Standardization of EPDM characterization tests for QC and specification purposes.(This is the conclusion of a two-part series. The first installment appeared in May, 1997). Unsaturation un·sat·u·rat·ed adj. 1. Of or relating to an organic compound, especially a fatty acid, containing one or more double or triple bonds between the carbon atoms. 2. Capable of dissolving more of a solute at a given temperature. content of EPDM EPDM Ethylene-Propylene-Diene-Monomer EPDM Enterprise Product Data Management EPDM Ethylene Propylene Dimonomer (industrial/commercial piping/plumbing components) EPDM Engineering Product Data Management : Ethylidene norbornene and dicyclopentadiene The exercise carried out in this context was restricted to EPDMs containing ethylidene norbornene (ENB), dicyclopentadiene (DCPD DCPD Dicyclopentadiene DCPD Direct Current Potential Drop DCPD Direct Compensation Property Damage (automobile insurance coverage) DCPD Daly City Police Department (California) DCPD Directional Canister Passage Detector ) or both as third monomers for unsaturation. Because of the more readily analytically accessible unsaturated unsaturated /un·sat·u·rat·ed/ (un-sach´ur-at?ed) 1. not holding all of a solute which can be held in solution by the solvent. 2. denoting compounds in which two or more atoms are united by double or triple bonds. -C = C- carbon-carbon double bond, the the amount of unsaturation of EPDM has always been somewhat easier than that of the ethylene ethylene (ĕth`əlēn') or ethene (ĕth`ēn), H2C=CH2, a gaseous unsaturated hydrocarbon. It is the simplest alkene. or propylene propylene /pro·pyl·ene/ (pro´pi-len) a gaseous hydrocarbon, CH3CHdbondCH2. propylene glycol a colorless viscous liquid used as a humectant and solvent in pharmaceutical preparations. content. For this reason, a variety of analytical test methods has based developed (refs. 15 and 16), of which the most important are: * iodometric titration titration (tītrā`shən), gradual addition of an acidic solution to a basic solution or vice versa (see acids and bases); titrations are used to determine the concentration of acids or bases in solution. , resulting in a so-called iodine iodine (ī`ədīn, –dĭn) [Gr.,=violet], nonmetallic chemical element; symbol I; at. no. 53; at. wt. 126.9045; m.p. 113.5°C;; b.p. 184.35°C;; sp. gr. 4.93 at 20°C;; valence −1, +1, +3, +5, or +7. number, or expressed as C = C double bonds per 1,000 C-atoms; * refractive index A property of a material that changes the speed of light, computed as the ratio of the speed of light in a vacuum to the speed of light through the material. When light travels at an angle between two different materials, their refractive indices determine the angle of transmission measurement of the pure polymer; * pyrolysis py·rol·y·sis n. Decomposition or transformation of a chemical compound caused by heat. pyrolysis (pīrol´isis), n gas chromatography gas chromatography (GC) Type of chromatography with a gas mixture as the mobile phase. In a packed column, the packing or solid support (held in a tube) serves as the stationary phase (vapour-phase chromatography, or VPC) or is coated with a liquid stationary phase ; * infrared spectroscopy spectroscopy Branch of analysis devoted to identifying elements and compounds and elucidating atomic and molecular structure by measuring the radiant energy absorbed or emitted by a substance at characteristic wavelengths of the electromagnetic spectrum (including gamma ray, ; * [sup.1H]-NMR spectroscopy. Infrared determination of the unsaturation content of EPDM In the same way as in the exercise concerning ethylene determination, the five participating laboratories measured the ethylidene norbornene content of the samples 1-3, using their own in-house test method. The results are shown in table 6. Again, quite unsatisfactory results were obtained between the different laboratories as regards differences in average numbers as well as regards repeatability. Part of the reason again was the lack of a unified set of calibration standards. Based on an analysis of the coefficients of variation, the infrared spectroscopic spec·tro·scope n. An instrument for producing and observing spectra. spec tro·scop measurement seemed to be the best method. Therefore, this method was
further pursued within the working group by the exchange of information.Table 6 - ENB content of the three selected EPDM samples 1-3, measured by own in-house
Sample 1 Sample 2
x s v x s v
Lab 1
Refractive index 4.61. .148 3.2 2.20 .149 6.8
Lab 2
Infrared spectr. 2.56 .048 1.8
Lab 3
Infrared spectr. 3.80(*) .034 0.9 2.45 .025 1.0
Lab 4
Refractive index 4.7 .03 0.6 2.4 .07 2.9
Lab 5
Iodine number 4.8 .13 2.7 2.45 .08 3.3
Sample 3
x s v
Lab 1
Refractive index 9.10 .126 1.4
Lab 2
Infrared spectr. 6.73 .043 0.6
Lab 3
Infrared spectr. 9.60 .077 0.8
Lab 4
Refractive index 8.9 .05 0.6
Lab 5
Iodine number 9.0 .165 1.8
x = average (mass%) s = standard deviation In statistics, the average amount a number varies from the average number in a series of numbers. (statistics) standard deviation - (SD) A measure of the range of values in a set of numbers. (mass%) v = coefficient of varation (%) * Sample 1 contained some extender See Media Center Extender, bus extender and DOS extender. oil; result not corrected for oil content. Further study showed that the determination of the film thickness of the sample used for the infrared technique was the main cause of the variation in the results. It is rather difficult to press a film with a homogeneous thickness, particularly for polymers with a high ethylene content. The quantitative interpretation of an infrared spectrum Noun 1. infrared spectrum - the spectrum of infrared radiation infrared, infrared frequency - the infrared region of the electromagnetic spectrum; electromagnetic wave frequencies below the visible range; "they could sense radiation in the infrared" is therefore hampered by the inaccuracy in·ac·cu·ra·cy n. pl. in·ac·cu·ra·cies 1. The quality or condition of being inaccurate. 2. An instance of being inaccurate; an error. in the sample thickness at the point where the actual spectrum is measured. Different methods were used to determine the sample thickness: * mechanical means, such as thickness measurement using callipers Cal`li`pers n. pl. 1. See Calipers. callipers, calipers (US) npl (MED) → aparato ortopédico; (MATH) → calibrador m ; * amount of absorption of beta radiation Beta radiation Streams of electrons emitted by beta emitters like carbon-14 and radium. Mentioned in: Pinguecula and Pterygium radiation ; * measurement of the thickness included in the infrared technique by measurement of the infrared absorbance absorbance /ab·sor·bance/ (-sor´bans) 1. in analytical chemistry, a measure of the light that a solution does not transmit compared to a pure solution. Symbol . 2. at a so-called isotopic i·so·tope n. One of two or more atoms having the same atomic number but different mass numbers. [iso- + Greek topos, point, where no specific absorption of chemical groups takes place. One of the members of the working group (Exxon Chemical) has recently developed the latter method, apparently with very positive results. This method provides automatic film thickness determination at the same point of the film at which the spectrum is taken. Depending on the particular sample used, the thickness gauge is the net absorbance difference between: * Group 1: 2,708 [cm.sup.-1] (isotopic point) and 2,450 [cm.sup.-1] (anchor point Anchor Point may refer to:
* Group 2: 2,668 [cm.sup.-1] (isotopic point) and 2,450 [cm.sup.-1] (anchor point). The contents of unsaturation are measured by using the following response peaks: * ENB: 1,690 [cm.sup.-1], or rather the second derivative in the region 1,681 - 1,688 [cm.sup.-1]; * DCPD: 1,611 [cm.sup.-1], or rather the second derivative in the region 1,612 - 1,619 [cm.sup.-1]. This method has been published (ref. 17), and the reader is referred to this article for further details. As infrared spectroscopy is only a relative technique, this method also needs standard calibrants. See the section on calibration standard for ASTM ASTM abbr. American Society for Testing and Materials D3900 for a detailed description. For ENB-containing polymers, which generally cover a broader range of contents than DCPD-containing polymers, four standard calibrants are available, as opposed to two for DCPD. In fitting the infrared responses to these standard calibrants, the experience of the working group was that the use of the least square fit of these ENB standard calibrants gave slightly better results overall than a linear extrapolation (mathematics, algorithm) extrapolation - A mathematical procedure which estimates values of a function for certain desired inputs given values for known inputs. If the desired input is outside the range of the known values this is called extrapolation, if it is inside then for EPDMs with intermediate ENB contents. The working group therefore recommends the least square fit. In addition, this method for determining unsaturation content sets certain requirements for the Fourier transform Fourier transform In mathematical analysis, an integral transform useful in solving certain types of partial differential equations. A function's Fourier transform is derived by integrating the product of the function and a kernel function (an exponential function raised to infrared instrument; it should have spectral accumulation, averaging and subtracting capabilities and should be fitted with a sample shuttle to permit alternating and repetitive collection of single beam sample and "empty sample compartment" spectra. If a shuttle is not available, this method cannot be implemented exactly as described. In such cases, the method may be modified. but only at the expense of some development work. A preferred (but not mandatory) instrument for the procedure is the Perkin Elmer 1760 series spectrophotometer spectrophotometer, instrument for measuring and comparing the intensities of common spectral lines in the spectra of two different sources of light. See photometry; spectroscope; spectrum. with its dedicated 7500 or 7600 computer. In the case of oil-extended EPDM polymers, the oil must be extracted before the unsaturation is determined. A suitable procedure is described later. Further caution has to be taken with stearic acid stearic acid /ste·a·ric ac·id/ (ste-ar´ik) a saturated 18-carbon fatty acid occurring in most fats and oils, particularly of tropical plants and land animals; used pharmaceutically as a tablet and capsule lubricant and as an emulsifying or stearates contained in the EPDM polymer, for which the spectrum needs to be "cleaned." Based on this Fourier transform infrared method and corresponding standard calibrants, three laboratories (one laboratory did not have the right instrument at that time) compared their ability to achieve the same results and sufficient repeatability on the four standard calibrants belonging to this technique as well as on the samples 1-3 (tables 7 and 8). Table 7 - Fourier transform infrared determination of ENB content of four ENB standard calibrants
ENB standard 2 ENB standard 3
x s v x s v
Assigned ENB
content (mass %) 2.26 5.11
Lab 1 2.15 .028 1.3 5.29 .057 1.1
Lab 2 2.18 .061 2.8 5.09 .080 1.6
Lab 3 2.25 .058 2.6 5.02 .082 1.6
ENB standard 4
x s v
Assigned ENB
content (mass %) 9.99
Lab 1 9.91 .264 2.7
Lab 2 9.80 .088 0.9
Lab 3 9.80 .118 1.2
ENB standard 1 contains 0.0 mass % ENB and was not measured, but used only for calibration purposes. Table 8 - ENB content of three selected EPDM samples 1-3, determined using the new Fourier transform infrared method
Sample 1 Sample 2 Sample 3
x s v x s v x s v
Lab 1 2.45 .025 1.0 9.60 .077 0.8
Lab 2 4.53 .042 0.9 2.42 .034 1.4 9.62 .046 0.5
Lab 3 4.64 .045 1.0 2.36 .008 0.4 9.68 .171 1.8
The correspondence between laboratories is remarkably good. The conclusion can be drawn that the participating laboratories managed repeatability levels using this method which were on average three times smaller (as expressed by the coefficient of variation Coefficient of Variation A measure of investment risk that defines risk as the standard deviation per unit of expected return. ) than those achieved using their own in-house method. On this basis, the HSRP (Hot Standby Router Protocol) A protocol from Cisco for switching to a backup router in the event of failure. See VRRP. HSRP - Hot Standby Routing Protocol working group decided to recommend the Exxon test method for the determination of mass % unsaturation in EPDM. Calibration standards for unsaturation It is commonly accepted that [sup.1H]-NMR is the best technique for obtaining quantitative data on the content of unsaturation in EPDM, irrespective of irrespective of prep. Without consideration of; regardless of. irrespective of preposition despite whether the unsaturation is based on ENB or DCPD. Therefore, a set of standard calibrants were certified using [sup.1H]-NMR, using samples dissolved in deuterated o-dichlorobenzene at 120 [degrees] C. The following standard calibrants were developed: Ethylidene norbornene * Standard 1: 0 mass % * Standard 2: 2.26 mass % * Standard 3: 5.11 mass % * Standard 4: 9.99 mass % Dicyclopentadiene * Standard 1: 0 mass % * Standard 5: 4.80 mass % All contents were based on the recommended compositional definition given in the section on the description of molecular composition of EPDM. These standard calibrants cover the range of commonly available commercial EPDM rubbers EPDM rubber (ethylene propylene diene monomer rubber) is an elastomer which is characterized by wide range of applications. EPDM rubber is used in vibrators and seals; glass-run channel; radiator, garden and appliance hose; tubing; washers; belts; and electrical insulation. , where slight extrapolation to somewhat higher numbers may be necessary in exceptional cases. Determination of the stabilizer stabilizer: see airplane. content (phenolic phe·no·lic adj. Of, relating to, containing, or derived from phenol. n. Any of various synthetic thermosetting resins, obtained by the reaction of phenols with simple aldehydes and used as adhesives. antioxidants Antioxidants Substances that reduce the damage of the highly reactive free radicals that are the byproducts of the cells. Mentioned in: Aging, Nutritional Supplements antioxidants, n. ) During the course of the study. the question was raised as to whether harmonization har·mo·nize v. har·mo·nized, har·mo·niz·ing, har·mo·niz·es v.tr. 1. To bring or come into agreement or harmony. See Synonyms at agree. 2. Music To provide harmony for (a melody). of the determination of stabilizer content of EP(D)M would be desirable. An overview of the different methods in use at the different suppliers revealed the following: * infrared spectroscopy: three suppliers; typical coefficient of variation 4-13%; * HPLC HPLC high-performance liquid chromatography. HPLC high performance liquid chromatography. HPLC High-performance liquid chromatography Lab instrumentation A highly sensitive analytic method in which analytes are placed (high pressure liquid chromatography): one supplier; typical coefficient of variation 3%; * photometric pho·tom·e·try n. Measurement of the properties of light, especially luminous intensity. pho to·met detection: one supplier; no further details.The fact that the determination of stabilizer content can be included in the infrared determination of ethylene content and unsaturation content without much extra effort is a strong point. The HPLC method involves a great deal of work. In addition, the amount of stabilizer does not normally need to be determined with such great accuracy, as stabilizer content is usually specified as a one-sided minimum requirement. It was agreed, therefore, that the possibility of a slight gain in repeatability with HPLC compared with infrared spectroscopy was not really worth the effort. Infrared determination of the stabilizer content EP(D)M In the case of commonly used stabilizers of the hindered phenol phenol (fē`nōl), C6H5OH, a colorless, crystalline solid that melts at about 41°C;, boils at 182°C;, and is soluble in ethanol and ether and somewhat soluble in water. type, the determination can be included in the infrared determination of ethylene content, making use of the infrared absorbance at 1,740 [cm.sup.-1]. The height of the peak is measured and is quantified by calibrating the instrument with known standards. In the same way as for the ethylene content determination, the film thickness gauge may be determined as the net absorbance difference between the isotopic points and the anchor point for groups 1 and 2. The test method as described above has to be the only preferred test method, due to certain restrictions: * it is useful only for sterically hindered phenolic antioxidants; * it may show an interference between ester additives and the EP(D)M (stearates); * in the case of oil-extended products, it can be used only with paraffinic oil. The IISRP IISRP International Institute of Synthetic Rubber Producers working group recommends infrared spectroscopy for the determination of the stabilizer content of EP(D)M. Calibration standards for stabilizer content Calibration standards are best achieved by mill-mixing known amounts of the actual stabilizer in pure EP(D)M samples. Because of the limited stability of the stabilizer content during the shelf-life of such calibration standards, it was not considered useful to develop a unified set of standard calibrants. Determination of the oil content of oil-extended EPDM Determination of the extender oil content of EPDM is another important analytical technique An analytical technique is a method that is used to determine the concentration of a chemical compound or chemical element. There are a wide variety of techniques used for analysis, from simple weighing (gravimetric) to titrations (titrimetric)to very advanced techniques using . In addition, many of the aforementioned analytical techniques (ethylene content and unsaturation content) require the oil to be removed before the measurement can be carried out. Methods commonly in use can be divided into two categories - extraction methods and precipitation methods. The first category makes use of a solvent which dissolves the extender oil but which does not dissolve the EPDM. Depending on the experimental configuration, extraction is achieved either by step-by-step extractions in flasks with regular renewal of the extraction medium or using Soxhlet apparatus. The second category makes use of suitable solvent/non-solvent combinations, where first of all the oil-extended polymer is dissolved completely, then the EPDM polymer is precipitated and separated from the liquid phase containing the extender oil. For both categories either the EPDM moiety moiety: see clan. or the oil moiety after evaporation evaporation, change of a liquid into vapor at any temperature below its boiling point. For example, water, when placed in a shallow open container exposed to air, gradually disappears, evaporating at a rate that depends on the amount of surface exposed, the humidity of the extraction medium can be used for the calculation of the oil content. There was no reason beforehand to prefer either of the two categories, as can be seen from the first attempt to standardize this analytical method. Each participating laboratory measured the oil content of sample 1, using its own in-house method (table 9). Table 9 - oil content of sample 1, measured by own in-house method
Method Solvent Non-solvent Appr. duration
Lab 1 Extraction 2-propanol 4 hours
16 hours
Lab 2 Extraction MEK 2 hours
Lab 3 Extraction Ethanol/toluene
azeotrope
Lab 4 Precipitation Toluene Methanol/acetone 3 hours
50/50
Lab 5 Precipitation Toluene Acetone 2.5 hours
Extraction Acetone/cyclohexane 1 hour
2/1
Method x s v
Lab 1 Extraction 23.6 .31 1.3
via extract
24.1 .03 .14
via extract
Lab 2 Extraction 24.5 .25 1.0
Lab 3 Extraction 23.9 .17 3.4
Lab 4 Precipitation 22.3 .77 3.4
Lab 5 Precipitation 23.6 .32 1.4
Extraction 23.8 .10 .42
x = average (mass %); s = standard deviation (mass %); v = coefficient variation (%) An important criterion in this test is the duration, particularly as this test is to be used for quality-control purposes. On die other hand, as can be seen from table 9, the repeatability of the extraction method seems to improve when longer extraction times are employed. This is of particular importance for EPDMs with a relatively high ethylene content, for which it is well known that the extraction of extender oil can be a time-consuming exercise: the diffusion of the extender oil from the EPDM into the solvent is apparently retarded by crystallinity or by a lower degree of swelling of the EPDM in the non-solvent. After thorough consideration the decision was taken to pursue standardization by improving the extraction methods employed by laboratories 2 and 5. Laboratory 2 made use of Soxhlet apparatus with boiling MEK Noun 1. MEK - a terrorist organization formed in the 1960s by children of Iranian merchants; sought to counter the Shah of Iran's pro-western policies of modernization and opposition to communism; following a philosophy that mixes Marxism and Islam it now attacks the , laboratory 5 of a one-off extraction of [+ or -] 200 mg of EPDM rubber in a boiling azeotrope azeotrope /azeo·trope/ (a´ze-o-trop?) a mixture of two substances that has a constant boiling point and cannot be separated by fractional distillation.azeotrop´ic a·ze·o·trope n. of 40 ml acetone acetone (ăs`ĭtōn), dimethyl ketone (dīmĕth`əl kē`tōn), or 2-propanone (prō`pənōn), CH3COCH3 and 20 ml cyclohexane cyclohexane (sī'kləhĕk`sān), C6H12, colorless liquid hydrocarbon. It is a cyclic alkane that melts at 6°C; and boils at 81°C;. It is nearly insoluble in water. using a conical conical /con·i·cal/ (kon´i-k'l) cone-shaped. con·i·cal or con·ic adj. Of, relating to, or shaped like a cone. flask flask (flask) 1. a laboratory vessel, usually of glass and with a constricted neck. 2. a metal case in which materials used in making artificial dentures are placed for processing. . The main difference was the speed with which the Soxhlet method reached equilibrium compared to the conical flask method for a high-ethylene-containing EPDM (sample 4) extended with 100 phr of paraffinic oil (see figure 2). The higher speed of equilibrium achieved with the Soxhlet method gives an extra guarantee for better repeatability within as short a time-span as possible for the analysis. Why the Soxhlet method, using MEK as the extraction agent, systematically tends to give higher extraction numbers than the conical flask method, using the 2/1 mix of acetone/ cyclohexane, could not be explained. It may well be that the different extraction agents tend to differ a little, depending on their solvent power for EPDM, as regards the amount of very-low-molecular-weight, oligomeric EPDM they extract from the rubber along with the oil. This is a systematic problem with all oil-determination tests, irrespective of whether they are extraction or precipitation methods. To the committee's knowledge there is no fundamental solution to this problem, according to according to prep. 1. As stated or indicated by; on the authority of: according to historians. 2. In keeping with: according to instructions. 3. the present state of the art. As long as everybody uses the same method, the same systematic error (bias) will be made by everybody and this will not lead to major differences in comparison of the data. Although it was felt that the conical flask method had not been fully exploited, it was generally felt that overall all the laboratories achieved the best reproducibility and the same average results with the Soxhlet extraction method. The duration of the extraction, either 30 or 45 minutes, had only a small influence on the results. The IISRP working group decided to recommend the Soxhlet extraction method, using MEK as the solvent, as the preferred method for the determination of the oil content of oil-extended EPDM. This method has been published (ref. 18), and for a detailed description of the method the reader is referred to the pertinent article. Mooney viscosity Sample preparation An important factor affecting the precision of the Mooney measurement has over the last few years proved to be the sample preparation (refs. 6-8). Two types of sample preparations in particular have become more or less officially accepted: * massing on a mill; * no mill-massing: by cutting a sample from a bale bale 1. a package of wool in a wool pack weighing 150-250 lb depending largely on whether it is greasy or scoured. 2. a compressed bundle of hay, either about 100 lb tied with wire or twine, or large, round, untied bales, as big as a small hay stack and referred to as 'big bales'. or, in the case of the rubber's being in crumb or pellet form, by compacting a pre-dimensioned sample in a press. The obvious question was whether either of these two sample preparations could also be recommended for EP(D)M. A review of existing standards for Mooney measurement and/or sample preparation shows the following: * ISO (1) See ISO speed. (2) (International Organization for Standardization, Geneva, Switzerland, www.iso.ch) An organization that sets international standards, founded in 1946. The U.S. member body is ANSI. 4097 (ref. 19) and ISO 1795 (ref. 20): no mill-massing preferred, if massed: 35 [+ or -] 5 [degrees] C; * DIN 53 523 (ref. 21): no mill-massing, vacuum-compacting or mill-massing at 35 [+ or -] 5 [degrees] C; * ASTM D1646-90 (ref. 22): massed or unmassed, if massed: 50 [+ or -] 5 [degrees] C. A review of the actual conditions used by the different laboratories revealed the following: * Lab 1: mill-massing at 105 [degrees] C; * Lab 2: no massing: compacting; * Lab 3: mill-massing at approx. 125 [degrees] C; * Lab 4: mill-massing at 140 [degrees] C; * Lab 5: mill-massing at 50 [degrees] C. In any case, on the basis of their own experience of reducing the test variability, most members do not adopt the lower temperatures recommended in the standard test methods. A review of the experiences of the participants of this study concerning the influences of mill-massing and a lack of mill-massing on the actual Mooney values measured and on the repeatability of the Mooney measurement gives rise to conflicting evidence. Depending on the type of EPDM rubber being tested and on the particular participant carrying out the test, several cases were observed: * average Mooney value is the same; massing at high temperature gives lower standard deviation; * average Mooney value is a few units lower with massing; massing at high temperature gives lower standard deviation; * average Mooney value is a few units lower with massing; standard deviation the same; * average Mooney value is somewhat higher with massing, particularly for free-flowing polymers. It is quite clear that no consistent picture can be derived from these observations. In any case, the advantage of the use of unmassed samples does not show up clearly if compared to the high temperature mill-massing. The IISRP working group decided to support the unmassed sample preparation as the preferred method, simply because of ease of handling. The choice of whether or not to use mill-massing is left to the discretion of the various producers. It is recommended that the sample preparation be specified in reporting data. Mooney test conditions In the course of time, a wide variety of test conditions for Mooney viscosity has come to be used by the various suppliers. A good overview can be gained from the Synthetic Rubber synthetic rubber: see rubber. Manual (ref. 23), wherein eight different conditions are quoted as being in common use for measuring Mooney viscosities of EP(D)M: * ML(1+4) 100 [degrees] C; * ML(1+8) 100 [degrees] C; * ML(1+8) 120 [degrees] C; * ML(1+4) 121 [degrees] C; * ML(2+10) 121 [degrees] C; * ML(1+4) 125 [degrees] C; * ML(1+8) 125 [degrees] C; * ML(1+8) 150 [degrees] C. A summary of the preferred conditions recommended in the appropriate standard test methods either for EPDM or for the Mooney measurement itself shows tile following: * IS0 4097-1980 (ref. 19): ML(1+4) 125 [degrees] C * DIN 53 523 (ref. 21): ML(1+4) 100 [degrees] C * ASTM D1646-90 (ref. 22): ML(1+4) 125 [degrees] C The origin of all these conditions is impossible to trace but they are to a large extent the result of difficulties encountered in accurately measuring the Mooney viscosities for the range of EP(D)M polymers on the market (refs. 5, 24 and 25). In particular, the fact is often overlooked that all pertinent standard test methods for Mooney viscosity call for caution if measurements are carried out on high-molecular-weight polymers. They put the provisional restrictions on the Mooney value that it should not exceed 80, due to mechanical equipment limitations which may result in erratic results. It is quite clear that one condition is by no means sufficient to cover the whole range of EP(D)Ms available and therefore different conditions have originated. The question may be posed though whether the amount as cited above is practical. The main point of discrimination between die various test conditions is the temperature at which the test is carried out. This is the main tool for bringing the Mooney viscosity within the necessary experimental range. And because there are various EP(D)M types being marketed with extremely low Mooney viscosities ([is less than] 10 at 100 [degrees] C), there is sometimes a tendency to lower the temperature conditions to such a temperature and even to standardize the test at that temperature. However, a complicating factor at a temperature of 100 [degrees] C is that for polymers with a high ethylene content remnants of crystallinity may persist up to temperatures above 100 [degrees] C, which will have a rather significant but irreproducible influence on the Mooney viscosity measured at that temperature. There is basically no way out of this dilemma. In order to standardize the Mooney measurement for EP(D)M rubber, the IISRP working group based its considerations on the experimental work already carried out on earlier occasions (refs. 5, 23 and 24) and reached agreement on a recommendation to reduce the various possible conditions to mainly two: * ML(1+4) 125 [degrees] C; * ML(1+8) 150 [degrees] C, leaving the possibility open to use other conditions if the need should arise. The first condition is the one which is most widely used and which is generally recommended in the appropriate standard test methods. The second condition is recommended for high Mooney and/or heavily branched grades. A clear guideline as to when there should be a switch to the second condition cannot be given because it depends on equipment factors in combination with the particularities of the various grades. This is left to the discretion of the particular supplier. Availability of standard calibrants The standard calibrants for the ethylene and unsaturation determination were produced in commercial plants, and sizeable quantities have been set aside. Samples of these calibrants are available from: Exxon Chemical Technology Center Brussels Hermeslaan 2 B-1831 Machelen Belgium and Exxon Chemical Co., Baytown Polymers Center Vistalon Technology Manager 5200 Bayway Drive P.O. Box 5200 Baytown, TX 77522 USA Conclusions In a joint project between European EPDM producers under the auspices of the IISRP a set of recommended characterization tests In computer programming, a characterization test is a means to describe (characterize) the actual behavior of an existing piece of software, and therefore protect existing behavior of legacy code against unentended changes via automated testing. was developed for quality-control and specification purposes for EPDM rubber. These test methods include procedures for: * ethylene content; * unsaturation content (ENB and/or DCPD level); * stabilizer content (phenolic type); * oil content of oil-extended EPDM types; * Mooney viscosity. For ethylene content, a new set of standard calibrants was developed in cooperation with the ASTM D11.11 rubber-testing sub-committee. These are partially based on the older calibrants of ASTM Standard D3900, and involve a shift in mass % ethylene up to 4.5 mass %, relative to the formerly assigned numbers (standard) assigned numbers - The RFC STD 2 documenting the currently assigned values from several series of numbers used in network protocol implementations. This RFC is updated periodically and, in any case, current information can be obtained from the Internet Assigned Numbers . Standard calibrants were also developed for unsaturation content. All standard calibrants are available upon request. The decision to implement such test methods, as well as the timing thereof, are left to the discretion of the individual suppliers and consumers. References [1.] Worldwide rubber statistics, International Institute of Synthetic Rubber Producers Inc., Houston, TX. [2.] J.W.M. Noordermeer, Kirk-Othmer Encyclopedia of Chemical Technology - Fourth Ed., 8, 978 (1993), John Wiley John Wiley may refer to:
[3.] ASTM D3568-90. Standard Test Methods for Rubber - Evaluation of EPDM (ethylene propylene diene Dienes are hydrocarbons which contain two double bonds. Dienes are intermediate between alkenes and polyenes. Classes Dienes can be divided into three classes:
[4.] Prufung von Kautschuk und Elastomeren, DIN 53 670 Teil 10: Prufung von Kautschuk in Standard Testmischungen; ethylen-propylen-dien kautschuk EPDM, 1983 issue. [5.] International Standard, ISO 4097 rubber, ethylene-propylene-diene (EPDM), non-oil extended raw general-purpose rubber - evaluation procedures, 1991 issue. [6.] J. Markert, Gummi Asbest Kunststoffe, 9, 568 (1976). [7.] R. Koopmann, Kautschuk + Gummi, Kunststoffe, 38, 281 (1985). [8.] H. Kramer, Kautschuk + Gummi, Kunststoffe, 43, 912 (1990); Rubber World, 204, 35 1991). [9.] W. Breemhaar, R. Koopmann, J. Markert and J. Noordermeer, Kautschuk Gummi Kunststoffe, 46, 957 (1993). [10.] International Standard, ISO 5725, Precision of test methods - determination of repeatability and reproducibilily for a standard test method by inter-laboratory, tests. [11.] I.J. Gardner, C. Cozewith and G. Verstrate, Rubber Chemistry and Technology, 44, 1015 (1971). [12.] Standard Test Methods for Rubber, ASTM D3900, determination of ethylene units in EPM EPM equine protozoal myeloencephalitis. and EPDM, 1980 issue. [13.] S. DiMartino and M. Kelchtermans, Determination of the composition of ethylene-propylene-rubbers using [sup.13C]-NMR spectroscopy, J. Appl. Poly. Sci. in print. [14.] Standard Test Methods for Rubber, ASTM D3900, determination of ethylene units in EPM and EPDM, 1994 issue. [15.] J. Gardner and G. Verstrate, Rubb. Chem. and Techn., 46, 1019-1034 (1973). [16.] J. van Schooten and J.K. Evenhuis, Polymer 6 (11), 561-577 (1965). [17.] Determination of 5-ethylenenorbornene (ENB) by Fourier transform infrared spectroscopy, Materiaux et Techniques December (1991), 69-72. [18.] J. Fourreau, Oil determination by MEK extraction, Materiaux et Techniques, 4-5, 80 (1992). [19.] ASTM D1646-93, Standard Test Method for Rubber - Viscosity and Vulcanization vulcanization (vŭl'kənəzā`shən), treatment of rubber to give it certain qualities, e.g., strength, elasticity, and resistance to solvents, and to render it impervious to moderate heat and cold. Characteristics (Mooney viscometer viscometer Instrument for measuring the viscosity (resistance to internal flow) of a fluid. In one type, the time taken for a given volume of fluid to flow through an opening is recorded. ). [20.] International Standard ISO 1795, Rubber, raw, natural and synthetic; sampling and further preparative pre·par·a·tive adj. Serving or tending to prepare or make ready; preliminary. n. Something that prepares for or acts as a preliminary to something following. procedures, 1992 issue. [21.] Prufung von Kautschuk und Elastomeren, DIN 53 523, Prufung mit den Scherscheiben-Viscosimeter nach Mooney, 1992 issue. [22.] International Standard, ISO 289-1, Determination of viscosity of natural and synthetic rubbers by the shearing disk viscometer, 1994 issue. [23.] The Synthetic Rubber Manual, International Institute of Synthetic Rubber Producers, Inc., Houston 1992, 12th Ed. [24.] R.J.H. America, G.W. Visser and W. Breemhaar, PRI PRI: see Institutional Revolutionary party. (Primary Rate Interface) An ISDN service that provides 23 64 Kbps B (Bearer) channels and one 64 Kbps D (Data) channel (23B+D), which is equivalent to the 24 channels of a T1 line. Conference, Belgian Section, Leuven, Belgium; 16-17 April 1991. [25.] E.T. Italiaander, EPR EPR Electron Paramagnetic Resonance EPR Extended Producer Responsibility EPR Electronic Patient Record(s) EPR Emergency Preparedness and Response (US DHS) EPR Endpoint Reference EPR Ethylene-Propylene Rubber raw polymer Mooney and its importance of using the correct test temperature to ensure reproducibility of test data; Polysar, technical memorandum, Antwerp, 8 November 1988. Dr. Noordermeer authored this on behalf of the International Institute of Synthetic Rubber Procedures. The objective is to achieve standardization among EPDM producers of testing procedures for the basic properties. This article was originally published in Kautschuk + Gummi, Kunststoffe. The first part will ran in May. |
|
||||||||||||||||||

tro·scop
Printer friendly
Cite/link
Email
Feedback
Reader Opinion