Embedded microheating elements in polymeric micro channels for temperature control and fluid flow sensing.


This paper describes the first demonstration of temperature control and flow sensing of fluids using integrated circuit integrated circuit (IC), electronic circuit built on a semiconductor substrate, usually one of single-crystal silicon. The circuit, often called a chip, is packaged in a hermetically sealed case or a nonhermetic plastic capsule, with leads extending from it for  (IC)-based microheating elements embedded in microchannels molded in polydimethylsiloxane (PDMS (Product Data Management System) See PDM. ). Fluid channels and connections to capillary tubing are molded in PDMS using a silicon wafer template. The PDMS film is then bonded to an IC that contains the micromachined microheating elements. Capillary tubes are inserted and fluids are externally pumped through the channels. Heating of the fluid is observed by the formation of bubbles on the microheating element. Sensing of fluid flow is demonstrated by measuring a change in the large signal resistance of the microheater analogous to a hot wire anemometer anemometer: see wind.
anemometer

Instrument for measuring the speed of airflow. The most familiar instruments for measuring wind speeds are the revolving cups that drive an electric generator (useful range approximately 5–100 knots).
 with a detection limit of [+ or -] 320 pL/s.

Keywords: flow sensor A flow sensor is a device for sensing the rate of fluid flow. Typically a flow sensor is the sensing element used in a flow meter, or flow logger, to record the flow of fluids. ; microfluidics; microheating elements; polymer microchannels.

**********

1. Introduction

Over the past several years there has been substantial work in the development of the fluidic flu·id·ic  
adj.
1. Of, relating to, or characteristic of a fluid.

2. Relating to or controlled by fluidics.
 components for micro-total analytical system ([mu]TAS TAS
abbr.
1. telephone answering system

2. true airspeed
) and biochip biochip

Small-scale device, analogous to an integrated circuit, constructed of or used to analyze organic molecules associated with living organisms. One type of theoretical biochip is a small device constructed of large organic molecules, such as proteins, and capable of
 technologies. These components include separation channels [1,2,3,4], polymerase chain reaction polymerase chain reaction (pŏl`ĭmərās') (PCR), laboratory process in which a particular DNA segment from a mixture of DNA chains is rapidly replicated, producing a large, readily analyzed sample of a piece of DNA; the process is  (PCR PCR polymerase chain reaction.

PCR
abbr.
polymerase chain reaction


Polymerase chain reaction (PCR) 
) [5,6], monolithic micro-columns [7], electro-osmotic pumps [8], external pressure driven pumps, bubble pumps [9], valves, and fluidic interconnects and interfaces [10].

These fluidic components have been realized in a variety of substrate materials including silicon, glass, and polymers. A number of trade-offs have been recognized to exist between the choices of substrate materials. For example, silicon/glass systems have the advantages of well-understood surface chemistries and compatibility with IC fabrication fabrication (fab´rikā´shn),
n the construction or making of a restoration.
 processes and electronics. In contrast, plastics offer advantages in cost reduction and ease of fabrication but do not have a clear path towards the integration of electronic and active components.

Approaches for integration of fluidic components have been pursued by a hybrid-package [10] approach and by direct integration in an IC compatible process [11,12]. Both of these methods require the use of silicon as a substrate material.

Integration of thermal elements in microfluidic systems has important applications including the control of chemical reaction rates, temperature cycling, and fluid flow measurement and control. Recently, a pulsed mode thermal fluid flow sensor for microfluidic systems was reported [13]. The sensor was fabricated by a silicon micromachining process and then anodically bonded to a glass plate for sealing and for providing inlet and outlet fluid ports. Pulsed mode operation of the device was used to reduce the sensitivity of the sensor to fluctuations in ambient temperature Outside temperature at any given altitude, preferably expressed in degrees centigrade. .

Our work has been focused on the development of fabrication methods for integration of active components on plastic substrate materials by pick-and-place tools. We call this approach to integration drop-in functionality. Using drop-in functionality, a plastic substrate can be patterned with some network of fluid channels and other passive fluidic elements. Bonding to a top plastic plate that contains a fluid network and fluid interconnects would then provide a seal to the plastic substrate. Silicon/glass/IC-based active elements (with appropriate fluidic and electrical interface) are embedded in the locations to perform the desired functions, e.g., sample manipulation and detection. The advantages of this approach are that active elements important for control and detection of chemical reaction can be integrated in the locations where their functionality is required while still preserving the advantages of using plastic materials for the passive fluidic elements.

In this paper, we present results of integration of microheating elements fabricated on an integrated circuit chip with fluid microchannels in a molded polydimethylsiloxane (PDMS). The PDMS film is cut into a small rectangular chip and positioned over the integrated circuit chip containing a microheating element. The two chips are aligned and bonded so that the microchannel is directly over the microheater. The microchannel is interfaced to an external pump and capillary tubing using a novel method that we have developed to create a capillary interface using commercially available fittings for the mold.

Results for the operation of the assembled device are presented. Fluid flow rate is measured by observing the time of flight of particles in the channel using a florescence microscope. Fluid heating by the microheating element is observed by the formation of steam bubbles in the channel. Finally, the microheating element is demonstrated as a thermal fluid flow sensor with operation analogous to a hot wire anemometer.

2. Experimental Method

We present a fabrication method for the integration of the silicon IC-based microreactive elements with molded fluid channels. The fluid channels are molded [4] using a micromachined silicon wafer template [14] as a cast. Ferules are attached to the template to create a mold for an interface to capillary tubing. The IC chips containing the microheating elements are prepared and then bonded to the fluid channels. The steps to fabricate and assemble the microreactive elements will be described in this section, beginning with the fabrication of the silicon template.

2.1 Fabrication of Silicon Templates

We first reported use of silicon templates for imprinting imprinting, acquisition of behavior in many animal species, in which, at a critical period early in life, the animals form strong and lasting attachments. Imprinting is important for normal social development.  channels in plastic in a previous paper [14]. The process sequence for making the silicon templates will be described in detail in this paper. Low-doped (100) silicon wafers are placed in a wet oxidation Wet oxidation is a form of hydrothermal treatment. It is the oxidation of dissolved or suspended components in water using oxygen as the oxidizer. It is referred to as "Wet Air Oxidation" (WAO) when air is used.  furnace at 980[degrees]C for 20 min to create a 100 nm thick oxide film (1). Following this, photoresist is spun-on the wafer and the wafer is soft baked. Hexamethyldisilazane is spun-on before application of the photoresist as an adhesion promoter. In addition, photoresist is spun-on to the backside of the wafer. Finally, the wafers are rinsed in deionized de·i·on·ize  
tr.v. de·i·on·ized, de·i·on·iz·ing, de·i·on·iz·es
To remove ions from (a solution) using an ion-exchange process.



de·i
 (DI) water and spin dried.

A photomask is prepared for patterning the photoresist. For this work, our designs require a 20 [micro]m minimum feature size. We create the photomasks by high-resolution printout onto clear transparency film. The mask contains the drawings of the fluid channel elements and an alignment mark for rotational alignment to the flat of the silicon wafer. All features are drawn in manhattan format, i.e., lines or boxes parallel or perpendicular to the alignment mark, so that they will be aligned to the {111} planes of the silicon wafer. This alignment is required for anisotropic Refers to properties that differ based on the direction that is measured. For example, an anisotropic antenna is a directional antenna; the power level is not the same in all directions. Contrast with isotropic.  etching of the silicon wafer. The mask is placed over the front side of the silicon wafer and aligned to the flat using the alignment mark. The photoresist is exposed using a UV light source and patterned in developer, then rinsed and dried. Once the photoresist is patterned, it is used to pattern the silicon dioxide silicon dioxide: see silica.


(SiO2) A hard, glassy mineral found in such materials as rock, quartz, sand and opal. In MOS chip fabrication, it is used to create the insulation layer between the metal gates of the top layer and the silicon elements below.
 film by a 2 min wet etch in a 6% buffered HF oxide etch solution, then rinsed and dried. Following this, the photoresist is stripped in acetone acetone (ăs`ĭtōn), dimethyl ketone (dīmĕth`əl kē`tōn), or 2-propanone (prō`pənōn), CH3COCH3  and then the wafer is again rinsed and dried.

At this point, the wafers are ready for the anisotropic etch in tetramethylammonium hydroxide Tetramethylammonium hydroxide (TMAH) is a quaternary ammonium salt with the molecular formula (CH3)4NOH. It is used as an anisotropic etchant of silicon.  (TMAH TMAH Tetra-Methyl Ammonium Hydroxide ) solution. We prepare a TMAH etch solution, based on the work reported by Tabata et al. [15], consisting of 450 mL of 25% by mass electronic grade TMAH in aqueous solution mixed with 900 mL of DI water (1 part 25% TMAH to 2 parts DI water). The solution is placed in a reflux etch container and heated to 80[degrees]C on a hotplate. The wafers are placed in a Teflon holder and immersed in the solution. Following the work by Klassen et al. [16], 1 g of ammonium peroxydisulfate (APODS APoDS Adaptive Processing of Data Structures
APODS Ammonium Peroxy Disulphate
) in powder form is added every 10 min during etching in order to eliminate the formation of hillocks. (2) The etch rate for this solution in the <100> direction was measured as 0.9 [micro]m/min.

Finally, the silicon dioxide mask is stripped by a second 2 min wet etch in 6% buffered HF, then rinsed and dried. The cross section of the template a trapezoidal-shaped line with a width at the top surface of 20 [micro]m. The sides of the trapezoid trapezoid, closed plane figure bounded by four line segments, or sides, two of which are parallel and two of which are nonparallel. The parallel sides of a trapezoid are called bases and the nonparallel sides legs; in an isosceles trapezoid the legs are of equal  are aligned to the {111} planes of the silicon crystal forming a 54.74[degrees]C angle from the wafer surface plane [17].

2.2 Casting PDMS With Capillary Interconnects

The silicon templates were used to mold PDMS films following similar procedures described by Effenhauser [18] and Duffy [4]; however, we developed a method to form an interface for capillary tubing using commercial off-the-shelf Commercial off-the-shelf (COTS) is a term for software or hardware, generally technology or computer products, that are ready-made and available for sale, lease, or license to the general public.  capillary tubing hard-ware. Commercially available capillary tubing, tubing sleeves, and fittings were used to create the fluid interconnects. The tip of a capillary tube fitting was cut and glued (using a quick setting epoxy) to the template at each end of the microchannels for the fluid connection. Teflon sleeves were then inserted into fittings.

Figure 1 shows a schematic of the process used to mold the PDMS film. The silicon template with the attached fittings was placed in a dish. The uncured polymer was poured over the template to a thickness of approximately 0.5 cm and cured at room temperature for 24 h. After curing, the polymer membrane was cut using a scalpel into discrete PDMS chips that would later be bonded to the ICs containing the microheating elements. The PDMS chips were individually peeled away from the silicon template using tweezers tweezers An instrument with pincers used to grasp or extract. See Optical tweezers. . As the chips were peeled away, the fitting tips remained bonded to the silicon template creating a fluid reservoir while the tubing sleeves remained embedded in the membrane. The PDMS chips incorporated a microchannel configuration with two fluid reservoirs, which is designed for alignment to the microheating elements. Capillary tubing could then later be inserted into the tubing sleeves to interface the external pump to the channel.

[FIGURE 1 OMITTED]

2.3 MicroHeating Element Fabrication

The IC-based microheating elements that were used in this work have been previously developed, studied, and described extensively for applications such as thermal radiation thermal radiation

Process by which energy is emitted by a warm surface. The energy is electromagnetic radiation and so travels at the speed of light and does not require a medium to carry it.
 elements [19], thermal flat panel displays [20], and microhotplate gas sensors [21] at NIST (National Institute of Standards & Technology, Washington, DC, www.nist.gov) The standards-defining agency of the U.S. government, formerly the National Bureau of Standards. It is one of three agencies that fall under the Technology Administration (www.technology. . The microheating elements were operated in air and vacuum environments. They have been shown to have a linear relationship with operating temperature and power dissipation of 40[degrees]C/mW throughout their operating temperature of ambient to 1000[degrees]C. It has also been reported that they can be controlled in a pulsed temperature mode with millisecond One thousandth of a second. See space/time and ohnosecond.

(unit) millisecond - (ms) One thousandth of a second, one thousand microseconds. A long time for a modern computer.
 time constants [22] throughout their operating temperature.

The fabrication process for the microheating elements is briefly described in this paragraph. The microheating elements were designed using IC layout software that was modified to include an open tile [23]. The ICs were fabricated through a commercial IC foundry through the MOSIS MOSIS Metal Oxide Semiconductor Implementation System
MOSIS Marine Corps Operating and Support Information System (US Marine Corps)
MOSIS Micro-Optical Silicon System
 Service [24]. The ICs received from the foundry contained precursors of the microheating element that contained openings in the Si[O.sub.2] passivation passivation

the final stage in instrument manufacture, passing the finished instruments through a bath of nitric acid which removes foreign particles and promotes the formation of a protective coating of chromium oxide.
 film created by the layout of the open tiles. To complete their fabrication, the microheating element precursors were silicon micromachined by wet anisotropic silicon etching in TMAH [25] in order to mechanically release them from the silicon substrate without dissolving the aluminum bond pads.

Figure 2 shows an optical micrograph micrograph /mi·cro·graph/ (-graf)
1. an instrument used to record very minute movements by making a greatly magnified photograph of the minute motions of a diaphragm.

2.
 and a scanning electron micrograph electron micrograph
n.
A micrograph made by an electron microscope.
 of a fully fabricated microheating element. The microheating element was designed to be 200 [micro]m square in size. The element consists of a trampoline-shaped membrane of glass that is suspended over a trapezoidal shaped pit in the silicon substrate. The "tethers" at the four corners of the trampoline trampoline

Resilient sheet or web (often of nylon) supported by springs in a metal frame and used as a springboard and landing area in tumbling. Trampolining is an individual sport of acrobatic movements performed after rebounding into the air from the trampoline.
 mechanically support the square plate in the center. The plate in the center contains a polysilicon resistive resistive /re·sis·tive/ (re-zis´tiv) pertaining to or characterized by resistance.  filament filament, in astronomy: see chromosphere.  that is encapsulated in the glass membrane. Two of the tethers contain electrical connections from the filament to external bond pads on the IC. The sides of the trapezoidal pit are aligned with the {111} planes of the silicon substrate. The openings in the Si[O.sub.2] passivation film are determined by the layout of the open tiles in the IC design. Following the fabrication process, the IC chip containing the microheating element was attached and wire bonded in a standard 40-pin dual-in-line package (DIP).

2.4 Assembly

The PDMS chip containing the fluid microchannel was bonded and sealed to the IC chip containing the microheating elements using an oxidation process described by Schueller et al. [26]. After packaging, the IC was placed in an oxygen plasma reactor at 26.7 Pa (200 mTorr) for 5 min. Following this, the PDMS chip was placed with the IC chip in oxygen plasma for 5 s. It was found that longer treatments of the PDMS chip in the oxygen plasma hardened and discolored dis·col·or  
v. dis·col·ored, dis·col·or·ing, dis·col·ors

v.tr.
To alter or spoil the color of; stain.

v.intr.
To become altered or spoiled in color.
 it and also resulted in weak bonds.

[FIGURE 2 OMITTED]

Figure 3 shows a diagram of the assembly process. After exposure to the plasma, the IC chip and the PDMS chip were hand assembled under a stereo optical microscope optical microscope

See under microscope.
 using tweezers to position the microchannel patterned in the PDMS chip over the microheating element. The devices were left undisturbed overnight before handling. It was found that leaving the chips undisturbed overnight resulted in strong bonds between the PDMS and the Si[O.sub.2] surface of the IC chip. Figure 4 is a photograph of the fully assembled device with capillary tubing inserted into the tube sleeves.

[FIGURE 3 OMITTED]

[FIGURE 4 OMITTED]

The experimental setup for this study was designed as follows: the two capillary tubes that were inserted into the capillary interface tubing sleeves shown in Fig. 4 were used to vacuum pump Vacuum pump

A device that reduces the pressure of a gas (usually air) in a container. When gas in a closed container is lowered from atmospheric pressure, the operation constitutes an increase in vacuum in this container.
 fluid from a reservoir through the microchannel. One tube was run into a fluid reservoir that was filled with deionized and degassed water. The second capillary tube was run to the vacuum pump. The vacuum had a pressure indicator display.

2.5 Liposome liposome (lī`pəsōm', lĭp`ə–), microscopic, fluid-filled pouch whose walls are made of layers of phospholipids identical to the phospholipids that make up cell membranes.  Preparation

Liposomes Liposomes

Aqueous compartments enclosed by lipid bilayer membranes; liposomes are also known as lipid vesicles. Phospholipid molecules consist of an elongated nonpolar (hydrophobic) structure with a polar (hydrophilic) structure at one end.
 were prepared by the injection method described [27] in previous publications and encapsulated 100 mmol [L.sup.-1] of the fluorescent dye Noun 1. fluorescent dye - a yellow dye that is visible even when highly diluted; used as an absorption indicator when silver nitrate solution is added to sodium chloride in order to precipitate silver chloride (turns pink when no chloride ions are left in solution and , 5-(and -6) carboxyfluorescein (CF) (Molecular Probes Molecular Probes is a biotechnology company located in Eugene, Oregon specializing in fluorescence. The company was founded in 1975 by Richard and Rosaria Haugland in their kitchen in Minnesota, then moved briefly to Texas and finally to Oregon in the early 1980s. , Eugene OR). (3)

3. Results and Discussion

In this section, we present the measurement results for integrated heating elements in molded fluid channels. The pumping of fluids through the capillary interface was first tested. Following this, the temperature of the fluid was thermally cycled using the embedded heating elements. Finally, the microheating elements were used to measure fluid flow.

3.1 Observation of Fluid Flow

In this experiment, liposomes that encapsulated a fluorescent dye [28] were mixed to form a colloidal suspension Noun 1. colloidal suspension - a colloid that has a continuous liquid phase in which a solid is suspended in a liquid
colloidal solution, sol

colloid - a mixture with properties between those of a solution and fine suspension
 with the DI water in the fluid reservoir. Fluid was vacuum pumped from the reservoir into the microchannel.

Figure 5a is an optical micrograph of the fluid channel in the PDMS taken looking down on the device. The cross sectional area of the fluid channel is in the shape of a trapezoid corresponding to the trapezoidal shape of the lines created on the silicon template. The channel had a dimension at the top of the trapezoid of 50 [micro]m and 210 [micro]m at the bottom. This corresponded to a 110 [micro]m channel depth and a 9500 [micro][m.sup.2] cross-sectional area.

Figure 5b is an optical florescence micrograph of the liposomes encapsulating florescent flo·res·cence  
n.
A condition, time, or period of flowering. See Synonyms at bloom1.



[New Latin fl
 dye flowing through the channel with a vacuum pressure of 5 cm Hg measured at the pump. The particles moving through the channel traversed the 250 [micro]m distance indicated in the figure in 1 s corresponding to a particle velocity Particle velocity is the velocity v of a particle (real or imagined) in a medium as it transmits a wave. In many cases this is a longitudinal wave of pressure as with sound, but it can also be a transverse wave as with the vibration of a taut string.  of 250 [micro]m/s. It was noted that particles close to the walls of the channel appeared to move at a slightly slower velocity than particles in the center of the channel; however, assuming a constant flow velocity In fluid dynamics the flow velocity, or velocity field, of a fluid is a vector field which is used to mathematically describe the motion of the fluid. Definition
The flow velocity of a fluid is a vector field

, the flow rate was calculated to be 2.4 nL/s.

Figure 5c is an optical florescence micrograph of the liposomes flowing through the channel at a higher rate corresponding with a vacuum pressure of 20 cm Hg measured at the pump. In this case, the particle flow velocity was too fast for visual measurement of fluid flow rate.

3.2 Thermal Cycling of Fluids

The microheating element was thermally cycled in a stagnant liquid, i.e., no fluid flow in the channel, by switching on and off a current source that was electrically connected to the filament. The on-state current was increased to a point where bubbles could be observed forming to the heater.

[FIGURE 5 OMITTED]

Figure 6 is an optical micrograph of the microheating element in the fluid channel. Figure 6a corresponds to the heater off-state (no current flow). Figure 6b corresponds to the heater operated in the on-state with a current flow of 10 mA.

It was noted that many bubbles (as opposed to one bubble) formed above the microheating element and were much smaller in size in comparison to the device. This indicated that many nucleation sites nucleation sites

the ends of microtubules in the cytoplasmic skeleton; contributes to the growth of protofilaments.
 were available for the bubbles to form on the heating element. This might be attributed to the geometry of the device since it contained a cavity beneath the membrane with openings that could pass liquid, as shown in Fig. 2b.

By visual observation, the bubbles would dissipate instantaneously when the power was switched off to the microheating element. This indicated that the liquid was in a sub-critical boiling condition.

[FIGURE 6 OMITTED]

3.3 Thermal Fluid Flow Sensor

The electrical resistance Electrical resistance

Opposition of a circuit to the flow of electric current. Ohm's law states that the current I flowing in a circuit is proportional to the applied potential difference V.
 of the microheating element was measured to be 3.2 k[ohm ohm (ōm) [for G. S. Ohm], unit of electrical resistance, defined as the resistance in a circuit in which a potential difference of one volt creates a current of one ampere; hence, 1 ohm equals 1 volt/ampere. ] at low current, e.g., negligible heating of the filament so that it was approximately at ambient temperature. As the power dissipation of the filament is increased, its temperature correspondingly increases causing an increase in the electrical resistance. This effect is characterized by a linear temperature coefficient The temperature coefficient is the relative change of a physical property when the temperature is changed by 1 K.

In the following formula, let R be the physical property to be measured, let T be the temperature of at which the property is measured.
 of resistance. The temperature coefficient of resistance (TCR TCR

T cell receptor.
) of the polysilicon filament of microheating elements fabricated using this approach has been previously investigated and characterized to be 0.001/[degrees]C. The electrical resistance of the microheating elements follow the linear relationship of

R = [R.sub.0][1 + TCR (T - [T.sub.0])], (1)

where [R.sub.0] is the resistance at low power dissipation at temperature [T.sub.0], and T is the temperature of the polysilicon filament resulting in a resistance R.

The relationship between the temperature of the filament and the power dissipation is related to the thermal conductivity of its surrounding medium. If the surrounding medium is composed of a fluid (or gas) that is flowing over the heating element, a steady state equilibrium condition will be established between the temperature of the filament and the power dissipation. If the flow rate of the liquid is changed, a new equilibrium temperature will be established.

The change in heater temperature from the first state of equilibrium to the second will result in a corresponding change in the electrical resistance on the microheater according to according to
prep.
1. As stated or indicated by; on the authority of: according to historians.

2. In keeping with: according to instructions.

3.
 Eq. (1). This change in resistance, [DELTA]R = [R.sub.0] - R, is plotted as a function of pump pressure in Fig. 7.

[FIGURE 7 OMITTED]

In this measurement, the power dissipation of the heater was adjusted to give rise to a 3.3 k[ohm] electrical resistance. Using Eq. (1), this corresponds to an approximate (4) 31[degrees]C temperature rise of the microheating element from ambient (room) temperature. A high precision current-voltage meter was used to measure the resistance of the heater. In this measurement, the fluctuation of the resistance was on the order of [+ or -] 0.2 [ohm].

The resistance change in Fig. 7 is plotted with respect to zero fluid flow, i.e., [R.sub.0] corresponds to no flow. As the flow rate is increased, the resistance changes as plotted in the figure to a maximum difference of 7 [ohm] at a pump pressure of 40 cm Hg. This corresponds to a fractional resistance change on the order of 0.2% at 40 cm Hg.

The observation of particle velocity in the preceding section yielded a 2.4 nL/s flow rate at 5 cm Hg. The resistance change of the microheating element at this corresponding pressure is on the order of 1.5 [ohm]. These results indicate that the thermal flow sensor is capable of measuring fluid flow in this channel with a lower detection limit of approximately 320 pL/s. The results in this experiment show a qualitative agreement with measurements shown in [13] even though the device and measurement configurations are substantially different. The PDMS film may be providing significantly higher thermal isolation allowing the measurement of fluid flow without pulsed mode techniques. However, pulsed mode operation will have an advantage of reduced sensitivity to ambient temperature fluctuations of the surrounding medium.

4. Conclusion

This paper presents a first demonstration of temperature control and fluid flow sensing in microchannels molded in polydimethylsiloxane (PDMS) using embedded IC-based microheating elements. The paper also describes a novel method to create a fluid interface to capillary tubing using commercially available capillary tubing connectors. Future work will be in optimization of the IC-based microheating elements, further thermal characterization, and application of the microheating elements for control of microchannel chemical reactions. The silicon machining can be performed from the backside of the chip so that microheating elements could be designed with openings on the topside; this would that allow solution to flow beneath the element. It is expected that optimization of the heater design will further increase the sensitivity of the sensor. In addition, since circuits can be monolithically integrated in this process, an improvement in the detection limit could be achieved by integration of a low-noise on-chip circuit and amplifier.

Acknowledgement

This research is supported by the NIST Advanced Technology Program's DNA DNA: see nucleic acid.
DNA
 or deoxyribonucleic acid

One of two types of nucleic acid (the other is RNA); a complex organic compound found in all living cells and many viruses. It is the chemical substance of genes.
 Tools for Diagnostics and by the Microanalytical Laboratory NIST Competence Project. The authors acknowledge informative discussions with Jon Geist for the optimization of the anisotropic etching process to fabricate the silicon templates.

Accepted: June 1, 2004

Available online: http://www.nist.gov/jres

(1) Note: Silicon wafers can be purchased from the manufacturer with oxide predeposited.

(2) It was noted that hillocks did form without the addition of the APODS, resulting in an extremely rough surface.

(3) Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation or endorsement by NIST, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.

(4) This assumes that the heating element is at the same temperature at each point on its surface where in actuality it is expected that there will be some distribution of temperature.

5. References

[1] D. J. Harrison, K. Fluri, K. Seiler, Z. Fan, C. S. Effenhauser, and A. Manz, Micromachining A Miniaturized Capillary Electrophoresis-Based Chemical-Analysis System On A Chip, Science 261, 895-897 (1993).

[2] S. C. Jacobson, R. Hergenroder, L. B. Koutny, R. J. Warmakc, and J. M. Ramsey, Effects of Injection Schemes and Column Geometry on the Performance of Microchip Electrophoresis Devices, Anal, Chem. 66 (7) 1107-1113 (1994).

[3] A. T. Woolley and R. A. Mathies, Ultra-High-Speed Dna Fragment Separations Using Microfabricated Capillary Array Electrophoresis Chips, Proc. Natl. Acad. Sci. U.S.A. 91 (24) 11348-11352 (1994).

[4] D. C. Duffy, J. C. McDonald J. C. McDonald was an American racecar driver. Indy 500 results

Year Car Start Qual Rank Finish Laps Led Retired
1930 35 13 98.953 21 18 112 0 Fuel tank leak |colspan=6|Totals |112 |0 |
Starts 1
Poles 0
Front Row 0
Wins 0
Top 5 0
, O. J. A. Schueller, and G. M. Whitesides, Rapid prototyping of microfluidic systems in poly(dimethylsiloxane), Anal. Chem. 70 (23), 4974-4984 (1998).

[5] A. T. Woolley, D. Hadley, P. Landre, A. J. deMello, R. A. Mathies, R.A.S.; and M. A, Northrup, Functional integration of PCR amplification and capillary electrophoresis in a microfabricated DNA analysis DNA analysis Any technique used to analyze genes and DNA. See Chromosome walking, DNA fingerprinting, Footprinting, In situ hybridization, Jeffries' probe, Jumping libraries, PCR, RFLP analysis, Southern blot hybridization.  device. Anal. Chem. 68 (23), 4081-4086 (1996).

[6] M. A. Burns, C. H. Mastrangelo, T. S. Sammarco, F. P. Man, J. R. Webster, B. N. Johnson, B. Foerster, D. Jones, Y. Fields, A. R. Kaiser, and D. T. Burke, Microfabricated structures for integrated DNA analysis, Proc. Natl. Acad. Sci. USA 93(11), 5556-5561 (1996).

[7] B. He, N. Tait, and F. Regnier, Fabrication of nanocolumns for liquid chromatography, Anal. Chem. 70 (18), 3790-3797 (1998).

[8] P. H. Paul, D. W. Arnold, and D. J. Rakestraw, ElectroKinetic Generation of High Pressures Using Porous Microstructures. MicroTotal Analysis Systems '98. Harrison, D. J. and Van Den Berg Van den Berg is the surname of:
  • Rudolf van den Berg (born 1949), Dutch director
  • Albert van den Berg (born 1976), South African rugby player
  • Jan Hendrik van den Berg (born 1914), Dutch psychologist
  • Janwillem van den Berg (1920-1985), Dutch speech scientist
, A., eds., Kluwer Academic Publishers, Dordrecht (1998) p. 49.

[9] L. Lin, Microscale Thermal Bubble Formation: Thermophysical Phenomena And Applications, Microscale Thermophys. Eng. 2 (2), 71-85 (1998).

[10] D. Jaeggi, B. L. Gray, N. J. Mourlas, B. P. van Drieenhuizen, K. R. Williams, N. I. Maluf, and G. T. A. Kovacs, Novel Interconnection Technologies for Integrated Microfluidic Systems, Proc. Solid-State Sensor and Actuator Workshop, Hilton Head, SC (1998) pp. 112-115.

[11] J. R. Webster, C. H. Mastrangelo, M. A. Burns, and D. T. Burke, An inexpensive plastic technology for microfabricated capillary electrophoresis chips. MicroTotal Analysis Systems '98, Harrison, D. J. and Van Den Berg, A., eds., Kluwer Academic Publishers, Dordrecht (1998) pp. 249-252.

[12] S. N. Brahmasandra, J. R. Webster, D. T. Burke, C. H. Mastrangelo, and M. A, Burns, Microfabricated fluidic reaction and separation system for integrated DNA analysis, MicroTotal Analysis Systems '98, Harrison, D. J. and Van Den Berg, A., eds., Kluwer Academic Publishers, Dordrecht (1998) pp. 267-270.

[13] N. Okulan, H. T. Henderson, and C. H. Ahn. A Pulsed Mode Micromachined Flow Sensor with Temperature Drift Compensation, IEEE (Institute of Electrical and Electronics Engineers, New York, www.ieee.org) A membership organization that includes engineers, scientists and students in electronics and allied fields.  Trans. Elec. Dev. 47 (2), 340-347 (2000).

[14] L. Martynova, L. E. Locascio, M. Gaitan, G. Kramer, R. Christensen, and W. MacCrehan, Fabrication of Plastic Microfluid Channels by Imprinting Methods, Anal, Chem. 69, 4783-4789 (1997).

[15] O. Tabata, R. Asahi, H. Funabashi, K. Shimaoka, and S. Sugiyama, Anisotropic etching of silicon in TMAH Solutions, Sensors and Actuators A 34, 51-57 (1992).

[16] E. H. Klassen, R. J. Reay, C. Storment, J. Audy, P. Henry, A. P. Brokaw, and G. T. A. Kovacs, Micromachined Thermally Isolated Circuits, Solid-State Sensor And Actuator Workshop Technical Digest, Hilton Head, SC (1996) pp. 127-131.

[17] H. Seidel sei·del  
n.
A beer mug.



[German, from Middle High German sdel, from Latin situla, bucket.]

Noun 1.
, The Mechanism of Anisotropic Silicon Etching and its Relevance for Micromachining, Transducers '87 (1987) pp. 120-125.

[18] C. S. Effenhauser, G. J. M. Bruin, A. Paulus, and M. Ehrat, Integrated capillary electrophoresis on flexible silicone microdevices: Analysis of DNA restriction fragments and detection of single DNA molecules on microchips, Anal. Chem. 69, 3451-3457 (1997).

[19] M. Parameswaran, A. M. Robinson, D. L. Blackburn, M. Gaitan, and J. Geist, Micromachined Thermal Radiation Emitter form a Commercial CMOS (Complementary Metal Oxide Semiconductor) Pronounced "c-moss." The most widely used integrated circuit design. It is found in almost every electronic product from handheld devices to mainframes.  Process, IEEE Elec. Dev. Lett. 12 (2), 57-59 (1991).

[20] M. Gaitan, M. Parameswaran, R. Barry Johnson, and Ronald Chung, Commercial CMOS Foundry Thermal Display, Proc. SPIE SPIE International Society for Optical Engineering
SPIE Society of Photo-Optical Instrumentation Engineers
SPIE Source Path Isolation Engine
SPIE Special Purpose Insertion Extraction
SPIE Software Process Improvement Experimentation
SPIE Standard Protocols in Effect
, Vol. 1969 (1993) pp. 363-369.

[21] J. S. Suehle, R. E. Cavicchi, M. Gaitan, and S. Semancik, Tin Oxide tin oxide (SnO),
n a polishing agent in the form of a purified white powder, prepared as a paste with glycerine or water.
 Gas Sensor Fabricated Using CMOS Micro-hotplates an In-Situ Processing, IEEE Elec. Dev Lett. 14 (3), 118-120 (1993).

[22] R. E. Cavicchi, J. S. Suehle, K. E. Kreider, M. Gaitan, and P. Chaparala, Fast Temperature Programmed Sensing of Micro-Hotplate Gas Sensors, IEEE Elec. Dev. Lett. 16 (6), 286-288 (1995).

[23] J. C. Marshall, M. Parsmeswaran, M. E, Zaghloul, and M. Gaitan, High-Level CAD Melds Micromachined Devices with Foundries, IEEE Circuits and Devices 8 (6), 10-17 (1992).

[24] http://www.mosis.org/Products/mems.html.

[25] N. H. Tea, V. Milanovic, J. S. Suehle, M. Gaitan, M. E. Zaghloul, and Jon Geist, Hybrid Post-processing Etching for CMOS-Compatible MEMS (MicroElectroMechanical Systems) Tiny mechanical devices that are built onto semiconductor chips and are measured in micrometers. In the research labs since the 1980s, MEMS devices began to materialize as commercial products in the mid-1990s. , J. MEMS 6 (4), 363-372 (1997).

[26] O. J. A. Schueller, D. C. Duffy, J. A. Rogers, S. T. Brittain, and G. M. Whitesides, Reconfigurable diffraction gratings based on elastomeric microfluidic devices, Sensors and Actuators A 78, (2-3), 149-159 (1999).

[27] L. Locascio-Brown, A. L. Plant, V. Horvath, R. A. Durst, Liposome Flow-Injection Immunoassay--Implications For Sensitivity, Dynamic-Range, And Antibody Regeneration, Anal, Chem. 62 (23), 2587-2593 (1990).

[28] S. J. Choquette, L. Locascio-Brown, R. A. Durst, Planar Waveguide waveguide, device that controls the propagation of an electromagnetic wave so that the wave is forced to follow a path defined by the physical structure of the guide.  Immunosensor with Fluoresent Liposome Amplification, Anal. Chem. 64 (1), 55-60 (1992).

Michael Gaitan and Laurie E. Locascio

National Institute of Standards and Technology National Institute of Standards and Technology, governmental agency within the U.S. Dept. of Commerce with the mission of "working with industry to develop and apply technology, measurements, and standards" in the national interest. , Gaithersburg, MD 20899-8124

gaitan@nist.gov

About the authors: Dr. Michael Gaitan is an electrical engineer and the Project Leader of the MicroElectroMechanical Systems (MEMS) Project, Semiconductor Electronics Division, NIST Electronics and Electrical Engineering Laboratory. He directs the MEMS Project activities in providing industry with standardized MEMS test structures and test methods for characterizing the thermo-electro-mechanical properties of thin-films used in IC and MEMS tehnologies; works with IC foundries to improve accessibility of MEMS manufacturing; and heads research and development of novel measurement applications of MEMS technology for our new programs in BioElectronics Bioelectronics

A discipline in which biotechnology and electronics are joined in at least three areas of research and development: biosensors, molecular electronics, and neuronal interfaces.
 and Single Molecule Manipulation and Measurement (SM3).

Dr. Laurie E. Locascio is a Project Leader and biomedical bi·o·med·i·cal
adj.
1. Of or relating to biomedicine.

2. Of, relating to, or involving biological, medical, and physical sciences.
 engineer in the Analytical Chemistry Division within the Chemical Science and Technology Laboratory at NIST. Her current research efforts involve the design and application of microfluidic chemical systems, also known as "lab-on-a-chip" devices. This work focuses on the development of new methods for microfabrication and microsystems integration; development of fundamental methods for accurately measuring flow and temperature in microsystems; development of new methods for improved microchemical mi·cro·chem·is·try  
n.
Chemistry that deals with minute quantities of materials, frequently less than one milligram in mass or one milliliter in volume.



mi
 separations and detection; and the development of microscale methods to facilitate single molecule measurement and manipulation. The National Institute of Standards and Technology is an agency of the Technology Administration, U.S. Department of Commerce.
COPYRIGHT 2004 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2004, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Locascio, Laurie E.
Publication:Journal of Research of the National Institute of Standards and Technology
Date:May 1, 2004
Words:4744
Previous Article:Uncertainties in small-angle measurement systems used to calibrate angle artifacts.
Next Article:Uncertainty and traceability for the CEESI Iowa natural gas facility.
Topics:



Related Articles
Temperature sensors, monitors, controls: product lines reviewed. (plastics machinery) (Buyers Guide)
Temperature control for polymer processing.
Mold cooling: should you try a more direct approach?
Temperature sensors, monitors and controls.(1998/1999 Manufacturing Handbook & Buyers' Guide)(Buyers Guide)
Liquid temperature-control equipment.(1998/1999 Manufacturing Handbook & Buyers' Guide)(Buyers Guide)
Transparent pipes shape microstructures. (Light Splash).
Non-newtonian polymer solution for rheology now available. (Standard Reference Materials).(Brief Article)
The new AI: Europe embraces ambient intelligence.(On the Forefront)(Ambient Intelligence )
Cleaning fluid maintains thermal oil systems.(Materials)
The ups and downs of routing fluids on chips.(microfluidic chips)(Brief article)

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles