Detector development for the abBA experiment.


We have developed a new type of field-expansion spectrometer to measure the neutron beta decay correlations (a, b, B, and A). A precision measurement of these correlations places stringent requirements on charged particle detectors. The design employs large area segmented silicon detectors to detect both protons and electrons in coincidence. Other requirements include good energy resolution (<5 keV), a thin dead layer to allow observation of 30-keV protons, fast timing resolution (~1 ns) to reconstruct electron-backscattering events, and nearly unity efficiency. We report results of testing commercially available surface-barrier silicon detectors for energy resolution and timing performance, and measurement of the dead-layer thickness of ion-implanted silicon detectors with a 3.2 MeV alpha source.

Key words: dead layer; electronbackscattering event; neutron beta decay; silicon detector.

1. Introduction

Precision measurements of neutron beta decay correlations, a, b, B, and A, provide important tests of the electroweak e·lec·tro·weak  
adj.
Of or relating to the combination of the electromagnetic and weak nuclear forces in a unified theory.
 standard model: a test of unitarity of the first row of the CKM CKM Cabibbo-Kobayashi-Maskawa (quark mixing matrix)
CKM Certified Knowledge Manager (trademark of Hudson Associates Consulting, Inc.
 matrix, a test of the CVC See CSC.  hypothesis, and a sensitive search for right-handed currents or scalar or tensor tensor, in mathematics, quantity that depends linearly on several vector variables and that varies covariantly with respect to some variables and contravariantly with respect to others when the coordinate axes are rotated (see Cartesian coordinates).  bosons.

The abBA collaboration has developed a conceptual design for a new type of field-expansion magnetic spectrometer to measure neutron decay correlation coefficients, a, b, B, and A, with a polarized A one-way direction of a signal or the molecules within a material pointing in one direction.  pulsed cold neutron beam [1,2]. abBA aims to measure the correlations to an order of magnitude A change in quantity or volume as measured by the decimal point. For example, from tens to hundreds is one order of magnitude. Tens to thousands is two orders of magnitude; tens to millions is three orders of magnitude, etc.  better accuracy than previous measurements, where the accuracy was limited by systematic errors caused by 1) neutron polarimetry Polarimetry

The science of determining the polarization state of electromagnetic radiation (x-rays, light or radio waves). Radiation is said to be linearly polarized when the electric vector oscillates in only one plane.
, 2) poor detector properties such as resolution, efficiency, stability, and homogeneity, and 3) large backgrounds. To improve systematic uncertainties, abBA will utilize the pulsed nature of the neutron beam at SNS SNS sympathetic nervous system. , which allows an accurate beam polarization measurement by a time-of-flight technique for A and B [3]. Coincidences between electrons and protons from the neutron decay will be used to suppress backgrounds.

The decay rate for polarized neutrons as a function of electron energy ([E.sub.e]) and neutron life time ([[tau].sub.n]) is given by [4],

dW [proportional] [1/[[tau].sub.n]]F([E.sub.e])[1 + a[[[P.sub.e] * [P.sub.v]]/[[E.sub.e][E.sub.v]]] + b[[m.sub.e]/[E.sub.e]]+B[[[[sigma].sub.n] * [P.sub.v]]/[E.sub.v]]+A[[[[sigma].sub.n] * [P.sub.e]]/[E.sub.e]], (1)

where [P.sub.e] and [P.sub.v] are the outgoing electron and neutrino neutrino (ntrē`nō) [Ital.,=little neutral (particle)], elementary particle with no electric charge and a very small mass emitted during the decay of certain other particles.  momenta, [[sigma].sub.n] is the neutron spin, [E.sub.v] is the neutrino energy, and F([E.sub.e]) is electron energy spectrum. To measure the decay parameters both electrons (50 keV to 780 keV) and protons (< 30 keV) must be detected.

The charged decay particles, electrons and protons, are guided through a magnetic field into two large area and 2[pi] solid angle Si detectors mounted on both ends of the field-expansion spectrometer. The protons are accelerated by an electric field up to 30 keV before they reach the detectors. The dead layer of the Si detector has to be thin enough ([approximately equal to] 100 nm) to allow a low-energy proton detection with energy resolution better than 5 keV. Measurement of the electron energy depends only on the characteristics of the detectors. Electrons drift into the detectors following the magnetic field in their spiral path. However, about 20 % of them backscatter backscatter

in radiology, radiation deflected by scattering processes at angles greater than 90 degrees to the original direction of the beam of radiation. Important in radiotherapy when estimating surface exposure dose.
. Most of these electrons drift back to the original detector because of the magnetic field pinch, but a small fraction have enough kinetic energy kinetic energy: see energy.
kinetic energy

Form of energy that an object has by reason of its motion. The kind of motion may be translation (motion along a path from one place to another), rotation about an axis, vibration, or any combination of
 to cross the spectrometer and hit the opposite detector. The timing of electron hits must be resolved sufficiently to determine which detector was hit first in order to reconstruct events for a [2] and A. The required timing resolution is about 1 ns.

The size of the detector is defined by the magnetic field expansion for charged particles. The conceptual design of the Si detector has 15 cm diameter and 2 mm thickness and consists of 100 active pixels. Each pixel has a hexagonal shape with a 1 [cm.sup.2] area. To verify that the conceptual detector meets design criteria, we chose commercially available surface-barrier Si detectors to measure timing resolution and ion-implanted Si detectors to measure available dead layers.

2. Timing Measurement

In the abBA experiment, some backscattered electrons will deposit their energy in both detectors. In such cases, in order to reconstruct the events, good timing resolution is required to distinguish which detector the electrons hit first. Using a [.sup.90]Sr beta source with maximum energy of 2.28 MeV, timing resolution was measured for surface-barrier Si detectors using a fast plastic scintillator scin·til·la·tor
n.
A substance that glows when hit by high-energy particles or photons.
 as a time reference.

In the measurement, the electrons from the source travel through the 2.8 mm thick plastic scintillator producing fast signals that are used as a timing reference. They then enter a vacuum chamber through a thin Al window, and finally hit a 1 [cm.sup.2] area and 2 mm thick Si detector [5], which was cooled close to L[N.sub.2] temperature. The Si detector was biased with 160 V. Electrons interacting with silicon create electron-hole pairs. Due to the applied electric field, electrons move towards the anode anode (ăn`ōd), electrode through which current enters an electric device. In electrolysis, it is the positive electrode in the electrolytic cell.
anode

Terminal or electrode from which electrons leave a system.
 and holes move towards the cathode to generate electrical signals, which were amplified with the circuit shown in Fig. 1.

[FIGURE 1 OMITTED]

Si-detector events with a scintillator coincidence were recorded with an oscilloscope oscilloscope (əsĭl`əskōp'), electronic device used to produce visual displays corresponding to electrical signals. Displays of such nonelectrical phenomena as the variations of a sound's intensity can be made if the phenomena are . In order to determine the time of the signal with the respect to the scintillator, a linear fit was used to determine slopes in the preamp signal and in the discriminator dis·crim·i·na·tor  
n.
1. One that discriminates.

2. Electronics A device that converts a property of an input signal, such as frequency or phase, into an amplitude variation, depending on how the signal differs from a
 output of the scintillator. The time difference between the signals was obtained by comparing the times to reach half of maximum signal amplitude. The time difference between the two signals ranged from 9.02 ns to 11.06 ns with standard deviation In statistics, the average amount a number varies from the average number in a series of numbers.

(statistics) standard deviation - (SD) A measure of the range of values in a set of numbers.
 of 0.59 ns for 12 waveforms from events with energy deposition of 170 keV as shown by the histogram histogram
 or bar graph

Graph using vertical or horizontal bars whose lengths indicate quantities. Along with the pie chart, the histogram is the most common format for representing statistical data.
 in Fig. 2. The experimental FWHM FWHM Full Width at Half Maximum  is 2.35[sigma] = 1.4 ns, with uncertainty of 0.5 ns. The expected FWHM can be calculated using charge collection time, a FET FET: see transistor.


(Field Effect Transistor) One of two major categories of transistor; the other is bipolar. FETs use a gate element that, when charged, creates an electromagnetic field that changes the conductivity of a silicon
 noise, and a signal amplitude and capacitance, using the following formula,

[GAMMA] = 2.35[[[V.sub.n]C[delta]E]/E][square root of [[T.sub.rise]/2]],

[FIGURE 2 OMITTED]

where [V.sub.n] is the FET noise, [approximately equal to] [10.sup.-9] [V/sqrt(Hz)], C = 20 pF, [delta]E is the energy per electron-hole pair at room temperature, 3.62 eV/pair, and [T.sub.rise] is measured rise time, 55.4 ns. The expected value Expected value

The weighted average of a probability distribution. Also known as the mean value.
, [GAMMA] = 1.02 ns, agrees with measured time resolution. We expect that timing resolution would be improved by increasing the detector bias to increase drift velocities of both electron and hole in the silicon and by cooling the FET to reduce electronic noise [6].

With 59.5 keV gamma rays from a [.sup.241]Am source, we measured 8.1 keV energy resolution with an FET at room temperature and energy resolution was improved to 3.6 keV with the cooled FET.

3. Measurement of Detector Dead Layer Thickness

The dead layer on the surface of a detector serves to protect the Si surface. The detector dead layer is a consequence from the fabrication method, which starts with a homogeneous crystal of p-type materials and then converts a region of the crystal near the surface from a p-type to an n-type material. This creates a junction at some distance from the surface, typically 100 nm to 200 nm. A particle interacting with this layer will lose its energy without detection and therefore it is important to minimize the thickness of the dead layer. This is one of the main issues in detector development for the abBA experiment where the dead layer must be thin enough to detect 30 keV protons.

We have demonstrated the capability to measure thin dead layers with four ion-implanted sample detectors and each segmented into four 1 [cm.sup.2] active areas and having [approximately equal to] 300 [micro]m depletion depth [7]. Two of the detectors are not coated and two are coated with gold on the surface.

The detector dead layers were determined by measuring the energy loss of 3.2 MeV [alpha]-particles from a radioactive [.sup.148]Gd source and by using a calculated dE/dx [8]. The detectors were mounted on a rotating stage in a vacuum chamber and spectra were taken for several angles. The measured [alpha]-particle energy was fitted to E = [E.sub.0] - [DELTA]E/cos[theta Theta

A measure of the rate of decline in the value of an option due to the passage of time. Theta can also be referred to as the time decay on the value of an option. If everything is held constant, then the option will lose value as time moves closer to the maturity of the option.
], where [E.sub.0] is the incident particle energy, [DELTA]E is the energy loss in the detector dead layer, and [theta] the angle of the detector, from nominal incidence. The results of the dead layer measurements are shown in Fig. 3. We obtained 93 nm, 80.5 nm, 236 nm, and 230.1 nm for the non-coated detectors and coated detectors respectively. Detectors coated with gold have twice thicker dead layers than detectors without a gold coating.

[FIGURE 3 OMITTED]

4. Conclusion

Detector development for the abBA experiment has made significant progress. We measured the timing resolution with commercially available surface-barrier Si detectors. Measured timing resolution of 1.4 ns agrees with the expected timing resolution of 1.02 ns within 30 % uncertainty. The dead layers of commercially available ion-implanted Si detectors were measured and 100 nm dead layers are sufficiently thin to allow detection of 30 keV protons. Further development works such as dead layer measurement with 30 keV protons, improvement of timing resolution remain before a large area, segmented, and thin dead-layer detector can be produced.

5. References

[1] W. S. Wilburn et al., in this special issue.

[2] J. D. Bowman, in this special issue.

[3] S. I. Penttila et al., in this special issue.

[4] J. D. Jackson
This page is on the physicist. For the basketball player see J. D. Jackson (basketball).


John David Jackson (born 1925) is a Canadian-American physics professor emeritus at the University of California, Berkeley and a senior staff physicist at
, S. B. Treiman, and H. W. Wyld, Jr., Phys Rev. 106, 517 (1957).

[5] PerkinElmer Instruments, 100 Midland Rd., Oak Ridge, Tennessee Oak Ridge is an incorporated city in Anderson and Roane Counties in East Tennessee, about 25 miles northwest of Knoxville. Oak Ridge's population was 27,387 people at the 2000 census.  37830, USA.

[6] C. Jacoboni et al., Solid-State Electronics 20, 77 (1979).

[7] Micron Semiconductor Limited, Lancing, Sussex, England.

[8] J. Ziegler, Computer program SRIM-2003 (2003).

P.-N. Seo, J. D. Bowman, G. S. Mitchell, S. I. Penttila, and W. S. Wilburn

Los Alamos National Laboratory Los Alamos National Laboratory (LANL) (previously known at various times as Site Y, Los Alamos Laboratory, and Los Alamos Scientific Laboratory) is a United States Department of Energy (DOE) national laboratory, managed and operated by Los Alamos National , Los Alamos, NM 87545, USA

Accepted: August 11, 2004

Available online: http://www.nist.gov/jres
COPYRIGHT 2005 National Institute of Standards and Technology
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2005, Gale Group. All rights reserved. Gale Group is a Thomson Corporation Company.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Author:Wilburn, W.S.
Publication:Journal of Research of the National Institute of Standards and Technology
Geographic Code:1USA
Date:Jul 1, 2005
Words:1720
Previous Article:Detecting the radiative decay mode of the neutron.
Next Article:A backscatter suppressed electron detector for the measurement of "a".
Topics:



Related Articles
Radiation detectors fall short of standards.(SECURITY BEAT: Homeland Defense Briefs)
Precision Measurements with Slow Neutrons: April 5-7, 2004.(Department of Energy. Los Alamos National Laboratory's conference)(Illustration)
Precision measurements with slow neutrons: April 5-7, 2004.(Calendar)
Project of neutron beta-decay A-asymmetry measurement with relative accuracy of (1-2) X [10.sup.-3].
Measurement of neutron decay parameters--the abBA experiment.
Proposed measurement of the beta-neutrino correlation in neutron decay.
Detecting the radiative decay mode of the neutron.
Monte Carlo study of the abBA experiment: detector response and physics analysis.
Simulation of charged particle trajectories in the neutron decay correlation experiment abBA.
Development of a position sensitive neutron detector with high efficiency and energy resolution for use at high-flux beam sources.

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles