St. Jude Medical Discovers Cosmic Radiation Can Affect Memory Chip in Limited Number of Older Generation ICDs.


ST. PAUL St. Paul

as a missionary he fearlessly confronts the “perils of waters, of robbers, in the city, in the wilderness.” [N.T.: II Cor. 11:26]

See : Bravery
, Minn. -- St. Jude Medical St. Jude Medical, Inc. NYSE: STJ is a $2.9 billion global cardiovascular device company, with headquarters in St. Paul, Minnesota, United States. The company sells products in more than 100 countries and has over 20 operations and manufacturing facilities worldwide.  Inc. (NYSE NYSE

See: New York Stock Exchange
:STJ STJ Superior Tribunal de Justica (Brazil)
STJ Supremo Tribunal de Justiça (Portugal)
STJ Superconducting Tunnel Junction
STJ San Giljan (postal locality, Malta) 
) announced today that it has discovered that background levels of atmospheric ionizing cosmic radiation, more commonly known as cosmic rays cosmic rays, charged particles moving at nearly the speed of light reaching the earth from outer space. Primary cosmic rays consist mostly of protons (nuclei of hydrogen atoms), some alpha particles (helium nuclei), and lesser amounts of nuclei of carbon, nitrogen, , can affect a limited number of its older generation implantable cardioverter defibrillator defibrillator, device that delivers an electrical shock to the heart in order to stop certain forms of rapid heart rhythm disturbances (arrhythmias). The shock changes a fibrillation to an organized rhythm or changes a very rapid and ineffective cardiac rhythm to a  (ICD ICD International Classification of Diseases (of the World Health Organization); intrauterine contraceptive device.

ICD
abbr.
) products.

Although the incidence rate is low, and there have been no serious patient injuries or deaths reported to St. Jude Medical attributable to this anomaly, the Company is taking a conservative approach in advising the medical community and regulatory agencies.

Specifically, St. Jude Medical has identified through extensive investigation, including testing at an independent nuclear laboratory, that a particular vendor-supplied static random access memory Static random access memory (SRAM) is a type of semiconductor memory. The word "static" indicates that the memory retains its contents as long as power remains applied, unlike dynamic RAM (DRAM) that needs to be periodically refreshed (nevertheless, SRAM should not be confused with  (SRAM See static RAM.

SRAM - static random-access memory
) chip can be affected, at a low frequency rate, by background levels of atmospheric ionizing cosmic radiation. A full copy of the physician communication, which includes a description of the anomaly and its related clinical risks, as well as St. Jude Medical's recommendations to physicians for patient monitoring, can be found at http://www.sjm.com/companyinformation/physicianletter.html.

This particular memory chip component was used in the following older generation St. Jude Medical ICDs:

--Photon DR (Model V-230HV) (certain serial numbers)

--Photon Micro VR/DR (Models V-194/V-232)

--Atlas VR/DR (Models V-199/V-240)

To date, an incidence of only 0.00167 of the devices at issue (60 out of 36,000) have been found to have been affected by background levels of cosmic radiation. An estimated 26,000 of the device models remain in service. The estimated incidence of an anomaly of this type in the affected device models is 0.00257 over the five year projected life of the device.

"St. Jude Medical has taken a conservative, proactive approach in notifying the medical community of this anomaly," said Dr. Mark Carlson, Professor of Medicine at Case Western Reserve University in Cleveland, Ohio, and a member of an independent panel of physicians that reviewed the issue for St. Jude Medical. Dr. Carlson also recently served as Chairman of a joint Policy Conference on Pacemaker and ICD Performance.

Among the information provided in the physician advisory is St. Jude Medical's recommendation that, if it is not already their practice, physicians should perform routine device monitoring every three months for patients with these devices. St. Jude Medical is also offering its remote monitoring product Housecall Plus(TM), where available, at no charge for models compatible with the remote monitoring system.

"Because the incidence associated with this anomaly is low, St. Jude Medical's recommendation for monitoring at three-month intervals will be appropriate for the vast majority of patients," said Dr. Bruce Wilkoff, director of cardiac pacing and tachyarrhythmia tachyarrhythmia /tachy·ar·rhyth·mia/ (tak?e-ah-rith´me-ah) any disturbance of the heart rhythm in which the heart rate is abnormally increased.

tach·y·ar·rhyth·mi·a
n.
 devices at The Cleveland Clinic and Heart Center and a member of the St. Jude Medical independent Medical Advisory Board.

As part of a new ICD product platform introduced in 2002, and prior to St. Jude Medical having any knowledge of this particular anomaly, St. Jude Medical began using a different vendor and a different design of the SRAM memory chip component. Laboratory testing and clinical experience indicate that this newer generation memory chip component does not share the same susceptibility to background cosmic radiation as the earlier generation. Consequently, other St. Jude Medical ICDs and all models of CRT-D CRT-D Cardiac Resynchronization Therapy - Defibrillators  devices, including the Atlas DR model V-242 and all Epic, Epic HF, Epic +, Epic + HF, Atlas + and Atlas + HF product families, are NOT affected by this issue.

Overall reliability information about the affected ICDs and other St. Jude Medical products can be found in the company's Product Performance Report on the St. Jude Medical web site at www.sjm.com.

St. Jude Medical has notified the U.S. Food and Drug Administration (FDA FDA
abbr.
Food and Drug Administration


FDA,
n.pr See Food and Drug Administration.

FDA,
n.pr the abbreviation for the Food and Drug Administration.
) of this physician advisory. The FDA may determine this communication action to be a recall.

The Company does not expect this physician communication to have any material financial impact.

St. Jude Medical is dedicated to the design, manufacture and distribution of innovative medical devices of the highest quality, offering physicians, patients and payers outstanding clinical performance and demonstrated economic value.

Any statements made regarding the Company's anticipated future product launches, regulatory approvals, revenues, earnings, market shares, and potential clinical success are forward-looking statements which are subject to risks and uncertainties, such as those described in the Financial Section of the Company's Annual Report to Shareholders for the fiscal year ended December 31, 2004 (see page 16). Actual results may differ materially from anticipated results.
COPYRIGHT 2005 Business Wire
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2005, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Publication:Business Wire
Geographic Code:1USA
Date:Oct 7, 2005
Words:732
Previous Article:Low Health Literacy Costs Europe Millions.
Next Article:China's Fruit and Vegetable Market Remains One of the Last to be Affected by the Development of New Retail Formats in the Mainland Market.
Topics:



Related Articles
Air-crew radiation doses climbing.
Shrinking the incredible universal magnet.
New radiation belt spotted around Earth.
Moon may radio cosmic rays' biggest hits.(Brief Article)
TISSUE VALVE GETS EUROPEAN AND CANADIAN APPROVALS.
Environmental Health in Space.
Big Bang: the Origin of the Universe.(Books: a selection of new and notable books of scientific interest)(Brief Article)(Book Review)
Fujitsu Laboratories Develops Technologies that Predict Software Error Rates on Logic Circuits.
Medical device makers: ranked by number of employees in the greater San Fernando Valley.(The List)
Medical device makers: ranked by number of employees in the greater San Fernando Valley.(HEALTH CARE)(Reprint)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles