The candela and photometric and radiometric measurements.


The national measurement system for photometric pho·tom·e·try  
n.
Measurement of the properties of light, especially luminous intensity.



photo·met
 and radiometric quantities is presently based upon techniques that make these quantities traceable to a high-accuracy cryogenic cryogenic /cry·o·gen·ic/ (-jen´ik) producing low temperatures.

cry·o·gen·ic
adj.
1. Relating to or producing low temperatures.

2.
 radiometer radiometer (rā'dēŏm`ətər), instrument for detection or measurement of electromagnetic radiation; the term is applied in particular to devices used to measure infrared radiation. . The redefinition Noun 1. redefinition - the act of giving a new definition; "words like `conservative' require periodic redefinition"; "she provided a redefinition of his duties"
definition - a concise explanation of the meaning of a word or phrase or symbol
 of the candela candela (kăndĕ`lə), in weights and measures: see candle.


A unit of measurement of the intensity of light. Part of the SI system of measurement, one candela (cd) is the monochromatic radiation of 540THz with a radiant intensity
 in 1979 provided the opportunity for national measurement laboratories to base their photometric measurements on optical detector technology rather than on the emission from high-temperature blackbody blackbody

Theoretical surface that absorbs all radiant energy that falls on it, and radiates electromagnetic energy at all frequencies, from radio waves to gamma rays, with an intensity distribution dependent on its temperature.
 optical sources. The ensuing en·sue  
intr.v. en·sued, en·su·ing, en·sues
1. To follow as a consequence or result. See Synonyms at follow.

2. To take place subsequently.
 technical developments of the past 20 years, including the significant improvements in cryogenic radiometer performance, have provided the opportunity to place the fundamental maintenance of photometric quantities upon absolute detector based technology as was allowed by the 1979 redefinition. Additionally, the development of improved photodetectors has had a significant impact on the methodology in most of the radiometric measurement areas. This paper will review the status of the NIST (National Institute of Standards & Technology, Washington, DC, www.nist.gov) The standards-defining agency of the U.S. government, formerly the National Bureau of Standards. It is one of three agencies that fall under the Technology Administration (www.technology.  implementation of the technical changes mandated by the 1979 redefinition of the c andela and its effect upon the maintenance and dissemination dissemination Medtalk The spread of a pernicious process–eg, CA, acute infection Oncology Metastasis, see there  of optical radiation measurements.

Key words: cryogenic radiometer; laser power; photodiodes; photometry photometry (fōtŏm`ətrē), branch of physics dealing with the measurement of the intensity of a source of light, such as an electric lamp, and with the intensity of light such a source may cast on a surface area. ; radiation temperature; radiometry Radiometry

A branch of science that deals with the measurement or detection of radiant electromagnetic energy. Radiometry is divided according to regions of the spectrum in which the same experimental techniques can be used.
; standards.

Available online: http://www.nist.gov/jres

1. Introduction

The science of performing light measurements accounting for human visual response is referred to as photometry. Photometric quantities, such as luminous flux The energy field generated by light. See lumen.
Luminous flux

The time rate of flow of light. It is radiant flux in the form of electromagnetic waves which affects the eye or, more strictly, the time rate of flow of radiant energy evaluated according to
 [International System of Units International System of Units, officially called the Système International d'Unités, or SI, system of units adopted by the 11th General Conference on Weights and Measures (1960). It is based on the metric system.  (SI) unit lumens], are designed as a realistic metric for human perception of light by appropriately accounting for the human eye's varying response as a function of wavelength. Human visual response decreases its sensitivity at short wavelengths in the blue and at long wavelengths in the red and reaches its greatest sensitivity in the green region of the spectrum. Based in part upon work by Gibson and Tyndall at NBS (National Bureau of Standards) See NIST.

NBS - National Bureau of Standards: part of the US Department of Commerce, now NIST.
 in 1923, the International Commission on Illumination The International Commission on Illumination (usually known as the CIE for its French-language name Commission internationale de l'éclairage) is the international authority on light, illumination, color, and color spaces.  (CIE (Commission Internationale de l'Eclairage, International Commission on Illumination, Vienna, Austria, www.cie.co.at) An international organization that sets standards for all aspects of lighting and illumination, including colorimetry, photometry and the measurement of visible and ) defined an average spectral response The variable output of a light-sensitive device that is based on the color of the light it perceives.  for the human eye for well-illuminated conditions, which is termed the spectral spectral /spec·tral/ (spek´tral) pertaining to a spectrum; performed by means of a spectrum.

spec·tral
adj.
Of, relating to, or produced by a spectrum.
 luminous efficiency Luminous efficiency

Visual efficacy of visible radiation, a function of the spectral distribution of the source radiation in accordance with the “spectral luminous efficiency curve,’’ usually for the light-adapted eye or photopic vision,
 for photopic vision photopic vision
n.
See photopia.
. This spectral response is usually denoted as V([lambda]) in recognition of the fact that this distribution was originally called the "visibility curve" (1, 2). This international standard for spect ral luminous efficiency is shown in Fig. 1. Inspection of this distribution indicates that the human visual response peaks at about 555 nm and decreases to zero in the ultraviolet (UV) and the infrared (IR) wavelength regions. By definition the response has been set to unity at the peak of the response at 555 nm. Different response functions are used as standards for the spectral luminous efficiency for lesser illuminations, which are called mesopic vision Mesopic vision is a combination of photopic vision and scotopic vision in low but not quite dark lighting situations.

The combination of the higher total sensitivity of the rods in the eye for the blue range with the color perception through the cones results in a very
 for intermediate levels of illumination and scotopic vision scotopic vision
n.
Vision that occurs when the eye is dark-adapted. Also called scotopia.
 for very low levels of illumination. These matters are not central to the discussion here and hence will not be expanded upon. The interested reader is referred to the literature for the details and applications of these definitions (3, 4).

Prior to 1948, there was no widely agreed upon Adj. 1. agreed upon - constituted or contracted by stipulation or agreement; "stipulatory obligations"
stipulatory

noncontroversial, uncontroversial - not likely to arouse controversy
 international standard for luminous intensity Luminous intensity

The solid angular luminous flux density in a given direction from a light source. It may be considered as the luminous flux on a small surface normal to the given direction, divided by the solid angle (in steradians) which the surface
 (roughly, the perceived brightness of a light source). The international units international units,
n.pl a unit of measurement that evaluates the potency of a substance. Because it measures potency instead of quantity, there is a different international unit-to-mg conversion ratio for each particular substance.
 for luminous intensity and other photometric quantities were maintained by comparisons of artifacts artifacts

see specimen artifacts.
, including candles, lamps, and other types of light sources, none of which were based on fundamental physical laws. In 1948, the General Conference on Weights and Measures The General Conference on Weights and Measures is the English name of the Conférence générale des poids et mesures (CGPM, never GCWM). It is one of the three organizations established to maintain the International System of Units (SI) under the terms of the Convention  (GCPM) adopted a definition of the unit of luminous intensity, called the "new candle," derived from the radiation properties of platinum at the temperature where it froze froze  
v.
Past tense of freeze.


froze
Verb

the past tense of freeze

froze, frozen freeze
 from a liquid to a solid. This definition, for the first time, based photometry on a repeatable physical process or phenomena that could be independently recreated and observed anywhere in the world. The 1967 CGPM CGPM Conférence Générale des Poids et Mesures (French: General Conference on Weights and Measures)  changed the name of the unit of measurement for luminous intensity from "new candle" to "candela" but did not change its physical definition, hence the following definition was used for the can dela from 1948 to 1979 (5):

The candela is the luminous intensity, in the perpendicular direction, of a surface of 1/600000 square meter Noun 1. square meter - a centare is 1/100th of an are
centare, square metre

area unit, square measure - a system of units used to measure areas
 of a blackbody at the temperature of freezing platinum under a pressure of 101325 new-tons per square meter. (13th CGPM, 1967, Resolution 5).

Most national laboratories found it difficult and expensive to maintain a platinum freezing-point blackbody and a consensus developed to change the definition of the candela to one based solely upon the measurement of optical power (6). In 1979 the CGPM adopted a recommendation from the International Committee on Weights and Measures weights and measures, units and standards for expressing the amount of some quantity, such as length, capacity, or weight; the science of measurement standards and methods is known as metrology.  (CIPM CIPM Comité International des Poids et Mesures (International Committee of Weights and Measures)
CIPM Center for Integrated Pest Management
CIPM Certificate in Investment Performance Measurement
) and redefined the candela in terms of a specific amount of optical power:

The candela is the luminous intensity, in a given direction, of a source that emits monochromatic monochromatic /mono·chro·mat·ic/ (-kro-mat´ik)
1. existing in or having only one color.

2. pertaining to or affected by monochromatic vision.

3. staining with only one dye at a time.
 radiation of frequency 540 X [10.sup.12] hertz hertz (hûrts) [for Heinrich R. Hertz], abbr. Hz, unit of frequency, equal to 1 cycle per second. The term is combined with metric prefixes to denote multiple units such as the kilohertz (1,000 Hz), megahertz (1,000,000 Hz), and gigahertz  and that has a radiant intensity In radiometry, radiant intensity is a measure of the intensity of electromagnetic radiation. It is defined as power per unit solid angle. The SI unit of radiant intensity is watts per steradian (W·sr-1).  in that direction of 1/683 watts per steradian ste·ra·di·an  
n. Abbr. sr
A unit of measure equal to the solid angle subtended at the center of a sphere by an area on the surface of the sphere that is equal to the radius squared:
. (16th CPGM, 1979, Resolution 3)

The specified frequency was chosen to be at the peak of the V([lambda]) distribution (555 nm) and the factor 1/683 watts per steradian was chosen based upon the best estimate at that time of the freezing point freezing point

Temperature at which a liquid becomes a solid. When the pressure surrounding the liquid is increased, the freezing point is raised. The addition of some solids can lower the freezing point of a liquid, a principle used when salt is applied to melt ice on
 of platinum. These choices allowed continuity between the new definition and the old one. This definition also decoupled the photometric units from the radiation properties of a source and allowed the photometric units to be determined by the measurement of optical power with appropriate photodetectors.

It might appear that the definition of the candela only pertains to light at a wavelength near 555 nm and thus a monochromatic light source must be used to determine the candela. The central point of photometry, however, is to measure all light according to according to
prep.
1. As stated or indicated by; on the authority of: according to historians.

2. In keeping with: according to instructions.

3.
 human visual perception. To that end, the CIPM adopted the CIE definition for spectral luminous efficiency, V([lambda]) to determine luminous intensity at other wavelengths on a relative basis.

This extension of the definition also provides a practical prescription for measuring photometric units by using optically filtered photodetector A device that senses light. It uses the principle of photoconductivity, which is exhibited in certain materials that change their electrical conductivity when exposed to light. See photoelectric, photocell and photodiode.  systems. These photodetectors are constructed to have a wavelength-dependent response that is closely proportional to the V([lambda]) distribution. In Sec. 2 of this paper we will discuss the foundation of detector-based metrology, and in Sec. 3 we will show how filtered photometers provide methods for determining photometric quantities according to the candela definition within SI.

Table 1 provides a listing of photometric quantities and their radiometric counterparts. For historical reasons the fundamental photometric quantity in SI has been luminous intensity which, although it is a property of a light source, can be related to detector quantities by measuring the luminous flux of a source under conditions of known geometry. Luminous flux, measured in lumens, is a more commonly used quantity in photometry. In illumination applications, where a lamp fixture is present that might include reflectors and diffusers, the luminous flux of all the light emitted by a lamp in all directions is a key measure of the lamp's performance. The number of lumens produced by a lamp proportioned to its electrical power consumed is a figure of merit Noun 1. figure of merit - a numerical expression representing the efficiency of a given system, material, or procedure
efficiency - the ratio of the output to the input of any system
 for the lamp's efficiency.

2. Absolute Detector Radiometry

The 1979 redefinition of the candela was a major driving force in radiometry and photometry that spurred the need for, and subsequent development of, improved optical radiation detectors in the visible wavelength region. While light sensors such as photomultiplier tubes A vacuum tube that converts light into electrical energy and amplifies it. Photomultiplier tubes are used in high-end drum scanners, because they are more sensitive to light than the CCD elements used in lower-cost devices.  and cadmium sulfide Noun 1. cadmium sulfide - a yellow sulfide used chiefly as a pigment
sulfide, sulphide - a compound of sulphur and some other element that is more electropositive

zinc cadmium sulfide - cadmium sulfide containing zinc
 devices were well known, such technologies were not suitable as fundamental standards. Fortunately, at the time the redefinition was formulated, technology had progressed. Newer types of photodetectors allowed photometry and radiometry to become detector-based, rather than based on traditional thermal source techniques.

By the 1970s, silicon photodiodes had become available with unprecedented stability (7). Advances in semiconductor technology provided a wealth of new types of solid-state photodetectors, both from silicon and other materials. Solid-state devices solid-state device

Electronic device that operates on the basis of the electric, magnetic, or optical properties of a solid material, especially one that uses a solid crystal in which an orderly three-dimensional arrangement of atoms, ions, or molecules is repeated
 are particularly attractive as sensors since they can directly supply an electrical signal, usually current, which can be proportional to the input optical signal over many orders of magnitude. Their internal impedance impedance, in electricity, measure in ohms of the degree to which an electric circuit resists the flow of electric current when a voltage is impressed across its terminals.  and other electronic characteristics lend themselves to easy mating with modern solid-state electronics and thereby provide a high-quality and inexpensive sensor system for many applications of optical radiation measurement.

At about the same time silicon devices were being perfected for use in radiometry, better electrical-substitution radiometers were also developed. These included cryogenic devices operating near liquid-helium temperatures with a relative combined standard uncertainty of 0.01 % or better. Electrical-substitution radiometers are constructed by devising an absorbing receiver that collects optical power and as a result undergoes a temperature rise. The optical power is determined by using electrical power to produce the same temperature rise or, in most practical implementations, to maintain a steady-state temperature as the optical power load varies. The equality of the electrical and optical power is implied because of their equivalent thermal effect on the system.

Electrical substitution devices are often referred to as absolute detectors because they determine the radiant flux radiant flux
n.
The rate of flow of radiant energy.

Noun 1. radiant flux - the rate of flow of radiant energy (electromagnetic waves)
flux - the rate of flow of energy or particles across a given surface
 incident upon them by direct reference to physical laws and do not depend upon another optical power-measuring device for their calibration. The presumption of the equivalence of electrical power and optical power heating of a system is verified by careful characterization of the radiometer. Similarly, in the case of silicon, the knowledge of the semiconductor physics is thought to be adequate for describing the internal quantum efficiency and hence the response of the device to optical radiation.

The measurement of high power and pulsed lasers poses challenging technical problems that have led to the development of specialized detectors, including electrically calibrated cal·i·brate  
tr.v. cal·i·brat·ed, cal·i·brat·ing, cal·i·brates
1. To check, adjust, or determine by comparison with a standard (the graduations of a quantitative measuring instrument):
 ones. We will review the NIST program in laser power and energy measurements and its detector technology in Sec. 5.

The national laboratories check the veracity veracity (vras´itē),
n
 of their absolute optical power measuring instruments through a continual series of international intercomparisons and other activities designed to establish equivalence of techniques. While a device such as a silicon photodiode A light sensor (photodetector) that allows current to flow in one direction from one side to the other when it absorbs photons (light). The more light, the more the current. Used to detect light pulses in optical fibers and other light-sensitive applications, it works the opposite of a  can be considered an absolute detector under specified conditions, the actual accomplishment of the task may require considerable expertise and careful attention to specified procedures. Checking the procedures and practices used in the realizations of absolute optical power measurement with a particular detector system is the main impetus for the intercomparisons carried out by the national laboratories.

Due to their importance in modern radiometry and photometry, we will briefly review the use of silicon photodetectors and cryogenic radiometers while leaving discussion of the many other detectors to the literature references cited in this paper.

2.1 Silicon Photodiodes

When photons are absorbed by silicon they create pairs of electrons and holes. When this occurs in the junction region of a silicon diode, an electrical current is generated which is proportional to the number of photons absorbed. The energy of the photon must exceed the band-gap energy of silicon, which is about 1.12 eV and which corresponds to light with wavelengths less than about 1200 nm. Silicon's band-gap allows it to be used as a photodetector from the near infrared into the soft x-ray region (8, 9). Other semiconductors with lower band-gap energies, such as germanium germanium (jərmā`nēəm) [from Germany], semimetallic chemical element; symbol Ge; at. no. 32; at. wt. 72.59; m.p. 937.4°C;; b.p. 2,830°C;; sp. gr. 5.323 at 25°C;; valence +2 or +4.  and indium-gallium arsenide ar·se·nide  
n.
A compound of arsenic with a more electropositive element.

Noun 1. arsenide - a compound of arsenic with a more positive element
, can be used for detectors further into the JR region.

The responsivity of a silicon photodiode is its output electrical signal per unit of input optical power. The electrical signal is usually the photocurrent pho·to·cur·rent  
n.
An electric current produced by illumination of a photoelectric material.
 measured in amperes and the optical power is measured in watts; hence, responsivity is typically expressed in units of A/W A/W Awaiting
A/W Alloy Wheels
A/W A-Wing (Star Wars)
A/W Actual Weight
A/W Alternate Weeks
A/W Automatic Win (gaming) 
. When this quantity is measured as a function of wavelength, it is referred to as the spectral responsivity of the detector. In most cases the silicon photodiode is coupled to a current-to-voltage amplifier that provides an output voltage that is easily measured with widely available high-quality voltmeters. (For other types of photodetectors and bolometers, the transducer transducer, device that accepts an input of energy in one form and produces an output of energy in some other form, with a known, fixed relationship between the input and output.  output may be more conveniently expressed in terms of other parameters that match their underlying physics.)

Top quality silicon photodiodes have proven to be generally stable, uniform, and sensitive. In addition, Geist and Zalewski showed that in the visible and near IR wavelength regions the absolute response of certain silicon detectors could be determined by a procedure that they called "self-calibration" (10, 11). This procedure relies upon the fact that the internal quantum efficiency of silicon is very close to unity over the wavelength region of 500 nm to 950 nm and that the reflectance re·flec·tance  
n.
The ratio of the total amount of radiation, as of light, reflected by a surface to the total amount of radiation incident on the surface.

Noun 1.
 of the diodes can be accurately measured using ordinary optical techniques. The internal quantum efficiency is the ratio of the number of electron-hole pairs created per absorbed photon. By accounting for the reflected light that is not absorbed, the number of photons incident on the silicon detector can be determined by measuring the electrical current produced. This procedure can, with some care, provide absolute calibrations of silicon detectors used to measure optical power, with a relative standard uncertainty of less th an 0.05 % (10).

The reflectance of a silicon photodiode can change as its surface changes, because of additional oxidation oxidation /ox·i·da·tion/ (ok?si-da´shun) the act of oxidizing or state of being oxidized.ox·idative

ox·i·da·tion
n.
1. The combination of a substance with oxygen.

2.
, contamination, or even humidity changes in ordinary laboratory atmospheres. Nevertheless, Zalewski and Duda showed how these effects can be minimized by using multiple photodiodes in a "trap" configuration (12). In a trap configuration, a collimated beam See collimated.  of light reflects off of one photodiode onto another, and then another, until a sufficient number of photodiodes absorb substantially all of the light.

A trap detector is so-named because almost none of the incident light escapes. Most importantly Adv. 1. most importantly - above and beyond all other consideration; "above all, you must be independent"
above all, most especially
, small changes in the reflectance of an individual surface do not significantly affect the total absorption of the whole device. The light flux not absorbed [[PHI phi
n.
Symbol The 21st letter of the Greek alphabet.


PHI,
n See health information, protected.
].sub.na] is related to the reflectance as shown in Eq. (1):

[[PHI].sub.na] [approximately equal to] [[PHI].sub.0] [[rho].sup.n], (1)

where [rho] is the reflectance, n is the number of reflections, and [[PHI].sub.0] is the incident flux. Since [rho] is typically about 0.2, [[PHI].sub.na] is a small fraction of the incident flux for n [greater than or equal to] 3. Typically, over 99.9 % of the light is absorbed and converted to electrical signal.

Since the internal quantum efficiency is near unity for a substantial wavelength region, trap detectors can be used as absolute standards based upon their known physical properties. In many other cases the inherent stability of trap detectors leads to their use as transfer standards. These are first calibrated against another radiometer, typically a cryogenic radiometer.

A schematic A graphical representation of a system. It often refers to electronic circuits on a printed circuit board or in an integrated circuit (chip). See logic gate and HDL.  diagram of a photodetector system used at NIST is shown in Fig. 2. The device is modular and constructed in four stages. This allows us to assemble a detector system including only those stages required for an application.

In many applications a quantity like irradiance ir·ra·di·ant  
adj.
Sending forth radiant light.



[Latin irradi
 or illuminance illuminance: see photometry.
Illuminance

A term expressing the density of luminous flux incident on a surface. This word has been proposed by the Colorimetry Committee of the Optical Society of America to replace the term illumination.
 needs to be measured, and hence a precision aperture An orifice. It often refers to an opening in which light is allowed to pass in optical systems such as cameras and lasers. See f-stop and numerical aperture.  is required to define the acceptance area of the detector system. In radiance or luminance The amount of brightness, measured in lumens, that is given off by a pixel or area on a screen. For example, dark red and bright red would have the same chrominance, but a different luminance.  measurements it is sometimes necessary to define the configuration factor of a system. This also requires a precision aperture in order to account properly for the flux transferred from one region to another. The design shown in Fig. 2 allows for the attachment of apertures for these purposes.

When a filter is used, as shown in Fig. 2, the system is referred to as a filter radiometer (FR). Filters are important when the application requires that the spectral responsivity of the detector be different than that of native silicon. This is the case in photometry where it is necessary to filter the light such that the response of the photodetector system is closely proportional to the V([lambda]) function. In other applications, narrow band filters may be necessary to radiometrically determine the temperature of a source or to simply monitor a process with optical emission at some particular wavelength. This is often the case in remote sensing applications A remote sensing application is a software application that processes remote sensing data. Remote sensing applications are similar to graphics software, but they enable generating geographic information from satellite and airborne sensor data. . For stability it is often necessary to temperature control the filter with thermoelectric ther·mo·e·lec·tric   also ther·mo·e·lec·tri·cal
adj.
Characteristic of, resulting from, or using electrical phenomena occurring in conjunction with a flow of heat.
 coolers or other means to avoid wavelength shifts. In many applications unwanted infrared radiation can be blocked with appropriate absorbing materials. However, in a few applications such as monitoring the output of a monochromator A monochromator is an optical device that transmits a mechanically selectable narrow band of wavelengths of light or other radiation chosen from a wider range of wavelengths available at the input. , no filter is used. In this c ase the spectral responsivity of the detector module is first calibrated and then used to determine the spectral output of the monochromator. Similarly, a calibrated detector can be used for directly measuring the power output from a single frequency laser.

Figure 2 shows a detector system composed of six silicon photodiodes arranged in trap configuration. In this configuration, the collimated light Collimated light is light whose rays are parallel and thus has a planar wavefront. The word is derived from "co-linear" and implies light that does not disperse, even over an infinite distance.  beam entering from the right would make six reflections and be mostly absorbed with only a small residual amount exiting the sensor system after the last reflection [Eq. (1)]. This transmission-trap design allows the untrapped optical flux to be measured, and hence small corrections to the responsivity can be made for extremely accurate measurements.

Instead of a trap detector system, a single silicon detector is often used for simplicity of construction and use. Since filters and other absorbers must be calibrated for transmittance in order to determine the spectral responsivity of the system, when they are used the advantage of a trap detector as an absolute standard is lost. (While one could measure the filter transmittance separately from a well-characterized detector, it is most often expedient ex·pe·di·ent  
adj.
1. Appropriate to a purpose.

2.
a. Serving to promote one's interest: was merciful only when mercy was expedient.

b.
 to calibrate To adjust or bring into balance. Scanners, CRTs and similar peripherals may require periodic adjustment. Unlike digital devices, the electronic components within these analog devices may change from their original specification. See color calibration and tweak.  the entire system together.)

To enhance the stability of the system, the photodetectors can be temperature controlled with a thermoelectric cooler or other suitable temperature-control system. In some applications it is sufficient to use a regulated heater to maintain the detector at a constant temperature above ambient (13).

Silicon photodiodes are available with shunt To divert, switch or bypass.  resistances of greater than 1 G[OMEGA 1. (programming) Omega - A prototype-based object-oriented language from Austria.

["Type-Safe Object-Oriented Programming with Prototypes - The Concept of Omega", G. Blaschek, Structured Programming 12:217-225, 1991].
2.
]. This large internal resistance allows the construction of high-gain current-to voltage converters A voltage converter changes the voltage of an electrical power source and is usually combined with other components to create a power supply. The term "voltage converter" is sometimes used as a generic term for a power supply. . Signals from silicon photodiode can have electrical noise floors as low as 0.1 fA (14). Silicon photodiodes can also have a linear response over many orders of magnitude of input optical power (15). This stability, linearity, and ease of use have made the silicon photodetector the essential ingredient in contemporary photometry and in many aspects of spectral radiometry. The calibration of filter detector systems an their use will be discussed in later sections of this paper.

2.2 Electrical Substitution Radiometry

Electrical substitution radiometers (ESRs) are instruments that measure optical power by equating e·quate  
v. e·quat·ed, e·quat·ing, e·quates

v.tr.
1. To make equal or equivalent.

2. To reduce to a standard or an average; equalize.

3.
 it to a equivalent amount of electrical power. ESRs are some times referred to as electrically calibrated radiometer (ECRs) or simply as absolute radiometers.

The fundamentals of an ESR ESR - Eric S. Raymond  can be understood b reference to Fig. 3. The radiant flux (optical power) [PHI] impinges upon a receiving cavity that is designed to collect radiation optimally. Upon absorption of the power, the cavity will experience a temperature rise. The receiver cavity is coupled through a thermal conductor of conductance G to a constant temperature heat sink A material that absorbs heat. Typically made of aluminum, heat sinks are widely used in amplifiers and other electronic devices that build up heat. Small heat sinks are the most economical method for cooling microprocessors and other chips.  maintained at a reference temperature [T.sub.o]. Ignoring losses due to radiation, convection, and stray thermal conductance thermal conductance

A measure of the ability of a material to transfer heat per unit time, given one unit area of the material and a temperature gradient through the thickness of the material. It is measured in watts per meter per degree Kelvin.
, the equilibrium (long-time) temperature rise is given by T - [T.sub.o] = [PHI]/G. The same amount of power supplied by the electrical heater would have the same effect.

In practice, ESR performance can often be improved by maintaining the receiving cavity at a constant temperature with the electrical heater. When a shutter (1) An opaque window that is moved in one direction to let light in and in another to close off the light. In fixed-lens cameras, one shutter often suffices for aperture and speed.  is imposed to stop the light beam, power is supplied to the heater to create a set temperature rise. When the shutter is opened, a feedback loop decreases the electrical power a sufficient amount to maintain the same cavity temperature. Ignoring corrections and losses mentioned earlier, the optical power is given by [PHI] = ([i.sub.cl.sup.2] - [i.sub.op.sup.2]) R, where [i.sub.cl] is the current applied through the heater of resistance R when the shutter is closed and [i.sub.op] is the current when the shutter is opened.

While we have used a cavity absorber for this example, the idea of electrical substitution is used to calibrate other optical detectors Optical detectors

Devices that respond to incident ultraviolet, visible, or infrared electromagnetic radiation by giving rise to an output signal, usually electrical. Based upon the manner of their interaction with radiation, they fall into three categories.
 such as bolometers and pyroelectric py·ro·e·lec·tric  
adj.
Relating to or exhibiting pyroelectricity.

n.
A pyroelectric material.

Adj. 1. pyroelectric - relating to or exhibiting pyroelectricity
pyroelectrical
 devices. These are often used for laser power and energy measurements (discussed in Sec. 5). The practical challenge when utilizing ESRs is to carefully characterize the various loss mechanisms in order to apply appropriate corrections to the power equivalence relationship.

ESRs have been in use for 75 years or more, and their history and development have been described by Hengstberger in considerable detail (16). ESR technology ESR Technology was formerly the Engineering, Safety and Risk business of AEA Technology, which was formed from the commercial arm of the United Kingdom Atomic Energy Authority. It specialises in the provision of independent technical expertise, products and services to help customers:  was pioneered by Coblentz at NBS during the early part of this century (17). He developed a number of radiometers and used them for diverse purposes in photometry and radiometry, including an early measurement of the Stefan-Boltzmann constant (18). The limited wavelength range and difficulty in obtaining consistently high accuracy with solid-state devices spurred the development of much-improved ESRs during the 1970s and 1980s.

In this time period, a significant innovation was the design and construction of an ESR that operated at cryogenic temperatures. This resulted in an increased sensitivity and a reduction in the uncertainties due to radiative and convective losses. The first cryogenic radiometer at NBS was constructed by Ginnings and Reilly in 1972 to measure thermodynamic ther·mo·dy·nam·ic
adj.
1. Characteristic of or resulting from the conversion of heat into other forms of energy.

2. Of or relating to thermodynamics.
 temperatures above 0 [degrees]C (19). For a variety of reasons this project did not achieve the desired results, but by building on the experience gained by Ginnings and Reilly, Yokley built a cryogenic radiometer system at NBS in the mid-1970s to measure the radiation temperature of low-temperature blackbodies used in a low-background environment (20, 21). This device was used for a number of years to perform specialized calibrations of low-flux sources, but it was not engineered to perform high-accuracy measurements. It yielded measurements with relative uncertainties of several percent.

Quinn and Martin at the National Physical Laboratory (NPL 1. NPL - New Programming Language. IBM's original (temporary) name for PL/I, changed due to conflict with England's "National Physical Laboratory." MPL and MPPL were considered before settling on PL/I. Sammet 1969, p.542.
2.
) in the UK developed a high-accuracy cryogenic radiometer for use in a radiometric determination of the Stefan-Boltzmann constant with a relative combined standard uncertainty of 0.013 % (22). In order to achieve this low uncertainty, it was necessary to measure radiometric flux with this same uncertainty. This important benchmark work contains a detailed analysis of the errors and uncertainties associated with a cryogenic ESR and has led to the world-wide adoption of these devices as fundamental radiometric standards with relative combined standard uncertainties of less than 0.01 %. NPL staff later developed this device into a radiometer for laser power measurements and thereby provided a technique to establish high accuracy calibrations of other stable detectors by comparison to the cryogenic radiometer (23). The commercial availability and intrinsic accuracy of these instruments has resulted in their employment by a number of national standards labo ratories to provide the technical basis of radiometric measurement (24).

The NIST High-Accuracy Cryogenic Radiometer (HACR HACR Hispanic Association on Corporate Responsibility
HACR Heating, Air Conditioning & Refrigeration
HACR High Accuracy Cryogenic Radiometer
HACR Helicopter Active Control Rotor
) is shown in Fig. 4 and is described in detail in the technical literature (25). The radiometer system is contained in a large stainless steel stainless steel: see steel.
stainless steel

Any of a family of alloy steels usually containing 10–30% chromium. The presence of chromium, together with low carbon content, gives remarkable resistance to corrosion and heat.
 vacuum vessel that contains reservoirs for cryogenic fluids. The outer cryogenic shield is maintained at 77 K with liquid nitrogen Noun 1. liquid nitrogen - nitrogen in a liquid state
atomic number 7, N, nitrogen - a common nonmetallic element that is normally a colorless odorless tasteless inert diatomic gas; constitutes 78 percent of the atmosphere by volume; a constituent of all living
, and the inner reservoir is filled with liquid helium Liquid helium  at 4.2 K. An intermediate shield stabilizes at an operating temperature of about 50 K. The heart of the instrument is the absorbing cavity that is connected to a thermally controlled heat sink held at 5 K. An enlargement enlargement,
n an increase in size.

enlargement, Dilantin,
n.pr See hyperplasia, gingival, Dilantin.

enlargement, idiopathic,
n
 of the cavity is shown in Fig. 5. Polarized A one-way direction of a signal or the molecules within a material pointing in one direction.  optical radiation from a laser system enters the vacuum vessel of the HACR through a window at Brewster's angle Brewster's angle (also known as the polarization angle) is an optical phenomenon named after the Scottish physicist, Sir David Brewster (1781–1868).

When light moves between two media of differing refractive index, generally some of it is reflected at the
. After passing through the window the light is incident upon the absorbing cavity, which is designed so that the optical radiation undergoes many reflections and is effectively trapped in the cavity. The cavity is coated with specular spec·u·lar  
adj.
Of, resembling, or produced by a mirror or speculum.



specu·lar·ly adv.

Adj. 1.
 black paint to improve the absorpt ance. The substantial distance the light must travel in the HACR before being absorbed requires that the light be highly collimated In a straight line. Collimated light beams are parallel rays of light. . This is most often accomplished by using laser radiation. Laser radiation also provides higher power Higher power is a term used in a 12-step program, such as Alcoholics Anonymous, to describe "a power greater than yourself." Although many participants equate their higher power with God, a belief in God or in formal religion is not mandatory; the higher power is intended as a  levels, typically milliwatts of power, than are commonly available with a monochromator system. This level of power is required by the first generation of cryogenic radiometers. Newer systems are now available which can be used successfully with radiation from a monochromator and conventional light sources.

The reference block, (c) in Fig. 5, is actively maintained at a constant temperature and acts as a heat sink for the cavity. The temperature of the present HACR reference block is maintained to within 0.15 mK over a typical measurement time of 45 mm. The temperature is sensed by germanium resistance thermometers resistance thermometer
n.
A device measuring temperature by the change of the electrical resistance of a metal wire.
, (d) in the figure. There is a thermal link (b) from the reference block to the helium helium (hē`lēəm), gaseous chemical element; symbol He; at. no. 2; at. wt. 4.0026; m.p. below −272°C; at 26 atmospheres pressure; b.p. −268.934°C; at 1 atmosphere pressure; density 0.  cryostat cryostat /cryo·stat/ (kri´o-stat)
1. a device by which temperature can be maintained at a very low level.

2. in pathology and histology, a chamber containing a microtome for sectioning frozen tissue.
 (a) and a second one (f) from the cavity to the reference block. There are two different heater arrangements on the cavity, one of them (in) is at the approximate position where the laser beam is incident and the second (1) is wrapped around the outer portion of the cavity. Both can be used for the electrical substitution equivalence measurement and their separate positions allow checking the effectiveness of the heating system for equivalence of optical and electrical power absorptance ab·sorp·tance  
n.
The ratio of absorbed to incident radiation.



[absorpt(ion) + -ance.]

Noun 1.
 by the cavity. This instrument achieves a measurement of optical power to within a 0.02 1 % relative combined standard un certainty (25, 26) for optical powers on the order of 1 mW. Because of this impressive accuracy, the HACR is a national standard for many optical power measurements in the visible wavelength region.

A typical mode of operation for the HACR is shown schematically sche·mat·ic  
adj.
Of, relating to, or in the form of a scheme or diagram.

n.
A structural or procedural diagram, especially of an electrical or mechanical system.
 in Fig. 6 and is described in more detail in the literature (25, 27). A secondary or transfer standard detector (TSD TSD Tay-Sachs disease. ) is inserted into the laser beam and intercepts the same beam that is measured by the HACR. After corrections for the HACR entrance window transmittance and other systematic effects, the known optical power of the beam is used in the computation of the responsivity of the TSD. (A TSD is used to calibrate other instruments and devices, as discussed in Sec. 2.3.)

A complete calibration for a TSD consists of determining the absolute spectral responsivity (ASR (Automatic Speech Recognition) Using voice recognition to replace keypad entry for telephone voice menus. Typically used to speak the digits 0 through 9 insted of keying them, ASR systems may be able to recognize a limited vocabulary. See voice recognition and AVSR. ) of the device, using both the HACR measurements and appropriate modeling. Here, the term "absolute" is used indicate that the spectral responsivity scale is derive from the national standards. This distinguishes these measurements from the more common situation where a relative responsivity function is shown to merely co pare the performance of a detector at different way lengths.

The ASR of a TSD can be calibrated directly using the HACR at selected laser wavelengths, and then t responsivity for regions between the laser wavelengt s can be interpolated interpolated /in·ter·po·lat·ed/ (in-ter´po-la?ted) inserted between other elements or parts.  based upon knowledge of t physics of the TSD. In the case of silicon photodiode a considerable amount of effort has gone into modeling their responsivity as a function of wavelength and hen the spectral responsivity from 400 nm to 1000 nm ca be accurately deduced (11, 28). Gentile and co-workers used this technique to develop the NIST scale of detect r spectral responsivity with a relative combined standard uncertainty of less than 0.04 % in the wavelength ran e of 406 nm to 920 nm.

Other types of detectors can be calibrated using this substitution technique in different wavelength region, including spectrally spec·tral  
adj.
1. Of or resembling a specter; ghostly.

2. Of, relating to, or produced by a spectrum.



spec·tral
 flat absorptive bolometer bolometer (bōlŏm`ətər, bə–), instrument for detecting and measuring radiation, e.g., visible light, infrared radiation, and ultraviolet radiation, in amounts as small as one millionth of an erg.  detectors and other semiconductor devices (7). A bolometer can be utilized as a TSD by carefully characterizing its stability and spectral absorption properties. The bolometer can be calibrated at selected wavelengths with the HACR and its overall spectral response characteristics inferred from knowledge of its relative spectral absorptance.

In some wavelength regions lasers may not be readily available for use in the manner shown in Fig. 6. So, cryogenic radiometers have been developed that operate with monochromatized light provided by conventional optical sources, which produce a continuous wavelength coverage for calibration (29). These monochromator-based sources generally have a more diverging di·verge  
v. di·verged, di·verg·ing, di·verg·es

v.intr.
1. To go or extend in different directions from a common point; branch out.

2. To differ, as in opinion or manner.

3.
 beam and less optical power per wavelength interval than do lasers. Their use with a cryogenic radiometer to characterize a TSD can, therefore, require greater care to achieve equivalent levels of uncertainty. Recently, cryogenic radiometers have been installed at synchrotron radiation facilities A list of storage rings and free electron lasers used as synchrotron radiation sources by country. Armenia
  • Center for the Advancement of Natural Discoveries using Light Emission (CANDLE), Yerevan (proposed) http://www.candle.am/index.
. With appropriate monochromators they afford accurate measurement over a wavelength region from the far UV to the IR spectral regions (30, 31).

The use of a cryogenic radiometer usually relies upon the availability of suitable window materials, which presents technical challenges in some wavelength regions. However, in a vacuum environment, such as with a synchrotron synchrotron: see particle accelerator.
synchrotron

Cyclic particle accelerator in which the particle is confined to its orbit by a magnetic field. The strength of the magnetic field increases as the particle's momentum increases.
 source, the necessity of using a window is avoided. While window corrections no longer contribute to the overall measurement uncertainty, other problems associated with ensuring that a particular TSD operates equivalently in a vacuum and in the air can pose a difficulty.

2.3 Detector Spectral Responsivity Measurements

The ultimate goal of detector-based radiometry is to determine the spectral responsivity of any photodetector system in terms of the spectral responsivity of an absolute national standard, its ASR. One approach might be to calibrate many detectors against a cryogenic radiometer standard directly, presuming pre·sum·ing  
adj.
Having or showing excessive and arrogant self-confidence; presumptuous.



pre·suming·ly adv.
 that both the cryogenic radiometer and the detectors under test could utilize the same light source. However it is usually preferable and more expedient, because of the complexity of a cryogenic radiometer, to make the comparison in two steps--first from the cryogenic radiometer to a TSD, and then from the TSD to the detector of interest. While the accuracy of a TSD-based calibration may be somewhat diminished, the ease of use and the possible avoidance of maintaining a cryogenic system make this approach useful in applications where the utmost calibration accuracy is not required. One also gains latitude to use a light source more appropriate for the absolute standard in the first step (e.g., a fixed-wave length laser), and another source more appropriate for the detector under test in the second step (e.g., a tunable monochromator).

For these reasons, and to provide calibrations for a wide range of customers and uses, NIST maintains a set of instruments that utilize monochromators and which are known as detector spectral comparators (DSCs) (32, 33). NIST has DSCs engineered for the UV, visible, near IR, and far IR spectral regions in addition to a facility at the NIST synchrotron that also can provide calibrations over a broad wavelength region.

Figure 7a shows a schematic outline of the NIST visible-near-IR DSC (1) (Digital Signal Controller) A microcontroller and DSP combined on the same chip. It adds the interrupt-driven capabilities normally associated with a microcontroller to a DSP, which typically functions as a continuous process. See microcontroller and DSP.  measurement facility. The essential features are a set of light sources chosen to cover the various wavelength regions, input optics, a double monochromator, and exit optics to refocus Verb 1. refocus - focus once again; The physicist refocused the light beam"
focus - cause to converge on or toward a central point; "Focus the light on this image"

2.
 the light on the detectors being calibrated or on a TSD being used as the calibration standard. In practice the TSD is calibrated using the HACR, mounted on a moveable carriage, and then placed in position to receive light from the monochromator. Its signal is determined at a particular wavelength, and then the test detector is moved to the same position. A monitor system ensures the stability of the monochromator output during the two measurements. Using this procedure, the ASR of the test detector is established in terms of the TSD, and hence in terms of the HACR. Typical results are shown in Fig. 7b.

Detector systems often have aperture areas of about 1 [cm.sup.2] or greater, and the ASR measurements are usually performed by using the focused beam from the monochromator, usually a few [mm.sup.2]. Hence, the entrance aperture of a detector is often underfilled. However, the carriage provides vertical and horizontal motion normal to the exit beam of the monochromator and allows for the mapping of the spatial uniformity of the detector's ASR.

Figure 8 shows typical spatial uniformity data for a silicon detector where the uniformity of response is within 1 % over the entire active area. This uniformity is usually a function of wavelength and hence a full characterization of a detector will include a study of the uniformity over the wavelength regions where it might be used. These measurements are important as not all detectors are this uniform and, depending on the active area used in a particular application, an appropriate average responsivity must be calculated.

This averaging is necessary to determine the wideaperture response for irradiance or illuminance measurements where the detector aperture is overfilled overfilled,
adj See overextended.
. Examples of this type of measurement are the determination of the illuminance from a lamp or the spectral irradiance of a source like the Sun. However, a lack of uniformity and the averaging process give rise to increased uncertainties in the ASR of detectors used in these applications.

Instead of calibrating the instrument by determining the responsivities of the different regions of the active area, and averaging them, a better strategy is to calibrate the instrument with a uniform field of known illuminance or irradiance. To improve the calibration capability for these types of applications and to avoid the averaging method mentioned above, NIST has developed a new laboratory, the Spectral Irradiance and Radiance Calibration using Uniform Sources (SIRCUS SIRCUS Standard Information Retrieval Capability for Users
SIRCUS SIDPERS Installation Retrieval Command/Unit Support
) facility, for these types of calibration (34).

Figure 9 gives a schematic diagram of the SIRCUS facility. It has as a core feature a suite of power-stabilized laser systems covering the wavelength region from 200 nm to beyond 20 [micro]m. The integrating sphere provides a spatially averaged beam that is uniform in intensity over a detector aperture. In this arrangement a TSD that has been carefully characterized for ASR and spatial uniformity, and fitted with a precision aperture, is used to determine the irradiance where test detectors are later substituted in position. Using this procedure, the irradiance responsivity of a test detector system can be determined in a manner that is insensitive in·sen·si·tive  
adj.
1. Not physically sensitive; numb.

2.
a. Lacking in sensitivity to the feelings or circumstances of others; unfeeling.

b.
 to the details of the test detector aperture, provided that the test detector intercepts approximately the same portion of the calibration beam that the TSD utilized.

An advantage of this facility is that by knowing the aperture areas of the detectors and the aperture area of the integrating sphere exit port, SIRCUS can be used as a radiance source to calibrate sensor systems that are designed to measure spectral radiance. A good example comes from pyrometry py·rom·e·ter  
n.
Any of various thermometers used for measuring high temperatures.



pyro·met
, where the determination of the spectral radiance of a blackbody source can be used to infer its temperature using the Planck law. In order to provide precision apertures for these purposes, NIST has developed a new facility to accurately determine aperture areas so that their uncertainties will not limit the accuracy of radiometric and photometric measurements (35).

The array of facilities briefly described here allows NIST to establish scales for the radiometric and photometric SI quantities shown in Table 1 with measurement chains linking back to the HACR. As the technology for using cryogenic radiometers progresses, we will implement these changes and thereby enhance our photometric and radiometric measurement standards for NIST's customers. This technical approach has been embraced worldwide. Most national laboratories engaged in maintaining their own radiometric and photometric units have established, or are establishing, a cryogenic-radiometer-based system of measurement (24).

3. Photometry

The SI base units for mass, length, time, temperature, electrical current, and the quantity of matter are completely described in terms of physical properties, but the base unit for luminous intensity is narrowly defined by optical power measurement at a single wavelength. However, within SI, the candela stands apart from the other six base units in that it is anthropocentric anthropocentric /an·thro·po·cen·tric/ (an?thro-po-sen´trik) with a human bias; considering humans the center of the universe.

an·thro·po·cen·tric
adj.
1.
. The purpose of photometry is to measure all light with a metric related to human vision, so one extends the definition of the candela to other wavelengths using V([lambda]) (5). This aspect of its definition also suggests methods for the realization of the unit.

3.1 Candela

The photometric quantities in Table 1 are related to the spectrally-dependent radiometric quantities by integrating them over the wavelengths where the eye is sensitive, as weighted by V([lambda]). (1) For example, the luminous flux is related to the radiant flux:

[[PHI].sub.v] = [K.sub.m][[integral of].sub.[lambda]] [PHI]([lambda]V([lambda])d[lambda], (2)

where [K.sub.m] is the maximum spectral efficacy of radiation for photopic vision. As a result of the 1979 definition of the candela, [K.sub.m] is equal to 683 lm/W [36] (2). The integral in Eq. (2) is evaluated over the range of wavelengths where V([lambda]) is non-zero, which as a practical matter is a wavelength range from 380 nm to 780 nm. This equation formally expresses the means of providing photometric information about light sources or visual illumination systems, such as building lighting, by integrating the appropriate radiometric quantities with a V([lambda]) weighting.

This integration can be done physically by constructing a photodetector system with a spectral responsivity proportional to the V([lambda]) function. That is, one need not make spectrally resolved measurements followed by a numerical integration In numerical analysis, numerical integration constitutes a broad family of algorithms for calculating the numerical value of a definite integral, and by extension, the term is also sometimes used to describe the numerical solution of differential equations. . This is shown schematically in Fig. 10a. A light source with some spectral distribution is placed at a distance r from a photodetector. The system is designed such that combined filter and detector give an overall responsivity proportional to the V([lambda]) distribution. Therefore, the output current [i.sub.o] is composed of signal from all the spectral components of the light, each receiving the appropriate weighting.

Figure 10b shows the data for the ASR calibration of one of NIST's photometers that is used for routine photometric measurements. Also plotted for reference is the V([lambda]) distribution where the vertical axes axes

[L., Gr.] plural of axis. The straight lines which intersect at right angles and on which graphs are drawn. Usually the horizontal axis is the x-axis and the vertical one the y-axis. Called also axes of reference.
 have been normalized for ease of comparison. In this instance, as can be seen from the figure, the combination of the filter and detector responsivity produced an ASR that matches the shape of the V([lambda]) function very closely. The small differences between V([lambda]) and the ASR for a photometer Photometer

An instrument used for making measurements of light, or electromagnetic radiation, in the visible range. In general, photometers may be divided into two classifications: laboratory photometers, which are usually fixed in position and yield results
 can be accounted for, and small corrections to a measurement result can be made (4, 37).

To understand how the photometers can be used to establish the photometric quantities, consider the arrangement of the lamp in Fig. 10 a that produces a spectral irradiance E([lambda]) on the plane of the detector whose aperture has area A. The area of the aperture is assumed to be small with respect to the distance r, hence the irradiance is assumed to be uniform over the aperture. (In an actual experiment or calibration these assumptions would be verified and an uncertainty assigned.) The result is that the spectral radiant flux is given by A E([lambda]). If the ASR of the photodetector is s([lambda]), the output current [i.sub.o] of the detector system will be given by,

[i.sub.o] = A [[intergral].sub.[lambda]] E([lambda])s([lambda])d[lambda] (3)

following directly from the definition of the ASR. The gain of the amplifier is ignored in this discussion and the output will be taken as a current even though in practice a current-to-voltage converter This article or section may contain original research or unverified claims.

Please help Wikipedia by adding references. See the for details.
This article has been tagged since September 2007.
 may be used. The transfer properties of the amplifier can be very accurately determined, but this is beyond the scope of the present discussion.

Alternatively we can express the output current in terms of a luminous flux responsivity [S.sub.v,f] expressed in the unit [lambda]/lm and defined in the following manner:

[S.sub.v,f] = [i.sub.o]/[[PHI].sub.v] = [i.sub.o]/A [K.sub.m] [[integral].sub.[lambda]] E([lambda])V([lambda])d[lambda]. (4)

This follows directly from Eq. (2), presuming that the irradiance is constant over the aperture area A. Combining Eqs. (3) and (4) gives:

[S.sub.v,f] = [[integral].sub.[lambda]] E([lambda])S([lambda]) d[lambda]/[K.sub.m] [[integral].sub.[lambda]] E([lambda])V([lambda])d[lambda]

S(555 nm) [[integral].sub.[lambda]] E([lambda])[S.sub.n]([lambda]) d[lambda]/[K.sub.m] [[integral].sub.[lambda]] E([lambda])V([lambda])d[lambda]. (5)

For convenience and to make our connection to the definition of the candela more apparent, we have factored S([lambda]) into S(555 nm) [S.sub.n]([lambda]). The term S(555 nm) is the absolute spectral responsivity at 555 nm (the peak position of the V([lambda]) function). The term [S.sub.n]([lambda]) is the relative spectral responsivity, normalized to unity at 555 nm (also the value of V([lambda]) at its peak).

Clearly, if [S.sub.n]([lambda]) were to equal V([lambda]) then the two integrals would cancel. In practice, with modern technology and optical fabrication fabrication (fab´rikā´shn),
n the construction or making of a restoration.
 skill, one can achieve a reasonably close match. This allows for the simplification

[S.sub.v,f] = S(555 nm)/[K.sub.m](1 + correction terms), (6)

where the correction terms depend upon the spectral mismatch mismatch

1. in blood transfusions and transplantation immunology, an incompatibility between potential donor and recipient.

2. one or more nucleotides in one of the double strands in a nucleic acid molecule without complementary nucleotides in the same position on the other
 between [S.sub.n]([lambda]) and V([lambda]) as well as the spectral distribution of the light.

A filtered detector system designed in this way, with a spectral responsivity that mimics V([lambda]), is called a photometer. Its luminous flux responsivity can be determined by knowing the ASR at a wavelength of 555 nm, and constants.

In the same way that we wrote the spectral radiant flux received by the photometer as A E([lambda]), we can write the luminous flux as [[PHI].sub.v] = A [E.sub.v], where [E.sub.v] is the illuminance expressed in lux (lx). (1 lx = 1 lm/[m.sup.2].) It follows that the illuminance responsivity [S.sub.v,i] of a photometer is:

[S.sub.v,i] = A [S.sub.v,f]. (7)

For the case of a point-like light source, which is realized by having the dimensions of a lamp and the photometer aperture small compared to r, the distance between them, the relationship between the illuminance [E.sub.v] and luminous intensity [I.sub.v] is:

[E.sub.v]=[I.sub.v]/[r.sup.2]=[i.sub.o]/[S.sub.v,i]=[i.sub.o]/A [S.sub.v,f]. (8)

Solving for [I.sub.v] we find,

[I.sub.v]=[i.sub.o][r.sup.2]/[S.sub.v,i]=[i.sub.o][r.sup.2]/A [S.sub.v,f]. (9)

Luminous intensity is thereby determined from the signal from a calibrated photometer and geometrical factors. Most importantly, the 1979 redefinition of the candela allows this to be an approach for the realization of a luminous intensity scale--quite a simplification from the previous requirement to use a platinum-point black-body source. This procedure has the additional attraction that uncertainty in a candela scale can be reduced as the ability to measure the illuminance responsivity of a photometer improves.

The illuminance responsivity, as can be seen from Eqs. (6) and (7), depends directly upon the ASR value at 555 nm, aperture area, a known constant ([K.sub.m]), an correction terms. Today, the stability and continuity of the SI unit relies upon the ability to construct and maintain photometers whose ASR is known with low uncertainty. The advent of cryogenic radiometer technology and the availability of stable silicon detectors have considerably advanced the ability to measure ASR, but this factor still is a limiting factor A factor or condition that, either temporarily or permanently, impedes mission accomplishment. Illustrative examples are transportation network deficiencies, lack of in-place facilities, malpositioned forces or materiel, extreme climatic conditions, distance, transit or overflight rights,  in improving the measurement of luminous intensity.

At NIST, the calibration of [S.sub.v,f] (or [S.sub.v,i]) is traced to the HACR. The current realization of the NIST candela has a relative expanded uncertainty of 0.41 % (coverage factor k = 2) with the major component of the uncertainty due to the uncertainty in the illuminance responsivity (38, 39). This is roughly a factor-of-ten increase in uncertainty over the uncertainty of a HACR measurement at the beginning of the measurement chain. The additional uncertainty accumulates through the TSD comparison steps and arises due to geometric issues in the transition from flux measurement to irradiance measurement. The development of SIRCUS is expected to improve the measurement of ASR and irradiance by a factor of five or so and thereby improve the photometric scales by a similar amount (34). Our initial effort using SIRCUS to calibrate filtered detectors for radiation temperature measurements indicates that this expectation is realistic and that we can safely predict ongoing improvement in measurement of photom etric quantities (40).

Figure 11a shows a schematic drawing Schematic drawing

Concise, graphical symbolism whereby the engineer communicates to others the functional relationship of the parts in a component and, in turn, of the components in a system.
 of the major features of the photometric bench that is used to maintain the NIST candela scale. This bench provides the ability to calibrate customers' photometers and lamps with respect to the NIST standards. Six photometers are mounted on a carousel that allows the rotation of any one of them into the light beam. The instrument has two alignment telescopes to correctly position the lamp on the optical axis In a lens element, the straight line which passes through the centers of curvature of the lens surfaces. In an optical system, the line formed by the coinciding principal axes of the series of optical elements. . The distance between the lamp and the photometer entrance aperture is measured with a digital linear encoder A linear encoder is a sensor, transducer or readhead paired with a scale that encodes position. The sensor reads the scale in order to convert the encoded position into an analog or digital signal, which can then be decoded into position by a digital readout (DRO).  that has been calibrated traceable to NIST length standards. The stability of the lamp is monitored with a stable monitor detector mounted on the baffle, and a shutter allows for the determination of the dark currents of the photometers. Dimensional stability dimensional stability,
n See stability, dimensional.
 is ensured by mounting the entire apparatus on commercially available optical tables. Figure 11b is a photograph of the instrument and ancillary equipment.

3.2 Lumen lumen: see photometry.


A unit of measurement of the amount of brightness that comes from a light source. Lumens define "luminous flux," which is energy within the range of frequencies we perceive as light.


While the candela is the SI base unit for photometry, the lumen, the unit of luminous flux, is generally of greater commercial interest. A key metric for lamps and illumination devices is their total luminous flux, which is their luminous intensity integrated over the entire solid angle, usually 4[pi], into which they radiate ra·di·ate
v.
1. To spread out in all directions from a center.

2. To emit or be emitted as radiation.



ra
.

This measurement is usually accomplished using an instrument called a goniophotometer. Goniophotometers are usually large, room-sized devices that provide movement of a photometer over a solid angle of 4[pi] and thereby provide the total luminous flux from direct integration of the luminous intensity (or illuminance) (2-4, 41). Until very recently this had been the primary method of directly determining the total luminous flux of a light source.

However, Y. Ohno at NIST has developed a new method of accurately integrating the total flux from a lamp, which uses an accurately characterized integrating sphere (39, 42-44). An integrating sphere is an instrument constructed by coating the inner surface of a spherical spher·i·cal
adj.
Having the shape of or approximating a sphere; globular.
 shell with a highly reflective and diffusive dif·fu·sive  
adj.
Characterized by diffusion.



dif·fusive·ly adv.

dif·fu
 material. The result is that the light from a lamp inside the sphere undergoes many reflections, which averages the spatial distribution of the lamp and results in a uniform average illuminance at the sphere's wall. (The integrating sphere concept was introduced to light measurement in 1892 by Sumpner in the course of his work on studying the reflection properties of surfaces and how the reflectances of walls affected the overall illumination in a room (45).)

It can be shown that if a source of total luminous flux [[PHI].sub.v] is placed in the interior of a sphere of radius r whose reflectance is [rho], then the illuminance [E.sub.v] at the surface of the sphere is given by Eq. (10) (3):

[E.sub.v] = [[PHI].sub.v] [rho]/4[pi][r.sup.2](1 - [rho]) (10)

The reflectance of a sphere can vary temporally, spatially, and with wavelength, and hence the integrating sphere has traditionally been employed only in the substitution mode for measurement with respect to a standard lamp standard lamp
Noun

a tall electric lamp that has a shade and stands on a base

standard lamp n (BRIT) → lámpara de pie

standard lamp n (
 whose total luminous flux is known by use of a goniophotometer. In the substitution method In optical fiber technology, the substitution method is a method of measuring the transmission loss of a fiber. It consists of:
  1. using a stable optical source, at the wavelength of interest, to drive a mode scrambler, the output of which overfills (drives) a 1 to
, a test lamp is compared with a standard lamp of known luminous flux and which has similar distribution of radiation.

The implementation of Ohno's method at NIST is shown in Fig. 12a, where a 2.5 m diameter integrating sphere is fitted with a cosine-corrected photometer for use as a monitor for purposes of the calibration. The integration sphere has a hole, shown in the lower right portion in the figure, to allow for the admission of a known external flux [[PHI].sub.in]. The monitor photometer signal is determined with the externally applied input flux. The external flux is shut off and the monitor photometer then determines the signal due to the test lamp whose flux is shown as [[PHI].sub.test]. The signals at the monitor photometer are [y.sub.in], and [y.sub.test], respectively. The external flux, [[PHI].sub.in] supplied to the sphere is given by A [E.sub.v], where [E.sub.v] is the illuminance measured at the precision aperture A using a calibrated photometer whose ASR calibration was discussed in the previous section. The test lamp flux is then given by

[[PHI].sub.test] = c [[PHI].sub.in] [y.sub.test]/[y.sub.in]. (11)

Here, c is a correction factor that contains terms to account for the spatial nonuniformity of the sphere, the oblique o·blique
adj.
Situated in a slanting position; not transverse or longitudinal.



oblique

slanting; inclined.
 angle of incidence of the external source radiation, and the spectral mismatch between the test lamp and the external source. This latter term arises because the combination of the integrating sphere and photometer can have a relative spectral responsivity that deviates from V([lambda]). These combined corrections are on the order of 1 % to 2 % and are known with some certainty; they are not limiting factors in the accuracy of the measurement. Determination of the correction factor involves a detailed angular angular /an·gu·lar/ (ang´gu-lar) sharply bent; having corners or angles.  mapping of the sphere's response as well as ascertaining its spectral characteristics. These procedures are extensively described in the literature (42-44).

Figure 12b is a photograph of the NIST 2.5 m diameter luminous flux integrating sphere and some of the associated laboratory instrumentation. The sphere built from preformed plastic sections, and its inside is coated with a fluorocarbon fluorocarbon /flu·o·ro·car·bon/ (floor´o-kahr?b?n) any of the class of organic compounds consisting of carbon and fluorine only.  composite material composite material or composite, any material made from at least two discrete substances, such as concrete. Many materials are produced as composites, such as the fiberglass-reinforced plastics used for automobile bodies and boat hulls, but the .

Measurement of luminous flux with Ohno's method has a relative expanded uncertainty of 0.53 % (coverage factor k = 2). The primary contribution to the uncertainty arises from the uncertainty in the ASR measurement of the standard photometer. As with the candela, the improvements that are envisioned in ASR measurements utilizing SIRCUS and other new facilities will have a direct impact upon this important photometric quantity.

3.3 Illuminance and Luminance

Illuminance and luminance, quantities used extensively in lighting engineering, are related to the spectro-radiometric quantities of irradiance and radiance by Eq. (2). Irradiance is a measure of optical power falling on a reference surface, and radiance quantifies light emission, per area and solid angle, by a source. We discussed above how the illuminance response (in A/lx) of a photometer can be determined. Figure 13 shows how this type of photometer can be used to characterize the luminance of a source that is constructed using two integrating spheres. The two integrating spheres are used to improve the uniformity of the light output as well as to allow the introduction of an aperture wheel between them, which provides the ability to change the luminance level over a large dynamic range. The source and sphere configuration produces a uniform illumination at the plane of the sphere aperture [A.sub.s]. A standard photometer with a precision aperture area [A.sub.d] is placed at a distance d from the sphere aperture. The luminance of the sphere source [L.sub.v] can be related to the illuminance [E.sub.v] measured by the photometer (39):

[L.sub.v] = k [E.sub.v][d.sup.2]/[A.sub.s], (12)

where,

k [approximately equal to] (1 + [A.sub.s]/[pi] [d.sup.2] + [A.sub.d]/[pi] [d.sup.2]).

Here, k is the configuration factor that accounts for the finite sizes of the apertures. Configuration factors have been tabulated in the literature for most illumination geometries ordinarily used (46, 47).

The luminance of the sphere source can be determined using a calibrated photometer whose illuminance response is known, and then another test instrument that is designed to measure luminance can be calibrated from it. In this manner luminance meters can be directly calibrated for NIST customers using a method based upon the NIST illuminance response determination described earlier in this paper. The monitor detector ensures temporal stability of the system for the transfer measurement and allows for the changing of the sphereconnecting aperture in a controlled way. The linearity of the silicon detectors used in the photometer allows for devices (such as shown in Fig. 13) to be operated over many orders of magnitude of signal output with the same detector configuration. This instrument is used to calibrate luminance meters for NIST customers with respect to the NIST maintenance of luminance that has a relative expanded uncertainty of 0.50 % (coverage factor k = 2). The largest source of uncertainty is in the illuminance responsivity of the photodetector and hence this quantity will attain lower uncertainty as the ASR measurements for detectors are improved. The actual calibration uncertainties for the customers' units may be slightly larger due to vagaries of the supplied commercial instruments.

The detector-based techniques described here for photometry have evolved to the point that NIST now recommends that its customers begin maintaining their photometric and radiometric standards on a detector basis whenever possible (48). Once adopted by the scientific establishment, this technology offers the prospect of a more economical and reliable method of maintaining photometric standards.

4. Spectral Sources and Pyrometry

While photometric units are integral measures over the whole spectrum, there is considerable value in standards and measurement services that provide spectral information too. Increasingly, detectors calibrated for spectral responsivity serve these needs (48). However, in many applications standard sources remain indispensable. In this section, we provide two examples.

4.1 Thermal Sources for Spectral Radiometric Quantities and Radiation Temperature

In the last century, the most prevalent standard light sources were electric lamps. With reasonable care, they could be calibrated and shipped to customers and field sites, and they were easier to operate than the flame standards of earlier times.

NIST continues to supply a variety of lamps for different needs. Some lamps are calibrated for spectral irradiance, the light that the lamp casts on a surface a specific distance away. Other lamps are calibrated for spectral radiance. These have broad, flat filaments that one observes directly. But no matter what radiometric property a lamp is designed to communicate, the ultimate goal is to allow a customer to calibrate radiometric instrumentation for its intended function.

Until recently, the physical foundation for spectral radiance and irradiance standards was the predictable spectral output of blackbody sources that obeyed Planck's radiation law Planck's radiation law

A law of physics which gives the spectral energy distribution of the heat radiation emitted from a so-called blackbody at any temperature.
 (49, 50). Of most importance were those blackbody sources that operated at the freezing-point temperatures of pure metals (that is, the temperature at which a molten metal solidified so·lid·i·fy  
v. so·lid·i·fied, so·lid·i·fy·ing, so·lid·i·fies

v.tr.
1. To make solid, compact, or hard.

2. To make strong or united.

v.intr.
). One could compute the spectral radiance of these blackbodies, but only as well as you knew what their temperatures actual were.

In 1990 at NIST, Mielenz et al. determined the freezing-point temperature of gold to be 1337.33 K with an expanded relative uncertainty of 0.34 K (coverage factor k = 3) using a filter radiometer (FR) similar to that shown in Fig. 2 (51). This achievement provided the impetus at NIST to decouple the spectral radiance and irradiance scales from any particular temperature scale (e.g., IPTS-68 or ITS-90) that might predefine the metal freezing-point temperatures. Instead, blackbody temperatures began to be measured ab initio [Latin, From the beginning; from the first act; from the inception.] An agreement is said to be "void ab initio" if it has at no time had any legal validity.  using FRs traceable to absolute detectors (52-54). Other national laboratories, including the National Physical Laboratory (NPL) in the U.K., the Physikalisch-Technische Bundesantalt (PTB PTB Physikalisch Technische Bundesanstalt (Germany)
PTB Partido Trabalhista Brasileiro (Brazilian Labor Party)
PTB Phosphotyrosine-Binding
PTB Powers That Be
PTB Power Tab
) in Germany, and the All-Russian Institute for Optophysical Measurements (VNIIOFI) in Russia have also pursued this area of research. Early results were reported as part of an international intercomparison in 1998 (55).

The basic technique for temperature measurements using an FR is shown in Fig. 14. A blackbody source at a temperature T is placed behind a precision circular aperture of known area A1 (radius [r.sub.1] such that this aperture determines the amount of the source viewed by an FR positioned at a distance d. The FR has an aperture of known area [A.sub.2] (radius [r.sub.2]) and has been calibrated such that its ASR, s [lambda], is determined over the wavelength region where the silicon detector is responsive. The dimensions shown in the figure are typical in the NIST implementation. The spectral radiant flux [PHI]([lambda]) at the detector is given by Eq. (13) in terms of the spectral radiance L ([lambda]) of the blackbody source and the geometrical parameters:

[PHI]([lambda] = L([lambda])[A.sub.1][A.sub.2]/[D.sup.2](1+[delta]+...), (13)

where [D.sup.2] = ([r.sup.2.sub.1] + [r.sup.2.sub.2] + [d.sup.2]), = and [delta] = [r.sup.2.sub.1] [r.sup.2.sub.2]/[D.sup.4].

The expressions for D and [delta] arise from the finite sizes of the apertures, similar to the configuration factor in Eq. (12) (46, 47).

Analogous to the treatment above for the FR used as a photometer, we can write the output current [i.sub.o] of the FR in terms of the ASR and the spectral radiant flux [PHI]([lambda]). Using the expression for spectral radiant flux from Eq. (13) we can write

[i.sub.o] = [[integral].sub.[lambda]] [PHI]([lambda])s([lambda])d[lambda], (14)

= [A.sub.1][A.sub.2]/[D.sup.2] (1 + [delta] + ...) [[integral].sub.[lambda]]L([lambda])s([lambda])d[lambda].

We can substitute into Eq. (14) Planck's well-known expression for the spectral radiance from a blackbody source and get the explicit dependence between the output current of the FR and the temperature of the blackbody source. The result is given in the following, where h is the Planck constant The Planck constant (denoted ) is a physical constant that is used to describe the sizes of quanta. , c is the speed of light in vacuum, k is the Boltzmann constant Boltzmann constant

Ratio of the universal gas constant (see gas laws) to Avogadro's number. It has a value of 1.380662 × 10−23 joules per kelvin.
, and [[epsilon].sub.[lambda]] is the emissivity Emissivity

The ratio of the radiation intensity of a nonblack body to the radiation intensity of a blackbody. This ratio, which is usually designated by the Greek letter ε, is always less than or just equal to one.
 of the blackbody cavity.

[i.sub.o] = [A.sub.1][A.sub.2]/[D.sup.2](1 + [delta] + ...) [[integral].sub.[lambda]] 2h [c.sup.2][[epsilon].sub.[lambda]]/[[lambda].sup.5](hc/e[lambda]kT -1) s([lambda])d[lambda] (15)

The importance of this equation is that it provides the relationship between the temperature being measured and the FR output, in a form that is dependent only upon known physical constants, geometric factors, and the ASR of the FR. Eq. (15) also indicates the important sources of uncertainty in the measurement. The largest component of the uncertainty usually arises from s([lambda]). This limiting factor, as in photometry, indicates the need for continued improvement in ASR determinations and improvements in the design of FRs. The defining apertures and distances are also important. However, the geometry of the experiment can usually be arrange such that the correction term [delta] is small and negligible. If the diameters of the apertures are small compared to the distance d, then D reduces to d, and the calculation is simplified considerably. The blackbody cavity must also be constructed such that the emissivity is a know quantity, which usually means the blackbody is designed to ensure the emissivity is n early unity and independent of wavelength. The blackbody must be designed so the calculated emissivity can be checked experimentally.

Figure 15 shows the ASRs of a number of FRs constructed at NIST to be used in temperature determination. Also shown in the figure is a plot of the spectral radiance from a 3000 K blackbody, typical of those used as the basis for spectral irradiance or spectral radiance lamp calibrations. The spectral radiance increases by orders of magnitude at longer wavelengths, hence the responsivities of the FRs must be well characterize over all wavelengths where silicon has a response. Even small light-leakage of the filter in the longer wavelength region can introduce a significant contribution to the integral in Eq. (15).

The understanding of the out-of-band response of an FR is one of the driving forces for the construction of SIRCUS, discussed earlier in this paper. SIRCUS will allow the measurement of the ASR of an FR far from the central wavelengths of transmittance, spanning a dynamic range of up to 10 orders of magnitude. Through an accurate determination of the tail portions of the ASR, uncertainties in temperature measurement will be minimized.

The FRs whose ASRs are shown in Fig. 15 have been used to measure temperatures of a variable-temperature blackbody that was independently measured using the NIST photoelectric Converting photons into electrons. When light is beamed onto a metal, electrons are released from its atoms. The higher the light frequency, the more electron energy released. Photonic sensors of all kinds work on this principle. They sense light and cause an electric current to flow.  pyrometer (56). The comparison of the results for three of the FRs is shown in Fig. 16. We see that the two methods gave the same temperatures to well within their combined experimental uncertainties. Of most significance was the uncertainty of the gold-point reference temperature, that is, gold's freezing-point temperature as previously determined at NIST. The calibration chain for the pyrometer traced back to this key parameter, and the effect of its uncertainty at different temperatures is shown explicitly in the figure. However, the uncertainty of the gold-point temperature does not affect the FR method, thus demonstrating how FRs can improve radiation temperature metrology.

That said, the results using the other FRs were not comparable because of larger uncertainties in the determination of their ASRs or, as in the case of FR1, the diminished signal from the blackbody in the ultraviolet region of the spectrum. The uncertainties in the NIST measurement of ASRs are larger in the UV and near-IR (800 nm to 1200 nm) and hence the uncertainties when using FR1, FR4, and FR6 were larger than when using the other devices. However, as NIST improves the ASR measurements in the UV and near JR by the use of SIRCUS, and improves the transfer-standard detectors used in the DSC laboratories, the uncertainty in the radiation temperature measurement will decrease significantly.

We expect to routinely perform the characterization of FR ASR's to within a relative standard uncertainty of 0.05 % using SIRCUS, which will provide the capability of determining temperature to within a relative uncertainty of about 0.1 % at 3000 K. This will improve the radiation temperature measurements discussed in Ref. (56), as well as the NIST scales of spectral irradiance and spectral radiance. Further, the pyrometers used for calibration of pyrometry lamps and other pyrometers can have their measurement scale set directly by viewing a blackbody that has been characterized with an FR system. When this is fully implemented in the pyrometry laboratory, we expect the uncertainty in the calibration of a pyrometer to decrease by a factor of two or more.

Once the temperature of a blackbody is determined, its spectral characteristics can be used to calibrate the spectral irradiance of lamps. This can be understood by using the definitions given above in Eq. (13). Using these we can rewrite re·write  
v. re·wrote , re·writ·ten , re·writ·ing, re·writes

v.tr.
1. To write again, especially in a different or improved form; revise.

2.
 an expression for the spectral irradiance E([lambda]) for the configuration shown in Fig. 14:

E([lambda]) = [PHI]([lambda])/[A.sub.2] = L([lambda])[A.sub.1]/[D.sup.2] (1 + [delta] + ...). (16)

NIST uses FEL FEL - Function Equation Language. Programs are sets of definitions. Sequences are lists stored in consecutive memory. "FEL Programmer's Guide", R. M. Keller, AMPS TR 7, U Utah, March 1982.  quartz-halogen lamps as calibration artifacts in spectral irradiance scale dissemination to its customers. High-temperature blackbodies that can operate near the distribution temperature of the FEL lamps (2800 K to 3100 K) have been developed, and hence their irradiance can be compared directly to the test lamp for a direct calibration. Previous techniques starting out from the gold freezing point required several more steps in the calibration because the gold freezing point is at a much lower temperature than the distribution temperature of most lamps used for calibration devices (55, 57). This disparity in signal levels in the visible wavelength region for calibration systems contributed significantly to the uncertainty in the calibrations for spectral irradiance and radiance.

Figure 17 shows the schematic layout for an instrument that determines the spectral irradiance of a test lamp using the method just described. The carriage is moved to place the FR in a position to view the output of a high-temperature blackbody (HTBB) source. Using Eq. (15) and measuring all the parameters allows the temperature of the HTBB, and thus its spectral radiance and irradiance to be deduced. The carriage then is moved so that the integrating sphere is in a position to view the HTBB. The system composed of the spectroradiometer, its input optics, and detector is calibrated from this known spectral radiance. The distance d' may differ in the placement of the integrating sphere and the FR because of their differing entrance apertures. An effort is made to have both of these devices view the same solid angle of the HTBB in order to eliminate problems from possible temperature variations within the HTBB cavity. The spectroradiometer has up to three different detectors that are chosen for optimal sensit ivity in different portions of the wavelength region from 200 nm to 2500 nm. Once the spectral irradiance responsivity of the spectroradiometer system is determined, the carriage is moved to a position where the integrating sphere aperture is illuminated il·lu·mi·nate  
v. il·lu·mi·nat·ed, il·lu·mi·nat·ing, il·lu·mi·nates

v.tr.
1. To provide or brighten with light.

2. To decorate or hang with lights.

3.
 by a test lamp, in this case an FEL quartz-halogen lamp. The spectral irradiance of the test lamp is then measured with the calibrated instrument. In practice, to avoid the continual use of the HTBB, a set of working-standard lamps is calibrated utilizing the blackbody, and they in turn can be used to calibrate routine test lamps for customers. With a proper choice and selection of working standard lamps there is no significant increase in uncertainty as compared to direct use of the blackbody on each calibration.

The carriage can move the integrating sphere into positions to view up to six different sources, and hence several different lamps can be calibrated in a series. The FR can be positioned to view the lamps as well as the HTBB, which offers a way of monitoring the stability of the working standard lamps. This lamp monitoring is a component of the quality-control regimen regimen /reg·i·men/ (rej´i-men) a strictly regulated scheme of diet, exercise, or other activity designed to achieve certain ends.

reg·i·men
n.
1.
 to ensure a stable and reliable measurement system.

A slight adjustment of the input optics provides the capability of creating a spectral radiance scale based upon an HTBB. NIST can provide calibrations for spectral radiance in two general ways: on a specialized lamp calibrated for its output, and on a detector system calibrated for its spectral radiance responsivity. The configuration shown schematically in Fig. 18 indicates how the lamp calibration is performed. The FR is again used to determine the temperature, and thus the spectral radiance of the HTBB. The carriage then moves a plane mirror to a point where a portion of the HTBB source area is focussed onto the entrance slit of the spectroradiometer, and the spectral radiance responsivity of the spectroradiometer system is determined. Finally, the carriage is moved so that the calibrated spectroradiometer views a portion of the filament filament, in astronomy: see chromosphere.  of a specialized radiance lamp.

Additionally, a second detector system can be mounted in a position equivalent to the FR shown in Fig. 18, and its spectral radiance responsivity determined by viewing either the HTBB or a calibrated lamp. Radiometers designed to measure spectral radiance generally have input optics that view a fixed solid angle. The calibration effort tries to have the test radiometer view the same source area that the reference spectroradiometer does in order to avoid errors due to any temperature gradients temperature gradient
n.
The rate of change of temperature with displacement in a given direction from a given reference point.



temperature gradient 
 in the sources. These issues of having uniform sources and providing the same viewing area for the device under test and the reference spectroradiometer are central to decreasing the uncertainty of these calibrations.

The transition to the FR-based spectral source scales is taking place as this paper is being written in early 2000, and we expect it to be fully implemented along with a new Facility for Automated Spectroradiometric Calibrations (FASCAL) within a year. These changes will improve the uncertainties in the calibration of both lamps and pyrometers. Based upon our initial results, we expect the uncertainties to improve as shown in Table 2 for spectral irradiance and in Table 3 for spectral radiance. The uncertainties shown are a result of a detailed treatment of the measurement equation for the calibration, using the uncertainties of temperature determination that we have presently demonstrated with the FR technique. The uncertainties for the spectral quantities are the relative expanded uncertainties with a coverage factor of k = 2. These decreases in uncertainty represent a significant improvement for our customers. The use of higher-quality FRs offers a mechanism for additional improvement as the state-of-the- art in detector calibration improves. Other improvements, such as better variable-temperature blackbodies that could be used routinely, instead of lamps, could also decrease the uncertainties significantly.

4.2 Synchrotron Radiation

NIST has maintained a synchrotron radiation source Coordinates:  The Synchrotron Radiation Source (SRS) at the Daresbury Laboratory in Cheshire, United Kingdom is a second generation[1] synchrotron radiation source.  since 1961, when a 180 MeV electron storage ring was commissioned as the Synchrotron Ultraviolet Radiation Source (SURF). It began the tradition at NIST of using synchrotron radiation to study a wide range of physical problems.

In 1974, SURF was upgraded to 240 MeV, with greater operating current. The original magnet was retained, but the rest of the major components were replaced or modernized mod·ern·ize  
v. mo·dern·ized, mo·dern·iz·ing, mo·dern·iz·es

v.tr.
To make modern in appearance, style, or character; update.

v.intr.
To accept or adopt modern ways, ideas, or style.
. Major changes included the replacement of the old glass vacuum envelope with a stainless steel vacuum chamber and the introduction of a new injection system. This instrument was named SURF II, and it underwent a number of additional improvements that enhanced beam lifetime, beam current, and operating energy (58). SURF II typically operated in a mode where the output radiant flux into an optical system could be known to within a relative standard uncertainty of about 2 %. This uncertainty, while appropriate for many uses, limited SURF II's usefulness as an absolute radiometric source.

In 1994 a decision was made for a complete upgrade and redesign of SURF to create a new high-accuracy facility that is called SURF III. The major change effected in the upgrade was the replacement of the original 127 metric ton laminated laminated /lam·i·nat·ed/ (-nat?ed) having, composed of, or arranged in layers or laminae.

laminated

made up of laminae or thin layers.
 iron magnet with a new 181 metric ton solid iron magnet. SURF I operated in a betatron betatron: see particle accelerator.
Betatron

A device for accelerating charged particles in an orbit by means of the electric field E from a slowly changing magnetic flux &PHgr;.
 mode and relied upon the laminated design to reduce electrical losses due to eddy currents Eddy current

An electric current induced within the body of a conductor when that conductor either moves through a nonuniform magnetic field or is in a region where there is a change in magnetic flux. It is sometimes called Foucault current.
. However, this design reduced the uniformity of the magnetic field, a major source of uncertainty.

A drawing of SURF III that shows its major features is shown in Fig. 19. Electrons are injected into the magnetic field at an energy of 10 MeV using a small microtron accelerator (not shown). The magnetic field bends the electrons into a circular orbit
For other meanings of the term "orbit", see orbit (disambiguation)


In astrodynamics or celestial mechanics a circular orbit is an elliptic orbit with the eccentricity equal to 0.
 which passes through a radio-frequency (RF) cavity. The electric field in the cavity accelerates the electrons each time through. This, in combination with a ramp-up of the magnetic field (using the electromagnetic field-coils), maintains the electrons in a steady orbit. The process continues until the magnetic field reaches 1.4 T, which holds electrons with an energy of about 350 MeY (59). At this point, an equilibrium is attained as the power supplied by the RF cavity balances the radiative synchrotron losses, discussed below. The electron beam A stream of electrons, or electricity, that is directed towards a receiving object. See electron beam imaging and electron beam lithography.  orbits inside a toroidal ultrahigh-vacuum vessel that has viewports for a number of electron-beam tangent tangent, in mathematics.

1 In geometry, the tangent to a circle or sphere is a straight line that intersects the circle or sphere in one and only one point.
 points, through which the synchrotron radiation exits the beam region for external use (not shown in detail). T he field coils field coil
n.
An electric coil around a field magnet used to generate a magnetic field, as in an electric motor. Also called field winding.

Noun 1.
 provide the magnetic field of the pole pieces, which have been machined to provide appropriate vertical and radial radial /ra·di·al/ (ra´de-al)
1. pertaining to the radius of the arm or to the radial (lateral) aspect of the arm as opposed to the ulnar (medial) aspect; pertaining to a radius.

2.
 field patterns for focussing and containment of the electron beam. The rest of the iron serves as a magnetic return-loop for the field.

The electrons in SURF III move in a circular orbit, as shown in Fig. 19, due to the familiar Lorentz force Lorentz force
n.
The orthogonal force on a charged particle traveling in a magnetic field.



[After Hendrik Lorentz.
 on charged particles charged particle
n.
An elementary particle, such as a proton or electron, with a positive or negative electric charge.
 in a magnetic field. In 1898, Lienard showed that when a charged particle undergoes centripetal acceleration centripetal acceleration

Property of the motion of an object traveling in a circular path. Centripetal describes the force on the object, directed toward the centre of the circle, which causes a constant change in the object's direction and thus its acceleration.
, such as in circular motion In physics, circular motion is rotation along a circle: a circular path or a circular orbit. The rotation around a fixed axis of a three-dimensional body involves circular motion of its parts. , it radiates electromagnetic radiation electromagnetic radiation, energy radiated in the form of a wave as a result of the motion of electric charges. A moving charge gives rise to a magnetic field, and if the motion is changing (accelerated), then the magnetic field varies and in turn produces an . The electrons in the SURF III storage ring are at highly relativistic rel·a·tiv·is·tic  
adj.
1. Of or relating to relativism.

2. Physics
a. Of, relating to, or resulting from speeds approaching the speed of light: relativistic increase in mass.
 energies, requiring relativistic electrodynamics Relativistic electrodynamics

The study of the interaction between electrically charged particles and electromagnetic fields when the velocities involved are comparable to that of light.
 to determine the spectral distribution of the radiation. The Schwinger theory shows that a narrow cone of light cone of light
n.
The bright triangular area of reflected light on the tympanic membrane during examination. Also called light reflex.
 is emitted forward, in the electrons' direction of motion, which contains very high harmonics of the orbital frequency (60). At SURF III, the synchrotron radiation is emitted with significant intensity at wavelengths as short as 1 nm and at wavelengths up into the far-infrared and RF range. The vertical angular spread of the radiation is very narrow and forward at the shortest wavelengths, but diverges at longer wavelengths.

Figure 20 shows the result of calculations using the Schwinger theory for two operating energies of SURF III. Increasing the energy of the stored beam increases the short wavelength output but does not substantially affect the long-wavelength output. The vertical scale in Fig. 20 is related to the radiant intensity of the beam per milliamperes of beam current. To calculate the radiance of SURF III it is necessary to also consider the spatial profile of the electron beam in orbit as well as other geometrical factors. For purposes here, the radiant intensity has been integrated over the vertical angle of emission, which at 100 nm is about 4 milliradians (full-width at half-maximum intensity). Depending upon the monochromator, the wavelength region utilized, and operating conditions, SURF III can deliver monochromatized photon fluxes of typically [10.sup.10] [s.sup.-1] to [10.sup.13] [s.sup.1] (61).

A cryogenic radiometer has been installed on a beam-line at SURF III, and it will be used to calibrate detectors and filters and to develop procedures to calibrate the output of SURF III. Theory indicates that if the magnetic field, radius of the orbit, and electron current are known, then the absolute radiant intensity of the beam can be calculated. At SURF III we will exploit two different techniques to measure the electron current and hence determine the output of the storage ring. The first will be to use an FR as described previously, although instead of the Planckian spectral distribution we will have the distribution shown in Fig. 20 derived by Schwinger. By measuring the radiant flux in a specific solid angle with a well-characterized FR, the beam current can be directly determined if the magnetic field and radius of orbit are known. A second method depends on measuring the radiation from a very small number of electrons and deducing the radiant intensity per electron. Scaling this up to normal opera ting ting  
n.
A single light metallic sound, as of a small bell.

intr.v. tinged , ting·ing, tings
To give forth a light metallic sound.
 currents relies on the linearity of silicon detectors (62). With the combination of these methods SURF m's radiant intensity can be predicted to within a relative uncertainty of 0.2 % to 0.5 %, depending upon wavelength and operating conditions.

With these innovations and improvements, SURF III serves as a primary optical power source from wave lengths as short as 1 nm to the far-infrared and terahertz ter·a·hertz  
n. Abbr. THz
One trillion (1012) hertz.

Noun 1. terahertz - one trillion periods per second
THz
 region of the spectra. In addition to serving as a primary spectral source, the radiation from SURF III is employed in a variety of calibration programs for UV detectors, optical properties of materials, reflectometry the soft x-ray region, and as a light source for a cryogenic radiometer program using synchrotron radiation (31).

5. Laser Power and Energy

In the early 1960s, as lasers were first being developed, NIST also began developing standards and measurement techniques required for this new technology Lasers, as opposed to previous light sources, produced narrow beams with very high spatial and temporal coherence, an extremely nonuniform irradiance profile (which was nonetheless significant in its detail), and in some cases, unprecedented irradiance or energy density levels, especially in short pulses. These were the properties that made lasers useful for many novel applications, but the traditional methods of radiometry were unable to measure them accurately and consistently.

Over the years, NIST has developed a substantial portfolio of standards and measurement services to characterize laser radiation at all commonly used wavelengths and over a wide range of powers and energies. The principal standards now in use (Fig. 21) are electrically calibrated calorimeters and radiometers. They accurately compare the power or energy in laser radiation to electrical power or energy, thus providing traceability to SI units (Système International d'Unites) A system of standard units of measurement finalized at the 14th General Conference on Weights and Measures in 1971. It is based on seven units of measure, including three from the MKS system (meter-kilogram-second), the ampere for .

The first NIST laser-energy standard was a "liquid-cell calorimeter calorimeter: see calorimetry.
calorimeter

Device for measuring heat produced during a mechanical, electrical, or chemical reaction and for calculating the heat capacity of materials.
," developed by Jennings in the early- to mid-1960s (63). This calorimeter used an aqueous aqueous /aque·ous/ (a´kwe-us)
1. watery; prepared with water.

2. see under humor.


a·que·ous
adj.
 solution of (CuSO.sub.4) in a cell with a quartz window. A small amount of dark ink was added to the solution to increase its wavelength-absorption range. Thermocouples measured a temperature rise caused by laser-light absorption as well as the corresponding temperature rise caused by an electrical heater. Through equivalence, this device measured the energy of individual laser pulses. Although designed for measuring the energy from pulsed lasers, such as ruby and Nd:glass varieties, the liquid cell calorimeters also worked well for measuring the average power from cw laser sources. For a number of years, these liquid cell calorimeters were used as primary standards at NIST and furnished to DoD primary standards laboratories for calibrating their laser power and energy meters.

Also during the early years of laser-power measurements, Phelan, Cook, Hamilton, and Day developed the electrically calibrated pyroelectric radiometer (ECPR ECPR European Consortium for Political Research (UK)
ECPR Efficient Component Pricing Rule
) as a laser power standard (64-66). The basic elements are shown in Fig. 22. A thin sample of pyroelectric material is sandwiched between two conducting electrodes Electrodes
Tiny wires in adhesive pads that are applied to the body for ECG measurement.

Mentioned in: Electrocardiography
. A pyroelectric material contains a permanent electric-dipole moment, or polarization polarization

Property of certain types of electromagnetic radiation in which the direction and magnitude of the vibrating electric field are related in a specified way.
 field, that varies as a function of temperature. Thus, when it is heated, the electrical potential across the crystal changes. This type of instrument is sensitive to changes in temperature (rather than to the absolute temperature) because the changing surface charges on the crystal are rapidly equalized by free charges attracted to the electrodes.

Here again, heating power can be applied either optically or electrically. The front electrode electrode, terminal through which electric current passes between metallic and nonmetallic parts of an electric circuit. In most familiar circuits current is carried by metallic conductors, but in some circuits the current passes for some distance through a  is coated with a gold-black that is both highly absorbing of laser radiation and somewhat electrically conductive conductive

having the quality of readily conducting electric current.


conductive flooring
flooring or floor covering made specially conductive to electrical current, usually by the inclusion of copper wiring that is earthed
. During operation, electrical heating is alternated with optical heating at about 15 Hz, and the electrical power is adjusted to obtain equivalence with the laser radiation. Because of its sensitivity and other advantageous properties, this device was commercialized, and it is in widespread use in laboratories throughout the world.

In the early 1970s, West developed a cavity-based calorimeter whose responsivity could be analyzed with well-established isoperibol (constant temperature environment) techniques (67, 68). Various models of isoperibol calorimeters have since been developed, and today they are used as primary standards for many NIST laser-measurement services (Fig. 21). For the best accuracy in low-power laser measurements, NIST also uses a laser-optimized cryogenic radiometer (LOCR LOCR Legends of Classic Rock (Canadian radio show)
LOCR Linux Optical Character Recognition
) (69) whose operation is similar to the HACR described earlier. However, the isoperibol calorimeters are used for higher powers, for pulsed lasers, and in other situations where comparison to the LOCR scale is not advantageous.

Both the LOCR and the isoperibol calorimeters contain cylindrical cyl·in·dri·cal
adj.
Of, relating to, or having the shape of a cylinder, especially of a circular cylinder.
 cavities that absorb laser radiation. Thermal sensors A device that detects temperature. Thermal sensors are found in many laptops and desktop PCs in order to sound an alarm when a certain temperature has been exceeded.  attached to the absorbing structure monitor the resulting temperature increase. A temperature-stabilized enclosure or jacket provides the environment required for isoperibol conditions, which allows the temperature sensors to accurately indicate the quantity of laser radiation received. As with the ECRs discussed earlier in this paper, electrical substitution provides the connection between optical power and the NIST electrical measurement scales.

The basic principles of cryogenic radiometers for power measurement were reviewed earlier in this article. Here, we explain the behavior and operation of the isoperibol calorimeters for energy measurement.

A cross section of a typical isoperibol calorimeter is shown in Fig. 23 (70). Its relatively simple geometry allows a straightforward analysis of its response to a laser-energy pulse (71, 72). The energy E absorbed by the calorimeter is related to voltage time-response V(t) of the temperature sensors,

E = [K.sub.el][V.sub.f] - [V.sub.i] - [epsilon] [[integral].sup.tf.sub.ti] (V(t) - [V.sub.[infinity]])dt], (17)

where [K.sub.el] is a constant determined by electrical-energy substitution; [V.sub.i] and [V.sub.f] are voltages that occur at times [t.sub.i] and [t.sub.f], respectively, that temporally bracket the laser-injection period; [V.sub.[infinity]] is the steady state or convergence voltage; and [epsilon] is the cooling constant described below. This equation essentially restates the first law of thermodynamics first law of thermodynamics

law dealing with the transformation of energy. States that energy can neither be created nor destroyed, only converted from one form to another.
: the energy E absorbed is equal to the change in internal energy [[K.sub.el] times ([V.sub.f] - [V.sub.i])], minus the heat transferred from the system during the measurement ([K.sub.el][epsilon] times the voltage integral between times [t.sub.i] and [t.sub.f]). The quantity in brackets is referred to as the corrected voltage rise [DELTA]V and represents the output of the calorimeter for an individual energy measurement. Figure 24 shows a typical voltage-response curve.

This equation assumes a linear system in which both heat capacity and thermal conductivity thermal conductivity

A measure of the ability of a material to transfer heat. Given two surfaces on either side of the material with a temperature difference between them, the thermal conductivity is the heat energy transferred per unit time and per unit
 are constant. To achieve this condition as closely as possible, the energy injection time is limited to keep the temperature rise of the cavity to less than a few degrees. As a linear system, the output response can be represented by a series of exponential 1. (mathematics) exponential - A function which raises some given constant (the "base") to the power of its argument. I.e.

f x = b^x

If no base is specified, e, the base of natural logarthims, is assumed.
2.
 terms, but before and sometime after the energy deposition, the voltage response approximates a single exponential function exponential function

In mathematics, a function in which a constant base is raised to a variable power. Exponential functions are used to model changes in population size, in the spread of diseases, and in the growth of investments.
 containing the cooling time (Law) such a lapse of time as ought, taking all the circumstances of the case in view, to produce a subsiding of passion previously provoked.
- Wharton.

See also: Cooling
 constant [epsilon].

The electrical calibration factor [K.sub.el] is determined by depositing a known amount of electrical energy [E.sub.el] through a heater thermally bonded to the absorbing cavity. Using the relationship

[K.sub.el] = [E.sub.el]/[DELTA]V, (18)

[K.sub.el] is determined for each calorimeter. Obviously, this implies that a calorimeter behaves the same for electrical-energy deposition as it does for laser-energy absorption. In reality, the temperature gradients within the calorimeter are different in the two cases, and consequently their voltage responses may differ as well. This electrical-optical inequivalence is minimized by careful design of the calorimeter. For example, in the calorimeter in Fig. 23, the heat (no matter where it originates in the innermost in·ner·most  
adj.
1. Situated or occurring farthest within: the innermost chamber.

2. Most intimate: one's innermost feelings.

n.
 cavity) is forced to travel through a common isothermal i·so·ther·mal
adj.
Of, relating to, or indicating equal or constant temperatures.



isothermal, isothermic

having the same temperature.
 element connecting the two inner cylindrical cavities before arriving at the temperature sensors on the outer cavity. This inequivalence can also be reduced by using cryogenic temperatures, as is done with cryogenic radiometers. However, for room temperature devices it remains a major source of uncertainty in the measurement.

Since not all the laser energy incident on the calorimeter causes heating, the actual measurement equation is

E = [K.sub.el][DELTA]V/[tau][alpha], (19)

where [tau] is the transmittance of the calorimeter window (unity when no window is used) and a is the absorptance of the cavity. This absorptance factor accounts for light scattered Scattered

Used for listed equity securities. Unconcentrated buy or sell interest.
 back Out of the calorimeter as well as radiative emission due to local heating. Indeed, thermal radiation thermal radiation

Process by which energy is emitted by a warm surface. The energy is electromagnetic radiation and so travels at the speed of light and does not require a medium to carry it.
 is generated primarily where the cavity is the hottest, in the neighborhood where the laser beam is absorbed. This heat pattern is generally not replicated through electrical substitution, potentially causing an inequivalence condition. However, to minimize such problems and to prevent damage to the cavity surfaces, for high-power laser measurements the laser beam is first reflected from a curved surface to spread its energy over a larger area. By careful application of optical spreading techniques, isoperibol calorimeters have been developed for power measurements at the kilowatt kilowatt: see watt.  (73) and multi-kilowatt (74, 75) levels.

If the incident laser-irradiance level is below 200 W/[cm.sup.2] then most black coatings (e.g., black paint, copper oxide Noun 1. copper oxide - an oxide of copper
oxide - any compound of oxygen with another element or a radical
, etc.) work fine and are not likely to suffer damage. However, pulsed laser radiation presents a problem. The energy from short pulses, such as those produced by Q-switched or mode-locked lasers, is deposited in a time interval too short for the heat to diffuse away from the absorption area. The resulting localized temperature increase can exceed the damage threshold of the absorbing material. Consequently, for calorimeters designed for measuring lasers with pulse-energy densities exceeding 300 mJ/[cm.sup.2] an absorbing volume is typically used instead of an absorbing surface. Typical volume-absorbing materials include liquids (as in the liquid-cell calorimeter discussed above), glasses, and certain ceramics.

NIST has designed and built several models of isoperibol calorimeters that use absorbing glasses for pulsed laser radiation. The absorption profile in these glasses can be expressed as

E(x) = [E.sub.0[e.sup.-[alpha]x]], (20)

where E(x) is the energy per area at a depth x into the material, [E.sub.0] is the energy per area at the incident surface, and [alpha] is the absorption coefficient absorption coefficient
n.
1. The milliliters of a gas at standard temperature and pressure that will saturate 100 milliters of liquid.

2. The amount of light absorbed in 1 atom or in 1 unit of thickness or mass of a given substance.
 of the material at the particular laser wavelength being used. The maximum associated temperature rise (assuming no heat transfer during the pulse duration In radar, measurement of pulse transmission time in microseconds; that is, the time the radar's transmitter is energized during each cycle. Also called pulse length and pulse width. ) is then given by

[DELTA]T = [alpha][E.sub.0]/[rho]c, (21)

where [DELTA]T is the induced temperature rise at the material's surface, [rho] is the absorber density, and c is the absorber specific heat. Materials with a low value of [alpha] have a lower associated [DELTA]T but require a greater thickness to absorb all the laser energy. Materials with a high value of [alpha] have a higher [DELTA]T but require thinner sample As a compromise between damage-threshold levels an desirable heat diffusion characteristics, the absorber material is chosen such that its inherent value of a allows complete absorption within a thickness of 1 mm to 2 mm. Using this guideline, glass absorbers have bee applied to infrared [76] and ultraviolet [77, 78] laser measurements. Even though volume-absorbing materials minimize thermal damage, materials used calorimeters must also be free from other effects, such as fluorescence fluorescence (flrĕs`əns), luminescence in which light of a visible color is emitted from a substance under stimulation or excitation by light or other forms of electromagnetic  and degradation, when exposed to laser pulses.

When comparing other detectors to the laser power and energy standards (e.g., during detector calibration), various types of beamsplitter systems are frequently used. They allow simultaneous measurement of the laser power or energy by two or more detectors and, as a result, help to minimize the effects of laser-amplitude variations. Figure 25 illustrates the basic concept. Note that beamsplitters usually have a wedge geometry to eliminate interference effects.

Beamsplitters also extend the operating power range of the standards. In power measurement systems, each beamsplitter is used either to sample part of the beam for a monitor detector or it is used as a calibrated standard itself. When used as a standard, the ratio of powers in the two beams is determined as follows. First, two detectors (usually standard calorimeters) are placed in the beams. Their power ratio is measured a number of times and averaged as [R.sub.1]. The two detectors are the interchanged, and the ratio of power in the two beams measured again to get an average [R.sub.2]. The absolute ratio is taken to be

R = [square root of ([R.sub.1] [R.sub.2])]. (22)

In this procedure, systematic effects cancel if the detectors are linear and spatially uniform. Once R is determined, other detectors can be placed in the beams and compared directly.

For optical fiber power measurements, the beamsplitter system is modified to incorporate fiber components such as optical fibers and fiber splitters (Fig. 26). Laser radiation, typically at low power in the infrared wavelength region, is delivered to the detectors through optical fibers and specialized connectors. Due to the many reflecting surfaces inherent in fiber systems, the coherence of laser radiation can lead to interference effects and power instability if the measurement system is not designed properly [79].

Today, NIST is working to extend its measurement capability and improve the measurement uncertainty in existing laser-measurement areas. In addition, we are developing new capabilities to meet the emerging needs of the laser community. For example, to support the next generation of high-resolution semiconductor photolithography A lithographic technique used to transfer the design of circuit paths onto printed circuit boards as well as the circuit paths and electronic elements of a chip onto a wafer's surface.

A photomask is created with the design for each layer of the board or wafer (chip).
 systems, calorimeter design efforts are in progress to develop standards for excimer-laser measurements at 157 nm. In addition to the power and energy measurements discussed above, related laser measurement capabilities are also maintained. These include source-based measurements such as irradiance profile (80, 81) and relative intensity noise Relative intensity noise (RIN), describes the instability in the power level of a laser. The noise term is important to describe lasers used in fiber-optic communication and LIDAR remote sensing.  (RIN) (82, 83), as well as detector-based measurements such as spatial uniformity (84), power linearity (85), spectral responsivity (86), frequency response (87), and impulse response In simple terms, the impulse response of a system is its output when presented with a very brief signal, an impulse. While an impulse is a difficult concept to imagine, and an impossible thing in reality, it represents the limit case of a pulse made infinitely short in time . Also, various laser-specific materials issues are studied (88). This combined capability allows NIST to give comprehensive coverage to the metrology needs of the laser community. As lasers with new wavelengths or power characteristics are employed by industry, NIST will continue to respond with useful standards and measurement techniques.

6. Conclusions

The redefinition of the candela in terms of optical power by the CPGM, and the adoption by the CIPM of the CIE V([lambda]) function for human visual response, has had a profound impact upon photometry and radiometry. As was intended, this definition shifted the emphasis in photometry and radiometry away from reliance upon blackbody sources operating at metal freezing-points to the reliance upon well-characterized optical radiation detectors.

This review has focussed on the impact of this new direction upon NIST activities, but the same basic story applies to major national metrology institutes around the world. While the redefinition pertained only to photometry, the impetus to develop better optical detectors has benefited the scientific disciplines of spectral radiometry and pyrometry as well. Over the same timespan that photometry has became dependent upon improved optical detectors, the use of lasers in communications and information technologies has also spurred the improvement in electrical-substitution radiometers used for laser-power measurements.

The real beneficiaries, however, are the customers of NIST calibration services, as the calibration ranges and uncertainties have improved significantly. Further, the stability of the measurement chain based upon the cryogenic radiometer provides a new level of robustness and continuity to the national measurement system. Customers, be they engaged in photography, display technology, electronics, optoelectronics, scientific instruments, aerospace, remote sensing Deriving digital models of an area on the earth. Using special cameras from airplanes or satellites, either the sun's reflections or the earth's temperature is turned into digital maps of the area. , or other fields, all enjoy these benefits. They translate into new capabilities for process control, simplified measurement tools, and improved product quality. Indeed, the ongoing shift towards detector-based methods provides widespread gains in accuracy, efficiency, and stability.

7. Symbols

c Heat capacity (specific heat) (J/K "Just kidding." See digispeak. )

d Distance (e.g., separation) (m)

i Current (electrical) (A)

[i.sub.o] Current, output (A)

r Distance (e.g., radius or separation) (m)

t Time (s)

x Distance (e.g., depth) (m)

[S.sub.v.f] Luminous flux responsivity (A/lm)

[S.sub.v,i] Illuminance responsivity (A/lx)

A Area (e.g., of an aperture) ([m.sup.2])

E Energy (J)

E Energy density (J/[m.sup.2])

[E.sub.v]: Illuminance (lx)

G: Thermal conductance (W/K W/K Watchkeeper )

[I.sub.v]: Luminous intensity (cd)

[K.sub.el]: Electrical calibration factor (J/V)

[K.sub.m]: Maximum spectral efficacy of radiation for photopic vision (683 lm/W)

[L.sub.v]: Luminance (cd/[m.sup.2])

R: Resistance (electrical) ([omega])

R: Ratio (e.g., power)

T: Temperature (K)

[T.sub.o]: Temperature, reference (K)

V: Voltage (V)

S([lambda]): Absolute spectral responsivity (A/W)

S(555): Absolute spectral responsivity at 555 nm (A/W)

[S.sub.n]([lambda]): Relative spectral responsivity (normalized to 1 at 555 nm)

E([lambda]): Spectral irradiance [W/([m.sup.2].nm)]

L([lambda]): Spectral radiance [W/([m.sup.2].sr.nm)]

V([lambda]): Spectral luminous efficiency function for photopic vision (originally, the Visibility function)

[alpha]: Absorptance (e.g., cavity); also:

[alpha]: Adsorption adsorption, adhesion of the molecules of liquids, gases, and dissolved substances to the surfaces of solids, as opposed to absorption, in which the molecules actually enter the absorbing medium (see adhesion and cohesion).  coefficient ([m.sup.-1])

[epsilon]: Cooling constant ([s.sup.-1])

[lambda]: Wavelength (nm)

[rho]: Reflectance (e.g., cavity or window); also:

[rho]: Density, absorber ([m.sup.-3])

[tau]: Transmittance (e.g., window)

[PHI]: Flux, optical (or radiant) (W) (historically, [[PHI].sub.c])

[[PHI].sub.v]: Luminous flux (lm)

[[PHI].sub.o]: Flux, incident optical (W)

[PHI]([lambda]): Spectral flux
This article is about the Digital signal processing term. For the term used in astronomy, which refers to a different measure, see Spectral flux density.


Spectral flux
, optical (or radiant) (W/nm)

8. Acronyms

ASR: Absolute spectral responsivity

CGPM: Conference Generale des Poids et Mesures (General Conference on Weights and Measures)

CIE: Commission Internationale de l'Eclairage (International Commission on Illumination)

CIPM: Comite International des Poids et Mesures (International Committee on Weights and Measures)

DoD: Department of Defense (U.S.)

DSC: Detector spectral comparitor

ECPR: Electrically calibrated pyroelectric radiometer

ECR ECR Efficient Consumer Response
ECR European Congress of Radiology
ECR Electron Cyclotron Resonance
ECR El Camino Real (Kings Highway; California)
ECR Electronic Cash Register
ECR East Coast Radio (South Africa) 
: Electrically calibrated radiometer

ESR: Electrical substitution radiometer

FASCAL: Facility for Automated Spectroradiometric Calibrations

FR: Filter radiometer

HACR: High-accuracy cryogenic radiometer (NIST)

HTBB: High-temperature blackbody

IR: Infrared

LOCR: Laser-optimized cryogenic radiometer (NIST)

NBS: National Bureau of Standards National Bureau of Standards: see National Institute of Standards and Technology.

National Bureau of Standards - National Institute of Standards and Technology


NPL: National Physical Laboratory (UK)

PTB: Physikalisch-Technische Bundesantalt

RF: Radio-frequency

RIN: Relative intensity noise

SI: Systeme international d'unites Sys·tème In·ter·na·tion·al d'U·ni·tés
n.
International System of Units.
 (International system of [metric] units)

SIRCUS: Spectral Irradiance and Radiance Calibrations using Uniform Sources

SURF: Synchrotron Ultraviolet Research Facility

TSD: Transfer standard detector

UV: Ultraviolet

VNIIOFI: All-Russian Institute for Optophysical Measurements

Acknowledgments

The dedication and achievements of the staff of the NIST Optical Technology Division over the last several decades are the reasons that this story can be told today. Space does not allow a complete accounting of my gratitude, but special mention is due for those most involved with the compilation of this report. Yoshi Ohno provided particular help and insight for the sections on photometry. Howard Yoon has been instrumental in the application of filtered radiometers and spectral source characterization. Tom Scott Tom Scott may refer to:
  • Tom Scott (businessman), UK businessman
  • Tom Scott (cartoonist), New Zealand cartoonist
  • Tom Scott (musician), jazz musician
  • Tom Scott (painter) (1854–1927), Scottish water-colourist.
, Leader of the Source and Detectors Group of the Optoelectronics Division, kindly supplied the excellent summary of the Group's laser power and energy work. Keith Lykke and Jonathan Hardis suggested many helpful improvements to the text.

About the author: Al Parr is a physicist and the Chief of the Optical Technology Division of the NIST Physics Laboratory. The Optical Technology Division has a broad mandate to provide accurate and uniform national measurement standards and support services support services Psychology Non-health care-related ancillary services–eg, transportation, financial aid, support groups, homemaker services, respite services, and other services  to advance the application of optical technologies for diverse industrial, governmental, and scientific use. The Division develops radiometric, photometric, spectroscopic spec·tro·scope  
n.
An instrument for producing and observing spectra.



spectro·scop
, and spectrophotometric measurement methods, standards, and data. The National Institute of Standards and Technology National Institute of Standards and Technology, governmental agency within the U.S. Dept. of Commerce with the mission of "working with industry to develop and apply technology, measurements, and standards" in the national interest.  is an agency of the Technology Administration, U.S. Department of Commerce.

(1.) In traditional notation, a subscript (1) In word processing and scientific notation, a digit or symbol that appears below the line; for example, H2O, the symbol for water. Contrast with superscript.

(2) In programming, a method for referencing data in a table.
 of [lambda] on radiometric quantities indicates their functional dependence on wavelength (e.g., [E.sub.[lambda]] for spectral irradiance and [V.sub.[lambda]] for the visibility function). In this paper, we use modern notation (e.g., E([lambda]) and V([lambda]), respectively).

(2.) The fact that 540 X [10.sup.12] Hz does not correspond exactly to a wavelength of 555 nm in air causes for factor [K.sub.m] to be slightly larger than 683 lm/W. See Ref. (2) for details. For the purpose of this paper, the differences are insignificant.

9. References

(1.) K. S. Gibson and E. P. T. Tyndall, Visibility of radiant energy radiant energy
n.
Energy transferred by radiation, especially by an electromagnetic wave.


radiant energy
Noun
 Scientific Papers of the Bureau of Standards Bureau of Standards
 since 1988 U.S. National Institute of Standards and Technology (NIST)

Agency of the U.S. Department of Commerce responsible for the standardization of weights and measures, timekeeping, and navigation.
 19, 131-191 (1923).

(2.) Commission Internationale de L'Eclairage, Sixieme Session, Geneve--Juillet 1924, Recucil des Travaux et Compte Rendu Compte´ ren`du

1. A report of an officer or agent.
 des Seances, Cambridge University Press Cambridge University Press (known colloquially as CUP) is a publisher given a Royal Charter by Henry VIII in 1534, and one of the two privileged presses (the other being Oxford University Press). , Cambridge (1926). See also: The Basis of Physical Photometry, CIE Publ. 18.. (Commission Internationale de L'Eclairage, Paris, 1983), ISBN ISBN
abbr.
International Standard Book Number


ISBN International Standard Book Number

ISBN n abbr (= International Standard Book Number) → ISBN m 
 92-9034018-5. (Currently available through the U.S. National Committee of the CIE, c/o Mr. Thomas M. Lemons, TLA-Lighting Consultants, Inc., 7 Pond Street, Salem, MA 01970-4819).

(3.) J. W. T. Walsh, Photometry, 3rd Ed., Constable and Company, London (1958).

(4.) C. DeCusatis, Ed., Handbook of Applied Photometry, SpringerVerlag, New York New York, state, United States
New York, Middle Atlantic state of the United States. It is bordered by Vermont, Massachusetts, Connecticut, and the Atlantic Ocean (E), New Jersey and Pennsylvania (S), Lakes Erie and Ontario and the Canadian province of
 (1997), ISBN 1-56396-416-3.

Note: Refs. (3) and (4) contain many references to the early literature and give the history and detail of all aspects of photometry. This paper is not intended to be an exhaustive review of the literature in photometry and radiometry, but hopefully it will act as a starting point Noun 1. starting point - earliest limiting point
terminus a quo

commencement, get-go, offset, outset, showtime, starting time, beginning, start, kickoff, first - the time at which something is supposed to begin; "they got an early start"; "she knew from the
 for the more interested reader.

(5.) Le Systeme International d'Unites (SI), 7e Edition, BIPM BIPM - Bureau International des Poids et Mesures , Pavillon de Breteuil Pavillon de Breteuil is a building located in Sèvres, France near Paris. It was inaugurated by Louis XIV in 1672. It is in the park of the former royal Château de Saint-Cloud, which was destroyed in 1870. , F-92312 Sevres Cedex, France (1998), ISBN 92-822-2154-7. (Published electronically as http://www.bipm.fr/pdf/si-brochure.pdf.) For a discussion of the use of the CIE V([lambda]) function in determining photometric quantities, see the monograph mon·o·graph  
n.
A scholarly piece of writing of essay or book length on a specific, often limited subject.

tr.v. mon·o·graphed, mon·o·graph·ing, mon·o·graphs
To write a monograph on.
: G. Wyszecki, W. R. Blevin, K. G. Kessler, and K. D. Mielenz, Principles Governing Photometry, BIPM, Pavillon de Breteuil, F-92312 Sevres Cedex, France (1983).

(6.) W. R. Blevin and B. Steiner, Redefinition of the Candela and Lumen, Metrologia 11, 97-104 (1975).

(7.) The literature on silicon detectors is too extensive to review here. For a review of the early NIST (NBS) efforts, see: J. Geist, M. A. Lind, A. R. Schaefer, and E. F. Zalewski, Spectral Radiometry: A New Approach Based on Electro-Optics, NBS Technical Note 954 (1977). (Available from U.S. Government Printing Office, Washington, DC 20402-9325. See http://nvl.nist.gov/pub/nistpubs/orderpub.htm for details.)

For a discussion of solid state devices, see the following publications and the references therein: W. L. Budde, Optical Radiation Measurements: Physical Detectors of Optical Radiation, Vol 4, Academic Press, New York (1983), ISBN 0-12-304904-0. G. H. Rieke, Detection of Light: From the Ultraviolet to the Submillimeter, Cambridge University Press, Cambridge (1994), ISBN 0-521-41028-2. E. L. Dereniak and D. G. Crowe, Optical Radiation Detectors, John Wiley John Wiley may refer to:
  • John Wiley & Sons, publishing company
  • John C. Wiley, American ambassador
  • John D. Wiley, Chancellor of the University of Wisconsin-Madison
  • John M. Wiley (1846–1912), U.S.
 & Sons, New York (1984), ISBN 0-471-89797-3. R. W. Boyd, Radiometry and the Detection of Optical Radiation, John Wiley & Sons, New York (1983), ISBN 0-471-86188-X. R. H. Kingston, Optical Sources, Detectors, and Systems, Academic Press, San Diego San Diego (săn dēā`gō), city (1990 pop. 1,110,549), seat of San Diego co., S Calif., on San Diego Bay; inc. 1850. San Diego includes the unincorporated communities of La Jolla and Spring Valley. Coronado is across the bay.  (1995), ISBN 0-12-408655-1.

(8.) R. Korde and J. Geist, Quantum efficiency stability of silicon photodiodes, Appl. Opt. 26, 5284-5290 (1987).

(9.) E. M. Gullikson, R. Korde, L. R. Canfield can·field  
n. Games
A form of solitaire.



[After Richard Albert Canfield (1855-1914), American gambler.]

Noun 1.
, and R. E. Vest, Stable silicon photodiodes for absolute intensity measurements in the VUV VUV

In currencies, this is the abbreviation for the Vanuatu Vatu.

Notes:
The currency market, also known as the Foreign Exchange market, is the largest financial market in the world, with a daily average volume of over US $1 trillion.
 and soft x-ray region, J. Electron Spectrose. Relat. Phenom phe·nom  
n. Slang
A phenomenon, especially a remarkable or outstanding person.
. 80, 313-316 (1996).

(10.) E. F. Zalewski and J. Geist, Silicon photodiode absolute spectral response self-calibration, Appl. Opt. 19, 1214-1216 (1980).

(11.) J. Geist, Silicon photodiode self-calibration, in Theory and Practices of Radiation Thermometry thermometry

Science of measuring the temperature of a system or the ability of a system to transfer heat to another system. Temperature measurement is important to a wide range of activities, including manufacturing, scientific research, and medicine.
, Cp. 14, D. P. DeWitt and G. D. Nutter, eds., John Wiley & Sons (1988), ISBN 0-471-81018-8.

(12.) B. F. Zalewski and C. R. Duda, Silicon photodiode device with 100 % external quantum efficiency, Appl. Opt. 22, 2867-2873 (1983).

(13.) G. Eppeldauer, Temperature monitored/controlled silicon photodiodes for standardization standardization

In industry, the development and application of standards that make it possible to manufacture a large volume of interchangeable parts. Standardization may focus on engineering standards, such as properties of materials, fits and tolerances, and drafting
, in Surveillance Technologies, S. Gowrinathan, R. J. Mataloni, and S. J. Schwartz, eds., Proc. SPIE SPIE International Society for Optical Engineering
SPIE Society of Photo-Optical Instrumentation Engineers
SPIE Source Path Isolation Engine
SPIE Special Purpose Insertion Extraction
SPIE Software Process Improvement Experimentation
SPIE Standard Protocols in Effect
 1479, 71-77 (1991).

(14.) G. Eppeldauer and J. E. Hardis, Fourteen-decade photocurrent measurements with large-area silicon photodiodes at room temperature, Appl. Opt. 30, 3091-3099 (1991).

(15.) W. Budde, Multidecade linearity measurements on Si photodiodes, Appl. Opt. 18, 1555-1558 (1979).

(16.) F. Hengstberger, Absolute Radiometry, Academic Press, San Diego (1989), ISBN 0-12-340810-5.

(17.) W. W. Coblentz, Various modifications of thermopiles having a continuous absorbing surface, Bull. Bur. Stand. 11, 131-187 (1914). W. W. Coblentz and W. B. Emerson, Studies of instruments for measuring radiant energy in absolute value: An absolute thermopile thermopile: see thermoelectricity. , Bull. Bur. Stand. 12, 503-551 (1915).

(18.) W. W. Coblentz, Present status of the determination of the constant of total radiation from a black body, Bull. Bur. Stand. 12, 553-582 (1915).

(19.) D. C. Ginnings and M. L. Reilly, Calorimetric cal·o·rim·e·ter  
n.
1. An apparatus for measuring the heat generated by a chemical reaction, change of state, or formation of a solution.

2.
 measurement of thermodynamic temperatures above 0 [degrees]C using total blackbody radiation blackbody radiation

The electromagnetic radiation that a perfect blackbody would give off at a given temperature. A warm blackbody would emit radiation with a higher average frequency than a cooler one.

Noun 1.
, In Temperature: Its Measurement and Control in Science and Industry, H. H. Plumb, ed., Vol. 4, Pt. I, 339-348, Instrument Society of America, Pittsburgh (1972).

(20.) C. R. Yokley, Long-wave infrared testing at NBS, in Applications of Optical Metrology: Techniques and Measurements II, R. C. Harney, ed., Proc. SPIE 416, 2-8 (1983).

(21.) C. R. Yokley, A Radiometric Calibration Facility for Low Temperature Blackbodies, Final Report to the NASA NASA: see National Aeronautics and Space Administration.
NASA
 in full National Aeronautics and Space Administration

Independent U.S.
 Lyndon B. Johnson Space Center The Lyndon B. Johnson Space Center (JSC) is the National Aeronautics and Space Administration's center for human spaceflight activities. The center consists of a complex of 100 buildings on 1,620 acres [1] , for the Earth Resources Aircraft Program (EUAP) Infrared Sensors Study, under Order No. T-9550C (November 23, 1976).

(22.) T. J. Quinn and J. E. Martin, A radiometric determination of the Stefan-Boltzmann constant and thermodynamic temperatures between--40 [degrees]C and +100 [degrees]C, Philos. Trans. R. Soc. London Ser. A 316, 85-189 (1985).

(23.) J. E. Martin, N. P. Fox, and P. J. Key, A cryogenic radiometer of absolute radiometric measurements, Metrologia 21, 147-155 (1985).

(24.) Author's notes from the last CCPR CCPR Covenant on Civil and Political Rights
CCPR California Center for Population Research
CCPR Central Council for Physical Recreation
CCPR Consultative Committee on Photometry and Radiometry
CCPR Calvert County Parks and Recreation
CCPR Co-Channel Power Ratio
 meeting in Paris in the spring of 1999 indicates that at least 15 of the 19 participants were using or starting to use a cryogenic radiometer for their measurement base.

(25.) T. R. Gentile, J. M. Houston, J. E. Hardis, C. L. Cromer, and A. C. Parr, National Institute of Standards and Technology high-accuracy cryogenic radiometer, Appl. Opt. 35, 1056-1068 (1996).

(26.) Uncertainties in this paper follow the nomenclature nomenclature /no·men·cla·ture/ (no´men-kla?cher) a classified system of names, as of anatomical structures, organisms, etc.

binomial nomenclature
 and suggestions in B. N. Taylor and C. E. Kuyatt, Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST Technical Note 1297, 2nd Ed. (1994). (See Ref. (7) for availability.) (Available electronically at http://physics.nist.gov/Document/tn1297.pdf.)

(27.) T. R. Gentile, J. M. Houston, and C. L. Cromer, Realization of a scale of absolute spectral response using the NIST High-Accuracy Cryogenic Radiometer, Appl. Opt. 35, 4392-4403 (1996). T. C. Larason, S. S. Bruce, and C. L. Cromer, The NIST high accuracy scale for absolute spectral response from 406 nm to 920 nm, J. Res. Natl. Inst. Stand. Technol. 101, 133-140 (1996).

(28.) N. P. Fox, Trap detectors and their properties, Metrologia 28, 197-202 (1991).

(29.) L. P. Boivin and K. Gibb, Monochromator-based cryogenic radiometry at the NRC NRC
abbr.
1. National Research Council

2. Nuclear Regulatory Commission

Noun 1. NRC - an independent federal agency created in 1974 to license and regulate nuclear power plants
, Metrologia 32, 565-570 (1996).

(30.) A. Lau-Frambs, U. Kroth, H. Rabus, E. Tegeler, G. Ulm, and B. Wende, First results with the new PTB cryogenic radiometer for the vacuum-ultraviolet spectral range, Metrologia 32, 571-574 (1996).

(31.) P.-S, Shaw, K. R. Lykke, R. Gupta, T. R. O'Brian, U. Arp, H. H. White, T. B. Lucatorto, J. L. Dehmer, and A. C. Parr, New ultraviolet radiometry beamline at the Synchrotron Ultraviolet Radiation Facility at NIST, Metrologia 35, 30 1-306 (1998). P.-S. Shaw, K. R. Lykke, R. Gupta, T. R. O'Brian, U. Arp, H. H. White, T. B. Lucatorto, J. L. Dehmer, and A. C. Parr, Ultraviolet radiometry with synchrotron radiation and cryogenic radiometry, Appl. Opt. 38, 18-28 (1999).

(32.) T. C. Larason, S. S. Bruce, and A. C. Parr, Spectroradiometric Detector Measurements: Part I--Ultraviolet Detectors and Part II--Visible to Near-Infrared Detectors, Natl. Inst. Stand. Technol. Spec. Publ. 250-41 (1998). (See Ref. (7) for availability.)

(33.) A. L. Migdall and G. P. Eppeldauer, Spectroradiometric Detector Measurements: Part III--Infrared Detectors, Natl. Inst. Stand. Technol. Spec. Publ. 250-42 (1998). (See Ref. (7) for availability.)

(34.) K. R. Lykke, P.-S. Shaw, L. M. Hanssen, and G. P. Eppeldauer, Development of a monochromatic, uniform source facility for calibration of radiance and irradiance detectors from 0.2 [micro]m to 12 [micro]m, Metrologia 36, 141-146 (1999).

(35.) J. B. Fowler, R. S. Durvasula, and A. C. Parr, High-accuracy aperture-area measurement facilities at the National Institute of Standards and Technology, Metrologia 35, 497-500 (1998).

(36.) International Lighting Vocabulary, CIE Publ. 17.4 (Commission Internationale de L'Eclairage, Vienna, 1987), ISBN 3-900734-07-0. (See Ref. (2) for availability.)

(37.) Methods of Characterizing the Performance of Radiometers and Photometers, CIE Publ. 53 (Commission Internationale de L'Eclairage, Paris, 1982), ISBN 92-9034-053-3. (See Ref. (2) for availability.)

(38.) C. L. Cromer, G. Eppeldauer, J. E. Hardis, T. C. Larason, and A. C. Parr, National Institute of Standards and Technology detector-based photometric scale, Appl. Opt. 32,2936-2948 (1993). C. L. Cromer, G. Eppeldauer, J. E. Hardis, T. C. Larason, Y. Ohno, and A. C. Parr, The NIST detector-based luminous intensity scale, J. Res. Natl. Inst. Stand. Technol. 101, 109-132 (1996).

(39.) Y. Ohno, Photometric Calibrations, Natl. Inst. Stand. Technol. Spec. Publ. 250-37 (1997). (See Ref. (7) for availability.)

(40.) H. Yoon, private communication on NIST work in progress.

(41.) The Measurement of Luminous Flux, CIE Publ. 84 (Commission Internationale de L'Eclairage, Vienna, 1989), ISBN 3-900734-21-6. (See Ref. (2) for availability.)

(42.) Y. Ohno, New method for realizing a total luminous flux scale using an integrating sphere with an external source, J. Illum. Eng. Soc. 24, 106-115 (1995).

(43.) Y. Ohno, Integrating sphere simulation--application to total flux scale realization, Appl. Opt. 33, 2637-2647 (1994).

(44.) Y. Ohno, Realization of NIST 1995 luminous flux scale using the integrating sphere method, J. Illum. Eng. Soc. 25, 13-22 (1996).

(45.) W. E. Sumpner, The diffusion of light, Phys. Soc. Proc. 12, 10-29 (1892).

(46.) F. E. Nicodemus, Ed., Self-Study Manual on Optical Radiation Measurements, NBS Technical Notes 910-1 through 910-8 (1976-1985). (See Ref. (7) for availability. On a limited basis some of these documents may be available from NIST, Optical Technology Division, 100 Bureau Drive Stop 8440, Gaithersburg, MD 20899-8440.)

(47.) R. Siegel and J. R. Howell, Thermal Radiation Heat Transfer, 3rd Ed., Hemisphere Publishing Corp., Washington (1992), ISBN 0-89116-271-2.

(48.) A. C. Parr, A National Measurement System for Radiometry, Photometry, and Pyrometry Based upon Absolute Detectors, NIST Technical Note 1421 (1996). (See Ref. (7) for availability.) See also http://physics.nist.gov/Pubs/TN142l/contents.html.

(49.) J. H. Walker, R. D. Saunders, J. K. Jackson, and D. A. McSparron, Spectral Irradiance Calibrations, Nat. Bur. Stand. (U.S.) Spec. Publ. 250-20 (1987). (See Ref. (7) for availability.)

(50.) J. H. Walker, R. D. Saunders, and A. T. Hattenburg, Spectral Radiance Calibrations, Natl. Bur. Stand. (U.S.) Spec. Publ. 250-1 (1987). (See Ref. (7) for availability.)

(51.) K. D. Mielenz, R. D. Saunders, and J. B. Shumaker, Spectroradiometric determination of the freezing temperature of gold, J. Res. Natl. Inst. Stand. Technol. 95, 49-67 (1990).

(52.) B. C. Johnson, C. L. Cromer, R. D. Saunders, G. Eppeldnuer, J. Fowler, V. I. Sapritsky, and G. Dezsi, A method of realizing spectral irradiance based on an absolute cryogenic radiometer, Metrologia 30, 309-315 (1993).

(53.) B. K. Tsai, B. C. Johnson, R. D. Saunders, and C. L. Cromer, Comparison of filter radiometer spectral responsivity with the NIST spectral irradiance and illuminance scales, Metrologia 32, 473-477 (1996).

(54.) H. W. Yoon and C. E. Gibson, Determination of radiance temperatures using detectors calibrated for absolute spectral power response, in Proceeding of TEMPMEKO '99, The 7th International Symposium on Temperature and Thermal Measurements in Industry and Science, Vol. 2, J. F. Dubbeldam and M. J. de Groot, eds., Delft Delft (dĕlft), city (1994 pop. 91,941), South Holland prov., W Netherlands. It has varied industries and is noted for its ceramics (china, tiles, and pottery) known as delftware. Founded in the 11th cent. , Netherlands, June 1999, pp. 737-742.

(55.) N. J. Harrison, N. P. Fox, P. Sperfeld, J. Metzdorf, B. B. Khlevnoy, R. I. Stolyarevskaya, V. B. Khromchenko, S. N. Mekhontsev, V. I. Shapoval, M. F. Zelener, and V. I. Sapritsky, International comparison of radiation-temperature measurements with filtered detectors over the temperature range 1380 to 3100 K, Metrologia 35, 283-288 (1998).

(56.) C. E. Gibson, B. K. Tsai, and A. C. Parr, Radiance Temperature Calibrations, Natl. Inst. Stand. Technol. Spec. Publ. 250-43 (1998). (See Ref. (7.) for availability.)

(57.) P. Sperfeld, J. Metezdorf, N. J. Harrison, N. P. Fox, B. B. Khlevnoy, V. B. Khromchenko, S. N. Mekhontsev, V. I. Shapoval, M. F. Zelener, and V. I. Sapritsky, Investigation of high-temperature black body BB3300, Metrologia 35, 419-422 (1998)

(58.) R. P. Madden mad·den  
v. mad·dened, mad·den·ing, mad·dens

v.tr.
1. To make angry; irritate.

2. To drive insane.

v.intr.
To become infuriated.
, A status report on the SURF II synchrotron radiation facility at NBS, Nucl. Instrum. Methods 172, 1-8 (1980).

(59.) A discussion of the theory of synchrotron radiation is outside the scope of this paper. The reader is referred to standard textbooks and the references therein for details on this subject: E.-E. Koch, Handbook on Synchrotron Radiation, Vol. la, D. E. Eastman and Y. Farge, general editors, North-Holland Publishing Co., Amsterdam (1983), ISBN 0-444-86709-0; H. Winick and S. Doniach, Synchrotron Radiation Research, Plenum In a building, the space between the real ceiling and the dropped ceiling, which is often used as an air duct for heating and air conditioning. It is also filled with electrical, telephone and network wires. See plenum cable.  Press, New York (1980), ISBN 0-306-40363-3.

(60.) J. Schwinger, On the classical radiation of accelerated electrons, Phys. Rev. 75, 1912-1925 (1949). Schwinger's paper was one of the first theoretical treatments of synchrotron radiation. Discussions of the fundamentals of the nature of the radiation from accelerated charged particles can be found in J. D. Jackson
This page is on the physicist. For the basketball player see J. D. Jackson (basketball).


John David Jackson (born 1925) is a Canadian-American physics professor emeritus at the University of California, Berkeley and a senior staff physicist at
, Classical Electrodynamics electrodynamics, study of phenomena associated with charged bodies in motion and varying electric and magnetic fields (see charge; electricity); since a moving charge produces a magnetic field, electrodynamics is concerned with effects such as magnetism, , 3rd Ed., John Wiley & Sons, New York (1998). See also books cited in Ref. (59.).

(61.) For details of the beamlines and available instruments at SURF III, consult the SURF Users' Guide. (Available from NIST, Electron and Optical Physics Division, 100 Bureau Drive Stop 8410, Gaithersburg, MD 20899-8410.)

(62.) L. R. Hughey and A. R. Schaefer, Reduced absolute uncertainty in the irradiance of SURF II and instrumentation for measuring linearity of x-ray, xuv, and uv detectors, Nucl. Instrum. Methods 195, 367-370 (1982).

(63.) D. A. Jennings, Calorimetric measurement of pulsed laser output energy, IEEE (Institute of Electrical and Electronics Engineers, New York, www.ieee.org) A membership organization that includes engineers, scientists and students in electronics and allied fields.  Trans. Instrum. Meas. IM-15(4), 161-164 (1966).

(64.) R. J. Phelan and A. R. Cook, Electrically calibrated pyroelectric optical-radiation detector, Appl. Opt. 12, 2494-2500 (1973).

(65.) C. A. Hamilton, G. W. Day, and R. J. Phelan, An Electrically Calibrated Pyroelectric Radiometer System, NBS Technical Note 678 (1976). (See Ref. (7) for availability.)

(66.) J. Geist and W. R. Blevin, Chopper-stabilized null A character that is all 0 bits. Also written as "NUL," it is the first character in the ASCII and EBCDIC data codes. In hex, it displays and prints as 00; in decimal, it may appear as a single zero in a chart of codes, but displays and prints as a blank space.  radiometer based upon an electrically calibrated pyroelectric detector, Appl. Opt. 12, 2532-2535 (1973).

(67.) E. D. West, Data Analysis for Isoperibol Laser Calorimetry calorimetry (kăl'ərĭm`ətrē), measurement of heat and the determination of heat capacity , NBS Technical Note 396 (1971). (See Ref. (7.) for availability.)

(68.) E. D. West, W. E. Case, A. L. Rasmussen, and L. B. Schmidt, A reference calorimeter for laser energy measurements, J. Res. Natl. Bur. Stand. (U.S.) 76A, 13-26 (1972).

(69.) D. J. Livigni, C. L. Cromer, T. R. Scott, B. C. Johnson, and Z. M. Zhang, Thermal characterization of a cryogenic radiometer and comparison with a laser calorimeter, Metrologia 35, 819-827 (1998). Continued: Metrologia 36, 161 (1999).

(70.) E. D. West and W. E. Case, Current Status of NBS Low-Power Laser Energy Measurement, in Proc. CPEM CPEM Conference on Precision Electromagnetic Measurements , Jul 1-5, 1974, London, England, IEEE Trans. Instrum. Meas. IM-23, 422-425 (1974).

(71.) Z. M. Zhang, D. J. Livigni, R. D. Jones, and T. R. Scott, Thermal modeling and analysis of laser calorimeters, J. Thermophys. Heat Transfer 10, 350-356 (1996).

(72.) E. D. West and K. L. Churney, Theory of isoperibol calorimetry for laser power and energy measurements, J. Appl. Phys. 41, 2705-2712 (1970).

(73.) E. D. West and L. B. Schmidt, A System for Calibrating Laser Power Meters for the Range 5-1000 W, NBS Technical Note 685 (1977). (See Ref. (7.) for availability.)

(74.) R. L. Smith, T. W Russell, W. E. Case, and A. L. Rasmussen, A calorimeter for high-power CW lasers, IEEE Trans. Instrum. Meas. IM-21, 434-438 (1972).

(75.) G. E. Chamberlain, P. A. Simpson, and R. L. Smith, Improvements in a calorimeter for high-power CW lasers, IEEE Trans. Instrum. Meas. IM-27, 81-86 (1978).

(76.) D. L. Franzen and L. B. Schmidt, Absolute reference calorimeter for measuring high power laser pulses, Appl. Opt. 15, 3115-3122 (1976).

(77.) R. W. Leonhardt, Calibration Service for Laser Power and Energy at 248 nm, NIST Technical Note 1394 (1998). (See Ref. (7.) for availability.)

(78.) M. L. Dowell, C. L. Cromer, R. W. Leonhardt, and T. R. Scott, Deep ultraviolet (DUV DUV Deep Ultraviolet
DUV Data-Under-Voice
DUV Design Under Verification
) laser metrology for semiconductor photolithography, in Proc. 1998 Intl. Conf. on Characterization and Metrology of ULSI (Ultra Large Scale Integration) More than one million transistors on a chip. See SSI, MSI, LSI and VLSI.  Tech., May 23-27, 1998, Gaithersburg, MD, pp. 530-541.

(79.) I. Vayshenker, X. Li, and T. R. Scott, Optical power meter calibration using tunable laser A laser that can change its frequency over a given range. In time, tunable lasers are expected to be capable of switching frequencies on a packet by packet basis.  diodes, in Proc. NatI. Conf. Stds. Labs., Jul 31-Aug 4, 1994, Chicago, IL (1994) pp. 363-372.

(80.) R. D. Jones and T. R. Scott, Error propagation The transmission (spreading) of signals from one place to another.  in laser beam spatial parameters, Opt. Quant Quant

A person with numerical and computer skills who carries out quantitative analyses of companies.


quant

A person who has strong skills in mathematics, engineering, or computer science, and who applies those skills to the securities
. Electron. 26, 25-34 (1994).

(81.) R. D. Jones and T. R. Scott, Characterization of a clipped Gaussian beam, in Beam Control, Diagnostics, Standards, and Propagation, L. W. Austin, A. Giesen, D. H. Leslie, and H. Weichel, eds., Proc. SPIE 2375, 360-374 (1995).

(82.) G. E. Obarski and P. D. Hale, How to measure relative intensity noise in lasers, Laser Focus World 35(5), 273-277 (1999).

(83.) G. E. Obarski, and R. D. Jones, Relative intensity noise correlates with beam profile in an [LP.sub.11] mode vertical-cavity surface-emitting laser The vertical-cavity surface-emitting laser (VCSEL; [v'ɪxl]) is a type of semiconductor laser diode with laser beam emission perpendicular from the top surface, contrary to conventional edge-emitting semiconductor lasers (also in-plane , Summaries of Papers Presented at the Conference on Lasers and Electro-Optics, 1997 (CLEO '97), Vol. 11, 228-229 (May 18-23, 1997), ISBN 0-7803-4125-2.

(84.) D. J. Livigni and X. Li, Spatial uniformity of optical detector responsivity, in Proc., Natl. Conf. Stds, Labs., Jul 31-Aug 4, 1994, Chicago, IL (1994) pp. 337-352.

(85.) I. Vayshenker, S. Yang, X. Li, and T. R. Scott, Nonlinearity of optical fiber power meters, in Technical Digest--Symposium on Optical Fiber Measurements 1966, Natl. Inst. Stand. Technol. Spec. Publ. 905 (1996) pp. 101-104.

(86.) J. H. Lehman, Pyroelectric trap detector for spectral responsivity measurements, Opt. Photon. News (Suppl.) 8(11), 35-36 (1997). Appl. Opt. 36, 9117-9118 (1997).

(87.) P. D. Hale, C.-M. Wang, R. Park, and W. Y. Lau, A transfer standard for measuring photoreceiver frequency response, J. Lightwave Tech. 14(11), 2457-2466 (1996).

(88.) T. R. Scott, R. J. Rockwell, Jr., and P. Batra, Optical density measurement of laser eye protection materials, in Proc. 1992 Intl. Laser Safety Conf., Dec 1-4, 1992, Cincinnati, OH (1993) Sec. 8-7 to 8-16.

[Figure 1 omitted]

[Figure 7 omitted]

[Figure 8 omitted]

[Figure 10 omitted]

[Figure 15 omitted]

[Figure 16 omitted]

[Figure 20 omitted]

[Figure 21 omitted]

[Figure 24 omitted]
Table 1

Corresponding radiometric and photometric quantitites. The radiometric
quantities, on the left, are shown with their usual symbols and SI Units
of measure. When integrated spectrally using V ([lambda]) weighting, one
obtains the photometric counterpart, on the right

Radiometric quantity   Symbol         Unit              Unit

Radiant energy           Q             J                lm s
Radiant flux (power)  P, [PHI]         W                 lm
Irradiance               E        W/[m.sup.2]      lm/[m.sup.2]=lx
Radiance                 L      W/([m.sup.2] sr)  lm/([m.sup.2] sr)
Radiant intensity        I            W/sr            lm/sr=cd

Radiometric quantity     Symbol      Photometric quantity

Radiant energy          [Q.sub.v]    Luminous energy
Radiant flux (power)  [[PHI].sub.v]  Luminous flux
Irradiance              [E.sub.v]    Illuminance
Radiance                [L.sub.v]    Luminance
Radiant intensity       [I.sub.v]    Luminous intensity

J=joule, lm=lumen, s=second, W=watt, m=meter, sr=steradian, cd=candela.
Table 2.

The present and expected relative expanded uncertainties (coverage
factor k = 2) for calibrations of spectral irradiance E ([lambda]) at
typical wavelengths

Wavelength  Present expanded  Expected expanded
   (nm)       uncertainty        uncertainty
                  (%)                (%)

   250            1.8                1.0
   350            1.1                0.6
   655            0.9                0.5
   900            1.1                0.5
   1600           1.4                0.5
   2400           4.4                1.1
Table 3.

The present and expected relative expanded uncertainties (coverage
factor k = 2) for calibrations of spectral radiance L([lambda]) at
typical wavelengths

Wavelength  Present expanded  Expected expanded
   (nm)       uncertainty        uncertainty
                  (%)                (%)

   225            1.5                1.2
   250            1.3                0.8
   350            1.0                0.6
   650            0.6                0.3
   900            0.6                0.3
   1550           0.5                0.3
   2400           0.4                0.3
COPYRIGHT 2001 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2001, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Parr, Albert C.
Publication:Journal of Research of the National Institute of Standards and Technology
Geographic Code:1USA
Date:Jan 1, 2001
Words:18687
Previous Article:The Kelvin and temperature measurements.
Next Article:Materials research with neutrons at NIST.(National Institute of Standards and Technology)
Topics:



Related Articles
Clearing the fog about fogging effects of liquid fire retardants in flexible foam.
New DSM Database Helps Select Stock Shapes.(Brief Article)
NIST DEVELOPS PHOTOMETRIC AND COLORIMETRIC STANDARDS FOR LEDS.(Brief Article)
Improved radiometric standards lead to high-accuracy measurements. (News Briefs).(Brief Article)
NIST publishes 2001 Edition of SP 330. (News Briefs).(Brief Article)
NIST develops an improved modular radiometer/photometer system. (News Briefs).(National Institute of Standards and Technology)
Development of a tunable LED-based colorimetric source.
Self-Study Manual now available on Compact Disc. (General Developments).(Brief Article)
NIST characterizes NOAA instrument for measuring global chlorophyll concentrations. (General Developments).
Radiometric measurement comparison on the integrating sphere source used to calibrate the moderate resolution imaging spectroradiometer (MODIS) and...

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles