Circuit board tests, compilers figure in Vitalograph audit.


Failure to validate the process for making printed circuit boards (PCB PCB: see polychlorinated biphenyl.
PCB
 in full polychlorinated biphenyl

Any of a class of highly stable organic compounds prepared by the reaction of chlorine with biphenyl, a two-ring compound.
), plus the equipment to test the boards, were key problems noted in last May's inspection of Vitalograph, an Ennis, Ireland, maker of spirometers and peak-flow meters.

Center for Devices investigator Kevin Vogel noted in the EIR EIR n. popular acronym for environmental impact report, required by many states as part of the application to a county or city for approval of a land development or project. (See: environmental impact report)  that Vitalograph's PCB process was not validated, but the observation did not make the 483 in that manner.

Instead, Vogel wrote in the inspection report that the company tested the PCBs with testing equipment and conducted visual examinations as well. He stated that the firm insisted that "because of this 100% inspection, no formal validation process had been completed." Vogel nonetheless insisted that they needed to "control the process in some manner."

The closest observation on this issue in the 483 was that Vitalograph could not produce documentation for PCB test equipment. An Aug. 6 warning letter (Doc. 7926W) noted that the firm "failed to validate and document a manufacturing process," and went on to refer to the unvalidated test equipment, apparently equating validation of test devices with validation of the production process.

In a June 25 response to the 483, Ger Fennel fennel, common name for several perennial herbs, genus Foeniculum vulgare of the family Umbelliferae (parsley family), related to dill. The strawlike foliage and the seeds are licorice-scented and are used (especially in Italian cooking) for flavoring. , QA director at the Ennis, facility, seemed to be on the same page when he referred to the comment regarding documents on test equipment validation. Fennel promised that Vitalograph would make every effort to "acquire validation documentation for existing boards, all future boards introduced will have validation documentation."

However, the warning letter deemed Fennel's response inadequate due to lack of documentation.

The remaining validation issues covered software, opening with the observation that Vitalograph had not documented its coding standards and had conducted no white-box testing. The EIR went into no detail on these observations and the warning letter only commented that the firm had not "adequately validated software for its intended use." Code walk-throughs and stress testing Determining the durability of a system by pushing it to its limits. Stress testing a network is performed by transmitting excessive numbers of packets or attempting to break in illegally.  were similarly documented.

During the inspection, Vogel inquired as to the validation status of the firm's compiler. According to according to
prep.
1. As stated or indicated by; on the authority of: according to historians.

2. In keeping with: according to instructions.

3.
 the EIR, an unnamed Vitalograph employee insisted that "the compiler is not required to be validated." Vogel disagreed and Fennel's letter promised to validate compilers in the future.

An unspecified computer problem triggered an observation that never made it past the EIR. Vogel wrote that he inquired as to why the firm failed to validate software for "boundaries covering the maximum number of records which could be migrated from the firm's software database."

An unnamed Vitalograph employee responded that the boundary had been tested, but that it was not anticipated that any user would "include the large number of patients...as [a] complainant A plaintiff; a person who commences a civil lawsuit against another, known as the defendant, in order to remedy an alleged wrong. An individual who files a written accusation with the police charging a suspect with the commission of a crime and providing facts to support the allegation  did." Complaint handling was a major facet of this audit. Vogel recommended notifying users of the "maximum number of records the software can handle."

FDA FDA
abbr.
Food and Drug Administration


FDA,
n.pr See Food and Drug Administration.

FDA,
n.pr the abbreviation for the Food and Drug Administration.
 also deemed deficient Vitalograph's procedures for electrostatic discharge (hardware, testing) Electrostatic Discharge - (ESD) One kind of test that hardware usually has to pass to prove it is suitable for sale and use. The hardware must still work after is has been subjected to some level of electrostatic discharge.  reduction, employee training and soldering soldering

Process that uses metal alloys with low melting points to join metallic surfaces without melting them. Tin-lead solders, once widely used in the electrical and plumbing industries, are now replaced by lead-free alloys.
 for process control deviation, according to the 483.

Vitalograph, Ennis, Ireland, 5/31-6/3/99, Doc. 108739M, $8.50 plus retrieval.

RELATED ARTICLE: The Checklist - Vitalograph, foreign audit

* Software and computer validation

* Cleaning validation The introduction to this article provides insufficient context for those unfamiliar with the subject matter.
Please help [ improve the introduction] to meet Wikipedia's layout standards. You can discuss the issue on the talk page.
 (water)
COPYRIGHT 2000 Washington Information Source, Inc.
No portion of this article can be reproduced without the express written permission from the copyright holder.
Copyright 2000 Gale, Cengage Learning. All rights reserved.

 Reader Opinion

Title:

Comment:



 

Article Details
Printer friendly Cite/link Email Feedback
Publication:Validation Times
Geographic Code:1USA
Date:Jan 1, 2000
Words:502
Previous Article:Bioburden, process controls noted in Diametrics' audit.
Next Article:Anthrax vaccine audit gives Pentagon, Congress concern.
Topics:



Related Articles
APTIX DEBUTS SOC DESIGN MAPPING TOOL FOR SYSTEM EMULATION.(the Expedition interactive emulation software environment for system-on-chip...
510(k), design control gaps trigger warning for Sun.(Brief Article)
IRISH EYES UNHAPPY OVER WARNING LETTER.(Vitalograph inspection finds process control problems)
Software validation at Drager, CRS-USA, Vitalograph cited.(Brief Article)
HP monitor recall tied to GMP deficiencies.(Hewlett-Packard GmbH)(Brief Article)
AOI testing positions in comparison: an extensive study compares AOI following printing, placement and reflow, with surprising results.(Test and...
Ninth Circuit reverses findings on reasonable compensation.
Stray voltage on low standoff components: the Doctor diagnoses an acute weak organic acid problem.(Process Doctor)
An OEM guideline for selecting PCB suppliers: when choosing a PCB supplier, site inspection and qualification testing are important, but doing your...

Terms of use | Copyright © 2009 Farlex, Inc. | Feedback | For webmasters | Submit articles