Printer Friendly
The Free Library
22,741,889 articles and books

Standardization of [.sup.68] Ge/[.sup.68] Ga using three liquid scintillation counting based methods.

A solution containing [.sup.68] Ge in equilibrium with its daughter, [.sup.68] Ga, has been standardized for the first time at the National Institute of Standards and Technology (NIST) using 3 liquid scintillation-based techniques: live-timed 4[pi] [beta]-1[lambda] anticoincidence (LTAC) counting, the Triple-to-Double Coincidence Ratio (TDCR) method, and [.sup.3] H-standard efficiency tracing with the CIEMAT (1) /NIST (CNET) method. The LTAC technique is much less dependent on level scheme data and model-dependent parameters and was thus able to provide a reference activity concentration value for the master solution with a combined standard uncertainty of about 0.3%. The other two methods gave activity concentration values with respective differences from the reference value differences from the reference value of + 12% and - 1.5% which were still within the experimental uncertainties.

Measurements made on the NIST "4[Pi]" [lambda] secondary standard ionization chamber allowed for they determination of calibration factors for that instrument, allowing future calibrations to be made for [.sup.68] Ge/[.sup.68]Ga without the needs for a primary measurement. The ability to produce standardized solutions of [.sup.68)] Ge presents opportunities of the development of a number of NIST-traceable calibration sources with very low (< 1%) relative standard uncertainties that can be used in diagnostic medical imaging.

Key words: anticoincidence counting; CIEMAT/NIST method; germanium-68; liquid scintillation counting; positron emitter; standardization; TDCR method.

Accepted: September 10, 2008

Available online: http://www.nist.gov/jres

1. Introduction

The use of Positron Emission Tomography (PET) as a tool for diagnosing disease, particularly cancer; continues to rise at a rate of about 20% per year (1), partially due to technological advances that allow for more quantitative data to be collected. The ability to consistently acquire truly quantitative imaging data depends on the use of radioactivity standards traceable to the National Institute of Standards and Technology (NIST).

Currently the most widely used radionuclide in PET imaging is (.sup.18] F. Although NIST has previously standardized [.sup.18] F [2, 3], its short half-life (1.83 h) makes it nearly impossible to prepare and distribute Standard Reference Materials to most users. As a potential solution, [.sup.68] Ge, in equilibrium with its daughter, [.sup.68] Ga, has been proposed as a surrogate.

A simplified scheme for the decays of [.sup.68] Ge and its [.sup.68] Ga daughter is shown in Fig. 1. Germanium-68 decays by pure electron capture (EC) to the ground state of [.sup.68] Ga with half-life of 270.95(16) d (4). Gallium-68 in turn decays with a half-life of 67.71(9) min by a combination of EC and positron emission primarily to the ground state of [.sup.68] Zn, but also with a branch to an excited state at 1077 keV with a probability of about 3% and a number of higher excited states with a combined probability of under 0.4%.

The nature of the decay scheme of [.sup.68] Ge/[.sup.68] Ga makes it amenable to a variety of different standardization techniques. In 1994, Schonfeld, et al. (5) reported on the results of measurements made with liquid scintillation (LS) counting using the CIEMAT/NIST (3) H standard efficiency tracing method (CNET) (6), (7), 4 [pi][beta]-[gamma] coincidence, and a calibrated ionization chamber (IC). The data showed good agreement between all three methods, giving activity concentration values within the respective experimental uncertainties (nominally 1% relative standard uncertainty).

[FIGURE 1 OMITTED]

More recently, Grigorescu, et al. (8), reported on the results of measurements using 4 [pi][beta] - [gamma] coincidence counting. As with Schonfeld, et al., the coincidence spectrometer consisted of a proportional counter and NaI (T1) detector for the [beta] and [gamma] detection channels, respectively. Because this experimental arrangement requires the use of dried sources, corrections for the loss of [.sup.68] Ge due to chemical volatility were necessary in both studies. This effect is reported by Grigorescu to be on the order of 20% to 26%. Nonetheless, they were able to obtain a measurement result with about a 1% relative standard uncertainty.

Liquid scintillation (LS) counting has been the method of choice in our laboratory for the measurement of [beta]-emitting radionuclides, primarily due to the high LS detection efficiency and the relative ease of sample preparation. Methods based on LS counting have another advantage in the context of measuring [.sup.68] Ge because the sample is introduced into the cocktail while still in solution, thereby eliminating the need to prepare dry sources. Seeking to take advantage of this, we have measured a single solution of [.sup.68] Ge/[.sup.68] Ga using three LS-based method: live-timed 4 [pi] [beta] - [gamma] anticonincidence (LTAC) using LS as the [beta] counting channel, the Triple-to-Double Coincidence Ratio Method (TDCR) (9), (10), and CNET (6), (7).

2. Experimental

All evaluation of measurement uncertainties throughout this work follow accepted conventions used by the NIST Radioactivity Group and are in accordance with hose recommended by the principal metrology organizations (11). All individual uncertainty components are given as estimated experimental standard deviations (or standard deviations of the mean, if appropriate), or quantities assumed to correspond to standard deviations regardless of the method used to evaluate their magnitude. Unless explicitly stated, all uncertainties cited in this paper are "standard uncertainties," corresponding to one uncertainty interval. One particular exception is the uncertainty reported for the activity concentration of the calibrated [.sup.68] Ge solution, which is given as an "expanded combined standard uncertainty." In accordance with NIST policy (12), the combined standard uncertainty (calculated by combining the individual uncertainty components in quadrature) is multiplied by a "coverage factor" of k = 2 to obtain an "expanded uncertainty" assumed to give an uncertainty interval having a confidence level of 90% to 95%.

2.1 Initial Solution Preparation

The master solution used in these experiments contained nominally 125 MBq [.sup.68]Ge in 5 mL of 0.5 mol. [L.sup.-1] HCl and was prepared by International Isotopes Idaho, Inc. [(Idaho Falls, ID) (2) using [.sup.68]Ge produced at the 100 MeV Isotope Production Facility at Los Alamos National Laboratory using the [.sup.nat]Ga(p,2n) [.sup.68]Ge reaction.

A generalized scheme for the preparation of the counting samples is shown in Fig. 2. The first step involved transfer of the master solution out of the shipping vial into a NIST standard 5 mL flame-sealed ampoule while at the same time performing the first of three serial dilutions that would be needed in order to bring the activity level in one of the ampoules down to that suitable for LS counting. The ampoule used for this study, labeled A1, was prepared by volumetrically adding 1 mL of the stock solution to an ampoule containing 4 g of gravimetrically added carrier solution having nominally 45 [micro]g each of nonradioactive [Ge.sup.+4] and [Ga.sup.+3] per gram of solution using 0.5 mol * [L.sup.-1] HCl as the solvent. The ampoule was weighed again after the addition of the [.sup.68]Ge to determine the mass of added radioactive solution. Ampoule A2, shown in Fig. 2, was held in reserve for future experiments.

Ampoule A1 was measured in the NIST-maintained radionuclide activity calibrators ("dose calibrators") and the NIST "4[pi]" [gamma] IC (13) to allow for the determination of calibration factors in this specific measurement geometry. The solution in A1 was then diluted by a factor of about 200 through two serial gravimetric dilutions, giving two additional ampoules, A1D1 and A1D2. As an additional check of the dilution factor between A1 and A1D1, the latter was also measured in the NIST IC.

2.2 Liquid Scintillation Source Preparation

All counting sources for the three counting techniques were prepared using solution AID2. A total of 18 LS cocktails containing [.sup.68]Ge were prepared for these studies. For the LTAC experiments, 4 mL of HiSafe-3 (Perkin Elmer, Waltham, MA) or PCS (GE Healthcare Biosciences, Piscataway, NJ) scintillant were added to each of two 3 cm diameter, glass pseudo-hemispheres. Nominally 0.04 g of [.sup.68]Ge solution were gravimetrically added and the hemispheres were sealed using epoxy.

For the TDCR experiments, two cocktails each of HiSafe-3 and PCS were prepared by dispensing 10mL of the scintillant into four 22 mL borosilicate glass LS vials (two per scintillant), followed by the gravimetric addition of nominally 0.04 g of solution from AID2. Similarly, 10 cocktails were prepared for the CNET experiments (five vials per scintillant). To vary the counting efficiency of the CNET cocktails, between 2 drops and 18 drops of a 10:1 (by volume) dilution of nitromethane in ethanol were added as a quenching agent to the CNET cocktails in addition to the scintillant and radioactive solution. In order to perform the efficiency tracing, a separate set of 10 LS vials having identical composition to the [.sup.68]Ge cocktails were prepared using a dilution of a NIST tritiated water Standard Reference Material 4927F (14) in place of the (68) Ge. In order to make the [.sup.68]Ge and [.sup.3]H cocktails as close in composition as possible, nominally 1 mL of the [Ge.sup.4+]/[Ga.sup.3+] carrier was added to each of the cocktails.

Two background blanks (one for each scintillant) were prepared for the TDCR and LTAC measurements in their appropriate vials. In order to properly mimic the composition of the radioactive cocktails, an equivalent mass of nonradioactive carrier solution was added to each blank. For the CNET measurements, four blanks were prepared so as to have the identical sample compositions of the least- and most-quenched of the radioactive [.sup.68]Ge cocktails. As with the TDCR and LTAC blanks, nonradioactive Ge/Ga carrier was substituted for the [.sup.68]Ge solution.

2.3 4[pi] [beta] - [gamma] Anticoincidence Counting (LTAC)

The system constructed at NIST uses an LS source optically coupled to an appropriate photomultiplier tube for the beta channel and a thallium-dopted sodium iodide [NaI(TI)] detector for the [gamma]-ray channel, as described previously (15), (16). The LS-based system is well suited to this case since the source solution does not have to be dried, and therefore, the large (20% to 26%) correction for Ge loss reported by Grigorescu (8) is avoided.

The four active sources were each measured for between 2 and 4 cycles and the blank sources up to 3 cycles during the period from 24 April to 5 May 2007. Each counting cycle consisted of measurements at between 8 to 12 threshold levels on the LS detector for between 200 and 1000 seconds. A minimum of 5 * [10.sup.6] LS and 1 * [10.sup.5] anti-coincident Nal detector counts were recorded for each non-blank measurement. The LS signal-to-background ratio for the lowest threshold (highest background) data points was about 950:1, while the signal-to-background ratio for the NaI detector was about 350:1. Further systematic tests demonstrated that the background variability during the span of the experiment, the variation of extending dead-time, and the presence or absence of the aluminum absorber did not affect the measurement results.

The position decay of the [.sup.68]Ga was detected in the LS channel, with count rate [N.sub.[beta]], while electron capture events from both [.sup.68]Ga and [.sup.68]Ge were avoided by constraining the lower level discriminator (LLD) on the amplified signal to be above about 20 V beta energy. In this way, the determined activity value was independent of any atomic transitions (all below 11 ke V), and directly proportional to the total positron emission probability. The LS positron efficiency, [[epsilon].sub.[beta]], was varied between about 0.5 and 0.95 using the LLD and extrapolated to 1.0. The NaI (Tl) detector was gated on the 511 ke V region using a single channel analyzer and the total [gamma]-ray ([N.sub.[gamma]] and anticoincidence ([N.sub.AC]) count rates were recorded. The extrapolation parameter used was Y [equivalent. to] [N.sub.AC] / [N.sub.[gamma]] [approximately equal to] (1 - [[epsilon].sub.[beta[). Most of the [gamma]-ray counts were due to positron-annihilation decays, detected with efficiency,

[[epsilon].sub.ann][congruent to][[N.sub.[gamma]]/[[N.sub.0]([b.sub.1] + [b.sub.2])], (1)

where [N.sub.0] is the activity and ([b.sub.1] + [b.sub.2]) is the total positron emission probability. There was an additional, approximately 0.2%, contribution from Compton scattering of 1077 ke V [gamma]-rays, detected with efficiency [[epsilon].sub.[gamma] 1077]. Since some of these 1077 keV [gamma]-rays correspond to electron capture events, and not positron emission, a small (0.2%) correction to the intercept was necessary. The modified extrapolation formula is,

[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII] (2)

where the numerator and denominator of the correction term correspond to total [gamma]-rays and anticoincident [gamma]-rays, respectively, and the branching probabilities [b.sub.1], [b.sub.2], and [b.sub.3] are illustrated and enumerated in Fig. 1. Note that the extrapolation is linear in Y, and the Y = 0 ([[epsilon].sub.[beta]] = 1]) intercept is given by,

[MATHEMATICAL EXPRESSION NOT REPRODUCIBLE IN ASCII] (3)

Equation (3) is similar to Eq. (4) IN (8), with a significant difference that here we account for the fact that some of the 1077 keV [gamma]-rays do correspond to a positron branch, and thus do not need to be subtracted. This effect is accommodated by the presence of [b.sub.2] in the numerator of the correction term (final term in Eq. (3)). Corrections due to the LS efficiency for [gamma]-rays and for coincidences due to those events were not necessary, as described below.

An estimate of [[epsilon].sub.[gamma]1077] was obtained during a separate set of measurements with a [.sup.60]Co point-source and [[epsilon].sub.ann] was estimated using Eq. (1). The correction term was checked by exaggerating the effect during additional measurements made with various [gamma]-ray energy gates. Gates Gl, G2, and G3 corresponded to a gate around the Compton region of the annihilation spectrum, the annihilation photopeak (511 keV) and the annihilation sum peak (1022 keV) respectively. The sum peak was unresolved from the 1077 keV peak in the NaI (Tl) detector events. As designed, the data from gates G1 and G3 required large corrections due to reduced [[epsilon]sub.ann] and enhanced [[epsilon].sub.1007] respectively. The uncorrected and corrected [N.sub.0] values are shown in Table 1. The fact that the large corrections for G1 and G3 gave consistent results with G2, supports the use of this method for the small (0.2 %) correction to the final [N.sub.0] value, based on G2 alone.

Table 1. Calculated [.sup.68] activity from various [gamma]-ray gates
and using Eq. (3), with and without the final correction term, and
relative to the corrected G2 value, (used for the final activity
determination). The uncertainties (k = 1) on the uncorrected activities
are standard deviations of the intercepts from the least-squares fits
to the various data sets. The uncertainties (k = 1) of the corrected
values are from estimates of the uncertainties in [[epsilon].sub.ann]
and [[epsilon].sub.[gamma]1077]

[gamma]-ray gate  Uncorrected [N.sub.0] (Bq)  Corrected [N.sub.0] (Bq)

G1 - Comptons        10020 [+ or -] 0.022       1.002 [+ or -] 0.004
G2 - photopeak      1.0021 [+ or -] 0.001      1.0000 [+ or -] 0.0004
G3 - sum peak       1.033  [+ or -] 0.002       1.000 [+ or -] 0.007

No correction in the final result was made for the gamma efficiency of the LS detector, or for coincidences between such events and the NaI (TI) detector. For gates G1 and G2, these effects tend to cancel out due to the two-photon annihilation process. If one photon is detected in the LS detector, that efficiency can be monitored by the other photon interacting in the NaI detector (15). If such an effect were present it would lead to a non-linear efficiency extrapolation. A typical G2 data set and residuals from a linear least-squares fit are shown in Fig. 3 and it is evident that a linear fit is satisfactory. Yet, a quadratic extrapolation was needed to fit the entire range of G3 (sum peak) data due to the unmonitored LS efficiency for [gamma]-rays in that configuration. Thus, for G3, a smaller [[epsilon].sup. [beta]] range (0.9-0.95) was employed such that linear and quadratic fits gave consistent results. This value was only used for the [[beta].sub.[gamma]1077] sensitivity test.

[FIGURE 3 OMITTED]

Another possible cause for a non-linear extrapolation would be if both the LS and NaI(T1) efficiencies differed depending on whether the positron was stopped in the LS hemisphere, or escaped before annihilating. This effect was mitigated by three factors: (1) the fact that most positrons annihilated within the hemisphere, (2) the high LS efficiency, and (3) the well-type geometry of the NaI(T1) detector. The sensitivity of the result to this effect was tested by placing an approximately 0.5 cm thick aluminum foil ever the hemisphere and comparing the resulting activity determination. No change in the goodness of the linear fit was detected and the ration of the intercept with to without the foil was 1.000 [+ or -] 0.001, where the uncertainty is a standard (k = 1) uncertainty on the linear fit coefficients.

2.4 Liquid Scintillation Counting Using the Triple-to-Double Coincidence Ratio (TDCR) Method

Each counting source was counted in the NISTTDCR system (17) on at least two separate occasions over the course of 27 days. Counting times were typically 1200 s, which allowed for the accumulation of at least [10.sup.6] counts in each of the three doubles counting channels. For each counting experiment, data were acquired at a minimum of 4 efficiency; points, which were varied through the use of a set of grey filters that were fitted over the LS vials. Data were acquired in triplicate at each efficiency point. The experimental efficiencies for the logical sum of double photon coincidence events, [[beta].sub.LSD], ranged from 0.89 to 1.14.

The counting data were analyzed using a program developed in-house for use with the Mathematica (18) symbolic mathematics package. Details of the program and the computation strategy will be published separately. However, it should be noted that the program calculates the total detection efficiency for the case of decay of [.sup.68]Ge in equilibrium with its [.sup.68]Ga daughter. To do this, the program was required to solve the TDCR equations (9), (10), (19), (20) for the EC branch of the [.sup.68]Ge parent, as well as both the EC and positron decay of the [.sup.68]Ga daughter. A relatively simplistic model, considering twelve possible decay pathways, was adopted to describe the atomic transitions encountered in the EC decay of [.sup.68]Ge and [.sup.68]Ga. These are depicted in Fig. 4. The values of the various nuclear and atomic input data were taken from the evaluation of the Decay Data Evaluation Project (DDEP) (4).

[FIGURE 4 OMITTED]

The analysis program calculates the individual phototube efficiencies, thereby allowing for correction due to asymmetry in the counting; rates in each of the detection efficiency due to detection of the 511 KeV annihilation photons was taken into account by using the positron spectrum calculated by the program SPEBETA (21) as input for the Monte Carlo simulation package PENELOPE (22) using the techniques described in (23). The resulting spectrum of energy (positrons+annihilation photons) absorbed in the LS cocktail was then used as input data for the TDCR analysis code.

The stopping power, dE/dx, for electrons in the LS cocktail was calculated by fitting a function of the form

(dE/dx) = a + bE + c[(lnE).sup.2] + dlnE/E + e/E (4)

(E is the value of the midpoint energy for each bin of the calculated beta spectrum and a, b, c, d, and e are fitting parameters) to data from the NIST ESTAR (24) database using previously published LS cocktail compositions (25).

A separate program, assuming equal phototube efficiencies, was developed for evaluating the effects of varying different input and model parameters. Calculations of [[epsilon].sub.LSD] were made as a function of the TDCR for kB values between 0.009 cm * Me[V.sup.-1] and 0.018 cm * Me[V.sup.-1] and the resulting [[epsilon].sub.LSD] values were found to be insensitive to the of kB. For consistency with previous measurements made in this laboratory (26), the value of kB for all analyses was taken to be 0.012 cm * Me[V.sup.-1]. A plot of the theoretical [[epsilon].sub.LSD] values as a function of TDCR at kB = 0.012 cm. Me[V.sup.-1] is shown in Fig. 5.

[FIGURE 5 OMITTED]

2.5. Liquid Scintillation Counting Using the CIEMAT-NIST [.sup.3]H-Standard Efficiency Tracing (CNET) Method

Each LS cocktail was sequentially counted for 10 cycles of 25 min per source on a Packard (Perkin Elmer, Waltham, MA) 2500TR LS spectrometer. Samples were then removed from the counter, agitated and sequentially counted for 10 cycles of 30 min per source in a Beckman LS6500 (Beckman Coulter, Fullerton, CA) spectrometer.

Efficiency tracing involves calculating a relationship between the measured [.sup.3]H LS efficiencies and the LS efficiencies expected for [.sup.68]Ge, in equilibrium with its daughter [.sup.68]Ga, over a range of experimental quench indicating parameters (7). The efficiency tracing computer program CN2004 (27) was used in the analysis of the LS data after changing the default input file to include the nuclear and atomic data found in the DDEP evaluation (4). The average calculated [.sup.68]Ge/[.sup.68]Ga efficiency was nominally 138% in the Packard LS counter and 147% in the Wallac LS counter using a kB value of 0.0075 cm * Me[V.sup.-1] and assuming that the cocktail had the composition of Ultima Gold as specified in the default CN2004 input files. Plots of the calculated theoretical [.sup.68]Ge, [.sup.68]Ga, and total efficiencies as a function of [.sup.3]H tritium are given in Fig. 6.

[FIGURE 6 OMITTED]

2.6. Ionization Chamber Measurements

For the NIST IC measurements, both A1 and A1D1 were measured 40 times each, in four groups of 10 measurements, alternating with 5 groups of 10 measurements of either radium ([.sup.226]Ra) reference sources RRS100 or RRS500b. Results are analyzed as a ratio of the response of the ampoule to the response of the RRS. After correction for background, the resulting ratio is used to derive a calibration factor, or K-value, defined as the activity of a given radionuclide that would produce the same response as the RRS. The relative values of the RRS100 and RRS500b are well characterized. By determining the K-value using the activity derived from different ampoules of different activity levels, it is also possible to verify the gravimetric dilution factor. The dilution factor from A1 to A1D1 was verified by this method to within 0.022%. The LTAC activity values and the mass dispensed into A1 were used to determine K-values that can be used for future measurements of [.sup.68]Ge in the NIST ampoule geometry.

2.7 Gamma Ray Spectrometry

The solution that remained in A1D2 after making the LS cocktails was analyzed for possible photon-emitting radionuclidic impurities using two calibrated High-Purity Germanium (HPGe) photon spectrometers at two different counting distances each. In addition, the data provided an additional, confirmatory measurement of the activity concentration using the 1078 keV gamma ray from the decay of [.sup.68]Ga. Characteristics of the detectors used in this study are given in Table 2.

Table 2. Characteristics of HPGe detectors used in the present study

Detector parameter           X-detector               B-detector

Detector diameter       43.6 [+ or -] 0.1 mm    54.9 [+ or -] 0.1 mm

Detector length         36.2 [+ or -] 0.1 mm    54.2 + 0.05 mm

End cap window          Beryllium               Beryllium
material

Window thickness        0.5 [+ or -] 0.05 mm    0.5 [+ or -] 0.05 mm

Crystal-window          3 [+ or -] 0.5 mm       3 [+ or -] 0.5 mm
distance

Crystal top dead zone   0.3 [+ or -] 0.03       0.3 [+ or -] 0.03
thickness               [micro]m                [micro]m

Crystal material        Germanium               Germanium

Crystal hole depth      32.6 [+ or -] 0.5 mm    47.2 [+ or -] 0.5 mm

Crystal hole diameter   10.4 [+ or -] 0.5 mm    12 [+ or -] 0.1 mm

Detector side cap       1.3 [+ or -] 0.1 mm     1.3 [+ or -] 0.1 mm
thickness

Detector side cap       70 [+ or -] 1 mm        63.5 [+ or -] 0.5 mm
diameter

Detector side cap       Aluminum                Magnesium
material

Detector type           n-type                  n-type

Calibration Geometries  Ampoule; side-mount,    Ampoule; 24 cm, 35 cm
(distances are          end-on 24 cm
source-to-detector

Data were collected using the Gamma Vision-32 (Ortec, Oak Ridge, TN) software package and analyzed using both Gamma Vision-32 and Genie 2000 (Canberra, Meriden, CT). Detection efficiencies were calculated from efficiency-energy relationships determined using solutions previously calibrated at NIST and measured in the 5 mL NIST ampoule geometry.

3. Results and Discussion

3.1 Impurity Analyses

No Photon-emitting radionuclidic impurities were detected in solution A1 to within the following limits (at the reference time) of the massic photon emission rate:

785[s.sup.-1] * [g.sup.-1] for 30 KeV [less then or equal to] E [less then or equal to] 507 KeV; and 285[s.sup.-1] * [g.sup.-1] for 515 KeV [less then or equal to] E [less then or equal to] 1800 KeV;

where E is the gamma-ray energy.

3.2 Activity Measurements Results

The results of the massic activity determinations for the solution contained in A1 as of the reference time are given in Table 3. The values in Table 3 take into account the dilution factor of 204.903231 between the solution in A1 and that used in the assays, A1D2. The uncertainties given in the table are expanded (k = 2) uncertainties based on the components given in Tables 4-7.

Table 3. Results of massic activity determinations ([C.sub.A], in Bq *
[g.sup.-1]) for the (68)Ge solution contained in ampoule A1 as of the
reference time of 12:00 EST 1 May 2007. The uncertainties, given in
parentheses, are expanded (k = 2) uncertainties based on the evaluated
uncertainty components listed in Tables 4-7 for the respective
techniques

Technique                                           [C.sub.A],
                                                    [10.sup.6] Bq *
                                                    [g.sup.-1]

4[pi][beta] - [gamma] anticoincidence counting        3.104(18)
(LTAC)

LS counting with the Triple-to-Double Ratio (TDCR)    3.141(25)
method

LS counting with the CIEMAT/NIST (3)H-standard        3.058(44)
efficiencytracing method (CNET)

Gamma-ray spectrophotometry with High Purity          3.2(9)
Germanium (HPGe) detectors

Table 4. Uncertainty components evaluated in the determination of the
massic activity, [C.sub.A], for [.sup.68]Ge solution A1 by 4[pi][beta]
- anticoincidence [gamma] counting (LTAC]

Component,            Comment                   Evaluation    %
[u.sub.i]

Measurement        Standard deviation of the         A       0.03
variability        mean of the determination
                   of [C.sub.A] for 8 trials
                   encompassing 4 samples, 3
                   backgrounds, 2 extending
                   dead-times and absorber/no
                   absorber-over a 2-week
                   period

Background         Additional variation              A       0.03
variability        estimate based on 3 trials
                   and within-run variation

Additional         Additional variability not        B       0.05
variability in     embodied by "random
background         sources"

Half-life          Standard uncertainty in           B       0.0008
                   half-life. (0.059%) over
                   the measurement decay
                   interval

Live time          Estimated standard                B       0.1
                   uncertainty on [C.sub.A]
                   due to uncertainty in
                   counting livetime.

Branching ratio    Estimated standard                B       0.13
                   uncertainty due to
                   uncertainty in published
                   decay branching ratios

Extrapolation      Estimated standard                B       0.2
                   uncertainty due to
                   extrapolation to zero
                   non-detection efficiency;
                   based on sensitivity tests,
                   previous measurements and
                   models

Correction due to  Estimated standard                B       0.1
detection of 1077  uncertainty on [C.sub.A]
keV photons`       due to efficiency of
                   detecting 1077 keV photons
                   based on sensitivity tests

Mass               Estimated standard                B       0.05
determinations     uncertainty of mass for any
                   single LS cocktail

Dilution factor    Uncertainty in [C.sub.A] of       B       0.04
                   solution Ge1A1 due to
                   uncertainty in
                   gravimetrically-determined
                   dilution factor between
                   solutions in ampoules Ge1A1
                   and Ge1A1D2

Combined                                                     0.29
([u.sub.c] =
[square root of]
([SIGMA]
[u.sub.i.sup.2])])

Expanded                                                     0.58
([U.sub.c] =
[u.sub.c] * k; k
= 2)

Table 5. Uncertainty components evaluated in the determination of the
massic activity, [C.sub.A], for [.sup.68]Ge solution A1 by liquid
scintillation counting using (TDCR) method

Component,                    Comment            Evaluation      %
[u.sub.i]

Sample              Standard deviation of the        A         0.06
repeatability       mean on the determination
                    of massic activity for a
                    single LS cocktail (n= 3 -
                    5) determinations of
                    [C.sub.A] per source)

LS cocktail         Standard deviation on the        A         0.09
composition         determination of [C.sub.A]
variability         of three LS cocktail
                    compositions (n = 6 - 19
                    determinations of [C.sub.A]
                    per composition)

Efficiency          Median difference between        B         0.20
dependence          maximum and minimum value
                    of [C.sub.A] determined for
                    a single source at between
                    3 and 4 efficiency values,
                    varied by use of grey
                    filters (n = 12 independent
                    measurements)

Effect of (68)Ga    Standard uncertainty in          B         0.18
beta endpoint       efficiency calculation due
energy,             to standard uncertainties
[E.sub.[beta],max]  on positron endpoint
on efficiency       energies of (68)Ga
calculations

Effect of other     Standard uncertainty due to      A         0.26
atomic and nuclear  uncertainties on data used
input data          as input to the TDCR
                    analysis code as determined
                    by Monte Carlo methods. A
                    total of 20 data sets were
                    generated from normal
                    distributions defined by
                    the published nuclear and
                    atomic data and their
                    associated standard
                    uncertainties, which were
                    taken as the standard
                    deviation of the respective
                    distributions. Each data
                    set was used to calculate
                    [C.sub.A] using a single
                    experimental data set

Half-life           Standard uncertainty in          B        6 * 10-3
                    half-life (0.059%) over the
                    measurement decay interval

Mass                Estimated standard               B         0.05
determinations      uncertainty of mass for any
                    single LS cocktail

Livetime            Standard uncertainty             B           7 *
                    arising from an estimated                [10.sup.-3]
                    uncertainty of 0.007% on
                    the determination of the
                    live time

Background          Standard deviation on the        A           5 *
                    determination of [C.sub.A]               [10.sup.-3]
                    determined via Monte Carlo
                    simulation. A total of 5
                    background data sets were
                    constructed from random
                    data arising from normal
                    distributions defined by
                    the average and standard
                    deviation of experimental
                    backgrounds at 4 efficiency
                    points having 3 repetitions
                    each; calculations were
                    carried out with all 5
                    background data sets for a
                    single experimental data
                    set

Dilution factor     Uncertainty in [C.sub.A] of      B         0.04
                    solution Ge1A1 due to
                    uncertainty in
                    gravimetrically-determined
                    dilution factor between
                    solutions in ampoules Ge1A1
                    and Ge1A1D2

Combined                                                       0.39
([u.sub.c] =
[square root of
([SIGMA]
[u.sub.i.sup.2])])

Expanded                                                       0.79
([U.sub.c] =
[u.sub.c] * k; k
= 2)

Table 6. Uncertainty components evaluated in the determination of
the massic activity, [C.sub.A], for (68) Ge solution A1 by liquid
scintillation counting

Component,                    Comment            Evaluation      %
[u.sub.i]                                           type

Sample              Standard deviation of the        A          0.05
repeatability       mean on the determination
                    of massic acitivity for a
                    single LS cocktail (n = 10
                    determinations of [C.sub.A]
                    per source)

LS measurement      Standard deviation on the        A          0.24
reproducibility     determination of [C.sub.A]
                    for 10 cockatils of 2
                    compositions

Mass                Estimated standard               B          0.05
determinations      uncertainty of [.sup.68]Ge
                    mass for any single LS
                    cocktail

Dilution factor     Uncertainty in [C.sub.A] of      B          0.04
                    solution Ge 1A1 due to
                    uncertainty in
                    gravimetrically-determined
                    dilution factor between
                    solutions in ampoules Ge1A1
                    and Ge 1A1 D2

(68) Ge decay       Standard uncertainty in          B          0.001
corrections         half-life (0.059 %) over
                    the measurement decay
                    interval

(68) Ge             Estimated uncertainty in         B          0.65
efficiency          [C.sub.A] due to step size
                    in CN2004 calculations

Livetime            Estimated uncertainty in         B          0.05
determinations      the correction to the LS                    (and
                    counting interval                           PE) (3)

Background          Estimated uncertainty due        B          0.004
                    to an average 4%
                    uncertainty in background
                    determination

                    Estimated uncertainty due        B          0.18
                    to 0.36% uncertainty in
                    [.sup.3]H standard
                    activity

Branching ratios    Estimated uncertainty due        B          0.08
                    to uncertainty in branching
                    ratios

Combined                                                        0.73
([u.sub.c] =
[square root of
([SIGMA]
[u.sub.i.sup.2])])

Expanded (U.sub.c]                                              1.45
= [u.sub.c] .k; k
= 2)

(3) The relative uncertainty for this component is partially embodied
(PE) in the relative standard uncertainties of the repeatability and
reproducibility components.

Table 7. Uncertainly components evaluated in the determination of the
massic activity, [C.sub.A], for [.sup.68]Ge solution Ge1A1 by
[gamma]-ray spectrometry using HPGe detectors

Component,                    Comment            Evaluation       %
[u.sub.i]                                           type

Measurement         Standard deviation on            A           0.98
repeatability       determination of [C.sub.A]
                    for 3 repeated measurement
                    of a single source at a
                    single geometry

Efficiency curve    Standard deviation of the        B           0.12
                    mean n determination of
                    detection efficiency for 4
                    sample geometries

Sample geometry     Typical uncertainly due to       B           0.33
                    change of sample geometry
                    (detector and
                    source-to-detector
                    distance) for a single
                    counting source

Decay correction    Standard uncertainly in          B           2.6 *
                    half-life (0.059%) over the              [10.sub.-3]
                    measurement decay interval

Decay data          Standard uncertainly             B           0.93
                    (0.93%) on emission
                    probablity of 1078 keV
                    gamma-ray in the decay of
                    [.sup.68]Ge

Dilution factor     Estimated standard               B           0.04
                    uncertainly in [C.sub.A] of
                    solution Ge1A1 due to
                    uncertainty in
                    gravimetrically-determined
                    dilution factor between
                    solutions in ampoules Ge1A1
                    and Ge1A1D2

Combined                                                         1.4
([u.sub.c] =
[square root of
[sigma]
[u.sub.i.sup.2])])

Expanded                                                         2.8
([U.sub.c] =
[u.sub.c] * k;
k = 2)

Of the different techniques used in this study to determine the activity concentration of the [.sup.68] Ge solution, the LTAC technique is much less dependent on level scheme data and model parameters not directly measured in the experiment. For this particular measurement, the only input parameter significantly impacting the activity calculation that was not directly measured in the experiment was the positron branching ratio. The other branching ratios only contributed to the minor (0.2%) correction for the leakage of 1077 KeV [gamma]-rays into the annihilation [gamma]-ray gate. And even this small correction was checked experimentally by modifying the experimental design to exaggerate the effect and then verifying that the corrected activity agreed with the original value.

On the other hand, our implementations of the TDCR and CNET efficiency tracing methods are unable to separate the positron and EC decay signals and must therefore account for all possible decay paths, including atomic rearrangement following electron capture. From a practical standpoint, a compromise between treating all possible paths and reasonable computation times must be made. While this certainly introduces some small amount of uncertainty, it is not expected that the weak contributions due to paths not considered in Fig. 4 would be significant, at least for the TDCR method. Instead, as seen in Table 5, the uncertainties on the input data play a very significant role.

Because of the more direct nature of the measurement in the LTAC technique, the LTAC activity value for the solution in A1 was adopted as the reference value for this study and was used in the calculation of the K-value for the NIST IC. The fact that the LTAC and TDCR measurement agree to within their respective experimental uncertainties is encouraging, given the complexity of the TDCR efficiency calculation. Nonetheless, one would hope that improvement in the NIST TDCR spectrometer would lead to higher EC detection efficiencies, thereby providing better results in the measurement of radionuclides that decay by this mode. The CNET results indicate that some improvements in the method are still needed to be able to reliably measure nuclides that undergo EC decay.

3.3 Determination of K-Value for NIST IC

In order to avoid the need to perform a primary standardization every time a NIST-calibrated solution of [.sup.68]Ge/[.sup.68]Ga is required, we determined a calibration factor (K-value) for the NIST IC. This K-value is not to be confused with the coverage factor, k, applied to uncertainty evaluations. Using the LTAC reference activity value and the measured responses in the IC against radium reference sources (RR) 500B and 100, the K-values were found to be 2.695 X [10.sup.7] [+ or -] 1.7 X [10.sup.5] Bq and 5.032 X [10.sup.6] [+ or -] 3.2 X [10.sup.4] Bq, respectively. The uncertainties on the K-values are expanded (k = 2) uncertainty and include relative standard uncertainty components due to the original primary standardization (0.29%), repeatability on 40 measurements in the IC (0.015%), source mass (0.05%) decay correction (0.002%), and source positioning (0.1%).

4. Conclusion

A solution containing [.sup.68]Ge in equilibrium with its decay daughter [.sup.68]Ga has been standardized for the first time at NIST, with a combined standard uncertainty of 0.29% using LTAC. Measurements made with two other LS techniques, TDCR and CNET, confirmed the LTAC result to within 1.1% and 1.5% respectively. The differences between results obtained with the latter two methods and the LTAC technique indicate that improvements in the models and/or their applications are needed, particularly for EC nuclides.

Data collected on the NIST 4[pi][gamma] ionization chamber allowed for the determination of calibration factors for that chamber in the 5 mL NIST ampoule geometry, thereby enabling future calibrations of solutions having the same solution composition without the need for the measurements to be made by a primary method.

Acknowledgments

The authors wish to thank Ms. Michelle Hammond and Dr. Leticia Pibida for performing the HPGe measurements.

A portion of this work was supported by RadQual, LLC.

5. References

(1) G. Wiley, PET/CT: Market & Business Models for Ownership, Imaging Economics 6 (2006).

(2) B. M. Coursey, D. D. Hoppes, M. P. Unterweger, A. GrauMalonda, R. A. Manning, Standardization of (18)F for use in Positron-Emission Tomography, Int. J. Appl. Radiat. Isot. 34, 1181-1189 (1983).

(3) B. E. Zimmerman, G .J. Kubicek, J. T. Cessna, P. S. Plascjak, and W. C. Eckelman, Radioassays and experimental evaluation of dose calibrator settings for [.sup.18]FDG, Appl.. Radiat. Isot. 54, 113-122 (2001).

(4) DDEP, Decay Data Evaluation Project Data, http://www.nucleide.org/DDEP_WG/DDEPdata.htm (2008).

(5) E. Schonfeld, U. Schotzig, E. Gunter, and H. Schrader, Standardization and decay data of [.sup.68]Ge/[.sup.68]Ga, Appl. Radiat, Isot. 45, 955-961 (1994).

(6) B. M. Coursey, W. B. Mann, A. Grau Malonda, E. Garcia-Torano, J. M. Los Arcos, J. A. B. Gibson, and D. Reher, standardization of carbon-14 by 4[pi][beta] liquid scintillation efficiency tracing with hydrogen-3, Appl. Radiat. Isot. 5, 403-408 (1986).

(7) B. E. Zimmerman and R. Colle, Standardization of [.sup.63]Ni by 4[pi] [beta] liquid scintillation spectrometry with (3) H-standard efficiency tracing, J. Res. Natl. Inst. Stand. Technol. 102, 455-477 (1997).

(8) E. L. Grigorescu, C. D. Negut, A. Luca, A. C. Razdolescu, and M. Tanase, Standardization of [.sup.68](Ge + Ga), Appl. Radiat. Isot. 60, 429-432 (2004).

(9) R. Broda and K. Pochwalski, the enhanced triple to double coincidence ratio (ETDCR) method for standardization of radionuclides by liquid scintillation counting, Nucl. Inst. Meth. Phys. Res. A312, 85-89 (1992).

(10) K. Pochwalski, R. Broda, and T. Radoszewski, Standardization of pure beta emitters by liquid scintillation counting, Appl. Radiat. Isot, 39, 165-172 (1988).

(11) International Organization for Standardization (ISO), Guide to the Expression of Uncertainly in Measurement (ISO, Geneva, 1993).

(12) B. N. Taylor and C. E. Kuyatt, Guidelines for evaluating and expressing the uncertainty of NIST measurement results, NIST Technical Note 1297 (National Institute of Standards and Technology, Gaithersburg, MD, 1994).

(13) J. M. Calhoun, NBS Special Publication 250-10: Radioactivity calibrations with the '4[pie]' gamma ionization chamber and other radioactivity calibration capabilities, NBS Special Publication 250-10 (National Institute of Standards and Technology, Gaithersburg, MD, 1987).

(14) National Institute of Standards and Technology, Standard Reference Material 4927F, Radioactivity Standard, Hydrogen-3, (National Institute of Standards and Technology, Gaithersburg, MD, 2000).

(15) R. Fitzgerald and M. K. Schultz, Liquid scintillation-based anticoincidence counting of Co-60 and Pb-210, Appl. Radiat. Isot. 66, 937-940 (2008).

(16) L. L. Lucas, Calibration of the massic activity of a solution of Te-99, Appl. Radit Isot. 49, 1061-1064 (1998).

(17) B. E. Zimmerman, R. Colle, and J. T. Cessna, Construction and implementation of the NIST triple-to-double coincidence ratio (TDCR) spectrometer, Appl. Radiat. Isot. 60, 433-438 (2004).

(18) Wolfram Research, Mathematica, v.4.0, (2001).

(19) R. Broda, A review of the triple-to-double coincidence ratio (TDCR) method for standardizing radionuclides, Appl. Radiat, Isot. 58, 585-594 (2003).

(20) P. Cassette, M. M. Be, F. Jaubert, and M. C. Lepy, Measurement of a [.sup.103]Pd solution using the TDCR method by LSC, Appl. Radiat. Isot. 60, 439-446 (2004).

(21) P. Cassette, SPEBETA programme de calcul du spectre en energie des electrons emis par des radionucleides emetteurs beta, CEA Technical Note (Commissariat a 1'Energie Atomique (CEA), Saclay, France (1992).

(22) J. Sempau, E. Acosta, J. M. Fernandez-Varea, and F. Salvat, An algorithm for Monte Carlo simulation of the coupled electronphoton transport., Nucl. Inst, Meth. B132, 377-390 (1997).

(23) B. E. Zimmerman, Monte Carlo calculations of spectra and interaction probabilities for photons in liquid scintillators for use in the standardization of radionuclides, Appl. Radiat. Isot. 64, 1492-1498 (2006).

(24) National Institute of Standards and Technology, Electron Stopping Power and Range Tables, http://physics.nist.gov/STAR (2008).

(25) Labortoire National Henri Beequerel, LS Sources Data, http://www.nucleide.org/ICRM_LSC_WG/icrmlsdata.htm (2008).

(26) R. Colle, B. E. Zimmerman, P. Cassette, and L. LaureanoPerez, [.sup.63]Ni, its half-life and standardization: revisited, Appl. Radiat. Isot. 66, 60-68 (2008).

(27) E. Gunther, CIENIST 2004, private communication (2004).

About the authors: Dr. B. E. Zimmerman is a research chemist in the Radioactivity Group (Ionizing Radiation Division) of the NIST Physics Laboratory. Mr. J. T. Cessna and Dr. R. Fitzgerald are physicists in NIST Radioactivity Group. The National Institute of Standards and Technology is an agency of the U.S. Department of Commerce.

(1) CIEMAT is an acronym for Centro de Investigaciones Energeticas, Medioambientales y Technologicas, which is the National Metrology Institute of Spain.

(2) Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

B. E. Zimmerman, J. T. Cessna, and R. Fitzgerald

Ionization Radiation Division, National Institute of Standards and Technology, Gaithersburg, MD 20899

bez@nist.gov

jeffrey.cessna@nist.gov

rayn.fitzgerald@nist.gov

COPYRIGHT 2008 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2008 Gale, Cengage Learning. All rights reserved.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Zimmerman, B.E.; Cessna, J.T.; Fitzgerald, R.
Publication:Journal of Research of the National Institute of Standards and Technology
Date:Sep 1, 2008
Words:7132
Previous Article:Modeling of photochemical reactions in a focused laser beam, II.
Next Article:Long-term stability of the NIST standard ultrasonic source.

Terms of use | Copyright © 2014 Farlex, Inc. | Feedback | For webmasters