Henley Healthcare, Sugar Land, TX.


Henley Healthcare, Sugar Land, TX, Sept. 20 (Dallas). The letter notified Henley that its Fluidotherapy devices were misbranded mis·brand  
tr.v. mis·brand·ed, mis·brand·ing, mis·brands
To brand or label misleadingly or fraudulently.

Adj. 1.
 due to violations of medical device reporting (MDR MDR,
n See multidrug resistance.

MDR,
n the abbreviation for minimum daily requirement, specifically the Minimum Daily Requirements for Specific Nutrients compiled by the United States Food and Drug Administration.
) regulations. FDA FDA
abbr.
Food and Drug Administration


FDA,
n.pr See Food and Drug Administration.

FDA,
n.pr the abbreviation for the Food and Drug Administration.
 noted two unreported incidents involving a fire triggered by contact between a foam distributor and a heater, and a short circuit leading to "spark/flame" caused by bolts that contacted electrical leads for a heater. The agency took the firm to task for eight claims made in promotional materials including several the firm had agreed to delete. These included "continuously kills bacteria," "post-surgical therapy" and "causes mechanical and thermal stimulation which produces counter-irritation." The letter continued with citations for failure to: have an adequate device master record for component specifications; establish acceptance activities to include inspections, tests, or other verification activities to assure conformance con·for·mance  
n.
Conformity.

Noun 1. conformance - correspondence in form or appearance
conformity

agreement, correspondence - compatibility of observations; "there was no agreement between theory and
; and monitor and control process parameters, component and device characteristics during production. FDA called on the firm to submit a description of Fluidotherapy models for which no 510(k) had been submitted. The 13-page warning letter covered 13 other deviations, including those for complaint handling, design control and validation. The inspection took place from Feb. 24 to Mar. 19. C-H, Design, F-B F-B Forward - Backward (radar image analysis) , Val
COPYRIGHT 1999 Washington Information Source, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 1999 Gale, Cengage Learning. All rights reserved.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Publication:Warning Letter Bulletin
Date:Nov 8, 1999
Words:195
Previous Article:Accutech Orthodontic, Chesapeake, VA.
Next Article:Angeion, Minneapolis, MN.



Related Articles
HUMAN DRUGS: 6548W
New Phase Technologies Relocates to Texas.
Fires in warehouse, devices trip Fluidotherapy maker.
INSPECTION LOG.(Brief Article)(Statistical Data Included)
SOPs, instructions noted in Henley warning letter.(Henley Healthcare)(Brief Article)
HENLEY FLIES IN SOLO SHOW.(L.A. Life)(Review)
New Phase Technologies.
Apartment permit activity in 10 most active markets during first half of 2005.(INDUSTRY TRENDS)
NYC metro area continues to lead nation in new apartment construction.(INDUSTRY TRENDS)
List of top construction markets sees little change.(INDUSTRY TRENDS)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles