The analysis of particles at low accelerating voltages ([less than or equal to] 10 kV) with energy dispersive x-ray spectroscopy (EDS).


In recent years, there have been a series of advancements in electron beam A stream of electrons, or electricity, that is directed towards a receiving object. See electron beam imaging and electron beam lithography.  instruments and x-ray detectors which may make it possible to improve significantly the quality of results from the quantitative electron-probe analysis of individual particles. These advances include: (1) field-emission gun electron beam instruments such as scanning electron microscopes scan·ning electron microscope
n. Abbr. SEM
An electron microscope that forms a three-dimensional image on a cathode-ray tube by moving a beam of focused electrons across an object and reading both the electrons scattered by the object and
 (FEG-SEMs) that have high brightness electron guns A device that creates a fine beam of electrons that is focused on a phosphor screen in a CRT.  with excellent performance at low beam low beam
n.
The beam of a vehicle's headlight that provides short-range illumination.

Noun 1. low beam - the beam of a car's headlights that provides illumination for a short distance
 energies, [E.sub.0] [less than or equal to] 10 keV and (2) high-resolution energy-dispersive x-ray spectrometers x-ray spectrometer
n.
A spectrometer using x-rays to separate the chemical constituents of a substance into their characteristic spectral lines for identification and determination of their concentration.
, like the microcalorimeter detector, that provide high-resolution (< 10 eV) parallel x-ray collection. These devices make it possible to separate low energy (<4 keV) x-ray lines including the K lines of carbon, nitrogen and oxygen and the L and M lines for elements with atomic numbers atomic number, often represented by the symbol Z, the number of protons in the nucleus of an atom, as well as the number of electrons in the neutral atom. Atoms with the same atomic number make up a chemical element.  in the range of 25 to 83. In light of these advances, this paper investigates the possibility of using accelerating voltages [less than or equal to] 10 kV, as a method to improve the accuracy of elemental analysis Elemental analysis is a process where a sample of some material (e.g., soil, waste or drinking water, bodily fluids, minerals, chemical compounds) is analyzed for its elemental and sometimes isotopic composition.  for micrometer-sized particles.

Key words: electron probe analysis; low voltage Low voltage is an electrical engineering term that broadly identifies safety considerations of an electricity supply system based on the voltage used. While different definitions exist for the exact voltage range covered by "low voltage", the most commonly used ones include "mains  analysis; particle analysis; scanning electron microscopy electron microscopy

Technique that allows examination of samples too small to be seen with a light microscope. Electron beams have much smaller wavelengths than visible light and hence higher resolving power.
; x-ray microanalysis microanalysis /mi·cro·anal·y·sis/ (-ah-nal´i-sis) the chemical analysis of minute quantities of material.

microanalysis

the chemical analysis of minute quantities of material.
.

1. Introduction

In classical electron probe analysis schemes employing either a ZAF ZAF South Africa (ISO Country code)
ZAF Zambia Air Force
ZAF Zombie Army Forums
ZAF Zero Alignment Feature
ZAF Zombie Alliance Force (gaming group) 
, bulk-sample [phi]([rho]z), or Bence-Albee approach, both sample and standard are assumed to be infinitely thick with respect to the penetration of the electron beam and have flat polished surfaces. For such a semi-infinite plate, the corrections for the interaction of the electron beam with the sample and the subsequent x-ray emission can be calculated from simple geometric relationships. In the quantitative analysis Quantitative Analysis

A security analysis that uses financial information derived from company annual reports and income statements to evaluate an investment decision.

Notes:
 of particles, the size and shape of the particle often cannot be controlled or accurately measured which results in two first-order particle effects that influence the generation and measurement of x rays from these samples. The first particle effect is the result of the finite size (mass) of the particle. The mass effect is related to the elastic scattering In scattering theory and in particular in particle physics, elastic scattering is one of the specific forms of scattering. In this process, the energy of the incident photon or particle (electron, positron, or neutron) is conserved and its propagating direction is changed by the  of the electrons and is strongly affected by the average atomic number of the particle. The mass effect is important when the particle size Particle size, also called grain size, refers to the diameter of individual grains of sediment, or the lithified particles in clastic rocks. The term may also be applied to other granular materials.  is smaller than the range of the primary electron beam so that a significant fraction of the beam escapes the particle before exciting x rays. The second particle effect is the result of x-ray absorption, which is dependant on Adj. 1. dependant on - determined by conditions or circumstances that follow; "arms sales contingent on the approval of congress"
contingent on, contingent upon, dependant upon, dependent on, dependent upon, depending on, contingent
 the shape of the analyzed an·a·lyze  
tr.v. an·a·lyzed, an·a·lyz·ing, an·a·lyz·es
1. To examine methodically by separating into parts and studying their interrelations.

2. Chemistry To make a chemical analysis of.

3.
 particle. In the analysis of most particles, the x-ray emergence angle and therefore the absorption path length cannot be determined as it can for polished specimens. The magnitude of this effect is largest when there is high absorption as is typically the case for "soft" x rays from elements like Al or Si that have energies less than about 2 keV (1,2). (A third particle effect due to secondary fluorescence fluorescence (flrĕs`əns), luminescence in which light of a visible color is emitted from a substance under stimulation or excitation by light or other forms of electromagnetic  is considered at most a second order effect and is not covered not covered Health care adjective Referring to a procedure, test or other health service to which a policy holder or insurance beneficiary is not entitled under the terms of the policy or payment system–eg, Medicare. Cf Covered.  in this paper.)

Over the years, several researchers have developed correction procedures to minimize the particle effects and reduce the uncertainties associated with particle analysis. These procedures range from simple normalization In relational database management, a process that breaks down data into record groups for efficient processing. There are six stages. By the third stage (third normal form), data are identified only by the key field in their record.  of the elemental elemental

emanating from or pertaining to elements.


elemental diet
see elemental diet.
 weight fractions obtained with conventional bulk-sample analysis procedures (3) to elaborate correction procedures based on modeling particle size and shape (4). Although these correction procedures significantly reduce the effects of particle geometry, elemental concentrations determined from the quantitative analysis of particles by EDS (Electronic Data Systems, Plano, TX, www.eds.com) Founded in 1962 by H. Ross Perot (independent candidate for the President of the U.S. in 1992), EDS is the largest outsourcing and data processing services organization in the country.  are often characterized by relative uncertainties on the order of [+ or -] 0.10 to [+ or -] 0.20. This compares to relative uncertainties of [+ or -] 0.02 to [+ or -] 0.05 for the EDS analysis of bulk polished samples as shown in Fig. 1 for the analysis of glass particles and bulk glass samples where RD refers to the relative differences between the experimentally determined concentrations and the known concentrations (see Sec. 1.3.1) (5).

1.1 Particle Analysis at Lower Electron Beam Energies

In conventional electron probe microanalysis the accelerating voltage is generally in the range of 15 kV to 25 kV, which provides the necessary overvoltage and sufficient current from the thermionic therm·i·on  
n.
An electrically charged particle, especially an electron, emitted by a conducting material at high temperatures.



therm
 source to excite efficiently the K and L x-ray lines for elements with atomic numbers as high as Z = 83. In general, this same criterion has been applied to the analysis of particles and the development of the methods used to analyze them.

The FEG-SEM in conjunction with a high resolution energy-dispersive x-ray detector such as the microcalorimeter detector will make it possible to use accelerating voltages in the range of 1 kV to 10 kV depending on the specific elements to be measured, to excite lower-energy characteristic x-ray lines with energies less than about 4 keV. Employing the L and M x-ray lines (energies from 0.640 keV to 2.42 keV) for the elements with atomic numbers in the range 25 to 83, rather than the K and L x-ray lines of these elements (energies 5.85 keV to 10.83 keV) should reduce significantly the magnitude of both the particle mass and absorption corrections.

The advantage of reducing the electron beam energy when analyzing particles is that the electron beam interaction volume within the particle is reduced and therefore the x-ray generation volume for the particle will be similar to that from a bulk material, despite the particle geometry. This is illustrated in Figs. 2 and 3 which are Monte Carlo Monte Carlo (môNtā` kärlō`), town (1982 pop. 13,150), principality of Monaco, on the Mediterranean Sea and the French Riviera.  plots from the Electron Flight Simulator flight simulator, device providing a controlled environment in which a flight trainee can experience conditions approximating those of actual flight. A simulator generally consists of an enclosure housing a working replica of the interior of the cockpit of an  program version 3.1E (Small World Inc. (1)) based on the Monte Carlo model of Joy (6). Figure 2 shows the electron trajectories (Fig. 2a) and the Mg K(alpha) x-ray generation locations (Fig. 2b) for a 20 kV analysis of a 3 (micro)m K-411 glass microspheres Not to be confused with Microsphere.

Glass microspheres are spheres of glass technically manufactured with a diameter in the range 1 to 1000 micrometres, although the term is also used for a wider range of 100 nanometres to 5 millimetres.
 (7). At 20 kV, the electrons interact throughout most of the particle volume with a significant fraction scattering scattering

In physics, the change in direction of motion of a particle because of a collision with another particle. The collision can occur between two charged particles; it need not involve direct physical contact.
 out of the sides and bottom of the particle. In addition, a significant number of x rays are generated along the sides of the particle. As a result, the characteristic x-ray intensities emitted from this particle will be influenced by the particle's shape and si ze, and will require corrections for both particle mass and particle absorption effects. The plots in Fig. 3(a) and 3(b) show that the electron interaction and Mg K[alpha] x-ray generation volumes at 5 kV are considerably smaller than the volumes in the 20 kV plots. In fact they are small enough that the effects of particle shape and size are minimal and the particle analyzed under these conditions closely approximates a bulk material.

1.2 Experimental

To study the effects of reducing the electron beam energy for particle analysis, three different morphological mor·phol·o·gy  
n. pl. mor·phol·o·gies
1.
a. The branch of biology that deals with the form and structure of organisms without consideration of function.

b.
 forms of the analytical glass K-411, NBS (National Bureau of Standards) See NIST.

NBS - National Bureau of Standards: part of the US Department of Commerce, now NIST.
 SRM (1) (Storage Resource Management) The management of the storage resources in an organization in order to avoid duplication of files and to determine space utilization across all servers.  470, which has the composition listed in Table 1, were analyzed (8).

The morphological forms included bulk-polished glass, microspheres, and randomly shaped shards. The shards were prepared by dry grinding grinding, process by which surface material is removed from an object, usually metal, by the abrasive action of a rotating wheel or a moving belt that contains abrasive grains.  glass K-411 in a tungsten-carbide mortar and the microspheres were prepared by injecting the glass shards into a tube furnace In solid state chemistry, a tube furnace is a heating device for conducting syntheses and purifications of inorganic compounds. The usual design consists of a cylindrical cavity surrounded by heating coils, which are imbedded in a thermally insulating matrix.  (8,9). Figure 4 is a secondary electron secondary electron
n.
An electron produced in secondary emission.



secondary electron

An electron produced by secondary emission.
 image of representative microspheres, which range in size from < 1 [micro]m to > 100 [micro]m in diameter. The shards and the microspheres were dispersed dis·perse  
v. dis·persed, dis·pers·ing, dis·pers·es

v.tr.
1.
a. To drive off or scatter in different directions: The police dispersed the crowd.

b.
 onto conductive conductive

having the quality of readily conducting electric current.


conductive flooring
flooring or floor covering made specially conductive to electrical current, usually by the inclusion of copper wiring that is earthed
 carbon tape; the bulk glass was mounted in Ag epoxy epoxy

Any of a class of thermosetting polymers, polyethers built up from monomers with an ether group that takes the form of a three-membered epoxide ring. The familiar two-part epoxy adhesives consist of a resin with epoxide rings at the ends of its molecules and a curing
 and polished. All samples were coated with approximately 10 nm of carbon to reduce charging effects.

The bulk glass and the microspheres were analyzed in an Hitachi S-4500 FEG-SEM equipped with a PGT PGT Public Guardian and Trustee
PGT Procuradoria Geral do Trabalho (Brazil general attorney's office of the work)
PGT Pistol Grip Tool
PGT Post Graduate Training
PGT Princeton Gamma-Tech, Inc.
 IMIX IMIX Internet Mix
IMIX Interactive Multimodal Information Extraction
 system including a 60 [mm.sup.2] Prism x-ray detector. Analyses were done at accelerating voltages of 5 kV and 10 kV, for 500 s and at 15 kV, and 25 kV for 300 s. The deadtime during spectrum collection was generally less than 20 %. Microspheres 1 [micro]m to 30 [micro]m in diameter were analyzed using an area raster The horizontal lines (scan lines) displayed on a TV or computer monitor. This is the origin of the term "raster graphics," which is the major category that all bitmapped images and video frames fall into (GIF, JPEG, MPEG, etc.).  and the results reported as the ratios of the background-corrected, elemental x-ray peak intensities from the microspheres divided by the corresponding bulk-glass intensities, (particle/bulk) ratios. Background correction was performed with a sequential simplex fitting procedure (10). All measurements were corrected for drift in beam current. Different microspheres were analyzed for each voltage in an attempt to minimize contamination from the beam raster.

Four irregularly shaped K-4 11 glass shards ranging in size from about 1 [micro]m to 24 [micro]m were analyzed in spot mode for 300 s at 10 keV, 17 keV, 20 keV, 22 keV, and 25 keV in a JEOL JEOL Japan Electron Optics Laboratory  8600 electron probe equipped with a Tracor Northern x-ray detector. Quantitative electron probe analysis was run on the shards employing the ZAF quantitative correction procedure contained in NIST-NIR Desk Top Spectrum Analyzer A hardware device or software used to examine the frequency and power components of a signal. It provides more information than an oscilloscope, because it can display the signals over a range of frequencies.  (DTSA DTSA Defense Technology Security Administration
DTSA Defense Technical Security Agency (now DTRA)
DTSA Desk Top Spectrum Analyzer
DTSA Seaman Apprentice, Dental Technician Striker (Naval Rating) 
) (10). Standards consisted of bulk, polished materials and included MgO for Mg, [SiO.sub.2] for Si, K-412 glass, SRM 470, (8) for Ca, and K-411 glass for Fe. An MLLSQ-fitting procedure was used with reference peak shapes for background subtraction subtraction, fundamental operation of arithmetic; the inverse of addition. If a and b are real numbers (see number), then the number ab is that number (called the difference) which when added to b (the subtractor) equals  (10).

In addition to the K-411 shards, shards of K-3189 glass (Table 2) were also analyzed in the JEOL 8600 electron microprobe The electron microprobe is an analytical tool used to non-destructively determine the chemical composition of small volumes of solid materials. It uses a high-energy focused beam of electrons to generate X-rays characteristic of the elements present within a sample volumes 1 to 3 . The K-3189 particles were analyzed for 500 s at 10 kV and 20 kV utilizing both an overscan A signal that reaches beyond the viewing area of a display screen. A slight overscan is generally not noticeable on a TV set, but is apparent on a computer monitor where taskbars are typically at the extreme ends of the viewing area.  of the particle area and a spot analysis on the particle center.

1.3 Results and Discussion

1.3.1 Particle Mass Effect

The effect of reducing the acceleration voltage on the magnitude of the particle mass correction can be seen in Fig. 5 which shows the particle/bulk ratios for the K[alpha] x-ray lines of Ca (3.47 keV) and Fe (6.40 keV) vs particle diameter at the different beam energies. These x-ray lines are energetic enough that the particle absorption correction is minimal and the shapes of the various plots are dominated by the particle mass correction. The 25 kV plot, Fig. 5(a), shows that the Fe and Ca curves approach unity only for particles with diameters greater than about 5 p.m. Below this diameter, the particle mass effect dominates due to electron scatter scat·ter
v.
1. To cause to separate and go in different directions.

2. To separate and go in different directions; disperse.

3. To deflect radiation or particles.

n.
 out the bottom and sides of the particles. This results in a 50 % loss in x-ray intensity for a 2 [micro]m sphere and a 75% loss for the smallest microspheres [approximately equal to] 1 [micro]m in diameter.

At 15 kV, Fig. 5(b), the magnitude of the particle mass effect is significantly reduced. At this voltage, the maximum loss of x-ray intensity is only about 13 % for a 1.8 [micro]m diameter sphere compared to the 50 % drop for a 2 [micro]m phere observed in the 25 kV data. In addition, the particle/bulk curves for the 15 kV data approach unity (values within 0.9 and 1.1) for the particles greater than about 3 [micro]m in diameter compared to 5 [micro]m observed at 25 kV.

Lowering the accelerating voltage to 10 kV further reduces the magnitude of the particle mass correction as shown in Fig. 5(c). The maximum loss of x-ray intensity for the 10 kV data is only about 10 % for all of the particles (excluding the one anomalous a·nom·a·lous  
adj.
1. Deviating from the normal or common order, form, or rule.

2. Equivocal, as in classification or nature.
 Fe value for the 18 [micro]m sphere which is at 0.87). Unlike the 15 kV and 25 kV plots, the Fe and Ca particle/bulk values at 10 kV remain near unity for all of the particles in the size range analyzed, 0.9 [micro]m to 18 [micro]m with no obvious drop off even for the smaller diameter particles.

1.3.2 Particle Absorption Effect

The effect of reducing the acceleration voltage on the magnitude of the particle absorption effect can be seen in Fig. 6, which plots the particle/bulk ratios as a function of particle size for the K[alpha] x-ray lines of Si (1.74 keV) and Mg (1.24 keV). The energies of these x-ray lines are low enough that x-ray absorption is significant and the particle absorption effect dominates for microspheres larger than about 4 [micro]m in diameter. For particles less than about 4 [micro]m the particle mass effect would still dominate for the higher beam energies.

At 25 kV, Fig. 6(a), the Mg particle/bulk values are > 1 for all particles larger than 3 [micro]m reaching a maximum of about 1.6 for a 5 [micro]m sphere. This ratio decreases to approach 1 only for the largest spheres with diameters [greater than or equal to] 25 [micro]m. The Si particle/bulk values behave similarly to the Mg values reaching a maximum of 1.4 for a 6 [micro]m sphere, and decreasing to near 1 for the larger spheres. As expected, the Si particle/bulk ratios have a lower maximum compared to the Mg values since the Si x rays are slightly more energetic than the Mg x rays resulting in a smaller absorption effect. As mentioned the particle mass effect dominates for particles less than about 4 [micro]m in diameter, as evidenced in both the Mg and Si plots with the particle/bulk values < 1 for the smaller particles.

The 15 kV results, Fig. 6b, indicate a significant reduction in the particle absorption effect compared to 25 kV results. The maximum particle/bulk value for Mg at 15 kV is only about 1.25 for a 4 [micro]m sphere compared to the value of 1.5 for the 5 [micro]m sphere at 25 kV. In addition, the Mg particle/bulk values now approach 1 for spheres in the 10 [micro]m size range with particle/bulk values of 1.1 for a 10.5 [micro]m sphere and 1.05 for a 15.5 [micro]m sphere. The Si data at 15 kV has a maximum particle/bulk value of 1.18 for a 4 [micro]m sphere compared to the 1.4 for roughly the same size sphere at 25 kV. In addition, the Si particle/bulk ratios for the two large spheres, 10.5 [micro]m and 15.5 [micro]m. are equal to 1 within the experimental uncertainties.

Similar to the particle mass effect, the particle absorption effect is even further reduced at 10 kV, Fig. 6(c). With the exception of a drop in the particle/bulk ratio for the smallest particle at 0.93 [micro]m, due to the particle mass effect, the particle/bulk curves for both Si and Mg at 10 kV are relatively flat for all particle sizes compared to the 15 kV and 25 kV plots. In addition, the particle/bulk values for the 10 kV plot do not show any significant trend above 1 for particles in the 2 [micro]m to 4 [micro]m size range as was evident in the 15 kV and 25 kV plots. For the 10 kV data, the particle/bulk values for Si and Mg are all between 0.9 and 1.1 for all particles greater than 1.4 [micro]m.

The results for 5 kV, Fig. 6d, are similar to the 10 kV data with relatively flat curves. However it is not possible to determine if there is any significant improvement compared to the 10 kV data since the Mg and Si particle/bulk values show an unexpected and significant bias <1 for the smaller particles. This behavior is also evident in the Ca K[alpha] particle/bulk ratios, shown in Fig. 7, for which x-ray absorption is not an issue. This anomaly Abnormality or deviation. Pronounced "uh-nom-uh-lee," it is a favorite word among computer people when complex systems produce output that is inexplicable. See software conflict and anomaly detection. , which is not evident in the 10 kV plots for these elements, is believed to be the result of sample contamination rather than the result of particle mass or absorption effects and is explained in greater detail in the section on contamination later in this paper.

Attempts to use the Fe L line at 5 kV for tracking the absorption effect were not successful because the Fe L intensities were too low resulting in high variations in the net (background corrected) peak intensities even for bulk glass analyses.

1.3.3 Quantitative Analysis of Glass Shards

In addition to the measurement of particle/bulk ratios for the glass microspheres, four K-411 glass shards were quantitatively analyzed in the electron probe. Quantitative analysis was performed with a conventional bulk-sample correction algorithm that involved the comparison of the x-ray intensities from the particles to the intensities from bulk standards. The relative difference (RD) between the particle values and the certified See certification.  bulk values for all elements were then calculated from the following equation:

RD = [(exPerimental-known)/known]. (1)

The shards were analyzed at five different accelerating voltages ranging from 10 kV to 25 kV. The results of the analysis are shown in Fig. 8, which is a plot of the RD s observed in the quantitative results for all elements at each of the different beam energies. The RD values in Fig. 8 are for quantitative results without any normalization or corrections for particle effects. At 25 kV, the standard deviation In statistics, the average amount a number varies from the average number in a series of numbers.

(statistics) standard deviation - (SD) A measure of the range of values in a set of numbers.


of the RD values is 0.238 with RD values ranging from + 0.26 to - 0.60. The variations of the RD values remain large for 22 kV and 20 kV with standard deviations of 0.2 16 and 0.203, respectively. In contrast, the standard deviations of RD distributions for beam energies less than 20 kV drop dramatically. The standard deviations for the RD distributions at 17 kV and 15 kV are 0.120 and 0.094, respectively. At 10 kV the range of RD values is only from + 0.08 to - 0.13 with a standard deviation of 0.054.

In addition to the pooled results for all the elements, the RD values for Mg, the lowest energy x-ray line for this glass composition, and Fe the highest energy line are shown in Figs. 9 and 10. The RD values for the unnormalized Mg analyses on the four shards are plotted in Fig. 9(a) as a function of beam energy. The variations of RD values for 22 kV and 25 kV are the largest with ranges of 0.63 and 0.64, respectively. As expected, the smallest range is 0.20 for 10 kV since the lower acceleration potential minimizes the particle effects.

Figure 9(b) shows a plot of the RD values for Mg analyses in which the sum of the various elemental concentrations have been normalized to 100 %. For these calculations oxygen was determined by stoichiometry stoichiometry

Determination of the proportions (by weight or number of molecules) in which elements or compounds react with one another. The rules for determining stoichiometric relationships are based on the laws of conservation (see
. Normalizing the particle data reduces the range of the RD values, for all voltages. For example, at 25 kV the distribution of RD values ranges from 0.63 (- 0.37 to + 0.26), to 0.34 (+ 0.036 to + 0.30) and for 10 kV the RD values only range from 0.20 (- 0.12 to + 0.08) to 0.13 (- 0.035 to + 0.095). However, while the range of the results improves with normalization, the normalized results show a systematic positive offset with the majority of the RD values for the different voltages being > 1. The positive offset is due to the particle absorption effect and is expected since simple normalization corrects mainly for the particle mass effect, i.e., the negative RD values in the unnormalized data, but not for absorption (Small 1981).

The unnormalized Fe data are shown in Fig. 10(a). For the Fe data the magnitude of the RD distributions follows a similar trend to the Mg data with the largest variation in RD values exhibited for the 25 kV data at 0.50 and the smallest variation for the 10 kV data at 0.065. Unlike the Mg data, the RD values in the unnormalized Fe values are all negative except the 15 kV data which are distributed around zero. The observed systematic negative RD values are the result of the particle mass effect but only for the voltages 17 kV and above. The anomalous negative displacement displacement, in psychology: see defense mechanism.


Same as offset. See base/displacement.
 for the 10 kV Fe data is believed to be the result of contamination similar to the anomalous results shown in Figs. 5d and 6 for the 5 keV data on Mg, Si, and Ca and not related to the particle mass or absorption effects (see the discussion of contamination below).

Looking at the normalized data, Fig. 10(b), the range of RD values is smaller than the unnormalized data with ranges in the RD values of 0.21 for 25 kV to only 0.05 for the 10 kV data. Since the dominant effect on the Fe data is the mass effect, the normalization reduces the spread in RD values and also significantly reduces the systematic negative offset in the unnormalized Fe RD distributions for the higher beam energies. It is important to note that for the 10 kV data the normalization does not correct the negative offset in the data. This further indicates that the observed trend is not related to the particle mass effect.

1.3.4 Surface Contamination Effects in Low-Voltage Analysis

The use of lower accelerating voltages for particle analysis can significantly reduce the magnitude of the particle mass and absorption effects and improve the quantitative results as shown above. However, reducing the accelerating voltage does introduce systematic errors in the measured x-ray intensities, unrelated to particle effects, which must be taken into account to obtain the highest accuracy in quantitative analysis. The systematic errors are the result of the formation of a contamination layer on the particle surfaces during analysis with the electron beam. This problem, although not unique to particles, can have a pronounced effect when analyzing micrometer-sized or smaller particles because the electron beam is often focused to a spot or rastered over a very small area for relatively long counting times often in excess of 500 s for the smaller particles. In this case, comparing the particle x-ray intensities to bulk-standard intensities, where the beam is often rastered over a much larger area and for a shorter time, results in systematic negative bias in the k-ratios. The primary effect of the contamination layer is not an increased absorption of generated x-rays but a retardation retardation: see mental retardation.  or reduction of the average energy of the electron beam that actually enters the specimen and hence the reduction of the overvoltages for the various analytical lines. The overvoltage is defined as: U = [E.sub.0]/[E.sub.crit 'crit A widely used short form for hematocrit ], where [E.sub.0] is the energy of the electron beam and [E.sub.crit] is the critical excitation excitation

Addition of a discrete amount of energy to a system that changes it usually from a state of lowest energy (ground state) to one of higher energy (excited state). For example, in a hydrogen atom, an excitation energy of 10.
 energy of the x-ray line. The x-ray intensity is observed to follow a general form, I [alpha] [(U - 1).sup.n], where n ~ 1.5.

The recommended minimum overvoltage is 2 V for conventional quantitative electron probe analysis with values as low as 1.5 V in special circumstances special circumstances n. in criminal cases, particularly homicides, actions of the accused or the situation under which the crime was committed for which state statutes allow or require imposition of a more severe punishment.  [11]. As the acceleration voltage is lowered to minimize the particle effects, the overvoltages for the various characteristic x-ray lines also decrease. In many cases, this necessitates choosing a characteristic line for which the overvoltage is less than the recommended value. The lower the overvoltage the greater will be the effect of electron retardation resulting in lower than expected characteristic x-ray intensities for all elements.

In this study, the anomalies observed in the quantitative analyses for Fe in the shards (Fig. 10), and Ca particle/bulk plot at 5 kV (Fig. 7) are thought to be manifestations of this contamination effect. A beam energy of 5 keV is reasonably low but for Mg and Si the overvoltages are greater than 2.7 V. In this instance, the effect of contamination is relatively modest with particle/bulk ratios on the order of 5 % low for the microspheres <5 [micro]m as seen in Fig. 6(d). Monte Carlo calculations [12] of the elemental k-ratios for K-411 glass coated with a varying thickness of carbon (Fig. 11) indicates that there is a 2 % drop in the Mg and Si k-ratios from a 5 nm thick layer of C increasing to roughly 5 % for a 10 nm thick layer.

The systematic offset in the Fe data from the shards analyzed at 10 keV, Fig. 10(a), is on the order of--10 % relative. Since 10 kV is not a particularly low-accelerating voltage the effect of a contamination layer on the average beam voltage is minimal. However, the overvolatge for Fe at 10 kV is only 1.5 V, at the lower end of the recommended values for conventional probe analysis, so that even a relatively small reduction in the average beam voltage results in a noticeable loss of Fe intensity from the particles. The contamination effect is evident only for the data collected on the electron probe and not for the analyses performed in the FEG-SEM. This was attributed in part to the formation of a thicker contamination layer in the probe due to the poorer chamber vacuum, which is on the order of 6 X [10.sup.-4] Pa in the probe compared to 6 X [10.sup.-5] Pa in the FEG-SEM. In addition a second possible source of the sample contamination may be the deposition, under beam bombardment, of hydrocarbon hydrocarbon (hī'drōkär`bən), any organic compound composed solely of the elements hydrogen and carbon. The hydrocarbons differ both in the total number of carbon and hydrogen atoms in their molecules and in the proportion of hydrogen  residue residue n. in a will, the assets of the estate of a person who has died with a will (died testate) which are left after all specific gifts have been made. Typical language: "I leave the rest, residue and remainder [or just residue] of my estate to my grandchildren.  from particle mounting procedures as well as from the sample itself. The k-ratios for the Fe K[alpha] x rays at 10 keV (Fig. 11) indicate that the Fe k-ratio has the most significant drop of all the elements with a 1.6 % decrease for a 5 nm C layer increasing to 3 % for a 10 nm layer.

To test further the effects of sample contamination on the analysis for elements such as Fe with low overvoltages, shards of glass K-3189 were analyzed at 10 kV and 20 kV using both an overscan of the particle area and a spot analysis on the particle center. The results are reported as the mean and standard deviation for Fe/Ca and Fe /Si intensity ratios in Table 3 and plotted in Figs. 13 and 14. The Fe intensities were normalized to the intensities for Si and Ca which have overvoltages that range from 5.44 for Si to 2.48 for Ca both well above the 1.5 limit so that their peak intensities will not be significantly affected by a reduction in the electron beam energy due to sample contamination.

The results from the K-3189 shards are similar to the results observed for the quantitative analysis of Fe in the K-411 shards. A comparison of the intensities ratios for the spot and area scans taken at 20 kV and 10 kV are shown in Figs. 12 and 13, respectively, A statistical T test was performed on the data and the results indicate that the means for the 20 kV area and spot analyses are indistinguishable at the 95 % confidence level while the means for 10 kV area and spot analyses are different at the 95 % confidence level with the mean for the spot analyses lower than the mean for the area analyses.

Since subsurface sub·sur·face  
adj.
Of, relating to, or situated in an area beneath a surface, especially the surface of the earth or of a body of water.

Adj. 1.
 charging effects in the non-conducting glass particles may possibly explain the anomalous Fe data for the glass shards, the Duane Hunt limit was determined from the x-ray continuum for the K3189 shards analyzed at 10 kV. The values ranged from a low of 9.96 keV to a high of 10.01 keV and do not indicate any significant static charging that would cause the observed anomaly (13).

Finally, the severity of the contamination effect can also be seen in the dramatic reduction of the Ca particle! bulk ratios at 5 kV in Fig. 7. Many of the particle/bulk ratios are less than 0.8 for the smaller particles. In this case, not only is the average beam energy significantly lowered by the contamination but also the overvoltage for Ca K[alpha] x-ray at 5 kV is only 1.24, well below the recommended value. In this instance, the overvoltage is so low that the Ca intensity is extremely sensitive to contamination even in the FEG-SEM with the better chamber vacuum. Monte Carlo calculations for the Ca k-ratio shown in Fig. 14 indicate that a 5 nm C-layer reduces the Ca k-ratio by 7 % and a 10 am layer reduces it by 13%.

1.4 Conclusions

The results of this study indicate that reducing the accelerating voltage of the electron beam for quantitative particle analysis significantly reduces the magnitude of the particle mass and absorption effects. By reducing the beam energy to 10 kV it was possible to quantitatively analyze irregularly-shaped shards of K-411 glass using bulk standards and a conventional ZAF correction procedure. At 10 kV the variation in mass fraction X was only + 0.08 to 0.12 without any corrections for particle size or geometry. This compares to a variation in mass fraction of + 0.26 to 0.60 relative for a similar analysis at 25 kV.

Although the application of low-voltage analysis reduces the effects of particle size and shape it also introduces systematic trends in the analytical results that must be taken into account for accurate quantitative analysis. These trends, which can be significant, are not related to particle effects but are associated with the buildup build·up also build-up  
n.
1. The act or process of amassing or increasing: a military buildup; a buildup of tension during the strike.

2.
 of contamination under bombardment of the electron beam. The contamination lowers the intensity of the measured x rays by lowering the average energy of the electron beam. This effect increases in magnitude as the accelerating potential of the beam is lowered, being on the order of 5 % for the elements Si and Mg in the K-411 glass microspheres analyzed at 5 kV. The systematic negative errors resulting from contamination are particularly severe for elements that also have low overvoltages such as seen in this study for the Fe data from the shards at 10 kV and the Ca data from the microspheres at 5 kV.

Further reduction of accelerating voltages will only exacerbate the problems associated with contamination buildup on particle surfaces, especially in conventional electron beam instruments where the sample-chamber vacuum is on the order of [10.sup.-4] Pa to [10.sup.-5] Pa and on samples where contamination arises from preparation steps. In samples where the particles are inherently "dirty" (for example atmospheric particles with hydrocarbon residues) low-voltage analysis may not be a viable option for quantitative analysis. Contamination will also be a problem for all low-voltage analyses, including bulk materials, in which a high electron dose is used for sample analysis such as a spot beams A spot beam, in telecommunications parlance, is a satellite signal that is specially concentrated in power (i.e. send by a high-gain antenna) so that it will cover only a limited geographic area on earth.  or small-area rasters.

[FIGURE 1 OMITTED]
Table 1

Elemental concentrations (mass fraction X 100) for glass K-411

Constituent  Concentration [+ or -] 2-sigma

[SiO.sub.2]       54.30 [+ or -] 0.20
    FeO           14.42 [+ or -] 0.20
    MgO           14.67 [+ or -] 0.20
    CaO           15.47 [+ or -] 0.20

Table 2

Elemental glass K-3189 (concentrations in mass fraction x 100)

    Constituent      Concentrations

    [SiO.sub.2]           40.0
[Al.sub.2][O.sub.3]       14.0
        CaO               14.0
        MgO               10.0
    [TiO.sub.2]            2.0
[Fe.sub.2][O.sub.3]       20.0

Table 3

Fe/Si and Fe/Ca intensities ratios for glass K-3189 at 20 kV and 10 kV


10kV Fe/Si area         10 kV Fe/Si spot        10 kV Fe/Ca area
0.0537 [+ or -] 0.0069  0.0443 [+ or -] 0.0016  0.221 [+ or -] 0.021
20kV Fe/Si area         20 kV Fe/Si spot        20 kV Fe/Ca area
0.282 [+ or -] 0.034    0.284 [+ or -] 0.028    0.650 [+ or -] 0.034


10kV Fe/Si area         10 kV Fe/Ca spot
0.0537 [+ or -] 0.0069  0.194 [+ or -] 0.0079
20kV Fe/Si area         20 kV Fe/Ca spot
0.282 [+ or -] 0.034    0.628 [+ or -] 0.037


Accepted: August 27, 2002

(1.) Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology National Institute of Standards and Technology, governmental agency within the U.S. Dept. of Commerce with the mission of "working with industry to develop and apply technology, measurements, and standards" in the national interest. , nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

2. References

(1.) J. A. Small, Quantitative particle analysis in electron beam instruments. Scanning Electron Micros. 1, 447-461 (1981).

(2.) J. T. Armstrong, Quantitative elemental analysis of individual microparticles with electron beam instruments, in Electron Probe Quantitation, K. F. J. Heinrich and D. E. Newbury, eds., Plenum In a building, the space between the real ceiling and the dropped ceiling, which is often used as an air duct for heating and air conditioning. It is also filled with electrical, telephone and network wires. See plenum cable.  Press, New York New York, state, United States
New York, Middle Atlantic state of the United States. It is bordered by Vermont, Massachusetts, Connecticut, and the Atlantic Ocean (E), New Jersey and Pennsylvania (S), Lakes Erie and Ontario and the Canadian province of
 (1991) PP. 261-315.

(3.) F. W. Wright, P. W. Hodge, and C. G. Langway, Studies of particles of extraterrestrial origin: chemical analysis of 118 particles, J. Geophys. Res. 68, 5575-5587 (1966).

(4.) J. T. Armstrong and P. R. Buseck, Quantitative chemical analysis Noun 1. quantitative chemical analysis - chemical analysis to determine the amounts of each element in the substance
quantitative analysis

chemical analysis, qualitative analysis - the act of decomposing a substance into its constituent elements
 of individual microparticles using the electron microprobe: theoretical. Anal anal (a´n'l) relating to the anus.

a·nal
adj.
1. Of, relating to, or near the anus.

2.
. Chem. 47, 2 178-2192 (1975).

(5.) J. A. Small and J. T. Armstrong, Quantitaive Particle Analysis: fact or fiction, in Microbeam Analysis-1996, G. W. Bailey, J. M. Corbett, R. V. W. Dimlich, J. R. Michael, N. J. Zaluzec, eds., San Francisco San Francisco (săn frănsĭs`kō), city (1990 pop. 723,959), coextensive with San Francisco co., W Calif., on the tip of a peninsula between the Pacific Ocean and San Francisco Bay, which are connected by the strait known as the Golden  Press, San Francisco (1996) p. 496.

(6.) D. C. Joy, An introduction to Monte-Carlo simulations. Scanning Microsc. 5, 329-337 (1991).

(7.) R. B. Marinenko, S. Roberson, J. A. Small, B. B. Thorne, D. Blackburn, D. Kauffman, and S. Leigh, Preparation and certification of K-411 glass microspheres. Microsc. Microanal. 6,542-550 (2000).

(8.) R. B. Marinenko, Preparation and characterization A rather long and fancy word for analyzing a system or process and measuring its "characteristics." For example, a Web characterization would yield the number of current sites on the Web, types of sites, annual growth, etc.  of K-411 and K-412 mineral glasses for microanalysis: SRM 470, NBS Spec. Pub. 260-74 (1982).

(9.) J. A. Small, K. F. J. Heinrich, C. E. Fiori, R. L. Myklebust, D. E. Newbury, and M. F. Dilmore, The production and characterization of glass fibers and spheres for microanalysis, Scanning Electron Microsc. 1, 445-454 (1978).

(10.) C. E. Fiori and C. R. Swyt, Desk Top Spectrum Analyzer (DTSA), U.S. Patent 5,299,138 (1994).

(11.) J. I. Goldstein, D. E. Newbury, P. Echlin, D. C. Joy, A. D. Romig, C. E. Lyman, C. Fiori, and E. Lifshin, Scanning Electron Microscopy and X-ray Microanalysis, 2nd Ed., Plenum Press, New York (1992) p. 349.

(12.) J. T. Armstrong, CITZAF--A package of correction programs for quantitative electron microbeam x-ray analysis, Microbeam Anal. 4, 177-200 (1995).

(13.) D. E. Newbury, Measures for Spectral Quality in Low-Voltage X-ray Microanalysis, Scanning 22, 345-351 (2000).

About the author: John A. Small is a research chemist (jargon) chemist - (Cambridge) Someone who wastes computer time on number crunching when you'd far rather the computer were working out anagrams of your name or printing Snoopy calendars or running life patterns. May or may not refer to someone who actually studies chemistry.  in the Surface and Microanalysis Science Division of the NIST (National Institute of Standards & Technology, Washington, DC, www.nist.gov) The standards-defining agency of the U.S. government, formerly the National Bureau of Standards. It is one of three agencies that fall under the Technology Administration (www.technology.  Chemical Science and Technology Laboratory. His research has been in the general area of accuracy in quantitative analysis of materials focusing on the quantitative chemical analysis of individual part ides. Research activities have included the development of a peak-to-background method for the quantitative analysis of particles and rough surfaces; the establishment of an accuracy base for the measurement of environmental asbestos asbestos, mineral
asbestos, common name for any of a variety of silicate minerals within the amphibole and serpentine groups that are fibrous in structure and more or less resistant to acid and fire.
; the modeling of bremsstrahlung bremsstrahlung (brĕm`shträ'ləng): see X ray.
bremsstrahlung

(German; “braking radiation”)
 radiation; and the development of an experimental electron-optical bench for measuring electron probe parameters. The National Institute of Standards and Technology is an agency of the Technology Administration, U.S. Department of Commerce.
COPYRIGHT 2002 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2002, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Small, J.A.
Publication:Journal of Research of the National Institute of Standards and Technology
Geographic Code:1USA
Date:Nov 1, 2002
Words:5733
Previous Article:Averaging of backscatter intensities in compounds.
Next Article:X-ray microanalysis in the variable pressure (environmental) scanning electron microscope.
Topics:



Related Articles
NEW MICROSCOPY CAPABILITY AT NIST.(National Institute of Standards and Technology)(Brief Article)
Chlorination of NR: surface and failure analyses in a post-vulcanization bonded composite. (Cover Story).(natural rubber )
Uncertainty in quantitative electron probe microanalysis.
Optimization of wavelength dispersive x-ray spectrometry analysis conditions.
Limitations to accuracy in extracting characteristic line intensities from x-ray spectra.
X-ray microanalysis in the variable pressure (environmental) scanning electron microscope.
Barriers to quantitative electron probe x-ray microanalysis for low voltage scanning electron microscopy.
The microcalorimeter for industrial applications.
RAL expands services.(Industry News)(Raytheon Co. Reliability Analysis Laboratory)(Brief Article)
Defects in molded silicone medical part identified.(Case Studies)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles