MicroSurgical Tech slapped with 483 for quality system, corrective/preventive actions.


A six-item 483 went to MicroSurgical Technologies, Redmond, WA, because quality data was not appropriately analyzed an·a·lyze  
tr.v. an·a·lyzed, an·a·lyz·ing, an·a·lyz·es
1. To examine methodically by separating into parts and studying their interrelations.

2. Chemistry To make a chemical analysis of.

3.
, and CAPA CAPA California Alternate Performance Assessment
CAPA Captaris, Inc (stock symbol)
CAPA Confederation of Asian and Pacific Accountants
CAPA Creative and Performing Arts (school) 
 procedures were not adequate. The audit was conducted from the Bothell, WA, District Office by investigator Deborah Nebenzahl.

FDA FDA
abbr.
Food and Drug Administration


FDA,
n.pr See Food and Drug Administration.

FDA,
n.pr the abbreviation for the Food and Drug Administration.
 documents faulted MicroSurgical because appropriate sources of quality data were not adequately analyzed to identify existing and potential causes of nonconforming product and other quality problems.

The firm noted its belief that its quality system "has been designed and constructed to ensure that sources of quality data are monitored and analyzed. MicroSurgical systematically collects data on all sources of quality data. The methods for coping with these types of exceptions includes immediate correction, risk assessment and escalation es·ca·late  
v. es·ca·lat·ed, es·ca·lat·ing, es·ca·lates

v.tr.
To increase, enlarge, or intensify: escalated the hostilities in the Persian Gulf.

v.intr.
 criteria."

Regarding comments on internal audits, the company noted the audit procedures at the firm until mid-2005 included a corrective action A corrective action is a change implemented to address a weakness identified in a management system. Normally corrective actions are instigated in response to a customer complaint, abnormal levels if internal nonconformity, nonconformities identified during an internal audit or  pathway that was separate from the general CAPA flow. Audit findings and the associated corrective/preventative actions from the period through mid-2005 were not visible within the CAPA historical log.

Another repeat violation was the following: "The firm does not have a mechanism for identifying existing or potential causes of quality problems through their internal quality audits."

Not all actions needed to correct and prevent the recurrence recurrence /re·cur·rence/ (-ker´ens) the return of symptoms after a remission.recur´rent

re·cur·rence
n.
1.
 of nonconforming product and other quality problems were identified, the 483 added. For example, the investigation conducted as part of CAPA request No. 392 did not include documented analysis of the scope of the issue to include whether multiple components, devices or lots were affected by the contamination found.

Also, the investigation conducted as part of CAPA 425 did not include documented analysis of the scope of the issue to include whether multiple components, devices or lots were affected by the failure of the EDM (Engineering Data Management) An information system that maintains the details of all engineering data while the product is in the design and concept phase. This includes geometry and changes to geometry. See PLM.

EDM - Electronic Data Management
 process.

In addition, the investigation conducted as part of CAPA 448 did not include documented analysis of the scope of the issue to include whether multiple components, devices or lots were affected by the expired sterilization sterilization

Any surgical procedure intended to end fertility permanently (see contraception). Such operations remove or interrupt the anatomical pathways through which the cells involved in fertilization travel (see reproductive system).
 certificate.

CAPA procedures also were inadequate, the report stated, because CAPA activities were not documented, including investigations of causes of nonconformities, the actions needed to correct or prevent recurrence of nonconforming product and other quality problems, the verification or validation of corrective actions and implementation of corrective and preventive actions Corrective and Preventive Action (CAPA) is a concept within Good Manufacturing Practice (GMP). CAPA focuses on the systematic investigation of discrepancies (failures and/or deviations) in an attempt to prevent their reoccurrence. .

Specifically, several CAPAs did not include documentation of corrective cor·rec·tive
adj.
Counteracting or modifying what is malfunctioning, undesirable, or injurious.

n.
An agent that corrects.


corrective,
n
 and preventative activities, including investigation of causes of nonconformities, actions needed to correct or prevent recurrence, implementation and verification or validation of corrective and preventative actions, as well as risk analysis.

For example, the following CAPAs lacked documentation of CAPA activities:

* CAPA 425--Lacked documentation including investigation of nonconformity non·con·form·i·ty  
n. pl. non·con·form·i·ties
1.
a. Refusal or failure to conform to accepted standards, conventions, rules, or laws.

b.
, the scope of actions needed to correct or prevent recurrence, as well as documentation for implementation, and verification or validation of corrective and preventative actions, and risk analysis that could have affected indications for use or safety.

* CAPA 285--Lacked documentation for implementation and verification or validation of CAPA actions, as well as risk analysis that could have affected indications for use or safety; and

The firm replied: "CAPA 285 is an open CAPA and for that reason, does not contain documentation for implementation and verification or validation of corrective and preventative actions. The issue is the lack of a formal method for providing QA reference samples for use in inspection. An informal method has been successfully used to provide reference samples, but the method has not been documented, keeping the CAPA open."

The firm stated that CAPA 425 was triggered as a result of a rejection of seven parts in a manufacturing batch for a single defect attribute. The suspect parts were detected and scrapped. "The defect in question is highly detectable, and other lots were not implicated im·pli·cate  
tr.v. im·pli·cat·ed, im·pli·cat·ing, im·pli·cates
1. To involve or connect intimately or incriminatingly: evidence that implicates others in the plot.

2.
 by these defects. The process involved in producing the attribute in question is part specific, thus no other components or devices were implicated," the letter noted.

The firm was not available for comment. MicroSurgical Technologies, Redmond, WA, 11/8-21/05, Doc. 109920M, $10 plus retrieval.
COPYRIGHT 2007 Washington Information Source, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2007 Gale, Cengage Learning. All rights reserved.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:Medical devices
Publication:Validation Times
Date:Mar 1, 2007
Words:653
Previous Article:Validation process procedures nets 11-item 483 for Electrode Store.(Medical devices)
Next Article:Actavis Totowa nets warning for QC, cleaning validation.(Human drugs)



Related Articles
Perry letter may have been result of lost paper-work, firm believes. (Medical Devices).
Medical devices.
Umbilical cord clamp manufacturer fails to implement corrective action.(Medical devices)
Satelec Acteon Group hit with 483 and warning letter for process validation, QA deficiencies.(Medical devices)
NeuroMetrix receives 2-item 483 for flaws in CAPA systems.(Medical devices)
DMRs, CAPAs and MDRs highlight 14-item 483 for Ram Precision.(MEDICAL DEVICES)
CAPAs, cleaning, calibration cited in 14-item 483 for Ram Precision.(Medical devices)
Repackager slapped with 10-item 483 for no design and development plan.(MEDICAL DEVICES)
Repeated complaint handling violation results in 6-item 483 for MicroSurgical Tech.(MEDICAL DEVICES)
Berkline cited for quality policy flaws in 12-item FDA write-up.(MEDICAL DEVICES)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles