Printer Friendly
The Free Library
6,672,202 articles and books
Member login
User name  
Password 
 
Join us Forgot password?

Intramural comparison of NIST laser and optical fiber power calibrations.


The responsivity of two optical detectors Optical detectors

Devices that respond to incident ultraviolet, visible, or infrared electromagnetic radiation by giving rise to an output signal, usually electrical. Based upon the manner of their interaction with radiation, they fall into three categories.
 was determined by the method of direct substitution in four different NIST (National Institute of Standards & Technology, Washington, DC, www.nist.gov) The standards-defining agency of the U.S. government, formerly the National Bureau of Standards. It is one of three agencies that fall under the Technology Administration (www.technology.  measurement facilities. The measurements were intended to demonstrate the determination of absolute responsivity as provided by NIST calibration services at laser and optical-communication wavelengths; nominally 633 nm, 850 nm, 1060 nm, 1310 nm, and 1550 nm. The optical detectors have been designated as checks standards for the purpose of routine intramural intramural /in·tra·mu·ral/ (-mu´r'l) within the wall of an organ.

in·tra·mu·ral
adj.
Occurring or situated within the walls of a cavity or organ.
 comparison of our calibration services and to meet requirements of the NIST quality system, based on ISO (1) See ISO speed.

(2) (International Organization for Standardization, Geneva, Switzerland, www.iso.ch) An organization that sets international standards, founded in 1946. The U.S. member body is ANSI.
 17025. The check standards are two optical-trap detectors, one based on silicon and the other on indium gallium arsenide Indium gallium arsenide (InGaAs) is a semiconductor composed of indium, gallium and arsenic. It is used in high-power and high-frequency electronics because of its superior electron velocity with respect to the more common semiconductors silicon and gallium arsenide.  photodiodes. The four measurement services are based on: (1) the laser optimized cryogenic radiometer radiometer (rā'dēŏm`ətər), instrument for detection or measurement of electromagnetic radiation; the term is applied in particular to devices used to measure infrared radiation.  (LOCR LOCR Legends of Classic Rock (Canadian radio show)
LOCR Linux Optical Character Recognition
) and free field collimated In a straight line. Collimated light beams are parallel rays of light.  laser light; (2) the C-series isoperibol calorimeter calorimeter: see calorimetry.
calorimeter

Device for measuring heat produced during a mechanical, electrical, or chemical reaction and for calculating the heat capacity of materials.
 and free-field collimated laser light; (3) the electrically calibrated cal·i·brate  
tr.v. cal·i·brat·ed, cal·i·brat·ing, cal·i·brates
1. To check, adjust, or determine by comparison with a standard (the graduations of a quantitative measuring instrument):
 pyroelectric py·ro·e·lec·tric  
adj.
Relating to or exhibiting pyroelectricity.

n.
A pyroelectric material.

Adj. 1. pyroelectric - relating to or exhibiting pyroelectricity
pyroelectrical
 radiometer and fiber-coupled laser light; (4) the pyroelectric wedge trap detector, which measures light from a lamp source and monochromator A monochromator is an optical device that transmits a mechanically selectable narrow band of wavelengths of light or other radiation chosen from a wider range of wavelengths available at the input. . The results indicate that the responsivity of the check standards, as determined independently using the four services, agree to within the published expanded uncertainty ranging from approximately 0.02% to 1.24%.

Key words: absolute responsivity; calorimeter; cryogenic radiometer; intercomparison; laser; optical power; optical fiber; pyroelectric detector; spectral responsivity.

**********

1. Introduction

The responsivity of a single optical detector determined from independent comparisons is a means of assuring that our stated uncertainties for a given measurement service are reasonable. Furthermore, such comparisons are a means of complying with the ISO 1725 quality system, the acceptance of which has been agreed upon Adj. 1. agreed upon - constituted or contracted by stipulation or agreement; "stipulatory obligations"
stipulatory

noncontroversial, uncontroversial - not likely to arouse controversy
 by the world's National Measurement Institutes [1]. Through these intercomparisons, we have become a customer of our own services and are able to rigorously evaluate our performance.

We presently have four measurement systems for measuring continuous-wave (CW) laser power at relatively low power levels (milliwatts and less). The oldest among the services was established in the late 1960s and is based on an isoperibol calorimeter, which we call the C-series calorimeter. This measurement device is electrically calibrated and is traceable to electrical standards (the volt and ohm ohm (ōm) [for G. S. Ohm], unit of electrical resistance, defined as the resistance in a circuit in which a potential difference of one volt creates a current of one ampere; hence, 1 ohm equals 1 volt/ampere. ). Since the C-series calorimeter was developed and as the demand for higher accuracy continues, we have more recently developed a measurement service based on a cryogenic radiometer as a primary standard; the Laser Optimized Cryogenic Radiometer (LOCR). To meet growing customer demand for routine calibration of laser and optical-fiber power meters (OFPMs), we have developed two additional calibration services based on comparisons with pyroelectric detectors for absolute responsivity of fibercoupled power meters and relative spectral responsivity. Among these four calibration services, absolute responsivity of fiber-coupled power meters, or OFPMs, at common telecommunications wavelengths (for example, 850 nm, 1310 nm, and 1550 nm) is the most frequently requested. The demand for OFPM calibrations is approximately 75 calibrations per year and continues to increase.

For the intramural comparison we used transfer standards capable of low measurement uncertainty [2,3]. These transfer standards are intended to have very high coupling efficiency, so that they may be used with the four measurement systems having various input-beam geometries, as shown in Fig. 1. These input geometries are: (1) free-field, nearly collimated laser light input; (2) laser light transmitted by single-mode fibers coupled with FC-type fiber connectors; (3) moderately diverging light from a lamp and monochromator. Where possible, we sought to repeat the responsivity measurements with the three laser-based measurement systems, using laser sources with nearly the same wavelengths. Nominally these wavelengths are: 514 nm, 633 nm, 850 nm, 1064 nm, 1310 nm, and 1550 nm. The spectral responsivity measurement system is capable of wavelength adjustment precision of about [+ or -]0.1 nm to approximate the wavelength of any of the laser sources, but the bandwidth is approximately 6 nm [4].

The results of this intramural comparison are given in several subject areas: a description of the two transfer standards, description of the four measurement systems with a statement of the measurement uncertainty, and a summary of results. The uncertainty is given with coverage factor, k = 2, in every case. The coverage factor corresponds to a level of confidence for the relative expanded uncertainty that is approximately 95% [5].

2. Transfer Standards

The transfer standards, or check standards, for this comparison are photodiode-based optical detectors designed and built at NIST [2,3]. For convenience, we use the colloquial col·lo·qui·al  
adj.
1. Characteristic of or appropriate to the spoken language or to writing that seeks the effect of speech; informal.

2. Relating to conversation; conversational.
 term "device under test," or DUT DUT Dutch (language)
DUT Device Under Test
DUT Diplôme Universitaire de Technologie (French University Graduation in Technology)
DUT Dalian University of Technology (also seen as DLUT) 
, to identify these detectors. The optical configuration of each DUT is based on an optical trap having two photodiodes and a spherical mirror. This basic design has been employed in the past using three diodes (and no mirror) [6]. The presence of the spherical mirror reduces the external quantum efficiency of the trap (compared with the three-diode design), but increases the coupling efficiency for larger values of numerical aperture The measurement of the acceptance angle of an optical fiber, which is the maximum angle at which the core of the fiber will take in light that will be contained within the core. Taken from the fiber core axis (center of core), the measurement is the square root of the squared refractive  (NA) [3]. The trap based on silicon (Si) photodiodes is suitable for measurements requiring an NA as large as 0.26. The trap based on indium gallium arsenide (InGaAs) photodiodes is suitable for a slightly lower NA because of the diode packaging constraints (the size of the chip carrier), and the choice of the spherical mirror having a larger radius of curvature Noun 1. radius of curvature - the radius of the circle of curvature; the absolute value of the reciprocal of the curvature of a curve at a given point
radius, r - the length of a line segment between the center and circumference of a circle or sphere
. DUT1 designates the detector based on Si photodiodes and DUT2 designates the detector based on InGaAs photodiodes.

We evaluated the detector responsivity in units of amperes per watt (A/W A/W Awaiting
A/W Alloy Wheels
A/W A-Wing (Star Wars)
A/W Actual Weight
A/W Alternate Weeks
A/W Automatic Win (gaming) 
). The current generated by each DUT was measured by a commercially available picoammeter.

3. C-series Calorimeter

The C-series calorimeter is an isoperibol calorimeter first developed at the National Bureau of Standards National Bureau of Standards: see National Institute of Standards and Technology.

National Bureau of Standards - National Institute of Standards and Technology
 around 1968 [7]. In principle, the measurement device and data analysis have changed very little since then. The calorimeter is considered to be isoperibol when the ambient temperature Outside temperature at any given altitude, preferably expressed in degrees centigrade.  is constant while the calorimeter optical cavity An optical cavity or optical resonator is an arrangement of mirrors that forms a standing wave cavity resonator for light waves. Optical cavities are a major component of lasers, surrounding the gain medium and providing feedback of the laser light.  temperature changes with time. The amount of optical power measured by the calorimeter is derived from knowledge of the exponential decrease in temperature of the calorimeter optical cavity, following a laser injection of optical energy for a known duration [8]. The calorimeter is electrically calibrated by substituting the laser injection with electrical heating, which may be measured accurately. The measurement procedure for the two check standards was identical, except for the laser wavelength used for each calibration.

The test procedure employs a beam splitter A beam splitter is an optical device that splits a beam of light in two. It is the crucial part of most interferometers.

In its most common form, a cube, it is made from two triangular glass prisms which are glued together at their base using Canada balsam.
 that reflects a portion of the laser source to a reference calorimeter and transmits a portion to a primary calorimeter as shown in Fig. 2. From this, the splitter ratio may be determined, which is simply the ratio of optical power in the transmitted beam to the reflected beam. With knowledge of the splitter ratio and the amount of optical power incident on the primary calorimeter, the responsivity of the DUT is determined.

[FIGURE 2 OMITTED]

The DUTs were compared to two NIST standard C-series calorimeters, designated C4-1 and C4-4, at three wavelengths: 632.8 nm (HeNe laser), 859.4 nm (diode laser See laser diode. ), and 1064 nm (Nd: YAG laser YAG laser Yttrium-aluminum-garnet laser, Nd:YAG–neodymium:yttrium-aluminum-garnet–laser. See Laser. ), by use of the measurement scheme shown in Fig. 2. The laser beam had a diameter of 2 mm or less, and was centered on the DUT input aperture. The power impinging upon the test instrument was measured concurrently by means of the calibrated beam splitter and the NIST reference calorimeter. The splitter ratio of the calibrated beam splitter was determined before and after each detector calibration using the two standard calorimeters.

The uncertainty analysis of this measurement is given in detail elsewhere [9]. The expanded uncertainty (with a coverage factor of k = 2) of calibrations based on the primary standard typically ranges from approximately 0.5% to 1%. This value varies largely as a result of measurement noise due to laser power instability, which depends on the laser wavelength and power level.

4. Laser Optimized Cryogenic Radiometer

The NIST LOCR is based on a commercially available cryogenic radiometer [10], which relies on a servo control system to maintain a constant temperature during laser heating of the radiometer cavity. The electrical heating compensation (decrease in heating power), is equal to the amount of optical power absorbed by the radiometer cavity [11].

The responsivities of DUT1 and DUT2 were determined by direct substitution of the LOCR using the calibration system shown in Fig. 3. Each DUT was in turn calibrated with a nominal power Nominal power is a measurement of a mediumwave radio station's output used in the United States. AM broadcasters are licensed by the Federal Communications Commission to operate at a specific nominal power, which may be (and usually is) different from the transmitter power output.  level of 1 mW. The calibration system used a laser power stabilizer stabilizer: see airplane.  and a spatial filter A spatial filter is an optical device which uses the principles of Fourier optics to alter the structure of a beam of coherent light or other electromagnetic radiation. Spatial filtering is commonly used to "clean up" the output of lasers, removing aberrations in the beam due to  to remove scattered light. The optical power applied to the DUT was calculated by interpolating between power measurements performed with the LOCR before and after the test detector measurement, and then applying the appropriate correction factors.

[FIGURE 3 OMITTED]

Four correction factors were used: the LOCR window transmittance (TW), the LOCR receiver absorptance ab·sorp·tance  
n.
The ratio of absorbed to incident radiation.



[absorpt(ion) + -ance.]

Noun 1.
 (AR), the relative aperture transmittance (TA), and the LOCR electrical calibration (kL). The detector's absolute responsivity (R) in A/W is given by the equation:

R = [O.sub.M][T.sub.W][A.sub.R]/[P.sub.S][T.sub.A][k.sub.L], (1)

where [P.sub.S] is the applied power in watts, interpolated interpolated /in·ter·po·lat·ed/ (in-ter´po-la?ted) inserted between other elements or parts.  from bracketing primary standard measurements, and [O.sub.M] is the detector output in amps.

The calibrations were performed using four laser sources with vacuum wavelengths of 514.6744 [+ or -] 0.0044 nm, 632.9918 [+ or -] 0.0054 nm, 1064.4209 [+ or -] 0.0054 nm, and 1550.4142 [+ or -] 0.0055 nm (all uncertainties k = 2). The laser radiation was contained in a single spectral line spectral line
n.
An isolated bright or dark line in a spectrum produced by emission or absorption of light of a single wavelength.



spectral line  
 having an approximately Gaussian intensity profile with known 1/[e.sup.2] diameter at the detector's entrance aperture.

The uncertainty analysis of this measurement is given in detail elsewhere [11]. The expanded uncertainty (with a coverage factor of k = 2) of calibrations based on the primary standard typically ranges from approximately 0.02% to 0.12%.

5. Spectral Responsivity

The spectral responsivity of the DUT was determined by direct substitution with a temperature-stabilized wedge-trap pyroelectric detector [11]. The operation of the pyroelectric element is based on the volume average of the change in temperature with respect to time [12]. The wedge-trap is a secondary standard, with traceability to the C-series calorimeter at several laser wavelengths and a known value for the reflectance of the detector coating over a range of wavelengths. The spectral responsivity of each DUT was measured using the system shown schematically in Fig. 4. The measurement system is designed to accommodate a variety of commercial instruments that cover the spectrum of wavelengths ranging from 400 nm to 1800 nm.

[FIGURE 4 OMITTED]

A tunable monochromatic monochromatic /mono·chro·mat·ic/ (-kro-mat´ik)
1. existing in or having only one color.

2. pertaining to or affected by monochromatic vision.

3. staining with only one dye at a time.
 light source composed of a filament filament, in astronomy: see chromosphere.  lamp, a grating monochromator, and a set of bandpass filters was used to calibrate To adjust or bring into balance. Scanners, CRTs and similar peripherals may require periodic adjustment. Unlike digital devices, the electronic components within these analog devices may change from their original specification. See color calibration and tweak.  the DUT. The output beam from the monochromator (transmitted through air) was directed alternately onto the DUT and the NIST transfer standard with a two-position mirror. The beam was focused (f# [congruent to] f/4) to a diameter of approximately 2 mm at the position of, and normal to, the plane of the test meter. The bandpass of the monochromator was less than 6 nm. The typical uncertainty value of this measurement is 1.24% (k = 2) and the uncertainty analysis is given in detail elsewhere [4]. The largest contribution to the uncertainty is calibration of the transfer standard with the primary standard, stray light, and the fact that the bandwidth is as large as 6 nm.

6. Fiber-Coupled Absolute Responsivity

The absolute responsivity of each DUT was determined by direct substitution of the DUT and an electrically calibrated pyroelectric radiometer (ECPR ECPR European Consortium for Political Research (UK)
ECPR Efficient Component Pricing Rule
). The amount of optical power measured by the ECPR is based on electrical substitution. The ECPR is considered a secondary standard traceable to the LOCR. In addition to being used to quantify the optical power absorbed by the ECPR, the electrical substitution provides thermal compensation for the pyroelectric response. The electrical substitution method In optical fiber technology, the substitution method is a method of measuring the transmission loss of a fiber. It consists of:
  1. using a stable optical source, at the wavelength of interest, to drive a mode scrambler, the output of which overfills (drives) a 1 to
 is accomplished by heating the detector with an amount of electrical power that is equal to, and 180[degrees] out of phase with, the optical beam transmitted through the chopper.

Figure 5 shows the measurement system configuration for calibration of optical fiber power meters (OFPMs). The fiber-based measurement system is based on light emitted from a variety of fiber-coupled laser diodes at wavelengths 672.4 nm, 851.5 nm, 1306.5 nm and 1549.6 nm. Each laser source contains a laser diode whose output is transmitted through a fiber to a fiber splitter, from which about 1% of the energy travels to a monitor detector. All system optical fibers are single mode. The remaining 99% of the energy is transmitted through another fiber to the DUT. All the lasers (except for 1550 nm) are Fabry-Perot types and have several longitudinal spectral modes. The coherence length In physics, coherence length is the propagation distance from a coherent source to a point where an electromagnetic wave maintains a specified degree of coherence. The significance is that interference will be strong within a coherence length of the source, but not beyond it.  of each of these lasers is approximately a few centimeters. The 1550 nm laser is a distributed-feedback (DFB DFB

acronym for dark, firm, dry meat. Called also dark cutting beef.
) laser with a coherence length of a few hundred meters.

A collimation collimation /col·li·ma·tion/ (kol?i-ma´shun)
1. in microscopy, the process of making light rays parallel; the adjustment or aligning of optical axes.

2.
 fixture (not shown in Fig. 5) forms a gap in the fiber between the diode laser and the splitter. This fixture contains two lenses; one to collimate col·li·mate  
tr.v. col·li·mat·ed, col·li·mat·ing, col·li·mates
1. To make parallel; line up.

2. To adjust the line of sight of (an optical device).
, the other to collect and focus light that is transmitted a short distance (free field) through the ECPR chopper wheel. When the chopper wheel is inserted into the gap, a chopped beam is then incident on the detectors (that is, the monitor and the ECPR). The chopper wheel is inserted into the collimation fixture gap each time the ECPR is used for measurements in this system but is removed when not using the ECPR.

The uncertainty value of this measurement is 0.4% (k = 2) and the uncertainty analysis is given in detail elsewhere [13]. The largest contribution to the uncertainty is calibration with the primary standard.

[FIGURE 5 OMITTED]

7. Results

The results are summarized two ways. First the absolute spectral responsivity of DUT1 and DUT2 are shown graphically in Figs. 6 and 7. The absolute responsivity, determined from the spectral responsivity at wavelengths corresponding to the laser-based measurements, was calculated by linear interpolation Linear interpolation is a method of curve fitting using linear polynomials. It is heavily employed in mathematics (particularly numerical analysis), and numerous applications including computer graphics. It is a simple form of interpolation. . Second, the results from all four measurement systems are summarized in Table 1. The maximum difference among the responsivity values acquired with the four measurement systems is stated with a single number (percentage) calculated from

[DELTA][([lambda]).sub.max] = (R[([lambda]).sub.hi] - R[([lambda]).sub.lo])/R[([lambda]).sub.hi]100 (2)

where R([lambda])[.sub.hi] is the maximum responsivity value and R([lambda])[.sub.lo] is the minimum responsivity value at wavelength [lambda].

8. Discussion

We expect that the responsivity values obtained from services having the lowest uncertainty to be bracketed within the range of values from the other services having a larger uncertainty. In every instance, we find that that our expectation is met. Therefore one can say that the measurements agree to within the stated uncertainty.

The results at 1549.6 nm show the greatest maximum difference between the spectral-responsivity measurement system and the OFPM measurement system. In this case, the difference is approximately 1%. The reason for this is difficult to know. One possible explanation is based on the fact that DUT2 may have a field of view narrower than is necessary to completely capture light diverging from the end of the fiber connector at longer wavelengths (for example, near 1550 nm). The coupling efficiency and spatial uniformity are both wavelength dependent. It is possible that the coupling efficiency is low for the divergence at 1550 nm, but not at shorter wavelengths (for single mode fiber, the divergence increases with wavelength [14]). The properties of DUT2 have been evaluated, but the divergence for the 1550 nm laser and fiber used for this measurement has not. With knowledge that the coupling efficiency decreases with increasing divergence, we expect a priori a priori

In epistemology, knowledge that is independent of all particular experiences, as opposed to a posteriori (or empirical) knowledge, which derives from experience.
 the fiber-coupled responsivity to be less than the free-field responsivity. We find just the opposite. The fiber coupled responsivity is greater. Other possible explanations are related to the detector's temperature dependence and the relatively wide bandwidth of the monochromator compared to laser sources. The 1550 nm wavelength is sufficiently shorter than the wavelength of the photodetector's band edge (which is near 1650) so that we suspect neither temperature biasing, nor the weighting errors that are introduced by the monochromator's 6 nm bandpass. To make any statements beyond speculation will require further investigation.

[FIGURE 6 OMITTED]

[FIGURE 7 OMITTED]

9. Conclusion and Future Work

The work described in this paper is important in several ways. Since its inception in 1968, there has been only one documented intramural comparison of the Cseries laser calorimeter with another primary laserpower measurement standard [15]. This is the first intramural comparison of our own check standards against all four measurement systems that are the basis of national traceability for OFPM calibrations. We have found that measurement results from the various services agree to within our stated uncertainties and thus we have further support that our stated uncertainties are reasonable.

In the coming months, another pair of optical-trap detectors (one based on two temperature-controlled germanium germanium (jərmā`nēəm) [from Germany], semimetallic chemical element; symbol Ge; at. no. 32; at. wt. 72.59; m.p. 937.4°C;; b.p. 2,830°C;; sp. gr. 5.323 at 25°C;; valence +2 or +4.  photodiodes [2], the other based on six Si photodiodes [16]) will also be designated as intramural comparison standards. Thus, two pairs of detectors will be available for measurements during the course of a year: one pair will be evaluated during the first half of the year, the other pair for the last half of the year. In time we will develop a history of the detectors while providing a basis for evaluation and recalibration of the various components of the measurement systems.
Beam geometry         Detector  Measurement system

Nearly collimated       [??]    1. C-series
                                2. LOCR
Diverging from fiber    [??]    3. Fiber-based responsivity
                                  with FC connector
Diverging from          [??]    4. Spectral responsivity
  monochromator
  (6 nm bandwidth)

Fig. 1. Relationship of the beam geometry and the detector input among
the four measurement systems.

Table 1. Summary of absolute responsivity values (units are A/W, unless
stated)

                           Relative                     DUT1
Measurement  Approximate   uncertainty
system (a)   power level   (%, k = 2)   514     632.8   672.4

SR            10 [micro]W  1.24         0.4055  0.5013  0.5332
C4-1           1 mW        0.8                  0.5032
C4-4                                            0.5037
OFPM         100 [micro]W  0.4                          0.5379
LOCR           1 mW        0.05         0.4046  0.5014

Maximum
difference
[Eq. (2)]                               0.22%   0.48%   0.87%

                                                   DUT2
Measurement  Nominal wavelength (nm)
system (a)   851.5    859.4       1064       1306.5   1549.6     1550.4

SR           0.6763   0.6829      0.2113     0.8877   .9988      0.9990
C4-1                  0.6848      0.2116
C4-4                  0.6849      0.2114
OFPM         0.6822                          0.8883  1.0080
LOCR                              0.2106                         1.0062

Maximum
difference
[Eq. (2)]    0.87%    0.29%       0.47%      0.07%   0.91%       0.72%

(a) SR = Spectral responsivity measurement system (6 nm bandwidth),
C4-1 = C calorimeter, C4-4 = C calorimeter, OFPM = Optical fiber power
(fiber coupled), LOCR = Laser optimized cryogenic radiometer.


Accepted: March 2, 2004

Available online: http://www.nist.gov/jres

10. References

[1] International Standard, ISO/IEC ISO/IEC International Organization for Standardization/International Electrotechnical Commission (ITU-T M 3000)  17025, General requirements for the competence of testing and calibration laboratories, pp. 1-26, First Edition, 1999-12-15, reference number 17025:1999(E).

[2] J. H. Lehman and X. Li, A transfer standard for optical fiber power metrology, Eng. Lab. Notes in Opt. & Phot. News 10, (5) (1999), archived in Appl. Opt. 38, 7164-7166 (1999).

[3] J. H. Lehman and C. L. Cromer, Optical trap detector for calibration of optical fiber powermeters: coupling efficiency, Appl. Opt. 31, 6531-6536 (2002).

[4] J. H. Lehman, Calibration Service for Spectral Responsivity of Laser and Optical-Fiber Power Meters at Wavelengths Between 0.4 [micro]m and 1.8 [micro]m, NIST Spec. Publ. 250-53 (1999) pp. 1-39.

[5] B. N. Taylor and C. E. Kuyatt, Guidelines for Evaluating and Expressing the Uncertainty of NIST Measurement Results, NIST Technical Note 1297 (1994) p. 5.

[6] N. P. Fox, Trap detectors and their properties, Metrologia 28, 197-202 (1991).

[7] E. D. West, W. E. Case, A. L. Rasmussen, and L. B. Schmidt, A reference calorimeter for laser energy measurements, J. Res. Natl. Bur. Stand. (U.S.) 76A, 13-26 (1972).

[8] E. D. West and K. L. Churney, Theory of Isoperibol Calorimetry calorimetry (kăl'ərĭm`ətrē), measurement of heat and the determination of heat capacity  for Laser Power and Energy Measurements, J. Appl. Phys. 41. 2705-2712 (1970).

[9] E. G. Johnson, Jr., Evaluating the inequivalence and a computational simplification for the NBS (National Bureau of Standards) See NIST.

NBS - National Bureau of Standards: part of the US Department of Commerce, now NIST.
 laser energy standards, Appl. Opt. 16, 2315-2321 (1977).

[10] D. L. Livigni, High accuracy laser power and energy meter calibration service, NIST Spec. Publ. 250-62 (2003) pp. 1-144.

[11] J. H. Lehman, Pyroelectric trap detector for spectral responsivity measurements, Eng. Lab. Notes, Opt, in Phot. News 8, (11), archived in Appl. Opt. 36. 9117-9118 (1997).

[12] S. B. Lang, Sourcebook of pyroelectricity Pyroelectricity (biology)

Electrical polarity in a biological material produced by a change in temperature. Pyroelectricity is probably a basic physical property of all living organisms.
, Gordon and Breach, New York New York, state, United States
New York, Middle Atlantic state of the United States. It is bordered by Vermont, Massachusetts, Connecticut, and the Atlantic Ocean (E), New Jersey and Pennsylvania (S), Lakes Erie and Ontario and the Canadian province of
 (1974) pp. 40-45.

[13] I. Vayshenker, X. Li, D. J. Livigni, T. R. Scott, and C. L. Cromer, NIST Measurement ervices: Optical Fiber Power Meter Calibrations at NIST, NIST Spec. Publ. 250-54 (1999) pp. 1-36.

[14] M. Young, Optics and Lasers Including Fibers and Optical Waveguides, 4th Ed., Springer-Verlag, New York (1993) pp. 260-275.

[15] D. J. Livigni, C. L. Cromer, T. R. Scott, B. Carol Johnson Carol Alfred Johnson (1903 - 30 July 2000) was a British Labour politician. He was Member of Parliament for Lewisham South from 1959 until the general election of February 1974, when the constituency was abolished by boundary changes. References , and Z. M. Zhang, Thermal characterization of a cryogenic radiometer and comparison with a laser calorimeter, Metrologia 35, 819-827 (1998).

[16] J. H. Lehman and C. L. Cromer, Optical tunnel trap detector for radiometric measurements, Metrologia 37, 477-480 (2000).

John H. Lehman, Igor Vayshenker, David J David J. Haskins (b. April 24, 1957, in Northampton, England) is a British alternative rock musician. He was the bassist for the seminal gothic rock band Bauhaus. Life and work . Livigni, and Joshua Hadler

National Institute of Standards and Technology National Institute of Standards and Technology, governmental agency within the U.S. Dept. of Commerce with the mission of "working with industry to develop and apply technology, measurements, and standards" in the national interest. , Boulder, CO 80305

lehman@boulder.nist.gov

igor@boulder.nist.gov

livigni@boulder.nist.gov

hadler@boulder.nist.gov

About the authors: The authors are with CW laser radiometry Radiometry

A branch of science that deals with the measurement or detection of radiant electromagnetic energy. Radiometry is divided according to regions of the spectrum in which the same experimental techniques can be used.
 project, which is part of the Sources and Detectors Group, Optoelectronics Division, NIST Electrical and Electronic Engineering Laboratory. John Lehman is the project leader and is responsible for spectral responsivity of laser and optical fiber power meters. Igor Vayshenker is an electrical engineer responsible for the absolute responsivity of laser and optical fiber power meters. David Livigni is an electrical engineer responsible for calibration services using the laser optimized cryogenic radiometer. Josh Hadler is a physicist responsible for C-series calibration services. The National Institute of Standards and Technology is an agency of the Technololgy Administration. U.S. Department of Commerce.
COPYRIGHT 2004 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2004, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Hadler, Joshua
Publication:Journal of Research of the National Institute of Standards and Technology
Date:Mar 1, 2004
Words:3660
Previous Article:Software architecture for a Virtual Environment for Nano Scale Assembly (VENSA).
Next Article:Erratum: statistical interpretation of key comparison reference value and degrees of equivalence.(Errata)(Correction Notice)
Topics:



Related Articles
Fibers with flare. (heat and optical fibers)
IMPROVED OPTICAL FIBER POWER MEASUREMENT SERVICES AT NIST.(Brief Article)
NEW MEASUREMENT SERVICES FOR RELATIVE INTENSITY NOISE (RIN) OF LASERS.(Brief Article)
NEW MEASUREMENT SERVICE FOR RELATIVE INTENSITY NOISE OF LASERS.(Brief Article)(Statistical Data Included)
NEW CALIBRATION SERVICE FOR HIGH-POWER LASER DETECTORS.(Brief Article)
NIST researchers measure wavelength-dependent gain of raman fiber optical amplifiers. (News Briefs).(National Institute of Standards and...
NIST produces two new wavelength calibration standards for the wavelength division multiplexing L-band. (Standard Reference Materials).(Brief Article)
Nist researchers report special test of high-speed photoreceiver to 110 GHz. (General Developments).(Brief Article)
NIST researchers complete international intercomparison of optical fiberpower measurements. (General Developments).(Brief Article)
Nonlinearity measurements of high-power laser detectors at NIST.

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles