Printer Friendly
The Free Library
14,709,671 articles and books
Member login
User name  
Password 
 
Join us Forgot password?

Limitations to accuracy in extracting characteristic line intensities from x-ray spectra.


The early development of quantitative electron probe microanalysis microanalysis /mi·cro·anal·y·sis/ (-ah-nal´i-sis) the chemical analysis of minute quantities of material.

microanalysis

the chemical analysis of minute quantities of material.
, first using crystal spectrometers, then energy dispersive dispersive /dis·per·sive/ (-per´siv)
1. tending to become dispersed.

2. promoting dispersion.
 x-ray spectrometers x-ray spectrometer
n.
A spectrometer using x-rays to separate the chemical constituents of a substance into their characteristic spectral lines for identification and determination of their concentration.
 (EDXS EDXS Energy-Dispersive X-ray Spectroscopy ), demonstrated that elements could be detected at 0.001 mass fraction level and major concentrations measured within 2 % relative uncertainty. However, during this period of extensive investigation and evaluation, EDXS detectors were not able to detect x rays below 1 keV and all quantitative analysis Quantitative Analysis

A security analysis that uses financial information derived from company annual reports and income statements to evaluate an investment decision.

Notes:
 was performed using a Set of reference standards measured on the instrument. Now that EDXS systems are often used without standards and are increasingly being used to analyse an·a·lyse  
v. Chiefly British
Variant of analyze.


analyse or US -lyze
Verb

[-lysing, -lysed] or -lyzing,
 elements using lines well below I keV, accuracy can be considerably worse than is documented in standard textbooks. Spectrum processing techniques found most applicable to EDXS have now been integrated into total system solutions and can give excellent results on selected samples. However, the same techniques fail in some applications because of a variety of instrumental effects. Prediction o f peak shape, width and position for every characteristic line and measurement of background intensity is complicated by variations in response from system to system and with changing count rate. However, with an understanding of the fundamental sources of error, even a total system can be tested like a "black box" in areas where it is most likely to fail and thus establish the degree of confidence that should apply in the intended application. This approach is particularly important when the microanalysis technique is applied at lower electron beam A stream of electrons, or electricity, that is directed towards a receiving object. See electron beam imaging and electron beam lithography.  voltages where the extraction of line intensities is complicated by extreme peak overlap and higher background levels.

Key words: EDX EDX Energy Dispersive X-Ray (Spectroscopy)
EDX Electronic Data Exchange
EDX Extended Data Register
EDX Event-Driven Executive (IBM Series/1 OS)
EDX Event-Based Data Exchange (UPNet) 
; EDXS; energy dispersive; least squares fitting; microanalysis; spectrum processing; x rays.

1. Introduction

The technique of x-ray microanalysis relies on the fact that if a flat bulk sample and a standard are exposed to the same beam current and in the same geometry relative to an x-ray detector, then

[C.sub.i]/[C.sup.std.sub.i] = XRCF XRCF X-Ray Calibration Facility (NASA)  * [I.sub.i]/[I.sup.std.sub.i] (1)

where [C.sub.i] and [C.sup.std.sub.i] are the mass concentrations of element in sample and standard respectively, [I.sub.i] and [I.sup.std.sub.i] are the detected intensities of the characteristic line for element in sample and standard and XRCF is the x-ray correction factor that accommodates the differences in x-ray generation, absorption, and fluorescence fluorescence (flrĕs`əns), luminescence in which light of a visible color is emitted from a substance under stimulation or excitation by light or other forms of electromagnetic  enhancement caused by the different overall compositions of sample and standard. In the two decades following Castaing's 1951 thesis (1), x-ray intensities were invariably in·var·i·a·ble  
adj.
Not changing or subject to change; constant.



in·vari·a·bil
 measured with high resolution Bragg crystal "wavelength dispersive" spectrometers (WDS Wds Words
WDS Wireless Distribution System (Joint Common Database)
WDS Wide-area Data Services
WDS Wireless Domain Services (Cisco Systems technology)
WDS Wavelength Dispersive Spectroscopy
) so it was straightforward to measure [I.sub.i] In this period, XRCF calculations were extensively investigated and reported at international conferences. Consequently, relative uncertainties of 2 % could be expected for x-ray microanalysis of bulk materials in a dedicated WDS electron probe instrument (2).

When energy dispersive x-ray spectrometry x-ray spectrometry
n.
The use of an x-ray spectrometer, especially for chemical analysis of a substance.
 (EDXS) systems were first introduced, the inferior instrumental resolution was a barrier to accurate measurement of line intensities and EDXS was only used as a rough qualitative tool. However, in 1973 Reed and Ware (3) demonstrated that quantitative analysis of silicate minerals The silicate minerals make up the largest and most important class of rock-forming minerals. They are classified based on the structure of their silicate ion group.

Subclasses: Nesosilicates or Isosilicates
Nesosilicates (or orthosilicates
 for elements with atomic number atomic number, often represented by the symbol Z, the number of protons in the nucleus of an atom, as well as the number of electrons in the neutral atom. Atoms with the same atomic number make up a chemical element.  11 and above could be achieved by EDXS with limit of detection around 0.001 mass fraction and with an accuracy equivalent to WDS. While other authors subsequently corroborated cor·rob·o·rate  
tr.v. cor·rob·o·rat·ed, cor·rob·o·rat·ing, cor·rob·o·rates
To strengthen or support with other evidence; make more certain. See Synonyms at confirm.
 these accuracy claims (e.g. (4)), a survey of EDXS accuracy on a variety of instruments [5] showed that relative standard uncertainties of 6 % could be expected for major constituents and much higher uncertainties could be obtained for concentrations below 0.2 mass fraction. This early report suggested that the source of error was primarily deconvolution In mathematics, deconvolution is an algorithm-based process used to reverse the effects of convolution on recorded data.[1] The concept of deconvolution is widely used in the techniques of signal processing and image processing.  of overlapping peaks and background correction.

Spectrum processing and EDXS instrumentation in general have undoubtedly improved since that survey, particularly in terms of convenience of use and operator deskilling Deskilling is the process by which skilled labor within an industry or economy is eliminated by the introduction of technologies operated by semiskilled or unskilled workers. . "Standardless" analysis has become popular because the requirement to maintain a collection of analytical standards for intensity comparison is a major overhead. Although some systems provide an option to use a single reference standard so that an analytical total can be obtained, most standardless procedures normalise Verb 1. normalise - become normal or return to its normal state; "Let us hope that relations with this country will normalize soon"
normalize

change - undergo a change; become different in essence; losing one's or its original nature; "She changed completely
 the results. Although normalisation 1. (data processing) normalisation - A transformation applied uniformly to each element in a set of data so that the set has some specific statistical property. For example, monthly measurements of the rainfall in London might be normalised by dividing each one by the total  conveniently overcomes problems of beam current fluctuation Fluctuation

A price or interest rate change.
, analytical errors are concealed beneath a total that is always unity. A recent report reviews the performance of some standardless EDXS solutions showing error histograms for analysis of a set of known materials (6). While the commercial systems with fitted standards achieved relative standard deviations In probability theory and statistics, the Relative Standard Deviation (RSD or %RSD) refers to the absolute value of the coefficient of variation expressed as a percentage.

It is widely used in analytical chemistry to express the precision of an assay.

l
 of around 12 %, a first principles approach gave a relative standard deviation of about 25 %. In the samples used, the XRCF's wer e well known, all concentrations were above 0.05 mass fraction and most were above 0.20 mass fraction so the uncertainty due to background subtraction subtraction, fundamental operation of arithmetic; the inverse of addition. If a and b are real numbers (see number), then the number ab is that number (called the difference) which when added to b (the subtractor) equals  was not an issue. Standardless analysis requires more accurate spectrum processing and detector modelling because it does not benefit from the cancelling of first order error terms and implicit instrument calibration calibration /cal·i·bra·tion/ (kal?i-bra´shun) determination of the accuracy of an instrument, usually by measurement of its variation from a standard, to ascertain necessary correction factors.  that occurs when ratioing measured intensities from sample and standard as in Eq. (1).

In the early work on quantitative EDXS and in the above assessment of standardless accuracy (6), the lowest energy line used was close to 1 keV. Most EDXS systems are now capable of detecting x rays well below 1 keV and the advantage of reduced analytical volume provides an increasing attraction to work at low beam low beam
n.
The beam of a vehicle's headlight that provides short-range illumination.

Noun 1. low beam - the beam of a car's headlights that provides illumination for a short distance
 voltages (7). However, the poorer line energy separation, increased incidence of peak overlap, and higher back- ground provide severe challenges for spectrum processing at these low energies. The low energy region also suffers electronic artifacts artifacts

see specimen artifacts.
, incomplete charge collection (ICC ICC

See: International Chamber of Commerce
) distortions of peak shape and large absorption edges in the background.

The successful early usage in mineralogical min·er·al·o·gy  
n. pl. min·er·al·o·gies
1. The study of minerals, including their distribution, identification, and properties.

2. A book or treatise on mineralogy.
 analysis, the sophisticated appearance of modern software, and the perceived security of a unity total provided by normalisation, have led to the expectation that all EDXS systems are capable of accurate quantitative analysis. However, with the increased application of standardless analysis, "out-of-the-box" with no customised installation, results are vulnerable to error, particularly when microanalysis is used outside the well-investigated territory of 1 keV to 10 keV energy and 15 kV to 25 kV accelerating voltage. This paper will demonstrate fundamental sources of error, how to avoid or overcome them and some approaches to validate overall spectrum processing performance for a "black box" system where the details of implementation are unknown to the operator.

2. Background Correction

2.1 Required Accuracy

Spectrum processing requirements are illustrated by Fig. 1. At 20 kV, the K lines of Ti, Cr, Mn, Fe, Ni are excited and clearly visible in the spectrum between 4 eV and 9 keV. The background is fairly flat and there is some K[beta]/K[alpha] overlap to contend with, but most algorithms will have no difficulty in revealing K[alpha] line intensities. However, Fig. 2 shows that when the same sample is excited at 5 kV, the spectrum processing task becomes formidable. The L line series for Ti, Cr, Mn, Fe, Ni all overlap, and the background bulges beneath the peaks so that a simple linear interpolation Linear interpolation is a method of curve fitting using linear polynomials. It is heavily employed in mathematics (particularly numerical analysis), and numerous applications including computer graphics. It is a simple form of interpolation.  would give large percentage errors in composition.

The accuracy required in background correction can be understood with reference to Fig. 3. Whereas the bremsstrahlung bremsstrahlung (brĕm`shträ'ləng): see X ray.
bremsstrahlung

(German; “braking radiation”)
 background forms a smooth continuum with count rate per unit energy interval that is practically independent of detector resolution, the intensity within a single characteristic emission line is smeared smear  
v. smeared, smear·ing, smears

v.tr.
1.
a. To spread or daub with a sticky, greasy, or dirty substance.

b.
 into a peak with full width at half maximum A full width at half maximum (FWHM) is an expression of the extent of a function, given by the difference between the two extreme values of the independent variable at which the dependent variable is equal to half of its maximum value.  (fwhm) determined by detector resolution so that the peak height is inversely proportional See Directly proportional, under Directly, and Inversion, 4.

See also: Inversely
 to resolution [see Eq. (7)]. The spectral response The variable output of a light-sensitive device that is based on the color of the light it perceives.  in Fig. 3 has been calculated for a typical Si(Li) detector with fwhm = 133 eV at 5.9 keV. Samples with higher atomic number generate proportionately pro·por·tion·ate  
adj.
Being in due proportion; proportional.

tr.v. pro·por·tion·at·ed, pro·por·tion·at·ing, pro·por·tion·ates
To make proportionate.
 more bremsstrahlung background and three examples are shown. Characteristic x-ray intensities depend mainly on element concentration and several K lines from different elements are shown at the 1 % level for reference. If the sample has a light matrix like C then the total background at any energy is much less than 1. % of the K[alpha] peak height for a pure eleme nt. If the sample has a heavier matrix like Fe, then the background level represents about 0.01 mass fraction concentration and for samples with particularly heavy matrices like Au, the background can be equivalent to a mass fraction of several percent at higher energies. Therefore, to detect elements at the 0.001 mass fraction level using K lines, the estimate of background will usually have to be accurate to well within 10 %. If other lines are used, any error in the background correction has a greater effect because the peak height for L or M lines is less than for K at the same concentration, as shown in Fig. 4. At low kV, background correction has to be more accurate because the background represents a much higher mass fraction as can be seen by comparing Fig. 3 with Fig. 5. At 5 kV, for a sample with matrix atomic number of about 26, the background height at 2.6 keV is equivalent to about 0.03 mass fraction of CIK CIK Central Index Key (SEC)
CIK Commission Internationale de Karting (French)
CIK Crypto Ignition Key
CIK Contribution in Kind
CIK Confederazione Italiana Kendo
[alpha] and this is five times greater than at 20 kV. Moreover, fewer K lines are excited a t low kV and lower intensity L lines have to be used. Now the background height from an Fe matrix is equivalent to almost 0.10 mass fraction concentration for FeL. At 5 kV beam voltage, a background correction accurate to within 1 % is therefore required to detect elements at the 0.00 1 mass fraction level using L lines. The bremsstrahlung shape for Fe in Fig. 5 also demonstrates the type of large step over the LIII absorption edge that makes background prediction particularly difficult at low energies.

2.2 Interpolated interpolated /in·ter·po·lat·ed/ (in-ter´po-la?ted) inserted between other elements or parts.  Background

Interpolation interpolation

In mathematics, estimation of a value between two known data points. A simple example is calculating the mean (see mean, median, and mode) of two population counts made 10 years apart to estimate the population in the fifth year.
 and extrapolation (mathematics, algorithm) extrapolation - A mathematical procedure which estimates values of a function for certain desired inputs given values for known inputs.

If the desired input is outside the range of the known values this is called extrapolation, if it is inside then
 techniques are described in detail in Ref. (8). At higher energies where the background is fairly linear, subtracting an interpolated background is a reliable method to determine the area of an isolated peak. Figure 6 shows one approach that simplifies the equations. If we take the sum of (2M + 1) channels straddling strad·dle  
v. strad·dled, strad·dling, strad·dles

v.tr.
1.
a. To stand or sit with a leg on each side of; bestride: straddle a horse.

b.
 the peak and define two background regions of N channels exactly the same distance away from the peak centre, then the formula for the net window integral, P, is:

P = S-(B1 + B2) * (2M + 1)/(2N). (2)

The standard deviation In statistics, the average amount a number varies from the average number in a series of numbers.

(statistics) standard deviation - (SD) A measure of the range of values in a set of numbers.
, [[sigma].sub.p] is given by

[[sigma].sub.P.sup.2] = S + (B1 + B2) * [[(2M + 1)/(2N)].sup.2]. (3)

and this is minimised by choosing a large number of background channels N. In practice it is not easy to extend N indefinitely because of nearby peaks. In Fig. 6 the background window clearly needs to be far enough away from the nearby FeK[beta] peak just above 7 keV. The net integral P is proportional to peak area but the proportionality constant cancels when making comparisons from sample and standard as in Eq. (1). However, for standardless analysis, the proportionality constant has to be determined and will be altered if detector resolution changes for any reason.

2.3 Digital Filtering

Rather than measuring the net window integral, the average peak intensity over (2M + 1) channels can be used and the average background subtracted thus:

P' = S/(2M+ 1)- (B1 + B2)/(2N). (4)

This can be regarded as a weighted sum of channel counts where the channels over the peak are each multiplied by l/(2M + 1) and the background channels multiplied by--1/(2N) before summing. If the peak channels touch the background channels, then a graph showing the positive and negative weighting coefficients has a "Top-Hat" shape (Fig. 7). This "Top-Hat" can he positioned anywhere in the spectrum to calculate the net peak intensity at that position. If this is done for every channel position in the spectrum, the net peak intensity at each channel is effectively the output of a digital convolution convolution /con·vo·lu·tion/ (-loo´shun) a tortuous irregularity or elevation caused by the infolding of a structure upon itself.  filter. if 2M + 1 is chosen to be close to the fwhm of a peak and 2N covers roughly the same number of channels, then the filter will remove any background that is linear over the range of the top hat and will emphasise peaks. Figure 8 shows the result of running such a filter along the biggest of the background curves shown in Fig. 3. The K peaks at 0.01 mass fraction concentration have also been filtered the same way and it is clear that the residual background after filtering is equivalent to much less than 0.001 mass fraction every-where in the spectrum except in the vicinity of the AuM absorption edge near 2 keV and the AuN edges below 0.5 keV. The filtering operation fil·ter·ing operation
n.
The surgical creation of a fistula between the anterior chamber of the eye and the subconjunctival space, as for glaucoma.
 converts Gaussian peaks into peaks with negative side lobes In antenna engineering, side lobes are the lobes of the radiation pattern that are not the main beam. The power density in the side lobes is generally much less than that in the main beam.  but the filtered shapes can still be used in a least squares fitting procedure [9,10]. No background points have to be selected, the method is immune to any smooth background artifact A distortion in an image or sound caused by a limitation or malfunction in the hardware or software. Artifacts may or may not be easily detectable. Under intense inspection, one might find artifacts all the time, but a few pixels out of balance or a few milliseconds of abnormal sound  such as backscattered electrons or pile up continuum and no prior knowledge of sample, geometry or microscope parameters are needed in order to make a background correction. The technique therefore has widespread applicability and has been used successfully in commercial EDXS systems for more than 25 years.

2.4 Background Modelling

A drawback DRAWBACK, com. law. An allowance made by the government to merchants on the reexportation of certain imported goods liable to duties, which, in some cases, consists of the whole; in others, of a part of the duties which had been paid upon the importation.  with the digital filtering approach is that it cannot separate high background curvature curvature

Measure of the rate of change of direction of a curved line or surface at any point. In general, it is the reciprocal of the radius of the circle or sphere of best fit to the curve or surface at that point.
 from the curvature of a peak and this is particularly apparent in the vicinity of absorption steps as demonstrated in Fig. 8. Even in this rather extreme case, the measured peak will only suffer a maximum 0.003 mass fraction equivalent error if it is in a particular energy position relative to the AuM absorption edge. Nevertheless, accuracy can sometimes be improved by exploiting more prior knowledge of the background shape.

If the sample is flat, homogeneous The same. Contrast with heterogeneous.

homogeneous - (Or "homogenous") Of uniform nature, similar in kind.

1. In the context of distributed systems, middleware makes heterogeneous systems appear as a homogeneous entity. For example see: interoperable network.
, and semi-infinite, the beam voltage is known and an estimate of composition of the specimen is available, then the theoretical bremsstrahlung background, [B.sub.E], for a single channel at energy E, can be calculated by including terms for sample and detector effects:

[B.sub.E] = [(detector efficiency).sub.E] * [(absorption).sub.E] * [(generation).sub.E]. (5)

After computing computing - computer  [B.sub.E] for every channel in the spectrum, the result is convolved with a Gaussian function In mathematics, a Gaussian function (named after Carl Friedrich Gauss) is a function of the form:



for some real constants a > 0, b, and c.
 where fwhm varies according to according to
prep.
1. As stated or indicated by; on the authority of: according to historians.

2. In keeping with: according to instructions.

3.
 the detector resolution as a function of energy (see Sec. 3.3). Detector efficiency is straightforward to calculate but is a source of error because usually only nominal data nominal data

a type of data in which there are limited categories but no order.
 are available for window thickness and composition, there are manufacturing variations from window to window and there may be additional absorption if a thin layer of oil has condensed con·dense  
v. con·densed, con·dens·ing, con·dens·es

v.tr.
1. To reduce the volume or compass of.

2. To make more concise; abridge or shorten.

3. Physics
a.
 on the window, or ice has built up on the crystal surface. The other major source of error is in the generation term. Various authors have shown that the standard "Kramers" generation term [K.sub.1] * Z * ([E.sub.0] - E)/ E, where [K.sub.1] is a constant, Z is the mean atomic number of the specimen, [E.sub.0] is the energy of incident electrons, and E is the energy of the radiation, does not fit observed spectra very well. Lifshin (11) found that adding an additional quadratic quadratic, mathematical expression of the second degree in one or more unknowns (see polynomial). The general quadratic in one unknown has the form ax2+bx+c, where a, b, and c are constants and x is the variable.  term. [K.sub.2] * Z * [([E.sub.0] - E).sup .2]/E provided a considerably better fit to the distribution as a function of energy. This observation was exploited in the popular "FRAME" public domain software where measurements at two background energies were used to determine [K.sub.1] and [K.sub.2] (12) and this has also been adopted in some commercial EDXS systems. If the "Kramers" term is regarded as the "theory" then including a quadratic term is equivalent to modifying the theory with a multiplier multiplier

In economics, a numerical coefficient showing the effect of a change in one economic variable on another. One macroeconomic multiplier, the autonomous expenditures multiplier, relates the impact of a change in total national investment on the nation's total
 that varies with energy thus:

Modelled background = [B.sub.E] * (a + b * E) (6)

where a and b are constants which force the theoretical [B.sub.E] to match the observed background at two chosen energies. This procedure can then be used to improve the fit for any theoretical bremsstrahlung formulation. Figure 9 shows a theoretical background that has been fitted to a spectrum of pyrope py·rope  
n.
A deep red garnet, Mg3Al2Si3O12, used as a gem.



[Middle English pirope, from Old French, from Latin
 garnet garnet, name applied to a group of isomorphic minerals crystallizing in the cubic system. They are used chiefly as gems and as abrasives (as in garnet paper).  at 2.16 keV and 8.14 keV. The fit between 3 keV and 8 keV is clearly very good and in general the modelling approach works well above 3 keV where absorption effects are slight. At lower energies uncertainty in detector efficiency is a problem and the correct estimation of absorption effects, [(absorption).sub.E] in Eq. (5), requires the composition to be known and this has to be determinated by iteration One repetition of a sequence of instructions or events. For example, in a program loop, one iteration is once through the instructions in the loop. See iterative development.

(programming) iteration - Repetition of a sequence of instructions.
. To do this, major elements must first be correctly identified and if a peak from an element like Si is misidentified as a much heavier element like W or Ta, then the calculated background shape will be incorrect. Another difficulty is that peak-free background points have to be found for fitting, id eally either side of every cluster of peaks. To identify whether a peak is present typically requires a region at least 4 X fwhm in width to do a background subtraction as in Fig. 6. At low energy, such regions are rarely available so the background has to be extrapolated from a fit at higher energies. In the straightforward example shown in Fig. 10, the background at Na is underestimated. Fitting at 2.5 keV would raise the level, and the error in predicting the background would exceed the statistical uncertainty at Na.

The points chosen for background fitting must be representative of the underlying bremsstrahlung background. Figure 11 demonstrates a bad fit to a spectrum of Fe[S.sub.2] where the background has been fitted using points at 2 keV and 8 keV. The point at 2 keV is raised above the true background by a residual low energy tail on the SK[alpha] peak caused by incomplete charge collection (ICC); consequently the fit is poor both at low energies and above the SK peak. Whereas tailing is not present on every peak, this example demonstrates that difficulties arise with background modelling whenever there are smooth background artifacts that are not detected as peaks. In Fig. 12, the background model for the GaP spectrum fits well at all energies when points at 8 keV and 16 keV are used for scaling. In contrast, Fig. 13 shows a spectrum recorded from the same sample under the same conditions but using a detector with a defective electron trap. This time, when the theoretical model is force fit to the spectrum using t he same points at 8 keV and 16 keV, the background estimate is very poor. In this case, the continuum due to backscattered electrons that enter the detector has raised the background at 16 keV above the underlying bremsstrahlung x-ray continuum.

Another potential source of a smooth background artifact is pile up which occurs at high count rates. Most EDXS systems are fitted with pile up inspectors that are fast and effective for high energy photons. However, lower energy photons are more difficult to detect in the presence of noise. Therefore, at high rates a pile up continuum can appear in the spectrum as a tail on the high energy side of peaks and a sum peak may appear at the end of the tail (13). Figure 14 shows a good background fit to a spectrum from pure cobalt Cobalt, town, Canada
Cobalt (kō`bôlt), town (1991 pop. 1,470), E Ont., Canada, NE of Sudbury, near Lake Timiskaming. Once a center for cobalt and silver mining, the area is now economically depressed.
 at a modest count rate. In particular, note the presence of a strong CoL peak well below 1 keV. When the pulse processor setting is switched to a shorter time constant and the beam current increased, the high count rate spectrum shown in Fig. 15 exhibits strong pile up effects involving CoL photons. As a consequence, the same background modelling procedure now gives a very poor estimate of background.

Thus, background modelling can in principle produce the most accurate background correction but it is particularly sensitive to the choice of points for fitting. The digital filtering approach is less sensitive to smooth background artifacts and avoids the need to choose fitting points so has more widespread applicability and gives more reproducible results. Both approaches require considerable care in implementation.

2.5 Detection Limits

If a peak is isolated and on a fairly flat background as in Fig. 6 then the background can be fitted either side of the peak, interpolated, and subtracted to give a very accurate measure of total peak area. This is the ideal situation referred to in most textbook calculations of statistical detection limits. However, Figs. 10 to 15 demonstrate how systematic errors in background subtraction can exceed the statistical fluctuations from channel to channel. In addition, if peaks overlap, the background for a given peak is effectively raised by its neighbour. Not only does this worsen wors·en  
tr. & intr.v. wors·ened, wors·en·ing, wors·ens
To make or become worse.


worsen
Verb

to make or become worse

worsening adjn
 the statistical detection limit (14) but also any systematic error in overlap correction may become the limiting factor A factor or condition that, either temporarily or permanently, impedes mission accomplishment. Illustrative examples are transportation network deficiencies, lack of in-place facilities, malpositioned forces or materiel, extreme climatic conditions, distance, transit or overflight rights,  in reliable detection of low concentrations.

3. Peak Overlap Correction

3.1 Overlap Factors

Isolated peaks are rarely found in routine analysis. For example, K[beta]/K[alpha] overlap of the transition elements transition elements or transition metals, in chemistry, group of elements characterized by the filling of an inner d electron orbital as atomic number increases.  is commonplace in analysis of steels and combinations such as WM/TaM/SiK, and PbM/SK involve closely overlapping lines. If overlapping peaks have at least some energy regions where there is no overlap, the interference can be dealt with using "overlap factors" (3). To do this, a spectrum from a standard containing just one elemental elemental

emanating from or pertaining to elements.


elemental diet
see elemental diet.
 peak is acquired, then the net window integral for this peak and all the other elemental peaks is obtained from this spectrum. Thus, the relative fraction of the window integral picked up in the energy windows for other elements can be determined. A complete matrix of factors can be constructed provided enough pure elements or simple compounds are available to obtain the factors for each element. When the unknown spectrum is recorded, the net integral in each energy window is equated to the sum of the window integral for the element of interest plus some fraction of all the other elemental peaks present. The set of simultaneous equations is solved to find the net integrals in the absence of overlaps.

The overlap factor approach avoids any detailed knowledge of peak shape but requires a considerable amount of experimental work to characterise a particular system. If the x-ray detector is changed, or the pulse processing time constant is changed, or the resolution, linearity, zero position, or calibration changes with count rate or degrades over time, then overlap factors will change and corrections will be biased. This is less of a problem in mineralogical analysis where count rates on different materials are similar and there are relatively few elements to consider for analysis. The approach is thus suited to dedicated analysis tasks involving a fixed set of known elements but generally gives a very poor result with severe overlaps such as PbM/SK, BrL/AIK or TaM/WM/SiK. For routine analysis of a wide range of materials, a much more flexible approach is required.

3.2 Least Squares Fitting of Experimental Profiles

When the profile for each elemental peak is known, least squares fitting can be used to find the best combination of profiles that match the sample spectrum (8). Profiles can be experimentally determined using pure element or simple compound standards and the background is subtracted, either explicitly or by digital filtering with a zero-area correlator A Correlator commonly refers to:
  • correlation function (quantum field theory)
  • An optical correlator
 like the "Top-Hat" (9,10). The sample spectrum also must have background removed by the same technique and the linear least squares Linear least squares is a mathematical optimization technique to find an approximate solution for a system of linear equations that has no exact solution. This usually happens if the number of equations (m) is bigger than the number of variables (n).  algorithm will then find that combination of profile intensities that gives the minimum sum of squares of differences with the background-subtracted sample spectrum. A refinement to the technique is to give more weight to those channels where the statistical standard deviation is small and the overall statistical precision of the determined intensities can be calculated, even in the more complex case where digital filtering has been used (9,10).

Acquiring a comprehensive library of profiles for all elements involves a lot of time and expense. If x-ray detectors from the same manufacturing process have similar characteristics, then the same profile library can in principle be used with more than one detector. Thus, a set of "virtual standard profiles" can be provided as part of a packaged software See software package.  solution provided there is some method for correcting for changing resolution and calibration from system to system. In practice, this can be achieved if the system noise is automatically monitored by the electronics; variations in electronic noise can then be corrected by convolving the spectrum and profiles with the appropriate Gaussian functions to bring them all to the same effective noise value. Similarly, zero and gain of the energy scale can be periodically checked using a suitable calibration standard so that the energy scales can be brought into register.

3.3 Calculated Peak Profiles and Incomplete Charge Collection

Stored experimental profiles effectively characterise the energy response of a particular detector but suitable standards are not always available for every element and profiles also have to be obtained for all the EDXS configurations that are to be used. In principle, a mathematical model
Note: The term model has a different meaning in model theory, a branch of mathematical logic. An artifact which is used to illustrate a mathematical idea is also called a mathematical model and this usage is the reverse of the sense explained below.
 can overcome these difficulties. To first approximation approximation /ap·prox·i·ma·tion/ (ah-prok?si-ma´shun)
1. the act or process of bringing into proximity or apposition.

2. a numerical value of limited accuracy.
, the response function of an x-ray detector to monochromatic monochromatic /mono·chro·mat·ic/ (-kro-mat´ik)
1. existing in or having only one color.

2. pertaining to or affected by monochromatic vision.

3. staining with only one dye at a time.
 radiation of energy E is a Gaussian function,

[G.sub.i] = exp exp
abbr.
1. exponent

2. exponential
(- 2.773 * [[([x.sub.i] - [x.sub.E])/[fwhm.sub.E].sup.2]) * 0.9394 * [DELTA]/[fwhm.sub.E]. (7)

where [x.sub.i] is the energy corresponding to channel i and [DELTA] is the channel width for the digitised spectrum. If the energy scale is assumed to be linear then

[x.sub.E] = [x.sub.0] + g * E (8)

where [x.sub.0] is the electronic zero and g is the gain. The peak resolution, [fwhm.sub.E] is given by

fwhm[E.sup.2] = [fwhm.sub.0.sup.2] + [dispersion dispersion, in chemistry
dispersion, in chemistry, mixture in which fine particles of one substance are scattered throughout another substance. A dispersion is classed as a suspension, colloid, or solution.
.sub.E.sup.2] (9)

where [fwhm.sub.0] refers to the electronic noise contribution and dispersionE is the detector contribution to spectral resolution The spectral resolution or resolving power of say a spectrograph, or, more generally, of a frequency spectrum, is a measure of its power to resolve features, say in the electromagnetic spectrum. . For an ideal detector, [dispersion.sub.E.sup.2] = k * E, where k is a constant (typically around 2.48 for Si(Li) if energies are all in eV). These approximations work quite well for x-ray detectors in the energy region 4 keV to 15 keV but incomplete charge collection (ICC) near the front contact of the transducer transducer, device that accepts an input of energy in one form and produces an output of energy in some other form, with a known, fixed relationship between the input and output.  means that in silicon-based detectors, the response function can change significantly for x rays that are absorbed strongly in silicon. In practice this means for energies around 1.84 keV to 3 keY and for energies below 1 keV, ICC effects give peaks a low energy tail, shift the centroid centroid

In geometry, the centre of mass of a two-dimensional figure or three-dimensional solid. Thus the centroid of a two-dimensional figure represents the point at which it could be balanced if it were cut out of, for example, sheet metal.
 below XE and make the peak broader than [fwhm.sub.E] (15,16). Various theoretical models have been proposed and suitable formulae suggested for representing ICC distortion, (e.g., 17,18). However, there is no one model that works for all detector designs and the nature and degree of ICC m ay vary from detector to detector because of changes in the manufacturing process (19) and may also be affected if there is any ice build up on the detector surface. To date, the IEEE (Institute of Electrical and Electronics Engineers, New York, www.ieee.org) A membership organization that includes engineers, scientists and students in electronics and allied fields.  "Peak to background" test, which ratios the height of the MnK[alpha] peak from an Fe55 radioactive source to the background level at 1 keY, is the only official standard pertaining per·tain  
intr.v. per·tained, per·tain·ing, per·tains
1. To have reference; relate: evidence that pertains to the accident.

2.
 to charge collection (20) but is a very poor measure of tailing. For example, Fig. 16 shows three spectra of [FeS.sub.2] taken at 20 kV with different x-ray detectors. In the detector showing the worst tailing on SK[alpha], IEEE P:B was 18 000:1 whereas the one with the least tailing gave P:B = 16,000:1. Either value would normally be regarded as an indicator of excellent charge collection and in this case, the better detector had a worse value of P: B. At energies> 3 keV, the degree of tailing is less because x rays penetrate deeper into Si and some of the effect of ICC can be taken into account by shifting and broadening a standard Gaussian function . However, some tailing remains and Fig. 17 shows that even with a good detector with ICC performance equivalent to the best in Fig. 16, the residual non-Gaussian tail component still represent of the order of 1 % of the main Ti K[alpha] peak. Tailing can be accommodated by a suitable modification to the shape model as shown in Fig. 17. Any uncorrected residual tail may appear in the results as spurious spu·ri·ous
adj.
Similar in appearance or symptoms but unrelated in morphology or pathology; false.



spurious

simulated; not genuine; false.
 concentrations for elements not present in the sample. Furthermore, as shown in Fig. 11, a residual tail can affect accuracy of background subtraction if background modelling is used.

3.4 The Effect of Inaccuracies in Position and Width

Even if the peak profiles include ICC, either because they have been recorded experimentally or have been accurately modelled, then the position and width may still vary due to instrumental effects. Peak position is normally easier to control and a good design with precision components can deliver stabilities <0.01 %/[degrees]C for gain, g. However, stabilisation Noun 1. stabilisation - the act of making something (as a vessel or aircraft) less likely to overturn
stabilization

improvement - the act of improving something; "their improvements increased the value of the property"
 for zero, x0, is not available on all electronic pulse processors and the baseline may move with count rate, particularly if there are a lot of low energy x rays that go undetected by the electronic pile-up pile·up or pile-up  
n.
1. Informal A serious collision usually involving several motor vehicles.

2. An accumulation: "the pile-up of unsold autos" 
 inspector (21). The electronic noise, [fwhm.sub.o] in Eq. (9), will change with processing time constants used in the electronics and may also change with count rate. In fact some electronic processors have adaptive shaping that is designed to produce the best resolution possible for a given count rate (22); this not only guarantees that resolution changes with count rate, but also introduces a weighted series of Gaussians with different [fwhm.sub.o] values into the peak shape model. Noise contribution to a detector may also change over time and be temporarily affected by electronic interference. It is therefore imperative to have a some method of correcting for inevitable changes in position and resolution.

The sensitivity to errors in position and width will now be established using the simplifying assumptions that peaks are simple Gaussians, that the background subtraction is perfect and that profiles are fitted to the data by weighted least squares Weighted least squares is a method of regression, similar to least squares in that it uses the same minimization of the sum of the residuals:

. Figure 18 shows the result of fitting a profile with correct fwhm = 100 eV but shifted by 1 eV relative to a sample peak on a background of the same height (P/B P/B

See: Price to book ratio
 = 1). The residual is obtained by subtracting the background and the fitted profile result. The residual shows a characteristic bipolar (1) See bipolar transmission.

(2) One of two major categories of transistor; the other is "field effect transistor" (FET). Although the first transistors and first silicon chips were bipolar, most chips today are field effect transistors wired as CMOS logic, which
 shape with lobe lobe (lob)
1. a more or less well-defined portion of an organ or gland.

2. one of the main divisions of a tooth crown.
 amplitudes of about 1 % of the original peak height for a positional error of 1 % of fwhm. The residual has close to zero net area so that the area of the fitted peak is still correct provided the shift is a small percentage of fwhm. Figure 19 shows the result of fitting a profile with the wrong width, fwhm = 103 eV, to a peak with fwhm = 100 eV on a background of the same height. The least squares algorithm guarantees that the residual has the minimum wei ghted sum of squares but when the resolution is incorrect, the residual has net negative area. This means that even with an isolated peak and an easy background subtraction, the area (and therefore measured x-ray line intensity) will be overestimated when the profile used for fitting is too broad. In practice, there are likely to be both position and width errors and Fig. 20 shows an example where not only is the area of the peak overestimated, but also the residuals might be mistaken as small concentrations of elements with nearby peaks.

Statistical weighting does have an influence on the sensitivity to error. If a peak is very small compared to the background, then statistical weighting is uniform and the results are the same as if no weighting was used. If the peak-to-background ratio is very high, then statistical weighting will force the residual to be small in the wings of a peak and this produces a different result. Other techniques for peak area measurement are affected to some extent by shift and width errors. For example, if a peak is integrated over an energy window close to fwhm in width, the area represents a certain proportion of the true peak area. If the spectrometer spectrometer

Device for detecting and analyzing wavelengths of electromagnetic radiation, commonly used for molecular spectroscopy; more broadly, any of various instruments in which an emission (as of electromagnetic radiation or particles) is spread out according to some
 drifts or resolution changes, then this proportion will change.

In Fig. 21 the sensitivity to shift errors is compared for simple window integral and least squares fitting techniques. For an isolated peak, a small shift in position has hardly any effect on the measured area and even in the extreme weighted least squares case, a shift of 4 % of fwhm gives less than 1 % relative error in the measured intensity. In contrast, an error in peak width has a significant effect on peak areas as shown in Fig. 22. The case of weighted least squares with an isolated high peak on a low background will rarely occur so it is the curves for the window integral method and for peaks on a high background that are more likely to be relevant to practical situations of analytical interest. These show that for a 4 % error in fwhm, the relative error in area is about 2 % and a useful approximation is that the relative error in measured area is about half the relative error in fwhm assumed for the peak.

A much more common and serious problem for EDXS is the case of overlapping peaks. When peaks are less than a fwhm apart, a small error in position will seriously affect the distribution of areas in the fitted result. For example, if there was a large peak near 1.75 keV in the spectrum and it was not known whether there was some Si or W present, two profiles would be fitted to the spectrum corresponding to the expected positions of SiK[alpha] and WM[alpha] which are 35 eV apart. Figure 23 demonstrates the effect of a positional calibration error on the fitted results when the true peak contains only WM[alpha]. When there is no positional error then the result is the expected unity mass fraction for W, 0 for Si. However, if the profiles are both too high in energy, the W is underestimated and a spurious concentration of Si is reported. In these extreme overlap situations, a useful rule of thumb is that the spurious contribution as a fraction of the main peak is given by the ratio of Shift! Separation. Thus, fo r two peaks separated by 40 eV, a shift error of 4 eV gives an error in area for both peaks equivalent to about 10 % of the main peak area. Therefore, in an even more extreme overlap example such as Alk[alpha]/BrL[alpha], where the separation is only 7 eV, just a 1 eV error in position will produce errors equivalent to 14 % of the main peak area.

Shift and resolution errors can still be a problem in more modest overlap situations. As Fig. 20 suggests, if there are any peaks within [+ or -] 2 X fwhm of a large peak, then the residual due to position or width errors can be picked up as spurious concentration. As shown in Sec. 2.4, background modelling is subject to errors particularly when suitable background points cannot be found near to the peak being measured. The digital filtering approach avoids this problem because the "Top-Hat" effectively makes a local background estimate very close to the peak. However, if there are errors in position or width, the sort of residual shown in Fig. 20 now influences the "Top-Hat" background estimate for nearby peaks so the influence of these errors extends to [+ or -] 3 X fwhm from the main peak. For example, MgK[alpha] and AIK AIK As I Know
AIK Assistance in Kind (host nation support)
AIK Allmäna Idrottsklubben (Swedish sports club)
AIK American Institute of Kenpo (Tucson, AZ marital arts) 
[alpha] peaks are separated by 230 eV and if a trace of Mg is to be determined in the presence of a large Al concentration, then position and width errors can still be an issue if the fwhm exceeds 80 eV.

3.5 Nonlinear A system in which the output is not a uniform relationship to the input.

nonlinear - (Scientific computation) A property of a system whose output is not proportional to its input.
 Fitting and Derivative Profiles

The discussion so far has assumed that peak width and position are predetermined pre·de·ter·mine  
v. pre·de·ter·mined, pre·de·ter·min·ing, pre·de·ter·mines

v.tr.
1. To determine, decide, or establish in advance:
 in which case the least squares solution can be obtained in closed form without iteration. Position and width can also be included as parameters in a either a non-linear least squares fit by sequential simplex for example (23) or by including first and second derivative profiles in a linear fit (24). However, if we let position and resolution vary without constraint, the best fit to the data can be obtained with combinations of position and resolution that do not correspond to the real solution [25]. This problem occurs because statistical noise can make it impossible to differentiate between plausible alternatives as the following example will demonstrate. Figure 24 shows a mixture of two Gaussian peaks, fwhm 100 eV, separated by 30 eV on a uniform background. This could correspond to a TaM peak next to a SiK peak for example and a representative level of statistical noise is shown. If the candidate peak profiles are allowed to shift along the energy axis while maintaining a constant separation, then another good fit is obtained with a shift of 13 eV as shown in Fig. 25. In this case, even though the fit is good, the peak heights are completely reversed and the fitted result for the left peak is now twice the height of the peak on the right! If the profile width is also allowed to vary, then another good fit can be obtained by fitting a single peak to the sum as shown in Fig. 26. In this case, if the sample did contain Ta and Si, the good fit with just a single peak would suggest that only one of these elements were present. Therefore, even qualitative identification of elements present would fail.

Whereas varying position and width can be used to obtain a good fit to a single peak, it is not suitable for resolving severe overlaps. If there are always going to be strong isolated peaks present in the spectrum and only [x.sub.0], [fwhm.sub.0] and g are varied in the fitting procedure, then the dominant peaks will effectively provide an internal calibration and constrain con·strain  
tr.v. con·strained, con·strain·ing, con·strains
1. To compel by physical, moral, or circumstantial force; oblige: felt constrained to object. See Synonyms at force.

2.
 the fit to provide sensible solutions (26). However, in low kV spectra in particular, there are unlikely to be any strong isolated peaks. Furthermore, if the spectrum contains only a few counts, non-linear methods will not converge con·verge  
v. con·verged, con·verg·ing, con·verg·es

v.intr.
1.
a. To tend toward or approach an intersecting point: lines that converge.

b.
 on any sensible result because residual errors (Mensuration) See Error, 6 (b).

See also: Residual
 in the fit will be totally masked A state of being disabled or cut off.  by statistical noise. Therefore, for an entirely general solution to microanalysis, a reliable energy calibration and specific measurement of resolution are required in order to overcome the accuracy barriers described in Sec. 3.4.

3.6 Calibration Requirements

If a Gaussian peak is on a low background, the background can be subtracted by linear interpolation and a quadratic curve fitted to the log intensity by weighted least squares using all points above 10 % of peak maximum. Thus, Eq. (8) can provide estimates for [X.sub.E] and [fwhm.sub.E]. If the area of the peak is N counts, then the standard deviation in measured position is approximately [[sigma].sub.x] = 0.43 [fwhm.sub.E]/sqrt (N) and the standard deviation in measured resolution is approximately [[sigma].sub.fwhm] = [fwhm.sub.E]/sqrt (N). (These values have been confirmed by simulation and are close to the precision obtainable using the full peak (27)). With two Gaussian peaks, A and B corresponding to energies [E.sub.A] and [E.sub.B], the fitted values of position, [x.sub.A] and [x.sub.B] can be used to determine [x.sub.0] and g using Eq. (8). Then the predicted peak position for a line at energy E will be

[x.sub.E] = [x.sub.A] * (1 - p) + [x.sub.B] * P (10)

where p = (E - [E.sub.A])/([E.sub.B] - [E.sub.A]). The standard deviation in this predicted position will then be

[[sigma].sub.E] = [[[sigma].sub.A.sup.2] * [(1 - p).sup.2] + [[sigma].sub.B.sup.2] * [p.sup.2]].sup.05] (11)

For EDXS systems without an automatic zero measurement, calibration requires a spectrum with two well defined x-ray peaks of similar area, usually AlK[alpha] and CuK[alpha] at 1.49 keV and 8.04 keV. If a spectrum with 300 000 total counts is recorded at 20 kV from a typical Si(Li) detector with [florin]whm = 133 eV at 5.9 keV and there are 50 000 counts in each peak, then the precision in calibration at energy E will from Eq. (H) be [+ or -] 0.19 eV at the OK[alpha] energy. Some EDXS systems have an automatic "strobed" zero energy measurement which effectively provides a reference peak corresponding to zero energy. Therefore, calibration can be achieved using a spectrum with just a single x-ray peak. If this is CuK[alpha] and has 100 000 counts (again assuming a total spectrum area of about 300 000 counts) and the strobe strobe  
n.
1. A strobe light.

2. A stroboscope.

3. A spot of higher than normal intensity in the sweep of an indicator, as on a radar screen, used as a reference mark for determining distance.
 zero peak has an area of 100 000 counts, then the precision in calibration at energy E will be [+ or -] 0.07 eV at the OK[alpha] energy. Resolution calibration is achieved in a similar manner where [florin]wh[m.sub.A] and [florin]wh[m.sub.B] are used to determine [florin]wh[m.sub.o] and k using Eq. (9). Using standard methods for error propagation The transmission (spreading) of signals from one place to another. , standard deviation for the predicted fwhm at energy E will be

[[sigma].sub.[fwhm.sub.E]] = [[sigma].sub.[fwhm.sub.A]].sub.2] * [([fwhm.sub.A]/ [fwhm.sub.E]).sup.2] * [(1-p).sup.2] + [[sigma].sub.[fwhm.sub.B].sub.2] * [([fwhm.sub.A]/ [fwhm.sub.B]).sup.2] * [p.sup.2].sup.0.5] (12)

Using the same example with a total spectrum area of 300 000 counts the relative uncertainty in fwhm for OK[alpha] would be 0.89 % for the method using AIK[alpha] and CuK[alpha] peaks with 50 000 counts and 0.24 % using strobe zero peak and CuK[alpha] peak of 100 000 counts.

Given the magnitude of effects demonstrated in Sec. 3.4, it would appear that calibration for position and resolution should not be a barrier to accuracy provided a suitable standard is used and at least 300 000 counts are acquired in the calibration spectrum. If the temperature changes by only 1[degree]C after the calibration, positional errors could typically be 1 eV or more. Although some sophisticated electronic designs claim high stability, in many systems, variation in position and resolution can be as much as 5 eV between low and high count rate and some electronic designs may also be sensitive to the balance of low and high energy x rays in the spectrum. Consequently, position and resolution errors are more likely to be dominated by instrumental effects rather than the counting statistics for calibration described by Eqs. (11) and (12).

4. Total System Stability and Reproducibility

Figures 14 and 15 indicate how the measured spectrum may change when count rate or electronic processor settings are altered. As well as background artifacts, peaks may shift in position and resolution may deteriorate de·te·ri·o·rate
v.
1. To grow worse in function or condition.

2. To weaken or disintegrate.
 at high count rates so that any calibration performed at low count rate or a particular processor setting may become invalid at higher count rates or a different setting. Figure 16 points out some of the variability in ICC that may be seen from detector to detector so that even at low count rates, the same software and procedure may produce different results with different detectors. Complete EDXS analysis systems are designed for ease of use and it is often difficult to get access to individual components to make specific tests. Furthermore, manufacturers may include specialised correction software to correct for certain types of instability and variation. Often the only way to test for stability and reproducibility is to treat the total system as a "black box" and validate it in real situation s that are representative of the analytical problems to be tackled. There will always be some statistical fluctuation in results due to the operation of subtracting background so that both positive and negative results are to be expected for any element with zero concentration. Any system that does not report small negative concentrations should therefore be treated with suspicion. Usually, tests similar to the following can be performed on any EDXS system and can be adapted to suit the particular application.

A severe test of spectrum processing is to acquire a spectrum from a sample of Al or [Al.sub.2][O.sub.3] and force the system to analyse for MgK, AlK, SiK, and BrL . Of course, only Al should be detected, but any error in background modelling or overlap correction will show up as spurious concentrations of Mg, Si, and Br. Since BrL is only 7 eV apart from AIK, this test is extremely sensitive to peak position accuracy whereas Mg and Al provide a check for less severe overlaps. Table 1 demonstrates the test in use. Analysis of an [Al.sub.2][O.sub.3] sample has been performed at different pulse processor settings and at a variety of input count rates. The statistical errors reported by the software package are shown next to each analysis result. Both positive and negative results are reported and any result within [+ or -] 3 standard deviations could reasonably be regarded as the result of statistical variation. However, the 35 kcps result at PT3 shows some significant spurious concentrations of Mg, Si, and Br.

The Al/Br overlap is particularly severe but is typical of the L/K overlaps that occur at low energies such as Ti/O, V/O V/O Voice Over (cinematography) , Cr/O. A less severe test is to use a sample of pure Si and analyse for WM, SiK, TaM, checking for spurious amounts of W or Ta. In this "null A character that is all 0 bits. Also written as "NUL," it is the first character in the ASCII and EBCDIC data codes. In hex, it displays and prints as 00; in decimal, it may appear as a single zero in a chart of codes, but displays and prints as a blank space. " testing approach, it is essential that the system reports the results for all elements and their standard deviations. Then it is possible to establish what the real limit of detectability is for a low concentration of any element. As Fig. 2 shows, analysis at low kV presents a severe challenge for any EDXS system but again, null tests can easily be devised to establish how effective spectrum processing is under these conditions. Even if peak intensities can be determined with some confidence, there is no guarantee that chemical effects will not affect XRCFs and some testing with representative compounds is always desirable. While it is important to explore and gain confidence in all likely analytical configurations, there is little point in pushing t he EDXS to extremes that are never likely to be used.

5. Conclusions

Some time in the future, spectrometers may have sufficiently good resolution to leave some background between every characteristic line so that simple interpolation could be used to obtain peak areas and this would be insensitive in·sen·si·tive  
adj.
1. Not physically sensitive; numb.

2.
a. Lacking in sensitivity to the feelings or circumstances of others; unfeeling.

b.
 to peak shape and position errors. Meanwhile, all current EDXS systems have to cope with peak overlap and whatever the sophistication so·phis·ti·cate  
v. so·phis·ti·cat·ed, so·phis·ti·cat·ing, so·phis·ti·cates

v.tr.
1. To cause to become less natural, especially to make less naive and more worldly.

2.
 in spectrum processing algorithms, the following factors always affect accuracy in determining characteristic peak intensities:

1) Prediction of peak shape, width and position for every characteristic line

2) Measurement of background intensity

3) Stability of spectrometer characteristics with time and changing count rate

Accuracy with relative errors less than 2 % for major constituents (mass fractions greater than 0.1) and errors less than 0.001 mass fraction at low concentrations can usually be achieved by restricting count rates to below 3 kcps, selecting specific elements on the basis of prior knowledge of the sample, working at 15 kV or higher so that the well-separated K lines for transition elements are available, avoiding analysis of any line below 1 keV in energy and making regular checks of energy calibration and beam current. To achieve this accuracy beyond this restricted range of application, particularly towards low keV, requires excellent electronic and detector stability and improved methods for modelling peak and background shape.

[FIGURE 1 OMITTED]

[FIGURE 2 OMITTED]

[FIGURE 3 OMITTED]

[FIGURE 4 OMITTED]

[FIGURE 5 OMITTED]

[FIGURE 6 OMITTED]

[FIGURE 7 OMITTED]

[FIGURE 8 OMITTED]

[FIGURE 9 OMITTED]

[FIGURE 10 OMITTED]

[FIGURE 11 OMITTED]

[FIGURE 12 OMITTED]

[FIGURE 13 OMITTED]

[FIGURE 14 OMITTED]

[FIGURE 15 OMITTED]

[FIGURE 16 OMITTED]

[FIGURE 17 OMITTED]

[FIGURE 18 OMITTED]

[FIGURE 19 OMITTED]

[FIGURE 20 OMITTED]

[FIGURE 21 OMITTED]

[FIGURE 22 OMITTED]

[FIGURE 23 OMITTED]

[FIGURE 24 OMITTED]

[FIGURE 25 OMITTED]

[FIGURE 26 OMITTED]
Table 1

Black box testing using spectra from [Al.sub.2][O.sub.3] sample obtained
in SEM at 20 kV and various beam currents. Results for mass fraction
with statistical standard deviation estimates are shown for various
electronic settings for process time (PT) and input count rates.
Statistically significant "mistakes" are shown in italics.


Spectrum            Mg                 AI                 Si
                                  Mass fraction
PT6, 1 kHz   -0.07[+ or -]0.13  53.90[+ or -]2.13  0.08[+ or -]0.21
PT5, 10 kHz   0.04[+ or -]0.02  53.20[+ or -]0.41  0.07[+ or -]0.04
PT3, 1 kHz    0.13[+ or -]0.15  48.71[+ or -]2.87  0.11[+ or -]0.30
PT3, 10 kHz   0.16[+ or -]0.03  53.84[+ or -]0.56  0.08[+ or -]0.05
PT3, 35 kHz   0.19[+ or -]0.02  54.48[+ or -]0.31  0.15[+ or -]0.03


Spectrum            Br

PT6, 1 kHz   -1.90[+ or -]3.35
PT5, 10 kHz  -0.73[+ or -]0.65
PT3, 1 kHz    7.51[+ or -]4.44
PT3, 10 kHz  -2.15[+ or -]0.88
PT3, 35 kHz  -3.58[+ or -]0.49


Accepted: August 22, 2002

6. References

(1.) R. Castaing, Application des sondes electroniques a one methode d'analyse ponctuelle chimique et cristallographiue, Thesis, University of Paris, O.N.E.R.A. Publ. No 55 (1951).

(2.) S. J. B. Reed, Electron Microprobe The electron microprobe is an analytical tool used to non-destructively determine the chemical composition of small volumes of solid materials. It uses a high-energy focused beam of electrons to generate X-rays characteristic of the elements present within a sample volumes 1 to 3  Analysis, 2nd Ed., Cambridge Univ. Press (1993) p. xvii.

(3.) S. J. B. Reed and N. G. Ware, Quantitative Electron Microprobe Analysis Using a Lithium lithium (lĭth`ēəm) [Gr.,=stone], metallic chemical element; symbol Li; at. no. 3; at. wt. 6.941; m.p. about 180.54°C;; b.p. about 1,342°C;; sp. gr. .534 at 20°C;; valence +1. Lithium is a soft, silver-white metal.  Drifted Silicon Detector, X-Ray Spectrometry 2, 69-74 (1973).

(4.) A. C. Dunham and F. C. F. Wilkinson, Accuracy, Precision and Detection Limits of Energy-dispersive Electron-microprobe Analyses of Silicates, X-Ray Spectrom. 7 (2), 50-56 (1978).

(5.) D. R. Beaman and L. F. Solosky, Accuracy of Quantitative Electron Probe Microanalysis with Energy Dispersive Spectrometers, Anal anal (a´n'l) relating to the anus.

a·nal
adj.
1. Of, relating to, or near the anus.

2.
. Chem. 44 (9), 1598-1610 (1972).

(6.) D. E. Newbury, Standardless Quantitative Electron-Excited X-ray Microanalysis by Energy-Dispersive Spectrometry spectrometry /spec·trom·e·try/ (spek-trom´e-tre) determination of the wavelengths or frequencies of the lines in a spectrum.

spec·trom·e·try
n.
: What Is Its Proper Role?, Microsc. Microanal. 4 (6), 585-597 (1999).

(7.) E. D. Boyes Boyes is a chain of department stores in the UK. William Boyes founded the firm in 1881 and his sons, grandsons and great-grandchildren have carried on the business. It is still family owned today and has grown from one small shop in Scarborough, North Yorkshire to a chain of 33 , I. R. Hartmann, F. Gooding, D. Sokola, L. Hanna, and D. L. Smith, EDX chemical microanalysis of bulk specimens in the SEM at low beam voltage and high spatial resolution (Data West Research Agency definition: see GIS glossary.) A measure of the accuracy or detail of a graphic display, expressed as dots per inch, pixels per line, lines per millimeter, etc. It is a measure of how fine an image is, usually expressed in dots per inch (dpi). , Inst. Phys. Conf. Ser. No. 119, EMAG EMAG Electromagnetics
EMAG Electronic Magazine
EMAG Equitable Members Action Group (UK)
EMAG Expatriate Medical Assistance Group
 91 Bristol, IOP IOP

intraocular pressure.

IOP Intraocular pressure, see there
 Publishing (1991) pp. 441-444.

(8.) P. J. Statham, A Comparative Study of Techniques for Quantitative Analysis of the X-Ray Spectra Obtained with a Si(Li) Detector, X-Ray Spectrom. 5, 16-28 (1976).

(9.) F. H. Schamber, A modification of the linear least-squares fitting method which provides continuum suppression, Proc. Symp. X-Ray Fluorescence X-ray fluorescence (XRF) is the emission of characteristic "secondary" (or fluorescent) X-rays from a material that has been excited by bombarding with high-energy X-rays or gamma rays.  Analysis of Environmental Samples, Chapel Hill, N.C. 1976, T. Dzubay, ed., Ann Arbor Ann Arbor, city (1990 pop. 109,592), seat of Washtenaw co., S Mich., on the Huron River; inc. 1851. It is a research and educational center, with a large number of government and industrial research and development firms, many in high-technology fields such as  Science Pub., Ann Arbor, Mich. (1977) pp. 241-257.

(10.) P. J. Statham, Deconvolution and Background Subtraction by Least-Squares Fitting with Prefiltering of Spectra, Anal. Chem. 49, 2149-2154 (1977).

(11.) E. Lifshin, The use of solid state x-ray detectors for obtaining fundamental x-ray data, Proc. 9th Ann. Conf. Microbeam Analysis Society, 53A-53B (1974).

(12.) C. E. Fiori, R. L. Myklebust, K. F. J. Heinrich, and H. Yakowitz, Prediction of Continuum Intensity in Energy-Dispersive X-Ray Microanalysis, Anal. Chem. 48, 172-176 (1976).

(13.) P. J. Statham, Pile-Up Rejection: Limitations and Corrections for Residual Errors in Energy-dispersive Spectrometers, X-Ray Spectrom. 6 (2), 94-103 (1977).

(14.) P. J. Statham, Quantifying benefits of resolution and count rate in EDX microanalysis, in X-Ray Spectrometry in Electron Beam Instruments, D. B. Williams, J. 1. Goldstein, and D. E. Newbury, eds., Plenum In a building, the space between the real ceiling and the dropped ceiling, which is often used as an air duct for heating and air conditioning. It is also filled with electrical, telephone and network wires. See plenum cable.  Press, NY (1995) pp. 101-126.

(15.) P. J. Statham, X-Ray Microanalysis with Si(Li) detectors, J. Microsc. 123 (1), 1-23 (1981).

(16.) A. J. Craven CRAVEN. A word of obloquy, which in trials by battle, was pronounced by the vanquished; upon which judgment was rendered against him. , C. P. McHardy, and W. A. P. Nicholson, The effect of incomplete charge collection on the peak shapes from Si(Li) x-ray detectors, Proc. EMAG 87, Manchester, UK 1987, Inst. Phys. Conf. Ser. No. 90 (1987) Chap. 11, pp. 345-348.

(17.) D. C. Joy, Modelling the Energy Dispersive X-ray Detector, in X-Ray Spectrometry in Electron Beam Instruments, D. B. Williams, J. I. Goldstein, and D. E. Newbury, eds., Plenum Press, NY (1995) pp. 53-65.

(18.) J. L. Campbell, L. McDonald, T. Hopman, T. Papp, Simulations of Si(Li) x-ray detector response, X-Ray Spectrom. 30, 230-241 (2001).

(19.) J. J. McCarthy James Joseph McCarthy (1817-1882) was an Irish architect, often referred to as the 'Irish Pugin'. Early years
James Joseph McCarthy was born in Dublin on 6 January 1817, son of Charles McCarthy who came of a Co. Kerry family settled in Dublin.
, The Effect of Detector Dead Layers on Light Element Detection, in X-Ray Spectrometry in Electron Beam Instruments, D. B. Williams, J. I. Goldstein, and D. E. Newbury, eds., Plenum Press, NY (1995) pp. 67-81.

(20.) IEEE Standard Test Procedures for Semiconductor X-Ray Energy Spectrometers, ANSI/IEEE Std 759-1984 (1984).

(21.) D. E. Newbury, Artifacts in Energy Dispersive X-Ray Spectrometry in Electron Beam Instruments. Are Things Getting Any Better?, in X-Ray Spectrometry in Electron Beam Instruments, D. B. Williams, J. I. Goldstein, and D. E. Newbury, eds., Plenum Press, NY (1995) pp. 167-201.

(22.) R. B. Mott and J. J. Friel, Improved EDS Performance with Digital Pulse Processing, in X-Ray Spectrometry in Electron Beam Instruments, D. B. Williams, J. I. Goldstein, and D. E. Newbury, Plenum Press, NY (1995) pp. 127-157.

(23.) C. E. Fiori, R. L. Myklebust, and K. Gorlen, Sequential simplex: a procedure for resolving specral interference in energy dispersive x-ray spectrometry, Proc. Workshop on Energy Dispersive X-Ray spectrometry, NBS (National Bureau of Standards) See NIST.

NBS - National Bureau of Standards: part of the US Department of Commerce, now NIST.
, Gaithersburg, MD, USA, 1979, NBS Pub. 604 (1981) pp. 23-25.

(24.) T. Kitazawa, H. Shuman, and A. P. Somlyo, Quantitative electron probe analysis: problems and solutions, Ultramicroscopy 11, 251-262 (1983).

(25.) P. J. Statham, Pitfalls in Linear and Non-linear Profile-fitting Procedures for Resolving Severely Overlapped Peaks, X-Ray Spectrom. 7 (3), 132-137 (1978).

(26.) H. Nullens, P. Van Espen, and F. Adams, Linear and non-linear peak fitting in energy-dispersive x-ray fluorescence, X-Ray Spectrom. 8, 104 (1979).

(27.) V. Raznikov, A. F. Dodonov, and E. V. Lanin, Data acquisition and processing in high resolution mass spectrometry mass spectrometry
 or mass spectroscopy

Analytic technique by which chemical substances are identified by sorting gaseous ions by mass using electric and magnetic fields.
 using ion counting, Internatl. J. Mass Spectrom. Ion Phys. 25, 295-313 (1977).

About the author: Peter J. Statham is Director of Research at Oxford Instruments Oxford Instruments plc is a United Kingdom manufacturing and research company operating in the fields of instrumentation, analysis, plasma processing, cryogenics and superconductivity.  Analytical Limited.
COPYRIGHT 2002 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2002, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Statham, Peter J.
Publication:Journal of Research of the National Institute of Standards and Technology
Date:Nov 1, 2002
Words:9046
Previous Article:Decomposition of wavelength dispersive X-Ray spectra.
Next Article:Averaging of backscatter intensities in compounds.
Topics:



Related Articles
Direct non-destructive characterization of tire materials by spectroscopy.
Application of x-ray spectroscopic scattering topography to rubber based composites.
NEW NIST MONOGRAPH OF SPECTRAL DATA FOR HIGHLY IONIZED ATOMS.(National Institute of Standards and Technology)(Brief Article)
Characterization of binary/tertiary blends of SBR, NBR and PVC by IR spectroscopy.(mid-IR and near-IR spectroscopy to estimate polymer compositions...
Uncertainty in quantitative electron probe microanalysis.
High count rate electron probe microanalysis.
X-ray microanalysis in the variable pressure (environmental) scanning electron microscope.
Handbook of basic atomic spectroscopic data available online.(General Developments)
X-ray spectrometry of copper: new results on an old subject.
A new X-ray eye on the cosmos.(ASTRONOMY)(X-ray Spectrometer-2 launched by Japan)(Brief Article)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles