Quantitating fluorescence intensity from fluorophore: assignment of MESF values.


A procedure is presented to convert the comparison of measured fluorescence signals into a comparison of fluorescence yields (FY). The fluorescence yield, which is a property of a solution or a suspension, is defined as the product of the fluorophore concentration and the molecular quantum yield The quantum yield of a radiation-induced process is the number of times that a defined event occurs per photon absorbed by the system. Thus, the quantum yield is a measure of the efficiency with which absorbed light produces some effect. . The paper revises the measurement model which relates the measured fluorescence signal to the FY. The equality of FY of two solutions provides an equivalence between the concentrations of fluorophore in the two solutions. The equivalence is the basis for quantitation in terms of molecules of equivalent soluble fluorophore (MESF MESF Mobile Electronics and Security Federation (UK trade association)
MESF Maritime Expeditionary Security Force
MESF Minimum Engineered Safety Feature
). The quantitation procedure starts with the measurement of fluorescence signals from a serial dilution of fluorescein fluorescein /flu·o·res·ce·in/ (fldbobr-res´en) a fluorescing dye; its sodium salt is used as a tracer in retinal angiography and as a diagnostic aid for revealing corneal trauma and fitting contact lenses.  solutions to obtain a calibration of a fluorometer fluorometer /flu·o·rom·e·ter/ (fldbobr-rom´e-ter) the instrument used in fluorometry, consisting of an energy source (e.g., a mercury arc lamp or xenon lamp) to induce fluorescence, filters or monochromators for selection of the . The fluorometer is used to measure the fluorescence signal of a suspension of microspheres with immobilized fluorescein isothiocyanate Noun 1. fluorescein isothiocyanate - a fluorochrome commonly conjugated with antibodies for use in indirect immunofluorescence
fluorescein isocyanate

fluorochrome - any of various fluorescent substances used in fluorescence microscopy to stain specimens
 (FITC FITC

fluorescein isothiocyanate; used as a fluorescent label for proteins, especially antibodies.
). The calibration is used to obtain the concentration of soluble fluorophores which gives the same fluorescence signal as the microsphere Not to be confused with Glass microphere.
This article largely refers to micropheres or protein protocells as small spherical units postulated by some scientists as a key stage in the origin of life.
 suspension. The number concentration of microspheres is measured and the equality of fluorescence yields is used to obtain the number of soluble fluorescein moleculcs equivalent to a single microsphere.

Key words: cytometry; FITC; fluorescein; fluorescence yield; MESF; microspheres.

**********

1. Introduction

RM 8640 consists of six plastic bottles each containing a 2 mL suspension of polymethyl methacrylate polymethyl methacrylate /poly·meth·yl meth·ac·ryl·ate/ (pol?e-meth´il meth-ak´ril-at) a thermoplastic acrylic resin formed by polymerization of methyl methacrylate. Abbreviated PMMA. Written also polymethylmethacrylate.  (PMMA PMMA polymethyl methacrylate. ) microspheres with a specified amount of immobilized fluorescein isothiocyanate (FITC). The microspheres are intended for calibrating the fluorescence response of flow cytometers [1]. This paper describes the procedures used for assigning values of MESF (molecules of equivalent soluble fluorophore) to the microspheres with immobilized FITC. There are three major measurements in this procedure. First, the concentration of microspheres is measured using a Multisizer 3 (Coulter Corporation, Miami FL) particle counter A particle counter is an instrument that detects and counts particles. Applications of particle counters are separated into two primary categories:
  • Aerosol particle counters
  • Liquid particle counters
Aerosol particle counters
 (1). Second, a fluorometer is calibrated cal·i·brate  
tr.v. cal·i·brat·ed, cal·i·brat·ing, cal·i·brates
1. To check, adjust, or determine by comparison with a standard (the graduations of a quantitative measuring instrument):
 using SRM (1) (Storage Resource Management) The management of the storage resources in an organization in order to avoid duplication of files and to determine space utilization across all servers.  1932, a fluorescein solution. Third, the fluorescence signal is measured for each of the microsphere suspensions. Finally, the data from the three measurements are used to calculate the MESF values of the microspheres.

In practice, we made serial dilutions of SRM 1932 and calibrated the fluorometer response as a function of fluorescein concentration. Since the concentration of fluorescein varies from [10.sup.-12] mol/L to [10.sup.-9] mol/L, it was necessary to pay special attention to contamination, linearity, photodegradation, and background subtraction subtraction, fundamental operation of arithmetic; the inverse of addition. If a and b are real numbers (see number), then the number ab is that number (called the difference) which when added to b (the subtractor) equals . The measurements of the concentration of the microsphere suspension constitutes the operational definition of particle concentration. There are at present no particle number The particle number, N, is the number of so called 'elementary particles' (or elementary constituents) in a thermodynamical system. The particle number is a fundamental parameter in thermodynamics and it is conjugate to the chemical potential.  standards to validate the concentration measurement.

A cytometer was used to measure the fluorescence signals associated with the five microsphere populations each with a different amount of immobilized FITC. A valid assignment of MESF values should yield a linear relation between the measured fluorescence signal in the cytometer and the assigned MESF values. A linearization In mathematics and its applications, linearization refers to finding the linear approximation to a function at a given point. In the study of dynamical systems, linearization is a method for assessing the local stability of an equilibrium point of a system of nonlinear differential  procedure was used to impose a linear relation between the cytometer response of the five microsphere populations and their MESF values.

2. Revised Measurement Model and the Assignment of MESF

It was pointed out [2] that the quantum yield as defined in Eqs. (A3) and (A14) in Ref. [3] is the ratio of fluorescent radiant flux radiant flux
n.
The rate of flow of radiant energy.

Noun 1. radiant flux - the rate of flow of radiant energy (electromagnetic waves)
flux - the rate of flow of energy or particles across a given surface
 to absorbed radiant flux. Since the radiant flux is the product of the number flux and average spectral energy, the ratio of radiant fluxes is not the same as the ratio of number fluxes. In the following we introduce a modification to the measurement model which allows a consistent use of quantum yield, a molecular property defined in terms of number flux.

In the previous paper [3] we expressed the fluorescence spectral radiance, [L.sub.f]([[lambda].sub.m],[[lambda].sub.x]), as

[L.sub.f]([[lambda].sub.m],[[lambda].sub.x]) = [[[S.sub.f]([[lambda].sub.m])]/[S'([[lambda].sub.m])]]L'([[lambda].sub.m]) = 2.3FNly([[lambda].sub.m],[[lambda].sub.x])[epsilon]([[lambda].sub.x])[[PHI].sub.i]([[lambda].sub.x])[DELTA][[lambda].sub.x] (1)

[[lambda].sub.m] and [[lambda].sub.x] are the emission and excitation wavelengths, [S.sub.f], S', are measured signals, L' is the spectral radiance of a reference source, F, l, [[PHI].sub.i]([[lambda].sub.x]), and [DELTA][[lambda].sub.x] are instrument characteristics, and N, y, and [epsilon] are sample properties. The quantity y([[lambda].sub.m],[[lambda].sub.x]) relates the absorbed radiant flux at wavelength [[lambda].sub.x] to the fluorescent radiant flux at wavelength [[lambda].sub.m]. In other words Adv. 1. in other words - otherwise stated; "in other words, we are broke"
put differently
, the radiant flux from fluorescence is a fraction y of the absorbed radiant flux. The radiant flux can be converted to a photon number flux by dividing the radiant flux by the energy of a single photon. Thus the quantity y([[lambda].sub.m],[[lambda].sub.x]) can be converted into a relation between fluorescence photon number flux and absorbed photon number flux by multiplying it by the ratio of the respective wavelengths.

y'([[lambda].sub.m],[[lambda].sub.x]) = [[[lambda].sub.m]/[[lambda].sub.x]] y([[lambda].sub.m],[[lambda].sub.x]). (2)

The quantity y'([[lambda].sub.m],[[lambda].sub.x]) is conveniently separated into a quantum yield [phi] and a normalized relative photon emission function s'([[lambda].sub.m],[[lambda].sub.x]) (1/nm), where

[integral]s'([[lambda].sub.m], [[lambda].sub.x])d[[lambda].sub.m] = 1. (3)

Thus Eq. (1) can be rewritten as

[[[S.sub.f]([[lambda].sub.m])]/[S'([[lambda].sub.m])]][[[lambda].sub.m]/[[lambda].sub.x]]L'([[lambda].sub.m]) = 2.3FNl[phi]s' ([[lambda].sub.m],[[lambda].sub.x])[epsilon]([[lambda].sub.x])[[PHI].sub.i]([[lambda].sub.x])[DELTA][[lambda].sub.x]. (4)

Integrating over all emission wavelengths gives

[integral][S.sub.f]([[lambda].sub.m])[[[[lambda].sub.m]/[[lambda].sub.x]][[L'([[lambda].sub.m])]/[S'([[lambda].sub.m])]]d[[lambda].sub.m] = [OMEGA]N[phi][epsilon]([[lambda].sub.x])[I.sub.0]. (5)

The change in the measurement model amounts to a modification of the spectral correction function given by the term in brackets on the left side of Eq. (5). Eq. (5) supplants a similar equation in Ref. [3]. The left side of Eq. (5) is found experimentally by performing the specified operation on the measured fluorescence emission spectrum emission spectrum: see spectrum. . The reference source used to calibrate To adjust or bring into balance. Scanners, CRTs and similar peripherals may require periodic adjustment. Unlike digital devices, the electronic components within these analog devices may change from their original specification. See color calibration and tweak.  the detector wavelength response was unpolarized. In order to minimize the effects due to polarization of the fluorescence emission, the incident laser beam polarization was confined to the plane defined by the incident and detected light beams. We made no estimate of possible artifacts artifacts

see specimen artifacts.
 due to polarization differences.

Suppose that the operation in Eq. (5) is performed on the emission spectrum from a reference solution with known concentration of fluorophore and the emission spectrum from a suspension of microspheres with immobilized FITC. Furthermore, the number concentration of microspheres is known. In the case that the two numbers on the left of Eq. (5) are equal, the corresponding solution and suspension properties on the right side of Eq. (5) are also equal.

[OMEGA][N.sub.sol][[phi].sub.sol][[epsilon].sub.sol]([[lambda].sub.x])[I.sub.0] = [OMEGA][N.sub.sus][[phi].sub.sus][[epsilon].sub.sus] ([[lambda].sub.x])[I.sub.0]. (6)

The subscripts "sol" and "sus" in Eq. (6) refer to solution and suspension, respectively. We assume that the experimental conditions used for the measurements on solution and suspension are the same and that the solution and suspension have equal indexes of refraction refraction, in physics, deflection of a wave on passing obliquely from one transparent medium into a second medium in which its speed is different, as the passage of a light ray from air into glass. . In that case, [I.sub.0] and [OMEGA] are the same on both sides of Eq. (6) and can be factored out.

[N.sub.sol][[phi].sub.sol][[epsilon].sub.sol] ([[lambda].sub.x]) = [N.sub.sus][[phi].sub.sus][[epsilon].sub.sus]([[lambda].sub.x]). (7)

It is known from measurements that there are shifts in the wavelength of maximum absorption. It is likely that the absolute value of the extinction coefficient is also different. The extinction coefficient of fluorophore immobilized on a microsphere has not been measured due to predominance of scattering. However, we make the major assumption that the molar molar /mo·lar/ (mo´lar)
1. pertaining to a mole of a substance.

2. a measure of the concentration of a solute, expressed as the number of moles of solute per liter of solution. Symbol M, , or mol/L.
 extinction coefficient is the same for fluorophore in solution and immobilized on the microsphere. Thus, Eq. (7) reduces to an equality of fluorescence yields.

[N.sub.sol][[phi].sub.sol] = [N.sub.sus][[phi].sub.sus]. (8)

Based on Eq. (8) [N.sub.sol] is equivalent to [N.sub.sus]. The calculation used to assign molecules of equivalent soluble fluorophore (MESF) values to the microspheres is given by

MESF = [[N.sub.A]/[1000]][[C.sub.eq]/[N.sub.spheres]] (9)

where [N.sub.sus] = [N.sub.sphere] ([mL.sup.-1]) is the number concentration of fluorescein labeled microspheres and [C.sub.eq] is the molar concentration Noun 1. molar concentration - concentration measured by the number of moles of solute per liter of solution
molarity, M

concentration - the strength of a solution; number of molecules of a substance in a given volume
 of soluble fluorescein which gives the same value for the left side of Eq. (5) as the suspension of microspheres. Avogadro's constant A·vo·ga·dro's constant
n.
See Avogadro's number.
, [N.sub.A] (6.022 X [10.sup.23]), in Eq. (9) is a conversion factor between molar and number concentrations. The equivalent concentration of soluble fluorescein is determined using the fluorescein calibration curve In analytical chemistry, a calibration curve is a general method for determining the concentration of a substance in an unknown sample by comparing the unknown to a set of standard samples of known concentration.

[C.sub.eq] = [10.sup.-intercept](FSX FSX Flight Simulator X
FSX Financial Services Exchange
FSX Fire Support Execution
FSX File System Xml
[P.sub.adj])[.sup.slope] (10)

where "intercept" and "slope" are the linear fit parameters describing the relationship between the logarithm logarithm (lŏg`ərĭthəm) [Gr.,=relation number], number associated with a positive number, being the power to which a third number, called the base, must be raised in order to obtain the given positive number.  of the observed fluorescence signal and the logarithm of the concentration of fluorescein. FS is the fluorescence signal of the microsphere suspension evaluated according to according to
prep.
1. As stated or indicated by; on the authority of: according to historians.

2. In keeping with: according to instructions.

3.
 the left side of Eq. (5). The value FS has to be adjusted for possible differences in illumination intensity between the calibration measurements and the microsphere measurements. The ratio of the average of the power readings taken during the calibration and microsphere measurements is set equal to the adjustment factor, [P.sub.adj], which multiplies FS in Eq. (10). The power readings varied by less than 1% during the calibration or the microsphere measurements. However, since the two measurements were taken on different days, the difference in average power could be as high as several percent. Eq. (9) and Eq. (10) constitute the operational definition of the MESF assignment.

In the following we describe the procedure used for obtaining the five factors needed in Eq. (9) and Eq. (10). As an example, the value of MESF is obtained for one specific suspension of microspheres.

3. Calibration of the Fluorometer

Figure 1 shows a schematic of the fluorometer used in the MESF assignments. A water-cooled argon argon (är`gŏn) [Gr.,=inert], gaseous chemical element; symbol Ar; at. no. 18; at. wt. 39.948; m.p. −189.2°C;; b.p. −185.7°C;; density 1.784 grams per liter at STP; valence 0.  ion laser An ion laser is a gas laser which uses an ionized gas as its lasing medium.[1] Like other gas lasers, ion lasers feature a sealed cavity containing the laser medium and mirrors forming a Fabry-Perot resonator.  (Lexel model 95) was the source of 488 nm light. A glass slide reflected a portion of the output beam and directed it towards a photo diode (Newport 818 UV) whose output was processed by a power meter (Newport 1815-C). The accuracy of the power reading was about 2% (product specifications). The output of the power meter was monitored as an index of relative illumination power which was used to determine the factor [P.sub.adj] in Eq. (10). The laser beam, transmitted through the glass slide, passed a neutral density (ND) filter wheel which contained ND filters of nominal optical density (OD) values 0, 1, 2, and 3. The beam passed through a laser line filter to eliminate plasma lines from the laser and then was directed via two mirrors to the entrance aperture of a 10X microscope objective. The two reflections changed the vertical polarization into horizontal. The objective lens focused the laser beam on a square capillary flow cell (WWP WWP Wounded Warrior Project
WWP Worms World Party (game)
WWP Workers World Party
WWP World Wide Packets (Veradale, WA)
WWP WebWorks Publisher
WWP Women Writers Project
100375 from Polymicro Technologies) with a inner dimension of 100 [micro]m. The portion of the capillary that was illuminated by the laser was stripped of the polyimide Pronounced "poly-ih-mid." A type of plastic (a synthetic polymeric resin) originally developed by DuPont that is very durable, easy to machine and can handle very high temperatures. Polyimide is also highly insulative and does not contaminate its surroundings (does not outgas).  coating which normally surrounds the glass capillary. The capillary flow cell was mounted on a rotary stage A rotary stage is a component of a motion system used to restrict an object to a single axis of rotation. The terms rotary table or rotation stage are often used interchangeably with rotary stage.  which in turn was mounted on a X-Y translation stage. The rotary stage was used to position the capillary so that the incident beam was perpendicular to one of the sides. The X-Y stage was used to position the capillary at the position of the monochromator A monochromator is an optical device that transmits a mechanically selectable narrow band of wavelengths of light or other radiation chosen from a wider range of wavelengths available at the input.  entrance slit image.

[FIGURE 1 OMITTED]

The Model 270M monochromator, made by JY Horiba, was equipped with a CCD-3000 Detector System. The CCD CCD
 in full charge-coupled device

Semiconductor device in which the individual semiconductor components are connected so that the electrical charge at the output of one device provides the input to the next device.
 chip used in the measurements was back illuminated with a minimum quantum efficiency of 65% at 550 nm. The operational temperature of the chip was 213 K with a dark current less than 4 [e.sup.-] [pixel.sup.-1] [min.sup.-1]. The pixel layout was 1024 by 256, and the dynamic range of the 16 bit analog to digital converter was 65535 digital number (DN). The measurements were performed by binning the 256 pixels in the short direction of the CCD chip. The entrance slit of the monochromator was equipped with a mechanical shutter, the single axial Model 227MCD MCD Minor Civil Division
MCD McDonalds (restaurant)
Mcd Macedonian (linguistics)
MCD Municipal Corporation of Delhi
MCD Magnetic Circular Dichroism
MCD Mad Cow Disease
 (JY Horiba). Appendix A describes the procedures used to validate the performance of the fluorometer [4].

3.1 Calibration of the Fluorometer With SRM 1932

SRM 1932 certifies the concentration of fluorescein as (60.97 [+ or -] 0.88) [micro]mol/kg. Given the density of the SRM buffer as 1.003 g/mL, the molar concentration of the SRM 1932 is (61.15 [+ or -] 0.88) [micro]mol/L. This value of the concentration was used as the initial concentration. Serial dilutions were made by combining previously made solution with additional buffer. All solutions were prepared gravimetrically using a calibrated balance (Sartorius 2024MP) with a resolution of 0.01 mg. The errors were obtained from the standard deviation In statistics, the average amount a number varies from the average number in a series of numbers.

(statistics) standard deviation - (SD) A measure of the range of values in a set of numbers.
 of four weighins. The standard error in the fluorescein concentrations was about 1.6% and originated mostly from the initial error in the SRM concentration. The weighin errors contributed a minimal error to the final fluorescein concentrations.

3.2 Measurement of the Fluorescence Signal

The fluorescein solution was pumped with a peristaltic pump A peristaltic pump is a type of positive displacement pump used for pumping a variety of fluids. The fluid is contained within a flexible tube fitted inside a circular pump casing (though linear peristaltic pumps have been made).  through a capillary flow cell [5]. The 10X objective and the capillary were mounted on appropriate mounts to provide the necessary adjustment. A good adjustment was indicated by the appearance of a clean circular beam cross section after the transit through the capillary. Poor adjustment was characterized by complex interference bands from light reflected by various surfaces of the capillary. The capillary tube (length = 0.7 m) was coupled to a plastic tube (length = 20 cm) via a stainless steel stainless steel: see steel.
stainless steel

Any of a family of alloy steels usually containing 10–30% chromium. The presence of chromium, together with low carbon content, gives remarkable resistance to corrosion and heat.
 pressure "Swagelok" coupling. The plastic tube was inserted into the pumping mechanism of a peristaltic pump (PP). A centrifuge centrifuge (sĕn`trəfyj), device using centrifugal force to separate two or more substances of different density, e.g., two liquids or a liquid and a solid.  vial vial

a small bottle.
 (1.5 mL capacity) contained the test solution and a small magnetic stirring bar. The ends of the capillary or the plastic tube were inserted in the solution and the pump direction set appropriately to pump the solution to the waste container A waste container (known more commonly in British English as a dustbin, rubbish-bin, ashcan or simply bin and American English as a trash can) is a container, which is usually made out of metal or plastic.[1]. . Neither the capillary nor the plastic tube touched the waste solution. Formation and detachment of small droplets above the waste container was an indicator of flow.

The spectrum was accumulated over a period of 40 s leading to appearance of sharp spikes in the CCD response. The spikes are pixels with unusually large charge content, the spikes are confined to one or two adjacent pixels. We removed the spikes by simply replacing the contents of the pixel containing a spike (attributed to cosmic ray cosmic ray

High-speed particle (atomic nucleus or electron) that travels through the Milky Way Galaxy. Some cosmic rays originate from the Sun, but most come from outside the solar system.
 events hitting a pixel in the CCD array) by an average of the contents of two nearby pixels. The necessary overall dynamic range was obtained by varying the integration time from 0.5 s to 40 s, and by the dynamic range of the CCD itself (about 60). The ND filter was not used to modify the illumination intensity.

The peristaltic pump produced a pulsating flow with a time period of approximately 1.6 s. The strong illumination caused substantial photodegradation of the fluorescein solution in the capillary. The photodegradation and the pulsating flow produced a time variation in the fluorescence signal. This variation was averaged adequately during integration times longer then 5 s. However, for shorter integration times we had to take multiple measurements and average the resulting fluorescence signals. The pulsating fluorescence intensity enlarged the measurement errors for the concentrated fluorescein solutions. We avoided changing the intensity of the illuminating beam since that would change the photodegradation rate for measurements performed on solutions with different fluorescein concentrations. The assumption is made that the photodegradation is the same for solution and fluorescein immobilized on microspheres. To minimize possible systematic errors due to photodegradation, the flow conditions for the calibration and microsphere measurements were made as similar as possible. The polarization anisotropy anisotropy /an·isot·ro·py/ (an?i-sot´rah-pe) the quality of being anisotropic.
anisotropy (an´āsôt´r
 for fluorescein solutions was approximately zero. Therefore, the measured spectra are characteristic of a solution of random emitters and systematic effects due to polarization are small.

3.3 Background Subtraction

Figure 2a shows the measured spectra of a pure phosphate buffer (solid circles) and a fluorescein solution with a concentration of approximately 16 pM (open circles) in phosphate buffer. The laser line filter shown in Fig. 1 was a critical component in the fluorometer since without it the spectra in Fig. 2a would be an order of magnitude A change in quantity or volume as measured by the decimal point. For example, from tens to hundreds is one order of magnitude. Tens to thousands is two orders of magnitude; tens to millions is three orders of magnitude, etc.  higher and dominated by laser plasma lines. The spectra in Fig. 2a have DN values larger than 2000, suggesting that CCD linearity correction was not significant. Figure 2b shows the spectrum when the buffer response is subtracted from the fluorescein solution response. As expected, the emission spectrum peaks at approximately 510 nm. The integration of the spectrum was performed by summing the DN values of the subtracted spectrum in Fig. 2b. The resulting truncation errors for wavelengths less than 595 nm and wavelengths greater than 620 nm were not evaluated. We estimate that these truncations lead to a systematic bias of less than -1%. Similar spectra were collected for solutions with higher values of fluorescein concentration and the integration time was lowered as required to insure that the resulting CCD response was not saturated. The subtracted spectrum was corrected for the spectral response The variable output of a light-sensitive device that is based on the color of the light it perceives.  of the detector as described in Appendix A. The corrected integration times were used to normalize normalize

to convert a set of data by, for example, converting them to logarithms or reciprocals so that their previous non-normal distribution is converted to a normal one.
 all integrated fluorescence signals (FS) to the fluorescence signal (FS) at an integration time of 1 s.

[FIGURE 2 OMITTED]

Figure 3 shows a log-log plot of the integrated DN values on the horizontal axis and five different concentrations of fluorescein plotted on the vertical axis. The best linear fit to the data in Fig. 3 is

log(concentration) = -15.20 + 0.957 X log(FS) (11)

The errors of the fit parameters were 0.09 and 0.009 for the intercept and slope, respectively. The errors were obtained from a linear regression Linear regression

A statistical technique for fitting a straight line to a set of data points.
 procedure in Mathcad. The lowest point (for a concentration of approximately 7 pmol/L) was systematically lower than expected from the linear trend set by the higher points. We rationalize this as an effect of adsorption adsorption, adhesion of the molecules of liquids, gases, and dissolved substances to the surfaces of solids, as opposed to absorption, in which the molecules actually enter the absorbing medium (see adhesion and cohesion).  of fluorescein on the capillary walls [4]. A calibration was accepted if the slope of the best fit fell between 0.95 and 1.05. A perfectly linear relation has a slope of 1.0, however a deviation of [+ or -] 0.05 was accepted. The values of "intercept" and "slope" are used in Eq. (10).

[FIGURE 3 OMITTED]

4. Measurement of Fluorescence From Microspheres With Immobilized FITC

The measurements of fluorescence signal were carried out in the identical apparatus as the calibration with serially diluted fluorescein solutions. Since the microsphere measurements were carried out after the fluorescein solution, great care was needed to eliminate possible contamination. Prior to the microsphere measurements, the capillary flow cell was washed for several hours. The washing was performed by pumping buffer through the capillary in alternating directions. The switch in pumping direction was important to clear possible dead spaces in the connection between the capillary tube and the plastic tube. As a rule of thumb, the cleaning was sufficiently good when the CCD signal was about 370 DN at 510 nm with an integration time of 10 s, and a power indicator of about 20. This number was obtained through experience. During all measurements the following pumping sequence was followed: first the suspension was pumped through the plastic tube into the capillary (this direction filled the capillary quickly); second the suspension was pumped through the capillary into the tube and fluorescence spectra were accumulated. Normal flow was indicated by the presence of a scattering diffraction pattern diffraction pattern

The interference pattern that results when a wave or a series of waves undergoes diffraction, as when passed through a diffraction grating or the lattices of a crystal.
 in the transmitted light. The pattern indicated the presence of spheres in the capillary as well as proper alignment. The intensity of the diffraction pattern fluctuated as expected since the number of microspheres in the sensing volume was of the order of twenty. On occasion, small bubbles passed through the illuminated region in the capillary. The passage of a bubble was obvious from the distortion of the transmitted laser beam. Clogging was obvious because flow stopped and the fluorescence signal decreased.

4.1 Fluorescence Measurement

Figure 4a shows the spectrum measured for a suspension of microspheres (Suspension #1, open circles) and a suspension of blank microspheres (solid circles). The suspensions were identical to those used in the microsphere concentration measurements. The measurements in Fig. 4a were performed with integrating time of 40 s, and power indicator displaying 19.2. Figure 4b shows the difference spectrum. The location of maximum emission shifts to the red, and the spectrum is broader. Both facts are typical of emission from immobilized FITC and serve as additional indicators that the flow cell is clean and free of fluorescein in solution. The quality of background subtraction was gauged by the disappearance of the water Raman line centered at 585 nm. In further analysis, the spectra in Fig. 4a were corrected for CCD non linearity ([approximately equal to]0.8% effect) and normalized to an integration time of 1 s. The spectrum was summed and the errors of the summed spectrum were estimated from multiple measurements. Spectral response corrections were made on the data in Fig. 4 ([approximately equal to]3 % effect). The resultant summed spectrum for Suspension #1 was (11220 [+ or -] 600) DN. The value of the summed spectrum was substituted for FS in Eq. (10). Taking the ratio of average power levels during calibration and microsphere fluorescence measurements gave [P.sub.adj] = 0.92 [+ or -] 0.02.

[FIGURE 4 OMITTED]

5. Measurement of Microsphere Concentration

The microsphere concentration was measured using a Coulter Multisizer 3 particle counter. The instrument detects small changes in conductivity between two reservoirs separated by a narrow orifice orifice /or·i·fice/ (or´i-fis)
1. the entrance or outlet of any body cavity.

2. any opening or meatus.orific´ial


aortic orifice
 of diameter 100 [micro]m. Whenever a particle passes through the orifice from one reservoir to the other, a transient change in conductivity is detected and recorded as a particle. In practice, the reservoir outside the orifice is the test reservoir into which samples are placed. A volumetric volumetric /vol·u·met·ric/ (vol?u-met´rik) pertaining to or accompanied by measurement in volumes.

vol·u·met·ric
adj.
Of or relating to measurement by volume.
 syringe withdraws fluid from the test reservoir through the orifice and the concentration is determined as the particle counts divided by the preset preset Cardiac pacing A parameter of a pacemaker that is programmed permanently when manufactured  volume of fluid withdrawn. The fluid that was placed into the test reservoir was obtained from the original fluid by diluting the original fluid about 500 times using Isotone Isotone

One of two or more atoms which display a constant difference A - Z between their mass number A and their atomic number Z.
 fluid. Isotone is a proprietary fluid which has optimal properties for the performance of the Multisizer 3. The composition of Isotone is sufficiently close to that of phosphate buffer saline (PBS PBS
 in full Public Broadcasting Service

Private, nonprofit U.S. corporation of public television stations. PBS provides its member stations, which are supported by public funds and private contributions rather than by commercials, with educational, cultural,
) so that microsphere suspension stability is not degraded. The concentration of the particles in the original fluid is found by multiplying the measured concentration by the reciprocal of the dilution.

5.1 Determination of the Dilution Factor of the Test Suspension

The microspheres were obtained from Bangs Laboratories as a special order and were delivered in 5 mL opaque plastic bottles. The nominal solids mass fraction was about 2 %. The microspheres, as shipped, were suspended in a solvent optimized for enhancing the stability of the suspension. For the purpose of measuring the fluorescence signal of the microspheres, it was necessary to re-suspend the microspheres in pH = 7.2 PBS. The first step was to take about 1 ml of the microsphere suspension directly from the plastic bottle and centrifuge it (Marathon 13K centrifuge from Fisher Scientific Fisher Scientific, formally Fisher Scientific International, Inc. and colloquially Fisher was a biotechnology company that provided products and services to the global scientific research and United States clinical laboratory markets.  at 2000 rpm). The pellet of microspheres was then re-suspended in 1 ml of PBS, pH 7.2, containing [approximately equal to]0.1 mg SDS 1. (company) SDS - Scientific Data Systems.
2. (tool) SDS - Schema Definition Set.
. An additional advantage of buffer exchange is that possible soluble fluorescent impurities in the bottle are minimized. The mass of about 200 [micro]g of the resulting suspension was placed into a container with a mass of Isotone equivalent to about 100 ml of Isotone. The mass of the buffer was determined using a balance, OHaus ED4130, while the suspension was transferred using calibrated 100 [micro]L or 200 [micro]L pipettes and the mass approximated as 0.1996 g using the known density of water. A gravimetric gravimetric /grav·i·met·ric/ (grav?i-me´trik) pertaining to measurement by weight; performed by weight, as a gravimetric method of drug assay.

grav·i·met·ric
adj.
1.
 check of the amount of solution delivered by the 100 [micro]L pipette pipette /pi·pette/ (pi-pet´) [Fr.]
1. a glass or transparent plastic tube used in measuring or transferring small quantities of liquid or gas.

2. to dispense by means of a pipette.
 gave (0.10016 [+ or -] 0.0006) g using water at 23 [degrees]C and a Sartorius 2024 MP balance. The error in the reciprocal of the dilution factor was estimated to be less than 1 %. The remaining suspension was used for fluorescence measurements as described above.

5.2 Concentration Measurement

Prior to each series of measurements the operation of the Multisizer 3 was verified by running a suspension of calibration microspheres. The quality of the measurement was evaluated by measuring the mean diameter of the microspheres, and the coefficient of variation Coefficient of Variation

A measure of investment risk that defines risk as the standard deviation per unit of expected return.
 (CV) of the diameter values. The diluted microsphere suspension was placed in the Multisizer 3 and stirred at an indicated rate of 13 rpm. An image of the orifice was always examined to verify that the orifice was free from debris. The instrument analytical volume was set to 500 [micro]L or 1 ml and a dilution factor (as determined above) was entered as a parameter in the acquisition software. Figure 5 shows a typical result of a single Multisizer 3 measurement. The horizontal axis gives the inferred particle diameter, and the vertical axis gives the number of times a particle with this diameter passed through the orifice. Using the dilution factor, the vertical axis can be converted into concentration. The inferred concentration of Suspension 1 was (1.427 [+ or -] 0.016) X [10.sup.6] [mL.sup.-1] for all particles with diameters between 6.3 [micro]m and 8.1 [micro]m. The peak to the right of the main peak corresponds to "doublets dou·blet  
n.
1. A close-fitting jacket, with or without sleeves, worn by European men between the 15th and 17th centuries.

2.
a. A pair of similar or identical things.

b. A member of such a pair.
". These are events associated with the passage of two particles. The concentration of "doublets" (diameters between 8.3 [micro]m and 9.7 [micro]m) was calculated to be (0.077 [+ or -] 0.002) X [10.sup.6] [mL.sup.-1]. Since the concentration of particles in the test solution is small, it is likely that the "doublets" correspond to permanent associations of two microspheres. For permanent associations, the radii ra·di·i  
n.
A plural of radius.


radii
Noun

a plural of radius
 of doublets and triplets are related to the particle radius by [r.sub.doublet dou·blet
n.
A pairing of two lenses to optically correct a chromatic and spherical aberration.
] = [2.sup.1/3][r.sub.sphere] and [r.sub.triplets] = [3.sup.1/3][r.sub.sphere] respectively. These relations are consistent with the observed size distributions of singlets, doublets, and triplets. This measurement of particle associations is relevant to the proper interpretation of the fluorescence signal as discussed below. The concentration measurement was repeated five times and the error associated with each concentration was found by dividing the standard deviation by the square root of 5, the number of trials. It should be noted that the error in the concentration is about 5% implying that the error due to the uncertainty of the dilution factor (about 1 %) is negligible. The error of the concentration measurement is due mainly to the statistics of the particles entering the orifice. Whenever possible, the errors in the concentration measurement were minimized by using more concentrated test suspensions and larger sampling volumes.

[FIGURE 5 OMITTED]

5.3 Correction for Microsphere Doublets

Both the Multisizer 3 and the cytometer measurements show the presence of microsphere doublets-permanent associations of two microspheres. Some aggregation of colloidal particles is expected and the amount depends on previous treatments (e.g., sonication sonication /son·i·ca·tion/ (son?i-ka´shun) exposure to sound waves; disruption of bacteria by exposure to high-frequency sound waves.

son·i·ca·tion
n.
, age, and solvent). The concentration measurements were performed with the same suspension as the fluorescence measurements. Furthermore the two measurements were performed on the same day. The simplest assumptions are that the Multisizer 3 gives the correct concentration of doublets in the suspension used for fluorescence measurements and that the fluorescence intensity from doublets is twice the fluorescence intensity from single microspheres. Therefore, in comparing the fluorescence yield of a suspension and solution we multiply the doublet concentration by a factor of two relative to the singlet concentration.

[N.sub.Total] = [N.sub.Single] + 2 [N.sub.Double], (12)

Where [N.sub.Single] is the number concentration of single microspheres, [N.sub.Double] is the number concentration of double microspheres, and [N.sub.Total] is the total concentration which should be used in comparing fluorescence yields in Eq. (9). The assumption that the fluorescence intensity from a double microsphere is twice the fluorescence from a single microsphere is qualified by such consideration as distortion of the illuminating light and changes in quantum yield at point of contact between the microspheres. The measured polarization anisotropy in microsphere suspensions was about 0.08. This small value was rationalized by the large tether tether

to tie an animal up by the head or neck so that it can graze but not move away. See also barton tether.
 length of the chain of seven carbon atoms that immobilized the FITC to the microsphere surface. A long tether permits considerable rotational freedom. The systematic effects due to polarization differences between solution and microsphere suspension were neglected.

6. Assignment of MESF Values

The assignment of values of molecules of equivalent soluble fluorophore (MESF) entails the comparison of the fluorescence signal from suspensions of microspheres with immobilized fluorophore and solutions of fluorophore [3]. The comparison requires the integration of the emitted fluorescence over all wavelengths. The comparison requires a correction for differences in the molar absorption coefficient molar absorption coefficient
n.
Symbol A spectrophotometric unit indicating the light a substance absorbs with respect to length, usually centimeters, and concentration,
 of soluble and immobilized fluorescein. Fluorescence excitation spectra show that the absorption spectra of the soluble and immobilized fluorescein are shifted relative to each other [3]. The fluorescence excitation spectra allow us to estimate the change in absorbance absorbance /ab·sor·bance/ (-sor´bans)
1. in analytical chemistry, a measure of the light that a solution does not transmit compared to a pure solution. Symbol .

2.
 at 488 nm due to the shift. However, we were not able to measure the absolute value of the molar absorption coefficient for the immobilized fluorescein. This adjustment is left for a future refinement. The values of FS, "intercept", "slope", Padj, and [N.sub.spheres] were used in Eq. (9) and Eq. (10) to calculate a value of MESF of 1667 [+ or -] 400. The same procedure was carried out for the other suspensions. Four independent series of measurements were carried out and the resulting MESF values averaged to give the final value reported in the Certificate of Investigation.

7. Linearization of MESF Assignments With a Cytometer

The objective of this measurement was to demonstrate that the MESF assignments obtained with the fluorometer were self-consistent. The microspheres were passed through a cytometer, and the response was measured. Each population of microspheres produced a population of fluorescence pulses which are characterized by a mean pulse area and a standard variation. Since the cytometer response is linear, the mean pulse height of the five populations of microspheres should correlate linearly with the assigned values of MESF. The cytometer does not provide a measurement of the absolute MESF values, however it does place a stringent constraint on relative MESF values. The MESF values obtained with the fluorometer were modified so that they correlate linearly with the mean channel measured with the cytometer. Appendix B gives details of the procedure used for validating the performance of the cytometer.

7.1 Microsphere Measurements

Alignment microspheres from Spherotech Corp. were used to align the cytometer laser beam. The alignment was sufficiently good when the fluorescence pulse mean channel was between 180 000 DN and 200 000 DN, and the CV was better than 4 %. After the alignment, the six populations of the microspheres in the reference material were mixed and pumped through the cytometer and the corresponding fluorescence and scattering peaks recorded. The suspensions were prepared by putting two drops of the suspension from each of the six plastic bottles into 1 ml of PBS buffer. Figure 6 shows typical results. Note that the single and double microsphere signals are resolved. The dense groupings with circular bounds correspond to single microsphere signals. Diagonally to the upper right of each dense grouping are less dense groupings (not enclosed by boundaries) corresponding to the passage of double microspheres. The table in Fig. 6 gives the properties of the groups of dots enclosed by the circular bounds. The Y geometric mean (mathematics) geometric mean - The Nth root of the product of N numbers.

If each number in a list of numbers was replaced with their geometric mean, then multiplying them all together would still give the same result.
 gives the mean scattering pulse amplitude, while the X geometric mean gives the mean fluorescence pulse height for each population. As expected, the scattering is relatively constant for the five populations, while the fluorescence signal differs substantially.

Figure 7 shows a plot of the log of the mean fluorescence channel (X geometric means in Fig. 6) associated with each microsphere as a function of the log of the assigned MESF value for each microsphere. As described above, the MESF assignments were performed using the fluorometer. A fit to a straight line gives a slope of 0.984, indicating that there is a small deviations from linearity. All points are within two standard deviations of the best straight line fit. Since the cytometer is a linear device, the measured mean channels should correlate linearly with the MESF values which are proportionate to the number of fluorophore on the microsphere and hence to the fluorescence signal. This fact provides a means to linearize lin·e·ar·ize  
tr.v. lin·e·ar·ized, lin·e·ar·iz·ing, lin·e·ar·iz·es
To put or project in linear form.



lin
 the assigned MESF values.

7.2 Linearization

The response of the cytometer was shown to be linear over the dynamic range encompassing the response of the five microsphere populations. Therefore, the MESF values have to correlate linearly with the observed mean channels in the cytometer measurement. The cytometer measurement provides a relative ordering of the MESF values but gives no measure of the absolute values. The procedure that was used to impose a linear relation on the MESF values determined by fluorometer was as follows. The solid circles in Fig. 7 show the MESF values obtained in Series 4 measurements as a function of mean cytometer channel. The data in Fig. 7 was fitted with a straight line whose slope was constrained to 1.027, which characterizes the cytometer linearity. Next, the MESF value given by the straight line was calculated for each mean channel. This calculated value is the linearized MESF value obtained for that series of measurements.

[FIGURE 7 OMITTED]

8. Certification of MESF Values

The MESF values were assigned by averaging the values obtained in four independent measurements each consisting of 1) calibration of the fluorometer using SRM 1932, 2) measurement of the microsphere fluorescence intensity, and 3) measurement of the microsphere concentration. The data in each set of four measurements gave an assignment of MESF values to the five populations of microspheres. The four different assignments provide a measure of reproducibility and an estimate of random error. The possible error obtained from the variation of the four MESF assignments was consistent with the error estimate for each of the four MESF assignments. The average values are reproduced in the Certificate of Investigation for RM 8640. The cytometer measurements were used to linearize the MESF values yielding values of linearized MESF. The four values of linearized MESF were averaged to give the average linearized MESF values in the Certificate of Investigation. The certificate includes both the average MESF values determined by the fluorometer, and the average linearized MESF values.

The MESF values were assigned under certain assumptions which are restated below in the order of decreasing relevance.

1. The molar extinction coefficient is the same for fluorescein in solution and fluorescein immobilized on the microspheres. It is known that there are differences in the wavelength of maximum absorption between fluorescein in solution and on the surface of the microsphere. It is expected that the magnitude at maximum absorption will also be different, however these have not been measured yet. The difference in molar extinction coefficient can be taken into account as soon as the values become available. The equality of fluorescence yield would imply that [N.sub.sus] microspheres are equivalent to a concentration of soluble fluorophore given by [N.sub.sol][[[epsilon].sub.sol]([[lambda].sub.x])/[[epsilon].sub.sus]([[lambda].sub.x])].

2. Photodegradation of fluorescein is the same in solution and at the surface of the microsphere. Systematic measurements of photodegradation in the two environments are not available. Differences in photodegradation rate would lead to systematic differences in the fluorescence signal between calibration measurements and microsphere measurements.

3. Residual polarization of fluorescence emission is the same for fluorescein in solution and at the surface of the microsphere. The measured polarization anisotropies are slightly different for fluorescein in solution and on the surface of a microsphere. Sensitivity of detectors to polarization would lead to small systematic differences in fluorescence signal between fluorescein in solution and on the surface of microspheres.

4. Adsorption on capillary flow cell walls is the same. If the adsorption (and holdup in dead spaces) of fluorescein in solution and fluorescein on microspheres is not the same, then there will be systematic differences in the fluorescence signal at low and high concentrations.

5. Illumination of FITC immobilized on microspheres is the same as the illumination of fluorescein in homogeneous solution.

9. Conclusions

A method was described for comparing the fluorescence yields of a solution of fluorescein and a suspension of microspheres with immobilized fluorescein isothiocyanate (FITC). The equality of fluorescence yields leads to an assignment of molecules of equivalent soluble fluorophore (MESF) to a microsphere with immobilized FITC. The MESF values may be the appropriate units for comparing fluorescence measurements. There is a need to investigate the validity of some of the assumptions that were made in carrying out the MESF assignment as described above.

10. Appendix A. Fluorometer Characterization

10.1 Wavelength Calibration

The wavelength calibration was performed using mercury lamp lines, and a neon lamp Noun 1. neon lamp - a lamp consisting of a small gas-discharge tube containing neon at low pressure; luminescence is produced by the action of currents at high frequencies that are wrapped a few turns around the tube
neon induction lamp, neon tube
 (Oriel 6032). The parameters in the data acquisition software were adjusted so that over the wavelength range 480 nm to 700 nm, the deviation between the true and measured wavelengths was less than 1 nm.

10.2 CCD Linearity

The output of a light emitting diode See LED.  (LED) was passed through a variable ND filter and focused on the end of a bifurcated bi·fur·cate  
v. bi·fur·cat·ed, bi·fur·cat·ing, bi·fur·cates

v.tr.
To divide into two parts or branches.

v.intr.
To separate into two parts or branches; fork.

adj.
 optical fiber. One of the arms of the optical fiber was placed at the location of the sample in the monochromator setup (see Fig. 1), and the other arm was placed in front of a photo diode (PD). Silicon photodiodes are known to be linear within 0.1% up to a photocurrent pho·to·cur·rent  
n.
An electric current produced by illumination of a photoelectric material.
 of 200 [micro]A [6]. The LED spectrum was recorded by the CCD for an integration time of 0.2 s. The spectrum was taken with automatic subtraction of the dark current. The CCD response was characterized by finding the average value of DN in a narrow range (518 nm to 522 nm) of wavelengths around the maximum response. The integrated CCD response was compared to the photo diode reading.

We examined the dependence of the average signal in DN as a function of the PD response. There was a substantial deviation from linearity at low values of DN. The most likely source of this deviation is trapping of electrons during the readout (1) A small display device that typically shows only a few digits or a couple of lines of data.

(2) Any display screen or panel.
 process [7]. During readout, the electrons are switched sequentially from pixel to pixel. The trapping centers prevent some of the electrons from reaching the final register pixel. Since the number of trapping sites is fixed, the relative importance of these sites increases with decreasing number of total electrons. The measured deviations provide a factor which can be used to linearize the CCD response. Thus the linearized DN value is given by

(DN)[.sub.linear] = (DN)[.sub.measured](1 + f)

f = [10.sup.[0.6418-0.718l'log(D[N.sub.measured])]] (1)

where (DN)[.sub.measured] is the DN value that is read out during a measurement and (DN)[.sub.linear] is the linearized value of the CCD response which is used in further data analysis. The factor f was obtained from the fit of the deviation of the measured CCD data from the linear PD response. We did not assign an error to the correction. The correction was applied to background measurements and sample measurements prior to subtraction of background. The correction is negligible for DN values above 600.

10.3 Integration Time Linearity

The CCD accumulates electrons for a preset integration time that is determined by a mechanical shutter located after the entrance slit of the monochromator. To measure the correspondence between the integration time setting in the software and the actual time, we illuminated the monochromator entrance slit with a constant light source and measured the CCD response for different indicated integration times [t.sub.indicated]. The reference light source (see Sec. 10.4) was used as the constant light source. The CCD response was integrated from 540 nm to 560 nm. The ratio of the CCD response divided by indicated time [t.sub.indicated] relative to the CCD response for 1 s of indicated integration time showed a bias in the indicated time setting. At indicated times longer than 1 s, the actual integration time is shorter than the indicted INDICTED, practice. When a man is accused by a bill of indictment preferred by a grand jury, he is said to be indicted.  time. The difference between the actual integration time, [t.sub.actual], and the indicated integration time, [t.sub.indicated], was corrected by multiplying the indicated time by the correction factor 1.00102-0.000782 X [t.sub.indicated].

10.4 Spectral Response

The reference lamp's output was calibrated at NIST (National Institute of Standards & Technology, Washington, DC, www.nist.gov) The standards-defining agency of the U.S. government, formerly the National Bureau of Standards. It is one of three agencies that fall under the Technology Administration (www.technology.  over the range of wavelengths 340 nm to 800 nm in steps of 20 nm. The output port of the calibrated lamp was placed at the location of the sample in the fluorometer. The variable iris of the calibrated lamp was set so that the CCD response was between 10 000 DN and 40 000 DN. The CCD linearity is excellent in this region.

We formed the ratio of calibrated output of the reference lamp to the output at 520 nm (normalized reference output) and compared it to the ratio of measured CCD response of the reference lamp to the CCD response at 520 nm (normalized CCD response). The sharp decline in the response at shorter wavelengths was due to the holographic See holographic storage.  filter which was used to reject the 488 nm excitation light. The spectral correction factor was obtained by dividing the normalized reference output by the normalized CCD response to the reference lamp. Multiplying the measured CCD response by the correction factor corrects for the variability of the detector response over the wavelength range.

11. Appendix B. Cytometer Characterization

The cytometer was constructed to be as simple as possible with all physical processes open to inspection. An air-cooled Argon ion laser (Omnichrome Model 150) provided the source of 488 nm illumination. The laser beam was focused by a spherical lens spherical lens
n.
Abbr. sph A lens in which all refracting surfaces are spherical.
 with a focal length Focal length

A measure of the collecting or diverging power of a lens or an optical system. Focal length, usually designated f
 of 50 mm. The focal point focal point
n.
See focus.
 was located in the flow channel of a cytometer flow cell provided by Becton Dickinson BD (NYSE: BDX), is a medical technology company that manufactures and sells medical devices, instrument systems and reagents. Founded in 1897 and headquartered in Franklin Lakes, New Jersey, BD employs 27,000 people in nearly 50 countries.  Biosciences. The sample was pumped by a syringe pump (Yale Apparatus Model YA-12), and the sheath sheath (sheth) a tubular case or envelope.

arachnoid sheath  the continuation of the arachnoidea mater around the optic nerve, forming part of its internal sheath.
 fluid was pumped by the pressure in the container vessel. A flow meter flow meter

Device that measures the velocity of a gas or liquid. It has applications in medicine as well as in chemical engineering, aeronautics, and meteorology. Examples include pitot tubes, venturi tubes, and rotameters (tapered graduated tubes with a float inside that is
 (Aalberg Model TMR TMR

total mixed ration.

TMR 1 Trainable mentally retarded 2 Transmyocardial revascularization, see there
1-010426) in the sheath flow line gave an indication of the flow rate (usually set to 90 scale units). The laser beam passed through a glass plate whose orientation provided a sensitive adjustment of the beam position in the scattering plane. The flow cell contained the collection optics that focused the emitted light about 25 cm from the lens where photomultiplier photomultiplier: see photoelectric cell.  (PMT See photomultiplier tube. ) detectors (Hamamatsu Model H6780) were placed. The usual arrangement of dichroic dichroic

characterized by dichroism.
 mirror (DM) and bandpass filters selected the fluorescence and elastically scattered light components. The outputs from the two PMTs were processed by digital electronics provided by Becton Dickinson Biosciences (BD FACSDiVa system). An oscilloscope oscilloscope (əsĭl`əskōp'), electronic device used to produce visual displays corresponding to electrical signals. Displays of such nonelectrical phenomena as the variations of a sound's intensity can be made if the phenomena are  provided a visual monitor of the pulses associated with the side scattering (SSC SSC Secondary School Certificate
SSC Standard Systems Center (USAF)
SSC State Services Commission (New Zealand)
SSC Swedish Space Corporation
SSC Salem State College (Massachusetts) 
) and the first fluorescence (FL1) channels of the detection electronics.

11.1 Linearity and Dynamic Range

The output of a green LED was focused on a slit of a chopper and then split by a glass plate reflector reflector: see telescope.  and passed to the inputs of two optical fiber (FO) bundles. One of the split beams (that transmitted through the glass plate) passed through a neutral density (ND) filter, and the FO routed the light to the PMT associated with the first fluorescence channel (FL1). The other beam (reflected from the glass plate) was incident on the FO that routed the light to the PMT associated with the side scattering channel In scattering theory, a scattering channel is a quantum state of the colliding system before or after the collision ().  (SSC). The chopper rotation was adjusted to give pulses that approximated the duration of the pulses from the microspheres in the flow cell. The adjustment was performed by visual inspection of the oscilloscope traces. The pulse rate pulse rate
n.
The rate of the pulse as observed in an artery, expressed as beats per minute.
 was about 190 pulses per second. The linearity measurements were performed by noting the mean channel of the detected pulses in FL1 for a given OD value of the ND filter. Six ND filters were purchased from Newport Corp., and the attenuation Loss of signal power in a transmission.
Attenuation

The reduction in level of a transmitted quantity as a function of a parameter, usually distance. It is applied mainly to acoustic or electromagnetic waves and is expressed as the ratio of power densities.
 values were used as provided by the manufacturer. The data was accumulated by recording events for different OD values of the ND filters. The six mean channels corresponding to OD values of 0, 0.51, 1.05, 1.50, 1.98, and 2.49 were recorded. The mean channels were plotted on a log-log scale versus the OD values. The average slope was found to be 1.027 [+ or -] 0.008. Therefore, we conclude that the cytometer response is linear. We assume that the photon pulses in the above simulation and from fluorescent microspheres behave in an identical fashion.

The noise properties of the cytometer are given by the coefficient of variation (CV) defined as the standard deviation of a pulse distribution divided by the mean pulse amplitude [8]. A linear relation between (CV)[.sup.2] and the inverse of the mean pulse amplitude was observed, and indicated that the main source of noise was the statistics of photon arrival at the photomultiplier cathode.
Region     # of        X              Y              % of
#          Events      Geometric      Geometric      Gated
                       Mean           Mean           Cells

None       30000       20987.58       63138.92       100.0
1           4769       80093.62       61790.11        15.9
2           4670       32087.41       61669.2         15.57
3           6639        5553.2        61640.29        22.13
4           4612        1326.39       61630.44        15.37
5           5242         308.8        61822.93        17.47

Fig. 6. Output window from FCS Express V2, a program for the analysis of
cytometer data. The upper diagram shows the distribution of side
scattering signals (SSC) and fluorescence signals (FL1) from six
populations of microspheres containing different amounts of immobilized
FITC. The broad peak closest to the SSC-A axis is the signal from the
blank microspheres with no immobilized FITC. The table below the diagram
contains the geometric means of the scattering signal (Y) and
fluorescence signal (X) for each of the five populations defined by the
circular regions in the graph above. The row label "1" denotes the
population with the highest fluorescence signal. The row labeled "None"
contains the geometric means for the entire data set. Other parameters,
such as the width of the distribution, can also be obtained for each
population.


Acknowledgment

The authors are indebted to Dr. John Lu for assistance in the statistical analysis of data. We are indebted to Dr. Gary Kramer for the use of a reference light source.

Accepted: January 31, 2005

Available online: http://www.nist.gov/jres

(1) Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by NIST, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

12. References

[1] A. K. Gaigalas, L. Li, et al., "The Development of Fluorescence Intensity Standards." J. Res. Natl. Inst. Stand. Technol. 106, 381-389 (2001).

[2] W. V. Prestwich, Unit of Medical Physics, McMaster University McMaster University, at Hamilton, Ont., Canada; nondenominational; founded 1887. It has faculties of humanities, science, social sciences, business, engineering, and health sciences, as well as a school of graduate studies and a divinity college.  Personal communication (2003).

[3] A. Schwartz, L. Wang, et al., "Quantitating Fluorescence Intensity from Fluorophore: The Definition of MESF Assignment." J. Res. Natl. Inst. Stand. Technol. 107, 83-91 (2002).

[4] R. Mavrodineanu, J. I. Shultz, et al., eds., Accuracy in Spectrophotometry spectrophotometry

Branch of spectroscopy dealing with measurement of radiant energy transmitted or reflected by a body as a function of wavelength. The measurement is usually compared to that transmitted or reflected by a system that serves as a standard.
 and Luminescence luminescence, general term applied to all forms of cool light, i.e., light emitted by sources other than a hot, incandescent body, such as a black body radiator.  Measurements, Washington D.C., National Bureau of Standards National Bureau of Standards: see National Institute of Standards and Technology.

National Bureau of Standards - National Institute of Standards and Technology
 (1973).

[5] L.-Q. Li and L. M. Davis, "Rapid and efficient detection of single chromophore chromophore /chro·mo·phore/ (kro´mo-for) any chemical group whose presence gives a decided color to a compound and which unites with certain other groups (auxochromes) to form dyes.  molecules in aqueous aqueous /aque·ous/ (a´kwe-us)
1. watery; prepared with water.

2. see under humor.


a·que·ous
adj.
 solution." Appl. Opt. 34(18), 3208-3217 (1995).

[6] T. Kubarsepp, A. Haapalinna, et al., "Nonlinearity measurements of silicon photodetectors." Appl. Opt. 37(13), 2716-2722 (1998).

[7] R. M. Smith. How Linear are Typical CCDs?, Experimental Astronomy 8, J. W. Beletic and P. Amico, eds., Kluwer Academic Publishers, Netherlands (1998).

[8] E. S. Chase and R. A. Hoffman, "Resolution of Dimly Fluorescent Particles: Measure of Fluorescence Sensitivity." Cytometry 33, 267-279 (1998).

A. K. Gaigalas and Lili Wang Lili Wang (March 6 1971 - October 12 2002) was a graduate student at North Carolina State University murdered by another graduate student named Richard Anderson. On that day, Anderson shot and killed her while she was playing tennis.

National Institute of Standards and Technology National Institute of Standards and Technology, governmental agency within the U.S. Dept. of Commerce with the mission of "working with industry to develop and apply technology, measurements, and standards" in the national interest. , Gaithersburg, MD 20899-8312

Abe Schwartz
For the wrestling character Abe "Knuckleball" Schwartz, see Steve Lombardi


Abe Schwartz (1881-1963) was a well-known klezmer musician of the 1920s.

Abe was born outside of Bucharest, Romania, and moved to the United States in 1899.


Center for Quantitative Cytometry PO Box 19344, San Juan San Juan, city, Argentina
San Juan (săn wän, Span. sän hwän), city (1991 pop. 353,476), capital of San Juan prov., W Argentina. It is a commercial and industrial center in an agricultural region.
, PR 00919

Gerald E. Marti

Center for Biologics Evaluation and Research The Center for Biologics Evaluation and Research (CBER) is one of six main centers for the Food and Drug Administration, which is in the United States Department of Health and Human Services. , FDA FDA
abbr.
Food and Drug Administration


FDA,
n.pr See Food and Drug Administration.

FDA,
n.pr the abbreviation for the Food and Drug Administration.
 Bethesda, MD 20892

and

Robert F. Vogt, Jr.

Division of Laboratory Sciences, CDC See Control Data, century date change and Back Orifice.

CDC - Control Data Corporation


Atlanta, GA 30341

adolfas.gaigalas@nist.gov

About the authors: A. K. Gaigalas is a physicist and Lili Wang a research chemist both in the Biotechnology Division of the NIST Chemical Science and Technology Laboratory. Abe Schwartz is a chemist at the Center for Quantitative Cytometry. Gerald E. Marti is a review and research officer at CBER CB·er  
n.
One that uses a CB radio.
, FDA and an attending physician at NCI See Liberate. . Robert F. Vogt is a chemist at the Center of Disease Control and Prevention, Atlanta, GA. The National Institute of Standards and Technology is an agency of the Technology Administration, U.S. Department of Commerce.
COPYRIGHT 2005 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2005, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:molecules of equivalent soluble fluorophore
Author:Vogt, Robert F., Jr.
Publication:Journal of Research of the National Institute of Standards and Technology
Geographic Code:1USA
Date:Mar 1, 2005
Words:8199
Previous Article:A summary of heat-flux sensor calibration data.(research of heat-flux sensors)
Next Article:Ba(OH)[.sub.2] equilibria in the system Ba-O-H-F, with application to the formation of [Ba.sub.2]Y[Cu.sub.3][O.sub.6.5 + x] from...
Topics:



Related Articles
The shape of more colors to come.(researchers mutate green florescent protein for use as a biological marker)(Brief Article)
TNT-sensing plastic exposes land mines.(Brief Article)
Detection scheme takes lesson from plants.(detecting molecules)(Brief Article)
The Development of Fluorescence Intensity Standards.
Temperature measurement in microfluidic devices. (News Briefs).(Brief Article)
Quantitating fluorescence intensity from fluorophore: the definition of MESF assignment.
Quantitating fluorescence intensity from fluorophores: practical use of MESF values.
Nanolights! Camera! Action! Tiny semiconductor crystals reveal cellular activity like never before.(quantum dots)
Detect spoilage with indicator.
Filters for fluorescence microscopy.(New Products)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles