MDR for improved fluoroelastomer cure testing.


Minnesota Mining and Manufacturing (3M0 selected Monsanto's MDR MDR,
n See multidrug resistance.

MDR,
n the abbreviation for minimum daily requirement, specifically the Minimum Daily Requirements for Specific Nutrients compiled by the United States Food and Drug Administration.
2000 moving die rheometer rhe·om·e·ter
n.
An instrument for measuring the flow of viscous liquids, such as blood.
 for cure testing of Fluorel fluoroelastomers. The selection of the MDR2000 resulted from its reduced testing variability and improved discrimination of compound differences in fluoroelastomers. The improved performance and productivity of this rheometer have justified the replacement of existing rheometers with the MDR2000.

3M evaluated an MDR2000 as compared to Monsanto's R100S type oscillating os·cil·late  
intr.v. os·cil·lat·ed, os·cil·lat·ing, os·cil·lates
1. To swing back and forth with a steady, uninterrupted rhythm.

2.
 disc rheometers and an ODR ODR Online Dispute Resolution
ODR On-Demand Routing
ODR One-Definition Rule (C++)
ODR Octal Data Rate (high speed memory interface transfers 8 bits of data per clock cycle)
ODR Office of Dispute Resolution
2000. The MDR2000 related twice as high as either the R100S or ODR2000 for discriminating compound differences. The variability of the MDR2000 was typically 2-4 times less than the other rheometers.

Two studies in 1991 compared the instruments first at the 3M specialty fluoropolymers laboratory and then at two additional 3M locations. All locations had access to each instrument type, so that location effects could be separated from machine effects.

Variations of a Fluorel compound were made with 10% change in Curative curative /cur·a·tive/ (kur´ah-tiv) tending to overcome disease and promote recovery.

cu·ra·tive
adj.
1. Serving or tending to cure.

2.
 A and Curative B. The experiment used a factorial factorial

For any whole number, the product of all the counting numbers up to and including itself. It is indicated with an exclamation point: 4! (read “four factorial”) is 1 × 2 × 3 × 4 = 24.
 design with the standard Fluorel FC-2174 gum, 30 parts of N990 carbon black, 3 parts of MgO and 6 parts of Ca(OH)2.

Starting with standard FC-2174 fluoroelastomer gum, levels of Curative A and Curative B were varied in a two level factorial design to help determine the discriminating power of each Machine-Arc combination. Table 1 shows the level number associated with the Curative A, Curative B combinations.
Table 1 - curative levels
Level  Curative A   Curative B
  1       -10%          10%
  2         0%           0%
  3        10%          10%
  4       -10%         -10%
  5        10%         -10%


Each instrument was calibrated cal·i·brate  
tr.v. cal·i·brat·ed, cal·i·brat·ing, cal·i·brates
1. To check, adjust, or determine by comparison with a standard (the graduations of a quantitative measuring instrument):
 prior to testing. All rheometer tests were run at 177 [degrees] C for 12 minutes using one mil polyester film. The test sample sizes were eight grams for the MDR2000 and 15 grams for the R100S and ODR2000.

The measured rheometer properties for FKM FKM Fluoroelastomer
FKM Fogarty Klein Monroe (Houston, Texas)
FKM Field Kitchen, Modular
 were minimum torque, maximum torque, ts2, t'50 and t'90. Tables 2 through 6 show the discriminating power (sensitivity) for each property. The variation for each machine was determined by pooling each location into one overall estimate. Evidence of unequal variation among locations for the same machine are noted at the bottom of each table.
Table 2 - minimum torque discrimination table
Machine               Average     Pooled    Ratio
                        range    std. dev.
MDR-2000, arc - 0.5  0.148333      0.010  14.833
ODR-2000, arc = 1.0   0.74357   0.073416  10.128(*)
ODR-100S, arc = 3.0   1.63333     0.1253  13.035(*)
(*) - Evidence of outliers (problem points) (location 2)
+ - Evidence of unequal variance between locations
    (location 2)
Table 3 - maximum torque discrimination table
Machine             Average      Pooled     Ratio
                      range     std. dev.
MDR-2000, arc = 0.5   3.875     0.1253     30.926
ODR-2000, arc = 1.0 7.38167   0.406202    18.172+
ODR-100S, arc = 3.0   15.75   0.806226    19.535+
+ - Evidence of unequal variance between locations
    (location 2, location 3)
Table 4 - ts2 discrimination table
Machine               Average      Pooled     Ratio
                        range     std. dev.
MDR-2000, arc = 0.5   0.31086   0.0106771    29.115
ODR-2000, arc = 1.0   0.45666      0.0402   11.36(++)
ODR-100S, arc = 3.0  0.466667      0.0683    6.8326
++ - Marginal evidence of unequal variance between
    locations (location 2)
Table 5 - t'50 discrimination table
Machine               Average       Pooled      Ratio
                        range    std. dev.
MDR-2000, arc = 0.5   0.38741   0.00844985      45.85
ODR-2000, arc = 1.0  0.608333     0.053386   11.395(++)
ODR-100S, arc = 3.0     0.650     0.065803      9.878
++ - Marginal evidence of unequal variance between
    locations (location 2)
Table 5 - t'90 discrimination table
Machine              Average   Pooled       Ratio
                      range    std. dev.
MDR-2000, arc = 0.5  0.57293   0.0211    27.153(+)
ODR-2000, arc = 1.0  0.65333   0.0543   12.032(++)
ODR-100S, arc = 3.0  0.66667   0.0585     11.396
+ - Evidence of unequal variance between locations
    (location 2)
++ - Marginal evidence of unequal variance between
    locations (location 2)


[Tabular Data Omitted]

The range of values, especially for minimum and maximum torques tor·ques  
n. Zoology
A band of feathers, hair, or coloration around the neck.



[Latin torqu
, change with instrument type. To compare instrument types for the ability to discriminate compound differences, the range for each type needs to be considered.

Taking a ratio of the range across levels (maximum minus minimum) divided by the testing pooled standard deviation Pooled standard deviation is a way to find a better estimate of the true standard deviation given several different samples taken in different circumstances where the mean may vary between samples but the true standard deviation (precision) is assumed to remain the same.  is a good way to measure the ability to discriminate. This ratio gives the number of test standard deviations In statistics, the average amount a number varies from the average number in a series of numbers.

(statistics) standard deviation - (SD) A measure of the range of values in a set of numbers.
 that fit between the maximum and minimum response levels for each change in curative level.

Table 7 (discriminate grading grid) grades each instrument type by rheometer data point. For each data point, the best discrimination (largest ration ration

a fixed allowance of total feed for an animal for one day. Usually specifies the individual ingredients and their amounts and the amounts of the specific nutriments such as carbohydrate, fiber, individual minerals and vitamins.
 number) is a grade of 3, and the next best a 2. The sum across data point grades gives a "Sum" for each instrument and oscillation Oscillation

Any effect that varies in a back-and-forth or reciprocating manner. Examples of oscillation include the variations of pressure in a sound wave and the fluctuations in a mathematical function whose value repeatedly alternates above and below some
 arc. This table shows the MDR2000 better discriminates between the cure systems and had less variability than the R100S or ODR2000 rheometers.

The method of ranking gives equal weight to each property and not accounting for ratios that are close seconds within a property. Asterisks were placed where potential outliers existed. One could also adjust the values base on potential outliers, asterisk (1) See Asterisk PBX.

(2) In programming, the asterisk or "star" symbol (*) means multiplication. For example, 10 * 7 means 10 multiplied by 7. The * is also a key on computer keypads for entering expressions using multiplication.
 values. Also no potential outliers were indicated with any of the MDR or ODR2000 series.
COPYRIGHT 1994 Lippincott & Peto, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 1994, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Title Annotation:MDR2000 moving die rheometer from Monsanto Co.
Author:Maier, Shirley J.
Publication:Rubber World
Date:Aug 1, 1994
Words:841
Previous Article:Record shipments predicted for OE, replacement tires. (original equipment) (based on data from the Tire Market Analysis Committee of the Rubber...
Next Article:Microencapsulation of antidegradants.
Topics:



Related Articles
Reinforcement with fluorplastic additives.
Monsanto: rubber chemicals, instruments and equipment. (company profile)
Monsanto: rubber chemicals, instruments and equipment. (Chemicals & Materials Corporate Profile '93: Special Advertising Section) (Company Profile)
Optimal measurement, use of dynamic properties from the MDR for compound analysis. (moving die rheometer)
A highly efficient cure system for BIMS.(isobutylene-co-p-bromomethylstyrene elastomers)
Instrumentation, test equipment suppliers.(Directory)
TiBTM: Some further applications of this unique retarder/kicker.(rubber industry research)
PRODUCT NEWS.(Brief Article)
Non-postcure fluoroelastomers.
Xanthate accelerators in vulcanizing isobutylene-based elastomers.

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles