indentor and foot vary between methods. The applied forces are the same for the normal, high and low tests, with only the micro test requiting the application of smaller forces. It is worth noting that the IRHD scale is non-linear.

The durometer range of hardness testers incorporates eight scale types: A, B, C, D, DO, O, OO and M. These are used for testing a wider range of materials. The A scale is used for soft rubbers and elastomers and type C for medium hard rubbers and plastics; both types use a truncated truncated adjective Shortened  cone cone, in botany
cone or strobilus (strŏb`ələs), in botany, reproductive organ of the gymnosperms (the conifers, cycads, and ginkgoes).
 shaped indentor. Type A is the most commonly used rubber scale. Type B is used to test moderately hard rubbers and type D is used to test hard rubbers and plastics. Both of these use a 30 [degrees] indentor. Type DO is used for very dense textile windings, type O is used for soft rubbers and medium density textiles and OO is used for low density textile windings and sponge. These three use a 3/32 inch spherically ended indentor. All types require samples more than 6 mm thick (unless it can be proved that smaller samples give equivalent results). Type M is used for testing thin and irregular HEIR, IRREGULAR. In Louisiana, irregular heirs are those who are neither testamentary nor legal, and who have been established by law to take the succession. See Civ. Code of Lo. art. 874.  rubbers of hardness in the range 20-90 and uses a very small round tipped indentor. Thinner samples may be used, although the support table starts to affect the value as thickness falls as the indentor penetrates the sample. Spring forces vary between instruments. Type A, B and O use the same spring force and it is recommended that a force equivalent of 1 kg is applied to the durometer to ensure that the spring force is repeatably overcome (note that the DIN standard uses 1.27 kg and tighter indentor dimensional limits). Type C, D and DO use the same spring, requiting a force equivalent of 5 kg to overcome the spring. Type OO uses a different spring and requires 400 g. Type M currently suggests a force suitable to overcome the calibrated cal·i·brate  
tr.v. cal·i·brat·ed, cal·i·brat·ing, cal·i·brates
1. To check, adjust, or determine by comparison with a standard (the graduations of a quantitative measuring instrument):
 spring force. All Shore scales are linear.

The IRHD method is based on the use of dead loads (weights). A foot is used to hold the sample in place with a force of 8.3 N (dead load) or 235 mN in the case of the micro hardness tester. A primary load of 0.3 N (dead load) or 8.3 mN (micro hardness tester) is then applied for five seconds, providing a datum The singular form of data; for example, one datum. It is rarely used, and data, its plural form, is commonly used for both singular and plural.  position. A secondary load of 5.4 N (dead load) or 145 mN (micro) is then applied for 30 seconds. The incremental Additional or increased growth, bulk, quantity, number, or value; enlarged.

Incremental cost is additional or increased cost of an item or service apart from its actual cost.
 displacement displacement, in psychology: see defense mechanism.


Same as offset. See base/displacement.
 from the datum is measured and converted to an IRHD value (a non-linear scale defined in the standard). The full-range displacement of (normal) dead load is 1.8 mm; the micro uses 0.3 mm.

In contrast, the durometers use calibrated springs. For example, the type A spring force varies from 0.5 N to 8.1 N (over the full displacement) and the type M from 0.3 N to 0.8 N. The presser foot the part of a sewing machine which rests on the cloth and presses it down upon the table of the machine.

See also: Presser
 applies a force sufficient to overcome the spring force. Once the presser foot contacts the sample, the indentation depth is recorded after a preset preset Cardiac pacing A parameter of a pacemaker that is programmed permanently when manufactured  dwell time The time cargo remains in a terminal's in-transit storage area while awaiting shipment by clearance transportation. See also storage. ; the standard ASTM dwell times are one and three seconds. The DIN standard uses three seconds, since the reading is usually still changing appreciably ap·pre·cia·ble  
adj.
Possible to estimate, measure, or perceive: appreciable changes in temperature. See Synonyms at perceptible.
 after one second. The force increases linearly with indentor displacement (full range is 2.5 mm for the A scale and 1.25 mm for the M scale).

The IRHD scale was set in 1948 to correspond to the durometer scale, in that a high number indicates a hard rubber and a low number indicates a softer rubber. The original micro hardness test was designed to be a scaled down version of the normal dead load test (displacements in the ratio 6 to 1). The forces applied were in the ratio 36 to 1. Therefore, if the limited thickness sample tested in the case of a micro instrument is 1/6th of the thickness of the dead load piece, 1/6th of the result is obtained. Scaling is set so that the same result should be obtained from both instruments. Similarly, results from the normal, high and low dead loads show correlation.

The type M durometer test was not designed as a scaled down version of the type A test, but merely as an instrument that was capable of testing smaller samples. It uses an unrelated indentor and spring; there is no easy relationship between the two instruments.

Experimental

It follows from the above discussion that results obtained from hand-held durometers are not reliable due to operator dependence. Therefore, only bench mounted instruments were used to obtain the experimental results; however, the conclusions drawn will also be relevant to hand held instruments. Since the majority of robber and elastomers use the A and M durometer scales, other durometer scales will be disregarded dis·re·gard  
tr.v. dis·re·gard·ed, dis·re·gard·ing, dis·re·gards
1. To pay no attention or heed to; ignore.

2. To treat without proper respect or attentiveness.

n.
 here. These two instruments are also the main counterparts of the normal and micro IRHD instruments.

All instruments were calibrated before starting and the calibration calibration /cal·i·bra·tion/ (kal?i-bra´shun) determination of the accuracy of an instrument, usually by measurement of its variation from a standard, to ascertain necessary correction factors.  was rechecked at the end. A standard temperature of 23 [+ or -] 2 [degrees] C was used (except where otherwise noted). The durometers were set to both one and three second dwell time (since the results from these times differ). Test times are defined by the standard for the IRHD instruments (five and 30 seconds). Each flat sample was tested in five different places, and curved samples were tested as specified below.

A previous work (ref. 20) compared the micro IRHD and micro durometer instruments, looking at the effects of sample thickness, lateral lateral /lat·er·al/ (-il)
1. denoting a position farther from the median plane or midline of the body or a structure.

2. pertaining to a side.


lat·er·al
adj.
1.
 dimensions, bent samples, temperature, retesting a previously measured spot and the effect of the foot force on the type M durometer. This article aims to continue and extend this work to include the dead load instrument and type A durometers, as well as incorporating results from curved surfaces.

Standard Wallace test blocks (varying compounds of natural rubber, supplied by MRPRA) for both dead load and micro instruments were used to provide comparative results for each instrument.

The previous work (ref. 20) on sample thickness was limited and therefore this is now investigated in further detail. The ISO standard (ref. 1) allows 1 mm thick samples to be used, but the preferred thickness is 2 [+ or -] 0.5 mm. It is known that the type M durometer has similar requirements. Tests were performed on a range of thinner materials. The durometer standard (ref. 22) suggests that samples be plied plied 1  
v.
Past tense and past participle of ply1.
 to increase their effective thickness; this was done to determine the effect of varying sample thickness. This was extended to similar work on the IRHD dead load instruments and type A durometers. The IRHD standard thickness is 8-10 mm while the type A is 6 mm. A selection of thinner samples was tested and plied to determine the effect of varying sample thickness.

Previous work (ref. 20) with the micro instruments included an investigation into the effect of increasing the ambient temperature Outside temperature at any given altitude, preferably expressed in degrees centigrade. . Therefore, tests were carded out on the dead load instrument and type A durometers at a raised temperature to determine any effect.

Curved samples, such as o-rings, are often tested, and the effect of testing these on different instruments was investigated. O-rings (of varying outer and core diameters Core Diameter can be defined as in the cross section of a realizable optical fiber, ideally circular, but assumed to a first approximation to be elliptical, the average of the diameters of the smallest circle that can be circumscribed about the core-cladding boundary, and the ) were placed on a specially designed table so that they could be accurately displaced displaced

see displacement.
 laterally lat·er·al  
adj.
1. Of, relating to, or situated at or on the side.

2. Of or constituting a change within an organization or a hierarchy to a position at a similar level, as in salary or responsibility, to the one being left:
 to determine the effect of testing away from the top dead center.

Results

Standard test blocks

The standard test blocks gave repeatable results using micro and normal dead load instruments. The one and three second dwell times (type A and M) produced equivalent and repeatable results. The dead load readings were consistently a few units higher than the type A durometer readings over the range tested (40-90 IRHD). However, there was an increasing tendency for the type M results to diverge diverge - If a series of approximations to some value get progressively further from it then the series is said to diverge.

The reduction of some term under some evaluation strategy diverges if it does not reach a normal form after a finite number of reductions.
 from the micro IRHD result with increasing hardness values.

Effects of thickness

The IRHD dead load and type A durometer instruments were used to test the standard Wallace micro samples (2 mm thick). As expected, the results differed from those obtained using the specified instrument for the sample thickness, i.e., the micro IRHD and the type M durometer. The softer rubbers, and also the IRHD dead load instrument, exhibited greater differences between the micro and macro instrument results. For the hardest rubber (76-79 IRHD), the IRHD dead load instrument gave a very close value (figure 1). The type A durometer read a few units lower, as expected, but the readings were closer between the type A and type M durometers. The type A durometer value of the hardest rubber differed by only one unit to the type M value (figure 1).

[GRAPH OMITTED]

Once the 2 mm thick samples were plied to 8 mm thick (the standard thickness required for the IRHD dead load tester) the results came within the specified tolerances of the test pieces. Increasing the thickness further made little difference to the result. Using the type A durometer, the results after plying Plying, in textile manufacture, is the activity of twisting, intermingling, or otherwise intimately combining two or more fibers or yarns into a combined yarn or fiber. Plying Yarns  pieces to provide a 6 mm thick sample (the standard thickness for type A) was not equivalent to the value when tested with a type M durometer (figure 2).

[GRAPH OMITTED]

In contrast, the standard dead load blocks of 8 mm thick, when tested on the micro IRHD and the type M durometer instruments, tended to give approximately the same results as the dead load IRHD and type A durometer instruments.

Various thinner samples were used with the micro instruments. Up to five pieces of polychloroprene (0.6 mm thickness) were plied, taking the thickness of the sample into the standard tolerance region (and slightly beyond). Both the IRHD micro and the type M durometer showed a continual decrease in hardness with increasing thickness (figure 3). As before, the IRHD hardness values were consistently higher than the type M durometer values. For a nitrile nitrile: see rubber.  sample, the readings at the initial thickness of 1.5 mm (within the tolerance given in the standard) were similar between instruments. The type M durometer results remained constant during the thickness increase, but the IRHD micro instrument showed a decrease in hardness with increasing thickness to 4.5 mm. In the case of a sample of silicone silicone, polymer in which atoms of silicon and oxygen alternate in a chain; various organic radicals, such as the methyl group, CH3, are bound to the silicon atoms. , the type M durometer results were consistently lower than the IRHD micro values, but both instruments exhibited a decrease in hardness of one unit, when doubling the thickness of the sample from 0.9 mm to 1.8 mm (within the standard).

[GRAPH OMITTED]

Effect of temperature

Raising the temperature by 10 [degrees] C appeared to make little difference to the results from the IRHD dead load on the standard test blocks (natural rubber compound). However, slightly lower values were observed on the harder samples tested on the type A durometer.

Effect of curved surfaces

The smaller diameter o-rings were laterally displaced in increments of 0.25 mm. The larger curved surfaces were displaced in increments of 0.5 mm. Graphs were plotted, showing the change in apparent hardness as the sample was displaced across the indentor. In general, the resulting curves on the graphs were flatter when testing the o-rings on a micro IRHD instrument than on the type M durometer. The type M durometer produced curves which were more peaked, with the hardness values rapidly decreasing on either side of top dead center (figure 4).

[GRAPH OMITTED]

A piece of pipe (8 mm diameter) was tested on the IRHD dead load and type A instruments; both gave fairly flat curves. An EPDM EPDM Ethylene-Propylene-Diene-Monomer
EPDM Enterprise Product Data Management
EPDM Ethylene Propylene Dimonomer (industrial/commercial piping/plumbing components)
EPDM Engineering Product Data Management
 o-ring O-ring,
n a doughnut-shaped flexible gasket made of synthetic material. Used as an overdenture attachment.
 of core diameter 7.8 mm produced a gentle curve when tested on the type A durometer, but when tested on the IRHD dead load, it was inverted inverted

reverse in position, direction or order.


inverted L block
a pattern of local filtration anesthesia commonly used in laparotomy in the ox.
.

Discussion

From the results it is clear that there is a correlation between the dead load and the micro IRHD instruments. This is apparent when the IRHD dead load result of plied micro samples corresponds with the standard result on an IRHD micro instrument. In contrast, the same cannot be said for the type A and M durometer scales.

The results indicate that the thickness of the sample used on the IRHD dead load affects the result more than on the type A durometer, in agreement with Bassi et al (ref. 3).

In general, when the nitrile, polychloroprene and silicone samples were plied, a trend of decreasing hardness with increasing thickness was observed. Some differences were noted and it appears that different rubber types influence the results in slightly different ways. More extensive work is required in this area.

Generally, the micro instruments can be used for testing both micro and macro samples, while the macro instruments (IRHD dead load and type A durometer) are better for macro samples. Indeed, many people now use the micro IRHD instead of the dead load instrument.

Flatter curves are produced with the IRHD dead load, micro and the type A durometer instruments when testing curved samples, implying that there is less critical dependence on accurate sample positioning with these. Since the graphs produced when using the type M durometer are more peaked, it is important to accurately place the sample (to within ~ 0.1 mm). However, this is controlled when using an instrument attachment to centralize cen·tral·ize  
v. cen·tral·ized, cen·tral·iz·ing, cen·tral·iz·es

v.tr.
1. To draw into or toward a center; consolidate.

2.
 o-rings, but remains important when testing curved shapes that cannot be held accurately in such an attachment.

Increasing the temperature by approximately 10 [degrees] C appears to make a greater difference to harder natural rubber samples only on the type A durometer, and little difference using the IRHD dead load instrument. This compares to a slight effect noted in the previous work (ref. 20) with both the IRHD micro and the type M durometer instruments.

From previous work, it is clear that repeated testing at the same location makes an appreciable ap·pre·cia·ble  
adj.
Possible to estimate, measure, or perceive: appreciable changes in temperature. See Synonyms at perceptible.
 difference to the results. This is more apparent when using the type M durometer. It is important to ensure that the sample is displaced between tests - this can be difficult for small samples.

It is interesting to note that the results obtained from the durometers with a dwell time of three seconds differ from those obtained using one second. This effect was demonstrated more effectively in previous work (ref. 20). Therefore, for durometers, although different timings are unimportant un·im·por·tant  
adj.
Not important; petty.



unim·portance n.
 for comparative work, it is important that the timing is accurate and repeatable. The time required (35 seconds specified by the standard) for an IRHD test places the IRHD instruments at a disadvantage. However, previous work by Lackovic et al (ref. 21) indicates that this time can be reduced by a predictive technique, taking it into direct competition with durometer timing, i.e., three seconds.

Conclusion

This article has taken a historical look at the IRHD and durometer hardness measurement instruments, as well as discussing and emphasizing the fundamental differences between the most common instruments used for rubber and elastomer elastomer (ĭlăs`təmər), substance having to some extent the elastic properties of natural rubber. The term is sometimes used technically to distinguish synthetic rubbers and rubberlike plastics from natural rubber.  hardness characterization A rather long and fancy word for analyzing a system or process and measuring its "characteristics." For example, a Web characterization would yield the number of current sites on the Web, types of sites, annual growth, etc. . As demonstrated in a previous work (ref. 20), the various instruments exhibit advantages and disadvantages with certain sample types. IRHD instruments are preferred for non-destructive testing and the micro IRHD is generally a better choice for testing curved surfaces. The type A durometer is preferable for testing non-standard thickness samples and when shorter test cycle times are required. Accurate and repeatable timing is critical to allow type A and M durometers to provide consistent and comparable results.

References

(1.) ISO 48: 1994, Physical Testing of Rubber, Methods for the Determination of Hardness.

(2.) ASTM 1415-88 (1994), Test Method for Rubber Property - International Hardness.

(3.) A.C a.c.,
adv the abbreviation for ante cibum, a Latin phrase meaning “before eating.”
. Bassi, F. Casa and R. Mendici, Polymer Testing 7, 165 (1987).

(4.) H.P. Gurney, India Rubber Journal 497 (1921).

(5.) "A new rubber hardness tester," The Rubber Age 29, 242 (1939-40).

(6.) J.R. Scott, Transactions I. R. 1. 11, 224 (1935).

(7.) J.R. Scott and R.G. Newton, Journal of Rubber Research 9, 91 (1940).

(8.) H.A. Daynes and J.R. Scott, Journal of Rubber Research 12, 94 (1943).

(9.) J.R. Scott, Journal of Rubber Research 17, 145 (1948).

(10.) R.G. Newton, Journal of Rubber Research 17, 178 (1948).

(11.) J.R. Scott, Transactions 1. R. 1.27 (5), 249 (1951).

(12.) J.R. Scott and A.L. Soden, Proceedings of the International Rubber Conference, Washington, paper 22, pp. 170-176 (1959).

(13.) K. Price, Progress of Rubber Technology 42, 59 (1979).

(14.) J.C. Warner and J.A. Jerdonek, European European

emanating from or pertaining to Europe.


European bat lyssavirus
see lyssavirus.

European beech tree
fagussylvaticus.

European blastomycosis
see cryptococcosis.
 Rubber Journal & Urethanes Today 162, II (1980).

(15.) W.V. Chang and S.C. Sun, Rubber Chemistry and Technology 64, 202 (1991).

(16) "Rubber and plastics testing," Klucklow, pp. 153-162, 1963.

(17.) "Physical testing of rubbers," J.R. Scott, pp. 91-110, 1965.

(18.) R.P. Brown and A. Soekarnein, Polymer Testing 10, 117 (1991).

(19.) B.J. Briscoe and K.S. Sebastian, Rubber Chemistry and Technology 66, 827 (1993).

(20.) R. Morgans, S. Lackovic and P. Cobbold, The International Rubber Exhibition & Conference Book of Papers, Manchester, UK, Materials Paper 12 (1999).

(21.) S. Lackovic, R. Morgans and B. McGarry, The International Conference on Rubbers, Calcutta, India, Paper AT-5 (1997).

(22.) ASTM D2240-00, Standard Test Method for Rubber Property - Durometer Hardness.
COPYRIGHT 2001 Lippincott & Peto, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2001, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Cobbold, P.
Publication:Rubber World
Date:Jan 1, 2001
Words:3962
Previous Article:Factory testing and control of raw natural rubber and mixing batches using the RPA.
Next Article:Deterioration of rubber articles in various exposures and their protection.



Related Articles
Video Showcase.
Suppliers Showcase: INSTRUMENTS.
Hardness testers. (Literature: instruments).(from H.W. Wallace and Co.)(Brief Article)
Hardness prediction.(Hardware/Software)
Hardness test blocks.(Products: instruments)
Hardness prediction.(Products: hardware/software)
Pocket durometers.(Instruments)
Materials testing equipment for rubber and plastics highlighted.(Literature: Instruments)
Rubber testing technologies.(Literature: Instruments)
Rubber testing.(Literature: Instruments)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles