Rapid plasticity testing of silicones.


When a plastic material is subjected to a force, it continues to deform for as long as the force is applied, with most of the deformation deformation /de·for·ma·tion/ (de?for-ma´shun)
1. in dysmorphology, a type of structural defect characterized by the abnormal form or position of a body part, caused by a nondisruptive mechanical force.

2.
 remaining once the force has been removed (ref. 1). Plasticity testing is used to ensure the batch quality of a compound (or stock) and to ensure material consistency. It is difficult to get an exact idea of how a material behaves in a mill or extruder, and therefore plasticity tests have mainly been used to check the uniformity/quality of repeat batches.

The simplicity of compression (parallel) plate plastimeters led to their early adoption and continuing popularity. The compression plate compression plate Orthopedics A plate used to stabilize fractures, used with lag screws to provide dynamic compression on the tension side of bone; dynamic CPs are the most common types of plates, and have special oval screw holes with a beveled floor and an  plastimeter consists of two parallel plates used to compress a sample with a constant applied force; the thickness thickness (thik´nes) a measurement across the smallest dimension of an object.

triceps skinfold (TSF) thickness
 reduction is measured after a set time. Ira Williams (ref. 2) first reported this type of plastimeter in 1924, and the Williams plastimeter is still widely used today. However, the Williams measurement took between three minutes "Three Minutes" is the 46th episode of Lost. It is the twenty-second episode of the second season. The episode was directed by Stephen Williams, and written by Edward Kitsis and Adam Horowitz. It first aired on May 17, 2006 on ABC.  and ten minutes, excluding sample preparation. For effectiveness, results had to be known quickly, and there was seen to be a need lot a rapid plasticity test. Experimental work was carried out in the 1940s to determine the optimum test conditions (for rubber), resulting in the manufacture of the first Wallace Wal·lace , Alfred Russel 1823-1913.

British naturalist who developed a concept of evolution that paralleled the work of Charles Darwin.
 rapid plastimeter in 1951. This instrument provided a test that was quicker and less sensitive to operator error. The test pre-compresses the sample for 15 seconds to 1 mm (at a temperature of 100[degrees]C for natural rubber) and then further compresses the sample for 15 seconds under a load of 100 N. The thickness reduction gives a plasticity value. Although the instrument design has changed over the years, the Years, The

the seven decades of Eleanor Pargiter’s life. [Br. Lit.: Benét, 1109]

See : Time
 principle has remained the same. In both instruments, the plasticity number is a measure of the final sample thickness, reflecting the degree of flow within the sample. In all industries, repeatability and the sensitivity of the instrument to slight compound variations are important. While these two plastimeters have been studied previously, much of the work relates to the natural rubber industry (ref. 3). Traditionally, the rapid plastimeter has been used for testing unvulcanized Adj. 1. unvulcanized - (used of rubber) not subjected to the process of vulcanization
unvulcanised

unprocessed - not altered from an original or natural state; "unprocessed commodities"
 rubber, particularly raw natural rubber. This article studies the repeatability and sensitivity of these instruments with respect to the silicone rubber Noun 1. silicone rubber - made from silicone elastomers; retains flexibility resilience and tensile strength over a wide temperature range
synthetic rubber, rubber - any of various synthetic elastic materials whose properties resemble natural rubber
 industry.

Experimental

Standard Williams (ASTM ASTM
abbr.
American Society for Testing and Materials
 D926-98 [ref. 4]) and Wallace rapid plasticity tests (ISO (1) See ISO speed.

(2) (International Organization for Standardization, Geneva, Switzerland, www.iso.ch) An organization that sets international standards, founded in 1946. The U.S. member body is ANSI.
 2007 [ref. 5]) were carried out on a range of silicone rubber samples. Both the instruments used were designed and manufactured by H.W. Wallace.

For the Williams plastimeter (Wallace P2), two 5 [cm.sup.2] pieces of 12 [micro]m mylar My·lar  

A trademark used for a thin strong polyester film.

Noun 1. Mylar - a thin polyester film
trademark - a formally registered symbol identifying the manufacturer or distributor of a product
 film were used to set the zero position. A 2.0 [cm.sup.3] (approximately 16 mm diameter and 10 mm thick) sample of silicone rubber was cut from a sheet using a rotary Rotary can refer to:
  • Rotary engine, a type of internal combustion engine from the early 20th century
  • Rotary Woofer, a type of loudspeaker capable of very low frequency sound
  • Rotary International, a service organization
  • Rotary milking shed
 cutter cutter, small, one-masted sailing vessel, with a rig similar to that of a sloop except that it usually has a sliding bowsprit and a topmast. From 1800 to 1830 cutters were in service between England and France. . The sample was placed between the two pieces of film and located between the plastimeter platens. The load was applied to the sample for three minutes. The final result was read from a dial gauge dial gauge

Any of a number of deviation-type gauges that indicate the amount by which an object being gauged deviates from the standard. This deviation is usually shown in units of measurement, but some gauges show only whether the deviation is within a certain range.
 and converted to a WP (Williams plasticity) value. One unit represents a change in thickness of 0.01 ram. For the rapid plastimeter (Wallace P12E), 0.4 [cm.sup.3] samples were prepared using the standard constant volume sample cutter (mylar film was used to maintain consistency). The rapid plastimeter was operated semi-automatically with unheated platens. The sample was pre-compressed to 1 mm for 15 seconds, and then the load was applied for a further 15 seconds. The result (in rapid plasticity units) was digitally displayed. One unit represents a change in thickness of 0.01 mm (ref. 6).

Several tests were carried out on a range of typical silicone rubber samples to determine result repeatability, sensitivity and standard deviation In statistics, the average amount a number varies from the average number in a series of numbers.

(statistics) standard deviation - (SD) A measure of the range of values in a set of numbers.
. The sample mix was altered slightly (by 1%) to determine the sensitivity of both instruments to small compound variations. Standard Williams and rapid plastimeter samples were taken from both the original and the new mix and tested on the instruments.

Results

A range of silicone rubber samples was tested on both types of instrument. The standard deviations were normalized to full range. While the samples tested on the Williams plastimeter were found to have slightly lower standard deviations than those given by the rapid plastimeter (table 1), the spread of results was found to be acceptable on both instruments.

The sample range was re-tested (with each sample having a slightly different mix as described above). Result averages were calculated and the percentage change in plasticity value before and alter the mix change was tabulated (table 2). It was found that, for three out of four samples measured with the rapid plastimeter, the percentage change was greater. Only the lower plasticity number sample showed a larger percentage change when measured with the Williams plastimeter.

The altered compound samples were also tested for repeatability, and the same trend for a small spread of results was noted on both instruments.

Other factors influence accuracy and repeatability, such as specimen SPECIMEN. A sample; a part of something by which the other may be known.
     2. The act of congress of July 4, 1836, section 6, requires the inventor or discoverer of an invention or discovery to accompany his petition and specification for a patent with specimens
 preparation and instrument operation. The rapid plastimeter uses standard samples, produced in a few seconds with a constant volume cutter. A range of silicone rubber samples was prepared with the rapid plastimeter sample cutter and then weighed. The sample weights (nominally nom·i·nal  
adj.
1.
a. Of, resembling, relating to, or consisting of a name or names.

b. Assigned to or bearing a person's name: nominal shares.

2. Existing in name only.
 0.5 g) were found to be consistent to within +0.005 g (i.e., [+ or -] 1%). In contrast, the Williams plastimeter uses samples prepared with a rotary cutter. In practice, a constant diameter is maintained, but the sample height varies. In order to obtain the correct sample volume (conforming to the standard), the sample's weight is adjusted by adding or removing small amounts of compound. Therefore, the accuracy depends on the operator and the balance used. The Williams sample preparation procedure can easily take two to three minutes.

The operator may influence the results when using both types of instrument. Several different samples were tested on both instruments by two different operators, and standard deviations were calculated from their results. In the majority of cases, the percentage difference between the two operators' standard deviations was greater for the Williams than for the same sample measured on the rapid plastimeter, indicating better repeatability between operators when using the rapid plastimeter.

Overall, rapid plastimeter repeatability was acceptable, but lower than expected. Previous work (ref. 7) indicated that, when testing natural rubber, the rapid plastimeter generally exhibited better repeatability than shown here. One factor highlighted in the previous work showed that the type of paper or film used affected the end result. Since mylar film was used on both instruments throughout the experimental work, some rapid plastimeter tests were repeated using Rizla RizLa+ (known commonly as "Rizla") is a brand of papers in which tobacco is rolled to form cigarettes.

Rizla rolling papers are available in a range of thicknesses and sizes, indicated by the colour of the packaging.
 paper (recommended in the natural rubber standard, ISO 2007 [ref. 51).

It was found that paper improved the standard deviation of the rapid plastimeter results. Paper use gave results with standard deviations that were comparable, or better than those obtained using mylar film. Different operators also experienced better repeatability when using paper in place of film (table 3). It was also noted that lower plasticity samples appeared to be more affected by the use of paper or film, tending to give similar standard deviations.

Discussion

For the range of silicone rubber samples tested, it was found that the results obtained from the Williams plastimeter had a lower normalized standard deviation than those obtained from the rapid plastimeter (when using mylar). However, the spread was found to be acceptable on both instruments. When Rizla paper was used instead of mylar film on the rapid plastimeter, the standard deviations were found to be better than, or comparable to, those obtained from the Williams plastimeter.

While the rapid plastimeter is semi-automatic Semi-automatic may refer to:
  • A semi-automatic firearm, a firearm which automatically reloads. See also:
  • Semi-automatic rifle
  • Semi-automatic pistol
 in operation, the Williams plastimeter result is more operator dependent, both in sample preparation and during measurement. Therefore, more care must be taken when using the Williams plastimeter.

One additional area of concern with the Williams plastimeter is the method of sample preparation. While the rapid plastimeter uses accurately prepared samples from special cutter pliers pliers,
n a tool of pincer design with jaws of varying shapes; used for holding, bending, stretching, contouring, and cutting.

pliers, contouring,
n
, the Williams uses samples of a constant diameter, with the correct volume attained at·tain  
v. at·tained, at·tain·ing, at·tains

v.tr.
1. To gain as an objective; achieve: attain a diploma by hard work.

2.
 by adding or subtracting small amounts of the sample.

It was found that the rapid plastimeter was generally more sensitive to sample mix variations than the Williams plastimeter. This is important when trying to discriminate dis·crim·i·nate  
v. dis·crim·i·nat·ed, dis·crim·i·nat·ing, dis·crim·i·nates

v.intr.
1.
a.
 between different samples. Additionally, if different operators obtain results, confusion can arise when small compound variations are observed on the Williams (these could be due to either the operator or the sample mix). Previous work (ref. 7) showed that when using a rapid plastimeter to test natural rubber samples, little variation between operator is seen.

While these tests were carried out at room temperature, testing may also take place at elevated temperatures. The rapid plastimeter uses electrically heated platens, allowing the sample temperature to be easily controlled. In contrast, the Williams plastimeter requires the entire instrument to be placed in an oven (sample temperature control is not straightforward or simple).

In addition, the sample used for the rapid plastimeter is only about 20% of the volume required for a Williams plastimeter. This is also an advantage when measuring samples at a higher temperature, since temperature equilibrium equilibrium, state of balance. When a body or a system is in equilibrium, there is no net tendency to change. In mechanics, equilibrium has to do with the forces acting on a body.  may be reached more quickly.

Conclusion

The measurements show that the spread of results between the two plastimeters is comparable for the range of silicone rubber tested. Several advantages in using the rapid plastimeter have been presented.

From the work carried out, it has been possible to see the effects of a change in sample mix and operator on the two plastimeters. The results indicate that for the range of samples tested, the rapid plastimeter is more sensitive to small changes in compound mix and less operator dependent than the Williams plastimeter.

Most importantly Adv. 1. most importantly - above and beyond all other consideration; "above all, you must be independent"
above all, most especially
, testing can be carried out in a fraction of the time on the rapid plastimeter, a 90% time saving is realizable, once sample preparation time is included.
Table 1--result repeatability

Sample    Williams std. dev.    Rapid plastimeter std. dev.

  1             0.005                      0.008
  2             0.017                      0.021
  3             0.007                      0.008
  4             0.004                      0.015
  5             0.002                      0.005
  6             0.003                      0.004

Table 2--sensitivity to compound change

Sample                    % change between sapmle mix
           Williams
            value           Williams           Rapid
          (original)      plastimeter       plastimeter

  1          684               43               54
  2          454               35               60
  3          317                5                7
  4          254               21               11

Table 3--result repeatability between operators
using mylar film and Rizla paper

Sample                        Rapid plastimeter s.d.

           Williams        Operator 1        Operator 2
          plastimeter
           std. dev.     Mylar    Rizla    Mylar    Rizla

  1          0.017       0.033    0.006    0.031    0.011
  2          0.004       0.011    0.010    0.005    0.017
  3          0.002       0.002    0.005    0.003    0.002
  4          0.003       0.010    0.004    0.010    0.005


References

(1.) RABRM Bulletin, Special Issue on Plasticity, Vol. 8, No. 4 (1954).

(2.) I. Williams, Industrial and Engineering Chemistry, Vol. 16, No. 4, pp. 362-364 (1924).

(3.) Physical Testing of Rubber, R.P. Brown, pp. 85-98, second edition (1986).

(4.) ASTM D926-98. Standard Test Method for Rubber Property--Plasticity and Recovery (Parallel Plate Method).

(5.) ISO 2007: 1991, Physical Testing of Rubber, Part A59, Methods using plastimeters, Section 59.1, Determination of the rapid plasticity number.

(6.) B.M. McGarry, "Rapid plasticity and plasticity retention index testing," National Rubber Conference, India India, officially Republic of India, republic (2005 est pop. 1,080,264,000), 1,261,810 sq mi (3,268,090 sq km), S Asia. The second most populous country in the world, it is also sometimes called Bharat, its ancient name. India's land frontier (c.  (1994).

(7.) R. Morgans Morgans can refer to:
  • Morgans, the inhabitants of coastal waters south west of Thailand in the Andaman Sea. They are also know as Moken or Morgan sea gypsies.
  • Morgans is an alternative rendering of the name for Welsh water sprites: Morgens.
, S. Lackovic, B. McGarry, G. Dinnage and B. Pearce Pearce may refer to:

In places:
  • Pearce, Australian Capital Territory
People with the surname Pearce:
  • Pearce (surname)
See also
  • The Division of Pearce, an Australian Electoral Division in Western Australia
, "Importance of experimental parameters on rapid plasticity testing for PRI PRI: see Institutional Revolutionary party.


(Primary Rate Interface) An ISDN service that provides 23 64 Kbps B (Bearer) channels and one 64 Kbps D (Data) channel (23B+D), which is equivalent to the 24 channels of a T1 line.
," International Rubber Conference, India (1998).
COPYRIGHT 2003 Lippincott & Peto, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2003, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Process Machinery
Author:Banner, S.
Publication:Rubber World
Date:Nov 1, 2003
Words:1902
Previous Article:Mixtures of olefin polymers and nitrile rubbers.(Patent News)
Next Article:Third generation metallocene EPDMs.(Tech Service)



Related Articles
Custom mixing of silicone rubber.
High performance precipitated silica equivalent to fumed silica in silicone reinforcement.
A processing guide to silicone rubber extrusions.(part 1)
INDUSTRY HOTLINKS @ www.rubberworld.com.
Rubber World Hotlinks @ rubberworld.com.(Martin Marietta Magnesia Specialties)
INDUSTRY HOTLINKS@www.rubberworld.com.
INDUSTRY HOTLINKS @ www.rubberworld.com.
Plasticity measurement.
Plasticity measurement.(Instruments)
Silicone rubber materials for high voltage transmission and distribution insulators.

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles