Russian electrical stimulation: the early experiments. (Perspective).


Introduction

Electrical stimulation is used extensively in physical therapy, and "Russian currents" have been advocated for use in increasing muscle force. (1,2) This form of electrical stimulation seems to us to be the least understood in terms of physiological effects. Russian currents are alternating currents (AC) at a frequency of 2.5 kHz that are burst modulated mod·u·late  
v. mod·u·lat·ed, mod·u·lat·ing, mod·u·lates

v.tr.
1. To adjust or adapt to a certain proportion; regulate or temper.

2.
 at a frequency of 50 Hz with a 50% duty cycle. The stimulus is applied for a 10-second "on" period followed by a 50-second "off" or rest period, with a recommended treatment time of 10 minutes per stimulation session. (1) This stimulation regimen regimen /reg·i·men/ (rej´i-men) a strictly regulated scheme of diet, exercise, or other activity designed to achieve certain ends.

reg·i·men
n.
1.
 (called the "10/50/10" regimen), applied once daily over a period of weeks, has been claimed to result in force gains, but many of the claims appear to be anecdotal anecdotal /an·ec·do·tal/ (an?ek-do´t'l) based on case histories rather than on controlled clinical trials.
anecdotal adjective Unsubstantiated; occurring as single or isolated event.
. (3)

Selkowitz (1) has reviewed the experimental evidence in the English-language literature for increasing muscle force by use of Russian electrical stimulation. He concluded that there is convincing evidence for increased muscle force, but little evidence that the force gains were greater than those produced by voluntary exercise or a combination of exercise and electrical stimulation. He also noted that the studies he reviewed may not have had sufficient statistical power to distinguish among the conditions that were compared. Selkowitz also contended that there is insufficient evidence insufficient evidence n. a finding (decision) by a trial judge or an appeals court that the prosecution in a criminal case or a plaintiff in a lawsuit has not proved the case because the attorney did not present enough convincing evidence.  to distinguish force enhancements produced using Russian electrical stimulation ("kilohertz-frequency" AC) from those produced by other forms of electrical stimulation (eg, low-frequency monophasic pulsed current [PC]).

Only a few studies (4-10) of a relevant nature have been published since the review by Selkowitz. (1) Delitto et al (4) reported a single-subject experiment using an elite weight lifter weight·lift·er or weight lift·er  
n.
One who lifts heavy weights for exercise or in an athletic competition.

weight lifter nlevantador(a) m/f de pesas 
 undergoing ongoing weight training who was given periods of Russian electrical stimulation during the course of training. Marked improvements in performance, over and above those measured as a result of the training, accompanied the periods of stimulation. Delitto et al (5) compared force gains produced by Russian electrical stimulation with gains produced using voluntary exercise following anterior cruciate ligament anterior cruciate ligament
n. Abbr. ACL
The cruciate ligament of the knee that crosses from the anterior intercondylar area of the tibia to the posterior part of the lateral condyle of the femur.
 surgery. The electrically stimulated group showed higher force gains than the group that received voluntary exercise. Subsequent studies (6,7) of force recovery following anterior cruciate ligament surgery confirmed the earlier findings and established a correlation between training intensity and amount of force recovery. One of the studies (6) also demonstrated that clinical (Russian) stimulators were more effective than portable, battery-powered (monophasic PC) units. Unfortunately, the researchers could not establish whether the difference was due to the current type or to the inability of the battery-powered unit to supply the needed current intensity for all subjects. Snyder-Mackler et al (8) compared the maximum electrically induced torque (EIT EIT erythrocyte iron turnover. ) of 3 stimulators: a Russian current stimulator, an interferential stimulator operating at a frequency of 4 kHz, and a low-frequency biphasic bi·pha·sic  
adj.
Having two distinct phases: a biphasic waveform; a biphasic response to a stimulus. 
 PC stimulator. The interferential stimulator produced less torque than the other 2 machines, but this may have been because its maximum current output was not high enough for all subjects. The highest average torque was produced by the Russian stimulator, but the difference between it and the low-frequency stimulator was not significant. Laufer et al (9) compared maximum EITs obtained using 50-Hz modulated 2.5-kHz AC, 50-Hz monophasic PC, and 50-Hz biphasic PC. The only difference found was between the biphasic PC and the 2.5-kHz AC, with the biphasic PC producing the higher torque. Ward and Robertson (10) used 50-Hz modulated currents and measured maximum EIT at different kilohertz One thousand cycles per second. See Hertz.  frequencies in the range of 1 to 15 kHz. Maximum EITs were produced with a 1-kHz current. There were no comparisons with low-frequency monophasic PC.

Our purpose in this article is not to re-evaluate the evidence of trials that have examined force gains using Russian electrical stimulation. The review by Selkowitz (1) remains relevant, and the later studies, while adding to our knowledge, do not contradict con·tra·dict  
v. con·tra·dict·ed, con·tra·dict·ing, con·tra·dicts

v.tr.
1. To assert or express the opposite of (a statement).

2. To deny the statement of. See Synonyms at deny.
 his conclusions. Our aim is to present and examine the pioneering work that was published in Russian (11,12) and that we believe laid the foundation for the clinical use of Russian electrical stimulation. The combination of the English-language studies and the earlier Russian work provides what we believe is compelling evidence for "Russian stimulation." Questions remain, however, as to whether, and to what extent, "Russian currents" may be more effective than low-frequency PC for increasing a muscle's force-generating capability.

We believe some of the popularity of Russian electrical stimulation stemmed stemmed  
adj.
1. Having the stems removed.

2. Provided with a stem or a specific type of stem. Often used in combination: stemmed goblets; long-stemmed roses.
 from a talk given by Russian scientist Dr Yakov Kots (13) at a conference in 1977. Kots is reported to have advocated a stimulus regimen for increasing muscle force that he claimed was able to increase the maximum voluntary contraction contraction, in physics
contraction, in physics: see expansion.
contraction, in grammar
contraction, in writing: see abbreviation.

contraction - reduction
 (MVC (Model View Controller) An architecture for building applications that separate the data (model) from the user interface (view) and the processing (controller). ) of elite athletes elite athlete Sports medicine An athlete with potential for competing in the Olympics or as a professional athlete; EAs are at ↑ risk for injuries, given the amount of training, for psychological abuse by coaches and parents, and self abuse.  by up to 40%. Unfortunately, the only details of Kots' work were brief conference notes, translated from Russian and not readily accessible. (13) Seikowitz (1) noted that this is secondhand and undocumented information. Other authors (in the studies reviewed by Selkowitz (1)) have quoted the same secondary source.

Dr Kots later participated in a Canadian study on the effects of Russian electrical stimulation. College students who were athletes were the subjects. (14) The results of the study were published in English. Kots was, as best we can determine, advised by his accompanying translator that he could not provide copies of his prior Russian-language published work, nor references, to his western counterparts (Taylor AW, personal communication). The article about the Canadian study, (14) in which Kots was a coauthor co·au·thor or co-au·thor  
n.
A collaborating or joint author.

tr.v. co·au·thored, co·au·thor·ing, co·au·thors
To be a collaborating or joint author of: "He and a colleague . . .
, contains no references to his previously published Russian work. We find this puzzling and difficult to explain. The British Library British Library, national library of Great Britain, located in London. Long a part of the British Museum, the library collection originated in 1753 when the government purchased the Harleian Library, the library of Sir Robert Bruce Cotton, and groups of manuscripts.  had at the time of the Canadian study, and still has, subscriptions to the Russian-language journals in which Kots published. The details of Kots' research were readily available, albeit printed in the Russian language Russian language, also called Great Russian, member of the East Slavic group of the Slavic subfamily of the Indo-European family of languages (see Slavic languages).  and located in the United Kingdom. Nonetheless, a cloak of secrecy secrecy

see confidentiality.
 seems to have been invoked.

In this article, we describe, in some detail, the contents of 2 key Russian-language publications (11,12) that provide the original research on which "Russian currents" are based. They were obtained from the British Library and translated by one of the authors (NS).

The "10/50/10" Treatment Regimen

Russian electrical stimulation is applied for a 10-second "on" period followed by a 50-second "off" period, with a recommended treatment time of 10 minutes per stimulation session. The objective is to increase a muscle's ability to generate force, but what is often ignored is Kots' recommendation that this form of electrical stimulation should be used as an adjunct adjunct (aj´ungkt),
n a drug or other substance that serves a supplemental purpose in therapy.

adjunct 
 to exercise, (11) rather than as an alternative to exercise, and with electrical stimulation sessions separate from bouts Bouts is the name of
  • Aelbrecht Bouts (c. 1452-1549), An early Netherlandish painter
  • Dirk Bouts, Netherlandish painter
 of voluntary exercise.

Kots' argument for the use of electrical stimulation combined with voluntary exercise was that the commonly used exercise programs (those used at the time) build muscle bulk and muscle force but ignore the role of skill and fine motor control in athletic performance. (11) Electrical stimulation, however, preferentially pref·er·en·tial  
adj.
1. Of, relating to, or giving advantage or preference: preferential treatment.

2.
 recruits the fast-twitch, fast-fatiguable motor units associated with sudden, rapid movement, precise motor control, and gracefulness grace·ful  
adj.
Showing grace of movement, form, or proportion: "Capoeira is a graceful ballet of power and control, artists kicking and jumping in synchronized movement" Alisa Valdes.
 of movement. Thus, Kots argued, by a combination of exercise and electrical stimulation, an optimal force-enhancing regimen can be effected--one that maintains athletic skills and coordination in line with increases in muscle force. Although Kots' claim of preferential pref·er·en·tial  
adj.
1. Of, relating to, or giving advantage or preference: preferential treatment.

2.
 recruitment by electrical stimulation is well documented, (15) as is the involvement of fast-twitch fibers in rapid or correctional movement, (16) the claims regarding gracefulness, athletic skill, and coordination are more open to question.

Kots and Xvilon (11) reported a 2-part study, not using 2.5-kHz AC, but rather using short-duration (1-millisecond) rectangular rec·tan·gu·lar  
adj.
1. Having the shape of a rectangle.

2. Having one or more right angles.

3. Designating a geometric coordinate system with mutually perpendicular axes.
 PC at a frequency of 50 Hz. In the first part of their study, they determined optimum "on" and "off" times for stimulation. Their findings provide the rationale for the "10/50/10" treatment regimen that is characteristic of treatment with Russian electrical stimulation. In the second part of their study, they examined the force-enhancing effect of a single 10-minute training session done daily or every second day for a period of 9 or 19 days.

For the study by Kots and Xvilon, (11) 37 young athletes (age range=15-17 years, no mean or standard deviation In statistics, the average amount a number varies from the average number in a series of numbers.

(statistics) standard deviation - (SD) A measure of the range of values in a set of numbers.
 given) were recruited and divided into 4 groups. Three groups received electrical stimulation of the biceps brachii muscle
For other uses, see biceps.


In human anatomy, the biceps brachii is a muscle located on the upper arm. The biceps has several functions, the most important simply being to flex the elbow and to rotate the forearm.
, and the fourth group received electrical stimulation of the triceps surae The triceps surae is a term given by some anatomists to the gastrocnemius and soleus muscles together as they both insert into the calcaneus, the bone of the heel of the human foot, and form the major part of the muscle of the back part of the lower leg (the calf; otherwise known  muscle. Current was applied using 4- x 4-cm metal electrodes Electrodes
Tiny wires in adhesive pads that are applied to the body for ECG measurement.

Mentioned in: Electrocardiography
 over the muscle belly, with a saline-soaked pad between the electrodes and the skin. Stimulation was applied while the arm or leg was secured in an apparatus built for measuring isometric isometric /iso·met·ric/ (-met´rik) maintaining, or pertaining to, the same measure of length; of equal dimensions.

i·so·met·ric
adj.
1.
 torque (Fig. 1). The apparatus was used to measure maximum EIT and MVCs. Muscle hardness also was measured for the groups that received electrical stimulation of the biceps brachii muscle, both during MVCs and during electrical stimulation. The device for measuring muscle hardness was not described in any detail. It was a skin-mounted device (Fig. 1b) that, we surmise, applied a controlled force to the skin surface and gave a "hardness" reading determined by the amount of indentation in·den·ta·tion
n.
A notch, a pit, or a depression.
 produced. Hardness, measured in this way, would give an indirect indication of muscle force but, we believe, would give readings that are unduly biased in favor of the part of the muscle closest to the measuring device.

[FIGURE 1 OMITTED]

For the first part of the study by Kots and Xvilon, (11) trains of 50-Hz pulses were applied at maximum tolerable tol·er·a·ble  
adj.
1. Capable of being tolerated; endurable.

2. Fairly good; passable. See Synonyms at average.



tol
 intensity for 15 seconds, and the evoked e·voke  
tr.v. e·voked, e·vok·ing, e·vokes
1. To summon or call forth: actions that evoked our mistrust.

2.
 muscle torque and stimulus intensity were monitored. Kots and Xvilon found no appreciable ap·pre·cia·ble  
adj.
Possible to estimate, measure, or perceive: appreciable changes in temperature. See Synonyms at perceptible.
 decrease in torque with trains of up to 10 seconds' duration. Electrically induced fatigue, defined as a visible decline in the torque record, was noted (Fig. 2a) at a mean of 12.5 seconds (SD=1.8), after which it progressed rapidly. Fatigue was not quantified but simply rated as present or absent. On the basis of their observations, Kots and Xvilon concluded that a maximum "on" time of 10 seconds was desirable to avoid fatigue during the pulse train.

[FIGURE 2 OMITTED]

Having settled on a 10-second "on" time, Kots and Xvilon (11) then established what "off" time was required to avoid fatigue between pulse trains. Fatigue, in this case, was defined as a visible decrease in torque between 2 consecutive 10-second stimulus trains. They compared "off" times of 10, 20, 30, 40, and 50 seconds and reported that with "off" times of 30 seconds or less (Fig. 2b), the average torque during the second train was less than the torque during the first train and that fatigue increased (torque declined) during the second 10-second train. They concluded that the "off" time needed to be 40 to 50 seconds. They then measured the torque variation over 10 consecutive 10-second trains and found that with a 40-second "off" time, signs of fatigue were evident, particularly in the last few trains. With a 50-second "off" period, no fatigue was evident over the 10 consecutive trains (Fig. 2c). Accordingly, they chose a nonfatiguing "10/50/10" (10 seconds "on" and 50 seconds "off" for 10 trains) protocol for the second part of their study.

Increasing Muscle Force Using the "10/50/10" Treatment Regimen

In the second part of their study, Kots and Xvilon (11) used a single "10/ 50/10" treatment applied once daily or on every second day, and they monitored changes in muscle torque and muscle hardness over 9 or 19 days. Before each stimulation session, muscle torque and muscle hardness were measured during each of 3 MVCs. Limb circumference was measured during each MVC and after every MVC with the subject relaxed. Electrically induced torque and applied current also were monitored during treatment. Table 1 provides details of the 4 series of tests.

Kots and Xvilon (11) noted that although their EIT values were only a fraction of MVC, muscle hardness, as measured by their indentation device, was always greater than that of an MVC (Tab. 1). Their conclusion, based on their hardness measurements, was that electrical stimulation produces greater force in the excited muscle than when recruited voluntarily. The greater MVC values, they suggested, were due to (automatic voluntary) recruitment of synergistic muscles synergistic muscles
pl.n.
Muscles having similar and mutually helpful functions or actions.
, which were not recruited electrically. That is, MVC measurements reflect the net effect of all synergistic muscles contributing to a contraction. Hardness values reflect the contribution of just the muscle directly under the measuring device.

Kots and Xvilon (11) further observed that their subjects tolerated progressively higher stimulus intensities over the 9-or 19-day training period and that there was a corresponding progressive increase in EIT. The increases are shown in Figure 3. Increases in MVC and limb circumference also were found. The findings are summarized in Table 2 and depicted de·pict  
tr.v. de·pict·ed, de·pict·ing, de·picts
1. To represent in a picture or sculpture.

2. To represent in words; describe. See Synonyms at represent.
 graphically as part of Figure 4.

[FIGURES 3-4 OMITTED]

The authors (11) expressed surprise at the rapid and large increases in force production. They also noted that the magnitude of the force gain appeared to depend on the number of stimulation sessions (in Tab. 2, compare series 1 and 2 where 9 treatment sessions were used with series 3 where 19 treatments were applied). There seemed to be little difference whether the treatments were done every day (series 2 [9 sessions]) or every second day (series 1 [9 sessions]).

Figure 4 shows MVC plotted against duration in the treatment program (in days). The changes in limb circumference with the muscle relaxed and when producing an MVC also are plotted. Both circumference and MVC values are expressed as a percentage of the initial (baseline) values prior to electrical stimulation.

Kuts and Xvilon (11) argued that increasing a muscle's force-generating capability can be achieved by 2 means. One means is by central nervous system (CNS See Continuous net settlement.

CNS

See continuous net settlement (CNS).
) adaptation whereby a greater MVC is produced by CNS "learning" and adaptation of the pattern of excitation excitation

Addition of a discrete amount of energy to a system that changes it usually from a state of lowest energy (ground state) to one of higher energy (excited state). For example, in a hydrogen atom, an excitation energy of 10.
. In this case, the force gains are achieved by greater and more effective recruitment of muscle fibers. The second means is by building the physical bulk of the muscle to produce a greater force output for the same neural neural /neu·ral/ (noor´al)
1. pertaining to a nerve or to the nerves.

2. situated in the region of the spinal axis, as the neural arch.


neu·ral
adj.
1.
 input. In this case, the muscle fibers grow in size and muscle volume increases. The increases in limb circumference (and thus, by inference (logic) inference - The logical process by which new facts are derived from known facts by the application of inference rules.

See also symbolic inference, type inference.
, muscle bulk) paralleled the increase in muscle force, so the authors concluded that the force gains were predominantly pre·dom·i·nant  
adj.
1. Having greatest ascendancy, importance, influence, authority, or force. See Synonyms at dominant.

2.
 of peripheral origin.

To establish whether the MVC testing that was part of the experimental protocol contributed to the force gains, a control group was used. These subjects performed MVCs 6 times per day for 19 days to match the experimental group, who performed 3 MVCs before each stimulation session and 3 MVCs after each stimulation session. No increase in force was produced. Although this finding demonstrates that the force gains were not a result of performing repetitive MVCs, the control group does not control for a placebo placebo (pləsē`bō), inert substance given instead of a potent drug. Placebo medications are sometimes prescribed when a drug is not really needed or when one would not be appropriate because they make patients feel well taken care of.  response, because there is no way the controls could be unaware of the presence or absence of electrical stimulation. Given that few of the later studies by a variety of authors showed such large force gains with stimulation sessions so few and short, we question whether the extreme motivation for the young Russian athletes was a factor in the force gains. Possibly the age of the subjects had a bearing on the outcome. Other studies (reviewed by Selkowitz (1)) used subjects who were more physically mature and less motivated mo·ti·vate  
tr.v. mo·ti·vat·ed, mo·ti·vat·ing, mo·ti·vates
To provide with an incentive; move to action; impel.



mo
.

Medium-Frequency Alternating Current

Andrianova et al (12) reported on the use of kilohertz-frequency sinusoidal sinusoidal /si·nus·oi·dal/ (si?nu-soi´dal)
1. located in a sinusoid or affecting the circulation in the region of a sinusoid.

2. shaped like or pertaining to a sine wave.
 alternating current for increasing a muscle's force-generating capability. Both continuous (unmodulated) AC and AC bursts, modulated at 50 Hz (10 milliseconds "on" and 10 milliseconds "off"), were used. Andrianova and colleagues examined "direct" stimulation, where the electrodes were placed over the muscle, and "indirect" stimulation, where they attempted to stimulate the nerve trunk nerve trunk
n.
The main stem of a nerve, consisting of a bundle of nerve fibers bound together by a tough sheath of connective tissue.
 supplying the muscle. Their article (12) reports a 4-part study involving either wrist and finger flexors or the calf muscles The calf or gastrosoleus is a pair of muscles—the gastrocnemius and soleus—at the back of the lower human leg.

The gastrosoleus complex is connected to the foot through the Achilles tendon, and contract to induce plantar flexion and stabilization of the
, or both. For direct stimulation of wrist and finger flexors, electrodes measuring 6 x 3 cm and 4 x 3 cm were applied to the palmar surface of the forearm forearm /fore·arm/ (for´ahrm) antebrachium; the part of the arm between elbow and wrist.

fore·arm
n.
The part of the arm between the wrist and the elbow.
, with the long side across the forearm and the larger electrode electrode, terminal through which electric current passes between metallic and nonmetallic parts of an electric circuit. In most familiar circuits current is carried by metallic conductors, but in some circuits the current passes for some distance through a  more proximal proximal /prox·i·mal/ (-mil) nearest to a point of reference, as to a center or median line or to the point of attachment or origin.

prox·i·mal
adj.
. For indirect stimulation, a thin electrode (2.5 x 0.5 cm) was positioned along the fissure fissure /fis·sure/ (fish´er)
1. any cleft or groove, normal or otherwise, especially a deep fold in the cerebral cortex involving its entire thickness.

2. a fault in the enamel surface of a tooth.
 of the elbow joint elbow joint
n.
A compound hinge joint between the humerus and the bones of the forearm. Also called cubital joint.
 and a larger electrode (3 x 1.5 cm or 3.5 x 1 cm, respectively) was positioned on the palmar surface of the forearm or on the inner surface of the shoulder (long side across the inner surface). No further details of electrode placement were given. The authors stated that the same size electrodes were used for the calf muscles, but no details of electrode placement were given. It is uncertain, therefore, how electrodes were located to activate the nerve trunk supplying the calf muscles. The number of subjects in each part of the study ranged from 7 to 10.

In the first part of the study reported by Andrianova et al, (12) continuous (unmodulated) AC at frequencies of 100, 500, 1,000, 2,500, and either 3,000 or 5,000 Hz was used for stimulation of the wrist and finger flexors. Motor thresholds, maximum tolerable current, and the current required to achieve 60% of the maximum EIT were measured at each frequency. The results are shown in Figure 5.

[FIGURE 5 OMITTED]

Andrianova et al (12) reported that although current levels increased with increasing frequency, the discomfort associated with the stimulation decreased. They did not state whether or how discomfort was quantified, so we conclude that this was an empirical observation. For direct stimulation of the calf muscles, a maximum force of 92.5 kg (SD=25.0), approximately 70% of MVC, was elicited e·lic·it  
tr.v. e·lic·it·ed, e·lic·it·ing, e·lic·its
1.
a. To bring or draw out (something latent); educe.

b. To arrive at (a truth, for example) by logic.

2.
 at 2.5 kHz. For indirect stimulation (of wrist and finger flexors), the maximum force was elicited at 1 kHz. Above 1 kHz, rapid fatigue was noted. The authors concluded that a frequency of 1 kHz was optimal for force production using indirect stimulation and 2.5 kHz was optimal when using direct stimulation.

The second part of the study reported force measurements made using wrist and finger flexors with direct and indirect stimulation and indirect stimulation with 10-millisecond bursts at 50 Hz. Table 3 shows the maximum force produced. The results indicate that for indirect stimulation, whether continuous or modulated at 50 Hz, maximum force was produced at an AC frequency of 1 kHz. For direct stimulation using a continuous stimulus, maximum force was produced at an AC frequency of 2.5 kHz. Direct stimulation using 50-Hz bursts does not seem to have been examined.

Whether 1 kHz is the optimal frequency for indirect stimulation, whereas 2.5 kHz is the optimal frequency for direct stimulation, was investigated in the third part of the study, (12) which used wrist and finger flexors and a continuous AC stimulus. Frequencies of 2.5 kHz and 1 kHz only were compared (Tab. 4). These results were in agreement with the findings of the previous part of the study, although only stimulation with a continuous waveform The shape of a signal. See wavelength, sine wave and square wave.  was used in this part of the study. The authors apparently did not examine 50-Hz burst modulation modulation, in communications
modulation, in communications, process in which some characteristic of a wave (the carrier wave) is made to vary in accordance with an information-bearing signal wave (the modulating wave); demodulation is the process by which
.

Andrianva et al (12) noted that both indirect and direct stimulation produced similar levels of maximum force, although at different frequencies. A frequency of 1 kHz was optimal for force production using indirect stimulation and a continuous waveform, and a frequency of 2.5 kHz was optimal when using direct stimulation and a continuous waveform. The observation that levels of maximum force was similar led the authors to suggest that direct stimulation was capable of exciting not only the superficial superficial /su·per·fi·cial/ (-fish´al) pertaining to or situated near the surface.

su·per·fi·cial
adj.
1. Of, affecting, or being on or near the surface.

2.
 muscle fibers but presumably pre·sum·a·ble  
adj.
That can be presumed or taken for granted; reasonable as a supposition: presumable causes of the disaster.
 also the deep muscle fibers excited by indirect (nerve trunk) stimulation.

50-Hz Burst Modulation

Andrianova et al (12) concluded that whether current is applied in continuous mode or in 10-millisecond, 50-Hz bursts, the maximum force induced and the optimal frequency are not affected. This conclusion is consistent with the report of Soloviev, (17) who stated there was little difference in the variation in motor threshold with frequency, whether the current applied was continuous or burst modulated at 50 Hz. Accordingly, Andrianova et al recommended 50-Hz burst modulation because it would result in halving of the electrical energy delivered to the patient while producing little or no decrease in the maximum force induced. Soloviev's findings are supported by a recent study (18) in which motor thresholds in the range 1 to 25 kHz were examined. Little difference was found between a continuous stimulus and one modulated at 50 Hz.

To verify that 50-Hz burst modulation did not diminish maximum EIT, Andrianova et al (12) carried out the fourth part of their study, comparing continuous and burst-mode stimulation using direct stimulation of the calf muscles and indirect stimulation of the wrist and finger flexors. The findings are shown in Table 5. The results support the contention that 50-Hz, 50% duty cycle, burst modulation does not diminish maximum EIT. For this reason alone, they argued, burst modulation should be preferred for patient treatment because the physiological response is indistinguishable, while the current levels are halved halve  
tr.v. halved, halv·ing, halves
1. To divide (something) into two equal portions or parts.

2. To lessen or reduce by half: halved the recipe to serve two.

3.
. What does not seem to have been directly established is whether 2.5 kHz is still an optimal frequency for force production when 50-Hz bursts, rather than continuous AC, are used.

Increasing Muscle Force Using 50-Hz Burst Modulation

Andrianova et al (12) reported force gains in 2 different groups of 16 young wrestlers See
  • list of amateur wrestlers
  • list of professional wrestlers
  • list of independent circuit, non-affiliated or retired professional wrestlers
and
. The first group had their calf muscles stimulated directly using a frequency of 2.5 kHz. Stimulation was once per day for 18 days. Maximum voluntary contraction, limb circumference, and jumping height were measured daily. Half of the second group had their tibialis anterior muscle In human anatomy, the tibialis anterior is a muscle in the shin that spans the length of the tibia. It originates in the upper two-thirds of the lateral surface of the tibia and inserts into the medial cuneiform and first metatarsal bones of the foot.  stimulated directly at 2.5 kHz, and the other half of the group had their tibialis anterior muscle stimulated indirectly at 1 kHz. For both groups, the stimulation regimen was the same as that described earlier (10 seconds "on," 50 seconds "off," and 10 stimulation cycles) but with the current burst modulated at 50 Hz with a 50% duty cycle. Current was applied at a maximum tolerable level. The results are shown in Figure 6.

[FIGURE 6 OMITTED]

Force gains achieved were largest for the group that underwent calf muscle stimulation, where the increase in MVC over the 18-day training period was 45%. These force gains were accompanied by an increase in limb circumference of 3% and by an increase in jumping height of almost 15%. The group that underwent stimulation of the tibialis anterior muscle had an increase in dorsiflexor MVC of 30% or more. Indirect stimulation at 1 kHz appeared to result in a more rapid force gains than direct stimulation at 2.5 kHz (days 1-10), but by the end of the training period the difference was small.

Discussion

Increasing Muscle Force

The force gains reported by Kots and Xvilon (11) (27%-56%) and Andrianova et al (12) (30%-45%) are at the high end of gains reported in the English-language literature (7%-48%). (1) This is perhaps not surprising given the likelihood of a placebo response. Kots and co-workers had subjects who were young (15-17 years of age, no mean and standard deviation stated) and had not reached maturity and who were also in training as potential Olympic athletes. Other researchers (1) used more physically mature participants who also might have had less personal incentive to achieve force gains. Thus, the placebo effect placebo effect
n.
A beneficial effect in a patient following a particular treatment that arises from the patient's expectations concerning the treatment rather than from the treatment itself.
 in the studies of Kots and co-workers would be expected to be large. The extent of the placebo response is uncertain, but there is little doubt that the placebo effect can increase force measurements. It is interesting to note that in a later study, (4) in which Russian electrical stimulation was used and the subject was an elite weight lifter, the authors reported performance gains comparable to those reported by Andrianova et al. (12)

Force gains have been shown with electrical stimulation, just as they have with voluntary exercise, and there is some evidence that a combination of voluntary exercise and electrical stimulation (applied on separate occasions) can produce greater force gains than either intervention used alone. (1) A problem with the studies in which electrical stimulation was compared with voluntary exercise or a combination of both interventions is that there may not have been enough subjects to have sufficient statistical power. Although the numbers of subjects (typically between 10 and 20 per group) may have been enough to distinguish a large effect between treatment and control, the numbers appear to be too small to distinguish lesser effects that might have existed between the different treatment groups.

Nonetheless, the balance of evidence, in our opinion, suggests that a combination of exercise and electrical stimulation is more effective than either intervention used alone. There are 2 possible explanations. The first explanation is one of experimental design. With the combination applied sequentially (voluntary exercise and separate electrical stimulation), the total amount of exercise is greater. The second explanation is that exercise and electrical stimulation preferentially recruit different fiber types. Kots and Xvilon (11) argued that traditional, voluntary exercise regimens promoted increased force production in slow-twitch, fatigue-resistant muscle fibers because they are the ones first recruited in a voluntary contraction and there is limited recruitment of fast-twitch fibers in all but the fastest and most forceful force·ful  
adj.
Characterized by or full of force; effective: was persuaded by the forceful speaker to register to vote; enacted forceful measures to reduce drug abuse.
 voluntary contractions contractions Obstetrics Volleys of tightening and shortening of myometrium–uterine muscle, which occur during labor, cause dilatation and thinning of the cervix and aid in the descent of the infant in the birth canal. See Labor. Cf Decelerations. . An electrical stimulation regimen, by contrast, preferentially recruits the fast-twitch muscle fibers, which are innervated innervated adjective Containing or characterized by nerves  by larger-diameter motoneurons. On this basis, they contended, an optimal force gain program should include both exercise and electrical stimulation to increase force production of both fiber types.

Kots and Xvilon (11) also argued that, because of differential recruitment, muscle force-generating regimens consisting of voluntary exercise alone run the risk of an increase in muscle force production at the expense of reducing the speed of muscle contraction Noun 1. muscle contraction - (physiology) a shortening or tensing of a part or organ (especially of a muscle or muscle fiber)
contraction, muscular contraction

shortening - act of decreasing in length; "the dress needs shortening"
. They argued that fast-twitch fiber force gains should accompany voluntary contraction force gains of slow-twitch fibers in order to maintain the balance, which they believed is needed for performance of skillful skill·ful  
adj.
1. Possessing or exercising skill; expert. See Synonyms at proficient.

2. Characterized by, exhibiting, or requiring skill.
, well-executed movements.

The "10/50/10" Stimulation Regimen

Kots and Xvilon (11) contended that to increase force production, electrical stimulation should be nonfatiguing, meaning, that there should be no decrease in force during the stimulus period. Their observations of force decline using low-frequency (50-Hz) monophasic PC with different "on" and "off" times during a 10-minute treatment period were their evidence that the "10/50/ 10" stimulation regimen is "nonfatiguing," provided that the stimulus is monophasic PC. Their argument for a nonfatiguing response was that further stimulation of an electrically fatigued muscle will not increase the muscle's force production capability. The argument has credibility. At a stimulus frequency of 50 Hz, the dominant fatigue mechanisms are neurotransmitter neurotransmitter, chemical that transmits information across the junction (synapse) that separates one nerve cell (neuron) from another nerve cell or a muscle. Neurotransmitters are stored in the nerve cell's bulbous end (axon).  depletion depletion n. when a natural resource (particularly oil) is being used up. The annual amount of depletion may, ironically, provide a tax deduction for the company exploiting the resource because if the resource they are exploiting runs out, they will no longer be able  and propagation The transmission (spreading) of signals from one place to another.  failure at the level of the t-tubule system, (19) processes that would not result in increased force production. (19,20) Fatigue induced by voluntary exercise involves much lower nerve fiber nerve fiber
n.
A threadlike process of a neuron, especially the axon that conducts nerve impulses.
 firing frequencies (20) and places greater stresses on the contractile contractile /con·trac·tile/ (kon-trak´til) able to contract in response to a suitable stimulus.

con·trac·tile
adj.
Capable of contracting or causing contraction, as a tissue.
 components of the muscle fibers. Such stresses are argued to be needed for strengthening. (19) Thus, we believe that the choice of a "10/50/10" stimulation regimen to avoid neuromuscular neuromuscular /neu·ro·mus·cu·lar/ (-mus´ku-ler) pertaining to nerves and muscles, or to the relationship between them.

neu·ro·mus·cu·lar
adj.
1.
 fatigue has a sound physiological basis.

The "10/50/10" protocol was established using short-duration monophasic PC at a frequency of 50 Hz. (11) Because a "10/50/10" regimen is optimal when using short-duration PC does not mean that the same would necessarily apply when using kilohertz-frequency bursts of AC modulated at 50 Hz. Andrianova et al (12) used 50-Hz bursts of kilohertz-frequency AC and the "10/50/10" protocol, and this has led to the assumption that this protocol is optimal when using kilohertz-frequency AC. Fatigue effects were not measured by Andrianova et al, (12) and their rationale for using the protocol was simply a reference to the study by Kots and Xvilon. (11) The focus was on optimal frequencies for maximum force production. Andrianova et al (12) reported that at higher frequencies, there was a rapid drop-off in force, which limited the maximum EIT, that is, that fatigue effects appeared to have an effect at higher frequencies, but this was apparently only a qualitative observation. Their observation echoes that of Djourno, (21) who in 1952 reported the occurrence of increasing rates of fatigue with increasing frequency When using kilohertz-frequency AC and continuous stimulation. Nonetheless, fatigue seems to have been all but ignored by Andrianova et al, (12) who chose a "10/50/10" protocol for both direct and indirect stimulation on the basis of results obtained by Kots and Xvilon (11) using low-frequency monophasic PC.

Some years after the study by Andrianova et al, (12) Stefanovska and Vodovnik (22) compared 50-Hz single-pulse stimulation and 50-Hz burst stimulation at 2.5 kHz using 10-second trains of stimulation. They reported that when using 50-Hz single pulses, what they called "negligible This article or section is written like a personal reflection or and may require .
Please [ improve this article] by rewriting this article or section in an .
 fatigue," defined as no visible decrease in EIT, occurred over a 10-second stimulation period, even during repetitive stimulation. By contrast, the force measured using 2.5-kHz AC showed appreciable decline during the 10 seconds of stimulation. Whether a "10/50/10" protocol is optimal when using 50-Hz bursts of kilohertz-frequency AC, therefore, is questionable.

Optimal Frequencies

Andrianova et al (12) compared continuous stimulation with 50-Hz burst stimulation in the frequency range of 100 Hz to 5 kHz but only using what they considered indirect (presumably via the nerve trunk) stimulation. Their conclusion was that burst modulation did not affect the optimal frequency for muscle force production. Both continuous and burst-modulated waveforms produced maximum force at 1 kHz (Tab. 3). Unfortunately, no comparison of continuous and burst-modulated waveforms using direct (over the muscle) stimulation was reported. Their conclusion was that burst modulation makes no difference to the optimal frequency and should be preferred for patient treatment because the physiological response is indistinguishable while the current levels are halved. Although this was demonstrated for indirect stimulation, whether 2.5 kHz is still optimal for direct stimulation when 50-Hz burst modulation is used was not demonstrated.

Only one subsequent study of the frequency dependence of force production using kilohertz-frequency AC has been reported. (10) Ward and Robertson (10) examined frequencies in the range of 1 to 15 kHz, burst modulated at 50 Hz, and found that maximum wrist extensor extensor /ex·ten·sor/ (-ser) [L.]
1. causing extension.

2. a muscle that extends a joint.


ex·ten·sor
n.
A muscle that extends or straightens a limb or body part.
 torque was elicited at 1 kHz. Lower frequencies were not examined. The proximal electrode was positioned over the nerve trunk, and the distal distal /dis·tal/ (-t'l) remote; farther from any point of reference.

dis·tal
adj.
1. Anatomically located far from a point of reference, such as an origin or a point of attachment.
 electrode was positioned over the muscle belly, so the stimulation could not be unequivocally identified as "direct" or "indirect." The finding of maximum torque production at 1 kHz suggests that indirect stimulation under the proximal electrode contributed most to torque production.

Data suggest to us and others that an AC frequency of 2.5 kHz is optimal for direct stimulation when 50-Hz burst modulation is used, but this is inference rather than observation. We believe that it would be desirable to test the hypothesis experimentally. The evidence for 1 kHz as an optimum frequency for indirect stimulation, in our view, is more compelling (Tab. 3).

Kilohertz-Frequency AC Bursts or Low-Frequency Monophasic PC?

Andrianova et al (12) stated that burst-modulated, kilohertz-frequency AC is preferable to low-frequency PC because the stimulation is more comfortable. They concluded, on the basis of their research, that the optimum frequencies for AC stimulation are 1 kHz for indirect stimulation and 2.5 kHz for direct stimulation. Their conclusions have an interesting historic basis. The ability to evoke e·voke  
tr.v. e·voked, e·vok·ing, e·vokes
1. To summon or call forth: actions that evoked our mistrust.

2.
 a strong, comfortable contraction with kilohertz-frequency AC was first noted by d'Arsonval, (23) who reported, in 1891, that with continuous AC at a fixed voltage, neuromuscular excitation became stronger up to 1,250 to 1,500 Hz, remained constant to 2,500 Hz, and decreased between 2,500 and 5,000 Hz. d'Arsonval also noted that physical sensation and discomfort decreased steadily with increasing frequency up to the maximum frequency that his stimulator could produce (5,000 Hz). The idea that kilohertz-frequency AC is able to produce strong, comfortable muscle contractions at an optimal frequency between 1.5 and 2.5 kHz had been advanced by d'Arsonval about 80 years earlier than Andrianova et al. (12)

Unfortunately, d'Arsonval (23) did not report details of the electrode placement for his experiments. His interpretation of his studies indicated to him that maximum force with least discomfort is elicited between 1.5 and 2.5 kHz. In the early days of electrical stimulation of human subjects, it was common practice to use 2 cylindrical cyl·in·dri·cal
adj.
Of, relating to, or having the shape of a cylinder, especially of a circular cylinder.
, metal, hand-held electrodes. (24) Stimulation with this technique, in our opinion, might be more like "direct" stimulation than "indirect" stimulation because the relatively bulky bulk·y  
adj. bulk·i·er, bulk·i·est
1. Having considerable bulk; massive.

2. Of large size for its weight: a bulky knit.

3. Clumsy to manage; unwieldy.
 muscles would be positioned closer to the electrodes and would be more susceptible to direct excitation, rather than via the more distantly located, small-volume nerve trunk.

The studies reported by Ward and Robertson (10,25) shed some light on to the question of comfort of stimulation and its relation to maximum torque production. These authors (25) measured sensory sensory /sen·so·ry/ (sen´sor-e) pertaining to sensation.

sen·so·ry
adj.
1. Of or relating to the senses or sensation.

2.
, motor, and pain thresholds Noun 1. pain threshold - the lowest intensity of stimulation at which pain is experienced; "some people have much higher pain thresholds than do other people"
absolute threshold - the lowest level of stimulation that a person can detect
 at different frequencies in the range 1 to 35 kHz using a 50-Hz burst-modulated stimulus. They found that the separation between motor and pain thresholds increased between 1 and 10 kHz and then decreased at higher frequencies. To the extent that separation between motor and pain thresholds is a predictor of comfort, we surmise that more comfortable contractions are produced as the frequency increases, up to an optimum frequency of 10 kHz. In a subsequent study, (10) Ward and Robertson found that maximum torque was elicited not at 10 kHz, but at 1 kHz (the lowest frequency examined). These findings call into question the relationship between comfort of stimulation (at low torque levels) and maximum EIT.

An assumption of Andrianova et al (12) was that if the stimulus is more comfortable, greater maximum force can be elicited. On this basis, they stated a preference for kilohertz-frequency AC rather than low-frequency PC. At face value, this seems to be a reasonable assumption. However, as we have contended, when comparing different frequencies, greatest comfort and maximum EIT are not at the same frequency. Thus, it does not necessarily follow that if kilohertz-frequency AC produces more comfortable contractions than low-frequency PC, greater maximal max·i·mal
adj.
1. Of, relating to, or consisting of a maximum.

2. Being the greatest or highest possible.
 contractions will be produced.

The limited number of studies that have directly compared low-frequency PC and 2.5-kHz AC (8,9,26) are inconclusive INCONCLUSIVE. What does not put an end to a thing. Inconclusive presumptions are those which may be overcome by opposing proof; for example, the law presumes that he who possesses personal property is the owner of it, but evidence is allowed to contradict this presumption, and show who is . A recent study by Laufer et al (9) demonstrated greater EITs with low-frequency PC than 2.5-kHz AC. Walmsley et al (26) reported no difference (calling into question the statistical power of their study). Snyder-Mackler et al (8) also reported no difference, again calling into question whether the study had sufficient statistical power. Each of these groups of investigators used a stimulus that was ramped or increased manually by the experimenters, and this may have resulted in muscle fibers ceasing to contract due to neurotransmitter depletion, with a consequent con·se·quent  
adj.
1.
a. Following as a natural effect, result, or conclusion: tried to prevent an oil spill and the consequent damage to wildlife.

b.
 underestimation of the peak torque that can be elicited using 2.5-kHz AC. (18,27)

Conclusion

What are called "Russian currents" are widely used in physical therapy, but the support for their use in the English-language literature is scant scant  
adj. scant·er, scant·est
1. Barely sufficient: paid scant attention to the lecture.

2. Falling short of a specific measure: a scant cup of sugar.
. The studies reported in the Russian literature Russian literature, literary works mainly produced in the historic area of Russia, written in its earliest days in Church Slavonic and after the 17th cent. in the Russian language.  by Kots and Xvilon (11) and Andrianova et al (12) provide some experimental data to support their use. Andrianova et al (12) concluded that 1 kHz, rather than 2.5 kHz, is preferable for maximum force production when muscles are stimulated indirectly (over the nerve trunk), and this conclusion is supported by a more recent study. (10) This finding suggests that "Russian current" stimulators should provide a choice of 1-kHz or 2.5-kHz stimulus waveforms. As we have noted, however, the early studies (11,12) have not appeared in the English-language literature. In addition, we have no idea as to the extent to which they may have undergone peer review before publication.

The question of whether the burst-modulated AC used in "Russian current" stimulators is more effective for force production than low-frequency PC remains open. The data (8,9,26) are inconclusive. Other questions also remain. The "10/50/10" protocol that is fundamental to Russian electrical stimulation was based on measurements made using a low-frequency monophasic PC stimulus and not kilohertz-frequency AC bursts. The "10/ 50/10" protocol was chosen because it produced no measurable force reduction during the 10-minute stimulation period. Yet 10 seconds of 50-Hz burst-modulated, kilohertz-frequency stimulation has been shown to produce a marked reduction in force. (22) There is a question as to whether the "10/50/10" regimen is still optimal when kilohertz-frequency AC is used. The force gains measured by Andrianova et al (12) using kilohertz-frequency AC, when compared with those of Kots and Xvilon (11) using low-frequency PC, in our opinion, lend support to the choice of a burst-modulated AC stimulus regimen, but the evidence is not conclusive Determinative; beyond dispute or question. That which is conclusive is manifest, clear, or obvious. It is a legal inference made so peremptorily that it cannot be overthrown or contradicted. . Direct comparisons of muscle force-generating regimens that use different "on/off" times and treatment schedules (duration and number of times per day per week) are needed, as are further direct comparisons of force production using low-frequency PC and modulated kilohertz-frequency AC.
Table 1.
Details of the Four Series of Trials Conducted by Kots and
Xvilon (11) Using the "10/50/10" Treatment Regimen (a)

Variable                     Series 1                Series 2

No. of subjects              11                      10

Age (y)                      15-16                   15-16

Muscle                       Biceps brachii          Biceps brachii

Stimulation                  Once every second day   Daily

No. of treatment sessions     9                       9

EIT (% Of MVC)
  [bar]X                     53.9                    46.5
  SD                          2.7                     0.7
  Range                      38.5-60.1               42.6-49.3

Muscle hardness (% of MVC)
  [bar]X                     106.4                   108.0
  SD                           0.4                     0.3
  Range                      104.0-110.0             105.0-111.0

Variable                     Series 3         Series 4

No. of subjects               8                8

Age (y)                      16-17            16-17

Muscle                       Biceps brachii   Triceps surae

Stimulation                  Daily            Daily

No. of treatment sessions    19               19

EIT (% Of MVC)
  [bar]X                     43.8             36.5
  SD                          1.1              1.9
  Range                      27.2-57.7        27.1-41.3

Muscle hardness (% of MVC)
  [bar]X                     108.0
  SD                           0.5
  Range                      106.0-109.0

(a) EIT=electrically induced torque, MVC=maximum voluntary
contraction. No mean and standard deviation values were
stated for age.

Table 2.
Overall Changes (a) in Maximum Voluntary Contraction (MVC) and
Limb Circumference for the Four Series of Tests Reported by
Kots and Xvilon (11)

                                    Series 1

Variable                            [bar]X   SD    Range

MVC force (kg)
  Before series                     25.6     3.5   21.6-32.0
  After series                      32.5     0.7   27.6-36.3
  % change                          27.0     3.9   12.0-52.9

Limb circumference, relaxed (cm)
  Before series                     26.4     0.5   24.5-29.0
  After series                      27.1     0.4   26.0-31.2
  Change                             0.7            0.5-1.6

Limb circumference, with MVC (cm)
  Before series                     29.1     0.5   27.0-32.0
  After series                      30.0     0.5   28.5-33.9
  Change                             0.9            0.6-2.0

                                    Series 2

Variable                            [bar]X   SD    Range

MVC force (kg)
  Before series                     25.3     1.4   17.7-33.0
  After series                      32.8     1.5   23.9-39.8
  % change                          29.8     2.4   19.3-40.6

Limb circumference, relaxed (cm)
  Before series                     25.5     0.7   21.5-28.8
  After series                      26.4     0.7   22.3-29.8
  Change                             0.9            0.6-1.6

Limb circumference, with MVC (cm)
  Before series                     28.2     0.9   24.0-32.9
  After series                      29.1     0.9   24.5-33.7
  Change                             0.9            0.5-1.2

                                    Series 3

Variable                            [bar]X   SD    Range

MVC force (kg)
  Before series                     28.8     2.8   18.7-36.7
  After series                      39.9     2.8   28.9-53.6
  % change                          38.4     3.6   19.8-48.8

Limb circumference, relaxed (cm)
  Before series                     25.8     1.1   21.0-29.0
  After series                      26.8     1.1   21.6-29.6
  Change                             1.0            0.5-1.3

Limb circumference, with MVC (cm)
  Before series                     28.8     1.1   23.0-32.0
  After series                      30.1     1.1   24.1-33.3
  Change                             1.3            1.0-1.7

                                    Series 4

Variable                            [bar]X   SD    Range

MVC force (kg)
  Before series                     57.5     0.8   47.6-70.6
  After series                      89.8     2.0   62.6-108.4
  % change                          56.1     5.9   30.0-76.0

Limb circumference, relaxed (cm)
  Before series                     34.4     0.2   33.0-35.0
  After series                      35.8     0.2   35.0-36.0
  Change                             1.4            0.8-1.6

Limb circumference, with MVC (cm)
  Before series
  After series
  Change

(a) Kots and Xvilon (11) did not report statistical analyses of
their data. The group-average data shown do not permit analysis,
which would require "before" and "after" data paired by subject.

Table 3.
Maximum Force (in Kilograms, at the Pain Tolerance Threshold) With
Stimulation of the Wrist and Finger Flexors at Different Alternating
Current Frequencies in the Range of 100 Hz to 5 kHz (a)

Stimulation            100 Hz   500 Hz   1 kHz   2 kHz   3 kHz   5 kHz

Direct, continuous
  [bar]X                9.6     16.2     19.5    23.4    20.2
  SD                    3.1      4.9      5.0     5.7     4.4

Indirect, continuous
  [bar]X               18.6     21.6     23.5    18.8            13.5
  SD                    3.7      6.1      6.2     6.2             4.2

Indirect, 50-Hz
    bursts
  [bar]X               22.1     24.4     25.5    18.7            18.4
  SD                    4.8      5.4      4.8     3.8             2.8

(a) "Direct" refers to stimulation with electrodes over the muscle
belly, "indirect" refers to stimulation of the nerve trunk. From
Andrianova et al. (12) Only mean and standard deviation values were
given by Andrianova et al. (12) Ranges for the force values were not
stated. Andrianova et al (12) did not report statistical analyses of
their data. The group-average data shown do not permit analysis, which
would require data across frequencies for each subject so that
pair-wise comparisons could be made.

Table 4.
Verification of the Choice of Optimal Frequencies for Direct and
Indirect Stimulation of Forearm Muscles: Maximum Electrically
Induced Force (in Kilograms) at 1 kHz and 2.5 kHz (a)

                       1 kHz          2.5 kHz

Stimulation            [bar]X   SD    [bar]X   SD

Direct, continuous     23.6     4.1   26.3     4.5
Indirect, continuous   27.7     7.0   19.8     5.4

(a) The footnote to Table 3 regarding statistical analysis is
also applicable here. From Andrianova et al. (12)

Table 5.
Average Values of the Force Induced Through Direct Stimulation at
2.5 kHz of the Extensors and Flexors of the Foot and Indirect
Stimulation at 1 kHz of the Flexors of the Hand and Fingers to
Compare (a) Continuous Stimulation With 50-Hz Modulated Stimulation

                                                     Force (kg)

                                         Frequency
Muscle          Stimulation              (kHz)       [bar]X   SD

Triceps surae
  muscles       Direct, continuous       2.5          97.5    14.0
Triceps surae
  muscles       Direct, 50-Hz bursts     2.5         109.2    10.0
Wrist/finger
  flexors       Indirect, continuous     1            33.3     7.2
Wrist/finger
  flexors       Indirect, 50-Hz bursts   1            32.8     6.2

(a) The footnote to Table 3 regarding statistical analysis is also
applicable here.


References

(1) Selkowitz DM. High frequency electrical stimulation in muscle strengthening. Am J Sports Med. 1989;17:103-111.

(2) Selkowitz DM. Improvement in isometric strength of the quadriceps femoris muscle
"Quads" redirects here. For other uses see Quad
The quadriceps femoris (quadriceps, quadriceps extensor, guads or quads) includes the four prevailing muscles on the front of the thigh.
 after training with electrical stimulation. Phys Ther. 1985;65:186-196.

(3) Kots YM. Electrostimulation (Canadian-Soviet exchange symposium on electrostimulation of skeletal muscles Skeletal muscles
Muscles that move the skeleton. All of the muscles under voluntary control are skeletal muscles.

Mentioned in: Creatine Kinase Test
, Concordia University, Montreal, Quebec, Canada, December 6-15, 1977). Quoted in: Kramer J, Mendryk SW. Electrical stimulation as a strength improvement technique. J Orthop Sports Phys Ther. 1982;4:91-98.

(4) Delitto A, Brown M, Strube MJ, et al. Electrical stimulation of quadriceps femoris Noun 1. quadriceps femoris - a muscle of the thigh that extends the leg
musculus quadriceps femoris, quadriceps, quad

extensor, extensor muscle - a skeletal muscle whose contraction extends or stretches a body part
 in an elite weight lifter: a single-subject experiment. Int J Sports Med. 1989;10:187-191.

(5) Delitto A, Rose SJ, McKowen JM, et al. Electrical stimulation versus voluntary exercise in strengthening thigh thigh (thi) femur; the portion of the leg above the knee.

thigh
n.
The part of the leg between the hip and the knee. Also called femur.
 musculature musculature /mus·cu·la·ture/ (mus´kul-ah-cher) the muscular apparatus of the body or of a part.

mus·cu·la·ture
n.
The arrangement of the muscles in a part or in the body as a whole.
 after anterior cruciate ligament surgery. Phys Ther. 1988;68:660-663.

(6) Snyder-Mackler L, Delitto A, Stralka SW, Bailey SL. Use of electrical stimulation to enhance recovery of quadriceps femoris muscle force production in patients following anterior cruciate ligament reconstruction This article or section needs copy editing for grammar, style, cohesion, tone and/or spelling.
You can assist by [ editing it] now.
. Phys Ther. 1994;74:901-907.

(7) Snyder-Mackler L, Delitto A, Bailey SL, Stralka SW. Strength of quadriceps femoris muscle and functional recovery after reconstruction of the anterior cruciate ligament. J Bone Joint Surg Am. 1995;77: 1166-1173.

(8) Snyder-Mackler L, Garrett M, Roberts M. A comparison of torque generating capabilities of three different electrical stimulating currents. J Orthop Sports Phys Ther. 1989;11:297-301.

(9) Laufer Y, Ries JD, Leininger PM, Alon G. Quadriceps femoris muscle torques tor·ques  
n. Zoology
A band of feathers, hair, or coloration around the neck.



[Latin torqu
 produced and fatigue generated by neuromuscular electrical stimulation with three different waveforms. Phys Ther. 2001;81: 1307-1316.

(10) Ward AR, Robertson VJ. The variation in torque production with frequency using medium-frequency alternating current. Arch Phys Med Rehabil. 1998;79:1399-1404.

(11) Kots YM, Xvilon VA. Trenirovka mishechnoj sili metodom elektrostimuliatsii: soobschenie 2, trenirovka metodom elektricheskogo razdrazenii mishechi. Teor Pract Fis Cult. 1971;4:66-72.

(12) Andrianova GG, Kots YM, Marmyanov VA, Xvilon VA. Primenenie elektrostimuliatsii dlia trenirovki mishechnoj sili. Novosti Meditsinskogo Priborostroeniia. 1971;3:40-47.

(13) Babkin D, Timtsenko N (trans). Electrostimulation: notes from Dr YM Kots' (USSR USSR: see Union of Soviet Socialist Republics. ) lectures and laboratory periods presented at the Canadian-Soviet exchange symposium on electrostimulation of skeletal muscles, Concordia University, Montreal, Quebec, Canada, December 6-15, 1977. [Available from Dr Ward.]

(14) St Pierre D, Taylor AW, Lavoie M, et al. Effects of 2,500-Hz sinusoidal current on fibre area and strength of the quadriceps femoris. J Sports Med. 1986;26:60-66.

(15) Nelson RM, Hayes KW, Currier DP. Clinical Electrotherapy electrotherapy /elec·tro·ther·a·py/ (-ther´ah-pe) treatment of disease by means of electricity.

e·lec·tro·ther·a·py
n.
Medical therapy using electric currents.
. 3rd ed. Stamford, Conn: Appleton & Lange; 1999.

(16) McComas AJ. Skeletal skeletal /skel·e·tal/ (skel´e-t'l) pertaining to the skeleton.

skeletal

pertaining to the skeleton. See also skeletal muscle.
 Muscle Form and Function. Champaign Champaign (shămpān`), city (1990 pop. 63,502), Champaign co., E central Ill.; inc. 1860. It adjoins the city of Urbana and is a commercial and industrial center in a fertile farm area. The Univ. , Ill: Human Kinetics kinetics: see dynamics.
Kinetics (classical mechanics)

That part of classical mechanics which deals with the relation between the motions of material bodies and the forces acting upon them.
; 1996.

(17) Soloviev EN. Nyekogorii osobnyennostii elektrostimulyatsii na povishennik chastotak. Trudi instituta M VNIIMIO. 1963;vi:3.

(18) Ward AR, Robertson VJ. Variation in motor threshold with frequency using kHz frequency alternating current. Muscle Nerve. 2001; 24:1303-1311.

(19) Jones DA. High- and low-frequency fatigue revisited. Acta Physiol Scand. 1996;156:265-270.

(20) Jones DA. Muscle fatigue due to changes beyond the neuromuscular junction Neuromuscular junction
The site at which nerve impulses are transmitted to muscles.

Mentioned in: Botulinum Toxin Injections, Myasthenia Gravis

neuromuscular junction
. In: Porter R, Whelan J, eds. Human Muscle Fatigue: Physiological Mechanisms. London, England: Pitman Medical; 1981: 178-196.

(21) Djourno A. Sur quelques singularites de la contraction musculaire en courant Cou`rant´   

a. 1. (Her.) Represented as running; - said of a beast borne in a coat of arms.
n. 1. A piece of music in triple time; also, a lively dance; a coranto.
2.
 tetanisant de moyenne frequence fre·quence  
n.
Frequency.



[Middle English, multitude, from Old French, from Latin frequentia; see frequency.]

Noun 1.
. Comptes Rendus Hebdomadaires des Seances et Memories de la Societe de Biologie et de ses Filiales. 1952;146:398-399.

(22) Stefanovska A, Vodovnik L. Change in muscle force following electrical stimulation: dependence on stimulation waveform and frequency. Scand J Rehabil Med. 1985;17:141-146.

(23) d'Arsonval A. Action physiologique des courants alternatifs. Comptes Rendus Hebdomadaires des Seances et Memories de la Societe de Biologie et de ses Filiales. May 2, 1891:283-287.

(24) Geddes LA. A short history of the electrical stimulation of excitable excitable /ex·ci·ta·ble/ (ek-sit´ah-b'l) irritable (1).

ex·cit·a·ble
adj.
1. Capable of reacting to a stimulus. Used of a tissue, cell, or cell membrane.

2.
 tissue including therapeutic applications. The Physiologist physiologist /phys·i·ol·o·gist/ (fiz?e-ol´ah-jist) a specialist in physiology.

physiologist

a specialist in physiology.
. 1984; 27(suppl):s1-s47.

(25) Ward AR, Robertson VJ. Sensory, motor, and pain thresholds for stimulation with medium frequency alternating current. Arch Phys Med Rehabil. 1998;79:273-278.

(26) Walmsley RP, Letts G, Vooys J. A comparison of torque generated by knee extension with a maximal voluntary contraction vis-a-vis electrical stimulation. J Orthop Sports Phys Ther. 1984;6:10-17.

(27) Ward AR, Robertson VJ. The variation in fatigue rate with frequency using kHz frequency alternating current. Med Eng Phys. 2001;22: 637-646.

AR Ward, PhD, is Senior Lecturer senior lecturer
n. Chiefly British
A university teacher, especially one ranking next below a reader.
, Department of Human Physiology Human physiology is the science of the mechanical, physical, and biochemical functions of humans in good health, their organs, and the cells of which they are composed. The principal level of focus of physiology is at the level of organs and systems.  and Anatomy anatomy (ənăt`əmē), branch of biology concerned with the study of body structure of various organisms, including humans. Comparative anatomy is concerned with the structural differences of plant and animal forms. , Faculty of Health Sciences, La Trobe University 1. u/r = unranked

2.AsiaWeek is now discontinued. Student life
During the 1970s and 1980s, La Trobe, along with Monash, was considered to have the most politically active student body of any university in Australia.
, Victoria 3086, Australia (a.ward@latrobe.edu.au). Address all correspondence to Dr Ward.

N Shkuratova, PT, is a practicing physiotherapist physiotherapist /phys·io·ther·a·pist/ (-ther´ah-pist) physical therapist.

physiotherapist

physical therapist.
 and postgraduate postgraduate

after first degree graduation, the registerable degree in veterinary science.


postgraduate degree
may be a research degree, e.g. PhD, or a course-work masterate with a vocational bias, or any combination of these.
 student in the School of Physiotherapy School of Physiotherapy is located in Lahore, Punjab, Pakistan. It is located in Mayo Hospital and is affiliated with King Edward Medical College. , Faculty of Health Sciences, La Trobe University.

Dr Ward provided concept/idea for this work. Both authors provided writing, data collection and analysis, and consultation (including review of manuscript before submission). Ms Shkuratova provided translation of original Russian-language publications. The authors are indebted in·debt·ed  
adj.
Morally, socially, or legally obligated to another; beholden.



[Middle English endetted, from Old French endette, past participle of endetter, to oblige
 to Dr Aneta Stefanovska of the University of Ljubljana The University of Ljubljana (in Slovenian, Univerza v Ljubljani; in Latin, Universitas Labacensis) is the first and the largest university in Slovenia; with 56,000 enrolled students, it ranks among the biggest universities in the world.  for helpful discussion of Kots' work and for providing a draft manuscript by Professor Luigi Divieti of the Polytechnic Institute of Milan that provided links to the original Russian-language publications of Kots and co-workers.

This article was submitted December 27, 2001, and was accepted April 28, 2002.
COPYRIGHT 2002 American Physical Therapy Association, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2002, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Shkuratova, Nataliya
Publication:Physical Therapy
Geographic Code:4EXRU
Date:Oct 1, 2002
Words:7974
Previous Article:"Russian electrical stimulation": putting this perspective into perspective. (Introduction).
Next Article:Article titles in rehabilitation literature.(Bibliography)
Topics:



Related Articles
Functional electrical stimulation and lower extremity bracing for ambulation exercise of the spinal cord individual: a medically prescribed system.
Two theories of muscle strength augmentation using percutaneous electrical stimulation.
Factors influencing quadriceps femoris muscle torque using transcutaneous neuromuscular electrical stimulation.
Research and the cumulation of knowledge in physical therapy. (includes commentary and author response)
The effect of transcutaneous electrical stimulation on spinal motor neuron excitability in people without known neuromuscular diseases: the roles of...
"Russian electrical stimulation": putting this perspective into perspective. (Introduction).
An investigation into the analgesic effects of interferential currents and transcutaneous electrical nerve stimulation on experimentally induced...
A home program of sensory and neuromuscular electrical stimulation with upper-limb task practice in a patient 5 years after a stroke.(Case Report)