Medtronic Announces a Nationwide, Voluntarily Recall of Small Subset of Two Implantable Cardioverter-Defibrillator Models.


Business Editors/Health/Medical Writers

MINNEAPOLIS--(BUSINESS WIRE)--April 16, 2004

Medtronic, Inc. today announced that it is voluntarily recalling two older models of implantable cardioverter-defibrillators (ICDs). The Class I recall involves a small subset of Micro Jewel(R) II Model 7223Cx and GEM(R) DR Model 7271 ICDs that may take a longer than normal time to charge before delivering therapy. Most of these devices, implanted in 1997 and 1998, are close to the normal replacement time. No other Medtronic devices are involved in this action.

The Micro Jewel II device is no longer sold, and the GEM DR device has limited distribution. Physicians are being notified that they should verify the charge time and battery voltage of each affected device. If any devices exhibit unsatisfactory charge times, Medtronic recommends scheduling ICD ICD International Classification of Diseases (of the World Health Organization); intrauterine contraceptive device.

ICD
abbr.
 replacement. The final decision whether to replace the device is based on the physician's medical judgment and specific patient needs.

Implantable cardioverter-defibrillators are medical devices used to shock the heart into normal rhythm after patients suffer ventricular tachycardia Ventricular Tachycardia Definition

Ventricular tachycardia (V-tach) is a rapid heart beat that originates in one of the lower chambers (the ventricles) of the heart.
 or fibrillation fibrillation /fi·bril·la·tion/ (fi?bri-la´shun)
1. the quality of being made up of fibrils.

2. a small, local, involuntary, muscular contraction, due to spontaneous activation of single muscle cells or muscle
, which are rapid, life-threatening arrhythmias originating in the lower chambers of the heart. The devices are surgically implanted in the chest in a minor procedure lasting less than one hour.

When a cardiac arrhythmia cardiac arrhythmia
n.
See cardiac dysrhythmia.


Cardiac arrhythmia
An irregular heart rate or rhythm.

Mentioned in: Holter Monitoring, Stress Test

cardiac arrhythmia 
 requiring cardioversion Cardioversion Definition

Cardioversion refers to the process of restoring the heart's normal rhythm by applying a controlled electric shock to the exterior of the chest.
 or defibrillation Defibrillation Definition

Defibrillation is a process in which an electronic device sends an electric shock to the heart to stop an extremely rapid, irregular heartbeat, and restore the normal heart rhythm.
 shock occurs, the capacitor is normally charged, and the device delivers the appropriate shock. With the cardioverter-defibrillators in this recall, the capacitors may take longer than normal to charge near the end of the battery service life and could cause a delay in delivery or a non-delivery of shock therapy. Such a delay or non-delivery could result in patient injury or death because patients are not receiving the appropriate therapy in time. Medtronic no longer uses this capacitor technology.

Recently, Medtronic became aware of one serious injury and four deaths that may be related to the failure of the capacitor in a small subset of Micro Jewel II devices.

A total of 6,268 of the affected ICDs were manufactured, of which about 1,800 are thought to be still implanted in patients worldwide. Medtronic is working with the U.S. Food and Drug Administration (FDA FDA
abbr.
Food and Drug Administration


FDA,
n.pr See Food and Drug Administration.

FDA,
n.pr the abbreviation for the Food and Drug Administration.
) to ensure that all physicians and their patients with these devices are notified of the issue. Medtronic began notifying physicians on April 5.

The FDA defines a Class I recall as a situation in which there is a reasonable probability that the use of the product will cause serious adverse health consequences or death.

Medtronic, Inc., headquartered in Minneapolis, is the world's leading medical technology company, providing lifelong solutions for people with chronic disease. Its Internet address There are two kinds of addresses that are widely used on the Internet. One is a person's e-mail address, and the other is the address of a Web site, which is known as a URL. Following is an explanation of Internet e-mail addresses only. For more on URLs, see URL and Internet domain name.  is www.medtronic.com.

Any statements made about the company's anticipated financial results and regulatory approvals are forward-looking statements subject to risks and uncertainties such as those described in Medtronic's Annual Report on Form 10-K Form 10-K

A report required by the SEC from exchange-listed companies that provides for annual disclosure of certain financial information.


Form 10-K

See 10-K.
 for the year ended April 25, 2003. Actual results may differ materially from anticipated results.
COPYRIGHT 2004 Business Wire
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2004, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Publication:Business Wire
Geographic Code:1USA
Date:Apr 16, 2004
Words:484
Previous Article:Provo Names Ventura Martinez Del Rio Sr., Chief Executive Officer.
Next Article:Bridge Bancorp, Inc. Reports First Quarter 2004 Earnings.
Topics:



Related Articles
Medtronic Announces First Implant of InSync Marquis System, the Company's Fourth Generation Heart Failure Device.
Medtronic Announces FDA Approval of InSync Marquis Cardiac Resynchronization Therapy System.
More People with Medical Devices Now Can Be Cared for from Home with Innovative Medtronic Internet-Based Technology.
Medtronic Announces First Implant of Sprint Fidelis Lead, Smallest Defibrillation Lead in the Industry.
Medtronic CFO Announces Preliminary Revenues for Third Quarter of FY2004.
U.S. FDA Approves New Devices for Use with Medtronic's Internet-Based Heart Device Monitoring System.
Company takes its market to heart.(Up Front)
New York Stock Exchange Delays Opening Trading of ICD Manufacturers' Stocks Pending Presentation of Results of Landmark Trial.
Nihon Kohden Introduces Two New Biphasic Defibrillator Series.
New litigation packet aids members pursuing heart device claims.

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles